K Number
K220152

Validate with FDA (Live)

Date Cleared
2022-04-19

(90 days)

Product Code
Regulation Number
882.5890
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

SM9089:
TENS(3,4,5,6,8,15,19,21):
To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities.
PMS(1,2,7,9,10,11,12,13,14,16,17,18,20,22,23,24):
It is intended to be used to stimulate healthy muscles in order to improve and facilitate muscle performance.

SM9587:
TENS(3.4.5.6 8.11.15.16.19.21.22):
To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities.
PMS(1,2,7,9,10,12,13,14,17,18,20,23,24):
It is intended to be used to stimulate healthy muscles in order to improve and facilitate muscle performance.

SM9592:
TENS(2,3,4,5,7,9,10,19,21,22,24):
To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities.
PMS(1,6,8,11,12,13,14,15,16,17,18,20,23):
It is intended to be used to stimulate healthy muscles in order to improve and facilitate muscle performance.

SM9555:
TENS(3-6, 13,14,19-21,23):
To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities.
PMS(1,2,7-12,15-18,22,24):
It is intended to be used to stimulate healthy muscles in order to improve and facilitate muscle performance.

Device Description

Not Found

AI/ML Overview

The provided document is an FDA 510(k) clearance letter for Transcutaneous Electrical Nerve Stimulators (TENS) and Powered Muscle Stimulators (PMS) devices. It confirms that the devices are substantially equivalent to legally marketed predicate devices.

However, this document does not contain information about the acceptance criteria or a study that proves the device meets specific acceptance criteria. The letter primarily addresses regulatory clearance based on substantial equivalence to existing devices, general controls provisions of the Federal Food, Drug, and Cosmetic Act, and specifies the Indications for Use for various models.

For a device's performance to be evaluated against acceptance criteria, a clinical study or performance testing data would typically be required to be referenced in the submission. This information is usually found in the 510(k) summary or the detailed technical documentation submitted to the FDA, which is not part of this public clearance letter.

Therefore, I cannot provide the requested details using only the provided text. The document is a regulatory approval, not a performance study report.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.

April 19, 2022

Hong Qiangxing (Shen Zhen) Electronics Limited % Doris Dong General manager Shanghai CV Technology Co., Ltd. Room 903, No. 19 Dongbao Road, Songjiang Area, Shanghai, Shanghai 201613 China

Re: K220152

Trade/Device Name: TENS & PMS (Model:SM9089, SM9587, SM9592, SM9555) Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous electrical nerve stimulator for pain relief Regulatory Class: Class II Product Code: NUH, NGX Dated: January 11, 2022 Received: January 19, 2022

Dear Doris Dong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

Page 2

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Pamela Scott Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K220152

Device Name

TENS & PMS (Model: SM9089, SM9587, SM9592, SM9555)

Indications for Use (Describe) SM9089:

TENS(3,4,5,6,8,15,19,21):

To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities. PMS(1,2,7,9,10,11,12,13,14,16,17,18,20,22,23,24):

It is intended to be used to stimulate healthy muscles in order to improve and facilitate muscle performance.

SM9587:

TENS(3.4.5.6 8.11.15.16.19.21.22):

To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities. PMS(1,2,7,9,10,12,13,14,17,18,20,23,24):

It is intended to be used to stimulate healthy muscles in order to improve and facilitate muscle performance.

SM9592:

TENS(2,3,4,5,7,9,10,19,21,22,24):

To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities. PMS(1,6,8,11,12,13,14,15,16,17,18,20,23):

It is intended to be used to stimulate healthy muscles in order to improve and facilitate muscle performance.

SM9555:

TENS(3-6, 13,14,19-21,23):

To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities. PMS(1,2,7-12,15-18,22,24):

It is intended to be used to stimulate healthy muscles in order to improve and facilitate muscle performance.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

|X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

§ 882.5890 Transcutaneous electrical nerve stimulator for pain relief.

(a)
Identification. A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.(b)
Classification. Class II (performance standards).