K Number
K220624

Validate with FDA (Live)

Device Name
AI4CMR v1.0
Date Cleared
2022-07-22

(141 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

AI4CMR software is designed to report cardiac function measurements (ventricle volumes, ejection fraction, indices etc.) from 1.5T and 3T magnetic resonance (MR) scanners. AI4CMR uses artificial intelligence to automatically segment and quantify the different cardiac measurements. Its results are not intended to be used on a stand-alone basis for clinical decision-making.

The user incorporating AI4CMR into their DICOM application of choice is responsible for implementing a user interface.

Device Description

AI4CMR v1.0 is a cloud-hosted service used with any third-party DICOM viewer application where the DICOM viewer serves as the user interface and the interface to a PACS or scanner for AI4CMR. AI4CMR is implemented as a plug-in to the DICOM viewer by the user and automatically processes and analyses cardiac MR images received by the DICOM viewer to quantify relevant cardiac function metrics and makes the information available to the user at the user's discretion.

AI/ML Overview

The provided text describes the AI4CMR v1.0 device and its performance evaluation for FDA 510(k) clearance. Here's a breakdown of the acceptance criteria and study details:

Acceptance Criteria and Reported Device Performance

The acceptance criteria are implicitly defined by the reported performance metrics, primarily Intraclass Correlation Coefficient (ICC) and bias, tested against a "consensus" ground truth. The targets for these metrics demonstrating adequate agreement are not explicitly stated as numerical thresholds for acceptance, but are demonstrated by the presented results.

MetricAcceptance Criteria (Implicit from Study Design)Reported Device Performance (Bias ± SD)Reported Device Performance (ICC)
LV end-diastolic volume (EDV)High agreement with expert consensus8.663 ± 17.28 mL0.990
LV end-systolic volume (ESV)High agreement with expert consensus-2.893 ± 17.52 mL0.991
LV ejection fraction (EF)High agreement with expert consensus3.867 ± 5.74 %0.956
LV myocardial massHigh agreement with expert consensus2.452 ± 18.04 g0.955
RV end-diastolic volume (EDV)High agreement with expert consensus8.355 ± 15.59 mL0.964
RV end-systolic volume (ESV)High agreement with expert consensus-9.083 ± 13.83 mL0.953
RV ejection fraction (EF)High agreement with expert consensus7.802 ± 7.32 %0.814
Myocardium Segmentation (DSC)High segmentation overlap0.72 per image (average Dice Similarity Coefficient)N/A

Note: The Accuracy table on page 5 also lists "LV bias (std): 3.87 (5.74) LV ICC: 0.96" and "RV bias (std): 7.80 (7.32) RV ICC: 0.81" for EF, which align with the reported MRMC study results for EF.

Study Information

1. Sample sizes used for the test set and the data provenance:
* Clinical Performance Assessment (MRMC Study Test Set): 146 CMR cases.
* Provenance: Retrospective. Patient data was acquired from Siemens, GE, and Philips 1.5T scanners. The text does not explicitly state the country of origin for this specific 146-case dataset, but the training data was from Hospital de Braga, Portugal.
* Bench Testing (Standalone Performance Test Set): 15 CMR cases from the Society of Cardiac Magnetic Resonance (SCMR) Consensus Contour Data.
* Provenance: This is a publicly available consensus dataset. The origin of the cases themselves within the SCMR dataset is not specified in the document, but it includes data from various 1.5T and 3T scanners (Siemens, GE, Phillips).

2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
* Clinical Performance Assessment (MRMC Study): 2 expert readers.
* Qualifications: "2 expert readers who achieved excellent interrater variability (ICC > 0.75)". Specific qualifications (e.g., radiologist, years of experience) are not stated beyond them being "expert readers."
* Bench Testing (SCMR Consensus Data): 7 independent expert readers.
* Qualifications: From "various core laboratories." Specific qualifications are not detailed beyond "expert readers."

3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
* Clinical Performance Assessment (MRMC Study): The ground truth was established by 2 expert readers. They manually segmented the myocardium and determined volumes, LV mass, and Ejection Fraction. The "consensus" against which the device was compared appears to be derived from these two readers, as the text states, "The primary objective was to evaluate agreement between the AI4MED device and 2 expert readers who achieved excellent interrater variability." No explicit 2+1 or 3+1 adjudication process is described; it seems to be a consensus of the two experts.
* Bench Testing (SCMR Consensus Data): The SCMR Consensus Contour Data is inherently a consensus dataset established by 7 independent expert readers. The specific adjudication method (e.g., averaging, voting) used to create this consensus is not detailed here, but it implies a robust consensus approach.

4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
* A Multi-Reader, Multi-Center (MRMC) retrospective study was performed.
* However, this was an agreement study, comparing the AI device's performance to that of human experts. It was not a comparative effectiveness study designed to assess how human readers improve with AI assistance vs. without it (i.e., a human-in-the-loop study). Therefore, no effect size of human reader improvement with AI assistance is provided or applicable from this specific study design.

5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
* Yes, a standalone performance test was done ("Bench Testing"). The AI4CMR performed segmentation and quantification (EDV, ESV, LVM, EF) on its own and its results were compared against the SCMR Consensus Contour Data.

6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
* Expert Consensus.
* For the MRMC study: Consensus of 2 expert readers.
* For the Bench Testing: Consensus of 7 expert readers (SCMR Consensus Contour Data).

7. The sample size for the training set:
* 824 anonymized cases were initially collected. This dataset was split:
* Training Set: 577 cases (70% of 824).
* Validation Set: 121 cases (15%).
* Test Set (internal, for model development): 126 cases (15%).
* Note: The clinical validation test set (146 cases) used for the MRMC study was independent from this training/validation/internal test split.

8. How the ground truth for the training set was established:
* The document states that the training data was "collected retrospectively from Hospital de Braga, Portugal." It implies that cardiac function measurements (ventricle volumes, ejection fraction, indices etc.) would have been part of the standard clinical reporting for these cases. However, the specific method for establishing the ground truth (e.g., manual segmentation by clinicians, expert review, consensus) for the training data is not explicitly detailed in the provided text. It is assumed to be derived from the clinical records or expert annotations used during the model's development phase.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is the Department of Health & Human Services logo, which features an abstract image of a human figure. To the right of this is the FDA logo, which has the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

AI4MedImaging Medical Solutions S.A. % Carla Almeida Regulatory Affairs and Quality Manager Rua do Parque Poente, Lote 35 Braga, Minho 4705-002 PORTUGAL

July 22, 2022

Re: K220624

Trade/Device Name: AI4CMR v1.0 Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: QIH Dated: June 17, 2022 Received: June 23, 2022

Dear Carla Almeida:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR

{1}------------------------------------------------

  1. for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jessica Lamb. Ph.D. Assistant Director Imaging Software Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K220624

Device Name AI4CMR v1.0

Indications for Use (Describe)

AI4CMR software is designed to report cardiac function measurements (ventricle volumes, ejection fraction, indices etc.) from 1.5T and 3T magnetic resonance (MR) scanners. AI4CMR uses artificial intelligence to automatically segment and quantify the different cardiac measurements. Its results are not intended to be used on a stand-alone basis for clinical decision-making.

The user incorporating AI4CMR into their DICOM application of choice is responsible for implementing a user interface.

Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for AI4MedImaging. The logo consists of a circular graphic on the left and the text "AI4MedImaging" on the right. The circular graphic is made up of a series of blue lines that radiate out from the center, creating a spiral effect. The text is also blue and is written in a sans-serif font.

Section 5. 510(k) Summary

1. General Information

510(k) SponsorAI4MedImaging Medical Solutions S.A.
AddressRua do Parque Poente, It 324705-002 Sequeira, Braga Portugal
Correspondence PersonRory A. CarrilloQuality and Regulatory ConsultantCosm
Contact InformationEmail: rory@cosmhq.comPhone: 562-533-7010
Date PreparedJune 17, 2022

2. Subject Device

Proprietary NameAI4CMR v1.0
Common NameAI4CMR
Classification NameSystem, Image Processing, Radiological
Regulation Number21 CFR 892.2050
Regulation NameMedical Image Management and Processing System
Product CodeOIH
Regulatory ClassII

3. Predicate Device

Proprietary NameImbio RV/LV Software
Premarket NotificationK203256
Classification NameSystem, Image Processing, Radiological
Regulation Number21 CFR 892.2050
Regulation NameMedical Image Management and Processing System
Product CodeQIH
Regulatory ClassII

Device Description 4.

AI4CMR v1.0 is a cloud-hosted service used with any third-party DICOM viewer application where the DICOM viewer serves as the user interface and the interface to a PACS or scanner for AI4CMR. AI4CMR is implemented as a plug-in to the DICOM viewer by the user and automatically processes and analyses cardiac MR images received by the DICOM viewer to quantify relevant cardiac function metrics and makes the information available to the user at the user's discretion.

The following are the cardiac function metrics quantified and reported by the software:

AI4CMR v1.0 Traditional 510(k)

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for AI4MedImaging. The logo consists of a circular graphic on the left and the text "AI4MedImaging" on the right. The circular graphic is made up of many small blue lines that form a spiral shape. The text "AI4MedImaging" is also in blue and is written in a sans-serif font.

Quantitative Analysis

The subject device performs the following anatomical measurements:

  • Anatomy and tissue segmentation ●
  • LV/RV stroke volume 0
  • LV/RV cardiac output
  • LV/RV ejection fraction ●
  • LV/RV end-diastolic volume
  • LV/RV end-systolic volume

Reporting

The subject device enables the following metrics to be reported as desired by the user:

MetricUnitAccuracy
LV/RV stroke volumemln/a1
LV/RV cardiac outputL/minn/a1
LV/RV ejection fraction (EF)%LV bias (std): 3.87 (5.74)LV ICC: 0.96RV bias (std): 7.80 (7.32)RV ICC: 0.81
LV/RV end-diastolic volume (EDV)mlLV bias (std): 8.66 (17.28)LV ICC: 0.99RV bias (std): 8.36 (15.59)RV ICC: 0.96
LV/RV end-systolic volume (ESV)mlLV bias (std): -2.90 (17.52)LV ICC: 0.99RV bias (std): -9.08 (13.83)RV ICC: 0.95
LV myocardial massgbias (std): 2.45 (18.04)ICC: 0.96
LV/RV end-systolic volume index2ml/m^2n/a1
LV/RV end-diastolic volume index2ml/m^2n/a1
LV/RV stroke volume index2ml/m^2n/a1
Myocardium mass index2g/m^2n/a1

AI4CMR v1.0 Traditional 510(k)

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image contains the logo for AI4MedImaging. The logo consists of a circular design on the left, resembling a stylized spiral or a series of concentric circles with varying shades of blue. To the right of the circular design, the text "AI4MedImaging" is written in a sans-serif font, with each letter in a light blue color. The text is aligned horizontally and appears to be the primary identifier for the organization or product.

MetricUnitAccuracy
Cardiac index $2$L/(min m^2)n/a $1$

Notes:

1 These values are derived by performing simple mathematical operations and are derived from EDV, ESV, EF, and Mass metrics.

2 These values are only provided if the patient's height and weight are included in the DICOM data.

Training Dataset

The AI4CMR training was performed on a dataset of 824 anonvmized cases collected retrospectively from Hospital de Braga, Portugal. Acquisition occurred between 2015 to January 2019 and consisted of male (63%) and female (37%) patients ranging in age from 13 to 89 (mean of 58) years old from Siemens acquisition system. This dataset is independent from the clinical validation set. This dataset was split into the 3 sets (training, validation, test). The splitting ratio is 70% for the "training set", 15% for the "validation set" and 15% for the "test set", resulting in 577, 121 and 126 cases each, respectively.

5. Indications for Use

AI4CMR software is designed to report cardiac function measurements (ventricle volumes, ejection fraction, indices etc.) from 1.5T and 3T magnetic resonance (MR) scanners. AI4CMR uses artificial intelligence to automatically segment and quantify the different cardiac measurements. Its results are not intended to be used on a stand-alone basis for clinical decision-making.

The user incorporating AI4CMR into their DICOM application of choice is responsible for implementing a user interface.

Substantial Equivalence & Technical Characteristics 6.

Subject DeviceAI4CMR v1.0Predicate Device:Imbio RV/LV (K203256)
Intended UseAI4CMR software is designed toreport cardiac functionmeasurements (ventricle volumes,ejection fraction, indices etc.) from1.5T and 3T magnetic resonance(MR) scanners. AI4CMR usesartificial intelligence toautomatically segment andquantify the different cardiacmeasurements. Its results are notintended to be used on aThe Imbio RV/LV Software deviceis designed to measure themaximal diameters of the right andleft ventricles of the heart from avolumetric CTPA acquisition andreport the ratio of thosemeasurements. RV/LV analyzescases using an artificialintelligence algorithm to identifythe location and measurements ofthe ventricles. The RV/LV software

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for AI4MedImaging. The logo consists of a circular graphic on the left and the text "AI4MedImaging" on the right. The circular graphic is made up of several blue lines that radiate out from the center. The text is also blue and is in a sans-serif font.

Subject DeviceAI4CMR v1.0Predicate Device:Imbio RV/LV (K203256)
stand-alone basis for clinicaldecision-making.The user incorporating AI4CMRinto their DICOM application ofchoice is responsible forimplementing a user interface.provides the user with annotatedimages showing ventricularmeasurements. Its results are notintended to be used on astand-alone basis for clinicaldecision-making or otherwisepreclude clinical assessment ofCTPA cases.
Feature/FunctionSubject Device:AI4CMR v1.0Predicate Device:Imbio RV/LV(K203256)SubstantiallyEquivalent?
Indication for UseSee table aboveSee table aboveYes
Input DataRequirementsCardiovascular images:multi-phase,multi-slice acquired fromMRI scannersNon-gated, CT PulmonaryAngiography imagesYes¹
DICOM CompliantYesYesYes
LV SegmentationYesYesYes
RV SegmentationYesYesYes
Diameter MeasurementsYesYesYes
Fully AutomatedSegmentationYesYesYes
Interface3rd party Viewer as aplug-inCommand lineYes
OutputsReport onlyReport, DICOMSecondary Capture SeriesYes

See discussion below

The subject device and predicate device have similar indications for use and technological characteristics. Differences with the input data requirement do not raise questions of safety or effectiveness as the underlying technology is similar with similar risks that are mitigated by the same general and special controls.

7. Performance Data

Safety and performance of the AI4CMR v1.0 has been evaluated and verified in accordance with software specifications and applicable performance standards through software verification and validation testing. Additionally, the software validation activities were performed in accordance

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image contains the logo for AI4MedImaging. The logo consists of a circular graphic to the left of the company name. The graphic is made up of several blue lines that form a spiral shape. The company name, "AI4MedImaging", is written in blue, sans-serif font to the right of the graphic.

with ANSI AAMI IEC 62304:2006/41:2016 - Medical device software - Software life cycle processes, in addition to the FDA Guidance documents, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" and "Content of Premarket Submission for Management of Cybersecurity in Medical Devices."

7.1 Bench Testing

AI4MED performed a standalone performance test on the Society of Cardiac Magnetic Resonance (SCMR) Consensus Contour Data' which consists of a total of 15 CMR cases annotated by seven (7) independent expert readers from various core laboratories. The dataset consisted of male and female patients with an age range of 42 to 77 (average of 61) years old across various 1.5T and 3T scanners (Siemens, GE, Phillips). Readers performed myocardial segmentation and quantified End-diastolic volume (EDV), End-systolic volume (ESV), LV mass (LVM), and Ejection Fraction (EF). Agreement was evaluated and achieved between AI4CMR and the SCMR Consensus data. For myocardium segmentation, an average dice similarity coefficient (DSC) of 0.72 per image was achieved. For LVM, EDV, ESV, and EF the intraclass correlation coefficient (ICC) was evaluated and the following performance was obtained:

LV parameterLOA (± 2 SD)Bias ± SDr 2ICC
EDV[-58.5912, 45.9097] ml-6.3407 ± 26.1252 ml0.850.95
ESV[-30.2173, 22.4768] ml-3.8703 ± 13.1735 ml0.970.99
EF[-7.2668, 6.9567] %-0.155 ± 3.5559 %0.950.99
LVM[-57.2331, 11.7108] g-22.7611 ± 17.236 g0.720.78

7.2 Clinical Performance Assessment

AI4MED performed a multi-reader multi-center (MRMC) retrospective study consisting of 146 CMR cases with patients ranging from 17 to 85 years old (average of 51) that were predominantly male (77%) - consistent with cardiovascular disease incidence". Patient data consisted of diseased (~60%) and non-diseased (~40%) where the diseased was spread across the prevalent cardiovascular diseases worldwide'. CMR cases was acquired and balanced across Siemens, GE, and Philips 1.5T scanners with slice thickness of 8mm and slice diameter ranging from 8 to 10.5mm.

The primary objective was to evaluate agreement between the AI4MED device and 2 expert readers who achieved excellent interrater variability (ICC > 0.75). Readers manually segmented the myocardium for each CMR case per standard of care and manually determined volumes, LV mass, and Ejection Fraction. The dataset was independent of the data used for model training and development.

https://www.cardiacatlas.org/studies/scmr-consensus-data.

1 SCMR. Cardiac Atlas Project - SCMR Consensus Contour Data [Internet]. Available from:

2 Walli-Attaei, Marjan, et al. "Variations between women in risk factors, treatments, cardiovascular disease incidence, and death in 27 high-income, middle-income, and low-income countries (PURE); a prospective cohort study: " The Lancet 396.10244 (2020): 97-109 3 Ischemic heart disease, cardiomyopathis, pericardial abnormality, valve disease, cardiac mass tumor and others

AI4CMR v1.0 Traditional 510(k)

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image contains the logo for AI4MedImaging. The logo consists of a circular graphic on the left and the text "AI4MedImaging" on the right. The circular graphic is made up of several concentric circles, each of which is composed of a series of small, blue lines. The text "AI4MedImaging" is written in a sans-serif font and is also blue.

A summary of the agreement for Left and Right Ventricular EDV, Left and Right Ventricular ESV, Left and Right Ventricular Ejection Fraction, LV Myocardial Mass is provided below:

Left Ventricular EDV

Cronbach'sAlphaCorrelationCoef. (ρ)BiasICC95% Confidence Interval
Lower BoundUpper Bound
AI4CMR vs consensus0.9920.9808.6630.9900.980.99

Image /page/8/Figure/4 description: The image contains two scatter plots, labeled A and B. Plot A shows the difference between AI4CMR and Readers' Consensus on the y-axis, and the mean of AI4CMR and Readers' Consensus on the x-axis. The plot includes a mean line at 8.66, and lines for +1.96 SD at 42.65 and -1.96 SD at -25.33. Plot B shows AI4CMR on the y-axis and Readers' Consensus on the x-axis, with a regression line described by the equation Y = 14.49 + 0.90 * X and an R-squared value of 0.98.

[see next page]

AI4CMR v1.0 Traditional 510(k)

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for AI4MedImaging. The logo consists of a circular design on the left and the text "AI4MedImaging" on the right. The circular design is made up of several blue lines that form a spiral shape. The text is also in blue and is written in a clear, sans-serif font.

Left Ventricular ESV

Image /page/9/Figure/2 description: The image contains two scatter plots, labeled A and B, and a table of statistical values. Plot A shows the difference between AI4CMR and Readers' Consensus against the mean of AI4CMR and Readers' Consensus, with a mean difference of -2.89 and limits of agreement at +1.96 SD (31.56) and -1.96 SD (-37.35). Plot B shows AI4CMR values plotted against Readers' Consensus values, with a regression line equation of Y = 14.38 + 0.91 * X and an R^2 value of 0.97. The table presents statistical measures such as Cronbach's Alpha (0.992), Correlation Coefficient (0.975), Bias (-2.893), ICC (0.991), and a 95% Confidence Interval with a lower and upper bound of 0.99.

Left Ventricular Ejection Fraction

Cronbach'sAlphaCorrelationCoef. (ρ)BiasICC95% Confidence Interval
Lower BoundUpper Bound
AI4CMR vs consensus0,9690,9093,8670,9560,870,98

Image /page/9/Figure/5 description: The image contains two scatter plots, labeled A and B. Plot A shows the difference between AI4CMR and Readers' Consensus on the y-axis, plotted against the mean of AI4CMR and Readers' Consensus on the x-axis. The plot includes a mean line at 3.87 and lines indicating +1.96 SD at 15.15 and -1.96 SD at -7.42. Plot B shows AI4CMR on the y-axis plotted against Readers' Consensus on the x-axis, along with a regression line described by the equation Y = 1.36 + 0.89 * X, with an R^2 value of 0.89.

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image contains the logo for AI4MedImaging. The logo consists of a circular graphic on the left and the text "AI4MedImaging" on the right. The circular graphic is made up of several blue lines arranged in a circular pattern. The text "AI4MedImaging" is written in a blue sans-serif font.

Left Ventricular Myocardial Mass

Image /page/10/Figure/2 description: This image shows a table comparing AI4CMR vs consensus. The table includes Cronbach's Alpha, Correlation Coef. (p), Bias, ICC, and 95% Confidence Interval. The Cronbach's Alpha is 0.956, the Correlation Coef. (p) is 0.936, the Bias is 2.452, the ICC is 0.955, and the 95% Confidence Interval is 0.94 to 0.97.

Image /page/10/Figure/3 description: The image contains two scatter plots, labeled A and B. Plot A shows the difference between AI4CMR and Readers' Consensus on the y-axis, and the mean of AI4CMR and Readers' Consensus on the x-axis. The plot includes a mean line at 2.45, and lines indicating +1.96 SD at 37.94 and -1.96 SD at -33.03. Plot B shows AI4CMR on the y-axis and Readers' Consensus on the x-axis, with a regression line and the equation Y = 22.18 + 0.82 * X, and R^2 = 0.85.

Right Ventricular EDV

Cronbach'sAlphaCorrelationCoef. (ρ)BiasICC95% Confidence Interval
Lower BoundUpper Bound
AI4CMR vs consensus0,9720,9248,3550,9640,920,98

Image /page/10/Figure/6 description: The image contains two scatter plots, labeled A and B. Plot A shows the difference between AI4CMR and Readers' Consensus on the y-axis, and the mean of AI4CMR and Readers' Consensus on the x-axis. The plot includes a mean line at 8.35, and lines indicating +1.96 SD at 39.02 and -1.96 SD at -22.31. Plot B shows AI4CMR on the y-axis and Readers' Consensus on the x-axis, with a regression line and the equation Y = 10.70 + 0.88 * X, with R^2 = 0.90.

AI4CMR v1.0 Traditional 510(k)

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the logo for AI4MedImaging. The logo consists of a circular graphic on the left and the text "AI4MedImaging" on the right. The circular graphic is made up of several blue lines that form a spiral shape. The text "AI4MedImaging" is also blue and is written in a sans-serif font.

Right Ventricular ESV

Cronbach'sAlphaCorrelationCoef. (p)BiasICC95% Confidence Interval
Lower BoundUpper Bound
AI4CMR vs consensus0,9670,888-9,0830,9530,870,98

Image /page/11/Figure/3 description: The image contains two scatter plots, labeled A and B. Plot A shows the difference between AI4CMR and Readers' Consensus on the y-axis, and the mean of AI4CMR and Readers' Consensus on the x-axis. The plot includes a mean line at -9.08, and lines for +1.96 SD at 18.12 and -1.96 SD at -36.28. Plot B shows AI4CMR on the y-axis and Readers' Consensus on the x-axis, with a regression line and the equation Y = 11.47 + 0.96 * X and R^2 = 0.88.

Right Ventricular Ejection Fraction

Cronbach'sAlphaCorrelationCoef. (ρ)BiasICC95% Confidence Interval
Lower BoundUpper Bound
AI4CMR vs consensus0,9020,7127,8020,8140,150,93

Image /page/11/Figure/6 description: The image contains two scatter plots, labeled A and B. Plot A shows the difference between AI4CMR and Readers' Consensus on the y-axis, and the mean of AI4CMR and Readers' Consensus on the x-axis. The plot includes a mean line at 7.80, and lines indicating +1.96 SD at 22.20 and -1.96 SD at -6.60. Plot B shows AI4CMR on the y-axis and Readers' Consensus on the x-axis, with a regression line described by the equation Y = 0.93 + 0.86 * X and an R-squared value of 0.68.

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image contains the logo for AI4MedImaging. The logo consists of a circular graphic on the left and the text "AI4MedImaging" on the right. The circular graphic is made up of several blue lines arranged in a circular pattern. The text "AI4MedImaging" is written in a blue sans-serif font.

8. Conclusion

Based on the information submitted in this premarket notification, and based on the indications for use, technological characteristics and performance testing, the AI4CMR v1.0 raises no new questions of safety and effectiveness and is substantially equivalent to the predicate device in terms of safety and effectiveness.

§ 892.2050 Medical image management and processing system.

(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).