K Number
K214090

Validate with FDA (Live)

Device Name
Back 3 COLOR
Manufacturer
Date Cleared
2022-11-01

(308 days)

Product Code
Regulation Number
878.4400
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The BACK3 COLOR device is intended to provide topical heating for the purpose of elevating tissue temperature for the treatment of selected medical conditions such as relief of pain, muscle spasms, and increase in local circulation. The BACK3 COLOR massage device is intended to provide a temporary reduction in the appearance of cellulite.

Device Description

The BACK3 COLOR generates high frequency sinusoidal current with a monopolar mode of application using two electrodes. A fixed electrode is placed in contact with the patient and a handheld electrode is manipulated by a therapist. When both electrodes are in contact with a patient the electrical circuit is closed and RF therapy can be provided. The device can be operated in a capacitor resistive monopolar mode and a multipolar mode.

The product consists of a power console on a moveable trolley, LCD monitor, and accessories including capacitive resistive electrodes and multipolar electrodes. The unit can be adjusted to provide various levels of treatment frequency ranging from 300kHz to 1MHz.

AI/ML Overview

The provided text describes the BACK3 COLOR device, which is an electrosurgical cutting and coagulation device. The 510(k) summary focuses on establishing substantial equivalence to a predicate device (Winback Back 3SE, K162828) rather than presenting specific performance criteria and a study to prove meeting those criteria in the traditional sense of a clinical trial for diagnostic devices.

Instead, the acceptance criteria and study detailed here are focused on demonstrating that the new device (BACK3 COLOR) functions safely and effectively, and is technologically equivalent to the previously cleared predicate device, without introducing new questions of safety or effectiveness.

Here's a breakdown of the information requested, based on the provided text:

1. A table of acceptance criteria and the reported device performance

Since this is a substantial equivalence submission, the "acceptance criteria" are generally that the new device performs equivalently to the predicate device in terms of safety and technical specifications. The "reported device performance" is essentially the confirmation that these specifications are met for the new device.

Acceptance Criteria (Equivalence to Predicate)Reported Device Performance (BACK3 COLOR)
Regulation and Product Classification Code: 21 CFR 878.4400, PBXMeets (Identical)
Indications for Use: Provides topical heating for pain relief, muscle spasms, increased local circulation, and temporary reduction in cellulite.Meets (Identical)
Massaging Handpiece: YesMeets (Identical)
Electrode Shapes: Square and circularMeets (Identical)
Infrared Light: NoMeets (Identical)
Vacuum (suction): NoMeets (Identical)
Treatment Activation: Finger selection on consoleMeets (Identical)
RF Type: Multipolar/UnipolarMeets (Identical)
RF Frequency: 300kHz – 1MHzMeets (Identical)
Max RF Power: 300WMeets (Identical)
Intensity Adjustment: 0-100%Meets (Identical)
Configuration: Cart mounted console with accessoriesMeets (Identical)
Patient Safety Switch: YesMeets (Identical)
Electrical Safety: Compliance with IEC 60601-1Satisfactory results reported
Electromagnetic Compatibility: Compliance with IEC 60601-1-2Satisfactory results reported
Biocompatibility: Tissue contacting probes tested for cytotoxicity, sensitization, intracutaneous reactivitySatisfactory results reported
Software Assessment: Compliance with FDA software validation guidelines (Levels of Concern, User & System Requirements, Hazard Analysis, Software Requirements, Architectural Design, Software Validation & Testing)Satisfactory results reported
Usability and Risk Management: Verification of user interface, safety features, and satisfactory performance using worse-case assumptionsSatisfactory results reported
Ability to maintain therapeutic temperature on human skin surface: Demonstrated for all applicators, operation modes, and power settingsStudy conducted with satisfactory results

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample Size: The "Human being testing" for tissue temperature elevation involved three different people.
  • Data Provenance: The document does not specify the country of origin or whether the study was retrospective or prospective. Given it's a testing requirement for the device's performance, it is highly likely to be a prospective study.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This type of information is not typically applicable or detailed in a substantial equivalence filing for an electrosurgical device like this. The specific "Human being testing" described focused on physical measurements (temperature) directly from the device's application, rather than subjective interpretation by experts to establish a "ground truth" for a diagnostic outcome. Therefore, no experts for "ground truth" establishment in this sense are mentioned.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. The human being testing involved direct measurement of physical parameters (skin and room temperature) rather than subjective assessments that would require adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is an electrosurgical and radiofrequency therapy device, not a diagnostic imaging device with AI assistance for human readers. No MRMC study was mentioned or required.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This device delivers energy therapy; it is not an algorithm-based diagnostic tool. The "Software Assessment" done was for the device's operational software, not for a standalone diagnostic algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For the "Human being testing," the "ground truth" was the direct measurement of skin and room temperatures to demonstrate the device's ability to maintain a therapeutic temperature. This is a direct physical measurement, not a clinical ground truth like pathology or expert consensus.

8. The sample size for the training set

Not applicable. This document describes a medical device for physical therapy, not a machine learning model that requires a training set in that context. The "Software Assessment" refers to validation testing of the device's operational software.

9. How the ground truth for the training set was established

Not applicable, as there is no mention of a training set for a machine learning model.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol is a stylized representation of a human figure, while the FDA name is written in blue and includes the words "U.S. Food & Drug Administration".

11/1/2022

SWIMS America Corp % Mr. Matthieu COMMEAU Managing Director SWIMS America Corp, 1133 Westchester Avenue Suite N 220 White Plains, New York 10604

Re: K214090

Trade/Device Name: BACK3 COLOR Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: PBX Dated: September 30, 2022 Received: October 3, 2022

Dear Mr. COMMEAU:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K214090

Device Name BACK3 COLOR

Indications for Use (Describe)

The BACK3 COLOR device is intended to provide topical heating for the purpose of elevating tissue temperature for the treatment of selected medical conditions such as relief of pain, muscle spasms, and increase in local circulation. The BACK3 COLOR massage device is intended to provide a temporary reduction in the appearance of cellulite.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
---------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff

PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Premarket Notification SWIMS America Corp K214090

510(k) Summary Pursuant to 21 CFR 807.92

Date: 10/31/2022

1.Submitted By:SWIMS America Corp1133 Westchester Avenue Suite N 220White Plains, NY 10604
2.Contact Person:Mr. Matthieu COMMEAU,Managing Director of SWIMS America Corp,1133 Westchester Avenue Suite N 220White Plains, NY 10604Phone: 917-371-7388Email: mat@winback.com
3.Common Name:Winback Back 3SE Massager, Radiofrequency Induced Heat
4.Trade Name:BACK3 COLOR
5.Classification:Class II
6.Device Product Code:PBX

Description:

The BACK3 COLOR generates high frequency sinusoidal current with a monopolar mode of application using two electrodes. A fixed electrode is placed in contact with the patient and a handheld electrode is manipulated by a therapist. When both electrodes are in contact with a patient the electrical circuit is closed and RF therapy can be provided. The device can be operated in a capacitor resistive monopolar mode and a multipolar mode.

The product consists of a power console on a moveable trolley, LCD monitor, and accessories including capacitive resistive electrodes and multipolar electrodes. The unit can be adjusted to provide various levels of treatment frequency ranging from 300kHz to 1MHz.

Intended Use:

The BACK3 COLOR device is intended to provide topical heating for the purpose of elevating tissue temperature for the treatment of selected medical conditions such as relief of pain, muscle spasms, and increase in local circulation. The BACK3 COLOR massage device is intended to provide a temporary reduction in the appearance of cellulite.

{4}------------------------------------------------

510(k) Premarket Notification SWIMS America Corp K214090

Substantial Equivalence/Technological Characteristics:

The BACK3 COLOR device is substantially equivalent to the Back 3SE device from Winback USA Corp which was cleared under premarket notification K162828. Both devices are cart mounted consoles with electrode accessories capable of operation in monopolar and multipolar modes in the range of 300kHz to 1MHz radiofrequency.

Both devices operate in the same treatment range and feature intensity adjustments from 0 to 100%. Electrical safety and biocompatibility have been established for both devices. No direct comparison was made since there are no significant differences in operation and test results indicate identical safety.

The table below summarizes the equivalence of the devices.

Element of Comparison510(k) Device:BACK3 COLORK214090Predicate Device:Winback Back 3SEK162828Explanation ofDifferences
Regulation and ProductClassification Code21 CFR 878.4400PBX21 CFR 878.4400PBXNone
Indications for UseThe BACK3 COLOR device isintended to provide topical heatingfor the purpose of elevating tissuetemperature for the treatment ofselected medical conditions such asrelief of pain, musclespasms, and increase in localcirculation.The BACK3 COLOR massagedevice is intended to provide atemporary reduction in theappearance of cellulite.The Winback Back 3SE deviceis intended to provide topicalheating for the purpose ofelevating tissue temperature forthe treatment of selectedmedical conditions such asrelief of pain, musclespasms, and increase in localcirculation.The Winback Back 3SEmassage device is intended toprovide a temporary reductionin the appearance of cellulite.Identical
Massaging Hand pieceYesYesIdentical
Electrode ShapesSquare and circularSquare and circularIdentical
Infrared LightNoNoIdentical
Vacuum (suction)NoNoIdentical
Treatment ActivationFinger selection on consoleFinger selection on consoleIdentical
RF TypeMultipolar/UnipolarMultipolar/UnipolarIdentical
RF Frequency300kHz – 1MHz300kHz – 1MHzIdentical
Max RF Power300W300WIdentical
Intensity Adjustment0-100%0-100%Identical
ConfigurationCart mounted console withaccessoriesCart mounted console withaccessoriesIdentical
Patient Safety SwitchYesYesIdentical

Predicate Device Comparison Table

{5}------------------------------------------------

510(k) Premarket Notification

K214090
Accessories providedwith the device (diameterof the electrodes-mm)CET handpieceCET HandpieceAddition of some newaccessories which doesnot impact the securityand the performance ofthe subject device.These accessories havebeen added to facilitatethe treatment of thetherapist depending onthe area to be treated.These accessories havehowever the sameutility, the sameintended uses and theassociated modes areidentical to those of thepredicate (CET, RET...etc).
RET handpieceRET Handpiece
CET straightCET electrodes (40, 60, 70mm)
CET electrodes (40, 60, 70mm)RET electrodes (40, 60,70mm)
RET electrodes (40, 60, 70mm)Active cable for CET electrodes
Convex CET electrodesActive cable for RET electrodes
Thai CET ElectrodeNeutral plate
Convex RET ElectrodeSafety switch
Protect Ring3-Polar applicator for Body
Neutral plate6- Polar applicator for Body
Physio Blade and Connector3- Polar applicator for Face
RET Bracelet
Capacitive Fix Pad
Resistive Fix Pad
Neutral Fix Pad
Finger handpiece
Flex Pack
TECAR 6.0 Handpiece
TECAR 6.0 Tip (M)
3-Polar applicator for Body
3-Polar applicator for Face
Safety Switch
Double Cable
Active cable for CET electrodes
Active cable for RET electrodes
ModesCETCETAddition of some newmodes which does notimpact the security andthe performance of thedevice. These modeshave been added tofacilitate the treatmentof the therapistdepending on the areato be treated. Thesemodes have howeverthe same utility and thesame intended uses asthose of the predicate(CET, RET, etc).
RETRET
DEEP RETDEEP CET
DEEP CETFACE 3-POLAR
PULSEBODY 3-POLAR
FRACTALBODY 6-POLAR
FACE 3-POLAR
BODY 3-POLAR
The FRACTAL mode isonly compatible withthe use of the accessoryTECAR 6.0.Its frequency is thesame as the RET. Thismode has for aim todeliver TECAR energy
in fractional form whichis divided into severalheating points for afocused and targeteddeep action.
FunctionsMINILOWPULSEAddition of some newfunctions which does
MEDIUMCONTINUOUSnot impact the security
BOOSTBOOSTSUPERPULSESUPERPULSEnot impact the security
SWAPNORMALand the performance of
DYNAMICDYNAMICthe device. These
BOOSTmodes have been added
to facilitate the
treatment of the
therapist depending on
the area to be treated.
These modes have

{6}------------------------------------------------

510(k) Premarket Notification

SWIMS America Corp

K214090
however the sameutility and the sameintended uses as thoseof the predicate (CET,RET, etc).Change of name offunctions:The name of thefunctions has beenmodified. However, thefunctions are the sameas those of thepredicate.LOWPULSE functionwas renamed by MINI.CONTINUOUS wasrenamed by MEDIUMandSUPERPULSE byBOOST.
DesignFront view:Image: Front view of device with annotationsFront view:Image: Front view of deviceAesthetic difference:The user interfacedesign evolved: graphicand color changes.
User Interface:Image: User interface of deviceUser Interface:Image: User interface of device
Size (height x width xdepth)- in millimeters148x306x358148x306x358Identical
Weight5 kg5 kgIdentical

{7}------------------------------------------------

510(k) Premarket Notification SWIMS America Corp K214090

Summary of Testing:

The technological characteristics of the BACK3 COLOR System has been verified based on assessments of electrical safety, performance, biocompatibility, software and usability.

The following testing has been conducted with satisfactory results:

  • BACK3 COLOR Usability and Risk Management: Usability and Risk Management ● assessments were done using worse-case assumptions to verify user interface, safety features and satisfactory performance.
  • Biocompatibility: Samples of the tissue contacting probes were tested for cytotoxicity, sensitization and intracutaneous reactivity.
  • Software Assessment: Software features were assessed in accordance with FDA software validation guidelines. Levels of Concern, User & System Requirements, Hazard Analysis, Software Requirements, Architectural Design, Software Validation & Testing were all addressed.
  • . Electromagnetic Compatibility: EMC testing was done to evaluate emissions and immunity to electromagnetic fields in accordance with IEC 60601-1-2.
  • . Electrical Safety: Full electrical safety testing was done in compliance with IEC 60601-1.
  • Human being testing : a tissue temperature elevation study was conduct to demonstrate ability of all applicators of subject device to maintain therapeutic temperature on the surface of the human skin. The study was conduct on three different people on three body parts with all device operation modes and at the lowest and highest power settings. The skin and room temperatures were measured.

Conclusion:

The BACK3 COLOR is substantially equivalent to the predicate device (K162828). Both devices operate in the same treatment range and feature intensity adjustments from 0 to 100%. Electrical safety and biocompatibility have been established for both devices.

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.