K Number
K220451

Validate with FDA (Live)

Date Cleared
2023-10-05

(595 days)

Product Code
Regulation Number
862.3650
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Evidence MultiSTAT DOA Urine MultiPlex is intended for use with the Evidence MultiSTAT. The Evidence MultiSTAT is an analyzer intended for the qualitative determination of parent drug molecule and metabolites of drugs in human urine at the associated cutoffs.

The Evidence MultiSTAT DOA Urine MultiPlex detects the following drugs at the following cut-offs:

AnalyteAnalyte in Cut Off MaterialCut-Off
Benzodiazepines IOxazepam200 ng/mL
MethamphetamineS-(+)-Methamphetamine500 ng/mL
NoroxycodoneNoroxycodone100 ng/mL
Methadone(±)-Methadone300 ng/mL

The Evidence MultiSTAT DOA Urine MultiPlex provides only a preliminary analytical test result. A more specific alternative chemical method must be used to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry (GC/MS) and/or Liquid Chromatography/Tandem Mass Spectrometry (LC/MS/MS) are the preferred confirmatory methods. Other chemical confirmation methods are available. Clinical consideration and professional judgement should be applied to any drug of abuse test result, particularly when preliminary positive results are obtained.

For in vitro diagnostic use only.

Device Description

The Evidence MultiSTAT analyzer is a benchtop fully automated Biochip Array System. It performs simultaneous detection of multiple analytes from a single sample. The core technology is the Randox Biochip, a solid-state device containing an array of discrete test regions containing immobilised antibodies specific to different Drugs of Abuse (DOA) compound classes, A competitive chemiluminescent immunoassay is used for the DOA assays with the drug in the specimen and drug labelled with horseradish peroxidase (HRP) being in direct competition for the antibody binding sites. Increased levels of drug in a specimen will lead to reduced binding of drug labelled with HRP and thus a reduction in chemiluminescence being emitted. The light signal generated from each of the test regions on the biochip is detected by a Charge Coupled Device (CCD) camera in the Evidence MultiSTAT system which, together with the analyzer software, is used to quantify the light output and produce meaningful results.

The immunoassay processes are performed automatically in a self-contained and sealed biochip cartridge, which holds the biochips, the reagents, wash buffer and other fluids required for the test to be conducted.

Evidence MultiSTAT assays employ a qualitative reporting method. Each test sample is assayed against the provided Cut Off material of known concentration, which is used to determine the classification of the samples based on the comparison of the signal output.

The Evidence MultiSTAT System uses Randox Biochip Technology and performs simultaneous detection of multiple analytes from a single sample, using the Evidence MultiSTAT Analyzer. The assays are diagnostic tests for qualitative determination of the parent molecule and metabolites of drugs in human urine. The qualitative tests are based on a cut off value for each analyte, as detailed in the table below.

AI/ML Overview

The provided text describes a 510(k) premarket notification for the Randox Laboratories Limited, Evidence MultiSTAT DOA Urine Multiplex, and Evidence MultiSTAT device. This device is an in-vitro diagnostic (IVD) for the qualitative determination of drugs of abuse in human urine.

Crucially, the document does NOT describe the acceptance criteria or a study that proves the device meets those acceptance criteria in the context of an Artificial Intelligence/Machine Learning (AI/ML) powered medical device.

The document details the device's intended use, methodology (competitive immunoassay), analytes tested, and cut-off values. It mentions that "Studies were performed to evaluate performance with regards to the precision, specificity, and accuracy of the candidate devices" but does not provide any specific data, acceptance criteria values, or details of these studies.

Therefore, I cannot provide the requested information regarding acceptance criteria and a study proving the device meets them, specifically for an AI/ML context, because the provided text is for an IVD immunoassay device, not an AI/ML-powered medical device.

To answer your request based on the provided text, I can only state what is missing from the document related to AI/ML acceptance criteria and studies:

  • No AI/ML Component: The device described is an immunoassay system (Evidence MultiSTAT DOA Urine Multiplex and Evidence MultiSTAT analyzer), which works based on chemical reactions and light detection, not artificial intelligence or machine learning.
  • No Acceptance Criteria or Performance Data Provided: The document states that "Studies were performed to evaluate performance with regards to the precision, specificity, and accuracy of the candidate devices," but it does not provide any of the acceptance criteria themselves, nor the results of these studies. Therefore, no table of acceptance criteria vs. reported performance can be generated from this text.
  • No Test Set Details: Since no performance study details are given, there is no information on:
    • Sample size for the test set.
    • Data provenance (country, retrospective/prospective).
    • Number or qualifications of experts.
    • Adjudication method.
    • Multi-reader multi-case (MRMC) comparative effectiveness study.
    • Standalone (algorithm-only) performance.
    • Type of ground truth used.
  • No Training Set Details: As this is not an AI/ML device, there is no mention of a training set or how ground truth for a training set would be established.

In summary, the provided FDA 510(k) clearance letter and summary pertain to a traditional in-vitro diagnostic immunoassay system and do not contain any information relevant to the acceptance criteria or performance studies of an AI/ML-powered medical device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Randox Laboratories Limited Karena Shaw Regulatory Affairs Manager 55 Diamond Road Crumlin, Co. Antrim BT29 4Qy United Kingdom

Re: K220451

Trade/Device Name: Evidence MultiSTAT DOA Urine Multiplex. Evidence MultiSTAT Regulation Number: 21 CFR 862.3650 Regulation Name: Opiate Test System Regulatory Class: Class II Product Code: DJG, DJC, JXM, DJR, JJE Dated: March 7, 2023 Received: April 14, 2023

Dear Karena Shaw:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Joseph A. Digitally signed by
Kotarek -S Joseph A. Kotarek -S
Date: 2023.10.05
14:19:18 -04'00'

Josepk Kotarek, Ph.D. Toxicology Branch Chief Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K220451

Device Name

Evidence MultiSTAT Urine DOA MultiPlex, Evidence MultiSTAT

Indications for Use (Describe)

The Evidence MultiSTAT DOA Urine MultiPlex is intended for use with the Evidence MultiSTAT. The Evidence MultiSTAT is an analyzer intended for the qualitative determination of parent drug molecule and metabolites of drugs in human urine at the associated cutoffs.

The Evidence MultiSTAT DOA Urine MultiPlex detects the following drugs at the following cut-offs:

AnalyteAnalyte in Cut Off MaterialCut-Off
Benzodiazepines IOxazepam200 ng/mL
MethamphetamineS-(+)-Methamphetamine500 ng/mL
NoroxycodoneNoroxycodone100 ng/mL
Methadone(±)-Methadone300 ng/mL

The Evidence MultiSTAT DOA Urine MultiPlex provides only a preliminary analytical test result. A more specific alternative chemical method must be used to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry (GC/MS) and/or Liquid Chromatography/Tandem Mass Spectrometry (LC/MS/MS) are the preferred confirmatory methods. Other chemical confirmation methods are available. Clinical consideration and professional judgement should be applied to any drug of abuse test result, particularly when preliminary positive results are obtained.

For in vitro diagnostic use only.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect

of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

1. Substantial Equivalence as required by 21 CFR 807.92 Statement

This summary of the 510(k) substantial equivalence information is being submitted in accordance with the requirement of 21 CFR 807.92.

2. 510(k) Number Owner's Name and Address

Name:Randox Laboratories Limited
Address:55 Diamond Road, Crumlin, County Antrim, BT29 4QY,United Kingdom
Telephone:+44 2894422413
Contact Person:Karena Shaw
E-mail:Karena.Shaw@randox.com

Date of Summary Preparation: 04 October 2023

3. Device Proprietary Names, Common Names, Purpose for Submission, Classification Name, Panel, Product Code and 21 CFR Number

510(k) Number:K220451
Device Proprietary Names:Evidence MultiSTAT Urine DOA MultiPlex,Evidence MultiSTAT
Common Names:Analyzer, Chemistry (Photometric, Discrete), For ClinicalUse
Classification Names:

Classification Names:

The Candidate Device Test System Regulatory Classification is Class II; the Classification Panel is Clinical Chemistry (75) and Clinical Toxicology (91). Regulatory information for the test system is detailed in table 1.

Table 1

Product CodeClassification NameClassificationRegulation SectionPanel
JJEAnalyzer, Chemistry(Photometric,Discrete), ForClinical UseI21 CFR 862.2160ClinicalChemistry(75)
DJCMethamphetamineTest SystemII21 CFR 862.3610ClinicalToxicology(91)
JXMBenzodiazepine TestSystem(Oxazepam)II21 CFR 862.3170ClinicalToxicology(91)
DJRMethadone TestSystem(Methadone)II21 CFR 862.3620ClinicalToxicology(91)
DJGOpiate Test System(Noroxycodone)II21 CFR 862.3650ClinicalToxicology(91)

{4}------------------------------------------------

4. Predicate Device Proprietary Names and 510(k) Numbers

The predicate device for the Evidence MultiSTAT System is the MEDTOXScan® Reader (K091454) and is detailed in table 2.

ItemCandidate Device(K220451)Evidence MultiSTATUrine DOA MultiPlex,Evidence MultiSTATPredicate Device(K091454)MEDTOXScan Reader
Indications for UseQualitative detection ofdrugs of abuse in urineSame
Intended Use SettingPrescription Use OnlySame
MethodologyQualitative,CompetitiveImmunoassayQualitativeimmunochromatographic
AnalytesMethamphetamine,NoroxycodoneSame
Cutoffs (ng/mL)Methamphetamine (500)Noroxycodone (100)Same

Table 2: Similarities and Differences to Predicate Device

4.1 The Evidence MultiSTAT analyzer is a benchtop fully automated Biochip Array System. It performs simultaneous detection of multiple analytes from a single sample. The core technology is the Randox Biochip, a solid-state device containing an array of discrete test regions containing immobilised antibodies specific to different Drugs of Abuse (DOA) compound classes, A competitive chemiluminescent immunoassay is used for the DOA assays with the drug in the specimen and drug labelled with horseradish peroxidase (HRP) being in direct competition for the antibody binding sites. Increased levels of drug in a specimen will lead to reduced binding of drug labelled with HRP and thus a reduction in chemiluminescence being emitted. The light signal generated from each of the test regions on the biochip is detected by a Charge Coupled Device (CCD) camera in the Evidence MultiSTAT system which, together with the analyzer software, is used to quantify the light output and produce meaningful results.

The immunoassay processes are performed automatically in a self-contained and sealed biochip cartridge, which holds the biochips, the reagents, wash buffer and other fluids required for the test to be conducted.

Evidence MultiSTAT assays employ a qualitative reporting method. Each test sample is assayed against the provided Cut Off material of known concentration, which is used to determine the classification of the samples based on the comparison of the signal output.

4.2 The Evidence MultiSTAT System uses Randox Biochip Technology and performs simultaneous detection of multiple analytes from a single sample, using the Evidence MultiSTAT Analyzer. The assays are diagnostic tests for qualitative determination of the parent molecule and metabolites of drugs in human urine. The qualitative tests are based on a cut off value for each analyte, as detailed in the table below.

{5}------------------------------------------------

Table 3: Analytes, Associated Analyte in the Cut Off Material and Cut Off concentration employed in the Evidence MultiSTAT DOA Urine MultiPlex

AnalyteAnalyte in Cut Off MaterialCut Off
MethamphetamineS-(+)-Methamphetamine500 ng/ml
NoroxycodoneNoroxycodone100 ng/ml
Benzodiazepines 1Oxazepam200 ng/ml
Methadone(+)-Methadone300 ng/ml

The Evidence MultiSTAT DOA Urine MultiPlex (EV4393) will be supplied as a test kit comprising:

  • 12 x Urine Test Cartridges ।
  • 6 x 1 ml Urine Cut Off Material (lyophilized) l
  • 4 x 1 ml Urine Positive Control Material (lyophilized) ー
  • 2 x 10 ml Reconstitution Buffer -
  • 1 x USB which contains batch specific update and Instructions for Use (IFU) l
  • 1 x Batch Barcodes ।

Each kit is supplied with the Evidence MultiSTAT Accessory kit (EV4116) which contains:

  • 12 x MultiSTAT Tip Cartridges —
    • 1 x Tip/Waste Cartridge o
    • 6 x 1000 ul Pipette Tip O
    • 1 x Liquid Absorber O

Reagent Composition

MultiSTAT DOA Urine MultiPlex Assay Diluent

20 mM phosphate buffer, pH 7.0 containing protein, detergents, and preservatives. This is contained within the cartridge.

MultiSTAT DOA Urine MultiPlex Conjugate

20 mM Tris based buffer, pH 7.0 containing protein, preservatives, and horseradish peroxidase - labelled drug derivatives. This is contained within the cartridge.

MultiSTAT DOA Urine MultiPlex Biochip

Solid substrate containing immobilized antibody discrete test regions. This is contained within the cartridge.

MultiSTAT DOA Urine MultiPlex Wash Buffer

20 mM Tris buffered saline, pH 7.4, containing surfactant and preservatives. This is contained within the cartridge.

LUM-EV934/PX

Luminol-EV934 and Peroxide are contained within the cartridge and are mixed in a ratio of 1:1 by the analyser to give the working signal reagent

MultiSTAT DOA Urine MultiPlex Cut Off

Lyophilised, 20 mM phosphate buffer, pH 7.2 containing stabilizers, preservatives and drug concentrations at the assay cut off values (detailed in Table 5 above).

MultiSTAT DOA Urine MultiPlex Positive Control

Lyophilised, 20 mM phosphate buffer, pH 7.2 containing stabilizers, preservatives, and drug concentrations.

{6}------------------------------------------------

MultiSTAT Reconstitution Buffer

A solution at a neutral pH containing preservatives.

5. Intended Use

The Evidence MultiSTAT DOA Urine MultiPlex is intended for use with the Evidence MultiSTAT. The Evidence MultiSTAT is an analyzer intended for the qualitative determination of parent drug molecule and metabolites of drugs in human urine at the associated cutoffs.

The Evidence MultiSTAT DOA Urine MultiPlex detects the following drugs at the following cutoffs:

AnalyteAnalyte in Cut Off MaterialCut-Off
Benzodiazepines IOxazepam200 ng/mL
MethamphetamineS-(+)-Methamphetamine500 ng/mL
NoroxycodoneNoroxycodone100 ng/mL
Methadone(±)-Methadone300 ng/mL

The Evidence MultiSTAT DOA Urine MultiPlex provides only a preliminary analytical test result. A more specific alternative chemical method must be used to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry (GC/MS) and/or Liquid Chromatography/Tandem Mass Spectrometry (LC/MS/MS) are the preferred confirmatory methods. Other chemical confirmation methods are available. Clinical consideration and professional judgement should be applied to any drug of abuse test result, particularly when preliminary positive results are obtained.

For in vitro diagnostic use only.

6. Performance

Studies were performed to evaluate performance with regards to the precision, specificity, and accuracy of the candidate devices.

7. Conclusions

  • Based on the information provided in this 510(k), the Evidence MultiSTAT Urine ● DOA MultiPlex, Evidence MultiSTAT, is substantially equivalent to the predicate devices and raise no new issues of safety and effectiveness.

§ 862.3650 Opiate test system.

(a)
Identification. An opiate test system is a device intended to measure any of the addictive narcotic pain-relieving opiate drugs in blood, serum, urine, gastric contents, and saliva. An opiate is any natural or synthetic drug that has morphine-like pharmocological actions. The opiates include drugs such as morphine, morphine glucoronide, heroin, codeine, nalorphine, and meperedine. Measurements obtained by this device are used in the diagnosis and treatment of opiate use or overdose and in monitoring the levels of opiate administration to ensure appropriate therapy.(b)
Classification. Class II (special controls). An opiate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).