K Number
K220466

Validate with FDA (Live)

Manufacturer
Date Cleared
2022-11-03

(259 days)

Product Code
Regulation Number
870.1875
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ibiomedi Electronic Stethoscope ES-2020 is used to detect sound from the heart, arteries, veins, breathing sounds from the anterior/lateral chest and throat sounds in the neck. The chest piece is designed for children who are over two years old, teenagers and adults. It can be applied to any body type and can only be used for the purpose of medical diagnosis in clinics or hospitals.

Device Description

The ibiomedi Electronic Stethoscope ES-2020 detects sounds from the heart, arteries, veins, breathing sounds from the anterior/posterior/lateral chest and throat sounds in the neck of patients. The sounds are transmitted to the user's ears through accessories such as earphones.

User interface includes A: Power button, frequency response mode button, volume button B: Power indicator, wireless transmission indicator, frequency response mode indicator, volume indicator.

Turn on the ibiomedi Electronic Stethoscope ES-2020 and connect to the wireless device via Bluetooth to transmit and store sounds. When the ibiomedi Electronic Stethoscope ES-2020 and the connected wireless device have walls, human bodies and other barriers, the effective range of Bluetooth transmission will be affected. It is recommended to reduce the distance between the ibiomedi Electronic Stethoscope ES-2020 and the connected wireless device to improve Bluetooth connection.Power to the device is provided by two AAA 1.5V batteries.

The associated accessories include.

  • Small chest piece: For children with diaphragm. .
  • Large chest piece: For adult with diaphragm.
  • Audio line: For connecting external speakers.
  • . 1MORE Earphones: For medical personnel to connect the ibiomedi Electronic Stethoscope ES-2020.
AI/ML Overview

The provided text describes the ibiomedi Electronic Stethoscope ES-2020 which is used to detect sounds from the heart, arteries, veins, breathing sounds, and throat sounds. The text focuses on establishing substantial equivalence to a predicate device, as required by a 510(k) submission.

Here's an analysis of the acceptance criteria and study information provided in the document:

1. A table of acceptance criteria and the reported device performance

The document does not explicitly present a table of acceptance criteria with corresponding device performance for direct clinical efficacy or diagnostic accuracy. Instead, the performance data provided focuses on:

Acceptance Criterion TypeReported Device Performance / Evaluation
Biocompatibility- O-rings and 3M 1525L Polyethylene underwent testing. - Followed modified ISO 10993-10:2010 (Irritation and Skin Sensitization), ISO 10993-5:2009 (Cytotoxicity). - Tests included Guinea Pig Maximization Test (GPMT), Skin irritation test in rabbits, and Cytotoxicity study.
Electrical Safety & Electromagnetic Compatibility (EMC)- Complies with IEC 60601-1:2005, COR1:2006, COR2:2007, AMD1:2012; EN 60601-1: 2006 + A1: 2013; AAMI/IEC 60601-1:2005 + AMD 1:2012; CAN/CSA-C22.2 No. 60601-1:14 (for safety). - Complies with IEC 60601-1-2: 2014 / EN 60601-1-2: 2015 (for EMC).
Software Verification & Validation- Level of concern identified as "Minor" (unlikely to cause injury). - Software passed tests and met requirements of relevant standards.
Performance Testing (Sound Amplification)- Tests determined that sound amplification was not affected by the use of earphones. - Performance data presented by traditional stethoscope and the subject device are "similar."
Usability Study- Acceptance Criteria: At least 85% of participants with no experience in this type of product will be able to use it properly after reading Instruction Manual for around 30 minutes. - Summary: The summative evaluation objective of usability was met.

2. Sample size used for the test set and the data provenance

  • Usability Study: The sample size was 15 doctors and 15 nurses.
  • Data Provenance: Not explicitly stated, but given that the applicant is Sound Land Corp. in Taiwan, it is likely that the usability study participants were from Taiwan. The document does not specify if the study was retrospective or prospective.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

The document does not describe an expert-based ground truth establishment for the "Performance Testing" or "Usability Study" beyond the participants in the usability study being "doctors and nurses."

For the "Performance Testing" which compared the device to a traditional stethoscope, it's mentioned that their performance data are "similar," implying an unquantified comparison, not a ground truth derived from expert adjudication of sound quality or diagnostic outcomes.

4. Adjudication method for the test set

There is no mention of an adjudication method for establishing a ground truth.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

There is no information about an MRMC comparative effectiveness study in the provided text. The device is an electronic stethoscope, not an AI-assisted diagnostic tool that would typically involve human readers interpreting images with or without AI assistance. The "doctors and nurses" in the usability study were likely evaluating ease of use, not diagnostic improvement with AI.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This section is not applicable as the ibiomedi Electronic Stethoscope ES-2020 is a sound detection and amplification device, not an algorithm that performs standalone analysis or diagnosis without human interaction.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The document does not report the use of a formal ground truth (like expert consensus, pathology, or outcomes data) for evaluating the diagnostic accuracy or clinical effectiveness of the stethoscope. The tests focused on safety, EMC, software functionality, and usability, as well as a general comparison to traditional stethoscopes regarding sound amplification.

8. The sample size for the training set

No information about a training set is provided. This is consistent with the nature of the device, which is an electronic stethoscope rather than a machine learning or AI-based diagnostic system that would require a distinct training set.

9. How the ground truth for the training set was established

Not applicable, as there is no mention of a training set or ground truth for such a set.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 3, 2022

Sound Land Corp. Chi-Hsun Hung QC Manager No.32, Keji 1st Rd., Guishan Dist. Taoyuan, 33383 Taiwan

Re: K220466

Trade/Device Name: ibiomedi Electronic Stethoscope ES-2020 Regulation Number: 21 CFR 870.1875 Regulation Name: Stethoscope Regulatory Class: Class II Product Code: DQD, Dated: October 6, 2022 Received: October 6, 2022

Dear Chi-Hsun Hung:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.

510(k) Number (if known)

K220466

Device Name

ibiomedi Electronic Stethoscope ES-2020

Indications for Use (Describe)

The ibiomedi Electronic Stethoscope ES-2020 is used to detect sound from the heart, arteries, veins, breathing sounds from the anterior/lateral chest and throat sounds in the neck. The chest piece is designed for children who are over two years old, teenagers and adults. It can be applied to any body type and can only be used for the purpose of medical diagnosis in clinics or hospitals.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

□ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.

FORM FDA 3881 (6/20)

Page 1 of 1

PSC Publishing Services (101) 443-6740 EF

{3}------------------------------------------------

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 C.F.R. §807.92.

1.Preparation Date:09/24/2021
2.Applicant/Sponsor:Sound Land Corp.No.32, Keji 1st Rd., Guishan Dist., Taoyuan City33383, Taiwan
3.Contact Person:Hung, Chi-HsunTel:+886-3-396-0158Fax:+886-3-396-0153Email:eric@soundland.com.tw
4.Proprietary Name:ibiomedi Electronic Stethoscope ES-2020
5.Common Name:Electronic Stethoscope
6.Classification Name:21 CFR 870.1875Stethoscope, Electronic
7.Classification Identification: Class II
8.Product Codes:DQD
9.Predicate Device:Stethoscope, Electronic K160023

10. Device Description:

The ibiomedi Electronic Stethoscope ES-2020 detects sounds from the heart, arteries, veins, breathing sounds from the anterior/posterior/lateral chest and throat sounds in the neck of patients. The sounds are transmitted to the user's ears through accessories such as earphones.

User interface includes A: Power button, frequency response mode button, volume button B: Power indicator, wireless transmission indicator, frequency response mode indicator, volume indicator.

{4}------------------------------------------------

Turn on the ibiomedi Electronic Stethoscope ES-2020 and connect to the wireless device via Bluetooth to transmit and store sounds. When the ibiomedi Electronic Stethoscope ES-2020 and the connected wireless device have walls, human bodies and other barriers, the effective range of Bluetooth transmission will be affected. It is recommended to reduce the distance between the ibiomedi Electronic Stethoscope ES-2020 and the connected wireless device to improve Bluetooth connection.Power to the device is provided by two AAA 1.5V batteries.

The associated accessories include.

  • Small chest piece: For children with diaphragm. .
  • Large chest piece: For adult with diaphragm.
  • Audio line: For connecting external speakers.
  • . 1MORE Earphones: For medical personnel to connect the ibiomedi Electronic Stethoscope ES-2020.

11. Indications for Use:

The ibiomedi Electronic Stethoscope ES-2020 is used to detect sound from the heart, arteries, veins, breathing sounds from the anterior/lateral chest and throat sounds in the neck. The chest piece is designed for children who are over two years old, teenagers and adults. It can be applied to any body type and can only be used for the purpose of medical diagnosis in clinics or hospitals.

12. Comparison of Technological Characteristics with The Predicate Device:

The ibiomedi Electronic Stethoscope ES-2020 uses a miniature microphone to pick up sound through the chest piece and uses noise-cancelling technology to filter murmurs to distinguish heart sounds with a frequency of 20200Hz, lung sounds with a frequency of 1001000Hz, and all sounds within 1200Hz. And it can amplify the sound up to 50X, which is helpful for medical personnel to make judgments.

At a high level, the subject and predicate devices are based on the following same technological elements:

  • . Chest piece & miniature microphone: To use a miniature microphone to pick up sound through the chest piece.
  • Noise-cancelling technology: To filter murmurs and distinguish the frequency.
  • . Sound amplification: To amplify sound of specific frequency.

The following technological differences exist between the subject and predicate devices:

  • Use earphones instead of ear tubes.
  • . Replaceable chest piece.
  • Use only AAA alkaline batteries.

{5}------------------------------------------------

13. Performance Data

The following performance data were provided in support of the substantial equivalence determination.

Biocompatibility testing

Tests included on O-rings and 3M 1525L Polyethylene from the chest piece that contact with the patient's skin.

The biocompatibility evaluation for the ibiomedi Electronic Stethoscope ES-2020 was conducted in accordance with modified ISO 10993-10:2010, Biological evaluation of medical devices-Part 10: Tests for irritation and skin sensitization. The Guinea Pig Maximization Test (GPMT) was used to evaluate the hypersensitivity-inducing effect of the test article, and ISO 10993-10: 2010, Biological evaluation of medical devices-Part 10: Tests for irritation and skin sensitization. The potential of the test article under test to produce irritation was valuated following dermal application of the test article extracts from polar and non-polar extractions, and ISO 10993-5: 2009, Biological evaluation of medical devices-Part 5: Tests for in vitro cytotoxicity.

  • Skin sensitization study guinea pig maximization test
  • Skin irritation test in rabbits
  • Cytotoxicity study for biocompatibility

Electrical safety and electromagnetic compatibility (EMC)

Electrical safety and EMC testing were conducted on the ibiomedi Electronic Stethoscope ES-2020. The system complies with the IEC 60601-1:2005, COR1:2006, COR2:2007, AMD1:2012 and EN 60601-1: 2006 + A1: 2013 and AAMI/IEC 60601-1:2005 + AMD 1:2012 and CAN/CSA-C22.2 No. 60601-1:14 standards for safety, and IEC 60601-1-2: 2014 / EN 60601-1-2: 2015 standard for EMC.

Software Verification and Validation Testing

Sound Land considers the level of concern to be " Minor ". According to the guidance " Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices " Document issued on: May 11, 2005, The definition of the " Minor " is as follows: " the level of concern in Minor if failures or latent design flaws are unlikely to cause any injury to the patient or operator ".

The software of the ibiomedi Electronic Stethoscope ES-2020 has passed the test and met the requirements of the above standards, the ibiomedi Electronic Stethoscope ES-2020 is quite safe and reliable in terms of software.

Performance Testing

Tests were conducted to determine whether the amplification of sound was affected by the use of earphones. According to the results, the performance data presented by traditional stethoscope and the subject device are similar.

{6}------------------------------------------------

Usabilitv Studv

Usability testing of the ibiomedi Electronic Stethoscope ES-2020 included usability study of 15 doctors and 15 nurses.

Usability evaluation objective with the acceptance criteria:

At least 85% of participants with no experience in this type of product will be able to use it properly after reading Instruction Manual for around 30 minutes.

Summary:

The summative evaluation objective of usability is met. For the purposes of ISO 14971, the residual risks associated with sability shall be presumed to be acceptable.

14. Conclusions

The results of the bench performance tests, usability studies, the hardware and software verification and validation demonstrate that the ibiomedi Electronic Stethoscope ES-2020 should perform as intended in the specified use conditions. The data demonstrate that the ibiomedi Electronic Stethoscope ES-2020 device is substantially equivalent to the predicate device that is currently marketed for the same intended use.

§ 870.1875 Stethoscope.

(a)
Manual stethoscope —(1)Identification. A manual stethoscope is a mechanical device used to project the sounds associated with the heart, arteries, and veins and other internal organs.(2)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.(b)
Electronic stethoscope —(1)Identification. An electronic stethoscope is an electrically amplified device used to project the sounds associated with the heart, arteries, and veins and other internal organs.(2)
Classification. Class II (special controls). The device, when it is a lung sound monitor, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.