K Number
K220213

Validate with FDA (Live)

Device Name
Empulse R90
Date Cleared
2022-03-30

(63 days)

Product Code
Regulation Number
890.3860
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Empulse R90 is intended to be used to provide power assistance to manual wheelchair users. It is designed to augment the manual propulsion provided by the user, reducing the effort required by the user to propel the wheelchair.

Device Description

The Empulse R90 is an auxiliary power module to provide augmented power to manual wheelchair users. The device utilizes a single electric drive wheel to provide the augmented power. The device mounts to a camber tube (rigid wheelchair) or a compressible axle (folding wheelchair) to provide the assisted power. On-board sensors, including an inclinometer, accelerometer and motor current sensing to determine when the user wishes to move forward and if the system is on an incline or decline. The removable Lithium lon battery pack is easily switched out to an alternate battery pack if additional range is required. The battery can also be removed if the user wishes to reduce the overall weight. The actuator is used to automatically engage and disengage the locking latch as well as compress the pre-load traction spring when the user wishes to disconnect the EMPULSE R90 from the wheelchair. In the extended position, the actuator allows the pre-load spring to provide sufficient pre-load to the drive wheel to ensure sufficient traction.

AI/ML Overview

The provided text is a 510(k) Premarket Notification for a medical device called "Empulse R90", a power assist device for manual wheelchairs. It focuses on demonstrating substantial equivalence to predicate devices through technical comparisons and non-clinical testing.

The document does not contain any information about acceptance criteria or a study proving the device meets acceptance criteria related to an AI/ML-driven medical device, ground truth establishment, sample sizes for training/test sets, expert adjudication methods, or MRMC studies.

Therefore, I cannot fulfill your request for the specific information regarding AI/ML device acceptance criteria and study details. The document pertains to a physical medical device and its equivalent performance to existing physical predicate devices through non-clinical mechanical, electrical, and performance testing, rather than an AI/ML diagnostic or predictive tool.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

March 30, 2022

Sunrise Medical (US) LLC Devin Mcelroy Senior Director Quality Assurance and Regulatory Affairs 2842 N Business Park Ave. Fresno, California 93727

Re: K220213

Trade/Device Name: Empulse R90 Regulation Number: 21 CFR 890.3860 Regulation Name: Powered Wheelchair Regulatory Class: Class II Product Code: ITI Dated: January 21, 2022 Received: January 26, 2022

Dear Devin Mcelroy:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Heather Dean, PhD Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K220213

Device Name Empulse R90

Indications for Use (Describe)

The Empulse R90 is intended to be used to provide power assistance to manual wheelchair users. It is designed to augment the manual propulsion provided by the user, reducing the effort required by the user to propel the wheelchair.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

DATE PREPARED

January 10, 2021

MANUFACTURER AND 510(K) OWNER

Sunrise Medical (US) LLC 2842 N Business Park Ave., Fresno, CA 93727 USA Telephone: 1-559-294-2374 Official Contact: Mr. Devin I. McElroy, Senior Director Quality Assurance and Regulatory Affairs

REPRESENTATIVE / CONTACT

Devin I. McElroy, Senior Director Quality Assurance and Regulatory Affairs Sunrise Medical (US) LLC 2842 N Business Park Ave., Fresno, CA 93727 USA Telephone: 1-559-294-2374 Email: devin.mcelroy@sunmed.com Website: www.sunmed.com

DEVICE INFORMATION

Proprietary Name/Trade Name:Empulse R90
Common Name:Power Assist Device
Regulation Number:21 CFR 890.3860
Class:II
Product Code:ITI: Powered Wheelchair
Premarket Review:Neurological and Physical Medicine Devices (OHT5)
Neuromodulation and Physical Medicine Devices (DHT5B)
Review Panel:Physical Medicine

PREDICATE DEVICE IDENTIFICATION

The Empulse R90 is substantially equivalent to the following predicates:

510(k) NumberPredicate Device Name / ManufacturerPrimary Predicate
K151199Smart Drive MX2 / Max Mobility (Permobil)$\checkmark$
K140204JWX-2 (Navione) / Yamaha

DEVICE DESCRIPTION

The Empulse R90 is an auxiliary power module to provide augmented power to manual wheelchair users. The device utilizes a single electric drive wheel to provide the augmented power. The device mounts to a camber tube (rigid wheelchair) or a compressible axle (folding wheelchair) to provide the assisted power. On-board sensors, including an inclinometer, accelerometer and motor current sensing to determine when the user wishes to move forward and if the system is on an incline or decline.

{4}------------------------------------------------

The removable Lithium lon battery pack is easily switched out to an alternate battery pack if additional range is required. The battery can also be removed if the user wishes to reduce the overall weight.

The actuator is used to automatically engage and disengage the locking latch as well as compress the pre-load traction spring when the user wishes to disconnect the EMPULSE R90 from the wheelchair. In the extended position, the actuator allows the pre-load spring to provide sufficient pre-load to the drive wheel to ensure sufficient traction.

INTENDED USE

The EMPULSE R90 is designed to be used on both indoor and outdoor environments, smooth flat surfaces, inclined surfaces up to 6 degrees and lightly rough surfaces such as asphalt. It is not designed to be used on highly rough surfaces such as grass or loose soil.

INDICATIONS FOR USE

The Empulse R90 is intended to be used to provide power assistance to manual wheelchair users. It is designed to augment the manual propulsion provided by the user, reducing the effort required by the user to propel the wheelchair.

COMPARISON OF TECHNOLOGICAL CHARACTERISTICS

Sunrise Medical (US) LLC believes that the Empulse R90 Power Assist Device is substantially equivalent to the predicate devices based on the information summarized here and in Table 5.1:

The subject device has a similar design and dimensions to the device cleared in K151199 and uses similar or identical materials as the devices cleared in K140204. The subject device has the same intended use and similar technological characteristics to the devices cleared in K151199 and K140204. The subject device has the same intended use environment as the devices cleared in K151199 and K140204. The subject device uses similar software to the devices cleared in K151199 and K140204. The Empulse R90 Power Assist Device has undergone testing to ensure that any differences in technological characteristics (e.g., battery, controls, and power assistance) do not negatively affect safety and effectiveness when compared to the predicate devices.

{5}------------------------------------------------

Table 5.1: Table of Technological Comparison
Primary PredicatePermobil (Max Mobility)Smart Drive MX2Secondary PredicateYamaha JWX-2 (Navione)Subject DeviceEmpulse R90Statement of Equivalence
1Intended UseK151199ITI 890.386030-Jul-15K140204ITI 890.386028-Aug-14K220213ITI 890.3860
1.1 Intended UseNot provided.The device JWX-2 (Navione) is a Power Assist Wheelchair Conversion Kit and suitable for the manual wheelchair users who are limited in their field of activities because of their physical conditions. The device can expand their field of activities by assisting their wheelchair operating force.The Empulse R90 is intended to be used to provide power assistance to manual wheelchair users. It is designed to augment the manual propulsion provided by the user, reducing the effort required by the user to propel the wheelchair.Substantially equivalent to Secondary Predicate. No new issues of safety or effectiveness.
1.2 Indications for UseThe SmartDrive MX2 Wheelchair Power Assist is intended to provide auxiliary power to manual wheelchairs to reduce the pushing power needed by their users, including pediatrics. It is intended to be used by users capable of operating and maneuvering a powered and manual wheelchairThe Yamaha JWX-2 is a Power Assist Wheelchair Conversion Kit and suitable for the manual wheelchair users who are limited in their field of activities because of their physical conditions. The device can expand their field of activities by assisting their wheelchair operating force.The Empulse R90 is intended to be used to provide power assistance to manual wheelchair users. It is designed to augment the manual propulsion provided by the user, reducing the effort required by the user to propel the wheelchair.Substantially Equivalent to Predicates. No new issues of safety or effectiveness.
Primary PredicatePermobil (Max Mobility)Smart Drive MX2Secondary PredicateYamaha JWX-2 (Navione)Subject DeviceEmpulse R90Statement of Equivalence
1.3 Where UsedIndoors and outdoors;care facilities and privateresidences.Indoors and outdoors;care facilities and privateresidences.Indoors and outdoors;care facilities and privateresidences.Identical to Predicates. Nonew issues of safety oreffectiveness
2 Safety
2.1 Mechanical SafetySubject to ANSI/RESNA orISO 7176 Requirementsfor Power Wheelchairs,where applicable.Subject to ANSI/RESNA orISO 7176 Requirementsfor Power Wheelchairs,where applicable.Subject to ANSI/RESNA orISO 7176 Requirementsfor Power Wheelchairs,where applicable.Identical to Predicates. Nonew issues of safety oreffectiveness.
2.2 Electrical SafetySubject to ANSI/RESNA orISO 7176 Requirementsfor Power Wheelchairs,where applicable.Subject to ANSI/RESNA orISO 7176 Requirementsfor Power Wheelchairs,where applicable.Subject to ANSI/RESNA orISO 7176 Requirementsfor Power Wheelchairs,where applicable.Identical to Predicates. Nonew issues of safety oreffectiveness.
2.3 BiocompatibilityNo indications ofcomplianceCompliant to ISO 10993-1,-5Compliant to ISO 10993-1,-5 (See Section 15 of thissubmission for details).Identical to SecondaryPredicate. No new issuesof safety or effectiveness.
3 Overall System
3.1 Weight Capacity14 kg to 150 kg130 kg136 kgSubstantially equivalent toPrimary Predicate. No newissues of safety oreffectiveness.
3.2 Expected Product Life5 yearsNot provided5 yearsIdentical to PrimaryPredicate. No new issuesof safety or effectiveness.
4 Mechanical Enclosure and Mounting
4.1 Structural MaterialsAluminum and SteelAluminum and steelAluminum and steelIdentical to Predicates. Nonew issues of safety oreffectiveness.
4.2 Width141 mm91mm (both left and rightwheels)150mmSubstantially equivalent toPrimary Predicate. No newissues of safety oreffectiveness. Slightly
Primary PredicatePermobil (Max Mobility)Smart Drive MX2Secondary PredicateYamaha JWX-2 (Navione)Subject DeviceEmpulse R90Statement of Equivalence
4.3 Depth389mmSame as height310mmwider width does notinterfere with safeoperation of manualwheelchairs.
4.4 Height242mm609 mm (24")559 mm (22")377mm – 428mmSubstantially equivalent toPrimary Predicate. No newissues of safety oreffectiveness.
4.5 Overall Mass5.7 kg17 kg (Ni-MH)17.7 kg (Li-Ion)5.5 kg + 1.1 kg (battery)Substantially equivalent toPrimary Predicate. No newissues of safety oreffectiveness.
4.6 Manual WheelchairMountRigid: Clamp mount tothe camber tube. Driveassembly pivotally mountonto clamp.Folding: Same as rigid butwith the addition of across member thatextends horizontallybetween the two axleplates, thus simulating acamber tube.Quick Release axle mount.One wheel per side.Rigid: Receiver fixedmounted to the cambertube. Drive assembly freeto move vertically relativeto mount.Folding: Same as rigid butwith the addition of across member thatextends horizontallybetween the two axleplates, thus simulating acamber tube.Substantially equivalent toPrimary Predicate. No newissues of safety oreffectiveness.
4.7 Housing MaterialPlasticAluminumPlasticIdentical to PrimaryPredicate. No new issuesof safety or effectiveness.
5 Performance
Primary PredicatePermobil (Max Mobility)Smart Drive MX2Secondary PredicateYamaha JWX-2 (Navione)Subject DeviceEmpulse R90Statement of Equivalence
5.1 Max Speed (fwd)5.5 mphNot Specified5.5 mphIdentical to PrimaryPredicate. No new issuesof safety or effectiveness.
5.2 Max Speed (rev)N/ANot SpecifiedN/AIdentical to Predicates. Nonew issues of safety oreffectiveness.
5.3 Max Safe SlopeAs specified by manualwheelchair manufacturerIdentical to SecondaryPredicate. No new issuesof safety or effectiveness.
5.4 Maximum Range onFull Charge19.8 km20 km (NiMH)40 km (Li-ion)19 ± 1.5 kmSubstantially equivalent toPrimary Predicate. No newissues of safety oreffectiveness.
5.5 BluetoothBluetooth LENot AvailableBluetooth LE 5.2Identical to PrimaryPredicate. No new issuesof safety or effectiveness.
6 Power System
6.1 Motor Power250 W2x 110 W250 WIdentical to PrimaryPredicate. No new issuesof safety or effectiveness.
6.2 Battery TypeLi-lonNi-MH or Li-IonLi-IonIdentical to PrimaryPredicate. No new issuesof safety or effectiveness.
6.3 Battery Capacity122 Whr160 Whr (Ni-MH)283 Whr (Li-ion)100 Whr – 160 WhrSubstantially equivalent toPredicates. No new issuesof safety or effectiveness.
6.4 Battery Charger Input100-240 Vac, 50/60Hz100-240 Vac, 50/60Hz100-240 Vac, 50/60HzIdentical to Predicates. Nonew issues of safety oreffectiveness.
6.5 Battery Charger Output2.0 A2.6 A2.0 AIdentical to PrimaryPredicate. No new issues
Primary PredicatePermobil (Max Mobility)Smart Drive MX2Secondary PredicateYamaha JWX-2 (Navione)Subject DeviceEmpulse R90Statement of Equivalence
6.6 Operating Voltage36 Vdc24 Vdc36 Vdcof safety or effectiveness.Identical to PrimaryPredicate. No new issuesof safety or effectiveness.
7 Control System
7.1 Control TypeConstant power on. Stopand start of power assistis controlled by wriststrap sensor withBluetooth communicationto the drive unit.Torque sensing handrimapplies proportionalpower based on torqueapplied to the handrim.Power is pulsed on forspecified duration. Gain ofapplied power isadjustable.The device will senseacceleration (torque) andapply a power pulse to thedrive motor for a specifiedduration. Gain of appliedpower is adjustable.Substantially equivalent toPredicates. No new issuesof safety or effectiveness.
7.2 Operating Voltage36 Vdc24 Vdc36 VdcIdentical to PrimaryPredicate. No new issuesof safety or effectiveness.
7.3 Moisture ResistanceISO 7176-9ISO 7176-9ISO 7176-9Identical to Predicates. Nonew issues of safety oreffectiveness
7.4 OperatingTemperature-25°C to +50°C-25°C to +50°C-25°C to +50°CIdentical to Predicates. Nonew issues of safety oreffectiveness
7.5 Storage Temperature-40°C to +65°C-40°C to +65°C-40°C to +65°CIdentical to Predicates. Nonew issues of safety oreffectiveness
8 EMC
8.1 EMC - ImmunityCompliant to ANSI/RESNAWC2-21 or ISO 7176-21Compliant to ANSI/RESNAWC2-21 or ISO 7176-21Compliant to ANSI/RESNAWC2-21 or ISO 7176-21Identical to Predicates. Nonew issues of safety oreffectiveness.
8.2 EMC - RadiatedEmissionsCompliant to ANSI/RESNAWC2-21 or ISO 7176-21Compliant to ANSI/RESNAWC2-21 or ISO 7176-21Compliant to ANSI/RESNAWC2-21 or ISO 7176-21Identical to Predicates. Nonew issues of safety oreffectiveness
Primary PredicatePermobil (Max Mobility)Smart Drive MX2Secondary PredicateYamaha JWX-2 (Navione)Subject DeviceEmpulse R90Statement of Equivalence
8.3 EMC - ESDCompliant to ANSI/RESNAWC2-21 or ISO 7176-21Compliant to ANSI/RESNAWC2-21 or ISO 7176-21Compliant to ANSI/RESNAWC2-21 or ISO 7176-21Identical to Predicates. Nonew issues of safety oreffectiveness.
8.4 Power FrequencyMagnetic Field ImmunityCompliant to ANSI/RESNAWC2-21 or ISO 7176-21Compliant to ANSI/RESNAWC2-21 or ISO 7176-21Compliant to ANSI/RESNAWC2-21 or ISO 7176-21Identical to Predicates. Nonew issues of safety oreffectiveness.
9 User Input Controls
9.1 On/Off ButtonLocated on the drive unitLocated on the drive unitLocated on the drive unitIdentical to Predicates. Nonew issues of safety oreffectiveness.
9.3 Speed Up/Speed DownAvailable by tapping onthe wrist control.Maximum speedadjustable through App.Maximum speedadjustable on the driveunit.Speed adjustable byapplying anacceleration/retardationtorque to the manualwheelchair handrims.Substantially equivalent toPredicates. No new issuesof safety or effectiveness.
9.4 Battery state of chargeindicationLED IndicatorLED IndicatorLED IndicatorIdentical to Predicates. Nonew issues of safety oreffectiveness.
9.5 Programming ToolSmart phone appNot programmableSmart phone appIdentical to PrimaryPredicate. No new issuesof safety or effectiveness.
9.6 Bluetooth functionalityParameter settingOver the air firmwareupdatesDistance, pushes and timetrackerRoute TrackingNot availableThe Control Box connectswith the primary driveunit through a wirelessBluetooth connection. TheControl Box providessecondary speed increase,speed decrease and motorde-energization throughthe control box. TheSubstantially equivalentto the PrimaryPredicate. No newissues of safety oreffectiveness.
Primary PredicateSecondary PredicateSubject DeviceStatement of Equivalence
Permobil (Max Mobility)Smart Drive MX2Yamaha JWX-2 (Navione)Empulse R90Control Box also providesa means to togglebetween lifting anddeploying the drive wheelactuator.

Table 5.1: Table of Technological Comparison

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

{10}------------------------------------------------

{11}------------------------------------------------

{12}------------------------------------------------

SUMMARY OF NON-CLINICAL TESTING

The following tests were performed to demonstrate substantial equivalence based on current industry and FDA recognized standards.

  • Energy consumption (per ISO 7176-4)
  • Maximum speed, acceleration, and deceleration (per ISO 7176-6)
  • . Static, impact, and fatigue (per ISO 7176-8)
  • Climatic test (per ISO 7176-9)
  • Obstacle climbing (per ISO 7176-10)
  • Power and control systems for power wheelchairs (per ISO 7176-14)
  • Documentation and labeling (per ISO 7176-15)
  • EMC testing (per ISO 7176-21)
  • . Vocabulary (per ISO 7176-26)
  • Alkaline or other non-acid electrolyte batteries safety testing (IEC 62133-2)
  • Evaluation and testing within a risk management process (per ISO 10993-1)
  • Software life cycle process (per IEC 62304) ●

CONCLUSION

Based on the testing performed, it can be concluded that the subject device does not raise new issues of safety or effectiveness compared to the predicate devices. The similar indications for use, technological characteristics, and performance characteristics for the proposed Empulse R90 Power Assist Device is assessed to be substantially equivalent to the predicate devices.

§ 890.3860 Powered wheelchair.

(a)
Identification. A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.(b)
Classification. Class II (performance standards).