K Number
K220673

Validate with FDA (Live)

Date Cleared
2023-03-16

(374 days)

Product Code
Regulation Number
N/A
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Prescription Use:
Sterile Silver Alginate Wound Dressing (Rx only) is indicated for the management of moderate to heavily exuding partial to full thickness wounds, including, postoperative wounds, leg ulcers, pressure ulcers, diabetic ulcers, graft and donor sites.

OTC:
Sterile Silver Alginate Wound Dressing (OTC) is first aid to help in minor cuts, minor lacerations, minor scrapes, minor scalds and minor burns.

Device Description

Sterile Silver Alginate Wound Dressing is a sterile, single-use dressing composed of calcium alginate and silver antibacterial agent. This highly absorbent dressing vertically absorbs wound exudate and release silver ions within the dressing in the presence of wound fluid to help reduce bacteria within the dressing. As wound exudate is absorbed, the alginate forms a gel, which assists in maintaining a moist environment and allows intact removal. Based on in-vitro testing, the silver in the dressing inhibits bacterial growth in the dressing for up to seven days. No clinical benefit has been demonstrated regarding the antibacterial effectiveness of the silver component.

AI/ML Overview

The provided document is a 510(k) Summary for a Sterile Silver Alginate Wound Dressing. It describes the device, its intended use, and a comparison to a predicate device to establish substantial equivalence.

Based on the document, here's an analysis of the acceptance criteria and supporting studies:

1. Table of Acceptance Criteria and Reported Device Performance:

The document refers to adherence to specific standards and conducted tests, implying that the acceptance criteria are defined by these standards and the device's ability to successfully meet them.

Acceptance Criteria (Implied by Standards/Tests)Reported Device Performance
Material CompositionSubject Device: Calcium alginate, Carboxymethylcellulose sodium, Silver sodium zirconium hydrogenphosphate. Predicate Device: Alginate and silver. (Not an explicit performance metric, but a characteristic compared.)
Antibacterial DurationSubject Device: 7 days. Predicate Device: 7 days.
Single UseSubject Device: Yes. Predicate Device: Yes.
Biocompatibility (per ISO 10993-1 for breached/compromised surfaces with prolonged contact (>24h to 30d))Tests Conducted (Subject Device): Cytotoxicity, Sensitization, Irritation, Acute systemic toxicity, Subacute systemic toxicity, Implantation, Material-mediated pyrogenicity. Reported Performance: "demonstrate the device is as safe as its predicate device." (Implies meeting the criteria for each of these tests, but specific results are not detailed in this summary).
Sterilization (Gamma Irradiation, SAL: 10-6)Subject Device: Gamma Irradiation, SAL: 10-6. Predicate Device: Radiation, SAL: 10-6.
Free Swell Absorption CapacitySubject Device: 15.7 g/g. Predicate Device: 14.2 g/g. (Subject device's absorption is higher, exceeding predicate).
Liquid Absorbency (per EN 13726-1-2002)Reported Performance: "The test results demonstrated that the Sterile Silver Alginate Wound Dressing complies with the following standards." (Implies successful adherence to the standard's requirements, specific metrics not detailed).
Loss on Drying (per USP <731>)Reported Performance: "The test results demonstrated that the Sterile Silver Alginate Wound Dressing complies with the following standards." (Implies successful adherence to the standard's requirements, specific metrics not detailed).
pH Value (per USP <791> pH)Reported Performance: "The test results demonstrated that the Sterile Silver Alginate Wound Dressing complies with the following standards." (Implies successful adherence to the standard's requirements, specific metrics not detailed).
Antimicrobial Efficacy (per AATCC 100-2019)Reported Performance: "The test results demonstrated that the Sterile Silver Alginate Wound Dressing complies with the following standards." (Implies successful adherence to the standard's requirements, specific metrics not detailed).
Minimum Bacteriostatic Concentration Test (per AATCC 100-2019)Reported Performance: "The test results demonstrated that the Sterile Silver Alginate Wound Dressing complies with the following standards." (Implies successful adherence to the standard's requirements, specific metrics not detailed).

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample Size: Not explicitly stated for any of the performance bench tests. The document only mentions that tests were "conducted to verify that the proposed device met all design specifications." For biocompatibility, it states "the subject device was evaluated for" the listed tests, but does not provide numbers of samples or animals used.
  • Data Provenance: Not specified. The submission is from Qingdao Bright Moon Biomedical Materials Co., Ltd. in China, suggesting the testing likely occurred in China, but this is not explicitly stated. The document indicates these are "non-clinical performance tests," implying they are laboratory/bench tests, not human data. It is inherently retrospective as the tests were completed prior to submission.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Not Applicable. This document describes the evaluation of a medical device (wound dressing) through non-clinical performance and biocompatibility testing. It does not involve diagnostic image analysis or clinical assessments that would require expert human interpretation to establish ground truth. The "ground truth" for these tests is defined by the physical or biological outcomes as measured by the testing protocols per the cited standards.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • None. As with point 3, this type of adjudication method is used for clinical studies or studies involving human readers/interpreters, which is not the case here. The results of the non-clinical tests are direct measurements or observations based on established protocols.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No. This is not an AI-enabled diagnostic device and no MRMC study was conducted or is applicable. The document explicitly states: "No clinical study is included in this submission."

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

  • Not applicable. This device is a physical wound dressing, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • The "ground truth" for the non-clinical performance tests is defined by the metrics and acceptance criteria outlined in the cited international and national standards (e.g., EN 13726-1-2002 for liquid absorbency, USP <731> for Loss on Drying, USP <791> for pH Value, AATCC 100-2019 for Antimicrobial Efficacy and MBC, and ISO 10993 for biocompatibility). The device's performance is compared against these scientifically established benchmarks.

8. The sample size for the training set

  • Not applicable. This device is a physical wound dressing, not an AI/machine learning model, so there is no training set.

9. How the ground truth for the training set was established

  • Not applicable. As there is no training set for an AI model, there is no corresponding ground truth establishment process for a training set.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.

March 16, 2023

Oingdao Bright Moon Biomedical Materials Co., Ltd. Deng Yunlong R&D Department Manager No. 788, Bright Moon Road, Huangdao District Qingdao, Shandong 266400 China

Re: K220673

Trade/Device Name: Sterile Silver Alginate Wound Dressing Regulatory Class: Unclassified Product Code: FRO Dated: February 13, 2023 Received: February 14, 2023

Dear Deng Yunlong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Julie A. Morabito -S

Julie Morabito, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K220673

Device Name Sterile Silver Alginate Wound Dressing

Indications for Use (Describe)

Prescription Use:

Sterile Silver Alginate Wound Dressing (Rx only) is indicated for the management of moderate to heavily exuding partial to full thickness wounds, including, postoperative wounds, leg ulcers, pressure ulcers, diabetic ulcers, graft and donor sites.

OTC:

Sterile Silver Alginate Wound Dressing (OTC) is first aid to help in minor cuts, minor lacerations, minor scrapes, minor scalds and minor burns.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.

{3}------------------------------------------------

510(K) Summary - K220673

This 510(K) Summary information is being submitted in accordance with 21 CFR 807.92.

I. SUBMITTER:

Qingdao Bright Moon Biomedical Materials Co., Ltd. No. 788, Bright Moon Road, Huangdao District, Qingdao, Shandong,China Contact Person: Deng yunlong Title: R&D Department Manager Tel: +86 18669829719 Email: yunlong.deng@bmsg.com

Submission Correspondent: Deng yunlong Email: yunlong.deng@bmsg.com Tel:+86 18669829719

Summary prepared: March 14th, 2023

II. SUBJECT DEVICE

Name of Device: Sterile Silver Alginate Wound Dressing Classification Name: Dressing, Wound, Drug FDA Panel: General and Plastic Surgery Regulatory Class: Unclassified Product Code: FRO III. PREDICATE DEVICE

Predicate Device:Silver Alginate Dressing (Prescription use), Antibacterial
Alginate Wound Dressing (OTC use) (K202982)
Classification Name:Dressing, Wound, Drug
FDA Panel:General and Plastic Surgery
Regulatory Class:Unclassified
Product Code:FRO

{4}------------------------------------------------

IV. DEVICE DESCRIPTION

Sterile Silver Alginate Wound Dressing is a sterile, single-use dressing composed of calcium alginate and silver antibacterial agent. This highly absorbent dressing vertically absorbs wound exudate and release silver ions within the dressing in the presence of wound fluid to help reduce bacteria within the dressing. As wound exudate is absorbed, the alginate forms a gel, which assists in maintaining a moist environment and allows intact removal. Based on in-vitro testing, the silver in the dressing inhibits bacterial growth in the dressing for up to seven days. No clinical benefit has been demonstrated regarding the antibacterial effectiveness of the silver component.

V. AVAILABLE MODELS

The proposed device is available in different sizes, as shown in the following table: Table 1 Device model of Sterile Silver Alginate Wound Dressing

Model NameShapeRefSpecification (cm)Tolerance
Sterile Silver AlginateWound DressingSheetYD05$5\times5$±5%
YD10$7.5\times12$±5%
YD15$10\times10$±5%
YD20$10\times15$±5%
YD25$10\times20$±5%
YD30$12.5\times12.5$±5%
YD40$15\times15$±5%
YD50$15\times20$±5%
YD60$15\times25$±5%
YD65$20\times20$±5%
YF01$2\times5$±5%
YF02$2\times10$±5%
YF03$2\times30$±5%
YF04$2\times44$±5%
YF05$3\times30$±5%
YF06$4\times30$±5%

{5}------------------------------------------------

VI. INTENDED USE per 21 CFR 807.92(A)(5)

Prescription Use:

Sterile Silver Alginate Wound Dressing (Rx only) is indicated for the management of moderate to heavily exuding partial to full thickness wounds, including, postoperative wounds, trauma wounds, leg ulcers, pressure ulcers, diabetic ulcers, graft and donor sites.

OTC:

Sterile Silver Alginate Wound Dressing (OTC) is first aid to help in minor abrasions, minor cuts, minor lacerations, minor scrapes, minor scalds and minor burns.

COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH VII. THE PREDICATE DEVICE per 21 CFR 807.92(a)(6)

The Sterile Silver Alginate Wound Dressing is compared with the predicate device Silver Alginate Dressing (Prescription use), Antibacterial Alginate Wound Dressing (OTC use) (K202982), manufactured by Winner Medical Co., Ltd.

The results are shown below in the Technological Characteristics Comparison Table:

Table 2 Technological Characteristics Comparison Table Between Subject Device and Predicate Device

ItemSubject Device (K220673)Predicate Device (K202982)Comparison
ClassificationProduct CodeFROFROSame
ClassificationRegulationUnclassifiedUnclassifiedSame
Indications for UsePrescription Use:Sterile Silver Alginate WoundDressing (Rx only) is indicatedfor the management of moderateto heavily exuding partial to fullthickness wounds, including,postoperative wounds, traumawounds, leg ulcers, pressureulcers, diabetic ulcers, graft anddonor sites.OTC:Sterile Silver Alginate WoundDressing (OTC) is first aid to helpin minor abrasions, minor cuts.Prescription:Silver Alginate Dressing (Rxonly) is indicated for themanagement of moderate toheavily exuding partial to fullthickness wounds, including,postoperative wounds, traumawounds, leg ulcers, pressureulcers, diabetic ulcers, graft anddonor sites.OTC:Antibacterial Alginate WoundDressing (OTC) is first aid to helpin minor abrasions, minor cuts.Same

{6}------------------------------------------------

minor lacerations, minor scrapes,minor scalds and minor burns.minor lacerations, minor scrapes,minor scalds and minor burns.
MechanismAlginate for absorbing liquid,silver present in the alginate forreducing bacteria in the dressing.Alginate for absorbing liquid,silver present in the alginate forreducing bacteria in the dressing.Same
MaterialCalcium alginate,Carboxymethylcellulose sodium,Silver sodium zirconiumhydrogenphosphateAlginate and silverDifferent
AntibacterialDuration7 days7 daysSame
Single UseYesYesSame
BiocompatibilityBiocompatibility in accordanceto 10993-1(breached orcompromised surfaces withprolonged contact(>24h to 30d))Biocompatibility in accordanceto 10993-1(breached orcompromised surfaces withprolonged contact(>24h to 30d))Same
SterilizationGamma IrradiationSAL:10-6RadiationSAL:10-6Same
Free SwellAbsorptionCapacity15.7g/g14.2g/gDifferent

VIII. SUBSTANTIAL EQUIVALENCE DISCUSSION per 21 CFR 807.92(b)

Sterile Silver Alginate Wound Dressing has been conducted biocompatibility studies in accordance with ISO 10993 to demonstrate the device is as safe as its predicate device. The performance bench testing was conducted to demonstrate that the subject device is as effective as its predicate device.

IX. PERFORMANCE DATA

Non-Clinical Performance Test Conclusion

Biocompatibility

Based on Table A.1 of ISO 10993-1 and Table A.1 of FDA Guidance "Use of International Standard ISO 10993-1, Biological evaluation of medical devices – Part 1 Evaluation and testing within a risk management process", the subject device is categorized as a surface device in contact with breached or compromised surface with prolonged duration. The subject device was evaluated for:

{7}------------------------------------------------

Cytotoxicity Sensitization Irritation Acute systemic toxicity Subacute systemic toxicity Implantation Material-mediated pyrogenicity

Performance Bench Testing

Performance testing were conducted to verify that the proposed device met all design specifications was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the Sterile Silver Alginate Wound Dressing complies with the following standards:

Liquid absorbency EN 13726-1-2002 Loss on Drying: USP <731> pH Value: USP <791> pH Antimicrobial Efficacy: AATCC 100-2019 Minimum Bacteriostatic Concentration Test: AATCC 100-2019

Clinical Test Conclusion

No clinical study is included in this submission.

X. CONCLUSION per 21 CFR 807.92(c)

The conclusion drawn from the nonclinical tests demonstrates that the subject device, the Sterile Silver Alginate Wound Dressing is as safe, as effective, and performs as well as or better than the legally marketed predicate device Silver Alginate Dressing (Prescription use), Antibacterial Alginate Wound Dressing (OTC use) (K202982).

N/A