K Number
K220175

Validate with FDA (Live)

Device Name
GM85
Date Cleared
2022-04-21

(90 days)

Product Code
Regulation Number
892.1720
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The GM85 Digital Mobile X-ray imaging System is intended for use in generating radiographic images of human anatomy by a qualified/trained doctor or technician. This device is not intended for mammographic applications.

Device Description

The GM85 Digital Mobile X-ray Imaging System is used to capture images by transmitting X-ray to a patient's body. The X-ray passing through a patient's body is sent to the detector and then converted into electrical signals. These signals go through the process of amplification and digital data conversion in the signal process on the S-Station, which is the Operation Software (OS) of Samsung Digital Diagnostic X-ray System, and save in DICOM file, a standard for medical imaging. The captured images are tuned up by an Image Post-processing Engine (IPE) which is exclusively installed in S-Station, and send to the Picture Archiving & Communication System (PACS) sever for reading images.

The GM85 Digital Mobile X-ray imaging System was previously cleared with K181626, and through this premarket notification, we would like to add more configurations in the previously cleared GM85 as an optional collapsible column type with a manual collimator, a tube, four detectors, and exposure switches (two wired and one wireless types) are optionally added, and software including the Image Post-processing Engine (IPE) is changed in order to support new hardware and apply new software features.

AI/ML Overview

The provided document is a 510(k) Premarket Notification for the Samsung GM85 Digital Mobile X-ray Imaging System. It describes the updated device and compares it to a legally marketed predicate device (K181626, also GM85). The document primarily focuses on demonstrating substantial equivalence rather than presenting detailed acceptance criteria and a comprehensive study for de novo device performance.

However, based on the information provided, here's an attempt to extract and synthesize what is available regarding acceptance criteria and the supporting study, focusing on the changes made to the device:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly present a table of acceptance criteria with numerical targets and corresponding device performance for the overall system in a typical format found in clinical studies. Instead, it describes differences and then asserts that these differences do not "contribute any adverse impact to the device's safety and effectiveness" or "do not impact safety and/or effectiveness."

The closest to acceptance criteria are the statements made about the new detectors and software features, aiming for equivalent or improved performance compared to the predicate device.

Aspect of Performance/CharacteristicAcceptance Criteria (Implicit)Reported Device Performance (Implicit)
Detectors: Image Characteristics (MTF, DQE)Equivalent to existing (predicate) detectors, without adverse impact on safety and diagnostic effectiveness.New detectors have "equivalent image characteristics as the existing ones." MTF is "slightly higher." DQE is "similar" or "a little lower" but "do not contribute any adverse impact to the device's safety and diagnostic effectiveness."
Detectors: Spatial Resolution (Pixel Pitch, High Contrast Limiting Resolution, Number of pixels)No adverse impact on safety and/or effectiveness.New detectors have "same or lower pixel pitch," "same or higher pixel number," and "same or higher resolution" compared to predicate. These "do not impact safety and/or effectiveness."
Detectors: Mechanical/Environmental (Dust/Water-resistance, Max. load capacity)No adverse impact on safety and effectiveness.New detectors have "same or better dust/water-resistance" and "same or higher max load capacity" than predicate. These changes "do not contribute any adverse impact to the device's safety and effectiveness."
SimGrid (Image Processing Software)No adverse impact on the device's safety and effectiveness. Improved functionality.Updated SimGrid provides a parameter for controlling strength, which "does not contribute any adverse impact to the device's safety and effectiveness."
IPE (Image Post-processing Engine)No impact to the device's safety and effectiveness.Upgraded IPE (Clinical Parameter Control) allows simultaneous comparison of editing image with current image, and this "does not contribute any impact to the device's safety and effectiveness."
Overall System (Clinical Equivalence)Equivalent to the predicate device.Phantom image evaluations were found to be "equivalent to the predicate devices." "There is no significant difference in the average score of image quality evaluation between the proposed device and the predicate device."

2. Sample Size for Test Set and Data Provenance

  • Sample Size for Test Set: Not explicitly stated as a numerical sample size for "cases" in the traditional sense of a clinical study. The document mentions "Anthropomorphic phantom images were provided." The number of phantom images or specific phantom types isn't detailed. For software elements, "Software System Test Case for verification and validation" was performed, but no numerical count is given.
  • Data Provenance: The study used "Anthropomorphic phantom images" and non-clinical data (MTF, DQE measurements, Software System Test Cases). This indicates testing in a controlled environment (laboratory/phantom studies). The country of origin for the data is not specified, but the manufacturer is based in the Republic of South Korea. The data is non-clinical/phantom-based, so it's not strictly "retrospective or prospective" in the human clinical trial sense.

3. Number of Experts and Qualifications for Ground Truth

  • Number of Experts: "A professional radiologist" (singular)
  • Qualifications of Experts: "professional radiologist" (no specific years of experience or subspecialty mentioned beyond being a radiologist).

4. Adjudication Method for the Test Set

  • Adjudication Method: Not applicable or not specified. A single "professional radiologist" evaluated the phantom images, implying no consensus or multi-reader adjudication process as there was only one reviewer mentioned.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • MRMC Study: No, a multi-reader multi-case (MRMC) comparative effectiveness study was not explicitly stated or described. The evaluation involved a single radiologist reviewing phantom images.
  • Effect Size: Not applicable, as no MRMC study was conducted.

6. Standalone (Algorithm only without human-in-the-loop performance) Study

  • Standalone Study: Yes, performance data such as MTF and DQE measurements (IEC 62220-1) are inherent standalone technical performance metrics of the detectors. Software System Test Cases for verification and validation would also be considered standalone testing. The "anthropomorphic phantom images" evaluation by a single radiologist could be considered a form of standalone performance evaluation for the system's image quality output, as it assesses the device's generated images directly.

7. Type of Ground Truth Used

  • Type of Ground Truth: The ground truth for the phantom image evaluation was implicitly the expected characteristics and quality of images of anthropomorphic phantoms, as assessed by a professional radiologist for equivalence to predicate devices. For technical metrics like MTF and DQE, the ground truth is against the measurement standards (IEC 62220-1). There was no pathology or outcomes data.

8. Sample Size for the Training Set

  • Sample Size for Training Set: The document does not provide information regarding the sample size of a training set for any machine learning components. While the device includes an "Image Post-processing Engine (IPE)" and features like "SimGrid," if these involve learned algorithms, their training data is not discussed. This submission is for an X-ray system, and the focus is on hardware and general image processing functionality rather than an AI/CADe device where training data would typically be detailed.

9. How the Ground Truth for the Training Set Was Established

  • How Ground Truth for Training Set was Established: Not applicable or not provided, as training set details are absent.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

April 21, 2022

Samsung Electronics Co. Ltd. % Jaesang Noh Senior Professional. Regulatory Affairs 129, Samsung-Ro, Yeongtong-Gu Suwon-Si, Gyeonggi-Do 16677 REPUBLIC OF SOUTH KOREA

Re: K220175

Trade/Device Name: GM85 Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile X-Ray System Regulatory Class: Class II Product Code: IZL Dated: January 14, 2022 Received: January 21, 2022

Dear Jaesang Noh:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Laurel Burk Assistant Director Diagnostic X-ray Systems Team Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K220175

Device Name GM85

Indications for Use (Describe)

The GM85 Digital Mobile X-ray imaging System is intended for use in generating radiographic images of human anatomy by a qualified/trained doctor or technician. This device is not intended for mammographic applications.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

SAMSUNG

510(k) Premarket Notification - Traditional

Section 5: 510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted accordance with requirements of 21 CFR 807.92

    1. Date: April 19, 2022

2. Submitter

  • Company Name: SAMSUNG ELECTRONICS Co., Ltd. A.
  • Address: 129, Samsung-ro, Yeongtong-gu, Suwon-si, Gyeonggi-do, 16677, B. Republic of Korea

3. Primary Contact Person

  • Name: JAESANG NOH A.
  • B. Title: Regulatory Affairs, Senior Professional
  • Phone Number: +82-2-2193-2444 ﻥ
  • D. FAX Number: +82-2-2194-0272
  • E-Mail: jaesang.noh@samsung.com ய்

4. Secondary Contact Person

  • Name: Ninad Gujar A.
  • B. Title: Vice President, Regulatory Affairs & Quality Control
  • C. Phone Number: 978-564-8503
  • D. FAX Number: 978-560-0602
  • ய் E-Mail: ngujar@neurologica.com

5. Proposed Device

  • Trade Name: GM85 A.
  • Device Name: GM85 B.
  • ﻥ Common Name: Digital Diagnostic Mobile X-ray System
  • D. Classification Name: Mobile X-ray System
  • Product Code: IZL ய்
  • Regulation: 21 CFR 892.1720 ட்

6. Predicate Devices

Predicate Device
Device NameGM85
Classification NameMobileX-ray system.
Product CodeIZL
Regulation21 CFR 892.1720
510(k)#K181626
510(K) Decision DateJuly 20, 2018

{4}------------------------------------------------

510(k) Premarket Notification - Traditional

7. Device Description

The GM85 Digital Mobile X-ray Imaging System is used to capture images by transmitting X-ray to a patient's body. The X-ray passing through a patient's body is sent to the detector and then converted into electrical signals. These signals go through the process of amplification and digital data conversion in the signal process on the S-Station, which is the Operation Software (OS) of Samsung Digital Diagnostic X-ray System, and save in DICOM file, a standard for medical imaging. The captured images are tuned up by an Image Post-processing Engine (IPE) which is exclusively installed in S-Station, and send to the Picture Archiving & Communication System (PACS) sever for reading images.

The GM85 Digital Mobile X-ray imaging System was previously cleared with K181626, and through this premarket notification, we would like to add more configurations in the previously cleared GM85 as an optional collapsible column type with a manual collimator, a tube, four detectors, and exposure switches (two wired and one wireless types) are optionally added, and software including the Image Post-processing Engine (IPE) is changed in order to support new hardware and apply new software features.

The cleared GM85 (K181626) has two types of columns, collapsible and fixed column having an automatic and manual collimator respectively. A manual collimator is optionally added for a collapsible column in proposed device GM85, where a new tube is available. Four detectors are added in the proposed device GM85; two are new and the others have been cleared with K200418.

It was determined that the level of concern for the software contained in the GM85 Digital Mobile X-ray Imaging System was Moderate in accordance with the FDA guidance, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Device" and its level is the same as the predicate device.

Note that when using the Mirror View, image transmitted via Miracast is not available for the purpose of diagnosis.

8. Intended Use

The GM85 Digital Mobile X-ray Imaging System is intended for use in generating radiographic images of human anatomy by a qualified/trained doctor or technician. This device is not intended for mammographic applications.

9. Summary of Technological characteristic of the proposed device compared with the predicate devices

The proposed device, GM85, has the same technological characteristics and hardware as its original predicate device, GM85 (K181626), and added an optional collapsible column type with a manual collimator, a tube, four detectors, exposure switches (two wired and one wireless types) and new software features. They do not have significant changes in materials, energy source or technological characteristics compared to the predicate device.

Comparisons of technological characteristics were executed and demonstrate the substantial equivalence to the predicates.

  • A. Comparing with Predicate Device

{5}------------------------------------------------

510(k) Premarket Notification - Traditional

The proposed device is shown as its parts are identical or equivalent with predicate device while some differences are made as below, which do not show significant difference in safety and effectiveness.

SpecificationProposed DevicePredicate DeviceDiscussion
Device NameGM85GM85-
ManufacturerSAMSUNG ELECTRONICSSAMSUNG ELECTRONICS-
510(k) Number-K181626-
AppearancesImage: [C-Type*]Collapsiblecolumn type(C-Type)Image: [F-Type*]Fixedcolumn type(F-Type)Image: [Fit-Type*]***Collapsiblecolumn type(Fit-Type)Image: [C-Type*]Collapsible columntype(C-Type)Image: [F-Type*]**Fixed column type(F-Type)Difference(1)
Intended UseThe GM85 Digital Mobile X-rayImaging System is intended for usein generating radiographic imagesof human anatomy by aqualified/trained doctor ortechnician. This device is notintended for mammographicapplications.Same
Manufacturer ContentsGM85GM85(K181626)Discussion
C-typeF-typeFit-typeC-typeF-type
(1)HighVoltage Generator
TypeHigh FrequencyHigh Frequency-
Max. Power32kW / 40kW32kW / 40kWSame
OutputRangekVp Range40 to 150kVp40 to 150kVpSame
mA Range10 - 500mA10 - 500mASame
Exposure Time1msec-10sec1msec-10secSame
Manufacturer ContentsGM85GM85(K181626)Discussion
(2)Tube assembly
MovingRangeHorizontal793~1355mm793~1355mmSame
Vertical550~2030mm550~2030or1850mm550~2030mm550~2030mm550~2030 or1850mmSame
RotationRangeColumn±315°±315°Same
Tube(Arm axis)±180°±180°Same
Tube-30°~90°-30°~90°Same

{6}------------------------------------------------

SAMSUNG

Manufacturer ContentsGM85GM85(K181626)Discussion
(Tube axis)C-typeF-typeFit-typeC-typeF-type
TubeModelLUC-13L,XRR-3332XLUC-11L,XRR-3332XLUC-13L,XRR-3332XDifference(2)
Focal spot0.6/1.2mm0.6/1.5mm,0.6/1.2mm0.6/1.2mmDifference(3)
Target Angle14°14°Same
Target MaterialRhenium-tungsten facedmolybdenumRhenium-tungsten facedmolybdenumSame
Nominal TubeVoltage150kVp125kVp150kVpDifference(4)
Max.Anode HU300kHU200kHU300kHUSame
CollimatorSDR-OGCL40USDR-OGCL41USDR-OGCL30LSDR-OGCL40USDR-OGCL41UDifference(5)
Automatic212X306X179mmManual222X271X140mmManual206X230X139mmAutomatic212X306X179mmManual222X271X140mm
Manufacturer ContentsGM85GM85(K181626)Discussion
(3) Detector
NameS4335-WS4343-WS3025-WS4335-AWS4343-AWS4335-AWMS4343-AWMS3025-AWS3025-AWMS4335-WS4343-WS3025-WS4335-AWS4343-AWDifference(6)
S4335-AWMS4343-AWMS3025-AW
S3025-AWS3025-AWS3025-AW
S3025-AWM
Detector TypeCsIIndirectCsIIndirectSame
Detector Area14"X17"(345mmX425mm)17"X17"(425mmX425mm)10"X12"(251mmX314mm)14"X17"(345mmX425mm)17"X17"(425mmmX425mm)10"X12"(245mmmX295mm)Difference(6)-1
Number of pixels2466X30403036X30402024X 25362466X30403036X30401750X2108Difference(6)-2
Pixel Pitch(um)140124140Difference(6)-3
High Contrast LimitingResolution (LP/mm)3.574.033.57Difference(6)-4
CommunicationWired /WirelessWirelessWired / WirelessSame
Dust/Water-resistanceIP54IP67IP54Difference(6)-5
Max.load capacity(uniform load / local load,,40 mm in diameter disk atthe center)400 kg/200 kg400 kg/200 kg150 kg/100 kgDifference(6)-6
DQE(Detective Quantum65%(0lp/65%(0lp/75%(0lp/60%(0lp/76%(0lp/mm,76%(0lp/m78%(0.1lp/Difference(6)-7

{7}------------------------------------------------

SAMSUNG

Manufacturer ContentsGM85GM85(K181626)Discussion
(3) Detector
Efficiency)mm,Typical)mm,Typical)mm,Typical)mm,Typical)Typical)m,Typical)mm,Typical)
MTF(Modulation TransferFunction)88%(0.5lp/mm,Typical)88%(0.5lp/mm,Typical)87%(0.5lp/mm,Typical)90%(0.5lp/mm,Typical)86%(0.5lp/mm,Typical)86%(0.5lp/mm,Typical)86%(0.5lp/mm,Typical)Difference(6)-8
Manufacturer ContentsGM85Fit-typeGM85(K181626)Discussion
(4) GridC-typeF-typeC-typeF-type
Lines/cm84.684.6Same
Grid mechanismStationaryStationarySame
RemovabilityRemovableRemovableSame
(5) Weight Distribution Cap
Model NameSDR-OGWD80USDR-OGWD80USame
Size(mm)505x553x37.4505x553x37.4Same
(6) Exposure Switch
NameC2U-45C2UW-DS-SA01L01 4C6ML03 4C6MSDR-OGRC30KTS_CES(SAM)C2UW-LP-ISDR-OGRC30KC2U-45C2UW-DS-SA01TS_CES(SAM)C2UW-LP-IDifference (7)
TypesWired2 or 3 ButtomDeadman typeWirelessInfra-red/BluetoothWired2 or 3 ButtomDeadman typeWirelessInfra-red/BluetoothSame
(7) Software Features
S-EnhanceS-EnhanceS-EnhanceSame
Pediatric ExposureManagement(PEM)PEMPEMSame
S-DAP(Dose Area Product)S-DAPS-DAPSame
S-AlignS-AlignN/AS-AlignN/ASame
S-ShareS-ShareS-ShareSame
Bone SuppressionImage(BSI)BSIBSISame
Remote ViewRemote ViewRemote ViewSame
Manual StitchingManual StitchingManual StitchingSame
SimGridUpgraded SimGridSimGridDifference(8)
Quick LinkQuick LinkQuick LinkDifference(9)

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image contains the word "SAMSUNG" in a bold, sans-serif font. The word is written in all capital letters and is a dark blue color. The background of the image is white, which provides a strong contrast to the blue text.

Manufacturer ContentsGM85GM85(K181626)Discussion
C-typeF-typeFit-typeC-typeF-type
Mirror ViewMirror View-Difference(10)
Clinical ParameterControlClinical Parameter Control-Difference(11)
Value-up PackageValue-up Package-Difference(12)
Remote SoftwareUpgradeRemote Software Upgrade-Difference(13)
NoDifferencesExplanation
(1)AppearancesThe collapsible column type which has a manual collimator isadded to the proposed device GM85 as an option in comparisonwith the predicate device's one that has only collapsible columntype with an automatic collimator. The collapsible column type hastwo options about the collimator, automatic and manual in theproposed device GM85 and this change does not contribute anyadverse impacts to the device's safety and effectiveness.
(2)Tube ModelTwo tubes are available in the optional collapsible column of theproposed device GM85. One is the same as the predicate deviceand the other is newly added. This change does not contributeany adverse impact to the device's safety and effectiveness.
(3)Focal spotThe focal spot size of the new tube used with an optionalcollapsible column of the proposed device GM85 is bigger thanthe predicate device's focal spot size and this change does notcontribute any adverse impact to the device's safety andeffectiveness.
(4)Nominal tubevoltageThe maximum voltage of tubes used with the optional collapsiblecolumn of the proposed device GM85 is limited to lower than thepredicate device's maximum tube voltage and these changes donot contribute any adverse impact to the device's safety andeffectiveness.
(5)CollimatorThe proposed device GM85 has different types of collimatordepending on column type. The fixed column type has the samecollimator (manual, SDR-OGCL41U) applied to the predicatedevice and the collapsible column type has two collimators. Oneis the same (automatic, SDR-OGCL40U) as the predicate deviceand the other is different (manual, SDR-OGCL30L). This changedoes not contribute any adverse impacts to the device's safetyand effectiveness.
(6)DetectorsFour detectors are added to the GM85 device. Two detector(S4335-AWM and S4343-AWM) are new and the others (S3025-AW and S3025-AWM) are cleared with K200418. These changesdo not contribute any adverse impact to the device's safety andeffectiveness.
(6)-1Detector AreaThe new detectors which are added to the GM85 device have the

{9}------------------------------------------------

same (S4335-AWM and S4343-AWM) or similar (S3025-AW andS3025-AWM) dimension as those of the predicate device andthese changes do not contribute any adverse impact to thedevice's safety and effectiveness.
(6)-2Number of pixelsThe new detectors which are added to the GM85 device have thesame (S4335-AWM and S4343-AWM) or higher (S3025-AW andS3025-AWM) pixel number as those of the predicate device andthese changes do not impact safety and/or effectiveness.
(6)-3Pixel Pitch(um)The new detectors which are added to the GM85 device have thesame (S4335-AWM and S4343-AWM) or lower (S3025-AW andS3025-AWM) pixel pitch as those of the predicate device andthese changes do not impact safety and/or effectiveness.
(6)-4High ContrastLimitingResolution(LP/mm)The new detectors which are added to the GM85 device have thesame (S4335-AWM and S4343-AWM) or higher (S3025-AW andS3025-AWM) resolution as those of the predicate device andthese changes do not impact safety and/or effectiveness
(6)-5DetectorDust/Water-resistanceThe new detectors which are added to the GM85 device have thesame (S4335-AWM and S4343-AWM) or better (S3025-AW andS3022-AWM) dust/water-resistance as those of the predicatedevice and these changes do not contribute any adverse impactto the device's safety and effectiveness.
(6)-6DetectorMax.load capacityThe new detectors which are added to the GM85 device have thesame (S4335-AWM and S4343-AWM) or higher (S3025-AW andS3022-AWM) max load capacity as those of the predicate deviceand these changes do not contribute any adverse impact to thedevice's safety and effectiveness.
(6)-7DQE (DetectiveQuantumEfficiency)The new detectors which are added to the GM85 device havesimilar (S3025-AW) or a little lower (S4335-AWM, S4343-AWM,and S3025-AWM) DQE as those of the predicate device but thesechanges do not contribute any adverse impact to the device'ssafety and diagnostic effectiveness.
(6)-8MTF (ModulationTransferFunction)The new detectors which are added to the GM85 device haveslightly higher (S4335-AWM, S4343-AWM, S3025-AW andS3025-AWM) MTF as those of the predicate device and thesechanges do not contribute any adverse impact to the device'ssafety and diagnostic effectiveness.
(7)Exposure SwitchTwo wired and one wireless exposure switches is added to theproposed device GM85 as an option, which are from differentmanufacturers in comparison with those of the predicate device,and these changes do not contribute any adverse impact to thedevice's safety and effectiveness.
(8)SimGridThe SimGrid in the predicate device, cleared with K181626, is anadditional image processing software option which is able tocompensate the contrast loss due to scatter radiations, primarilyacquisitions without a physical anti-scatter grid of all kinds ofanatomical regions The SimGrid modified to provide a parameter
(9)Quick Linkfor controlling strength which is the extent of the compensation of the contrast loss based on the individual preference is available in the proposed device GM85 and this change does not contribute any adverse impact to the device's safety and effectiveness. The function of Quick Link, which allows Quick Link to access to external server such as RIS/PACS server, is applied to the proposed device GM85 and this change does not contribute any impact to the device's safety and effectiveness.
(10)Mirror ViewThe function of the Mirror View, which shares the screen with other displays by using Miracast wireless communication, is applied to the proposed device GM85 and this change does not contribute any impact to the device's safety and effectiveness.
(11)ClinicalParameterControlThe captured images are tuned up by the preset value of an Image Post-processing Engine (IPE) in the predicate device and it is upgraded in the proposed device GM85 to provide the function, called as Clinical Parameter Control, to compare the editing image with the current image simultaneously during presetting the parameters. This change does not contribute any impact to the device's safety and effectiveness.
(12)Value-upPackageNew software features which is an option to provide convenience for use, called as the Value-up Package, is applied to the proposed device GM85 and these changes do not contribute any impact to the device's safety and effectiveness.
(13)Remote SoftwareUpgradeThe function of the Remote Software Upgrade, which enable the authorized user to download and update the software from the server by the user authentication, is applied the proposed device GM85 and this change does not contribute any impact to the device's safety and effectiveness

{10}------------------------------------------------

510(k) Premarket Notification - Traditional

B. Safety, EMC and Performance Data

Electrical, mechanical, environmental safety and performance testing according to standard ES 60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-2-28, IEC 60601-2-54, ISO14971, 21CFR1020.30 and 21CFR1020.31 were performed, and EMC testing was conducted in accordance with standard IEC 60601-1-2. Wireless function was tested and verified followed by guidance, Radio frequency Wireless Technology in Medical Devices. All test results were satisfying the standards.

C. Non-clinical data

Non-clinical data was provided in conformance to the FDA "Guidance for the Submission of 510(k)'s for Solid-State X-ray Imaging Devices", which includes MTF and DQE measurements as tested by IEC 62220-1.

The proposed device has new detectors that have equivalent image characteristics as the existing ones. Specific description is added to make it clear with the non-clinical data and phantom image evaluation report. And those detectors evaluated by Software

{11}------------------------------------------------

510(k) Premarket Notification - Traditional

System Test Case for verification and validation.

D. Clinical data

In clinical data, phantom image evaluations of the new detectors, upgraded SimGrid and Image Post-processing Engine (IPE) were performed in accordance with FDA guidance for the submission of 510(k)'s for Solid State X-ray Imaging Devices. Anthropomorphic phantom images were provided; these images were not necessary to establish substantial equivalence based on the modifications to the device (note X-ray flat-panel detector similar to the predicate detector) but they provide further evidence in addition to the performance data to show that the complete system works as intended. They were evaluated by a professional radiologist and found to be equivalent to the predicate devices. There is no significant difference in the average score of image quality evaluation between the proposed device and the predicate device. Therefore, these changes do not affect either the safety or the effectiveness, compared to the predicate device.

E. Summary of the Standards and Guidance Compliance

  • AAMI ANSI ES60601-1 2005/(R)2012 and A1:2012, C1:2009/(R)2012 and 1. A2:2010/(R)2012 Medical Electrical Equipment – Part 1: General requirements for basic safety and essential performance
    1. IEC 60601-1-2 Edition 4 Medical Electrical Equipment - Part 1-2: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Electromagnetic compatibility - Requirements and Tests
    1. IEC 60601-1-3 Edition 2.1 Medical Electrical Equipment - Part 1-3: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Radiation protection in Diagnostic X-ray Equipment
    1. IEC 60601-2-28 Edition 3 Medical Electrical Equipment – Part 2-28: Particular Requirements for the Basic Safety and Essential Performance of X-ray Tube Assemblies for Medical Diagnosis
    1. IEC 60601-2-54 Edition 1.2 Medical Electrical Equipment – Part 2-54: particular Requirements for the Basic Safety and Essential Performance of X-ray Equipment for Radiography and Radioscopy
    1. IEC 62220-1-1 Edition 1.0 Medical electrical Equipment - Characteristics of digital X-ray imaging devices Part 1-1: Determination of the detective quantum efficiency Detectors used in radiographic imaging
  • Content of Premarket Submissions for Management of Cybersecurity in Medical 7. Devices Guidance for Industry and Food and Drug Administration Staff issued on October 2, 2014
    1. Guidance for Industry and FDA Staff Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices issued on May 11, 2015
  • Guidance for the Submission for 510(k) for Solid State X-ray Imaging Devices 9. Guidance for Industry and Food and Drug Administration Staff issued on September 1, 2016
    1. Pediatric Information for X-ray Imaging Device Premarket Notifications Guidance for Industry and Food and Drug Administration Staff issued on November 28, 2017

{12}------------------------------------------------

510(k) Premarket Notification - Traditional

F. Conclusions

The non-clinical and phantom evaluation data demonstrate that the proposed device is as safe, as effective, and performs as well as the legally marketed predicate devices.

§ 892.1720 Mobile x-ray system.

(a)
Identification. A mobile x-ray system is a transportable device system intended to be used to generate and control x-ray for diagnostic procedures. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.