Intended Use

Y ADU Surgical Gowns are intended to be worn by operating room personnel during surgical procedure to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material.

The surgical gowns meet the requirements for Level 2 and Level 3 classification Per ANSI/AAMI PB-70.

Device Description

Y ADU Surgical Gowns provide Level 2 and Level 3 protections per ANSI/AAMI PB70 standard.

Y ADU Surgical Gowns (Level 2) are made from SMMS material. They are full length, constructed with raglan sleeves, hook and loop neck closures, and tie waist closures. They provide AAMI Level-2 protection.

The Level 2 Gowns have six different sizes: Small (S), Medium (M), Large (L), Extra Large (XL), Extra Extra Large (XXL), and Super Extra large (XXXL).

SMMS is a multi-ply material consisting of layers of spunbonded and meltblown polypropyle.

The body and sleeve is made from 35g Blue SMMS. Sleeve opening is made from pure polyester. The collar closure is made from dacron. The belt is made from 35g Blue SMMS.

Y ADU Surgical Gowns (Reinforced, Level 3) are made from SMMS material. They are full length, constructed with raglan sleeves, hook and loop neck closures, and tie waist closures. The chest area is reinforced with layer of PE+SPP material. They provide AAMI Level-3 protection.

The Level 3 Reinforced Gowns have six different sizes: Small (S), Medium (M), Large (L), Extra Large (XL), Extra Extra Large (XXL), and Super Extra large (XXXL).

SMMS is a multi-ply material consisting of layers of spunbond and meltblown polypropyle.

The body and sleeve is made from 45g Blue SMMS. Sleeve opening is made from pure polyester. The collar closure is made from dacron. The belt is made from 45g Blue SMMS. The reinforced area is made from PE +SPP material.

All gowns are sterilized with ethylene oxide.

AI/ML Overview

This document describes the YADU Surgical Gowns (K220092) and their conformance to safety and performance standards.

1. Table of Acceptance Criteria and Reported Device Performance:

Title of TestPurpose of TestAcceptance CriteriaResults (Level 2 gown)Results (Level 3 gown)
AATCC127: 2018Water resistance/Hydrostatic Pressure≥ 20 cm (Level 2)≥ 50 cm (Level 3)PassedPassed
AATCC42: 2017Water Resistance Impact Penetration≤ 1.0 gPassedPassed
ASTM D5034-09Tensile StrengthMD Mean ≥ 30 N;CD Mean ≥ 30 NPassedPassed
ASTM D5733-99Tearing StrengthMD Mean ≥ 10 N;CD Mean ≥ 10 NPassedPassed
ISO 9073-10: 2003LintingAverage ≤ 10000 (Log L < 4.0)PassedPassed
ASTM D1683-17Seam Strength≥ 30 NPassedPassed
ASTM F1868-17Evaporative Resistance of FabricsNo requirementPassedPassed
16 CFR 1610Flammability testingMeets Requirements Of Flame Resistant CPSC 1610 Class 1PassedPassed
ISO 10993-5: 2009Tests for In vitro cytotoxicityUnder conditions of the study, device extract is not cytotoxicPassedPassed
ISO 10993-10: 2010Tests for irritation And skinUnder the conditions of the study, not an irritant (irritation)PassedPassed
ISO 10993-10: 2010Tests for irritation and skinUnder conditions of the study, not a sensitizer (skin sensitization)PassedPassed
ISO 11135: 2014Sterilization ValidationSAL level of 10-6PassedPassed
ISO 11135: 2014 Annex BEthylene Oxide residualsEO: < 4 mg/deviceECH: < 9 mg/devicePassedPassed

2. Sample Size Used for the Test Set and Data Provenance:

The document does not explicitly state the exact sample sizes for each specific test mentioned. However, the tests are non-clinical and conducted to verify design specifications. The data provenance is implied to be from laboratory testing performed by the manufacturer (HeNan YADU Industrial Co., Ltd.) or a testing facility following international standards (e.g., ISO, ASTM, AATCC, CPSC CFR). The tests are retrospective in the sense that they are conducted on manufactured devices to confirm their properties.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:

Not applicable. This device is a surgical gown, and its performance criteria are established by recognized international and national standards (e.g., ANSI/AAMI PB-70, ASTM F2407, ISO 10993, 16 CFR 1610) rather than expert consensus on a test set in the way a diagnostic AI device would. The "ground truth" for each criterion is the objective measurement defined by the respective standard.

4. Adjudication Method for the Test Set:

Not applicable. The evaluation relies on standardized objective test methods rather than subjective human assessment requiring adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done:

No. This is a medical device (surgical gown), not a diagnostic or AI-assisted system for which MRMC studies are typically performed.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

Not applicable. This device is a physical product, not an algorithm.

7. The Type of Ground Truth Used:

The ground truth used for proving the device meets acceptance criteria is based on objective measurements and results from standardized laboratory tests according to established national and international consensus standards (e.g., AATCC, ASTM, ISO, CPSC CFR). These standards define the acceptable performance limits for various physical, chemical, and biological properties relevant to surgical gowns.

8. The Sample Size for the Training Set:

Not applicable. As this is a physical medical device and not a machine learning model, there is no "training set" in the context of data used to develop an algorithm. The "development" and "design" of the gown would involve engineering principles and material science, not data training.

9. How the Ground Truth for the Training Set was Established:

Not applicable, as there is no training set for this type of device. The design specifications and material selection would be based on existing industry knowledge, material properties, and the requirements of the relevant performance standards.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

HeNan YADU Industrial Co., Ltd. % Charles Shen Director Manton Business and Technology Services 37 Winding Ridge Oakland, New Jersey 07436

Re: K220092

Trade/Device Name: YADU Surgical Gowns Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FY A Dated: September 7, 2022 Received: September 8, 2022

Dear Charles Shen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For Bifeng Oian, M.D., Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K220092

Device Name YADU Surgical Gowns

Indications for Use (Describe)

Y ADU Surgical Gowns are intended to be worn by operating room personnel during surgical procedure to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material.

The surgical gowns meet the requirements for Level 3 classification Per ANSI/AAMI PB-70.

Model NumberStructureLevel of ProtectionSizeSterile/Non-SterileColor
SY0136Non-ReinforcedLevel-2SSterileBlue
SY0137Non-ReinforcedLevel-2MSterileBlue
SY0138Non-ReinforcedLevel-2LSterileBlue
SY0139Non-ReinforcedLevel-2XLSterileBlue
SY0172Non-ReinforcedLevel-2XXLSterileBlue
SY0173Non-ReinforcedLevel-2XXXLSterileBlue
SY0148ReinforcedLevel-3SSterileBlue
SY0149ReinforcedLevel-3MSterileBlue
SY0150ReinforcedLevel-3LSterileBlue
SY0151ReinforcedLevel-3XLSterileBlue
SY0178ReinforcedLevel-3XXLSterileBlue
SY0179ReinforcedLevel-3XXXLSterileBlue

Model Number and details for gowns involved in K220092:

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(K) Number: K220092

510(k) Summary:

This summary of 510(k) safety and effectiveness information is being submitted In accordance with the requirements of 21CFR 807.92

1.0 Submitter Information

HeNan YADU Industrial Co., Ltd. No.234, West Jianpu Road 453400 Changyuan, Henan Province, CHINA Tel: (086) 373-2157057

Submission Correspondent

Charles Shen Manton Business and Technology Services 37 Winding Ridge, Oakland, NJ 07436 Tel: 608-217-9358 Email: cyshen@aol.com

Date of Summary: October 6, 2022

2.0 Device Information

Proprietary Name:YADU Surgical Gowns
Common Name:Surgical Gown
Classification Name:Gown, Surgical

3.0 Device Classification

Classification Regulation:21 CFR 878.4040
Class:Class 2
Panel:General Hospital
Product Code:FYA

4.0 Predicate Device Information:

Manufacturer: Jiangsu Medplus Non-woven Manufacturer Co., Ltd

{4}------------------------------------------------

Product Name:Level 2 Standard Surgical Gown, Level 3 Standard SurgicalGown, Level 3 Reinforced Surgical Gown
510(K) #:K211422

5.0 Device description:

Y ADU Surgical Gowns provide Level 2 and Level 3 protections per ANSI/AAMI PB70 standard.

Y ADU Surgical Gowns (Level 2) are made from SMMS material. They are full length, constructed with raglan sleeves, hook and loop neck closures, and tie waist closures. They provide AAMI Level-2 protection.

The Level 2 Gowns have six different sizes: Small (S), Medium (M), Large (L), Extra Large (XL), Extra Extra Large (XXL), and Super Extra large (XXXL).

SMMS is a multi-ply material consisting of layers of spunbonded and meltblown polypropyle.

The body and sleeve is made from 35g Blue SMMS. Sleeve opening is made from pure polyester. The collar closure is made from dacron. The belt is made from 35g Blue SMMS.

Y ADU Surgical Gowns (Reinforced, Level 3) are made from SMMS material. They are full length, constructed with raglan sleeves, hook and loop neck closures, and tie waist closures. The chest area is reinforced with layer of PE+SPP material. They provide AAMI Level-3 protection.

The Level 3 Reinforced Gowns have six different sizes: Small (S), Medium (M), Large (L), Extra Large (XL), Extra Extra Large (XXL), and Super Extra large (XXXL).

SMMS is a multi-ply material consisting of layers of spunbond and meltblown polypropyle.

The body and sleeve is made from 45g Blue SMMS. Sleeve opening is made from pure polyester. The collar closure is made from dacron. The belt is made from 45g Blue SMMS. The reinforced area is made from PE +SPP material.

All gowns are sterilized with ethylene oxide.

6.0 Indications for Use:

{5}------------------------------------------------

Y ADU Surgical Gowns are intended to be worn by operating room personnel during surgical procedure to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material.

The surgical gowns meet the requirements for Level 2 and Level 3 classification Per ANSI/AAMI PB-70.

ModelNumberStructureLevel ofProtectionSizeSterile/Non-SterileColor
SY0136Non-ReinforcedLevel-2SSterileBlue
SY0137Non-ReinforcedLevel-2MSterileBlue
SY0138Non-ReinforcedLevel-2LSterileBlue
SY0139Non-ReinforcedLevel-2XLSterileBlue
SY0172Non-ReinforcedLevel-2XXLSterileBlue
SY0173Non-ReinforcedLevel-2XXXLSterileBlue
SY0148ReinforcedLevel-3SSterileBlue
SY0149ReinforcedLevel-3MSterileBlue
SY0150ReinforcedLevel-3LSterileBlue
SY0151ReinforcedLevel-3XLSterileBlue
SY0178ReinforcedLevel-3XXLSterileBlue
SY0179ReinforcedLevel-3XXXLSterileBlue

Model Number and details for gowns involved in K220092:

7.0 Comparison to Predicate Devices

Y ADU Surgical Gowns are compared with the following Predicate Devices in terms of intended use, design, material, specifications, and performance.

  • (1) K211422, "Level 2 Standard Surgical Gown, Level 3 Standard Surgical Gown, Level 3 Reinforced Surgical Gown", manufactured by "Jiangsu Medplus Nonwoven Manufacturer Co., Ltd" located in Siyang, Jiangsu, China.
    The following table shows similarities and differences of use, design, and material between our device and the predicate devices.

Table 5.1: Comparison of Intended Use, Design, and Material

DescriptionSubject Device (K220092)Predicate Device (K211422)Comparison
Product CodeFYAFYASame
RegulationNumber21 CFR 878.404021 CFR 878.4040Same
ClassificationClass2Class2Same

{6}------------------------------------------------

Indication forUseYADU Surgical Gowns areintended to be worn by operatingroom personnel during surgicalprocedure to protect both thesurgical patient and the operatingroom personnel from transfer ofmicroorganisms, body fluids, andparticulate material.The surgical gowns met therequirements for Level 2and Level 3 classificationPer ANSI/AAMI PB-70Surgical gown is intended to beworn by operating roompersonnel during surgicalprocedure to protect both thesurgical patient and the operatingroom personnel from transfer ofmicroorganisms, body fluids, andparticulate material.Per ANSI/AAMI PB70: 2012,Liquid barrier performance andclassification of protective appareland drapes intended for use inhealth care facilities, the Level 2standard surgical gowns met therequirements for Level 2classification, the Level 3 standardSurgical gowns and Level 3Same
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Reinforced surgical gowns met theRequirements for Level 3classification
Basic DesignFull length, constructed withraglan sleeves, neck closures, andwaist closures.Full length, constructed withraglan sleeves, neck closures, andwaist closures.Same
FeaturesRegular and ReinforcedRegular and ReinforcedSame
MaterialsSMMS nonwoven, Polyester, PE/SPPSMS nonwoven, Polyester andPolyamideSimilar
SizesS,M,L,XL,XXL,XXXLXS,S,M,L,XL,XXL,XXXLSimilar
Single UseYesYesSame
ColorBlueBlueSame
SterileSterileInformation not availableSame
LabelingConform with 21 CFR Part 801Conform with 21 CFR Part 801Same

The minor differences between the subject device and predicate device do not raise any concerns in terms of safety and effectiveness.

The following table shows similarities and differences of the performance between our device and the predicate devices. Tests were conducted following the recommended procedures outlined in the respective consensus standards.

Table 5.2: Comparison of Biocompatibility and Performance Testing

{7}------------------------------------------------

DescriptionSubject Device (K220092)Predicate Device (K211422)Comparison
CytotoxicityUnder conditions of the study,device extract is not cytotoxic(ISO10993-5)Under conditions of the study,device extract is not cytotoxic(ISO10993-5)Same
Skin Irritation andSensitizationUnder the conditions of thestudy, not an irritant or a sensitize(ISO 10993-10)Under the conditions of thestudy, not an irritant or asensitizer (ISO 10993-10)Same
Hydrostatic Pressure:Water Resistance≥ 20cm (Level 2)≥ 50cm (Level 3)≥ 20cm (Level 2)≥ 50cm (Level 3)Same
Impact PenetrationTest : WaterResistance≤ 1.0g≤ 1.0gSame
ProtectionLevel 2 and Level 3 perANSI/AAMI PB70Level 2 and Level 3 perANSI/AAMI PB70Same
Tensile Strength> 30 N> 20 NSimilar
Tear Strength> 10 N> 20 NSimilar
Seam Strength> 30 NData not availableSimilar
FlammabilityClass 1Class 1Same
Lint< 10000(Log particle < 4)Log (particle count) < 4Same
Ethylene Oxide andECH residualsEO: < 4 mg/deviceECH: < 9 mg/deviceEO: < 4 mg/deviceECH: < 9 mg/deviceSame

Both subject device and predicate device met the requirements of ANSI/AAMI PB70 and ASTM F2407. Their performance data are very similar.

Non-Clinical Study Summary 8.0

Non clinical tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards:

  • Protection Performance: ANSI/AAMI PB-70:2012 ●
  • Mechanical Performance: ASTMF2407-20
  • In Vitro Cytotoxicity Test: ISO10993-5:2009 ●
  • Skin Irritation Test: ISO10993-10:2010
  • Skin Sensitization Test: ISO10993-10:2010
  • Sterilization Validation: ISO-11135-1:2014 ●

{8}------------------------------------------------

Title of TestPurpose of TestAcceptance CriteriaResults(Level 2 gown)Results(Level 3 gown)
AATCC127: 2018Water resistance/Hydrostatic Pressure≥ 20 cm (Level 2)≥ 50 cm (Level 3)PassedPassed
AATCC42: 2017Water Resistanceimpact penetration≤ 1.0 gPassedPassed
ASTM D5034-09Tensile StrengthMD Mean ≥ 30 N;CD Mean ≥ 30 NPassedPassed
ASTM D5733-99Tearing StrengthMD Mean ≥ 10 N;CD Mean ≥ 10 NPassedPassed
ISO 9073-10: 2003LintingAverage ≤ 10000(Log L < 4.0)PassedPassed
ASTM D1683-17Seam Strength≥ 30 NPassedPassed
ASTM F1868-17Evaporative Resistanceof FabricsNo requirementPassedPassed
16 CFR 1610Flammability testingMeets RequirementsOf Flame ResistantCPSC 1610 Class 1PassedPassed
ISO 10993-5: 2009Tests for In vitrocytotoxicityUnder conditions of thestudy, device extract isnot cytotoxicPassedPassed
ISO 10993-10: 2010Tests for irritationAnd skin sensitizationUnder the conditionsof the study, not anirritantPassedPassed
ISO 10993-10: 2010Tests for irritationand skin sensitizationUnder conditions of thestudy, not a sensitizerPassedPassed
ISO 11135: 2014Sterilization ValidationSAL level of 10-6PassedPassed
ISO 11135: 2014Annex BEthylene OxideresidualsEO: < 4 mg/deviceECH: < 9 mg/devicePassedPassed

Clinical Study Summary 9.0

Clinical study is not performed for this product.

10.0 Conclusion

The conclusions drawn from the nonclinical tests demonstrate that the subject device is as safe, as effective, and performs as well as or better than the legally marketed predicate device.

§ 878.4040 Surgical apparel.

(a)
Identification. Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.(b)
Classification. (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.