K Number
K220493

Validate with FDA (Live)

Manufacturer
Date Cleared
2023-10-27

(612 days)

Product Code
Regulation Number
878.4400
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

PLAXPOT™ MULTI PLASMA is intended for the removal and destruction of skin lesions and coagulation of tissue.

Device Description

PLAXPOT™ MULTI PLASMA is a device that consists of a handpiece which is the main body and a charger (charging cradle, and power adapter). The high frequency current is exposed to the air and the air is ionized. The electrode tip can perform treatment without direct contact with the tissue. After the power is applied through the internal battery, the mode is set and the Shot button is pressed, a high voltage alternating current pulse is released according to the set value, and the air is ionized to form a plasma arc. This device uses plasma energy. This causes excess skink and sublimate (direct passage from the solid to the gaseous state of matter) without affecting the deeper skin layers. The difference in potential distance between device's handpiece and the skin represents the area where the voltaic arc will take place. In this gap, the gases in the air are ionized, thus Plasma is generated. The tissues touched by plasma "sublime" without heat transfer to other unwanted tissues.

AI/ML Overview

The provided text is an FDA 510(k) clearance letter and an associated 510(k) summary for the PLAXPOT™ MULTI PLASMA device. This documentation primarily focuses on establishing "substantial equivalence" of the device to legally marketed predicate devices, rather than a detailed study proving the device's clinical performance against specific acceptance criteria for a novel AI/ML device.

Therefore, the document does not contain information regarding acceptance criteria for an AI/ML device, nor does it describe a study that proves the device meets such criteria. The device in question, an electrosurgical device, is not an AI/ML product.

The 510(k) summary (specifically sections 5.8 and Table 2) outlines performance bench tests designed to demonstrate the device's functional requirements and safety as an electrosurgical device, but these are not the type of studies (e.g., MRMC, standalone AI performance) you've inquired about for AI/ML device validation.

Here's an explanation of why the requested information is not present based on the provided text:

  • Acceptance Criteria for AI/ML Performance: The document does not mention any AI/ML components. The device operates on the principle of generating plasma through high-frequency current to remove/destroy skin lesions and coagulate tissue. Its "Performance Testing" relates to electrical safety, EMC, and thermal effects, which are standard for electrosurgical devices, not AI/ML performance metrics like sensitivity, specificity, AUC, or F1-score.
  • Study Proving Device Meets Acceptance Criteria: No study in the document describes the formal validation of an AI/ML algorithm against specified performance thresholds. The "Performance Testing" section mentions "bench tests" and "functional requirements," which are engineering and electrical safety checks, not clinical performance studies for an AI/ML diagnostic or therapeutic aid.
  • Table of Acceptance Criteria and Reported Performance: No such table exists for AI/ML performance. The table (Table 2) lists "Performance Testing Summary" for bench tests with "Pass" results, indicating compliance with electrical safety and other engineering standards.
  • Sample Size for Test Set and Data Provenance: Not applicable as there's no AI/ML test set defined.
  • Number of Experts and Qualifications for Ground Truth: Not applicable, as there's no AI/ML ground truth generation process described.
  • Adjudication Method: Not applicable.
  • Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study: Not mentioned, as it's not an AI/ML device requiring human-in-the-loop performance evaluation.
  • Standalone (Algorithm Only) Performance: Not applicable, as there's no standalone algorithm.
  • Type of Ground Truth Used: Not applicable.
  • Sample Size for Training Set: Not applicable.
  • How Ground Truth for Training Set was Established: Not applicable.

In summary, the provided document details the FDA clearance process for an electrosurgical device (PLAXPOT™ MULTI PLASMA) based on substantial equivalence to existing predicate devices. It covers safety, electrical standards, and intended use as an electrosurgical device, but it does not contain any information about an AI/ML component or studies related to AI/ML device performance.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 27, 2023

GCS Co. Limited % Su Kyung Park Director Compliance Insight 497 Circle Freeway Dr. Unit 230 West Chester, Ohio 45246

Re: K220493

Trade/Device Name: Plaxpot Multi Plasma Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: QVJ Dated: [NOTE: Use date of most recent supplement] Received: September 29, 2023

Dear Su Kyung Park:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely, Digitally signed by Mark Mark Trumbore -S Trumbore -S Date: 2023.10.27
16:19:00 -04'00' Mark Trumbore, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K220493

Device Name PLAXPOT™ MULTI PLASMA

Indications for Use (Describe) PLAXPOT™ MULTI PLASMA is intended for the removal and destruction of skin lesions and coagulation of tissue.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

PLAXPOT™ MULTI PLASMA

510(k) Summary

General Information 5.1

18 February 2022 Preparation Date:

26 October 2023 Revision Date:

Submitter/Holder

Su Kyung, Park

Director

GCS Co., Ltd. #1008, 555, Dunchon-daero, Jungwon-gu, Seongnam-si, Gyeonggi-do, Korea Tel: 82-2-555-1536 Fax : 82-2-3142-7670 Email: kara21c(@gcsgroup.co.kr

Website: http://www.bestgcs.com/

Primary Contact

Cynthia Ipach President/CEO

Email: cipach@compliance-insight.com

Office +1 513-860-3512 ext. 306 Compliance Insight 497 Circle Freeway Dr Suite 230. Cincinnati, OH 45246

5.2 Regulatory Information

Subject Device Trade NamePLAXPOT TM MULTI PLASMA
Classification NameLow Power Electrosurgical Devices For Skin LesionDestruction
Device ClassificationII
Regulation DescriptionElectrosurgical cutting and coagulation device andaccessories
FDA Product CodeQVJ
CFR References21 CFR 878.4400
Review PanelGeneral & Plastic Surgery

{4}------------------------------------------------

5.3 Identification of Predicate Devices

The primary predicate device for PLAXPOT™ MULTI PLASMA is the SubNovii Advanced Plasma Technology (K201738) and the reference device is Plasma IQ (K192813).

5.4 Subject Device Description

PLAXPOT™ MULTI PLASMA is a device that consists of a handpiece which is the main body and a charger (charging cradle, and power adapter). The high frequency current is exposed to the air and the air is ionized. The electrode tip can perform treatment without direct contact with the tissue.

After the power is applied through the internal battery, the mode is set and the Shot button is pressed, a high voltage alternating current pulse is released according to the set value, and the air is ionized to form a plasma arc.

This device uses plasma energy. This causes excess skink and sublimate (direct passage from the solid to the gaseous state of matter) without affecting the deeper skin layers. The difference in potential distance between device's handpiece and the skin represents the area where the voltaic arc will take place. In this gap, the gases in the air are ionized, thus Plasma is generated. The tissues touched by plasma "sublime" without heat transfer to other unwanted tissues.

ર્સ્ટ Indications for Use

PLAXPOT™ MULTI PLASMA is intended for the removal and destruction of skin lesions and coagulation of tissue.

Substantial Equivalence Discussion 5.6

In Table 1 on the following page please find a characteristics comparison summary. The review of the indications for use and comparison characteristics provided in Table 1 demonstrate that PLAXPOT™ MULTI PLASMA is substantially equivalent to the primary predicate device, the SubNovii Advanced Plasma Technology K201738.

{5}------------------------------------------------

DeviceCharacteristicsProposedDevicePrimaryPredicateReferenceDevice
Product NamePLAXPOTTMMULTI PLASMASubNoviiAdvancedPlasmaTechnologyPlasma IQ
510(k)K220493K201738K192813
ManufacturerGCS Co. Ltd.CartessaAestheticsNeauvia NorthAmerica
FDA ProductCodeQVJGEIGEI
CFR Reference878.4400878.4400878.4400
Device ClassIIIIII
Prescription orOTCPrescriptionPrescriptionPrescription
ImplantedDeviceNoNoNo
Indications forUse StatementPLAXPOT™MULTI PLASMAis intended for theremoval anddestruction of skinlesions andcoagulation oftissue.The SubNovii isintended for theremoval anddestruction ofskin lesions andcoagulation oftissue.Plasma IQ isused in theremoval anddestruction ofskin lesions andcoagulation oftissue.
Mode ofOperationPlasma highfrequency currentionizing the aircreating a plasmastream.Plasmaradiofrequencyenergy ionizesthe air creating aplasma stream.Plasmaradiofrequencyenergy ionizesthe air creatinga plasmastream.
Output ModeContinuousContinuousContinuous
Energy TypeHigh FrequencyHigh FrequencyHigh Frequency
Power Level3 modes at 0.7W,1W, 2WMax 5WMax 5W
Reusable orSingle UseReusableReusableReusable
OutputMonopolarMonopolarMonopolar
Max PowerOutput3W5W5W
Power SupplyInternalRechargeableLithium IonBattery 7.4 VDC1000mAhInternalRechargeableLithium IonBattery 11.1VDCInternalRechargeableLithium IonBattery 11.1VDC
Frequency70 kHz40kHz40kHz
SystemComponentsA handpiece thatincorporates theelectrosurgicalgenerator unit,OLED display,software, anddocking station.A handpiece thatincorporates theelectrosurgicalgenerator unit,docking stationand an activeelectrode.A handpiecethatincorporates theelectrosurgicalgenerator unit,docking stationand an activeelectrode.
ElectricalSafetyStandardsIEC 60601-1IEC 60601-1-2IEC 60601-1IEC 60601-1-2IEC 60601-1IEC 60601-1-2
SoftwareYesNoNo
SoftwareComplianceIEC 62304N/AN/A

Table 1. Summary Comparison of Characteristics

{6}------------------------------------------------

PLAXPOT™ MULTI PLASMA

5.7 Sterilization and Shelf Life

The PLAXPOT™ MULTI PLASMA is sold non-sterile.

{7}------------------------------------------------

Shelf-life is not applicable because of low likelihood of time-dependent product degradation.

Performance Testing - Bench 5.8

Performance bench tests of PLAXPOT™ MULTI PLASMA have been performed, see Table 2. The results from the performance bench testing demonstrate that PLAXPOT™ MULTI PLASMA has met the functional requirements and is substantially equivalent to the predicate device. Performance bench testing is provided in detail.

StudyAttachmentTest MethodResults
Basic Safety andEssential Performance18-3EN 60601-1:2006+A1:2013 or EN60601-1:2006+A12:2014 andtest reportIEC60601_1KPass
ManufacturerTechnical File18-5N/APass
EMC Test18-6IEC 60601-1-2:2014Pass
EMC Test18-7IEC 60601-1-2:2014Pass
Thermal Effects TestNAFDA Submissions forElectrosurgical Devicesfor General SurgeryPass

Table 2. Performance Testing Summary

5.9 Conclusion

The subject device PLAXPOT™ MULTI PLASMA has the same technology, principle of operation, indications for use and technical specifications as the primary predicate device Subnovii and the reference device Plasma IQ. PLAXPOT™ MULTI PLASMA shares a substantially equivalent design, indications for use and technology (i.e. features, materials, and principles of operation) as the predicate device and no new elements pertaining to safety or effectiveness have been identified. Performance test results demonstrate that the subject device can perform the same intended use as safely as the predicate and reference devices. Therefore, the subject device is substantially equivalent to the predicate.

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.