Intended Use

The STERLINK mini sterilizer with STERLOAD™ mini cassette is a vaporized hydrogen peroxide sterilizer intended for use in the terminal sterilization of cleaned, rinsed, and dried reusable metal and nonmetal medical devices used in healthcare facilities. A pre-programmed sterilization lumen cycle operates at low temperature and is thus suitable for processing medical devices sensitive to heat and moisture.

The STERLINK mini can sterilize *:

  • · Instruments with diffusion-restricted spaces such as the hinged portion of forceps and scissors
  • · Medical devices with a single stainless-steel lumen with:
  • o An inside diameter of 1.6 mm or larger and a length of 200 mm or shorter

*The validation testing was conducted using a maximum of five (5) lumens per load. Hospital loads should not exceed the maximum number of lumens validated by this testing. The validation studies were performed using a validation load with a total weight of 1.54 lbs.

Tyvek® Roll with CI for STERLINK™ Sterilizer, when used in CHAMBER mode, is intended to provide health care workers with an effective method to enclose devices intended for sterilization in the STERLINK™ sterilizer. The device is intended to allow sterilization of enclosed devices and also to maintain sterility of the enclosed devices until used up to 1 month post sterilization.

The materials compatible for use in the Tyvek® Roll with CI for STERLINK™ Sterilizer, when used in Chamber mode, are as follows: Aluminum 5052, Aluminum 6061, Stainless Steel 316L, Titanium, Acrylonitrile Butadiene Styrene (ABS), High Density Polyethylene (HDPE), Polypropylene (PP), Polytetrafluoroethylene (PTFE) and Silicone (Hardness 50).

The maximum load weight that can be placed in the Tyvek® roll is:

  • · 3.97 pounds (1.8 kg) for CHAMBER mode of FPS-15s Plus sterilizer
  • · 1.54 pounds (0.7 kg) for CHAMBER mode of STERLINK mini sterilizer

The roll is printed with a chemical indicator bar which is a process indicator (ISO 11140 1:2005) that changes from red to blue (or lighter) when exposed to hydrogen peroxide vapor during processing in the STERLINK™ sterilizer.

The Tyvek® Roll with CI for STERLINK™ Sterilizer is offered in the follow 1 type: · Sterilization roll, Flat

Terragene Bionova® SCBI (BT96) is a self-contained biological indicator inoculated with viable 106 Geobacillus stearothermophilus bacterial spores and is intended for monitoring the efficacy of vaporized hydrogen peroxide sterilization processes. BT96 has Super Rapid readout at 30 minutes at 60°C.

Terragene Bionova® Reader Incubators (IC10/20FRLCD, Mini-Bio) incubate at 60°C and 37°C and read the Terragene Bionova® SCBI for fluorescent results at the times prescribed in the User Manuals.

Terragene Chemdye® (CD42) is a chemical process indicator intended for monitoring the efficacy of vaporized hydrogen peroxide sterilization processes. The chemical indicator changes from red to yellow for CD42 to indicate that the conditions of the cycle have been met.

Terragene Cintape® (CT40) is a chemical process indicator tape intended for monitoring the efficacy of Vaporized Hydrogen Peroxide sterilization processes. The indicating tape changes from purple to green when exposure to vaporized hydrogen peroxide.

The self-contained biological indicator and chemical processing indicators are intended for use with:

  • · CHAMBER mode of FPS-15s Plus sterilizer
  • · CHAMBER mode of STERLINK mini sterilizer
Device Description

The STERLINK mini sterilizer with STERLOAD™ mini cassette is a low temperature sterilizer which uses the STERLINK™ process to inactivate microorganisms on a broad range of medical devices and instruments. This system consists of a main device connected pump module and cassette which are called the STERLINK mini and STERLOAD™ mini cassette, respectively. The STERLOAD™ mini cassette contains 58-59.5% (weight concentration) of hydrogen peroxide (H2O2) which is utilized as the sterilant.

Tyvek® Roll with CI for STERLINK™ Sterilizer is intended to contain medical devices to be terminally sterilized in the STERLINK™ sterilization system. The medical devices are inserted into the roll, sealed, and then sterilized in the STERLINK™ sterilization system. After completion of the sterilization process, the roll maintains sterility of the enclosed medical devices until the seal is opened. The device is intended to allow sterilization of enclosed devices and also to maintain sterility of the enclosed medical devices for up to 1 month post sterilization. The roll is printed with a chemical indicator bar that changes from red to blue (or lighter) when exposed to hydrogen peroxide vapor during processing in the STERLINKTM sterilizer. The Tyvek® Roll with CI for STERLINKIM Sterilizer is offered in one type as a Flat Sterilization roll. The sterilization roll is made from a Tyvek® sheet and a clear plastic film that are heat sealed together on opposite two sides. After being cut into a suitable length, the product to be sterilized is placed inside and the two open ends are heat sealed. The process indicator printed on the Tyvek® will exhibit a color change after the roll is exposed to hydrogen peroxide (H2O2).

Terragene® Bionova® BT96 Fluorescence Super Rapid Readout Biological Indicators are singleuse Self-Contained Biological Indicators (SCBIs) that consist of a polypropylene tube, a spore carrier, and a glass ampoule with a culture medium, enclosed with a colored cap. Each tube contains a population of Geobacillus stearothermophilus ATCC 7953 spores inoculated on a spore carrier, a plastic cap with holes and a barrier permeable to Plasma or Vaporized Hydrogen Peroxide. Each BT96 has a Process Indicator on label that changes from purple to green when exposed to hydrogen peroxide. The Bionova® BT96 Biological Indicators have been designed for monitoring of Vaporized Hydrogen Peroxide sterilization processes when used in conjunction with Bionova® IC10/20FRLCD or MiniBio Auto-Readers Incubators.

Chemdye® CD42 Process Indicators (Type 1 according to ISO 11140-1:2014 standard) are singleuse chemical indicators that consist of plastic strips printed with indicator ink. These indicators have been designed to monitor Plasma or Vaporized Hydrogen Peroxide sterilization processes within loads, ensuring an adequate exposure to the sterilizing agent during the sterilization process and allowing to distinguish between processed and unprocessed items.

Cintape® CT40 Process Indicators (Type 1 according to ISO 11140-1:2014 standard) are singleuse chemical indicators that consist of a roll of self-adhesive plastic tape printed with indicator ink. These indicators have been designed to monitor Plasma or Vaporized Hydrogen Peroxide sterilization processes, ensuring an adequate exposure to the sterilizing agent during the sterilization process and allowing to distinguish between processed and unprocessed items. The adhesive component of the tape allows the adhesion to different types of packaging and wraps, such as cloth, paper and plastic.

AI/ML Overview

The provided text describes acceptance criteria and study results for medical sterilization devices (STERLINK mini Sterilizer with STERLOAD mini Cassette, Tyvek Roll with CI for STERLINK Sterilizer, and Sterilization Process Indicator for STERLINK Sterilizer).

Here's a breakdown of the requested information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Device: STERLINK mini Sterilizer with STERLOAD™ mini Cassette

TestPurposeAcceptable CriteriaReported Device Performance
Human factors and usability engineeringTo confirm whether the user error affects the risk to user safety and performance of the device when the device is used according to the intended use in accordance with IEC 62366-1.User safety risk and device performance degradation due to user errors should not occur.Pass
BiocompatibilityTo confirm the biological safety of the sterilized load in accordance with ISO 10993-5.≤ Grade 2Pass
Software validationTo verify that the software is compatible for the intended use of the device in accordance with IEC 62304.The risk of all identified software hazards should be reduced to an acceptable risk level and the device is appropriate for its intended use.Pass
Electrical safetyTo confirm the electrical safety of the device in accordance with IEC 60601-1, IEC 61010-1, and IEC 61010-2-040.The device conforms to the relevant standards.Pass
Electromagnetic compatibility (EMC)To confirm the electromagnetic wave safety of the device in accordance with IEC 60601-1-2 and EN 55011.The device conforms to the relevant standards.Pass
Resistance validation for biological indicator testTo validate the D-value and total kill endpoint of the biological indicators in accordance with ISO 11138-1.The total kill endpoint time should be shorter than the sterilization half cycle.Pass
Lumen sterilizationTo confirm the lumen sterilization performance in accordance with ISO 14937.The SAL (Sterility Assurance Level) should be at least 10^-6^.Pass
Surface sterilizationTo confirm the surface sterilization performance of various materials in accordance with ISO 14937, ISO 11737-1, and ISO11737-2.The SAL (Sterility Assurance Level) should be at least 10^-6^.Pass
Mated surface sterilizationTo confirm the mated surface sterilization performance of various materials in accordance with ISO 14937, ISO 11737-1, and ISO11737-2.The SAL (Sterility Assurance Level) should be at least 10^-6^.Pass
Simulated use testTo demonstrate the efficacy of the sterilization validation on medical devices contaminated by organic and inorganic matters with bacterial spores in accordance with ASTM E1837-96 and ISO 11737-1.The microorganisms should not survive after sterilization.Pass
In-use testTo demonstrate the sterilization validation of the used and clinically pretreated medical devices in accordance with ASTM E1837-96.The microorganisms should not survive after sterilization.Pass
Sporicidal activity testTo confirm the sporicidal activity of the device in accordance with AOAC Official Method 966.04.All spores inoculated on the carriers must be sterilized.Pass
Bacteriostasis testTo confirm the bacteriostasis action of sterilized various materials after sterilization process in accordance with ISO 11737-1.Sterilized materials should not exhibit the bacteriostatic effects.Pass
Material compatibility testTo demonstrate the compatibility of various materials in the sterilization cycle in accordance with ASTM D638, ASTM E8/E8M, ASTM D790, ASTM E290-14, ASTM D256, ASTM E23-18, ASTM E1164, ASTM D3985, and ASTM E1240.The physical properties of the material after sterilization should be similar to those before sterilization.Pass
Delivery validationTo validate the packaging and performance of the device after exposure to simulated shipping conditions in accordance with ASTM D4169-14.After the distribution cycle of the drop test, significant deformations and defects in the electrical and mechanical performance of the device should not be observed.Pass
Hydrogen peroxide gas detection testTo measure the concentration of the hydrogen peroxide emitted from the device in accordance with Occupational Safety and Health Administration (OSHA, TWA).< 1 ppm hydrogen peroxide gas detectionPass
Sterilant preservation testTo verify the shelf life of the sterilant cassette in accordance with the internal test standard.The sterilant should meet the specified weight, concentration, and pH standards during the shelf life.Pass

Device: Tyvek® Roll with CI for STERLINK™ Sterilizer

Performance TestPurposeAcceptable CriteriaReported Device Performance
Sterilization efficacy testTo ensure that the device is suitable for the sterilization process in accordance with ISO 14937.The SAL should be at least 10^-6^.Pass
CI of sterilization roll validationTo confirm that the shelf life before and after the sterilization of device, and to ensure CI of the device is compatible for the sterilization process in accordance with ANSI/AAMI/ISO 11140-1.- CI should not change completely with temperature, and sterilant exposure time of less than 7 seconds, during the shelf life.- CI should maintain the changed color after real-time aging during shelf life after sterilization.Pass
Internal pressurization testTo verify the internal pressurization of the sterilized device at the shelf life in accordance with ASTM F1980-16 and ASTM F1140/F1140M-13.- Burst pressure: $ \ge $ 1kPa- Creep time: > 30 sec in creep pressure (> 80% of burst pressure)Pass
Visual inspection testTo confirm the sterilized device integrity was maintained at shelf life in accordance with ASTM F1980-16 and ASTM F1886/F1886M-16.There should be no defects in the sealing area.Pass
Dye penetration testTo confirm the packaging of the sterilized device was maintained at shelf life in accordance with ASTM F1980-16 and ASTM F1929-15.There should be no dye leakage.Pass
Tensile strength of Tyvek®To confirm the physical properties of the sterilized device at shelf life in accordance with ASTM F1980-16, ASTM D5035-11, ASTM D882, ASTM F88, and ASTM D1922-20.The physical properties must meet the criteria specified by the manufacturer.Pass
Tensile strength of plastic film(Same purpose as above)(Same criteria as above)Pass
Seal strength(Same purpose as above)(Same criteria as above)Pass
Tear resistance(Same purpose as above)(Same criteria as above)Pass
Microbial Barrier TestTo confirm the microbial barrier performance of sterilized devices at shelf life in accordance with ASTM F1980-16 and DIN 58953-6.- Test with liquid inoculum: No microbial growth should appear.- Test with gas inoculum: Only a maximum 15 colonies should grow out of a total of 10 bacterial permeability testers. The number of colonies in each bacterial permeability tester should not be greater than five.Pass
Residual sterilant on Tyvek® validationTo confirm the time required for the residual sterilant to be reduced to a safe amount after sterilization in accordance with the internal test standard.≤ 0.05 ppmPass
Biocompatibility testTo confirm the biological safety after sterilization in accordance with ISO 10993-5.≤ Grade 2Pass

Device: Sterilization Process Indicator for STERLINK™ Sterilizer (BT96, CD42, CT40)

ItemTestPurposeAcceptable CriteriaReported Device Performance
Self-Contained Biological Indicator (BT96)Resistance validation for biological indicator testTo validate the D-value and total kill endpoint of biological indicators in accordance with ISO 11138-1.The total kill endpoint time should be shorter than the half cycle sterilant exposure time.Pass
BI & Test pack validation testTo verify the suitability of the test pack to be used as monitoring of the sterilization process in accordance with the internal test standard.The test pack with SCBI should fail to sterilize in half cycle and succeed in full cycle.Pass
Chemical Indicator (CD42, CT40)Performance characteristics obtained from resistometerTo evaluate the correct performance of chemical indicators, by testing them at different exposure conditions in accordance with ANSI/AAMI/ISO 11140-1.The CI should change color correctly only when exposed to the temperature and sterilant concentration conditions specified by the manufacturer.Pass
BiocompatibilityTo demonstrate that the CI does not offset or transfer during and after the sterilization process in accordance with ANSI/AAMI/ISO 11140-1.The CI does not release any substance or bleed when the sterilization process.Pass
Endpoint stabilityTo demonstrate the stability of the endpoint reaction of CI at the end of their shelf life and specified period after sterilization in accordance with ANSI/AAMI/ISO 11140-1.The endpoint reaction of CI should be maintained until the shelf life and specified period after sterilization.Pass
Shelf life studyTo verify the shelf life of the CI in accordance with ANSI/AAMI/ISO 11140-1.The endpoint reaction of the real time aged CI should be maintained until shelf life.Pass
Chemical indicator validationTo confirm the compatibility of the CI for the sterilization process in accordance with the internal test standard.The minimum time required for color change of CI within the sterilization process time should be confirmed.Pass

Additional Study Information:

The provided text describes non-clinical tests performed to demonstrate substantial equivalence to predicate devices, not clinical studies involving human patients or complex AI algorithms requiring ground truth from experts. The devices are sterilizers and sterilization indicators, which are typically evaluated through microbiological and engineering performance testing.

Therefore, many of the specific questions about "test set," "ground truth," "experts," "adjudication," and "AI assistance" are not directly applicable to the type of device and testing described in this document.

For completeness, here's what can be extracted or inferred:

  1. Sample size used for the test set and the data provenance:

    • Sample size: Not explicitly stated as a single "test set" size. The text mentions "maximum of five (5) lumens per load" for lumen sterilization validation and "a validation load with a total weight of 1.54 lbs." for the sterilizer. For the Tyvek roll microbial barrier test, it mentions "a total of 10 bacterial permeability testers." Specific quantities for other tests are not provided.
    • Data provenance: Not specified. These are non-clinical engineering and microbiological tests, likely conducted in a laboratory setting by the manufacturer (Plasmapp Co., Ltd. and Terragene S.A.). The country of origin of the manufacturer is South Korea for Plasmapp and presumably Argentina for Terragene (as it's a known Argentinian company). The studies are "non-clinical tests," implying they were conducted in a controlled environment, not on human subjects.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience): Not applicable. The tests are laboratory-based and use established microbiological and engineering standards (e.g., ISO, ASTM, AOAC). "Ground truth" is determined by scientific measurements and observation of bacterial kill, physical properties, or chemical indicator changes, not by expert consensus on subjective interpretations.

  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable. Adjudication methods like 2+1 or 3+1 are typically used for subjective assessments (e.g., radiological reads) where multiple experts might disagree. These tests are objective measurements against defined standards.

  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. These devices do not involve "human readers" interpreting "cases" with or without "AI assistance."

  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. There is no AI algorithm component to these devices described in the document.

  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): The "ground truth" for these sterilization and chemical indicator tests is based on:

    • Microbiological kill: Demonstrating the inactivation of a specified population of resistant spores (e.g., Geobacillus stearothermophilus) to a defined Sterility Assurance Level (SAL), typically 10^-6^. This is confirmed through laboratory culture methods.
    • Physical and chemical property measurements: Adherence to established engineering standards (e.g., tensile strength, burst pressure, color change properties of chemical indicators).
    • Safety standards: Compliance with international safety and electromagnetic compatibility standards.
  7. The sample size for the training set: Not applicable. These devices do not involve machine learning or AI models that require a "training set."

  8. How the ground truth for the training set was established: Not applicable, as there is no training set mentioned.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square are the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

April 14, 2023

Plasmapp Co., Ltd. % Candace Cederman Consultant CardioMed Device Consultants LLC 1783 Forest Drive, Suite 254 Annapolis, Maryland 21401

Re: K220345

Trade/Device Name: STERLINK mini Sterilizer with STERLOAD mini Cassette, Tyvek Roll with CI for STERLINK Sterilizer, Sterilization Process Indicator for STERLINK Sterilizer; Terragene Bionova SCBI (BT96), Terragene Bionova Reader Incubators (IC10/20FRLCD, Mini-Bio), Terragene Chemdye (CD42), Terragene Cintape (CT40) Regulation Number: 21 CFR 880.6860 Regulation Name: Ethylene Oxide Gas Sterilizer Regulatory Class: Class II Product Code: MLR, FRG, JOJ, FRC Dated: February 25, 2022 Received: February 28, 2022

Dear Candace Cederman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of

{1}------------------------------------------------

Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely. Christopher K. Duqard -2

for Clarence W. Murray, III, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K220345

Device Name

STERLINK mini Sterilizer with STERLOAD™ mini Cassette

Indications for Use (Describe)

The STERLINK mini sterilizer with STERLOAD™ mini cassette is a vaporized hydrogen peroxide sterilizer intended for use in the terminal sterilization of cleaned, rinsed, and dried reusable metal and nonmetal medical devices used in healthcare facilities. A pre-programmed sterilization lumen cycle operates at low temperature and is thus suitable for processing medical devices sensitive to heat and moisture.

The STERLINK mini can sterilize *:

  • · Instruments with diffusion-restricted spaces such as the hinged portion of forceps and scissors
  • · Medical devices with a single stainless-steel lumen with:
  • o An inside diameter of 1.6 mm or larger and a length of 200 mm or shorter

*The validation testing was conducted using a maximum of five (5) lumens per load. Hospital loads should not exceed the maximum number of lumens validated by this testing. The validation studies were performed using a validation load with a total weight of 1.54 lbs.

Type of Use (Select one or both, as applicable)
----------------------------------------------------------------------------------------------------------------------------------------------------------

| | Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Indications for Use

510(k) Number (if known)

K220345

Device Name Tyvek® Roll with CI for STERLINK™ Sterilizer

Indications for Use (Describe)

Tyvek® Roll with CI for STERLINK™ Sterilizer, when used in CHAMBER mode, is intended to provide health care workers with an effective method to enclose devices intended for sterilization in the STERLINK™ sterilizer. The device is intended to allow sterilization of enclosed devices and also to maintain sterility of the enclosed devices until used up to 1 month post sterilization.

The materials compatible for use in the Tyvek® Roll with CI for STERLINK™ Sterilizer, when used in Chamber mode, are as follows: Aluminum 5052, Aluminum 6061, Stainless Steel 316L, Titanium, Acrylonitrile Butadiene Styrene (ABS), High Density Polyethylene (HDPE), Polypropylene (PP), Polytetrafluoroethylene (PTFE) and Silicone (Hardness 50).

The maximum load weight that can be placed in the Tyvek® roll is:

  • · 3.97 pounds (1.8 kg) for CHAMBER mode of FPS-15s Plus sterilizer
  • · 1.54 pounds (0.7 kg) for CHAMBER mode of STERLINK mini sterilizer

The roll is printed with a chemical indicator bar which is a process indicator (ISO 11140 1:2005) that changes from red to blue (or lighter) when exposed to hydrogen peroxide vapor during processing in the STERLINK™ sterilizer.

The Tyvek® Roll with CI for STERLINK™ Sterilizer is offered in the follow 1 type: · Sterilization roll, Flat

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Indications for Use

510(k) Number (if known) K220345

Device Name

Sterilization Process Indicator for STERLINK™ Sterilizer

Terragene Bionova® SCBI (BT96), Terragene Bionova® Reader Incubators (IC10/20FRLCD, Mini-Bio), Terragene Chemdye® (CD42), Terragene Cintape® (CT40)

Indications for Use (Describe)

Terragene Bionova® SCBI (BT96) is a self-contained biological indicator inoculated with viable 10% Geobacillus stearothermophilus bacterial spores and is intended for monitoring the efficacy of vaporized hydrogen peroxide sterilization processes. BT96 has Super Rapid readout at 30 minutes at 60°C.

Terragene Bionova® Reader Incubators (IC10/20FRLCD, Mini-Bio) incubate at 60°C and 37°C and read the Terragene Bionova® SCBI for fluorescent results at the times prescribed in the User Manuals.

Terragene Chemdye® (CD42) is a chemical process indicator intended for monitoring the efficacy of vaporized hydrogen peroxide sterilization processes. The chemical indicator changes from red to yellow for CD42 to indicate that the conditions of the cycle have been met.

Terragene Cintape® (CT40) is a chemical process indicator tape intended for monitoring the efficacy of Vaporized Hydrogen Peroxide sterilization processes. The indicating tape changes from purple to green when exposure to vaporized hydrogen peroxide.

The self-contained biological indicator and chemical processing indicators are intended for use with:

  • · CHAMBER mode of FPS-15s Plus sterilizer
  • · CHAMBER mode of STERLINK mini sterilizer
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{5}------------------------------------------------

Image /page/5/Picture/2 description: The image shows the word "plasmapp" in a teal color. There is a teal asterisk to the left of the word. The font is sans-serif and the letters are lowercase.

510(k) Summary [as required by 21 CFR 807.92(c)]

STERLINK mini Sterilizer with STERLOAD™ mini Cassette Tyvek® Roll with CI for STERLINK™ Sterilizer Sterilization Process Indicator for STERLINK™ Sterilizer

510(k) K220345

General Information

Applicant/Submitter:Plasmapp Co., Ltd.BVC-111, 125, Gwahak-ro,Yuseong-gu, Daejeon, 34141,Rep. of Korea (South Korea)Tel: +82 42 716 2115
Contact Person:Candace Cederman
Address:CardioMed Device Consultants LLC1783 Forest Drive, Suite 254Annapolis, MD 21401Tel: +1 410 674 2060
Preparation Date:April 13, 2023

Device Name and Code

Device Trade Name:A. STERLINK mini Sterilizer with STERLOAD™ mini CassetteB. Tyvek® Roll with CI for STERLINK™ SterilizerC. Sterilization Process Indicator for STERLINK™ Sterilizer;Terragene Bionova® SCBI (BT96); Terragene Bionova®Reader Incubators (IC10/20FRLCD, Mini-Bio); TerrageneChemdye® (CD42); Terragene Cintape® (CT40)
Common Name:A. Vapor Phase Hydrogen Peroxide Sterilization SystemB. Tyvek® Roll for VH2O2 SterilizerC. Self-contained Biological Indicator, Self-Contained BiologicalIndicator Incubator, Chemical Indicator.

{6}------------------------------------------------

Classification Name:A. Ethylene Oxide Gas SterilizerB. Sterilization WrapC. Sterilization Process Indicator
Product Code:A. MLRB. FRG, JOJC. FRC, JOJ
Regulation Number:A. 21 CFR 880.6860B. 21 CFR 880.6850, 21 CFR 880.2800C. 21 CFR 880.2800
Classification:Class II
Review Panel:General Hospital

STERLINK mini Sterilizer with STERLOAD™ mini Cassette A.

Device Description A.1

The STERLINK mini sterilizer with STERLOAD™ mini cassette is a low temperature sterilizer which uses the STERLINK™ process to inactivate microorganisms on a broad range of medical devices and instruments. This sterilizer offers an effective, safe, fast, economical, easy to use, reliable, and flexible sterilization method.

This system consists of a main device connected pump module and cassette which are called the STERLINK mini and STERLOAD™ mini cassette, respectively. The STERLOAD™ mini cassette contains 58-59.5% (weight concentration) of hydrogen peroxide (H2O2) which is utilized as the sterilant.

A.2 Indications / Intended Use

The subject and predicate device have the same intended use. The specific indications for use differ only in the size and weight of the medical devices that can be sterilized in each chamber.

The STERLINK mini sterilizer with STERLOAD™ mini cassette is a vaporized hydrogen peroxide sterilizer intended for use in the terminal sterilization of cleaned, rinsed, and dried reusable metal and nonmetal medical devices used in healthcare facilities. A pre-programmed

{7}------------------------------------------------

sterilization cycle operates at low pressure and low temperature and is thus suitable for processing medical devices sensitive to heat and moisture.

  • I The STERLINK mini can sterilize*:
    • Instruments with diffusion-restricted spaces such as the hinged portion of forceps and scissors
    • · Medical devices with a single stainless-steel lumen with:

o An inside diameter of 1.6 mm or larger and a length of 200 mm or shorter

*The validation testing for all lumen sizes was conducted using a maximum of five (5) lumens per load. Hospital loads should not exceed the maximum number of lumens validated by this testing. The validation studies were performed using a validation load with a total weight of 1.54 lbs.

A.3 Technical Characteristics in Comparison to Predicate Devices

A summary of the technical characteristics of the subject device and predicate device can be found in the table below.

{8}------------------------------------------------

Subject DevicePredicate DeviceComparison
Plasmapp Co., Ltd.Plasmapp Co., Ltd.Same
Device NameSTERLINK miniFPS-15s Plus-
510(k) NumberK212200-
DeviceClassificationNameEthylene oxide gas sterilizerEthylene oxide gas sterilizerSame
ClassificationProduct CodeMLRMLRSame
RegulationNumber21 CFR 880.686021 CFR 880.6860Same
Intended UseThe STERLINK mini sterilizer with STERLOADTMmini cassette is a vaporized hydrogen peroxidesterilizer intended for use in the terminal sterilizationof cleaned, rinsed, and dried reusable metal andnonmetal medical devices used in healthcarefacilities. A pre-programmed sterilization lumen cycleoperates at low pressure and low temperature and isthus suitable for processing medical devices sensitiveto heat and moisture.The STERLINK mini can sterilize*:• Instruments with diffusion-restricted spaces such asthe hinged portion of forceps and scissors• Medical devices with a single stainless-steel lumenwith:o An inside diameter of 1.6 mm or larger and alength of 200 mm or shorter*The validation testing for all lumen sizes wasconducted using a maximum of five (5) lumens perload. Hospital loads should not exceed the maximumnumber of lumens validated by this testing. TheThe STERLINKTM FPS-15s Plus sterilizer withSTERLOADTM cassette is a vaporized hydrogenperoxide sterilizer intended for use in the terminalsterilization of cleaned, rinsed, and dried reusablemetal and nonmetal medical devices used in healthcarefacilities. A pre-programmed sterilization lumen cycleoperates at low pressure and low temperature and isthus suitable for processing medical devices sensitiveto heat and moisture.The STERLINKTM FPS-15s Plus can sterilize*:• Instruments with diffusion-restricted spaces such asthe hinged portion of forceps and scissors• Medical devices with a single stainless-steel lumenwith:o An inside diameter of 2.4 mm or larger and alength of 280 mm or shorter*The validation testing for all lumen sizes wasconducted using a maximum of five (5) lumens perload. Hospital loads should not exceed the maximumnumber of lumens validated by this testing. TheSimilar;Intended Use is thesame with theexception of themaximum load andlumen specification
Subject DevicePredicate DeviceComparison
Plasmapp Co., Ltd.Plasmapp Co., Ltd.Same
validation studies were performed using a validationvalidation studies were performed using a validation
load with a total weight of 1.54 lbs.load with a total weight of 3.97 lbs.
PhysicalCharacteristicSelf-contained, stand-alone deviceSelf-contained, stand-alone deviceSame
Design andConstructionWelded frame onto which is mounted thesterilization chamber along with a variety ofinstruments and components, and controls, piping, allof which is housed in covered frame as mainsterilizer.The vacuum pump is housed in a frame with a coverand configured separately, and connected to the mainsterilizer through a pumping pipe.Welded frame onto which is mounted the sterilizationchamber along with a variety of instruments andcomponents, controls, piping, and vacuum pump, allof which is housed in covered frameSame Components asPredicate with adifferent configuration
Chamber Volume7 L14 LSmaller than Predicate
WeightMain sterilizer: 44.1 lbs (20 kg)Pump module: 46.3 lbs (21 kg)147 lbs (67 kg)Lighter than Predicate
Max Power1000 W1000 WSame
Control SystemEmbedded LinuxEmbedded LinuxSame
Internal process monitor
TemperatureChamber and vaporizer thermocoupleChamber and vaporizer thermocoupleSame
PressureChamber pressure transducersChamber pressure transducersSame
SterilantConcentrationNoneNoneSame
OperationalPrincipleSterilization of the intended devices by exposureunder controlled conditions of pressure, temperature,and timeSterilization of the intended devices by exposureunder controlled conditions of pressure, temperature,and timeSame
OperationalParametersLow pressure (vacuum; sub-atmospheric down to 3Torr) and temperature (60°C)Low pressure (vacuum; sub-atmospheric down to 3Torr) and temperature (60°C)Same
Pre-processingRequirementsCleaned, rinsed, and dried devicesCleaned, rinsed, and dried devicesSame
Subject DevicePredicate DeviceComparison
Plasmapp Co., Ltd.Plasmapp Co., Ltd.Same
DevicesReusable metal and non-metal medical devices thatare used in healthcare facilities, including those thatare sensitive to heat and moistureReusable metal and non-metal medical devices thatare used in healthcare facilities, including those thatare sensitive to heat and moistureSame
Sterilization CyclesOne (1) pre-programmed; approximately 18 minutesOne (1) pre-programmed; approximately 36 minutesSimilar; Shortenedcycle compared toPredicate
Sterilant
Model NameSTERLOAD™ miniSTERLOAD™Different
TypeCassette type (unit dose)Cassette type (unit dose)Same
Sterilant58-59.5% aqueous solution of hydrogen peroxide58-59.5% aqueous solution of hydrogen peroxideSame
Shelf Life12 months12 monthsSame
Monitoring accessories*
BiologicalIndicatorSelf-contained biological indicator, GeobacillusstearothermophilusSelf-contained biological indicator, GeobacillusstearothermophilusSame
Process ChallengeDevice /Routine Test PackSelf-contained biological indicator, GeobacillusstearothermophilusSelf-contained biological indicator, GeobacillusstearothermophilusSame
Chemical IndicatorTerragene® CI Strips and TapesTerragene® CI Strips and TapesSame
Miscellaneous (Sterilization wrap) *
Load PackagingTyvek® and PET/LLDPE filmTyvek® and PET/LLDPE filmSame

{9}------------------------------------------------

{10}------------------------------------------------

  • Denotes revision made to correct manufacturing material (previously listed as Tyvek®/HDPE)

Technological differences between the STERLINK mini sterilizer and the predicate device are minimal and include:

  • · The STERLINK mini sterilizer has a smaller sterilization chamber volume, which is lighter and requires separated pump module.

{11}------------------------------------------------

Performance Data A.4

The non-clinical tests performed are listed in the table below.
-----------------------------------------------------------------------
TestPurposeAcceptable criteriaResult
Human factors andusability engineeringTo confirm whether the user error affects the risk to user safety andperformance of the device when the device is used according to theintended use in accordance with IEC 62366-1User safety risk and device performance degradation due to usererrors should not occur.Pass
BiocompatibilityTo confirm the biological safety of the sterilized load in accordancewith ISO 10993-5.≤ Grade 2Pass
Software validationTo verify that the software is compatible for the intended use of thedevice in accordance with IEC 62304The risk of all identified software hazards should be reduced toan acceptable risk level and the device is appropriate for itsintended use.Pass
Electrical safetyTo confirm the electrical safety of the device in accordance with IEC60601-1, IEC 61010-1, and IEC 61010-2-040The device conforms to the relevant standards.Pass
Electromagneticcompatibility (EMC)To confirm the electromagnetic wave safety of the device inaccordance with IEC 60601-1-2 and EN 55011The device conforms to the relevant standards.Pass
Resistance validation forbiological indicator testTo validate the D-value and total kill endpoint of the biologicalindicators in accordance with ISO 11138-1The total kill endpoint time should be shorter than thesterilization half cycle.Pass
Lumen sterilizationTo confirm the lumen sterilization performance in accordance withISO 14937The SAL (Sterility Assurance Level) should be at least 10-6.Pass
Surface sterilizationTo confirm the surface sterilization performance of various materialsin accordance with ISO 14937, ISO 11737-1, and ISO11737-2.The SAL (Sterility Assurance Level) should be at least 10-6.Pass
Mated surfacesterilizationTo confirm the mated surface sterilization performance of variousmaterials in accordance with ISO 14937, ISO 11737-1, andISO11737-2.The SAL (Sterility Assurance Level) should be at least 10-6.Pass
Simulated use testTo demonstrate the efficacy of the sterilization validation on medicaldevices contaminated by organic and inorganic matters with bacterialspores in accordance with ASTM E1837-96 and ISO 11737-1.The microorganisms should not survive after sterilization.Pass
In-use testTo demonstrate the sterilization validation of the used and clinicallypretreated medical devices in accordance with ASTM E1837-96The microorganisms should not survive after sterilization.Pass
Sporicidal activity testTo confirm the sporicidal activity of the device in accordance withAOAC Official Method 966.04All spores inoculated on the carriers must be sterilized.Pass
Bacteriostasis testTo confirm the bacteriostasis action of sterilized various materialsafter sterilization process in accordance with ISO 11737-1Sterilized materials should not exhibit the bacteriostatic effects.Pass
Material compatibilitytestTo demonstrate the compatibility of various materials in thesterilization cycle in accordance with ASTM D638, ASTM E8/E8M,ASTM D790, ASTM E290-14, ASTM D256, ASTM E23-18, ASTME1164, ASTM D3985, and ASTM E1240The physical properties of the material after sterilization shouldbe similar to those before sterilization.Pass
TestPurposeAcceptable criteriaResult
Delivery validationTo validate the packaging and performance of the device afterexposure to simulated shipping conditions in accordance with ASTMD4169-14.After the distribution cycle of the drop test, significantdeformations and defects in the electrical and mechanicalperformance of the device should not be observed.Pass
Hydrogen peroxide gasdetection testTo measure the concentration of the hydrogen peroxide emitted fromthe device in accordance with Occupational Safety and HealthAdministration (OSHA, TWA).< 1 ppm hydrogen peroxide gas detectionPass
Sterilant preservation testTo verify the shelf life of the sterilant cassette in accordance with theinternal test standard.The sterilant should meet the specified weight, concentration,and pH standards during the shelf life.Pass

{12}------------------------------------------------

{13}------------------------------------------------

Tyvek® Roll with CI for STERLINK™ Sterilizer B.

B.1 Device Description

Tyvek® Roll with CI for STERLINK™ Sterilizer is intended to contain medical devices to be terminally sterilized in the STERLINK™ sterilization system. The medical devices are inserted into the roll, sealed, and then sterilized in the STERLINK™ sterilization system. After completion of the sterilization process, the roll maintains sterility of the enclosed medical devices until the seal is opened. The device is intended to allow sterilization of enclosed devices and also to maintain sterility of the enclosed devices for up to 1 month post sterilization.

The roll is printed with a chemical indicator bar that changes from red to blue (or lighter) when exposed to hydrogen peroxide vapor during processing in the STERLINKTM sterilizer.

The Tyvek® Roll with CI for STERLINKIM Sterilizer is offered in one type as a Flat Sterilization roll. The sterilization roll is made from a Tyvek® sheet and a clear plastic film that are heat sealed together on opposite two sides. After being cut into a suitable length, the product to be sterilized is placed inside and the two open ends are heat sealed. The process indicator printed on the Tyvek® will exhibit a color change after the roll is exposed to hydrogen peroxide (H2O2).

B.2 Indications / Intended Use

The subject and predicate device have the same intended use. The specific indications for use differ only in the identification of the appropriate sterilization cycles and the weight of the medical devices that can be sterilized in each chamber. The revised indications for use are as follows:

Tyvek® Roll with CI for STERLINK™ Sterilizer, when used in CHAMBER mode, is intended to provide health care workers with an effective method to enclose devices intended for sterilization in the STERLINK™ sterilizer. The device is intended to allow sterilization of enclosed devices and also to maintain sterility of the enclosed devices until used up to 1 month post sterilization.

The materials compatible for use in the Tyvek® Roll with CI for STERLINK™ Sterilizer, when used in CHAMBER mode, are as follows: Aluminum 5052, Aluminum 6061, Stainless Steel 304, Stainless Steel 316L, Titanium, Acrylonitrile Butadiene Styrene (ABS), High Density Polyethylene (HPDE), Polypropylene (PP), Polytetrafluorethylene (PTFE) and Silicone (Hardness 50).

{14}------------------------------------------------

The maximum load weight that can be placed in the Tyvek® roll is:

  • 3.97 pounds (1.8 kg) for CHAMBER mode of FPS-15s Plus sterilizer .
  • 1.54 pounds (0.7 kg) for CHAMBER mode of STERLINK mini sterilizer

The roll is printed with a chemical indicator bar which is a process indicator (ISO 11140-1:2005) that changes from red to blue (or lighter) when exposed to hydrogen peroxide vapor during processing in the STERLINK™ sterilizer.

The Tyvek® Roll with CI for STERLINK™ Sterilizer is offered in the follow 1 type:

  • . Sterilization roll, Flat

{15}------------------------------------------------

B.3 Technical Characteristics in Comparison to Predicate Devices

A summary of the technical characteristics of the subject device can be found in the table below.

Subject DevicePredicate DeviceComparison
510(k) SponsorPlasmapp Co., Ltd.Plasmapp Co., Ltd.Same
ManufacturerSigma Medical Supplies Corp.Sigma Medical Supplies Corp.Same
Device NameTyvek® Roll with CI for STERLINK™ SterilizerTyvek® Roll with CI for STERLINK™ SterilizerSame
510(k) Number-K212198-
Device ClassificationName1) Sterilization Wrap2) Sterilization Process Indicator1) Sterilization Wrap2) Sterilization Process IndicatorSame
Classification ProductCode1) FRG2) JOJ1) FRG2) JOJSame
Regulation Number1) 21 CFR 880.68502) 21 CFR 880.28001) 21 CFR 880.68502) 21 CFR 880.2800Same
Intended UseTyvek® Roll with CI for STERLINK™ Sterilizer,when used in CHAMBER mode, is intended toprovide health care workers with an effectivemethod to enclose devices intended for sterilizationin the STERLINK™ sterilizer. The device isintended to allow sterilization of enclosed devicesand also to maintain sterility of the enclosed devicesuntil used up to 1 month post sterilization.The materials compatible for use in the Tyvek® Rollwith CI for STERLINK™ Sterilizer, when used inChamber mode, are as follows: Aluminum 5052,Aluminum 6061, Stainless Steel 304, Stainless Steel316L, Titanium, Acrylonitrile Butadiene Styrene(ABS), High Density Polyethylene (HDPE),Polypropylene (PP), Polytetrafluoroethylene (PTFE)and Silicone (Hardness 50).Tyvek® Roll with CI for STERLINK™ Sterilizer,when used in CHAMBER mode, is intended toprovide health care workers with an effectivemethod to enclose devices intended for sterilizationin the STERLINK™ sterilizer. The device isintended to allow sterilization of enclosed devicesand also to maintain sterility of the enclosed devicesuntil used up to 1 month post sterilization.The materials compatible for use in the Tyvek® Rollwith CI for STERLINK™ Sterilizer, when used inChamber mode, are as follows: Aluminum 5052,Aluminum 6061, Stainless Steel 304, StainlessSteel 316L, Titanium, Acrylonitrile ButadieneStyrene (ABS), High Density Polyethylene(HDPE), Polypropylene (PP),
Subject DevicePredicate DeviceComparison
The maximum load weight that can be placed in the Tyvek® roll is:• 3.97 pounds (1.8 kg) for CHAMBER mode of FPS-15s Plus sterilizer• 1.54 pounds (0.7 kg) for CHAMBER mode of STERLINK mini sterilizerPolytetrafluoroethylene (PTFE) and Silicone(Hardness 50).The maximum load weight that can be placed in the Tyvek® roll is 3.97 lbs (1.8 kg).
The roll is printed with a chemical indicator bar which is a process indicator (ISO 11140-1:2005) that changes from red to blue (or lighter) when exposed to hydrogen peroxide vapor during processing in the STERLINK™ sterilizer.The Tyvek® Roll with CI for STERLINK™The roll is printed with a chemical indicator bar which is a process indicator (ISO 11140-1:2005) that changes from red to blue (or lighter) when exposed to hydrogen peroxide vapor during processing in the STERLINK™ sterilizer.
Sterilizer is offered in the follow 1 type:• Sterilization roll, FlatThe Tyvek® Roll with CI for STERLINK™Sterilizer is offered in the follow 1 type:• Sterilization roll, Flat
Pouch TypesSterilization roll, FlatSterilization roll, FlatSame
Device models(Configurations/Dimensions)Mode: FR400100Dimensions: 400mm x 100MMode: FR400100Dimensions: 400mm x 100MSame
Material CompositionTyvek®, PET, PE, Water, CH3COOH, Alcohol, n-Heptane adhesive, Hydrogen peroxide vaporProcess Indicator Print InkTyvek®, PET, PE, Water, CH3COOH, Alcohol, n-Heptane adhesive, Hydrogen peroxide vaporProcess Indicator Print InkSame
Sterilization CycleSTERLINK™ FPS-15s Plus - Chamber mode(overall cycle: 36 minutes)STERLINK mini - Chamber mode (overall cycle: 18minutes)STERLINK™ FPS-15s Plus - Chamber mode(overall cycle: 36 minutes)Similar; Expanded toinclude additionaldevice and itsoperating mode
Design FeatureSterilization roll, Flat: This roll is made from aTyvek® and plastic film that are heat sealed onSterilization roll, Flat: This roll is made from aTyvek® and plastic film that are heat sealed onSame
Subject DevicePredicate DeviceComparison
opposite two sides. It will be cut into the suitablelength and the opened sides will be heat-sealed. Theindicators printed on the Tyvek® are the same withthe self-sealing sterilization roll.opposite two sides. It will be cut into the suitablelength and the opened sides will be heat-sealed.The indicators printed on the Tyvek® are the samewith the self-sealing sterilization roll.
Chemical IndicatorDevice DesignThe color of the Chemical Indicator changes fromred to blue (or lighter) when exposed to hydrogenperoxide.The color of the Chemical Indicator changes fromred to blue (or lighter) when exposed to hydrogenperoxide.Same
Shelf Life28 months for STERILINK mini – Chamber mode15 months for STERILINK FPS-15s Plus –Chamber mode15 months for STERILINK FPS-15s Plus –Chamber mode

{16}------------------------------------------------

{17}------------------------------------------------

B.4 Performance data

The non-clinical tests performed are listed in the table below.

Performance TestPurposeAcceptable criteriaResult
Sterilization efficacy testTo ensure that the device is suitable for the sterilizationprocess in accordance with ISO 14937.The SAL should be at least 10-6.Pass
CI of sterilization rollvalidationTo confirm that the shelf life before and after thesterilization of device, and to ensure CI of the device iscompatible for the sterilization process in accordancewith ANSI/AAMI/ISO 11140-1.- CI should not change completely with temperature,and sterilant exposure time of less than 7 seconds,during the shelf life.- CI should maintain the changed color after real-timeaging during shelf life after sterilization.Pass
Internal pressurization testTo verify the internal pressurization of the sterilizeddevice at the shelf life in accordance with ASTMF1980-16 and ASTM F1140/F1140M-13.- Burst pressure: $ ≧ $ 1kPA- Creep time: > 30 sec in creep pressure (> 80% ofburst pressure)Pass
Visual inspection testTo confirm the sterilized device integrity wasmaintained at shelf life in accordance with ASTMF1980-16 and ASTM F1886/F1886M-16.There should be no defects in the sealing area.Pass

{18}------------------------------------------------

Performance TestPurposeAcceptable criteriaResult
Dye penetration testTo confirm the packaging of the sterilized device wasmaintained at shelf life in accordance with ASTMF1980-16 and ASTM F1929-15.There should be no dye leakage.Pass
Tensile strength of Tyvek®Pass
Tensile strength of plastic filmTo confirm the physical properties of the sterilizeddevice at shelf life in accordance with ASTM F1980-The physical properties must meet the criteriaspecified by the manufacturer.Pass
Seal strength16, ASTM D5035-11, ASTM D882, ASTM F88, andASTM D1922-20.Pass
Tear resistancePass
Microbial Barrier TestTo confirm the microbial barrier performance ofsterilized devices at shelf life in accordance withASTM F1980-16 and DIN 58953-6.- Test with liquid inoculum: No microbial growth should appear.- Test with gas inoculum: Only a maximum 15 colonies should grow out ofa total of 10 bacterial permeability testers.The number of colonies in each bacterialpermeability tester should not be greater than five.Pass
Residual sterilant on Tyvek®validationTo confirm the time required for the residual sterilantto be reduced to a safe amount after sterilization inaccordance with the internal test standard.≤ 0.05 ppmPass
Biocompatibility testTo confirm the biological safety after sterilization inaccordance with ISO 10993-5≤ Grade 2Pass

{19}------------------------------------------------

C. Sterilization Process Indicator for STERLINK™ Sterilizer

C.1 Device Description

Terragene® Bionova® BT96 Fluorescence Super Rapid Readout Biological Indicators are singleuse Self-Contained Biological Indicators (SCBIs) that consist of a polypropylene tube, a spore carrier, and a glass ampoule with a culture medium, enclosed with a colored cap. Each tube contains a population of Geobacillus stearothermophilus ATCC 7953 spores inoculated on a spore carrier, a plastic cap with holes and a barrier permeable to Plasma or Vaporized Hydrogen Peroxide. Each BT96 has a Process Indicator on label that changes from purple to green when exposed to hydrogen peroxide. The Bionova® BT96 Biological Indicators have been designed for monitoring of Vaporized Hydrogen Peroxide sterilization processes when used in conjunction with Bionova® IC10/20FRLCD or MiniBio Auto-Readers Incubators.

Chemdye® CD42 Process Indicators (Type 1 according to ISO 11140-1:2014 standard) are singleuse chemical indicators that consist of plastic strips printed with indicator ink. These indicators have been designed to monitor Plasma or Vaporized Hydrogen Peroxide sterilization processes within loads, ensuring an adequate exposure to the sterilizing agent during the sterilization process and allowing to distinguish between processed and unprocessed items.

Cintape® CT40 Process Indicators (Type 1 according to ISO 11140-1:2014 standard) are singleuse chemical indicators that consist of a roll of self-adhesive plastic tape printed with indicator ink. These indicators have been designed to monitor Plasma or Vaporized Hydrogen Peroxide sterilization processes, ensuring an adequate exposure to the sterilizing agent during the sterilization process and allowing to distinguish between processed and unprocessed items. The adhesive component of the tape allows the adhesion to different types of packaging and wraps, such as cloth, paper and plastic.

C.2 Indications for Use / Intended Use

The subject and predicate device have the same intended use. The specific indications for use differ only in the identification of the appropriate sterilization cycles. The revised indications for use are as follows:

Terragene Bionova® SCBI (BT96) is a self-contained biological indicator inoculated with viable 106 Geobacillus stearothermophilus bacterial spores and is intended for monitoring the efficacy of vaporized hydrogen peroxide sterilization processes. BT96 has Super Rapid readout at 30 minutes at 60°C.

{20}------------------------------------------------

Terragene Bionova® Reader Incubators (IC10/20FRLCD, Mini-Bio) incubate at 60°C and 37°C and read the Terragene Bionova® SCBI for fluorescent results at the times prescribed in the User Manuals.

Terragene Chemdye® (CD42) is a chemical process indicator intended for monitoring the efficacy of vaporized hydrogen peroxide sterilization processes. The chemical indicator changes from red to yellow to indicate that the conditions of the cycle have been met.

Terragene Cintape® (CT40) is a chemical process indicator tape intended for monitoring the efficacy of Vaporized Hydrogen Peroxide sterilization processes. The indicating tape changes from purple to green when exposure to vaporized hydrogen peroxide.

The self-contained biological indicator and chemical processing indicators are intended for use with:

  • CHAMBER mode of FPS-15s Plus sterilizer
  • CHAMBER mode of STERLINK mini sterilizer

C.3 Technical Characteristics in Comparison to Predicate Devices

A summary of the technical characteristics of the subject device and predicate device can be found in the table below.

Subject DevicePredicate DeviceComparison
SponsorPlasmapp Co., Ltd.Plasmapp Co., Ltd.Same
Device NameSterilization Process Indicator forSTERLINKTM SterilizerTerragene Bionova® SCBI (BT96);Terragene Bionova® ReaderIncubators (IC10/20FRLCD, Mini-Bio);Terragene Chemdye® (CD42);Terragene Cintape® (CT40)Terragene Bionova® SCBI (BT96);Terragene Bionova® ReaderIncubators (IC10/20FRLCD, Mini-Bio);Terragene Chemdye® (CD42);Terragene Cintape® (CT40)Same
510(k)Number-K212193-
ManufacturerTerragene® S.A.Terragene® S.A.Same
DeviceClassificationNameSterilization Process IndicatorSterilization Process IndicatorSame
ClassificationProduct CodeFRC (biological indicators)JOJ (chemical indicators)FRC (biological indicators)JOJ (chemical indicators)Same
RegulationNumber21 CFR 880.280021 CFR 880.2800Same

{21}------------------------------------------------

Subject DevicePredicate DeviceComparison
Indications forUseTerragene Bionova® SCBI (BT96) isa self-contained biological indicatorinoculated with viable 106Geobacillus stearothermophilusbacterial spores and is intended formonitoring the efficacy of vaporizedhydrogen peroxide sterilizationprocesses. BT96 has Super Rapidreadout at 30 minutes at 60°C.Terragene Bionova® SCBI (BT96) is aself-contained biological indicatorinoculated with viable 106 Geobacillusstearothermophilus bacterial sporesand is intended for monitoring theefficacy of vaporized hydrogenperoxide sterilization processes. BT96has Super Rapid readout at 30 minutesat 60°C.
Terragene Bionova® ReaderIncubators (IC10/20FRLCD, Mini-Bio) incubate at 60°C and 37°C andread the Terragene Bionova® SCBIfor fluorescent results at the timesprescribed in the User Manuals.Terragene Bionova® Reader Incubators(IC10/20FRLCD, Mini-Bio) incubateat 60°C and 37°C and read theTerragene Bionova® SCBI forfluorescent results at the timesprescribed in the User Manuals.
Terragene Chemdye® (CD42) is achemical process indicator intendedfor monitoring the efficacy ofvaporized hydrogen peroxidesterilization processes. The chemicalindicator changes from red to yellowfor CD42 to indicate that theconditions of the cycle have been met.Terragene Chemdye® (CD42) is achemical process indicator intended formonitoring the efficacy of vaporizedhydrogen peroxide sterilizationprocesses. The chemical indicatorchanges from red to yellow for CD42to indicate that the conditions of thecycle have been met.Similar;Expanded toincludeadditionaldevice and itsoperatingmode
Terragene Cintape® (CT40) is achemical process indicator tapeintended for monitoring the efficacyof Vaporized Hydrogen Peroxidesterilization processes. The indicatingtape changes from purple to greenwhen exposure to vaporized hydrogenperoxide.Terragene Cintape® CT40 is achemical process indicator tapeintended for monitoring the efficacy ofVaporized Hydrogen Peroxidesterilization processes. The indicatingtape changes from purple to greenwhen exposure to vaporized hydrogenperoxide.
The self-contained biologicalindicator and chemical processingindicators are intended for use with:• CHAMBER mode of FPS-15s Plussterilizer• CHAMBER mode of STERLINKmini sterilizerThe self-contained biological indicatorand chemical processing indicators areintended for use with theSTERLINKTM FPS-15s Plus whenoperating in chamber mode.
Intended Use:CyclesVaporized Hydrogen Peroxide Models Cycle BT96,CD42,CT40 STERLINKTM FPS-15sPlus - Chamber modeSTERLINK mini -Chamber modeVaporized Hydrogen Peroxide Models Cycle BT96,CD42,CT40 STERLINKTM FPS-15sPlus - Chamber modeSimilar;Expanded toincludeadditionaldevice and itsoperatingmode

{22}------------------------------------------------

Subject DevicePredicate DeviceComparison
Terragene Bionova® SCBI (BT96)
Type ofBiologicalIndicatorSelf-ContainedSelf-ContainedSame
OrganismSporeSpeciesStrainGeobacillus stearothermophilusATCC 7953 spores inoculated on astrip (spore carrier)Geobacillus stearothermophilusATCC 7953 spores inoculated on astrip (spore carrier)Same
Viable SporePopulation≥ 106≥ 106Same
ResistancecharacteristicsD-valueSurvival time/Kill windowD-valueSurvival time/Kill windowSame
IntendedSterilizationCyclesSTERLINKTM FPS-15s Plus -Chamber mode (overall cycle: 36minutes)STERLINK mini - Chamber mode(overall cycle: 18 minutes)STERLINKTM FPS-15s Plus -Chamber mode (overall cycle: 36minutes)Similar;Expanded toincludeadditionaldevice and itsoperatingmode
Shelf Life2 years2 yearsSame
Terragene Chemdye® (CD42), Terragene Chemdye® (CT40)
IntendedSterilizationCyclesSTERLINKTM FPS-15s Plus -Chamber mode (overall cycle: 36minutes)STERLINK mini - Chamber mode(overall cycle: 18 minutes)STERLINKTM FPS-15s Plus -Chamber mode (overall cycle: 36minutes)Similar;Expanded toincludeadditionaldevice and itsoperatingmode
Device designStrip, TapeStrip, TapeSame
Color Changeupon Exposureto H2O2CD42: red to yellowCT40: purple to greenCD42: red to yellowCT40: purple to greenSame
RecommendedStorageConditionsDry place, away from sunlight, attemperature between 10-30°C, 30-80% relative humidity. Do not wet.Do not store close to sterilizingagents.Dry place, away from sunlight, attemperature between 10-30°C, 30-80%relative humidity. Do not wet. Do notstore close to sterilizing agents.Same
Shelf Life5 years5 yearsSame

The subject and predicate devices are identical with respect to the organism, accessories, spore population, resistance characteristics, culture conditions, carrier materials, packaging, storage conditions and claimed shelf life. The only difference between the subject and predicate devices

{23}------------------------------------------------

are the proposed indications for use, to label the indicators for use with the STERILINK mini when operating in chamber mode.

C.4 Performance Data

The non-clinical tests performed are listed in the table below.

ItemTestPurposeAcceptable criteriaResult
Self-ContainedBiologicalIndicator(BT96)Resistancevalidation forbiologicalindicator testTo validate the D-value andtotal kill endpoint of biologicalindicators in accordance withISO 11138-1.The total kill endpoint timeshould be shorter than the halfcycle sterilant exposure time.Pass
BI & Test packvalidation testTo verify the suitability of thetest pack to be used asmonitoring of the sterilizationprocess in accordance with theinternal test standard.The test pack with SCBIshould fail to sterilize in halfcycle and succeed in fullcycle.Pass
ChemicalIndicator(CD42,CT40)Performancecharacteristicsobtained fromresistometerTo evaluate the correctperformance of chemicalindicators, by testing them atdifferent exposure conditions inaccordance withANSI/AAMI/ISO 11140-1.The CI should change colorcorrectly only when exposedto the temperature andsterilant concentrationconditions specified by themanufacturerPass
BiocompatibilityTo demonstrate that the CIdoes not offset or transferduring and after thesterilization process inaccordance withANSI/AAMI/ISO 11140-1.The CI does not release anysubstance or bleed when thesterilization process.Pass
EndpointstabilityTo demonstrate the stability ofthe endpoint reaction of CI atthe end of their shelf life andspecified period aftersterilization in accordance withANSI/AAMI/ISO 11140-1.The endpoint reaction of CIshould be maintained until theshelf life and specified periodafter sterilization.Pass
Shelf life studyTo verify the shelf life of theCI in accordance withANSI/AAMI/ISO 11140-1.The endpoint reaction of thereal time aged CI should bemaintained until shelf life.Pass
ChemicalindicatorvalidationTo confirm the compatibility ofthe CI for the sterilizationprocess in accordance with theinternal test standardThe minimum time requiredfor color change of CI withinthe sterilization process timeshould be confirmed.Pass

{24}------------------------------------------------

Clinical Data

This submission does not contain any data from clinical testing

Conclusions

The conclusions drawn from the nonclinical tests demonstrate that the subject devices, STERLINK mini Sterilizer with STERLOAD mini Cassette, Tyvek Roll with CI for STERLINK Sterilizer, and Sterilization Process Indicator for STERLINK Sterilizer: Terragene Bionova SCBI (BT96), Terragene Bionova Reader Incubators (IC10/20FRLCD, Mini-Bio), Terragene Chemdye (CD42) and Terragene Cintape (CT40) are as safe, as effective, and perform as well as or better than the legally marketed predicate devices, STERLINK FPS-15s Plus Sterilizer with STERLOAD™ Cassette, Tyvek® Roll with CI for STERLINK™ Sterilizer, and Terragene Bionova® SCBI (BT96), Terragene Bionova® Reader Incubators (IC10/20FRLCD, Mini-Bio), Terragene Chemdye® (CD42), Terragene Cintape® (CT40).

§ 880.6860 Ethylene oxide gas sterilizer.

(a)
Identification. An ethylene gas sterilizer is a nonportable device intended for use by a health care provider that uses ethylene oxide (ETO) to sterilize medical products.(b)
Classification. Class II (performance standards).