K Number
K220101

Validate with FDA (Live)

Device Name
Pulse Oximeter
Date Cleared
2023-03-08

(420 days)

Product Code
Regulation Number
870.2700
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Pulse Oximeter is a handheld non-invasive device intended for spot-checking of oxygen saturation of arterial hemoglobin (SpO2) and Pulse Rate of adult, adolescent, child and infant patients in hospital-type facilities and homecare. The device is not intended to be used under motion and low perfusion. It is designed for finger thickness between 0.8cm and 2.2cm (0.3 inches to 0.9 inches).

Device Description

The proposed device, Pulse Oximeter, is a battery powered device, which can mainly detect and display the measured oxyhemoglobin saturation (SpO2) and pulse rate (PR) value. It available two models, MD300CN130 and MD300CN356. The model MD300CN130 is adopted LED screen to display SpO2 and Pulse Rate (PR) value, low power indication and pulse bar, the device has 2 display modes. The model MD300CN356 using LCD screen with 7 display modes, it can display SpO2, Pulse Rate (PR), Perfusion Index (PI), waveform, battery indicator, signal indicator and pulse bar, the brightness level can be adjusted 1-10 level.

The proposed device consists of power supply module, detector and emitter LED, signal collection and process module, display module, user interface and button control circuit.

The enclosure of the proposed device is made of ABS and the fingertip cushion is made of Silicone Gel.

The proposed device is not for life-supporting or life-sustaining, not for implant.

The device is not sterile, and the transducers are reusable and do not need sterilization.

The device is for prescription.

The device does not contain drug or biological products.

AI/ML Overview

Here's an analysis of the acceptance criteria and study detailed in the provided document:

1. Table of Acceptance Criteria and Reported Device Performance

The FDA clearance document indicates that the device's performance was compared to its predicate device and to standards. For SpO2 and PR accuracy, the acceptance criteria are generally established by standards like ISO 80601-2-61.

Feature / MetricAcceptance Criteria (from predicate/standards)Reported Device Performance (from studies)
SpO2 Display Range0%~100%0%~100%
SpO2 Measurement Range70%~100%70%~100%
SpO2 Accuracy (70%~100%)± 2%70%~100%, ± 2% (implied by meeting standards)
SpO2 Accuracy (0~69%)No definition0~69% no definition
SpO2 Resolution1%1%
PR Display Range30 bpm~250 bpm30 bpm~250 bpm
PR Measurement Range30 bpm~250 bpm30 bpm~250 bpm
PR Accuracy (30 bpm~99 bpm)± 2 bpm30 bpm~99 bpm, ± 2 bpm (implied by meeting standards)
PR Accuracy (100 bpm~250 bpm)± 2%100 bpm~250 bpm, ± 2% (implied by meeting standards)
PR Resolution1 bpm1 bpm
BiocompatibilityMeet ISO 10993-1, -5, -10Tests confirm meeting requirements (Section 5, Analysis 9)
Electrical SafetyConformed to IEC60601-1, IEC 60601-1-11Conformed (Section 7, Non-Clinical Test Conclusion)
EMCConformed to IEC60601-1-2Conformed (Section 7, Non-Clinical Test Conclusion)
Software ValidationCompliance with FDA Guidance for Software in Medical DevicesCompliance (Section 7, Non-Clinical Test Conclusion)
Operational Temp (MD300CN130)0℃~40℃0℃~40℃ (implied by meeting standard ISO 80601-2-61)
Operational Temp (MD300CN356 & Predicate)5°C~40°C5°C~40°C (implied by meeting standard ISO 80601-2-61)

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: 11 healthy adult volunteer subjects.
  • Data Provenance: The study was a prospective clinical study conducted to validate SpO2 accuracy. The country of origin is not explicitly stated, but the manufacturer is based in Beijing, China, suggesting the study likely occurred in China or an affiliated location. The subjects had "light to dark pigmentation, include male and female."

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

The document does not explicitly state the number of experts used to establish ground truth or their qualifications. However, the ground truth was established by "arterial blood co-oximeter," which is a laboratory instrument. This implies expert handling and interpretation of the co-oximeter results, but no specific human expert qualifications are provided for the ground truth establishment itself. Clinical personnel would have been involved in drawing blood samples and operating the co-oximeter.

4. Adjudication Method for the Test Set

Not applicable. This was a quantitative measurement comparison study against a reference standard (co-oximeter), not a qualitative assessment requiring adjudication of interpretations.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not done. The study focused on the standalone performance of the pulse oximeter against a clinical reference standard. It did not involve human readers interpreting results with or without AI assistance.

6. If a Standalone Study Was Done

Yes, a standalone study was done. The clinical study described in Section 7 (Clinical Test) was conducted to evaluate the SpO2 accuracy performance of the proposed pulse oximeter itself, without human-in-the-loop interaction for interpretation, other than operating the device.

7. The Type of Ground Truth Used

The ground truth used was outcomes data / reference standard measurement from an "arterial blood co-oximeter." This is considered a gold standard for measuring arterial oxygen saturation (SaO2).

8. The Sample Size for the Training Set

The document does not provide information regarding a training set or its size. As this is a traditional medical device (pulse oximeter) and not explicitly a machine learning/AI device in the context of image analysis or complex pattern recognition requiring a training set, such information would typically not be present in its 510(k) submission. Pulse oximeters operate based on established physiological principles and algorithms, rather than training on large datasets in the way modern AI models do.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as no training set information is provided or implied for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 8, 2023

Beijing Choice Electronic Technology Co., Ltd. Haiying Zhao Quality Director 2nd Floor 3rd Floor and Room 410-412 4th Floor No. 2 Building No. 9 Shuangyuan Road Shijingshan Beijing, Beijing 100041 China

Re: K220101

Trade/Device Name: Pulse Oximeter Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DOA Dated: February 21, 2023 Received: March 2, 2023

Dear Haiying Zhao:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

James J. Lee -S

James J. Lee, Ph.D. Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K220101

Device Name Pulse Oximeter

Indications for Use (Describe)

The Pulse Oximeter is a handheld non-invasive device intended for spot-checking of oxygen saturation of arterial hemoglobin (SpO2) and Pulse Rate of adult, adolescent, child and infant patients in hospital-type facilities and homecare. The device is not intended to be used under motion and low perfusion. It is designed for finger thickness between 0.8cm and 2.2cm (0.3 inches to 0.9 inches).

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Section II 510(k) Summary

This summary of 510(k) is being submitted in accordance with the requirements of 21 CFR 807.92.

There is no prior submission for the device.

Submitter Information 1.

Manufacturer

Establishment Registration Number: 3005569927

Beijing Choice Electronic Technology Co., Ltd.

2nd Floor, 3rd Floor and Room 410-412 4th Floor, No. 2 Building, No. 9 Shuangyuan Road, Shijingshan District BEIJING, 100041, P.R. China.

Contact Person

Haiying Zhao

Beijing Choice Electronic Technology Co., Ltd.

2nd Floor, 3rd Floor and Room 410-412 4th Floor, No. 9 Shuangyuan Road, Shijingshan District BEIJING, 100041, P.R. China.

Phone: +86-10-88204631

Fax: 861088204632

Email: cc@choicemmed.com

Date prepared: November 10, 2021

2. Identification of Proposed Device

Device Common Name: Oximeter

Device Trade/Proprietary Name: Pulse Oximeter

Model: MD300CN130, MD300CN356

Classification Name: Oximeter

Regulation Number: 870.2700

Product Code: DQA

Regulatory Class: Class II

Panel: Anesthesiology

3. Device Description

The proposed device, Pulse Oximeter, is a battery powered device, which can mainly detect and display the measured oxyhemoglobin saturation (SpO2) and pulse rate (PR) value. It available two models, MD300CN130 and MD300CN356. The model MD300CN130 is adopted LED screen to display SpO2 and Pulse Rate (PR) value, low power indication and pulse bar, the device has 2 display modes. The model MD300CN356 using LCD screen with 7 display modes, it can display SpO2, Pulse Rate (PR), Perfusion Index (PI), waveform, battery indicator, signal indicator and pulse bar, the brightness level can be adjusted

{4}------------------------------------------------

1-10 level.

The proposed device consists of power supply module, detector and emitter LED, signal collection and process module, display module, user interface and button control circuit.

The enclosure of the proposed device is made of ABS and the fingertip cushion is made of Silicone Gel.

The proposed device is not for life-supporting or life-sustaining, not for implant.

The device is not sterile, and the transducers are reusable and do not need sterilization.

The device is for prescription.

The device does not contain drug or biological products.

4. Identification of Predicate Device

Trade/Proprietary Name: Fingertip Pulse Oximeter

Model: MD300CI218

Common Name: Pulse Oximeter

K-number: K181503

Regulation Number: 21 CFR 870.2700

Device Class: II

Product Code: DQA

Panel: Anesthesiology

Manufactured by: Beijing Choice Electronic Technology Co., Ltd.

{5}------------------------------------------------

ડ. Substantially Equivalent (SE) Comparison

Table 1 Comparison Table between the Proposed Device and Predicate Device

Comparison ElementsProposed DevicePredicate Device(K181503)Remark
MD300CN130MD300CN356
Product NamePulse OximeterPulse OximeterFingertip Pulse Oximeter/
ModelMD300CN130MD300CN356MD300CI218/
Regulation No.21 CFR 870.270021 CFR 870.270021 CFR 870.2700Same
Regulatory ClassIIIIIISame
Classification NameOximeterOximeterOximeterSame
Product CodeDQADQADQASame
Intended UseThe Pulse Oximeter is a handheld non-invasive device intended for spot-checking of oxygen saturation of arterial hemoglobin (SpO2) and Pulse Rate of adult, adolescent, child and infant patients in hospitals, hospital-type facilities and homecare. The device is not intended to be used under motion and low perfusion. It is designed for finger thickness between 0.8cm and 2.2cm (0.3 inches to 0.9 inches).The Fingertip Pulse Oximeter MD300CI218 is a handheld non-invasive device intended for spot-checking of oxygen saturation of arterial hemoglobin (SpO2) and Pulse Rate of adult, adolescent, child and infant patients in hospitals, hospital-type facilities and homecare environment.Similar 1
Intended PopulationAdult, adolescent, child and infant patientsAdult, adolescent, child and infant patientsSame
ComponentsPower supply module, detector and emitter LED, signal collection and processor module, display module, user interface and button control.Power supply module, detector and emitter LED, signal collection and processor module, display module, Bluetooth module, user interface and button control.Similar, Analysis 2
Design PrincipleThe pulse oximeter works by applying a sensor to a pulsating arteriolar vascular bed. The sensor containsThe pulse oximeter works by applying a sensor to a pulsating arteriolar vascularSame
Comparison ElementsProposed DevicePredicate Device(K181503)Remark
MD300CN130MD300CN356
a dual light source and photo detector. The onewavelength of light source is 660nm, which is redlight; the other is 905nm, which is infrared-red light.Skin, bone, tissue and venous vessels normallyabsorb a constant amount of light over time. Thephoto detector in finger sensor collects and convertsthe light into electronic signal which is proportionalto the light intensity. The arteriolar bed normallypulsates and absorbs variable amounts of light duringsystole and diastole, as blood volume increases anddecreases. The ratio of light absorbed at systole anddiastole is translated into an oxygen saturationmeasurement. This measurement is referred to asSpO2.bed. The sensor contains a dual lightsource and photo detector. The onewavelength of light source is 660nm,which is red light; the other is 905nm,which is infrared-red light. Skin, bone,tissue and venous vessels normally absorba constant amount of light over time. Thephoto detector in finger sensor collectsand converts the light into electronicsignal which is proportional to the lightintensity. The arteriolar bed normallypulsates and absorbs variable amounts oflight during systole and diastole, as bloodvolume increases and decreases. The ratioof light absorbed at systole and diastole istranslated into an oxygen saturationmeasurement. This measurement isreferred to as SpO2.
MeasurementRed660± 3nm660± 3nm
WavelengthInfrared905± 10nm905± 10nm
DesignDisplay TypeLEDLCDOLEDDifferent □ Analysis 3
FeaturesWorking Timecontinuously operatedas long as 18 hourscontinuously operated aslong as 20 hourscontinuously operated as long as 18 hoursSimilar □ Analysis 4
Display Mode2 display modes7 display modes2 display modesSimilar, Analysis 5
Power supply2*AAA alkaline batteries2*AAA alkaline batteriesSame
Performance DataSpO2. PRSpO2. PR. PISpO2. PRSimilar, Analysis 6
Comparison ElementsProposed DevicePredicate Device(K181503)Remark
MD300CN130MD300CN356
SpO2SpO2 Display Range0%~100%0%~100%
SpO2 Measurement Range70%~100%70%~100%
SpO2 Accuracy70%100%, ± 2%;069% no definition70%100%, ± 2%;069% no definition
SpO2 Resolution1%1%
PRPR Display Range30 bpm~250 bpm30 bpm~250 bpm
PR Measurement Range30 bpm~250 bpm30 bpm~250 bpm
PR Accuracy30 bpm99 bpm, ± 2 bpm;100 bpm250 bpm, ± 2%30 bpm99 bpm, ± 2 bpm;100 bpm250 bpm, ± 2%
PR Resolution1 bpm1 bpm
PIDisplay rangeNA0.1%~20%NA
Measure rangeNA0.3~20.0%NA
ResolutionNA0.1%NA
TransmitterNABluetooth Compliance: Version 4.0Similar, Analysis 7
EnvironmentOperating Temperature0℃~40℃5°C~40°CSimilar, Analysis 8
Storage/Transport temperature-25℃~70℃-25℃~70℃
Relative Humidity15%~93% no condensation in operation;<=93% no condensation in storage/transport15%~93% no condensation in operation;<=93% no condensation in storage/transport
Atmosphere Pressure70kPa~106kpa70kPa~106kpa
Comparison ElementsProposed DevicePredicate Device(K181503)Remark
MD300CN130MD300CN356
ContactingMaterialBattery CoverABSABSDifferent 9
EnclosureABSABS
Screen protectfilmPMMAPMMA
Power ButtonSilica gelABS
FingertipCushionSilica gelSilica gel
TestingLaboratoryTestingThe laboratory tests include SpO2 PR and PI (onlyapply for MED300CN356) accuracy Test, WeakPerfusion Test, High and Low Temperature andHumidity Test, Performance Test After Cleaning andISO 80601-2-61The laboratory tests include SpO2 and PRaccuracy Test, Weak Perfusion Test, Highand Low Temperature and Humidity Test,Performance Test After Cleaning and ISO80601-2-61Similar, Analysis 10
Electrical SafetyConformed to IEC60601-1, IEC 60601-1-11Conformed to IEC60601-1, IEC 60601-1-11
ElectromagneticCompatibilityConformed to IEC60601-1-2Conformed to IEC60601-1-2
SoftwareCompliance with FDA Guidance for the content ofPremarket Submissions for Software Contained inMedical DevicesCompliance with FDA Guidance for thecontent of Premarket Submissions forSoftware Contained in Medical Devices
Comparison ElementsProposed DevicePredicate Device(K181503)Remark
MD300CN130MD300CN356
Label and LabelingCompliance with the Guidance of pulse oximeter-premarket notification issued on March 4,2013Compliance with the Guidance of pulse oximeter-premarket notification issued on March 4,2013Same

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

Analysis 1 - Intended Use

The intended use of proposed device is similar with the ntended finction and population of proposed device is the same. The intended use of proposed device provided more detailed description of the better understand for home use. This difference does not affect the substantially equivalence between proposed device on safety and effectiveness.

The intended use

Analysis 2 - Components

The components of proposed device are similar with does not have the Bluetooth module. The Bluetooth module that does not affect other function of device. In addition of proposed device have been verification and the results ould meet the requirement of IEC 60601-1-11, IEC60601-1-2 and ISO 80601-2-61, Submissions for Software Contained in Medical Devices. Therefore, this difference does not affect substantially equivalence between proposed device on safety and effectiveness.

Analysis 3 - Display Type

The proposed device has the different display type with the mode MD300CN 30 of proposed device is configurated with the LED display, the mode MD300CN356 are configurated with the predicate device is using the OLED display. The varies display type is due to different marked strategy. In addition, the final products have validation and the result could meet the requirement of IEC 60601-1-11, IEC60601-1-2 and ISO 80601-2-61. Therefore, this diffect substantially equivalence between proposed device and predicate device on safety and effectiveness.

Analysis 4 - Working Time

The working time of proposed device is different from the streent hardware and software design of each model. However, the final products have been verification and the result could meet the requirement of IEC 60601-1-11, IEC60601-1-2 and ISO 8001-2

{10}------------------------------------------------

  1. Therefore, this difference does not affect substantially equivalence between proposed device on safety and effectiveness.

Analysis 5 - Display Mode

The display mode of the proposed device is different from the varies display modes more choices for consumer. The display mode is depended on the display type and embedded software. However, the final products have been verification and the requrenent of ISO 80001-2-61. In addition, the embeded software could meet the requirements of Software Contained in Medical Devices. This difference does not affect substantially equivalence between proposed device on safety and effectiveness.

Analysis 6 - Performance Data

The mode MD300CN130 of proposed device can measure that is same with the predicate device. Meanwhile, the mode MD300CN356 has on more PI performance data compare with the parameters of SpO2 and PR for proposed device are totally same with the predicate device. The PI finction had been ver ISO 80601-2-61 and the result could meet the requirement. Therefore, this difference does not affect substantially equivalence between proposed device and predicate device on safectiveness.

Analysis 7 - Transmitter

The proposed device does not have the Bluetont from predicate device. The Bluetooth is the independent function module, it will not affect other functions. In addition, the embeded software could meet the requirements of Submissions for Software Contained in Medical Devices. Therefore, this difference does not affect substantially equivalence between proposed device and predicate device on safectiveness.

Analysis 8 - Environment

The Operaing Temperature of proposed device and other environment requirements are same. The lover Imit operating temperature of proposed device is 0°C which the peating temperature of proposed device has been verification according to standard ISO 80601-2-61. All the results candard requirements. Therefore, this difference does not affect substantially equivalence between proposed device and predicate device on safety and effectiveness.

Different 9 - Contacting Material

The contact materials of proposed device are different with the contact materials of the proposed device have been done the biocompatibility test per ISO 1099-1 Biological evaluation of medical devices — Part 1: Evaluation and testing within a risk management process and the results

{11}------------------------------------------------

can meet the standard requirements. Therefore does not affect substantially equivalence between proposed device and predicate device on safety and effectiveness.

Analysis 10 – Testing

The mode MD300CN356 have one more PI function compare with the predicate device. The PI function had been verified through the bench test per ISO 80601-2-61. The test result can meet the standard. Therefore, this difference does not affect the substantially equivalence between the proposed device and predicate device on safety and effectiveness.

{12}------------------------------------------------

Intended Use 6.

The Pulse Oximeter is a handheld non-invasive device intended for spot-checking of oxygen saturation of arterial hemoglobin (SpO2) and Pulse Rate of adult, adolescent, child and infant patients in hospitals, hospital-type facilities and homecare. The device is not intended to be used under motion and low perfusion. It is designed for fingers between 0.8cm and 2.2cm (0.3 inches to 0.9 inches).

7. Test Summary

Non-Clinical Test Conclusion

Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards and guidance:

  • IEC60601-1: 2005, AMD1:2012, AMD2:2020 Medical Electrical Equipment – Part 1: General Requirements for basic safety and essential performance
  • IEC60601-1-11: 2015, AMD1:2020 Medical Electrical Equipment-Part 1-11: General Requirements for basic safety and essential performance- Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.
  • IEC60601-1-2: 2020 Medical Electrical Equipment Part 1-2: General Requirements for Basic Safety and Essential Performance Collateral Standard: Electromagnetic disturbances - Requirements and tests.
  • ISO 80601-2-61:2017 Medical electrical equipment - Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment.
  • The Software Validation is in compliance with FDA Guidance to Compliance with FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices.
  • ISO10993-5:2009 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
  • ISO10993-10:2010 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization

Clinical Test

A clinical study was conducted per the requirement of Annex EE of ISO 80601-2-61 to validate the SpO2 accuracy of proposed pulse oximeter. The purpose of the clinical study was to evaluate the SpO2 accuracy performance of the pulse oximeter during stationary (non-motion) conditions over a wide range of arterial blood oxygen saturation levels as compared to arterial blood co-oximeter. 11 healthy adult volunteer subjects (ages 20-42yr, with light to dark pigmentation, include male and female) were included in the study conducted to evaluate the SpO2 accuracy performance of proposed three models. Each system was evaluated during steady state/non-motion conditions with various levels of induced hypoxia resulting in stable oxygen saturation levels between 100% and 70% SaO2. Arterial blood samples were drawn during simultaneous data collection from the test devices. The blood was immediately analyzed on reference cooximeter providing functional SaO2 for the basis of the SpO2 accuracy comparison. The results of the testing demonstrate that all requirements and performance specifications were satisfied and the subject device is substantially equivalent to its predicates.

{13}------------------------------------------------

Conclusion 8.

The conclusions drawn from the nonclinical tests demonstrate that the proposed subject device is as safe, as effective, and performs as well as the legally marketed predicate device.

§ 870.2700 Oximeter.

(a)
Identification. An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.(b)
Classification. Class II (performance standards).