K Number
K220683

Validate with FDA (Live)

Manufacturer
Date Cleared
2022-10-04

(210 days)

Product Code
Regulation Number
870.5150
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The INDIGO CAT RX Aspiration Catheters and INDIGO Separator 4 As part of the INDIGO Aspiration System, the INDIGO CAT RX Aspiration Catheters and INDIGO Separator 4 are indicated for the removal of fresh, soft emboli and thrombi from vessels in the coronary and peripheral vasculature. The INDIGO Aspiration Tubing As part of the INDIGO Aspiration System, the INDIGO Sterile Aspiration Tubing is indicated to connect the INDIGO CAT RX Aspiration Catheters to the Penumbra Aspiration Pump. Penumbra Aspiration Pump The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems.

Device Description

The INDIGO Aspiration System is comprised of several devices: - INDIGO CAT RX Aspiration Catheter . - INDIGO Separator 4 ● - . INDIGO Aspiration Tubing - . INDIGO Pump Canister/Tubing - . Penumbra Aspiration Pump The INDIGO Aspiration System is designed to remove thrombus from the vasculature using continuous aspiration. The INDIGO CAT RX Aspiration Catheter is a dual lumen rapid exchange catheter that targets aspiration from the Pump directly to the thrombus, removing it via the INDIGO Aspiration Tubing and depositing it in the Pump Canister. The INDIGO Separator 4 may be used, if needed, to clear the lumen of the INDIGO CAT RX Aspiration Catheter should it become blocked with thrombus. The INDIGO CAT RX Aspiration Catheter is introduced through a guide catheter or long introducer sheath and into the coronary or peripheral vasculature and guided over a guidewire to the site of the primary occlusion. The INDIGO CAT RX Aspiration Catheter may be provided with a rotating hemostasis valve and a peelable sheath. The INDIGO Separator 4 is provided with an introducer and torque device. The INDIGO CAT RX Aspiration Catheter and INDIGO Separator 4 are visible under fluoroscopy.

AI/ML Overview

The provided text is a 510(k) summary for the Penumbra INDIGO Aspiration System (CAT RX Aspiration Catheter and Separator 4). It is a submission to the FDA for market clearance, asserting substantial equivalence to a previously cleared predicate device.

Crucially, this document does not contain information about acceptance criteria or a study proving the device meets those criteria from an AI/ML perspective. The device in question is a medical aspiration system, not an AI/ML diagnostic or therapeutic device.

The "Performance Data - Non-Clinical" and "Performance Data - Clinical" sections explicitly state:

  • "The subject and predicate Indigo System (CAT RX Aspiration Catheter and Separator 4) devices are identical. Therefore, previous device performance data regarding substantial equivalence described below remain unchanged." (Section 1.10)
  • "No clinical study was conducted as previously performed bench and animal testing was determined sufficient for verification and validation purposes for substantial equivalence determination." (Section 1.16)

The submission leverages prior testing and clearance of an identical predicate device. Therefore, there are no new studies or acceptance criteria specifically for this 510(k) submission to describe. The original predicate device (K180412 and reference K163618) would have had its own performance data and acceptance criteria, but those are not detailed in this document.

Therefore, it is not possible to answer the requested questions based on the provided text. The text does not describe an AI/ML device, nor does it present new performance data or acceptance criteria for a study demonstrating adherence to such criteria. The entire premise of the 510(k) is that the device is identical to a predicate and thus no new performance testing is required for this submission.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration".

October 4, 2022

Penumbra, Inc. Sindokht (Sisi) Soltanzadeh Regulatory Specialist III One Penumbra Place Alameda, California 94502

Re: K220683

Trade/Device Name: INDIGO Aspiration System (CAT RX Aspiration Catheter and Separator 4) Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy Catheter Regulatory Class: Class II Product Code: QEW, QEX Dated: August 30, 2022 Received: August 31, 2022

Dear Sindokht (Sisi) Soltanzadeh:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act and the limitations described below. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at

https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

The OHT2: Office of Cardiovascular Devices has determined that there is a reasonable likelihood that this device will be used for an intended use not identified in the proposed labeling and that such use could cause harm. Therefore, in accordance with Section 513(i)(1)(E) of the Act, the following limitation must appear in the Precautions section of the device's labeling:

    1. The safety and effectiveness of this device for use in the treatment of ST-Elevation Myocardial Infarction (STEMI) have not been established. Complications from the use of this device in this

{1}------------------------------------------------

manner could lead to death, permanent impairment, and/or the need for emergency medical intervention.

Please note that the above labeling limitations are required by Section 513(i)(1)(E) of the Act. Therefore, a new 510(k) is required before these limitations are modified in any way or removed from the device's labeling.

The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to the market. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification if the limitation statement described above is added to your labeling.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

for Bram Zuckerman, M.D. Director OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K220683

Device Name

INDIGO Aspiration System (CAT RX Aspiration Catheter and Separator 4)

Indications for Use (Describe)

The INDIGO CAT RX Aspiration Catheters and INDIGO Separator 4

As part of the INDIGO Aspiration System, the INDIGO CAT RX Aspiration Catheters and INDIGO Separator 4 are indicated for the removal of fresh, soft emboli and thrombi from vessels in the coronary and peripheral vasculature.

The INDIGO Aspiration Tubing

As part of the INDIGO Aspiration System, the INDIGO Sterile Aspiration Tubing is indicated to connect the INDIGO CAT RX Aspiration Catheters to the Penumbra Aspiration Pump.

Penumbra Aspiration Pump

The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

1 510(k) Summary

(as required by 21 CFR 807.92)

Pursuant to Section 12, Part (a)(i)(3A) of the Safe Medical Devices Act of 1990, Penumbra, Inc. is providing the summary of Substantial Equivalence for INDIGO Aspiration System (CAT RX Aspiration Catheter and Separator 4).

1.1 Sponsor/Applicant Name and Address

Penumbra, Inc. One Penumbra Place Alameda, CA 94502 USA

1.2 Sponsor Contact Information

Sindokht (Sisi) Soltanzadeh Regulatory Specialist III Phone: (269) 330-9520 FAX: (510) 217-6414 Email: ssoltanzadeh@penumbrainc.com

1.3 Date of Preparation of 510(k) Summary

September 29, 2022

1.4 Device Trade or Proprietary Name

INDIGO Aspiration System (CAT RX Aspiration Catheter and Separator 4)

1.5 Device Classification

Regulatory Class: II Classification Panel: Cardiovascular Classification Name: Catheter, Embolectomy Regulation Number: 21 CFR §870.5150 Product Code: QEW, QEX (Catheter, Embolectomy)

{5}------------------------------------------------

Predicate/Reference Device 1.6

510(k) NumberName of Device
Predicate Device
K180412INDIGO Aspiration System (CAT RX Aspiration Catheter and Separator 4)- Penumbra Engine Pump and Canister
Reference Device
K163618INDIGO Aspiration System

Device Description 1.7

The INDIGO Aspiration System is comprised of several devices:

  • INDIGO CAT RX Aspiration Catheter .
  • INDIGO Separator 4 ●
  • . INDIGO Aspiration Tubing
  • . INDIGO Pump Canister/Tubing
  • . Penumbra Aspiration Pump

The INDIGO Aspiration System is designed to remove thrombus from the vasculature using continuous aspiration. The INDIGO CAT RX Aspiration Catheter is a dual lumen rapid exchange catheter that targets aspiration from the Pump directly to the thrombus, removing it via the INDIGO Aspiration Tubing and depositing it in the Pump Canister. The INDIGO Separator 4 may be used, if needed, to clear the lumen of the INDIGO CAT RX Aspiration Catheter should it become blocked with thrombus. The INDIGO CAT RX Aspiration Catheter is introduced through a guide catheter or long introducer sheath and into the coronary or peripheral vasculature and guided over a guidewire to the site of the primary occlusion. The INDIGO CAT RX Aspiration Catheter may be provided with a rotating hemostasis valve and a peelable sheath. The INDIGO Separator 4 is provided with an introducer and torque device. The INDIGO CAT RX Aspiration Catheter and INDIGO Separator 4 are visible under fluoroscopy.

1.8 Indications for Use

The INDIGO CAT RX Aspiration Catheters and INDIGO Separator 4

As part of the INDIGO Aspiration System, the INDIGO CAT RX Aspiration Catheters and INDIGO Separator 4 are indicated for the removal of fresh, soft emboli and thrombi from vessels in the coronary and peripheral vasculature.

The INDIGO Aspiration Tubing

As part of the INDIGO Aspiration System, the INDIGO Sterile Aspiration Tubing is indicated to connect the INDIGO CAT RX Aspiration Catheters to the Penumbra Aspiration Pump.

{6}------------------------------------------------

Penumbra Aspiration Pump

The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems.

1.9 Technological Comparison

The Indigo System subject devices are identical to the predicate device and differ only by the removal of a labeling condition per FDA clearance letter for K163618. The technological characteristics and intended use remain unchanged. There are no changes in the device design. materials, packaging, and sterilization methods. At a high level, the subject and predicate devices are based on the following same technological characteristics:

  • Device design and materials Identical for Indigo System Devices: CAT RX Aspiration . Catheters, Separator 4, and Aspiration Tubing
  • . Packaging materials - Commonly utilized for interventional devices
  • Sterilization Ethylene Oxide (EO) ●
  • Shelf life There are no changes to the shelf life of the subject devices due the labeling . change.
  • . Single Use, disposable
  • Aspiration Source Penumbra Aspiration Pump ●

1.10 Performance Data - Non-Clinical

The subject and predicate Indigo System (CAT RX Aspiration Catheter and Separator 4) devices are identical. Therefore, previous device performance data regarding substantial equivalence described below remain unchanged.

1.11 Biocompatibility Testing

There are no changes to the previously provided biocompatibility data of the Indigo System (CAT RX Aspiration Catheter and Separator 4) sterile device materials, which were reviewed and cleared under the premarket notification listed in Section 1.6.

1.12 Design Verification - Bench Top Testing

There are no changes to the previously provided bench-top data of the devices, which were reviewed and cleared under the premarket notification listed in Section 1.6.

1.13 Shelf-Life

There are no changes to the previously performed shelf-life testing for these devices, which were reviewed and cleared under premarket notification listed in Section 1.6.

{7}------------------------------------------------

1.14 Packaging

There are no changes to the previously provided packaging material listing or the packaging process for these devices, which were cleared under the premarket notification listed in Section 1.6.

1.15 Design Validation - Animal Study

There are no changes to the previously provided animal testing data of the devices, which were reviewed and cleared under the premarket notification listed in Section 1.6.

1.16 Performance Data – Clinical

No clinical study was conducted as previously performed bench and animal testing was determined sufficient for verification and validation purposes for substantial equivalence determination.

1.17 Conclusions

Based on leveraged non-clinical bench testing and a benefit-risk assessment, the subject device is substantially equivalent to the predicate device and does not raise any new questions of safety and effectiveness.

§ 870.5150 Embolectomy catheter.

(a)
Identification. An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.(b)
Classification. Class II (performance standards).