K Number
K220067

Validate with FDA (Live)

Date Cleared
2022-08-30

(232 days)

Product Code
Regulation Number
878.4040
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The medical surgery mask mask is intended to be worn to protect and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These masks are in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.

Device Description

The medical surgery mask is pleated three-layer mask with ear loops and nose piece. The outer layer is made of spun-bonded polypropylene (PP) non-woven fabric. The middle layer with filtration function is made of melt blown polypropylene (PP) fabric. The inner layer contact with face is made of spun-bonded polypropylene (PP) non-woven fabric. The medical surgery mask is only the outer layers' color is blue (colorant: Pigment Blue 15:3/Model:147-14-8), which is held to cover the users' mouth and nose by two spandex elastic bands ultrasonic welded to the medical surgery mask.The inner layer is white. The elastic ear loops are not made with natural rubber latex. The nose piece contained in the medical surgery mask is in the middle layer of medical surgery mask to allow the user to fit the medical surgery mask around their noses, which is made of malleable aluminum wire.

The dimensions of each medical surgery mask are length 175±5 mm and width 95±5 mm, The dimensions of nose piece is length 120±10 mm, and the ear loop is length 180±10 mm. The mask model 2626-7 meets level 3 performance requirements in ASTM F2100.The medical surgery masks are sold non-sterile and are intended to be single use,disposable devices.

AI/ML Overview

The provided text describes the acceptance criteria and the study that proves a medical surgery mask meets these criteria, specifically for a 510(k) premarket notification (K220067) for the Medical Surgery Mask, Model: 2626-7.

Here's a breakdown of the requested information:

1. Table of Acceptance Criteria and Reported Device Performance:

Test Item (Performance Level 3)Acceptance Criteria (Level 3)Reported Device Performance (Model 2626-7)Verdict
Bacterial filtration efficiency (BFE) ASTM F2101-19BFE ≥ 98%Lot A: 99.72%, Lot B: 99.82%, Lot C: 99.94%Pass
Differential pressure (Delta-P) EN 14683:2019, Annex CDelta-P < 6.0 mmH2O/cm²Lot A: 1.86 mmH2O/cm², Lot B: 1.87 mmH2O/cm², Lot C: 1.86 mmH2O/cm²Pass
Sub-micron particulate filtration efficiency at 0.1 micron ASTM F2299PFE ≥ 98%Lot A: 99.34%, Lot B: 99.56%, Lot C: 99.33%Pass
Resistance to penetration by synthetic blood ASTM F1862Fluid resistant claimed at 160 mmHg32 out of 32 per lot pass at 160 mmHgPass
Flame spread 16 CFR Part 1610Class 1Class 1Pass
Biocompatibility: Cytotoxicity (ISO 10993-5:2009)NoncytotoxicUnder the conditions of the study, the device is noncytotoxic.Pass
Biocompatibility: Irritation (ISO 10993-10:2010)NonirritatingUnder the conditions of the study, the device is nonirritating.Pass
Biocompatibility: Sensitization (ISO 10993-10:2010)NonsensitizingUnder the conditions of the study, the device is nonsensitizing.Pass

2. Sample size used for the test set and the data provenance:

  • Sample Size: The document does not explicitly state a consolidated "sample size" for the entire test set. Instead, it refers to multiple lots (Lot A, B, C) for filtration and differential pressure tests, and "32 out of 32 per lot" for synthetic blood penetration. This implies that for each test, specific sample sizes were used per the respective standard's requirements.
  • Data Provenance: The tests are "Non-Clinical Tests Performed" by the applicant, Dongguan Missadola Technology Co., Ltd. The document does not specify the country of origin of the raw data, but the manufacturer is based in China. The data provenance is retrospective in the sense that these tests were performed and compiled for the 510(k) submission, not as part of a prospective clinical trial.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • This information is not applicable as the "ground truth" for this medical device (a surgical mask) is established through adherence to recognized international and national performance standards (ASTM, EN, ISO, CFR), not through expert consensus on medical images or clinical outcomes. The tests measured physical and biological properties of the mask.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

  • None. Adjudication methods like 2+1 or 3+1 are typically used in studies involving human interpretation (e.g., radiology reads) where there might be disagreement among experts. For performance testing of a physical product against objective standards, the results are typically quantitative and directly compared against the established criteria, not subject to human adjudication for "ground truth."

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No. This type of study (MRMC, AI assistance) is not applicable to a surgical mask. It's relevant for AI-powered diagnostic tools where human interpretation is involved.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Yes, in spirit. The performance tests (BFE, Delta-P, PFE, synthetic blood penetration, flammability, biocompatibility) are "standalone" in the sense that they measure the intrinsic performance of the mask itself, without human interaction with the device for its primary intended function (i.e., it's not a diagnostic tool where a human interprets algorithmic output). These are direct measurements of the product's physical and biological properties.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

  • The ground truth is established by objective performance standards and laboratory test methods (e.g., ASTM F2101 for BFE, EN 14683 for differential pressure, ISO 10993 for biocompatibility). These standards define the acceptable range for each critical performance parameter for a Level 3 surgical mask.

8. The sample size for the training set:

  • Not applicable. This device is a physical product (surgical mask), not an AI/ML algorithm. Therefore, there is no "training set" in the context of machine learning.

9. How the ground truth for the training set was established:

  • Not applicable. As there is no training set for an AI/ML algorithm, this question is not relevant.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath. The logo is simple and professional, and it is easily recognizable.

August 30, 2022

Dongguan Missadola Technology Co., Ltd % Doris Chen Regulatory Affairs Staff Shanghai Jiushun Enterprise Management Technology Service Co.,Ltd Room 1502, BaoAn Buiding, No.800 Dongfang Road Shanghai. 200122 China

Re: K220067

Trade/Device Name: Medical Surgery Mask, Model:2626-7 Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: July 29, 2022 Received: July 29, 2022

Dear Ms. Chen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K220067

Device Name MEDICAL SURGERY MASK Model:2626-7

Indications for Use (Describe)

The medical surgery mask mask is intended to be worn to protect and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These masks are in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K220067

510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance

with the requirements of 21 CFR 807.92.

The assigned 510(k) Number:K220067

Summary Prepared Date:July 29, 2022

1. Applicant :

Company Name:Dongguan Missadola Technology Co., LTD

Establishment Registration Number:3020729769

Address:F4,bloc 1,No.289,Nanhuang Road,Zhongtang Town, Dongguan city,China.

Contact Person (including title): Adeline Hong(General Manager)

Phone:+86-769-88878117

Email:adeline@missadola.com

2. Submission Correspondent:

Contact Person: Doris Chen

Shanghai Jiushun Enterprise Management Technology Service Co., Ltd.

Address: Room 1502,BaoAn Buiding,No.800 Dongfang Road,Shanghai,China.

Tel: +86-21-50931939

Email: doris-chen@isosh.com

3. Subject Device Information

Type of 510(k):Traditional
Common Name:FaceMask
Trade Name:MEDICAL SURGERY MASK: Model 2626-7
Classification Name:Mask,Surgical
Review Panel:General Hospital
Product Code:FXX
Regulation Number:21 CFR 878.4040

K220067 510(k) Summary Page 1 of 7

{4}------------------------------------------------

Regulation Class:

Predicate Device Information 4.

=

Predicate Device

Sponsor:Foshan Xinbao Technology Co., Ltd.
Common Name:Surgical apparel
Trade Name:Surgical Mask
510(k) number:K202424
Review Panel:General Hospital
Product Code:FXX
Regulation Number:21 CFR 878.4040
Regulation Class:II

5. Device Description

The medical surgery mask is pleated three-layer mask with ear loops and nose piece. The outer layer is made of spun-bonded polypropylene (PP) non-woven fabric. The middle layer with filtration function is made of melt blown polypropylene (PP) fabric. The inner layer contact with face is made of spun-bonded polypropylene (PP) non-woven fabric. The medical surgery mask is only the outer layers' color is blue (colorant: Pigment Blue 15:3/Model:147-14-8), which is held to cover the users' mouth and nose by two spandex elastic bands ultrasonic welded to the medical surgery mask.The inner layer is white. The elastic ear loops are not made with natural rubber latex. The nose piece contained in the medical surgery mask is in the middle layer of medical surgery mask to allow the user to fit the medical surgery mask around their noses, which is made of malleable aluminum wire.

The dimensions of each medical surgery mask are length 175±5 mm and width 95±5 mm, The dimensions of nose piece is length 120±10 mm, and the ear loop is length 180±10 mm. The mask model 2626-7 meets level 3 performance requirements in ASTM F2100.The medical surgery masks are sold non-sterile and are intended to be single use,disposable devices.

K220067 510(k) Summary Page 2 of 7

{5}------------------------------------------------

Intended Use / Indications for Use 6.

The medical surgery mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.

7. Comparison with predicate device

Elements ofComparisonSubject DevicePredicate DeviceVerdict
ManufacturerDongguanMissadolaTechnology Co., LTDFoshan Xinbao TechnologyCo., Ltd.--
Product NameMEDICAL SURGERY MASKSurgical Mask--
K NumberK220067K202424--
Product CodeFXXFXXSame
RegulationNumber21 CFR 878.404021 CFR 878.4040Same
Intended use/Indications for UseThe medical surgery masks areintended to be worn to protectboth the patient and healthcarepersonnel from transfer ofmicroorganisms, body fluidsand particulate material. Thesemasks are intended for use ininfection control practices toreduce the potential exposureto blood and body fluids. This isa single use, disposabledevice(s), provided non-sterile.The Surgical Masks areintended to be worn toprotect both the patientand healthcare personnelfrom transfer ofmicroorganisms, bodyfluids, and particulatematerial. These masks areintended for use ininfection control practices toreduce potential exposureto blood and body fluids.This is a single use,disposable device, providednonsterile.Same
Mask styleFlat pleated, 3 layers.Flat pleated,3 layers.Same
Design featureEar loopEar loopSame
MaterialOuterfacinglayerSpun-bond polypropyleneSpun-bond polypropyleneSame
MiddlelayerMelt blown polypropylene filterMelt blown polypropyleneSame
InnerSpun-bond polypropyleneSpun-bond polypropyleneSame
facinglayer
NosepieceMalleable aluminum wireGalvanized iron wireDifferentNote 1
EarloopsSpandexNylon and SpandexDifferentNote 1
ColorBlueBlueSame
Dimension (Width)17.5cm±0.5cm17.5cm±1cmSimilar
Dimension(Length)9.5cm±0.5cm9.5cm±1cmSimilar
OTC useYesYesSame
SterilityNon-SterileNon-SterileSame
UseSingle Use, DisposableSingle Use, DisposableSame
ASTMLevelF2100Level 3Level 3Same
Fluid resistancePerformanceASTM F186232 out of 32 pass at 160mmHgPass at 160mmHgSame
Particle FiltrationEfficiencyASTM F2299Lot A: 99.34%Lot B: 99.56%Lot C: 99.33%Pass at ≥98%SimilarNote 2
Bacterial FiltrationEfficiencyASTM F2101Lot A:99.72%Lot B:99.82%Lot C:99.94%Pass at ≥98%SimilarNote 2
FlammabilityClass16 CFR 1610Class 1Class 1Same
DifferentialPressure(Delta -P)EN 14683 Annex CLot A: 1.86mmH2O/cm²Lot B: 1.87mmH2O/cm²Lot C:1.86mmH2O/cm²Pass at <6.0 mmH2O/cm²SimilarNote 2
Biocompatibility
CytotoxicityUnder the conditions of the study, the device is noncytotoxic.Under the conditions of the study, the device is noncytotoxic.Same
SensitizationUnder the conditions of the study, the device is nonirritating.Under the conditions of the study, the device is nonirritating.Same
IrritationUnder the conditions of the study, the device is nonsensitizingUnder the conditions of the study, the device is nonsensitizingSame

Table 1 General Comparison

K220067 510(k) Summary Page 3 of 7

{6}------------------------------------------------

{7}------------------------------------------------

Note 1

The materials of nose piece and the ear loop were different from the predicate device. There are not raise additional questions for safety and effectiveness.

The biocompatibility evaluation test of the subject devices have been performed on the final finished device. The test results shows pass the requirements. There is no new risk generated from the difference of the material.

Note 2

For the Performance testing, the test results are not identical to each other, but they are similar and they both meet the requirement of Level 3 medical mask according to the ASTM F2100.

8. Summary of Non-Clinical Tests Performed

Non-clinical tests were conducted to verify that the proposed device met all design specifications as to the predicate device. The test results demonstrated that the proposed device complies with the following standards and the requirements stated in the Guidance for Industry and FDA Staff: Surgical Masks - Premarket Notification [510(k)] Submission issued on March 5, 2004:

  • A ASTM F2299-03, Standard Test Method for Determining the Initial Efficiency of Materials Used in Surgical face masks to Penetration by Particulates Using Latex Spheres.
  • EN 14683:2019,Annex C.Method for determination of breathability (differential pressure)

  • ASTM F1862/ASTM F1862M-17, Standard test method for resistance of Surgical face masks to penetration by synthetic blood (Horizontal projection of fixed volume at a known velocity)

  • ASTM F2101-19, Standard Test Method For Evaluating The Bacterial Filtration Efficiency (BFE) Of Surgical face mask Materials, Using A Biological Aerosol Of Staphylococcus Aureus.

  • 16 CFR Part 1610, Standard for the flammability of clothing textiles.

  • ISO 10993-5:2009, Biological Evaluation Of Medical Devices -- Part 5: Tests For In Vitro Cytotoxicity.

{8}------------------------------------------------

  • A ISO 10993-10:2010, Biological Evaluation Of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization.
    Table 2:Performance Testing
Test item(PerformanceLevel 3)Proposed deviceLevel 3Acceptance criteriaLevel 3Testresults/Verdict
Bacterial filtrationefficiency (BFE)ASTM F2101-19Lot A:99.72%Lot B:99.82%Lot C:99.94%BFE≥98%.Pass
Differentialpressure,(Delta-P)EN 14683:2019,Annex CLot A: 1.86mmH2O/cm²Lot B: 1.87mmH2O/cm²Lot C:1.86mmH2O/cm²Delta-P<6.0H2O/cm²Pass
Sub-micronparticulatefilltration efficiencyat 0.1 micron.ASTM F2299Lot A: 99.34%Lot B: 99.56%Lot C: 99.33%PFE≥98%.Pass
ResistancetopenetrationbysyntheticbloodASTM F186232 out of 32 per lot pass at 160 mmHgFluid resistant claimed at 160 mm HgPass
Flame spread16 CFR Part 1610Class1Class 1Pass

Results:All tests were passed.

Biocompatibility evaluation and test

Biocompatibility evaluation conducted in accordance with the FDA's 2016 guidance and

ISO10993-1:2018 supports that the subject devices are biocompatible.

The biocompatibility test includes the following tests:

  • A In Vitro Cytotoxicity Test per ISO 10993-5:2009 Biological evaluation of medical devices
    Part 5: Tests for in vitro cytotoxicity.

  • 》 Skin Sensitization Tests per ISO 10993-10:2010 Biological evaluation of medical devices
    —Part 10: Tests for irritation and skin sensitization

{9}------------------------------------------------

  • ア Skin Irritation Tests per ISO 10993-10:2010 Biological evaluation of medical devices—
    Part 10: Tests for irritation and skin sensitization.
ItemProposed deviceResult
CytotoxicityUnder the conditions of the study, the device is noncytotoxic.Pass
IrritationUnder the conditions of the study, the device is nonirritating.Pass
SensitizationUnder the conditions of the study, the device is nonsensitizingPass

9. Summary of Clinical Performance Test

No clinical study is included in this submission.

10. Conclusion

The conclusions drawn from the nonclinical tests demonstrate that the subject device,

MEDICAL SURGERY MASK is as safe, as effective, and performs as well as or better than the

legally marketed predicate device, Surgical Mask (K202424).

K220067 510(k) Summary Page 7 of 7

§ 878.4040 Surgical apparel.

(a)
Identification. Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.(b)
Classification. (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.