K Number
K220099

Validate with FDA (Live)

Manufacturer
Date Cleared
2022-10-20

(281 days)

Product Code
Regulation Number
870.1875
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

eClinic Stethopod is an Electronic Stethoscope. It converts patient auscultation sound (heart, lungs, bowel, arteries, and veins) into audio signal and transmits it to clinically defined virtual appointment methods. It is intended to be used by professional users in a clinical environment or by lay users in a nonclinical or clinical environment to assist remote healthcare professionals' physical assessment. The device is not intended for self-diagnosis. eClinic Stethopod is intended for patients or customers who are 2 years and older. It is for both prescription use and over the counter use.

Device Description

eClinic Stethopod is an electronic stethoscope that is designed to be used with virtual communication systems for telehealth. It provides patient auscultation sounds for clinicians during virtual appointments while maintaining communication between clinicians and patients. The eClinic Stethopod is a passive electromechanical device. It converts physiological sounds into audio signals through a microphone. It further processes the signal through passive electronics for noise reduction and then transmits the real time data to clinically defined communication applications that connect patients and clinicians. eClinic Stethopod has a chest piece to pick up acoustic signal, a handle housing electronics that converts acoustic signal to audio signal and a cable to connect with communication devices. eClinic Stethopod needs to be connected to a communication device to achieve its intended use. There are two accessories: Apple ear pod and Apple adapter (lighting to 3.5 mm jack).

AI/ML Overview

The provided text describes the 510(k) summary for the eClinic Stethopod, an electronic stethoscope. It details the device's characteristics, intended use, and a comparison with a predicate device, the Eko Core (K200776).

However, the document does not provide specific acceptance criteria or details of a study that proves the device meets
these criteria in the way typically expected for an AI/ML medical device (e.g., performance metrics like sensitivity, specificity, AUC). Instead, it primarily focuses on demonstrating substantial equivalence to a predicate device based on design verification and validation tests related to sound quality and functionality.

Based on the provided text, here's an attempt to answer the questions, highlighting where information is not available:

  1. A table of acceptance criteria and the reported device performance
    The document does not present quantitative acceptance criteria or detailed performance metrics in the format of a table as might be expected for software performance (e.g., sensitivity, specificity). Instead, it states:
    "Based on design verification test, eClinic Stethopod is equivalent to the predicate device on sound rhythm, heartbeat rate, frequency spectrum and signal to noise ratio for its intended application."
    "eClinic stethopod sound quality are further evaluated by clinicians and volunteers in design validation. It is concluded eClinc stethopod sound quality through phone calls and Zoom is acceptable and it is easy to use."

    Acceptance Criterion (Inferred from testing claims)Reported Device Performance (as stated in the document)
    Equivalence on sound rhythmMet (equivalent to predicate device)
    Equivalence on heartbeat rateMet (equivalent to predicate device)
    Equivalence on frequency spectrumMet (equivalent to predicate device)
    Equivalence on signal to noise ratioMet (equivalent to predicate device)
    Sound quality through phone calls and ZoomAcceptable
    Ease of useEasy to use
  2. Sample sizes used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)
    The document mentions "design verification and design validation tests" and that "eClinic stethopod sound quality are further evaluated by clinicians and volunteers in design validation." However, no specific sample sizes for these tests (e.g., number of patients, number of sounds) are provided. The data provenance (country of origin, retrospective/prospective) is also not mentioned.

  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)
    The document states that "clinicians and volunteers" were involved in the design validation to evaluate sound quality. It does not provide the specific number of these individuals or their detailed qualifications (e.g., years of experience, specific medical specialty).

  4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
    No adjudication method is described for the evaluation of sound quality by clinicians and volunteers.

  5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
    No MRMC comparative effectiveness study is described. The device is a "passive electromechanical device" that converts sounds to audio signals for transmission, rather than an AI that assists human readers in interpretation. Therefore, a study of human readers improving with AI assistance is not applicable to the description provided.

  6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
    The device itself is not described as an algorithm performing a diagnostic task in isolation. Its function is to convert and transmit physiological sounds. The "design verification test" concluded equivalence on technical aspects like sound rhythm, heartbeat rate, frequency spectrum, and signal-to-noise ratio. This could be considered a form of "standalone" evaluation of the device's signal processing capabilities, but it's not an "algorithm only" performance in a diagnostic sense.

  7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
    For the technical aspects ("sound rhythm, heartbeat rate, frequency spectrum and signal to noise ratio"), the "ground truth" seems to be based on comparison to the predicate device's performance, implying a technical benchmark. For the "sound quality" and "ease of use," the ground truth was established by the subjective evaluation of "clinicians and volunteers." No objective clinical ground truth (like pathology or outcomes data) is mentioned for the device's performance given its intended function as a sound transmission tool.

  8. The sample size for the training set
    The document describes a physical medical device, not an AI/ML algorithm that requires a "training set" in the conventional sense. Therefore, this question is not applicable to the information provided.

  9. How the ground truth for the training set was established
    As the device is not an AI/ML algorithm with a training set, this question is not applicable to the information provided.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text on the right. The symbol is a stylized representation of a caduceus, a traditional symbol of medicine, with intertwined snakes and wings. To the right of the symbol is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue, with "FDA" in a larger font size than the rest of the text.

October 20, 2022

HomeDiagnostic LLC % Dave Yungvirt CEO Third Party Review Group, LLC 25 Independence Blvd Warren, New Jersey 07059

Re: K220099

Trade/Device Name: eClinic Stethopod Regulation Number: 21 CFR 870.1875 Regulation Name: Stethoscope Regulatory Class: Class II Product Code: DOD Dated: October 4, 2022 Received: October 7, 2022

Dear Dave Yungvirt:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.

510(k) Number (if known) K220099

Device Name eClinic Stethopod

Indications for Use (Describe)

eClinic Stethopod is an Electronic Stethoscope. It converts patient auscultation sound (heart, lungs, bowel, arteries, and veins) into audio signal and transmits it to clinically defined virtual appointment methods. It is intended to be used by professional users in a clinical environment or by lay users in a nonclinical or clinical environment to assist remote healthcare professionals' physical assessment. The device is not intended for self-diagnosis. eClinic Stethopod is intended for patients or customers who are 2 years and older. It is for both prescription use and over the counter use.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (6/20)

Page 1 of 1

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for HomeDiagnostic LLC. The logo features a stylized blue house with a curved line underneath it. The text "HomeDiagnostic LLC" is written in a gray sans-serif font to the right of the house icon.

K220099

510(K) SUMMARY

This 510(k) Summary is in accordance with the requirements of the Safe Medical Device Act (SMDA) of 1990. The content of this 510(k) summary is provided in conformance with 21 CFR §807.92.

5.1 Applicant's Information

Name:HomeDiagnostic LLC
Address:3160 De La Cruz Blvd Suite 100
Phone:18669764038
Contact Person:Jian Gong
Email:info@eclinicdx.com
Preparation Date:18 Oct 2022

5.2 Device Name

Product/Trade Name:eClinic Stethopod (K220099)
Common Name:Stethoscope
Classification Name:Electronic Stethoscope
Regulatory Class:Class II
Regulatory Number:21 CFR870.1875
Product Code:DQD

5.3 Legally Marketed Predicate Device

The predicate device for eClinic Stethopod is legally marketed Eko Core (K200776).

5.4 Device Description

5.4.1 Device Identification

eClinic Stethopod is an electronic stethoscope that is designed to be used with virtual communication systems for telehealth. It provides patient auscultation sounds for clinicians during virtual appointments while maintaining communication between clinicians and patients.

5.4.2 Device Characteristics

The eClinic Stethopod is a passive electromechanical device. It converts physiological sounds into audio signals through a microphone. It further processes the signal through passive electronics for noise reduction and then transmits the real time data to clinically defined communication applications that connect patients and clinicians.

5.4.3 Environment of Use

The eClinic Stethopod is used during physical assessment. The environment of use could be anywhere where physical assessment is done, for example non-clinical environment or healthcare facilities.

5.4.4 Brief Description of the Device

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image contains the logo for HomeDiagnostic LLC. The logo features a stylized house icon in blue, with a curved line underneath. The text "HomeDiagnostic LLC" is written in a sans-serif font to the right of the house icon.

eClinic Stethopod has a chest piece to pick up acoustic signal, a handle housing electronics that converts acoustic signal to audio signal and a cable to connect with communication devices. eClinic Stethopod needs to be connected to a communication device to achieve its intended use. There are two accessories: Apple ear pod and Apple adapter (lighting to 3.5 mm jack).

5.4.5 Materials of Use

eClinic Stethopod is not intended to have direct contact to body fluid. It is intended to contact human intact skin for a very short period. The eClinic Stethopod parts that contact the skin are composed of biocompatible materials and compliance to ISO 10993-Part 1 Biological Evaluation of Medical Devices.

5.4.6 Essential Performance Characteristics

The essential performance characteristics of the eClinic Stethopod are listed in Table 1.

FeatureSpecification
Chest PieceSingle Cup Chest piece
Chest Piece Diameter48 mm
Chest Piece RingAnti-Chill silicone
Frequency Range20-2000Hz
Weight<250g
Compatible SoftwareiOS 13 or later, Android 8 and above, Zoom 5.4.3 and above
Compatible DevicesiPhone 6 or later, iPad Pro, iPad Mini, iPad 2 or later; iPod touch 4th
generation;
Android 8.0 and above

Table 1: eClinic Stethopod Essential Performance Characteristics

5.5 Intended Use

eClinic Stethopod is an Electronic Stethoscope. It converts patient auscultation sound (heart, lungs, bowel, arteries, and veins) into audio signal and transmits it to clinically defined virtual appointment methods. It is intended to be used by professional users in a clinical environment or by lay users in a nonclinical or clinical environment to assist remote healthcare professionals' physical assessment. The device is not intended for selfdiagnosis.

eClinic Stethopod is intended for patients or customers who are 2 years and older. It is for both prescription use and over the counter use.

The predicate device is legally marketed Eko Core (K200776) manufactured by Eko Devices Inc.

Product/Trade Name:Eko Core
Common Name:Stethoscope
Classification Name:Electronic Stethoscope
Regulatory Class:Class II
Regulatory Number:21 CFR870.1875
Product Code:DQD

Eko Core is an electronic stethoscope that enables amplification, and transmission of auscultation sound data (heart, lungs, bowel, arteries, and veins), whereby a clinician at one location on network can listen to the auscultation sounds of a patient on site or at a different location on the network. It is intended for use on pediatric

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image contains the logo for HomeDiagnostic LLC. The logo features a stylized house icon with a blue roof and gray windows. The text "HomeDiagnostic LLC" is written in a simple, sans-serif font to the right of the house icon.

and adult patients. It is intended to be used by professional users in a clinical environment or by lay users in a nonclinical environment. The device is not intended for self-diagnosis.

eClinic Stethopod is equivalent to the predicate device in:

  • Both are electronic stethoscope: O
  • Both convert auscultation sound into audio signal and transmits it to clinicians; O
  • Both are used to be part of a physical assessment remotely; O
  • Both could be used by lay users in non-clinical environment or clinical environment; O The transmitted auscultation sound will be used by professionals sitting remotely to assist diagnoses.
  • Both are not used for self-diagnostics. O

The intended use of eClinic Stethopod is the same to the predicate devices for remote physical assessment. The statement variation between eClinic Stethopod and predicate device is not critical to the intended use of the device. The differences do not affect the safety and effectiveness of the device.

5.6 Comparison of Technological Characteristics with the Predicate Device

The principal operation of the eClinic Stethopod is the same to the predicate devices. Acoustic physiology sound is picked up by microphone and converted into audio signal. Then it is transmitted to advanced communication tools to be heard remotely by clinicians.

The following technological characteristics of the eClinic Stethopod is equivalent to the predicate devices:

  • Use mechanical chest piece to collect physiological sounds; O
  • Use microphone and electronics to digitalize the physiological sounds; O
  • Use electronics to reduce noise; O
  • O Use advanced communication tools (smart phones and operation system IOS or Android);

The following technological characteristics of the eClinic Stethopod is different to the predicate device:

  • eClinic Stethopod sends data via a cable to smartphones, tablets, or computers. The predicate device uses O RF wireless transmission.
  • eClinic Stethopod works with phone calls or virtual appointment systems defined by clinics or hospitals. O The recommended video system is Zoom. For the predicate device, its own App is needed to function.

The implementation differences such as wire/wireless transmission, and virtual appointment system software/ device's own App, do not impact the device's safety and efficacy for the intended use of the device.

5.7 Performance data

The performance of the eClinic Stethpod described in this submission was evaluated according to the quality system regulation (21 CFR §820). Design verification and design validation tests confirmed that all design requirements were met and that the device functions as intended.

Based on design verification test, eClinic Stethopod is equivalent to the predicate device on sound rhythm, heartbeat rate, frequency spectrum and signal to noise ratio for its intended application.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for HomeDiagnostic LLC. The logo features a stylized house icon in blue, with a curved line underneath. The text "HomeDiagnostic LLC" is written in a simple, sans-serif font to the right of the house icon.

eClinic stethopod sound quality are further evaluated by clinicians and volunteers in design validation. It is concluded eClinc stethopod sound quality through phone calls and Zoom is acceptable and it is easy to use.

5.8 Conclusions

The eClinic Stethopod's intended use, and technological characteristic are equivalent to those of the predicate device. Differences in implementation between the eClinic Stehopod and the predicate do not raise any concerns in regard to safety or effectiveness. It is concluded that the proposed eClinic Stethopod is substantially equivalent to the predicate device within the meaning of the Medical Device Amendments Act of 1976.

Elements ofComparisoneClinic Stethopod(K220099)Eko CORE(K200776)
General
Regulatory ClassClass IIClass II
Classification nameElectronic StethoscopeElectronic Stethoscope
Regulation Number21 CFR870.187521 CFR870.1875
Product CodeDQDDQD
ManufacturerHomeDiagnostic LLCEko Devices, Inc
FDA ClearancePending510K cleared (K200776)
Indications for UseeClinic Stethopod is an ElectronicStethoscope. It converts patient auscultationsound (heart, lungs, bowel, arteries, andveins) into audio signal and transmits it toclinicians by clinically defined virtualappointment methods.It is intended to be used by professionalusers in a clinical environment or by layusers in a nonclinical or clinicalenvironment to assist remote healthcareprofessionals' physical assessment. Thedevice is not intended for self-diagnosis.eClinic Stethopod is intended for patients orcustomers who are 2 years and older. It isfor both prescription use and over thecounter use.The Eko CORE is an electronicstethoscope that enables amplification,filtering, and transmission ofauscultation sound data (heart, lungs,bowel, arteries, and veins), whereby aclinician at one location on networkcan listen to the auscultation sounds ofa patient on site or at a differentlocation on the network. Eko CORE isintended for use on pediatric and adultpatients. The Eko CORE is intended tobe used by professional users in aclinical environment or by lay users ina nonclinical environment. The deviceis not intended for self-diagnosis.
Patient GroupPediatric and adult patientsPediatric and adult patients
Type of UsersPatients and Health-care professionalPatients and Health-care professional
Use EnvironmentNonclinical or Clinical settingsNonclinical or Clinical settings
Clinical ConditionsHeart, Lung and other human body soundsHeart, Lung and other human bodysounds
PrescribedOTC and RxOTC and Rx
Characteristics
Principles of OperationMicrophone and electronics for converting physiology sound to audio signal, reducing noise, and sending data to clinically defined virtual appointment systems through advanced communication tools, such as smartphones, tablets or computers.Eko CORE is a modified version of the previously cleared predicate device (K151319). The principles of operation is the same as K151319: Electronic stethoscope that gives sound amplification and audio transmission to a smartphone
Stethoscope TypeStand-alone electronic stethoscope with chest pieces sourced from analog stethoscopeAttachment to an analog stethoscope (Core)
Real time soundReal time auscultation soundReal time auscultation sound
Record and Playback SoundsNoYes
CommunicationPatient and clinician real time communication.Patient and clinician real time communication.
Compatibility (Chest piece)Chest Piece is part of eClinic StethopodsLittmann 3M Cardiology II/III-WelchAllyn Harvey Elite-ADC601 lines of analog stethoscopes
Compatibility AppDirect phone calls on iOS 6 and above or Android 8 and above;Zoom 5.4.3 and aboveEko App
Technical Equivalence
User Interface• 3.5mm headphone jack• Mobile App (Clinical defined)• On/Off Power button• Volume adjustment• LED status indicator• Mobile App
Analog/Digital InteroperabilityYesYes
ConnectivityCableBluetooth
Sound AmplificationNoYes
Data Transfer to compatible computing platformsYesYes
Energy SourcePower derived from connected digital device. No additional battery needed.Lithium Ion Battery
Compatible DevicesIOS and Android smartphones, ComputersIOS and Android smartphones, Computers
Frequency range20Hz -2kHz20Hz to 2kHz
StandardsN/AIEC 60601-1IEC 60601-1-2IEC 60601-1-11

Table 2. Comparison with the Predicate Device

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for HomeDiagnostic LLC. The logo features a stylized blue house icon on the left. To the right of the icon is the company name, "HomeDiagnostic LLC", written in a simple, sans-serif font.

§ 870.1875 Stethoscope.

(a)
Manual stethoscope —(1)Identification. A manual stethoscope is a mechanical device used to project the sounds associated with the heart, arteries, and veins and other internal organs.(2)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.(b)
Electronic stethoscope —(1)Identification. An electronic stethoscope is an electrically amplified device used to project the sounds associated with the heart, arteries, and veins and other internal organs.(2)
Classification. Class II (special controls). The device, when it is a lung sound monitor, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.