K Number
K220472

Validate with FDA (Live)

Date Cleared
2022-08-11

(174 days)

Product Code
Regulation Number
878.4400
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The NEUWAVE™ Microwave Ablation System and Accessories are indicated for the ablation (coagulation) of soft tissue in percutaneous, open surgical and in conjunction with laparoscopic surgical settings, including the partial or complete ablation of non-resectable liver tumors.

The NEUWAVE™ Microwave Ablation System is not indicated for use in cardiac procedures

Device Description

The NEUWAVE™ Microwave Ablation System is a self-contained stand-alone system of hardware and software designed for the ablation of soft tissue which applies microwave energy to produce tissue heating effects generating tissue necrosis.

The system consists of three major components (1) a cart which contains the components necessary to deliver microwave power to the microwave ablation probes, monitor and control system functions, a graphical user interface for the user to interact with the system and a CO2 based cooling system, (2) a Power Distribution Module (PDM), and (3) a range of microwave ablation probe accessories for energy delivery.

The system has a single 2.45 GHz signal microwave source with three 140W microwave power amplifiers, a touch-screen graphical user interface, and a CO2 based cooling system for limiting the temperature of the ablation probe, handle, and cable.

The CO2 cooling system enables the Tissu-Loc function, which involves formation of an ice ball to adhere the probe in place prior to starting ablation therapy. This helps eliminate probe migration during imaging (CT scans, ultrasounds, etc.) and additional probe placement. The cooling system is responsible for controlling the pressure of the incoming CO2 gas from two E-sized CO2 cylinders located on the back of the system.

The graphical user interface allows the user to set, adjust and activate the power setting per probe, time setting for each probe, ability to synchronize ablation initiation on probes, ablation activation, cauterization activation, audible volume, probe temperature, and Tissu-Loc function.

Up to three (3) NEUWAVE Ablation Probes can connect to the PDM at once and the PDM allows power to be transferred from the system generator to the ablation probes.

The microwave ablation probes are accessories which transfer microwave energy from the system cart to the target tissue to create regions of thermal necrosis. Each probe contains thermocouples that monitor the temperature of the probes can be used for surgical mode or ablation mode with various limits of maximum selectable power and time based on the probe type.

AI/ML Overview

The NeuWave Medical Inc. NEUWAVE™ Microwave Ablation System and Accessories (K220472) underwent non-clinical testing to demonstrate substantial equivalence to its predicate device (K200081). The study focused on assessing electrical safety, electromagnetic compatibility, and ex-vivo tissue ablation to ensure that updated hardware and software components did not compromise the device's design and performance specifications.

Here's a breakdown of the acceptance criteria and study details:

1. Table of Acceptance Criteria and Reported Device Performance:

Test PerformedAcceptance CriteriaReported Device Performance
Electrical Safety TestingCompliance with IEC 60601-1 3rd Edition (demonstrate electrical safety).Pass
Electromagnetic Compatibility (EMC) TestingCompliance with IEC 60601-1-2 4th Edition (requirements for immunity, radiated, and conducted emissions).Pass
Ex-vivo tissue ablation testingAblation zone dimensions of the subject device are substantially equivalent to the predicate device for identical power settings.Pass
Software Verification TestingCompliance with "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, dated May 11, 2005."Pass
CybersecurityConformance to FDA's "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices" guidance document, including strengthened cybersecurity posture.Assessment performed and updates implemented.

2. Sample Size Used for the Test Set and Data Provenance:

  • Sample Size: Not explicitly stated as a specific number of cases or samples. For electrical safety and EMC testing, "the NEUWAVE™ Microwave Ablation System with NEUWAVE™ Probes" was tested. For ex-vivo tissue ablation, the testing involved "ablation zone dimensions of the subject and predicate devices." This implies a sufficient number of tests were conducted to ensure comparability.
  • Data Provenance: The studies were non-clinical design verification tests, performed by the manufacturer (NeuWave Medical Inc.) as part of their premarket notification. The data is prospective, generated specifically for this submission. The country of origin is implicitly the USA, where NeuWave Medical Inc. is based.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

  • This information is not provided as the studies were non-clinical, hardware, and software verification tests. The "ground truth" was established by engineering specifications, relevant IEC standards, and direct comparison of a physical characteristic (ablation zone dimensions) rather than expert interpretation of medical images or outcomes.

4. Adjudication Method for the Test Set:

  • This information is not applicable as the tests were non-clinical engineering and performance evaluations, not involving human interpretation or adjudication of medical data.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done:

  • No, an MRMC comparative effectiveness study was not done. The submission explicitly states, "No clinical data was generated or is required to support of this Premarket 510(k) Notification." The focus was on demonstrating substantial equivalence through non-clinical performance data due to minor hardware and software updates to an already cleared predicate device.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

  • The device is a physical microwave ablation system with embedded software, not a standalone algorithm in the typical sense of AI image analysis. The "software verification testing" was done in a standalone manner (evaluating the software itself), but it's not an "algorithm only without human-in-the-loop performance" that would typically be described for diagnostic AI tools. The system requires human operation for its intended use.

7. The Type of Ground Truth Used:

  • For Electrical Safety and EMC: Compliance with international safety and electromagnetic compatibility standards (IEC 60601-1, IEC 60601-1-2) served as the ground truth.
  • For Ex-vivo tissue ablation: The "ground truth" was the physical measurement of ablation zone dimensions, compared directly between the subject and predicate devices for identical power settings. This is a direct physical measurement.
  • For Software Verification: Compliance with FDA guidance for software in medical devices (functional correctness, robustness, etc.) served as the ground truth.
  • For Cybersecurity: Conformance to FDA guidance on cybersecurity management.

8. The Sample Size for the Training Set:

  • Not applicable. This submission focuses on a modified device and its non-clinical verification. It does not describe the development or training of a new AI algorithm that would typically involve a "training set" of data for machine learning. The software updates were primarily to support updated hardware due to component obsolescence, not to introduce new AI-driven functionalities requiring extensive data training.

9. How the Ground Truth for the Training Set Was Established:

  • Not applicable. As no training set for an AI algorithm was described, the method for establishing its ground truth is not relevant to this submission.

{0}------------------------------------------------

August 11, 2022

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

NeuWave Medical Inc. Mohamed Shariff Associate Director, Regulatory Affairs 3529 Anderson Street Madison, Wisconsin 53704

Re: K220472

Trade/Device Name: NEUWAVE Microwave Ablation System and Accessories Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: NEY Dated: July 6, 2022 Received: July 7, 2022

Dear Mohamed Shariff:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Colin Kejing Chen, Ph.D. Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known)

K220472

Device Name

NEUWAVE™ Microwave Ablation System and Accessories

Indications for Use (Describe)

The NEUWAVE™ Microwave Ablation System and Accessories are indicated for the ablation (coagulation) of soft tissue in percutaneous, open surgical and in conjunction with laparoscopic surgical settings, including the partial or complete ablation of non-resectable liver tumors.

The NEUWAVE™ Microwave Ablation System is not indicated for use in cardiac procedures

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K220472

Image /page/3/Picture/1 description: The image shows the logo for NeuWave Medical, a part of the Ethicon family of companies. The logo features a green circular icon with a white wave-like design on the left. To the right of the icon is the text "NeuWave Medical" in a simple, sans-serif font, with the words "PART OF THE ETHICON™ FAMILY OF COMPANIES" appearing in a smaller font size below.

510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

Date:August 8, 2022
Submitter Name/Address:NeuWave Medical, Inc.Part of the Ethicon Family ofCompanies 3529 Anderson StreetMadison, WI 53704
Contact Person:Mohamed ShariffAssociate Director, Regulatory AffairsTel: (856) 449 9609Fax: (608) 512 1508Email: msharif3@its.jnj.com
FDA Establishment Registration #3008769756
Device Identification
Trade/Proprietary Name:NEUWAVE™ Microwave Ablation System
Classification Name:Electrosurgical Cutting andCoagulation Device and Accessories
Product Code:NEY
CFR Section:21 CFR 878.4400
Device Class:Class II
Classification Panel:General and Plastic Surgery Devices

{4}------------------------------------------------

Predicate Device

The NEUWAVE™ Microwave Ablation System and Accessories is substantially equivalent to the following currently marketed device:

NEUWAVE™ Microwave Ablation System and Accessories cleared under K200081 on November 13, 2020.

Indications for Use

The NEUWAVE™ Microwave Ablation System and Accessories are indicated for the ablation (coagulation) of soft tissue in percutaneous, open surgical and in conjunction with laparoscopic surgical settings, including the partial or complete ablation of non-resectable liver tumors.

The NEUWAVE" Microwave Ablation System is not indicated for use in cardiac procedures.

Device Description

The NEUWAVE™ Microwave Ablation System is a self-contained stand-alone system of hardware and software designed for the ablation of soft tissue which applies microwave energy to produce tissue heating effects generating tissue necrosis.

The system consists of three major components (1) a cart which contains the components necessary to deliver microwave power to the microwave ablation probes, monitor and control system functions, a graphical user interface for the user to interact with the system and a CO2 based cooling system, (2) a Power Distribution Module (PDM), and (3) a range of microwave ablation probe accessories for energy delivery.

The system has a single 2.45 GHz signal microwave source with three 140W microwave power amplifiers, a touch-screen graphical user interface, and a CO2 based cooling system for limiting the temperature of the ablation probe, handle, and cable.

The CO2 cooling system enables the Tissu-Loc function, which involves formation of an ice ball to adhere the probe in place prior to starting ablation therapy. This helps eliminate probe migration during imaging (CT scans, ultrasounds, etc.) and additional probe placement. The cooling system is responsible for controlling the pressure of the incoming CO2 gas from two E-sized CO2 cylinders located on the back of the system.

The graphical user interface allows the user to set, adjust and activate the power setting per probe, time setting for each probe, ability to synchronize ablation initiation on probes, ablation activation, cauterization activation, audible volume, probe temperature, and Tissu-Loc function.

Up to three (3) NEUWAVE Ablation Probes can connect to the PDM at once and the PDM allows power to be transferred from the system generator to the ablation probes.

The microwave ablation probes are accessories which transfer microwave energy from the system cart to the target tissue to create regions of thermal necrosis. Each probe contains thermocouples that monitor the temperature of the probes can be used for surgical mode or ablation mode with various limits of maximum selectable power and time based on the probe type.

{5}------------------------------------------------

NEUWAVE™ LK, and NEUWAVE™ PR ablation probes are available in either 17- gauge or 15- gauge cannulas and are available in 15 cm and 20 cm lengths. These probes have a cable length of 1.4m.

NEUWA VE™ LN ablation probes are available only in 17-gauge cannulas and are available in 15 cm and 20 cm lengths. These probes have a cable length of 1.4m.

The model NEUWAVE™ SR ablation probes have a 13-gauge cannula and are available in a 25 cm length only. NEUWAVE™ SR probes have a cable length of 1.4m.

NEUWAVE™ Surgical PR Ablation Probes have a 15-gauge cannula and a 15 cm length. These probes have a 2.9m cable length and the probe tip has a non-stick coating that facilitates repeat ablations/Planar Coagulation and helps prevent ablated tissue from adhering to the tip of the probe.

The antenna of the NEUWAVE™ PR probe is designed to limit the length of the ablation for instances when a shorter ablation zone is desired. NEUWAVE™ PR Probes were developed to provide physicians with an additional ablation probe option with a different ablation burn pattern compared with NEUWAVE™ LK/LN/SR probes. The NEUWAVE™ PR probes are designed to produce ablations that quickly encompass the tip of the probe while limiting the overall length of the ablation compared with other NEUWAVE ablation probes.

The wide selection of ablation probes provides physicians with the opportunity to select the length, stiffness, burn pattern and number of probes to use to create a wide variety of ablation zone sizes and shapes.

A surgical clip is a single use sterile surgical accessory designed to hold two microwave ablation probes together in a parallel configuration at a fixed distance apart to facilitate planar coagulation of soft tissue during a surgical resection.

An optional footswitch accessory can be connected to the system to control power delivery in Surgical Mode.

Ablation Confirmation software (previously cleared under K192427) is available as an optional feature for the NEUWAVE™ Microwave Ablation System. When this option is supplied, a second monitor is provided with the cart which hosts the Ablation Confirmation graphical user interface.

{6}------------------------------------------------

Comparison of Intended Use and Indications for Use

The subject device and predicate device have the same Intended Use and Indications for Use as shown in the table below.

Predicate DeviceSubject Device
DeviceCharacteristicsNEUWAVE™ Microwave Ablation Systemand Accessories (K200081)NEUWAVE™Microwave AblationSystem and Accessorieswith updated Hardwareand Software
Intended UseTo thermally ablate soft tissue using microwaveenergy.Same
Indications for UseThe NEUWAVE™ Microwave Ablation Systemand Accessories are indicated for the ablation(coagulation) of soft tissue in percutaneous, opensurgical and in conjunction with laparoscopicsurgical settings, including the partial or completeablation of non-resectable liver tumors.The NEUWAVE™ Microwave AblationSystem is not indicated for use in cardiacSame

{7}------------------------------------------------

Comnarison of Technological Characteristics with Predicate Device

The subject and the predicate devices are similar in terms of technological characteristics as microwave devices used to ablate soft tissue.

The subject device NEUWAVE™ Microwave Ablation System includes some updated but functionally equivalent hardware components which were implemented due to component obsolescence, and a corresponding software update to support the updated hardware. No changes were made that impact the delivery of microwave energy and no changes were made to NEUWAVE™ Ablation probes or accessories.

Predicate DeviceSubject Device
Feature/SpecificationNEUWAVE™ Microwave AblationSystem and Accessories (K200081)NEUWAVE™Microwave AblationSystem andAccessories withupdated Hardwareand Software
Software VersionV3.1.0 softwareV3.1.1
Probe ApplicationsPercutaneous, open surgical and in conjunctionwith laparoscopic surgical settings.Same
User Interface ModesSurgical and Ablation ModeSame
Power Delivery InitiationMethodUser Interface, Footswitch (Footswitchavailable in Surgical Mode only) or Finger-SwitchSame
Predicate DeviceSubject Device
Feature/SpecificationNEUWAVETM Microwave AblationSystem and Accessories (K200081)NEUWAVETMMicrowave Ablation System andAccessories with updated Hardwareand Software
Probe dimensions
NEUWAVETM LK ProbeDiameter: 15 gauge, Length: 15 cm and 20 cmDiameter: 17 gauge, Length: 15 cm and 20 cmSame
NEUWAVETM LN ProbeDiameter: 17 gauge, Length: 15 cm and 20 cmSame
NEUWAVETM SR ProbeDiameter: 13 gauge, Length: 25 cmSame
NEUWAVETM PR ProbeDiameter: 15 gauge, Length: 15 cm and 20 cmDiameter: 17 gauge, Length: 15 cm and 20 cmSame
NEUWAVETM SurgicalPR ProbeDiameter: 15 gauge, Length: 15 cmSame
Generator Maximum Output Power
NEUWAVETM LK ProbeAblation Modeand SurgicalMode:Single Probe is 140W,Two Probes is 95W,Three Probes is 65WSame
NEUWAVETM LN Probe
NEUWAVETM SR Probe
NEUWAVETM PR ProbeAblation Mode:Single Probe is 65W, TwoProbes is 65W, ThreeProbes is 65WSame
Surgical Mode:Single Probe is 95W, TwoProbes is 95W, ThreeProbes is 65W
NEUWAVETM SurgicalPR ProbeAblation Mode:Single Probe is 80W, TwoProbes is 80W, ThreeProbes is 80WSame
Surgical Mode:Single Probe is 120W,Two Probes is 120W,Three Probes is 80W
Antenna DesignTriaxial Antenna for NEUWAVETM LK Probe,NEUWAVETM LN Probe and NEUWAVETMSR ProbeSame
Modified Triaxial Antenna for NEUWAVETMPR Probe and NEUWAVEM Surgical PRProbeSame
Predicate DeviceSubject Device
Feature/SpecificationNEUWAVE™ Microwave AblationSystem and Accessories (K200081)NEUWAVE™Microwave AblationSystem andAccessories withupdated Hardwareand Software
Settable Parameters
PowerNEUWAVE™ LK Probe, NEUWAVE™ LNProbe and NEUWAVE™ SRProbe: 20W-140W in 5WNEUWAVE™ PR Probe:20W-95W in 5 W increments(Ablation Mode maximum is 65W)NEUWAVE™ Surgical PR Probes:20W-120W in 5 W increments(Ablation Mode maximum is 80W)Same
Target Ablation TimeUp to 10 minutes as limited by software.User may ablate for additional time after 10minutes of ablation completion.Same
Planar Coagulation Time5 seconds - 1 MinuteSame
AccessoriesDual Probe ClipCT Table Mounting Adapter for PDMSurgical Table Mounting Adapter for PDMFoot switch (Locking USB or Standard USB)Same

{8}------------------------------------------------

{9}------------------------------------------------

{10}------------------------------------------------

Performance Data

Nonclinical Testing

Design Verification testing has been completed to assure that the modified NEUWAVE™ Microwave Ablation System and accessories meets its design and performance specifications; this testing is summarized in the following table. Electrical safety testing, electromagnetic compatibility testing and ex-vivo ablation tissue testing was conducted in accordance "Premarket Notification (510(k)) Submissions for Electrosurgical Devices. The subject device passed the system level test performance criteria with providing evidence of substantial equivalence between the subject and predicate device.

Test PerformedDescriptionPass/Fail
Electrical SafetyTestingElectrical Safety testing was undertaken per IEC 60601-1 3rdEdition to demonstrate that the NEUWAVETM MicrowaveAblation System with NEUWAVETM Probes is electricallysafePass
Electromagneticcompatibility(EMC) testingElectromagnetic compatibility (EMC) testing wasundertaken per IEC 60601-1-2 4th Edition to demonstratethat the NEUWAVETM Microwave Ablation System withNEUWAVETM Probes complies with the requirements forimmunity, radiated and conducted emissionsPass
Ex-vivo tissueablation testingEx-vivo tissue testing was conducted to demonstrateablation zone dimensions of the subject and predicatedevices are substantially equivalent for identical powersettings.Pass

Software verification testing was conducted, and results passed as recommended by Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, dated May 11, 2005.

Clinical Data

No clinical data was generated or is required to support of this Premarket 510(k) Notification.

Cybersecurity

Updates to the underlying operating system were performed to strengthen the cybersecurity posture of the system. Also, an assessment was performed to assure conformance to the FDA's "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices" guidance document.

{11}------------------------------------------------

Sterilization

No sterilization data was generated or required to support of this Premarket 510(k) Notification.

Biocompatibility

No biocompatibility data was generated or required to support of this Premarket 510(k) Notification.

Standards

  • 60601-1:2005+C1;A2, Medical Electrical Equipment, Part 1: General Requirements For Basic . Safety And Essential Performance
  • 60601-1:2014 Ed.3, Medical Electrical Equipment, Part 1: General Requirements For Basic ● Safety And Essential Performance
  • 60601-1-6:2010 Ed.3+A1, Medical Electrical Equipment Part 1-6: General Requirements For ● Basic Safety And Essential Performance - Collateral Standard: Usability
  • 60601-2-6:2012 Ed.2, Medical Electrical Equipment Part 2-6: Particular Requirements For The Basic Safety & Essential Performance Of Microwave Therapy Equipment
  • 60601-2-18:2009 Ed.3, Medical Electrical Equipment Part 2-18: Particular Requirements For ● The Basic Safety And Essential Performance Of Endoscopic Equipment

Conclusion

There are no differences in fundamental technological characteristics of the subject and predicate devices, and the minor changes implemented do not impact the essential performance of the subject device system and accessories. Based on the intended use, technological characteristics, and validation/verification data, the modified NEUWAVE™ Microwave Ablation System and Accessories is substantially equivalent to the predicate.

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.