K Number
K214118

Validate with FDA (Live)

Device Name
Riva Cem Automix
Manufacturer
Date Cleared
2022-08-30

(243 days)

Product Code
Regulation Number
872.3275
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Permanent cementation of:

    1. Porcelain fused to metal (PFM) to crowns and bridges
    1. Prefabricated/cast posts
    1. Metals to crowns, bridges, inlays & onlays, orthodontic appliances & posts
    1. Ceramics (high strength)* to crowns & bridges, inlays & onlays
  • *E.g. Zirconia, Lithium Disilicate,
    1. Ceramics (low strength)* to inlays
  • *E.g. Feldspathic porcelains, Glass ceramics
Device Description

Riva Cem Automix™ is a radiopaque, fluoride releasing, resin modified glass ionomer luting cement.

Riva Cem Automix™ is intended for the cementation of crowns, bridges, inlays, posts (fibre-reinforced), and orthodontic appliances.

Riva Cem Automix™ consists of a base paste and a catalyst paste packaged in the automix delivery system for direct delivery in consistent mix ratios. Riva Cem Automix™ is available in a universal light-yellow shade in a double barrel syringe. The automix tip offers convenience over traditional hand mixed systems.

The product provides both self-cure and tack cure mechanisms which makes it versatile for the cementation of different restorative systems.

AI/ML Overview

The provided text is a 510(k) premarket notification for a dental cement called "Riva Cem Automix." It focuses on demonstrating substantial equivalence to predicate devices through bench testing and biocompatibility testing, rather than an AI-powered diagnostic device. Therefore, it does not contain the information requested regarding acceptance criteria and a study proving an AI device meets those criteria.

Specifically, the document lacks information on:

  1. Acceptance criteria for an AI device and reported device performance: The document details acceptance criteria and performance for a dental cement, focusing on physical and chemical properties like working time, setting time, film thickness, flexural strength, radio-opacity, shear bond strength, and fluoride release. These are not acceptance criteria for an AI device.
  2. Sample size for a test set and data provenance: No AI test set is mentioned. The "performance data" section refers to in-vitro bench tests for a physical product.
  3. Number of experts and their qualifications for ground truth: Not applicable as it's not an AI device study requiring human expert labeling of data.
  4. Adjudication method for the test set: Not applicable.
  5. Multi-reader multi-case (MRMC) comparative effectiveness study: Not conducted as it's not an AI device study.
  6. Standalone (algorithm-only) performance: Not applicable.
  7. Type of ground truth used: For the dental cement, the ground truth is established by standardized laboratory testing methods (e.g., ISO standards) measuring physical and chemical properties, not expert consensus or pathology on images.
  8. Sample size for the training set: Not applicable as no AI training set is mentioned.
  9. How ground truth for the training set was established: Not applicable.

In summary, the provided document describes the regulatory approval of a dental cement, not an AI medical device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and "ADMINISTRATION" in a smaller font size below.

August 30, 2022

SDI Limited Antonella Kotefski Regulatory Affairs Manager 3-15 Brunsdon Street Bayswater, Victoria 3153 AUSTRALIA

Re: K214118

Trade/Device Name: Riva Cem Automix Regulation Number: 21 CFR 872.3275 Regulation Name: Dental Cement Regulatory Class: Class II Product Code: EMA Dated: July 29, 2022 Received: August 3, 2022

Dear Antonella Kotefski:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K214118

Device Name Riva Cem Automix

Indications for Use ( Describe )
Permanent cementation of:
    1. Porcelain fused to metal (PFM) to crowns and bridges
    1. Prefabricated/cast posts
    1. Metals to crowns, bridges, inlays & onlays, orthodontic appliances & posts
    1. Ceramics (high strength)* to crowns & bridges, inlays & onlays
  • *E.g. Zirconia, Lithium Disilicate,
    1. Ceramics (low strength)* to inlays
  • *E.g. Feldspathic porcelains, Glass ceramics
Type of Use (Select one or both, as applicable)
-------------------------------------------------

× Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.qov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

1. Submitter Information

Submitter:SDI Limited3-15 Brunsdon StreetBayswater, Victoria 3153Australia
Contact Person:Quynh Jewell
Position:Regulatory Affairs Manager
Phone:+61 3 8727 7111
Email:Quynh.Jewell@sdi.com.au

24 August 2022 Date Prepared:

2. Device Details

Proprietary Name:Riva Cem Automix™
Common Name:Dental Cement, Resin Modified Glass Ionomer
Regulation Name:Dental Cement
Regulation Number:21 CFR 872.3275
Product Code:EMA
Regulatory Class:II

Predicate Device 3.

Permanent cementation for the following:

Primary Predicate Device510(k) Device NumberCompany Name
RelyX Luting Plus™K1111853M ESPE
Secondary Predicate Device510(k) Device NumberCompany Name
FujiCEM™ 2K182854GC Corporation

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the letters "SDI" in a black box on the left side of the image. To the right of the box are the letters "s". The letters are in a sans-serif font and are white against a black background. The image is simple and clear.

Device Description 4.

Riva Cem Automix™ is a radiopaque, fluoride releasing, resin modified glass ionomer luting cement.

Riva Cem Automix™ is intended for the cementation of crowns, bridges, inlays, posts (fibre-reinforced), and orthodontic appliances.

Riva Cem Automix™ consists of a base paste and a catalyst paste packaged in the automix delivery system for direct delivery in consistent mix ratios. Riva Cem Automix™ is available in a universal light-yellow shade in a double barrel syringe. The automix tip offers convenience over traditional hand mixed systems.

The product provides both self-cure and tack cure mechanisms which makes it versatile for the cementation of different restorative systems.

5. Indications of Use

Permanent cementation of:

Indicated MaterialType of Restoration
Porcelain fused to metal (PFM)Crowns & bridges
Prefabricated/castPosts
MetalsCrown, bridges, inlays & onlays, orthodontic appliances& posts
Ceramics (high strength)1Crowns & bridges, inlays & onlays
Ceramics (low strength)2Inlays

1 High strength ceramics, e.g. Zirconia, Lithium Disilicate.

2 Low strength ceramics, e.g. Feldspathic Porcelains, Glass Ceramics.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows a black square with the letters "SDI" in white. The letters are in a sans-serif font and are stacked vertically, with "S" on top and "I" on the bottom. The letters are bold and slightly blurred. The black square provides a high contrast background for the white letters.

Comparison of Technological Characteristics with the Predicate 6. Device

Luting cements, including resin modified glass ionomer cements, are widely used to cement crowns. Luting cements provide the mechanical interlock between the crown and the other types of restorative materials indicated, and the tooth structure.

Riva Cem Automix™ is identical to predicate devices RelyX™ Luting Plus Automix and FujiCEM™ 2 in terms of technology and delivery.

Two mechanisms provide the bonding between the cement and the tooth structure or restorative material through self-curing and light-curing.

Self-curing the luting cement occurs by an acid-base reaction between the two pastes, as well as polymerisation of the methacrylic polymer.

Light curing also provides the chemical reaction of the two pastes by the photoinitiators in the formulation, enhancing the reaction between the cement and the restorative material.

Substantial Equivalence

Information provided in this 510(k) submission shows that Riva Cem Automix™ is substantially equivalent to the predicate devices 3M ESPE RelyX™ Luting Plus Automix (K111185) and GC FujiCEM™ 2 Luting Cement (K182854). A comparison of the devices is provided below in Table 1.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the letters "SDI" in white font on a black square background. The letters are large and bold, filling most of the square. The font appears to be sans-serif. The overall impression is simple and graphic.

Establishment Registration No: 3004140838

510(k) NOTIFICATION (Traditional): Riva Cem Automix – Summary – July 2022 - K214118

Table 1 Comparison of Riva Cem Automix™ with the Primary Predicate and Secondary Predicate Devices

Submitter DevicePrimary Predicate DeviceSecondary Predicate DeviceDifferences
ProprietaryNameRiva Cem AutomixTM(K214118)RelyXTM Luting Plus Automix(K111185)FujiCEMTM 2(K182854)Name
Regulation No.21 CFR 872.327521 CFR 872.327521 CFR 872.3275None
Product CodeEMAEMAEMANone
RegulatoryClassIIIIIINone
Rx / OTCRxRxRxNone
RegulationNameDental cementDental cementDental cementNone
ProductCategoryResin Modified Glass Ionomer LutingCementResin Modified Glass Ionomer LutingCementResin Modified Glass Ionomer LutingCementNone
Intended UseDental Luting CementDental Luting CementDental Luting CementNone
ApplicationPaste-pastePaste-pastePaste-pasteNone
CuringSelf-cure, tack cureSelf-cure, tack cureSelf-cureAll are similar.Riva CemAutomixTM andRelyXTM LutingPlus have optionaltack cure.
CuringChemistryAcid-base and polymerisationreactionsAcid-base and polymerisationreactionsAcid-base and polymerisationreactionsNone
Indications forUse (IFU)Permanent cementation of:- Porcelain fused to metal (PFM) tocrowns and bridges- Prefabricated/cast posts- Metals to crowns, bridges, inlays &onlays, orthodontic appliances &posts510k indications:This device is intended for use as adental cement.RelyX Luting Plus Automix (Lexus-2)is indicated for Luting:- Luting porcelain fused to metalcrowns and bridges to toothFujiCEM 2 is indicated for:- Cementation of metal-based inlays,onlays, crowns and bridges- Cementation of resin inlays, onlays,crowns and bridges- Cementation of all ceramic inlays- Cementation of high strength (e.g.Similar - differentwording. Seebelow forcomparison of eachindication.
Submitter DevicePrimary Predicate DeviceSecondary Predicate DeviceDifferences
ProprietaryNameRiva Cem AutomixTM(K214118)RelyXTM Luting Plus Automix(K111185)FujiCEMTM 2(K182854)Name
- Ceramics (high strength)* tocrowns & bridges, inlays & onlaysE.g. Zirconia, Lithium Disilicate,- Ceramics (low strength) to inlays*E.g. Feldspathic porcelains, Glassceramicsstructure, amalgam, composite orglass ionomer core build ups;- Luting metal inlays, onlays orcrowns;- Luting pre-fabricated and cast postcementation;- Luting orthodontic appliances;- Luting crowns made with all-alumina or all zirconia cores such asProcera All Ceramzirconia based, lithium disilicate)ceramic onlays, crowns and bridges- Cementation of metal, ceramic andfiber posts
Comparison ofIFUPorcelain fused to metal (PMF) tocrowns and bridges510k indications:Luting porcelain fused to metalcrowns and bridges to toothstructure, amalgam, composite orglass ionomer core build ups;Cementation of metal-based inlays,onlays, crowns and bridgesNone (PFM andmetal-based arethe same regardingcementation asboth consist ofmetal).
Prefabricated/cast postsN/ACementation of resin inlays, onlays,crowns and bridgesCementation of metal, ceramic andfiber postsNone (except forRelyXTM LutingPlus which it is notindicated)
Metals to crowns, bridges, inlays &onlays, orthodontic appliances &posts510k indications:Luting metal inlays, onlays orcrowns,Luting orthodontic appliances,Luting prefabricated and cast postcementationCementation of metal-based inlays,onlays, crowns and bridgesNone
Ceramics (high strength)* to crowns& bridges, inlays & onlays*E.g. Zirconia, Lithium Disilicate510k indications:Luting crowns made with all-aluminaor all zirconia cores such as ProceraCementation of high strength (e.g.based, lithium disilicate) ceramiconlays, crowns and bridgesNone
Submitter DevicePrimary Predicate DeviceSecondary Predicate DeviceDifferences
ProprietaryNameRiva Cem AutomixTM(K214118)RelyXTM Luting Plus Automix(K111185)FujiCEMTM 2(K182854)Name
All Ceram
Ceramics (low strength)* - Inlays*E.g. Feldspathic porcelains, GlassceramicsN/ACementation of all ceramic inlaysNone (except forRelyXTM LutingPlus which it is notindicated)
Light CuringSpecificationComparisonThe parameters for tack curing are:1. Use a light curing deviceWavelength: 440-480 nmIntensity: 1500 mW/cm2[+5% / - 15%]2. Apply for 5 seconds per surfaceThe parameters for tack curing are:1. Use a light curing device:Wavelength: 430-480 nmIntensity: 1470 mW/cm2[-10% / +20%]2. Apply for 5 seconds per surface.N/AMinor differences
Contra-indications1. Pulp capping2. Product may cause skin irritationsto some people. In such cases,discontinue use and seek medicalattention. Any persons having knownresin allergies should immediatelydiscontinue useThis product contains substancesthat may cause an allergic reactionby skin contact in certain individuals.Avoid use of the product in patientswith known acrylate allergies.1. Pulp capping.2. Avoid use of this product inpatients with known allergies to glassionomer cement, methacrylatemonomer or methacrylate polymerNone (except forRelyXTM LutingPlus whichcontraindication forpulp capping is notspecified)
TechnologicalCharacteristics& Mode ofActionA two-component paste/pastesystem based on:1.An acid-base reaction betweenfluoroalumino-silicate glass in andpolyacrylic acid copolymer.2. Polymerization of methacrylatemonomers is through chemical cureor light cure.3. Photoinitiator system: • Tack-curing for excess removalA two-component paste/pastesystem.Glass Ionomer Tooth bondingproperties based on1. An acid-base reaction betweenfluoroalumino-silicate glass andmethacrylated polycarboxylic acid;2. A free radical polymerization ofthe methacrylated copolymer andethylenically unsaturatedmonomers/agents, through chemicalA two-component paste/pastesystem.The device is set by acid-basereaction and polymerization aftermixing 2 pastes.1. Acid-base reaction occursbetween fluoroaluminosilicate glassin Paste A and polyacrylic acid inPaste B.2. Polymerization of methacrylatemonomers is through chemical cure.None. Devices arepaste/pastesystems which forman acid-basereaction andpolymerisation ofmethacrylatecomponent.Riva CemAutomixTM andRelyXTM LutingPlus offer a tack
Submitter DevicePrimary Predicate DeviceSecondary Predicate DeviceDifferences
ProprietaryNameRiva Cem Automix™(K214118)RelyX™ Luting Plus Automix(K111185)FujiCEM™ 2(K182854)Name
The 2 pastes are delivered in aconsistent manner through anAutomix syringe delivery system.cure or light cure;3. Photoinitiator system withcamphorquinone (CPQ) and 4-(dimethylamino)- phenethyl alcohol(DMAPE): • Tack-curing for excessremovalThe 2 pastes are delivered in aconsistent manner through anautomix delivery systemcure and containphotoinitiators.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the letters "SDI" in white font on a black square background. The letters are large and bold, filling most of the square. The font appears to be a sans-serif typeface.

SDI Limited

Establishment Registration No: 3004140838

510(k) NOTIFICATION (Traditional): Riva Cem Automix – Summary – July 2022 -K214118

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows a black square with the letters "SDI" in white. The letters are stacked on top of each other, with "S" on top, "D" in the middle, and "I" on the bottom. The font is a simple, sans-serif typeface. The overall impression is clean and modern.

SDI Limited

Establishment Registration No: 3004140838

510(k) NOTIFICATION (Traditional): Riva Cem Automix – Summary – July 2022 - K214118

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows a black square with the letters "SDI" in white. The letters are in a simple, sans-serif font and are stacked horizontally. The background is a solid black color, providing a strong contrast to the white letters.

SDI Limited

Establishment Registration No: 3004140838

510(k) NOTIFICATION (Traditional): Riva Cem Automix – Summary – July 2022 - K214118

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the letters "SDI" in a black square, followed by the letter "S" in a larger font. The letters are all in white. The letters "SDI" are stacked on top of each other, with the "S" on top, the "D" in the middle, and the "I" on the bottom. The black square is slightly offset to the left of the letter "S".

Performance Data 7.

The following performance data were provided in support of the substantial equivalence determination.

Bench testing

The following in-vitro bench tests were performed on Riva Cem Automix™ as per ISO 9917-2 2017 - "Dental water-based cements - Part 2: Resin-modified cements" (Section 5), ISO 29022 - "Dentistry - Adhesive - Notched-edge sheer bond strength test", ISO/TS 11405 -"Dentistry - Testing of adhesion to tooth structure" and an in-house method to verify physical properties and performance in support of substantial equivalence:

  • Working time
  • . Setting time
  • . Film thickness
  • . Flexural strength
  • . Radio-opacity
  • . Shear Bond strength
  • Fluoride release

The performance of Riva Cem Automix™ satisfactorily met the requirements of the nonclinical bench testing conducted to support substantial equivalence.

Biocompatibility testing

Biocompatibility testing was conducted according to ISO 10993-1 - Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process, and ISO 7405- Dentistry - Evaluation of biocompatibility of medical devices used in dentistry.

The results of the biocompatibility testing and composition analysis conducted relating to the subject device Riva Cem Automix™ support its substantial equivalence.

Clinical Performance Data

No data from human clinical studies has been included to support the substantial equivalence of Riva Cem Automix™.

Electrical safety and electromagnetic compatibility (EMC)

This section does not apply.

Software verification and validation testing

This section does not apply.

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the logo for SDI Limited. The logo consists of the letters "SDI" in a bold, sans-serif font, with the letters in white against a black square. To the right of the square, the words "SDI Limited" are written in a smaller, sans-serif font.

Mechanical and acoustic testing

This section does not apply.

Animal studies

This section does not apply.

Clinical studies

No data from human clinical studies has been included to support the substantial equivalence of Riva Cem Automix™ .

Conclusion Regarding Substantial Equivalence 8.

Riva Cem AutomixTM has the same intended use, incorporates the same fundamental technology, and has similar indications for use as the primary predicate RelyX Luting Plus and secondary predicate FujiCEM 2. Test data to verify physical properties and the performance of Riva Cem AutomixTM has been provided including working time, setting time, film thickness, flexural strength and radio-opacity on various substrates. The results of the testing, combined with the design and intended use comparison with both predicate devices, RelyX Luting Plus and FujiCEM 2, support substantial equivalence.

§ 872.3275 Dental cement.

(a)
Zinc oxide-eugenol —(1)Identification. Zinc oxide-eugenol is a device composed of zinc oxide-eugenol intended to serve as a temporary tooth filling or as a base cement to affix a temporary tooth filling, to affix dental devices such as crowns or bridges, or to be applied to a tooth to protect the tooth pulp.(2)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 872.9.(b)
Dental cement other than zinc oxide-eugenol —(1)Identification. Dental cement other than zinc oxide-eugenol is a device composed of various materials other than zinc oxide-eugenol intended to serve as a temporary tooth filling or as a base cement to affix a temporary tooth filling, to affix dental devices such as crowns or bridges, or to be applied to a tooth to protect the tooth pulp.(2)
Classification. Class II.