K Number
K214115

Validate with FDA (Live)

Device Name
MightySat -OTC
Manufacturer
Date Cleared
2024-01-31

(762 days)

Product Code
Regulation Number
870.2700
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Masimo MightySat-OTC is intended for the spot-checking of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR).

The Masimo MightySat-OTC is indicated for use with individuals 18 years and older who are well or poorly perfused under no motion conditions.

The Masimo MightySat-OTC is not intended for diagnosis or screening of lung disease and treatment decisions using the device should only be under the advice of a healthcare provider.

Device Description

The subject device, MightySat-OTC, is a fingertip pulse oximeter that includes Masimo SET technology for the measurement of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) in individuals 18 years and older. The predicate MightySat that was previously cleared under K181956, has the same intended use as the subject device except it is cleared for prescription use only.

The subject device is identical to the predicate with the RRp and PVi parameters have been disabled in the subject device and the labeling has been modified to provide additional information on the role of pulse oximeters for OTC users. There were no hardware changes made to the MightySat Rx (predicate device, K181956) that resulted in the MightySat-OTC. Like the predicate device, the subject device, has the combined function of a pulse oximeter monitor and a reusable sensor. It includes a color display, enclosed by plastic housing, and powered by two alkaline "AAA" batteries. It also includes optional Bluetooth wireless technology for the wireless transfer of patient data to mobile device application.

AI/ML Overview

Here’s an analysis of the acceptance criteria and study details for the Masimo MightySat-OTC, based on the provided document:

1. Table of Acceptance Criteria and the Reported Device Performance

Criterion (Measurement Accuracy in Accordance with ISO 80601-2-61)Acceptance Limit (ARMS)Reported Device Performance (ARMS)
SpO2, No Motion (70 - 100%, 18 years and older)1.5%1.26%
SpO2, Low Perfusion (70 - 100%, 18 years and older)2%2% (Same as predicate, implied to meet)
Pulse Rate, No Motion (25 – 240 BPM, 18 years and older)3 BPM3 BPM (Same as predicate, implied to meet)
Pulse Rate, Low Perfusion (25 – 240 BPM, 18 years and older)3 BPM3 BPM (Same as predicate, implied to meet)
SpO2 across different skin pigments and gendersWithin 1.5%Met across all measured subgroups (Light Male: 1.27%, Dark Male: 1.44%, Light Female: 1.11%, Dark Female: 1.32%)

2. Sample Size Used for the Test Set and the Data Provenance

  • Sample Size: 39 healthy volunteer subjects.
  • Data Provenance: The study was a prospective clinical performance testing conducted in accordance with ISO 80601-2-61. The location of the study is not explicitly stated, but it is a clinical study involving human volunteers. Data includes varying skin pigmentation, measured using the color-based Massey-Martin scale.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

The document does not specify the number or qualifications of experts used to establish ground truth. For pulse oximetry, the ground truth for SpO2 is typically established through arterial blood gas analysis (reference oximeter), usually performed by trained medical professionals, but this detail is not provided.

4. Adjudication Method for the Test Set

The document does not explicitly state an adjudication method (like 2+1, 3+1). For pulse oximetry clinical studies, ground truth is generally obtained directly from standardized reference methods (e.g., CO-oximetry of arterial blood samples), not through expert adjudication of interpretations.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not done. The study focused on the standalone performance of the device against a reference method for SpO2 and PR accuracy. The document mentions human factors and usability testing to validate user understanding of labeling and device outputs for OTC use, but this is a different type of study than an MRMC comparing human reader performance with and without AI assistance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Yes, the primary clinical performance testing for SpO2 and Pulse Rate accuracy appears to be a standalone (algorithm only) evaluation. The device measures and displays physiological signals directly, and the clinical study evaluates the accuracy of these measurements against a reference standard. The "human factors and usability testing" is a separate evaluation of the user interface and labeling for OTC use, not directly related to the algorithm's diagnostic performance with human input.

7. The Type of Ground Truth Used

The ground truth for SpO2 accuracy is based on a reference standard, specifically arterial blood oxygen saturation (SaO2), as implied by the "SaO2 range" in the performance tables. For pulse oximeters, this typically involves drawing arterial blood samples and analyzing them with a laboratory CO-oximeter.

8. The Sample Size for the Training Set

The document does not provide a specific sample size for the training set. It mentions that "the calibration curve has been adjusted to allow for improved accuracy performance, including across different skin pigments and genders." This implies that data was used to train or refine the device's algorithms, but the size and characteristics of that dataset are not detailed in this submission.

9. How the Ground Truth for the Training Set Was Established

The document states that the calibration curve was "adjusted to allow for improved accuracy performance, including across different skin pigments and genders." While it doesn't describe the exact process for establishing ground truth for this internal adjustment/training, it would typically involve similar methods to the clinical validation: comparing the device's readings to arterial blood gas measurements from a diverse population.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the symbol of the Department of Health & Human Services on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out next to it.

January 31, 2024

Masimo Corporation Kertana Shankar Regulatory Specialist II 52 Discovery Irvine, California 92618

Re: K214115

Trade/Device Name: MightySat - OTC Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: OLK Dated: September 19, 2022 Received: September 20, 2022

Dear Kertana Shankar:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Bradley Q. Quinn -S

Bradley Quinn Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K214115

Device Name MightySat-OTC

Indications for Use (Describe)

The Masimo MightySat-OTC is intended for the spot-checking of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR).

The Masimo MightySat-OTC is indicated for use with individuals 18 years and older who are well or poorly perfused under no motion conditions.

The Masimo MightySat-OTC is not intended for diagnosis or screening of lung disease and treatment decisions using the device should only be under the advice of a healthcare provider.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Masimo Corporation. The logo consists of a red circle with a white checkmark inside, followed by the word "Masimo" in black, bold font. To the right of the word "Masimo" is the text "MASIMO CORPO" followed by the address "52 Discovery, Irvine, CA 92618".

Submitter and Address ofManufacturing Facility:Masimo Corporation52 DiscoveryIrvine, CA 92618Phone: (949) 297-7000FAX: (949) 297-7592
Date:December 15, 2023
Contact:Kertana ShankarRegulatory Specialist IIMasimo CorporationPhone: (949) 297-7260
Trade Name:MightySat - OTC
Common Name:Oximeter
Classification Regulation/Product Code:21 CFR 870.2700, Class II/OLK
Establishment RegistrationNumber:3011353843
Reason for PremarketNotification:Over-the-counter use
Primary Predicate:K181956 - Masimo MightySat Fingertip Pulse Oximeter
Performance StandardsNo performance standards for the above device have been promulgatedpursuant to Section 514.

1.0 Device Description

The subject device, MightySat-OTC, is a fingertip pulse oximeter that includes Masimo SET technology for the measurement of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) in individuals 18 years and older. The predicate MightySat that was previously cleared under K181956, has the same intended use as the subject device except it is cleared for prescription use only.

The subject device is identical to the predicate with the RRp and PVi parameters have been disabled in the subject device and the labeling has been modified to provide additional information on the role of pulse oximeters for OTC users. There were no hardware changes made to the MightySat Rx (predicate device, K181956) that resulted in the MightySat-OTC. Like the predicate device, the subject device, has the combined function of a pulse oximeter monitor and a reusable sensor. It includes a color display, enclosed by plastic housing, and powered by two alkaline "AAA" batteries. It also includes optional Bluetooth wireless technology for the wireless transfer of patient data to mobile device application.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Masimo Corporation. The logo consists of a red circle with a white checkmark inside, followed by the word "Masimo" in large, bold, black letters. To the right of the word "Masimo" is the text "Masimo Corporation" in smaller, gray letters, followed by the address "52 Discovery, Irvine, CA 92618".

As part of this submission, the MightySat-OTC is being submitted to be marketed for Over-The-Counter (OTC) use.

The following are specifications for MightySat-OTC:

FEATURESPECIFICATION
Display
Display TypeOLED color display
Display RangeOxygen Saturation (SpO2): 0-100%Pulse Rate (PR): 25-240 beats per minute (BPM)Perfusion Index (Pi): 0.02-20%
Display ResolutionSpO2: 1%PR: 1 BPM
Measurement Accuracy in Accordance with ISO 80601-2-61
SpO2, No Motion70 - 100%, 1.5%, ARMS, 18 years and older
SpO2, Low Perfusion70 - 100%, 2%, ARMS, 18 years and older
Pulse Rate, No Motion25 – 240 BPM, 3 BPM ARMS, 18 years and older
Pulse Rate, Low Perfusion25 – 240 BPM, 3 BPM ARMS, 18 years and older
Power
Internal batteryTwo Alkaline "AAA" batteries
Interface
WirelessBluetooth LE
Mechanical
Dimensions2.9" x 1.6" x 1.2" (7.4 cm x 4.1 cm x 3.0 cm)
Weight0.16 lbs (73 g)
Environmental
Operating Temperature5°C to +40°C, ambient humidity
Storage Temperature-40°C to +70°C, ambient humidity
Operating/ Storage Humidity10% to 95%, non-condensing
Atmospheric Pressure540 mBar to 1060 mBar
Mode of Operation per IEC 60601-1
Mode of OperationContinuous

2.0 Intended Use/ Indications For Use

The Masimo MightySat-OTC is intended for the spot-checking of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR).

The Masimo MightySat-OTC is indicated for use with individuals 18 years and older who are well or poorly perfused under no motion conditions.

The Masimo MightySat-OTC is not intended for diagnosis or screening of lung disease and treatment decisions using the device should only be under the advice of a healthcare provider.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Masimo Corporation. The logo consists of a red circle with a white checkmark inside, followed by the word "Masimo" in black, bold letters. To the right of the logo is the text "MASIMO CORPORATION" in gray, followed by the address "52 Discovery, Irvine, CA 92618" also in gray.

3.0 Technological Characteristics

Principle of Operation

There has been no change to the principle of operation as part of this submission. The subject device, MightySat-OTC, utilizes the same principle of operation as the predicate device cleared under (K181956). The MightySat-OTC still relies on the principle that light is absorbed differently for oxygenated hemoglobin and deoxygenated hemoglobin and the ratio of the absorptions can be used to monitoring the changes in oxygen saturation and pulse rate. However, as part of the MightySat-OTC, the calibration curve has been adjusted to allow for improved accuracy performance, including across different skin pigments and genders.

Mechanism of Action for Achieving the Intended Effect

There has been no change to the mechanism of action as part of this submission. The subject device MightySat-OTC utilizes the same mechanism of action as the predicate device cleared under (K181956). MightySat automatically turns on when the device is opened. The device is then positioned on the patient's finger. Once the device is applied to the finger, it collects and processes physiological signals, and then displays the measurements on the device's display screen. The MightySat automatically turns off after removing the device from the finger.

4.0 Summary of Technological Characteristics of the subject device compared to the predicate device

Similarities and Differences between Predicate and Subject Device

The subject device, MightySat-OTC, and the predicate device, MightySat (K181956), are identical in design and have the following key similarities:

  • . Both devices have the same intended use;
  • Both devices have the same technological characteristics;
  • Both devices use the same principle of operation and mechanism of action; ●
  • Both devices include wireless Bluetooth feature for transferring patient information from the device to a mobile device, such as a smartphone;

The subject device. MightySat-OTC, and the predicate device, MightySat (K181956), have the following key differences:

  • The subject device is to be marketed for Over-The-Counter (OTC) use;
  • The subject device is indicated for a subset of populations;
  • The subject device is provided with an updated calibration curve for improved performance, ● including across different skin pigments and genders;
  • . The subject device does not provide the RRp and PVi features.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Masimo Corporation. The logo consists of a red circle with a white checkmark inside, followed by the word "Masimo" in a bold, sans-serif font. To the right of the word "Masimo" is the text "MASIMO CORPORATION" in a smaller font, followed by the address "52 Discovery, Irvine, CA 92618".

Between the subject device and the predicate device, there was no change in the intended use. To address the difference in the OTC marketing, the indication for use has been modified to be more appropriate for lay users who would most benefit from the OTC availability of a medical pulse oximeter. To mitigate concerns around the OTC availability as compared to the predicate, the indications for use statement excludes the use for selfdiagnosis or self-screening of medical conditions.

To support the OTC availability does not affect the safety and effectiveness, consideration was made to the factors that may be necessary for a pulse oximeter to be used as an OTC device. These factors included the usability and the demonstrated effectiveness during low blood circulation, different genders, and dark to light skin. Additionally, the labeling was made appropriate for lay users to mitigate any remaining residual risks.

There are no changes to the principle of operation or the mechanism of action from the predicate device. However, there is an update to calibration curve to allow for improved performance, including across different skin pigments and genders. The subject device also provides a reduced number of parameters from the predicate. Otherwise, there are no technological differences for the parameters provided on both devices.

The differences did not change the intended use of the subject device from the predicate. Provided the same intended use and same technological characteristics, the subject device was found to be substantially equivalent to the predicate.

See Table 4.0-1 for the comparison between the subject and predicate devices.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Masimo Corporation. The logo consists of the word "Masimo" in a bold, sans-serif font, with a red circle containing a white checkmark to the left of the word. Below the logo is the address "52 Discovery, Irvine, CA 92618".

Table 4.0-1 Comparison between Subject and Predicate Device
FeatureMightySat - OTC, Subject DeviceMightySat Fingertip Pulse Oximeter,PredicateComparison
510(k) NumberPendingK181956
General Information
Indications for UseThe Masimo MightySat-OTC is intended for thespot-checking of functional oxygen saturation ofarterial hemoglobin (SpO2) and pulse rate (PR).The Masimo MightySat-OTC is indicated foruse with individuals 18 years and older who arewell or poorly perfused under no motionconditions.The Masimo MightySat-OTC is not intended fordiagnosis or screening of lung disease andtreatment decisions using the device should onlybe under the advice of a healthcare provider.The Masimo MightySat Fingertip PulseOximeter is intended for hospitals, hospital-typefacilities, home environments, and transport.The Masimo MightySat Fingertip PulseOximeter is indicated for the noninvasive spot-checking of functional oxygen saturation ofarterial hemoglobin (SpO2) and pulse rate (PR)for adult and pediatric patients during both nomotion and motion conditions, and for patientswho are well or poorly perfused.The Masimo MightySat Fingertip PulseOximeter is indicated for the non-invasive spotchecking of respiration rate (RRp) for adult.Same. The devices have the sameintended use, but the subjectdevice's indications for use hasbeen modified to be appropriatefor OTC lay users. There is nochange in the intended use fromthe predicate device.The subject device also providesa subset of the parameterssupported by the predicatedevice. The subject device doesnot provide PVi and RRpfeatures. These features aresupplemental to the SpO2 andPulse Rate monitoring andtherefore does not change theintended use from the predicate.
Classification Regulation/Product Code21 CFR 870.2700, Class II/OLK21 CFR 870.2700, Class II/DQA21 CFR 868.2375, Class II/BZQSame as predicate, except thesubject device is for OTC useand does not provide respirationrate monitoring.
OTCRxDifferent. Subject device isindicated for OTC use.
Table 4.0-1 Comparison between Subject and Predicate Device
FeatureMightySat - OTC, Subject DeviceMightySat Fingertip Pulse Oximeter, PredicateComparison
510(k) NumberPendingK181956
Principle of operationPulse oximetry is governed by the following principles:Oxyhemoglobin (oxygenated blood) and deoxyhemoglobin (non-oxygenated blood) differ in their absorption of red and infrared light (spectrophotometry).The amount of arterial blood in tissue changes with your pulse (photoplethysymography). Therefore, the amount of light absorbed by the varying quantities of arterial blood changes as well.Pulse oximetry is governed by the following principles:Oxyhemoglobin (oxygenated blood) and deoxyhemoglobin (non-oxygenated blood) differ in their absorption of red and infrared light (spectrophotometry).The amount of arterial blood in tissue changes with your pulse (photoplethysymography). Therefore, the amount of light absorbed by the varying quantities of arterial blood changes as well.Respiration rate (RRp) is governed by the following principle:Measures the respiration rate by analyzing cyclic variations in the photoplethysmogramSame as predicate, except the subject device does not provide RRp and PVi features.
Indicated Population18 years and olderAdult, PediatricSimilar. Subject device is indicated for a subset of populations supported by the predicate.
DisplayTypeColor displayColor displaySame as predicate
Table 4.0-1 Comparison between Subject and Predicate Device
FeatureMightySat - OTC, Subject DeviceMightySat Fingertip Pulse Oximeter,PredicateComparison
510(k) NumberPendingK181956
Measurement Ranges
Oxygen Saturation(SpO2)0-100%0-100%Same as predicate
Pulse Rate (PR)25-240 beats per minute (BPM)25-240 beats per minute (BPM)Same as predicate
Perfusion Index (Pi)0.02-20%0.02-20%Same as predicate
Accuracy
SpO2, No Motion70 - 100%, 1.5% ARMS, 18 years and older70 - 100%, 2% ARMS, adults/pediatricsDifferent. Subject deviceprovides improvedperformance.Testing is provided to supportsubstantial equivalence.
SpO2, Low Perfusion70 - 100%, 2%, ARMS, 18 years and older70 - 100%, 2%, ARMS, adults/pediatricsSame as predicate
Pulse Rate, No Motion25 – 240 BPM, 3 BPM ARMS, 18 years andolder25 – 240 BPM, 3 BPM ARMS, adults/pediatricsSame as predicate
Pulse Rate, LowPerfusion25 – 240 BPM, 3 BPM ARMS, 18 years andolder25 – 240 BPM, 3 BPM ARMS, adults/pediatricsSame as predicate
Display Resolution
SpO21%1%Same as predicate
Pulse Rate1 BPM1 BPMSame as predicate
Biocompatibility
MaterialsPatient contacting materials are biocompatibleper ISO 10993-1Patient contacting materials are biocompatibleper ISO 10993-1Same as predicate
Environmental
Operating Temperature41°F to 104 °F (5°C to +40°C)41°F to 104 °F (5°C to +40°C)Same as predicate
Table 4.0-1 Comparison between Subject and Predicate Device
FeatureMightySat – OTC, Subject DeviceMightySat Fingertip Pulse Oximeter, PredicateComparison
510(k) NumberPendingK181956
Storage Temperature-40°F to 158°F (-40°C to 70°C)-40°F to 158°F (-40°C to 70°C)Same as predicate
Operational/Storage Humidity10 to 95%, non-condensing10 to 95%, non-condensingSame as predicate
Atmospheric Pressure540 to 1060 mBar540 to 1060 mBarSame as predicate
Mechanical
General Overall Dimension2.9" x 1.6" x 1.2" (7.4 cm x 4.1 cm x 3.0 cm)2.9" x 1.6" x 1.2" (7.4 cm x 4.1 cm x 3.0 cm)Same as predicate
Weight0.16 lbs (73g)0.16 lbs (73g)Same as predicate
Electrical
Type of powerInternal batteries, two standard “AAA” batteriesInternal batteries, two standard “AAA” batteriesSame as predicate
Electrical Safety/EMCIEC 60601-1/ IEC 60601-1-2IEC 60601-1/ IEC 60601-1-2Same as predicate
Interface
WirelessBluetoothBluetoothSame as predicate
Mode of Operation per IEC 60601-1
Mode of OperationContinuousContinuousSame as predicate

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for Masimo Corporation. The logo consists of the word "Masimo" in a bold, sans-serif font, with a red circle containing a white checkmark to the left of the word. To the right of the logo is the text "MASIMO CORPORATION" followed by the address "52 Discovery, Irvine, CA 92618".

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for Masimo Corporation. The logo consists of a red circle with a white checkmark inside, followed by the word "Masimo" in bold, black letters. To the right of the logo is the text "MASIMO CORPORATION" in gray, followed by the address "52 Discovery, Irvine, CA 92618" also in gray.

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the logo for Masimo Corporation. The logo consists of a red circle with a white checkmark inside, followed by the word "Masimo" in black, bold letters. To the right of the logo is the text "MASIMO CORPORATION" in gray, followed by the address "52 Discovery, Irvine, CA 92618".

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the logo for Masimo Corporation. The logo consists of a red circle with a white checkmark inside, followed by the word "Masimo" in bold, black letters. To the right of the company name is the company's address, which is "52 Discovery, Irvine, CA 92618".

5.0 Performance Data

Performance Bench Testing

Performance bench testing is included as part of this submission to support the device performance and to support the substantial equivalence to the predicate device (K181956).

Additionally, to support the performance of the subject device with the updated calibration curve under low perfusion conditions, bench testing using a low perfusion simulator was provided as part of this submission. To support the dynamic range of the signal detection under low perfusion, additional bench testing with opacity filters to simulate different skin pigments was performed.

Biocompatibility Testing

Additional biocompatibility testing was not included as part of this submission as there was no change to the patient contacting materials from the previously cleared predicate device (K181956).

Electromagnetic Compatibility, Electrical Safety, Environmental, Mechanical and Cleaning

Additional electromagnetic compatibility, electrical safety, environmental, and cleaning testing was not included as part of this submission as there was no change to the mechanical, hardware or electrical components from the previous clearance of the predicate (K181956).

Software Verification and Validation Testing

Masimo has established, implemented, and maintains procedures for software design, development, review, verification, and validation of its products in accordance with FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (May 11, 2005).

Consistent with the previously cleared predicate device (K181956), the subject device software is considered "Moderate" level of concern. This is because a failure or latent flaw in the device software could indirectly result in minor injury to the patient or operator through incorrect or delayed information or through the action of a care provider.

Software verification and validation testing was included as part of this submission to support the substantial equivalence of the subject device.

Wireless and Cybersecurity Testing

There were no changes made to the wireless capabilities of the MightySat-OTC from the predicate device (K181956) and therefore no additional wireless or cybersecurity testing was included as part of this submission.

Human Factors and Usability Testing

There were no changes made to the usability of the MightySat-OTC as part of this submission from the

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image shows the logo for Masimo Corporation. The logo consists of a red circle with a white checkmark inside, followed by the word "Masimo" in black, bold letters. To the right of the word "Masimo" is the text "MASIMO CORPORATION" followed by the address "52 Discovery, Irvine, CA 92618".

previous clearance (K181956).

To support the OTC use of the subject device, additional human factors validation testing (self-selection and knowledge task) was performed to validate users can understand labeling materials and correctly interpret the device outputs. The validation testing included a self-selection process in accordance with the FDA Guidance, Self-Selection Studies for Nonprescription Drug Products, to identify the intended user to participate in the human factors and usability validation of the device.

Clinical Testing

To support the improved use by lay-users, the calibration curve was updated to improve the overall performance of the MightySat-OTC including across different skin pigments and genders to 1.5 % Arms.

To support the performance of the MightySat-OTC, clinical data is provided as part of this submission. The clinical performance testing included data from 39 healthy volunteer subjects with varying skin pigmentation, which was measured using the color-based Massey-Martin scale. The results of the clinical study were also separated by skin pigment groups to support the performance of the MightySat-OTC across different skin pigment groups.

The clinical performance testing was conducted in accordance with ISO 80601-2-61 and the prospective study included 39 healthy volunteer subjects, including 22 dark and 17 light pigmented subjects with at least 7 subjects in each gender and skin pigment bin, which exceeded the ISO 80601-2-61 minimum sample size requirements and the minimum number of dark pigmented subjects in accordance with the FDA Guidance for Pulse Oximeters.

The overall performance was found to be 1.26 % Arms across the 70%-100% SaO2 range and 1.22 % Arms in the 85%-95% SaO2 range. The summary of the performance in these saturation ranges are provided in the table below.

SaO2 RangeNumberSubjectsBiasPrecisionARMSLimits ofAgreementData Pairs
70%-100%39-0.021.261.26[-2.50 2.45]2030
85%-95%39-0.351.171.22[-2.63, 1.94]738

Additional sub-analysis was performed based upon skin pigment and gender to support the performance of the subject device across different skin pigments and genders was within the 1.5 % Arms:

Sub-GroupNumber SubjectsSaO2 RangeBiasPrecisionARMS (%)Limits of AgreementData Pairs
Light Male1070%-100%-0.131.261.27[-2.61, 2.35]384
Dark Male1570%-100%0.111.441.44[-2.70, 2.93]612
Light Female770%-100%-0.061.111.11[-2.11, 2.23]258
Dark Female770%-100%-0.181.301.32[-2.73, 2.38]216

The result of the clinical testing supports the performance of the subject device, MightySat-OTC, across

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image shows the logo for Masimo Corporation. The logo consists of a red symbol on the left, followed by the word "Masimo" in a bold, sans-serif font. To the right of the word "Masimo" is the text "MASIMO CORPORATION" in a smaller, sans-serif font, followed by the address "52 Discovery, Irvine, CA 92618".

different skin pigmentation and gender combinations. The subject device demonstrated equitable performance for the intended user population across all the pigment and gender sub-groups and the accuracy was within 1.5% ARMS, without any clinically significant bias compared to the blood reference SaO2, or between the groups.

To further support the equitable performance, the data was bootstrapped for 1001 iterations with replacements. The bootstrapping was conducted both using all the data and also taking account flatteming of the data across the bins. The supported the subject device still met the 1.5 % Arms across the different subgroups.

The data was also analyzed using linear mixed effects (LME) models to support the performance was not impacted gender and skin pigment.

6.0 Conclusion

The data supported the MightySat-OTC is substantially equivalent.

§ 870.2700 Oximeter.

(a)
Identification. An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.(b)
Classification. Class II (performance standards).