K Number
K220218

Validate with FDA (Live)

Device Name
Siren SGT
Date Cleared
2022-05-18

(112 days)

Product Code
Regulation Number
876.5980
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SIREN SGT™ is indicated for use in gastric and bariatric surgical procedures for gastric aspiration and lavage, and to assist in gastric sleeve formation by decompressing and stabilizing the stomach, and by serving as a sizing guide.

Device Description

The Tools for Surgery™ SIREN SGT™ gastric tube is a clean, non-sterile, single patient use device, comprising a 36 Fr insertion tube portion 109 cm long, a hub containing a suction regulating valve, an extension tube, pinch clamp, and connector. Tubular portions including the hub are flexible polyvinylchloride (PVC). There are multiple apertures in proximity to a rounded distal tip. Line and numeric markings are provided to indicate depth of insertion. The suction regulating valve limits negative pressure applied to the stomach to less than 175 mmHg, and produces an audible signal when the suction pressure in the stomach exceeds ≈160 mmHg.

AI/ML Overview

The document describes the Siren SGT™, a gastric tube used in bariatric and gastric surgical procedures. The provided content is a 510(k) Summary, which is a premarket notification to the FDA to demonstrate that the device is substantially equivalent to a legally marketed predicate device. This type of submission usually focuses on demonstrating equivalence through comparison of technical characteristics and performance testing rather than establishing clinical efficacy or a specific "acceptance criteria" measured using clinical outcomes.

Based on the provided information, the "acceptance criteria" for this device are implicitly tied to demonstrating substantial equivalence to the predicate device (Boehringer Laboratories Gastric Sizing Tube, ViSiGi 3D®, K130483). The performance testing described is intended to show that despite some differences in design, the Siren SGT™ performs comparably and safely for its intended use.

Here's an attempt to extract and organize the information according to your request, acknowledging that a typical "acceptance criteria" table with performance metrics as you might see for a diagnostic AI device isn't explicitly laid out in this type of submission.


Table of Implicit Acceptance Criteria and Reported Device Performance (based on Substantial Equivalence)

Since this is a 510(k) submission for a physical medical device, the "acceptance criteria" are not framed in terms of precision, sensitivity, or specificity as they would be for an AI/diagnostic tool. Instead, the criteria revolve around demonstrating equivalent performance to a predicate device. The "reported device performance" refers to the results of the non-clinical tests that show the Siren SGT™ functions as intended and comparably to the predicate.

Acceptance Criteria (Demonstrates Substantial Equivalence)Reported Device Performance (as demonstrated by testing)
Functional Equivalence: Device performs intended functions (suction, decompression, drainage, irrigation, sizing) safely and effectively.Demonstrated through various functional tests: - VALVE FUNCTION TEST - MAXIMAL NEGATIVE PRESSURE TEST: Confirmed the pressure relief valve opens at 103 to 155 mmHg and limits negative pressure to < 175 mmHg, similar to predicate's 100-150 mmHg. - AUDIBLE SIGNAL TEST: Confirmed the audible signal activates when suction pressure exceeds ≈160 mmHg. - VISIGI PRESSURE COMPARISONS (presumably comparing Siren SGT™ pressure performance to predicate). - PINCH CLAMP FUNCTION TEST (for manual suction control).
Dimensional Compatibility & Structural Integrity: Device dimensions are appropriate for indicated use, and the device is robust under anticipated use and stress.Bio-compatibility and Material Safety.Demonstrated through: - DIMENSIONAL ANALYSIS TEST - INSERTION TUBE KINK TESTS - INSERTION TUBE COMPRESSION TESTS - INSERTION TUBE STIFFNESS TEST - JOINT STRENGTH TEST - CAP BOND STRENGTH TEST - CONNECTOR TEST - SEAL TEST - BUBBLE TEST, SGT - L929 NEUTRAL RED UPTAKE TEST (ISO)- INTRACUTANEOUS INJECTION TEST (ISO)- KLINGMAN MAXIMIZATION TEST (ISO) (These ISO tests are for biocompatibility/material safety.)
Stability Over Time (Aging): Device maintains its performance characteristics after simulated aging.Demonstrated through "Aged" tests: - COLOR TEXTURE PRINT TEST, AGED - VALVE FUNCTION TEST, AGED - MAXIMAL NEGATIVE PRESSURE TEST, AGED - TUBE DEGRADATION TEST, AGED - INSERTION TUBE KINK TEST, AGED - PINCH CLAMP FUNCTION TEST, AGED - JOINT STRENGTH TEST, AGED - CAP BOND STRENGTH TEST, AGED
Sterility & Single-Use Compliance: Device is supplied non-sterile and is intended for single patient use, consistent with similar devices.Supplied non-sterile, single use, disposable: (Stated as "Same" as predicate in Table 1). No specific test reports are listed for this, but it's a declared characteristic that aligns with the predicate and likely involves manufacturing process controls.
Indications for Use Consistency: The device is suitable for the stated indications.The Indications for Use statement for SIREN SGT™ is "The SIREN SGT™ is indicated for use in gastric and bariatric surgical procedures for gastric aspiration and lavage, and to assist in gastric sleeve formation by decompressing and stabilizing the stomach, and by serving as a sizing guide." This is stated to be the "Same" as the predicate device, forming a key aspect of substantial equivalence.
Safety: Device operation does not introduce new safety concerns compared to the predicate, and incorporates features to mitigate risks (e.g., pressure regulation, audible signal).The pressure relief valve limits negative pressure, and the audible signal informs the operator about excessive suction, prompting manual adjustment. These features, while different from the predicate's mechanism (inline shut-off valve, no audible signal), are presented as providing equivalent or enhanced safety mechanisms. Performance tests like Maximal Negative Pressure Test and Audible Signal Test confirm these safety features operate as designed.

Regarding the other requested information:

  1. Sample sizes used for the test set and the data provenance:

    • The document describes non-clinical bench testing. It does not mention "test sets" in the context of patient data or typical AI validation; rather, it refers to a series of physical tests performed on the device itself.
    • Sample Size: The document does not specify the number of devices or components tested for each specific non-clinical test. This information is typically detailed in the individual test reports, which are "submitted, referred to, and relied on" but not elaborated upon in this summary.
    • Data Provenance: Not applicable in the context of patient data. These are laboratory/bench tests on manufactured devices.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. This is a physical device, and "ground truth" for patient data interpretations (e.g., radiology images) is not relevant here. The "ground truth" for device performance is established by engineering specifications, validated test methods, and comparison to the predicate device's known characteristics. The tests themselves, and their interpretation, would be performed by qualified engineers and technicians.
  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable. This is not a study involving human interpretation of data where adjudication is needed. Device performance is measured against predefined engineering and functional specifications.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, a TRADITIONAL MRMC study was not done. This is a non-AI physical medical device. The study design is focused on demonstrating substantial equivalence through engineering testing, not clinical effectiveness studies with human readers.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is a physical device, not an algorithm. The "standalone" performance refers to the device's functional integrity as measured in the described bench tests.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • For this type of device, the "ground truth" is defined by engineering specifications, established industry standards (e.g., ISO standards for biocompatibility), and the known functional characteristics of the predicate device. The purpose of the tests is to verify that the manufactured device meets these predefined functional and safety parameters.
  7. The sample size for the training set:

    • Not applicable. This is not an AI/machine learning device that requires a training set.
  8. How the ground truth for the training set was established:

    • Not applicable, as there is no training set for an AI model.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

May 18, 2022

Tools for Surgery, LLC Arnold Leiboff, M.D. President 8 Technology Drive, #100 East Setauket, NY 11773-3327

K220218 Trade/Device Name: Siren SGT Regulation Number: 21 CFR 876.5980 Regulation Name: Gastrointestinal tube and accessories Regulatory Class: Class II Product Code: KNT Dated: April 12, 2022 Received: April 19, 2022

Dear Arnold Leiboff:

Re:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling

{1}------------------------------------------------

(21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Je Hi An, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K220218

Device Name SIREN SGTTM

Indications for Use (Describe)

The SIREN SGT™ is indicated for use in gastric and bariatric surgical procedures for gastric aspiration and lavage, and to assist in gastric sleeve formation by decompressing and stabilizing the stomach, and by serving as a sizing guide.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

Submitter Name:Tools for Surgery, LLC
Submission Date:January 24, 2022
Submitter Address:8 Technology Drive, #100East Setauket, NY 11733-3327
Contact Person:Arnold R. Leiboff, M.D., President
Contact Email:aleiboff@toolsforsurgery.com
Phone Number:(631) 444-4448
Fax Number:(631) 444-5330
Prepared By:Arnold R. Leiboff, MD
Date Prepared:January 19, 2022
Device Trade Name:SIREN SGTTM
Device Common Name:Gastrointestinal Tube and Accessories
Device Product Code:KNT
Regulation:876.5980
Device Class:Class II
Classification Panel:Gastroenterology and Urology
Classification Name:Gastrointestinal Tube and Accessories
Predicate Device:Boehringer Laboratories Gastric Sizing Tube (ViSiGi 3D®)510(k) Number: K130483
Device Description:The Tools for Surgery™ SIREN SGTTM gastric tube is a clean, non-sterile, single patient use device, comprising a 36 Fr insertion tubeportion 109 cm long, a hub containing a suction regulating valve, anextension tube, pinch clamp, and connector. Tubular portionsincluding the hub are flexible polyvinylchloride (PVC). There aremultiple apertures in proximity to a rounded distal tip. Line andnumeric markings are provided to indicate depth of insertion. Thesuction regulating valve limits negative pressure applied to thestomach to less than 175 mmHg, and produces an audible signal whenthe suction pressure in the stomach exceeds ≈160 mmHg.
Indications for Use:The SIREN SGTTM is indicated for use in gastric and bariatricsurgical procedures for gastric aspiration and lavage, and to assist ingastric sleeve formation by decompressing and stabilizing thestomach, and by serving as a sizing guide.

{4}------------------------------------------------

Substantial Equivalence Comparison

Table 1

Table 1
DeviceProposed DeviceSIREN SGTTMPredicate DeviceViSiGi 3D®Comparison
ManufacturerTools for Surgery, LLCBoehringer Laboratories, LLC
Model (Catalog No.)SGT-365236
Intended Use
Indication for UseStatementThe Tools for Surgery™ SIRENSGTTM is indicated for use in gastricand bariatric surgical procedures forgastric aspiration and lavage, and toassist in gastric sleeve formation bydecompressing and stabilizing thestomach, and by serving as a sizingguide.The Boehringer LaboratoriesViSiGi 3D® is indicated for use ingastric and bariatric surgicalprocedures for the application ofsuction, stomach decompression,drainage of gastric fluids,irrigation and to serve as a sizingguide.Same
Typical UseGastric and bariatric proceduresGastric and bariatric proceduresSame
Environments of UseSurgery centers, hospitalsSurgery centers, hospitalsSame
Patient PopulationIndividuals undergoing bariatricand/or gastric proceduresIndividuals undergoing bariatricand/or gastric proceduresSame
Intraoperative UseYesYesSame
FunctionsSuction, decompression, drainage,irrigation, sizingSuction, decompression, drainage,irrigation, sizingSame
Technical/Performance Characteristics
Insertion TubeSingle lumen with rounded, closeddistal endSingle lumen with rounded, closeddistal endSame
Insertion Tube Length109 cm107 cmSame
Outer Diameter36 Fr36 FrSame
Distal Side Holes18 or 36 apertures108 aperturesDifferent
Internal Support of DistalPortionNoStainless steel coilDifferent
Connector for SuctionYesYesSame
Suction Regulating ValvePressure relief valveInline shut-off valveDifferent
Valve Actuating Pressure(Manufacture'sspecification)129 mmHg ± 26 mmHg125 mmHg ± 25 mmHgSimilar
Audible Pressure SignalYesNoDifferent
Manual Suction ValvePinch ClampSelection (Slide) ValveDifferent
Insertion Tube MaterialDEHP-free PVCSEBS Co-polymerDifferent
Line Markings to GuideInsertion DepthYesYesSame
Numeric Markings to GuideInsertion DepthYesNoDifferent
SterilitySupplied non-sterileSupplied non-sterileSame
Single Use, DisposableYesYesSame

{5}------------------------------------------------

Performance/Non-Clinical Testing

The following non-clinical test reports have been submitted, referred to, and relied on in this 510(k) submission:

  • VALVE FUNCTION TEST
  • MAXIMAL NEGATIVE PRESSURE TEST ●
  • AUDIBLE SIGNAL TEST ●
  • VISIGI PRESSURE COMPARISONS
  • DIMENSIONAL ANALYSIS TEST
  • INSERTION TUBE KINK TESTS
  • INSERTION TUBE COMPRESSION TESTS
  • INSERTION TUBE STIFFNESS TEST
  • JOINT STRENGTH TEST
  • CAP BOND STRENGTH TEST
  • CONNECTOR TEST
  • . SEAL TEST
  • BUBBLE TEST, SGT
  • BUBBLE TEST, VISIGI
  • PACKAGE INSPECTION TEST
  • . COLOR TEXTURE PRINT TEST, AGED
  • VALVE FUNTION TEST, AGED
  • . MAXIMAL NEGATIVE PRESSURE TEST, AGED
  • . TUBE DEGRADATION TEST, AGED
  • INSERTION TUBE KINK TEST, AGED
  • PINCH CLAMP FUNCTION TEST, AGED
  • JOINT STRENGTH TEST, AGED
  • CAP BOND STRENGTH TEST, AGED
  • THE L929 NEUTRAL RED UPTAKE TEST (1 CONCENTRATION) - ISO
  • THE INTRACUTANEOUS INJECTION TEST ISO ●
  • THE KLINGMAN MAXIMIZATION TEST ISO ●

Discussion

The Tools for Surgery™ SIREN SGT™ gastric tube is similar to the predicate device Boehringer Laboratories Gastric Sizing Tube (ViSiGi 3D®).

Intended Use

The indications for use of the Tools for Surgery™ SIREN SGT™ and the Boehringer Laboratories ViSiGi 3D® are the same.

Technical Characteristics

The Tools for Surgery™ SIREN SGT™ gastric tube has technical characteristics similar to that of the predicate device. Both devices are orogastric tubes comprised of biocompatible elastomeric materials. The predicate device incorporates a slide valve to interrupt suction, while

{6}------------------------------------------------

the SIREN SGT™ employs a pinch clamp for this purpose. The insertion tubes of both devices have interval markings to help the surgeon and anesthesiologist gauge the depth of insertion and the position of the tip of the tube. The markings on the predicate device are simple lines that are indistinguishable from one another, while the markings on the SIREN SGT™ include numeric figures that indicate the exact distance to the forward tip of the insertion tube. The predicate device includes an internal metal coil support to stiffen the distal portion of its insertion tube; the SIREN SGT™ does not. The SIREN SGT™ insertion tube comprises one or two rows of 18 apertures, whereas the predicate device, ViSiGi 3D®, comprises six rows of 18 apertures. Both the SIREN SGT™ and predicate device connect to a vacuum source in order to deliver negative pressure to the gastric lumen. The suction pressure delivered by the vacuum source is adjustable by means of an external vacuum source suction regulator. Both the SIREN SGT™ and the predicate device incorporate integral suction regulating valves, which serve as additional means for limiting negative pressure applied to the gastric mucosa. The predicate device achieves this by means of an inline shut-off valve, which closes when negative pressure exceeds 100-150 mmHg. The SIREN SGT™ contains a pressure relief valve, which opens at 103 to 155 mmHg and limits the pressure in the insertion tube and stomach to negative 175 mmHg when full suction (~360 mmHg) is applied. When negative pressure within the SIREN SGT™ insertion tube exceeds 160 mmHg an audible signal produced by mechanical means by the suction regulating valve informs the operator that excessive suction is being applied to the tube, and prompts the operator to lower the suction pressure by manually adjusting the vacuum source suction regulator.

Performance Testing

The performance tests referenced above substantiate that the proposed device performs equivalently to the predicate device, and validate the function regulating valve and audible signal mechanism.

Conclusion

As evidenced by the intended use, technical characteristics, and performance testing, the Tools for Surgery™ SIREN SGT™ gastric tube is substantially equivalent to the predicate device.

§ 876.5980 Gastrointestinal tube and accessories.

(a)
Identification. A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).(b)
Classification. (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.