Intended Use

The APX-24 Portable X-ray is intended for use by a qualified/trained physician on both adult and pediatric subjects for taking diagnostic X-rays.

Device Description

The APX-24 Portable X-ray is a portable unit that operates from120 V / 220-240V, 50 / 60 Hz. The unit utilizes a newly designed high-frequency inverter and can bemounted to a tripod or support arm. The operator must observe the usual safety precautions regarding the use of x- rays.

AI/ML Overview

The provided FDA 510(k) summary for the APX-24 Portable X-ray System does not contain the specific details required to answer all aspects of your request regarding acceptance criteria and a study proving the device meets those criteria. The document primarily focuses on demonstrating substantial equivalence to a predicate device through technical specifications and adherence to general safety and performance standards.

Here's a breakdown of what can be extracted and what information is missing:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state "acceptance criteria" for the device's performance in terms of diagnostic accuracy or image quality compared to a clinical gold standard. Instead, it relies on demonstrating that the device meets general performance standards for X-ray systems and exhibits similar technical specifications to its predicate device.

The table below summarizes the technical characteristics, but it's not an acceptance criteria table related to diagnostic performance.

CharacteristicPredicate Device (MinXray HF100H)Subject Device (Aspenstate APX-24)
Intended UseDiagnostic X-rays for adult/pediatric subjectsDiagnostic X-rays for adult/pediatric subjects
Size/Weight406 x 222 x 241 mm, 18.6 kgs414 x 230 x 215.5 mm, 12.7 kg
Energy Source100-140V 50-60 Hz AC120VAC/220-240VAC, 50/60Hz
Mounting methodUsually mounted to MinXray XGS MKIII Portable StandNot mounted to a Portable Stand
User InterfaceUp-Down pushbuttons for kVp/exposure time w/ LED/mAsUp/Down pushbuttons for kVp/exposure time w/ LED
Exposure switchDual-stage, deadman typeDual-stage, deadman type
ControlsSoftware-based, 2 CPUsSoftware-based, 2 CPUs
ConstructionMonobloc HF generator, Medical full-bridge inverter systemMonobloc HF generator, Medical full-bridge inverter system
High Voltage Energy SourceHigh frequency (60kHz) inverterHigh frequency (102kHz) inverter
Line Voltage AdjustmentAutomatic, dynamicAutomatic, dynamic
Exposure times0.03-4.0 sec (various steps)0.03-4.0 sec (various steps)
Tube potential (kV)40 - 100kV (2kV step)40-100kV (1kV step)
kV steps31 (2kV-step)31 (2kV-step)
Tube current (mA)30mA (40-60 kV), 25mA (62-80 kV), 20mA (82-100 kV)Varied (30mA-10mA depending on kV)
mA stepsConstantConstant
X-ray tubeCanon D-124SCanon D-125SB
Anode heat Storage20,000 HU50,000 HU
Focal Spot Size1.2 mm1.2 mm
mAs0.6-120 mAs0.32-100 mAs
Total filtration3.2mm AL equivalentMin 2.8 mm AL
CollimatorAdvantech R72, continuously adjustable light beamContinuously adjustable light beam
Source to Skin Distance (SSD)300 mm300 mm
Performance Standard21CFR 1020.30IEC60601-1-3, IEC60601-1-6, IEC60601-2-28, IEC60601-2-54, 21CFR 1020.30, 21CFR 1020.31
Electrical safetyUL2601, IEC60601-1IEC60601-1

2. Sample size used for the test set and the data provenance
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
4. Adjudication method for the test set
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
8. The sample size for the training set
9. How the ground truth for the training set was established

Missing Information:

The provided document describes a portable X-ray system, not an AI-powered diagnostic device or software. Therefore, the concepts of "test set," "training set," "ground truth," "experts for ground truth," "adjudication methods," "MRMC studies," "AI improvement," and "standalone algorithm performance" are not applicable to this submission.

The 510(k) summary focuses on demonstrating that the APX-24 Portable X-ray System is substantially equivalent to a predicate device (MinXray HF100H+) based on its intended use, technological characteristics, and compliance with recognized safety and performance standards for X-ray equipment. It does not involve a clinical study to assess its diagnostic accuracy or the performance of an AI algorithm.

Nonclinical Testing:

The document lists several nonclinical performance standards that the device successfully met (Section VII). These standards cover aspects like:

  • IEC 60601-1-3:2008+A1:2013: General requirements for basic safety and essential performance — Collateral Standard: Radiation protection in diagnostic X-ray equipment.
  • IEC 60601-2-65:2012 / EN 60601-2-65:2013: Particular requirements for the basic safety and essential performance of dental intra-oral X-ray equipment (though this seems somewhat misplaced if the device is general-purpose X-ray).
  • EN 60601-1-2:2015: General requirements for basic safety and essential performance — Collateral Standard: Electromagnetic disturbances — Requirements and tests.
  • EN 61000-3-2:2014, EN 61000-3-3:2013: Electromagnetic compatibility (EMC) standards.
  • IEC 62133:2012 / EN 62133:2013: Secondary cells and batteries containing alkaline or other non-acid electrolytes – Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications. (Suggests internal battery component).
  • EN 60601-1:2006/A1:2013: General requirements for basic safety and essential performance.
  • IEC 60601-2-54:2009+A1:2014: Particular requirements for the basic safety and essential performance of X-ray equipment for radiography and radioscopy. (This is a more direct standard for such a device).
  • 21CFR 1020.30 & 1020.31: Federal regulations for ionizing radiation emitting products, specifically related to diagnostic X-ray systems and their components.

These nonclinical tests are the "study" demonstrating the device meets the "acceptance criteria" inherent in these international and national standards. The acceptance criteria for these tests are the pass/fail limits defined within each standard, and the report states that "Testing was performed successfully" and "Performance evaluation (test) reports and the device inspection report confirmed that the APX-24 Portable X-ray System is suitable for the device's intended use."

{0}------------------------------------------------

March 31, 2022

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Aspenstate, Inc. % Mr. Dave Kim Regulatory Affairs Mtech Group 7505 Fannin St, Suite 610 HOUSTON TX 77054

Re: K220678

Trade/Device Name: APX-24 Portable X-ray System Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile x-ray system Regulatory Class: Class II Product Code: IZL Dated: February 28, 2022 Received: March 8, 2022

Dear Mr. Kim:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Laurel Burk, Ph.D. Assistant Director Diagnostic X-ray Systems Team Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K220678

Device Name APX-24 Portable X-ray System

Indications for Use (Describe)

The APX-24 Portable X-ray is intended for use by a qualified/trained physician on both adult and pediatric subjects for taking diagnostic X-rays.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

This summary of 510(k) information is submitted following the requirements of 21 CFR Part 807.92.

Date 510k summary prepared:February 28, 2022
------------------------------------------------

I. SUBMITTER

Submitter's NameSubmitter's AddressAspenstate, Inc801 North Jupiter Rd, Suite 200Plano TX 75074
Submitter's TelephoneContact personAlbert Kim/Managing Directoralbert.kim@aspenstate.comTel: +1- 214-257-0113
Official CorrespondentAddressTelephoneDave Kim (davekim@mtech-inc.net)7505 Fannin St, Suite 610, Houston, TX 77054+713-467-2607

II. DEVICE

Trade/Device Name:APX-24 Portable X-ray System
Regulation Number:21 CFR 892.1720
Regulation Name:Mobile x-ray system
Regulatory Class:II
Product Codes:IZL
Common/Usual Name:Digital Mobile Diagnostic X-Ray System

III. PREDICATE DEVICE

Trade/Device Name:
510(k) Number:
Regulation Number:
Regulation Name:
Regulatory Class:
Product Codes:
Common/Usual Name:

MinXray HF100H+™ K052721 21 CFR 892.1720 Mobile x-ray system II IZL Digital Mobile Diagnostic X-Ray System

{4}------------------------------------------------

DEVICE DESCRIPTION: IV.

The APX-24 Portable X-ray is a portable unit that operates from120 V / 220-240V, 50 / 60 Hz. The unit utilizes a newly designed high-frequency inverter and can bemounted to a tripod or support arm. The operator must observe the usual safety precautions regarding the use of x- rays.

V. Indications for Use: 21 CFR 807 92 (a) (5)

The APX-24 Portable X-ray is intended for use by a qualified/trained physician or technician on both adult and pediatric subjects for taking diagnostic X-rays.

CharacteristicMinXray HF100H (K052721)Aspenstate APX-24
Intended Use:Intended for use by aqualified/trained physician ortechnician on both adult andpediatric subjects for takingdiagnostic x-rays.The APX-24 Portable X-ray isintended for use by aqualified/trained physician ortechnician on both adult andpediatric subjects for takingdiagnostic X-rays.
Size/weight406 x 222 x 241 mm18.6kgs414 x 230 x 215.5mm, 12.7kg
Energy Source100-140V 50-60 HzAC3 0kVA120VAC/220-240VAC, 50/60Hz
Mounting methodUnit is usually mounted to aMinXray XGS MKIII PortableStandThis unit is not mounted to aPortable Stand
User InterfaceUp-Down pushbuttons for kVpselections and exposure timeselections with LED indictorsand mAs indicatorsUp/Down pushbuttons for kVpselections and exposure timeselections with LED indicators
Exposure switchDual-stage, deadman typeDual-stage, deadman type
ControlsSoftware-based, 2 CPUs.Software-based, 2 CPUs.
ConstructionMonobloc HF a generator,Medical full-bridge invertersystemMonoblock HF a generator, Medicalfull-bridge inverter system
High VoltageEnergy SourceHigh frequency (60kHz)inverterHigh frequency (102kHz) inverter
Line VoltageadjustmentAutomatic, dynamicAutomatic, dynamic
Exposure times0.03-0.2 sec(in 0.01 sec. Steps)0.2-04 sec(in 0.02 sec. Steps)0.4-1.0 sec(in 0.05 sec. Steps)1.0-4.0 sec(m 0.1 sec. Steps)0.03-0.2 sec(in 0.01 sec. Steps)0.2-04 sec(in 0.02 sec. Steps)0.4-1.0 sec(in 0.05 sec. Steps)1.0-4.0 sec(m 0.1 sec. Steps)
Tube potential(kV)40 - 100kV 2kVstep40-100kV (1kV step)
kV steps31(2kV-step)31(2kV-step)
Tube current(mA)30mA(40-60 kV)25mA(62-80 kV)20mA(82-100 kV)30 mA (40-50 kV)30, 35, 40 mA (51-60 kV)25, 30 mA (61-70 kV)15, 20, 30 mA (71-80 kV)13, 25 mA (81-90 kV)10, 20 mA (91-100 kV)
mA stepsConstantConstant
X-ray tubeCanon D-124SCanon D-125SB
Anode heatStorage20,000HU50,000 HU
Focal SpotSize1.2 mm1.2 mm
mAs0.6-120mAs0.32-100mAs
Total filtration3.2mm AL equivalentMin 2.8 mm AL
CollimatorAdvantech R72Continuously adjustablelight beam type withcentral x-ray indicatorContinuously adjustable light beam type with central x-ray indicator
Source to SkinDistance (SSD)300 mm300 mm
PerformanceStandard21CFR 1020.30IEC60601-1-3, IEC60601-1-6,IEC60601-2-28, IEC60601-2-5421CFR 1020.30, 21CFR 1020.31
Electrical safetyUL2601, IEC60601-1IEC60601-1

Technological characteristics: 21 CFR 807 92 (a) (6) Comparison Table

{5}------------------------------------------------

ASPENSTATE, INC

801 North JupiterRd, Suite 200, Plano, TX 70574 USA

Tel.: +214-257-0113

VI. Discussion of differences

The subject device is similar to the predicate device regarding the indications for use and technological application. The subject and predicate devices are portable X-ray systems for taking diagnostic X-rays of human anatomy using a fixed tube current and voltage (kVp).

The differences are the device appearance, size, and user interface.

{6}------------------------------------------------

ASPENSTATE, INC 801 North JupiterRd, Suite 200, Plano, TX 70574 USA Tel.: +214-257-0113

VII. Nonclinical testing

Testing was performed successfully according to the following standards:

  • IEC 60601-1-3:2008+A1:2013

  • IEC 60601-2-65:2012

  • EN 60601-2-65:2013

  • EN 60601-1-2:2015

  • EN 61000-3-2:2014

  • EN 61000-3-3:2013

  • IEC 62133:2012

  • EN 62133:2013

  • EN 60601-1:2006/A1:2013

  • IEC 60601-2-54:2009 (First edition) + A1:2014 for use in conjunction with IEC 60601-1:2015 (third edition) + A1:2012

Furthermore, the following Specific Guidance Document was utilized in the device development to ensure the safety of this device for both the operators and patients: "The Content of Premarket Submissions for Software Contained in Medical Devices"

VIII. Conclusion:

By the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided above comparison table. the APX-24 Portable X-ray System has little difference with its size and user interface as the information in the table. However, the system is substantially equivalent to the predicate devices with its design, mechanical and electrical performance as described.

Performance evaluation (test) reports and the device inspection report confirmed that the APX-24 Portable X-ray System is suitable for the device's intended use.

§ 892.1720 Mobile x-ray system.

(a)
Identification. A mobile x-ray system is a transportable device system intended to be used to generate and control x-ray for diagnostic procedures. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.