K Number
K220691

Validate with FDA (Live)

Date Cleared
2022-05-05

(57 days)

Product Code
Regulation Number
870.1250
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The LiquID Guide Catheter Extension is intended to be used in conjunction with guide catheters to access discrete regions of the coronary and/or peripheral vasculature, and to facilitate placement of interventional devices.

Device Description

The LiquID device is a guide catheter extension that is substantially equivalent to currently marketed devices with the distinction of having a larger intraluminal diameter. The device is available in two models: Liquid 061 and LiquID 071 which are designed to fit within 6F and 7F conventional guide catheters, respectively. The device has a 150 cm working length and a 0.018" diameter proximal stainless-steel shaft connected to a 15cm single lumen distal shaft (exchange length).

The LiquID device distal shaft includes an internal PTFE liner, a MP35N coil internally clad with a tantalum core for support and radiopacity, and an exterior nylon-based encapsulation layer. The rounded and atraumatic distal tip includes PTFE and Pebax that is monolithically attached (no joints) to the distal shaft. The distal shaft exterior is wiped with silicone fluid for lubricity.

The transition between the single lumen distal shaft and proximal shaft includes a stainless collar that is welded to the proximal shaft, positioned over the PTFE liner, and encapsulated in polymer. This construction provides a smooth transition and structural integrity.

The LiquID device has proximal positioning marks (laser created oxide marks) that are approximately 95cm and 105cm from the distal tip and has a colored proximal handle that follows the coronary guide catheter size color code. The proximal shaft is not coated.

The LiguID device is packaged in a HDPE hoop within a Tyvek poly sealed pouch. The pouch is contained in an SBS shelf box. Twenty-five (25) shelf boxes are contained within a shipping container. The shipping containers are used for transporting the devices between the manufacturer, the ethylene oxide (EO) sterilization contractor, and the distribution locations.

AI/ML Overview

The document provided is a 510(k) premarket notification for a medical device (guide catheter extension). It does not describe a study involving an AI algorithm or software. Therefore, the requested information regarding acceptance criteria and studies proving the device meets those criteria, pertaining to AI-related aspects like sample size, ground truth, expert adjudication, or MRMC studies, cannot be extracted from this document.

The document focuses on the physical and functional performance of the medical device itself, establishing its substantial equivalence to predicate devices. It lists verification tests, biocompatibility tests, and sterilization validation:

1. Table of Acceptance Criteria and Reported Device Performance:

Test CategoryTest NameAcceptance CriteriaReported Performance
Physical/FunctionalShaft Marker Location and Visibility(Not explicitly stated, but implied to be within design specifications and visible)Met specified acceptance criteria
Inside Diameter(Not explicitly stated, but implied to be within design specifications)Met specified acceptance criteria
Outside Diameter(Not explicitly stated, but implied to be within design specifications)Met specified acceptance criteria
Working Length(Not explicitly stated, but device specs mention 150 cm)Met specified acceptance criteria
Exchange Length(Not explicitly stated, but device specs mention 15 cm)Met specified acceptance criteria
Distal Tip Length(Not explicitly stated, but implied to be within design specifications)Met specified acceptance criteria
Device Preparation(Not explicitly stated, but implied to be functional and as intended)Met specified acceptance criteria
Functionality with a Hemostasis Valve(Not explicitly stated, but implied to be compatible and functional)Met specified acceptance criteria
Trackability(Not explicitly stated, but implied to meet performance for guiding in vasculature)Met specified acceptance criteria
Contrast Injection(Not explicitly stated, but implied to function effectively for contrast delivery)Met specified acceptance criteria
Compatibility with Interventional Devices(Not explicitly stated, but implied to allow placement of interventional devices)Met specified acceptance criteria
Radiopacity(Not explicitly stated, but implied to be visible under fluoroscopy)Met specified acceptance criteria
Kink Resistance(Not explicitly stated, but implied to resist kinking during use)Met specified acceptance criteria
Distal and Proximal Flexibility(Not explicitly stated, but implied to meet flexibility requirements for navigation)Met specified acceptance criteria
Surface Inspection(Not explicitly stated, but implied to be free of defects)Met specified acceptance criteria
Tensile and Torsional Strength(Not explicitly stated, but implied to withstand forces during use)Met specified acceptance criteria
Push Rod to Attachment Bend(Not explicitly stated, but implied to demonstrate structural integrity)Met specified acceptance criteria
Burst Pressure(Not explicitly stated, but implied to withstand internal pressure)Met specified acceptance criteria
Corrosion Resistance(Not explicitly stated, but implied to resist corrosion in physiological environment)Met specified acceptance criteria
Coating Durability(Not explicitly stated, but implied to maintain integrity during use)Met specified acceptance criteria
Particulate Evaluation(Not explicitly stated, but implied to have acceptable particulate levels)Met specified acceptance criteria
BiocompatibilityCytotoxicityIn accordance with ISO 10993-1Passed
SensitizationIn accordance with ISO 10993-1Passed
IrritationIn accordance with ISO 10993-1Passed
Systemic ToxicityIn accordance with ISO 10993-1Passed
PyrogenicityIn accordance with ISO 10993-1Passed
ASTM HemolysisIn accordance with ISO 10993-1Passed
Complement ActivationIn accordance with ISO 10993-1Passed
ThrombogenicityIn accordance with ISO 10993-1Passed
SterilizationSterility Assurance Level (SAL)SAL of 10-6 (implied standard, based on common medical device requirements)Validated to meet a SAL of 10-6 in accordance with ISO 11135 (document states 10e-9, which is a stricter SAL, confirming compliance).

2. Sample size(s) used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
This information is not provided in the document. The listed tests are likely laboratory bench tests and material characterization, not studies involving human or animal data sets in the context of AI.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
Not applicable. This document does not describe studies involving human interpretation or ground truth establishment by experts for data sets.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
Not applicable for this type of device submission.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
Not applicable. This device is a physical medical instrument, not an AI software/algorithm.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
Not applicable. This document is for a physical medical device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
Not applicable. The "ground truth" for this device's performance would be engineering specifications, material properties, and established biological safety standards, not medical diagnostic ground truth.

8. The sample size for the training set:
Not applicable. There is no AI algorithm being trained.

9. How the ground truth for the training set was established:
Not applicable. There is no AI algorithm being trained.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Seigla Medical, Inc. Chad Kugler President & CEO 7688 5th Street SE Buffalo, Minnesota 55313

Re: K220691

Trade/Device Name: LiquID 061 Guide Catheter Extension, LiquID 071 Guide Catheter Extension Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: March 8, 2022 Received: March 9, 2022

Dear Chad Kugler:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

for Lydia Glaw, Ph.D. Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K220691

Device Name

Seigla Medical LiquID Guide Catheter Extension

Indications for Use (Describe)

The LiquID Guide Catheter Extension is intended to be used in conjunction with guide catheters to access discrete regions of the coronary and/or peripheral vasculature, and to facilitate placement of interventional devices.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Date Prepared: 08 March 2022 510(k) Number: K220691

Legal Manufacturer:

Name:Seigla Medical, Inc.
Address:7688 5th St SEBuffalo, MN 55313
Email:ckugler@seiglamedical.com
Contact Person:Chad Kugler
Phone Number:(763) 615-9058
Facsimile Number:(612) 453-4774

Device Information:

Classification:Class II Percutaneous Catheter / DQY / 870.1250
Trade Name:Seigla Medical LiquID M Guide Catheter Extension
Common Name:Catheter

Reference Devices:

The Seigla Medical Liquid Guide Catheter Extension is substantially equivalent in intended use, method of operation and technical aspects to the following predicate devices:

Primarv Predicate: K172090 - Vascular Solutions/Teleflex Guideliner V3 Catheters

Additional Reference Devices: K163314 - Boston Scientific Guidezilla II Guide Extension Catheter K183353 - Medtronic Telescope Guide Extension Catheter K160561 - OX Medical Boosting Catheter

Summary Description:

The LiquID device is a guide catheter extension that is substantially equivalent to currently marketed devices with the distinction of having a larger intraluminal diameter. The device is available in two models: Liquid 061 and LiquID 071 which are designed to fit within 6F and 7F conventional guide catheters, respectively. The device has a 150 cm working length and a 0.018" diameter proximal stainless-steel shaft connected to a 15cm single lumen distal shaft (exchange length).

The LiquID device distal shaft includes an internal PTFE liner, a MP35N coil internally clad with a tantalum core for support and radiopacity, and an exterior nylon-based encapsulation layer. The rounded and atraumatic distal tip includes PTFE and Pebax that is monolithically attached (no joints) to the distal shaft. The distal shaft exterior is wiped with silicone fluid for lubricity.

{4}------------------------------------------------

The transition between the single lumen distal shaft and proximal shaft includes a stainless collar that is welded to the proximal shaft, positioned over the PTFE liner, and encapsulated in polymer. This construction provides a smooth transition and structural integrity.

The LiquID device has proximal positioning marks (laser created oxide marks) that are approximately 95cm and 105cm from the distal tip and has a colored proximal handle that follows the coronary guide catheter size color code. The proximal shaft is not coated.

The LiguID device is packaged in a HDPE hoop within a Tyvek poly sealed pouch. The pouch is contained in an SBS shelf box. Twenty-five (25) shelf boxes are contained within a shipping container. The shipping containers are used for transporting the devices between the manufacturer, the ethylene oxide (EO) sterilization contractor, and the distribution locations.

Indication for Use:

The LiquID Guide Catheter Extension is intended to be used in conjunction with guide catheters to access discrete regions of the coronary and/or peripheral vasculature, and to facilitate placement of interventional devices.

Comparison to Predicate Devices:

The LiquID Guide Catheter Extension is substantially equivalent to the Vascular Solutions/Teleflex Guideliner V3 Catheters (K172090). In addition, the Boston Scientific Guidezilla II Guide Extension Catheter (K163314), Medtronic Telescope Guide Extension Catheter (K183353) and the OX Medical Boosting Catheter (K160561) are included as reference devices.

Guide catheter extensions are designed to facilitate the placement of interventional devices. They are placed within the interior lumen and advanced beyond the distal tip of conventional guide catheters. These devices allow physicians to achieve deeper vascular intubation providing enhanced placement and support for subsequent interventional device delivery.

The LiquID device and, the predicate device, and the reference devices include a PTFE lined single lumen distal shaft that is metal (coil or braid) supported with an exterior polymer encapsulation layer. The device distal shafts have an exterior coating for lubricity and include radiopaque features. The devices transition from the distal shaft to a push rod with designs that allow the introduction of interventional devices from the guide catheter and into the device distal shaft. The device transitions are also designed for strength. The devices are EO sterilized and are conventionally packaged (hoop, pouch, and box).

{5}------------------------------------------------

Summary:

Based upon the intended use, descriptive information, and performance evaluation provided in this premarket notification, the Seigla LiquID Guide Catheter Extension device has been shown to be substantially equivalent to a currently marketed predicate device and reference devices.

    1. The LiquID device has been verified through the following tests:
    • Shaft Marker Location and Visibility ●
    • Inside Diameter ●
    • Outside Diameter ●
    • Working Length ●
    • Exchange Length
    • Distal Tip Length
    • Device Preparation
    • Functionality with a Hemostasis Valve
    • Trackability
    • Contrast Injection
    • Compatibility with Interventional Devices
    • Radiopacity
    • Kink Resistance
    • Distal and Proximal Flexibility
    • Surface Inspection
    • Tensile and Torsional Strength
    • . Push Rod to Attachment Bend
    • Burst Pressure ●
    • Corrosion Resistance ●
    • Coating Durability ●
    • Particulate Evaluation .
    1. The LiguID device passed the following biocompatibility tests performed in accordance with ISO 10993-1:
    • Cytotoxicity
    • Sensitization ●
    • Irritation ●
    • Systemic Toxicity
    • Pyrogenicity ●
    • ASTM Hemolysis ●
    • Complement Activation .
    • Thrombogenicity ●
    1. The LiquID device has been validated to meet a SAL of 10-9 in accordance with ISO 11135.

The results of these verification tests met the specified acceptance criteria and did not raise new safety or performance issues. Therefore, the LiquID device is a guide catheter extension that is substantially equivalent to the predicate devices.

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).