K Number
K214054

Validate with FDA (Live)

Date Cleared
2022-10-14

(291 days)

Product Code
Regulation Number
872.3750
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

TP Orthodontics Light Cure Adhesive System is indicated for use as an orthodontic bonding agent for metal brackets, ceramic brackets, and buccal tubes to the tooth's surface.

TP Orthodontics Pre-Applied Adhesive is indicated for use in orthodontic appliance application and bonding for orthodontic treatment.

TP Orthodontics Echant is indicated for etching enamel and dentin to promote adhesion of primer/bonding agent adhesives to tooth structure and restorative materials.

Device Description

TP Orthodontics Light Cure Adhesive System and TP Orthodontics Pre-Applied Adhesive are light cure orthodontic adhesive used for bonding the orthodontic brackets, molar tubes, and buccal tubes to the surface of teeth for orthodontic treatment. The variations available are:

  • . TP Orthodontics Light Cure Adhesive System, which consists of an adhesive paste (will be available in a pre-loaded syringe), and sealant.
  • . TP Orthodontics Pre-Applied Adhesive, consisting of bracket adhesive pre-applied to the bracket pad.

The adhesives prevent bracket drift during placement, and is easy to clean-up excess (flash') prior to curing. The adhesive is polymerized ('cured') by exposure to a dental curing light, and once cured it creates a bond, attaching the orthodontic appliance to the teeth.

TP Orthodontics Etchant is a phosphoric acid etchant used on tooth enamel or dentin in preparation for orthodontic bonding. It is a blue gel and will be contained in a syringe, which will use disposable dispensing tips for its controlled application.

AI/ML Overview

The provided text is a 510(k) summary for dental adhesive products and an etchant, not an AI/ML medical device. Therefore, it does not contain the information required to answer the prompt regarding acceptance criteria, study details (sample size, data provenance, expert ground truth, adjudication, MRMC, standalone performance), or training set information for an AI/ML device.

The document explicitly states:

  • "No clinical performance testing was conducted on TP Orthodontics Light Cure Adhesive System, TP Orthodontics Pre-Applied Adhesive, and TP Orthodontics Etchant."
  • "TP Orthodontics Light Cure Adhesive System, TP Orthodontics Pre-Applied Adhesive and TP Orthodontics Etchant bench testing was performed. The proposed device passed testing and performed similar to its predicate devices (test reports are attached to this submission)."

This means the submission relies on bench testing demonstrating similarity to predicate devices, rather than clinical studies or AI/ML performance evaluations.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 14, 2022

TP Orthodontics, Inc. Cristiane Muller Clinical & Regulatory Affairs Manager 100 Center Plaza La Porte, Indiana 46350-9672

Re: K214054

Trade/Device Name: TP Orthodontics Light-Cure Adhesive System, TP Orthodontics Pre-Applied Adhesive, TP Orthodontics Etchant208520 Regulation Number: 21 CFR 872.3750 Regulation Name: Bracket Adhesive Resin And Tooth Conditioner Regulatory Class: Class II Product Code: DYH, KLE Dated: September 7, 2022 Received: September 16, 2022

Dear Cristiane Muller:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For Michael E. Adjodha, M. ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K214054

Device Name

TP Orthodontics Light Cure Adhesive System, TP Orthodontics Pre-Applied Adhesive and TP Orthodontics Etchant

Indications for Use (Describe)

TP Orthodontics Light Cure Adhesive System is indicated for use as an orthodontic bonding agent for metal brackets, ceramic brackets, and buccal tubes to the tooth's surface.

TP Orthodontics Pre-Applied Adhesive is indicated for use in orthodontic appliance application and bonding for orthodontic treatment.

TP Orthodontics Echant is indicated for etching enamel and dentin to promote adhesion of primer/bonding agent adhesives to tooth structure and restorative materials.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for TP Orthodontics, Inc. The logo consists of a stylized globe with an arrow wrapping around it on the left side. To the right of the globe is the text "TP Orthodontics, Inc."

September 7, 2022

510(k) Summary - K214054

Submitter Information

TP Orthodontics, Inc. 100 Center Plaza, La Porte, IN 46350-9672 Contact Person: Cristiane Muller Phone Number: 219.785.2591 Fax: 219.324.3029 E-mail: regulatory@tportho.com

Device Information

Trade Name: TP Orthodontics Light Cure Adhesive System, TP Orthodontics Pre-Applied Adhesive and TP Orthodontics Etchant Common Name: Adhesive, Bracket and Tooth Conditioner, Resin Product Code: DYH Classification Name: Bracket Adhesive Resin and Tooth Conditioner Regulation Number: 872-3750 Device Class: II Classification Panel: Dental

Primary Predicate Device

DeviceApplicant510(k) Number
BracePaste AdhesiveAmerican OrthodonticsK160782

Secondary Predicate Devices

DeviceApplicant510(k) Number
APC Plus Adhesive3M Unitek Corp.K020394
Acid EtchantAmerican OrthodonticsK172953

Reference Predicate Devices

DeviceApplicant510(k) Number
Transbond XT Light Cured Orthodontic Adhesive/Transbond XT PrimerUnitek Corp.K880393
APC Flash-Free AdhesiveUnitek Corp.K113197

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for TP Orthodontics, Inc. The logo consists of a stylized globe with an arrow wrapping around it on the left. To the right of the globe is the text "TP Orthodontics, Inc." in a simple, sans-serif font. The logo is clean and professional, suggesting a global reach for the company.

Indications for Use

TP Orthodontics Light Cure Adhesive System is indicated for use as an orthodontic bonding agent for metal brackets, ceramic brackets, and buccal tubes to the tooth's surface.

TP Orthodontics Pre-Applied Adhesive is indicated use in orthodontic appliance application and bonding for orthodontic treatment.

TP Orthodontics Acid Etchant is indicated for etching enamel and dentin to promote adhesion of primer/bonding agent adhesives to tooth structure and restorative materials.

Device Description and Summary of Technological Characteristics

TP Orthodontics Light Cure Adhesive System and TP Orthodontics Pre-Applied Adhesive are light cure orthodontic adhesive used for bonding the orthodontic brackets, molar tubes, and buccal tubes to the surface of teeth for orthodontic treatment. The variations available are:

  • . TP Orthodontics Light Cure Adhesive System, which consists of an adhesive paste (will be available in a pre-loaded syringe), and sealant.
  • . TP Orthodontics Pre-Applied Adhesive, consisting of bracket adhesive pre-applied to the bracket pad.

The adhesives prevent bracket drift during placement, and is easy to clean-up excess (flash') prior to curing. The adhesive is polymerized ('cured') by exposure to a dental curing light, and once cured it creates a bond, attaching the orthodontic appliance to the teeth.

TP Orthodontics Etchant is a phosphoric acid etchant used on tooth enamel or dentin in preparation for orthodontic bonding. It is a blue gel and will be contained in a syringe, which will use disposable dispensing tips for its controlled application.

{5}------------------------------------------------

Table 5.1. Substantial Equivalence Comparison

CharacteristicTP OrthodonticsLight Cure AdhesiveSystemBracePasteTransbond XT (Light CuredOrthodontic Adhesive/Primer)
DeviceProposed DevicePredicate DeviceReference Predicate
510(k)To be determinedK160782K880393
ManufacturerTP Orthodontics, Inc.American OrthodonticsUnitek Corp.
Regulation Number21 CFR 872.375021 CFR 872.375021 CFR 872.3750
Device Classification NameBracket Adhesive Resin andTooth ConditionerBracket Adhesive Resin andTooth ConditionerBracket Adhesive Resin andTooth Conditioner
Product CodeDYHDYHDYH
Device ClassClass IIClass IIClass II
Indications for UseIndicated for use as anorthodontic bonding agent formetal brackets, ceramicbrackets, and buccal tubes tothe tooth's surface.Intended for use as anorthodontic bonding agent forMetal Brackets, CeramicBrackets and Buccal Tubesto the tooth's surface.Designed for direct bondingof ceramic orthodonticbrackets and metal brackets.
Dentition to be UsedMixed and PermanentMixed and PermanentMixed and Permanent
Parts that can be BondedFixed orthodontic appliance(Brackets, Tubes,Accessories).Fixed orthodontic appliance(Brackets, Tubes,Accessories).Fixed orthodontic appliance(Brackets, Tubes,Accessories).
Mode of UseLight Cure orthodonticadhesive intended for use inbonding appliances fororthodontic treatment.Adhesive is applied toappliance. Once part ispositioned, visible light isused for curing the adhesiveand by this affix part to tooth.Light Cure orthodonticadhesive intended for use inbonding appliances fororthodontic treatment.Adhesive is applied toappliance. Once part ispositioned, visible light isused for curing the adhesiveand by this affix part to tooth.Light Cure orthodonticadhesive intended for use inbonding appliances fororthodontic treatment.Adhesive is applied toappliance. Once part ispositioned, visible light isused for curing the adhesiveand by this affix part to tooth.
ComponentsPaste and Sealant (primer)PastePaste and Sealant (primer)
Delivery SystemSyringe (paste) and bottlewith dropper (sealant)Syringe/ CarpuleSyringe/ Carpule (paste) andbottle with dropper (sealant)
Curing MechanismOrthodontic Curing LightOrthodontic Curing LightOrthodontic Curing Light
Curing RecommendationsIFU recommends followingcuring light manufacturer'sdirections.Instructions are provided ifusing a 300mW/cm2 andaverage wavelength of 400-500nm curing lightIFU recommends followingcuring light manufacturer'sdirections.Instructions are provided ifusing Blu Ray 3 (3000mW/cm2), which is distributedby American OrthodonticsIFU recommends followingcuring light manufacturer'sdirections.Instructions are provided ifusing 3M's OrtholuxLuminous Curing Light (1600mW/cm2)
Resin BaseMethacrylate MonomersMethacrylate MonomersMethacrylate Monomers
Filler Composition (paste)Silane treated silica andsilane treated glassSilane treated silica andsilane treated glass(aluminum boron silicateglass)Silica and Quartz
CatalystCamphoroquinoneCamphoroquinoneCamphoroquinone
Prescription (Rx) or Over-the-Counter (OTC)RxRxRx
CharacteristicTP OrthodonticsPre-Applied AdhesiveAPC Plus Adhesive
DeviceProposed DevicePredicate Device
510(k)To be determinedK020394
ManufacturerTP Orthodontics, Inc.3M Unitek Corp.
Regulation Number21 CFR 872.375021 CFR 872.3750
Device ClassificationNameBracket Adhesive Resin and Tooth ConditionerBracket Adhesive Resin and Tooth Conditioner
Product CodeDYHDYH, EJF, DYW
Device ClassClass IIClass II
Indications for UseIndicated for use in orthodontic applianceapplication and bonding for orthodontictreatment.Indicated for use in orthodontic applianceapplication and bonding for orthodontictreatment.
Dentition to be UsedMixed and PermanentMixed and Permanent
Bracket TypesTP Orthodontics commercially available metaland ceramic brackets.Unitek commercially available metal andceramic brackets.
Mode of UseLight Cure orthodontic adhesive intended foruse in bonding appliances for orthodontictreatment. Adhesive is already pre-applied toappliance. Once part is positioned, visible lightis used for curing the adhesive and by this affixpart to tooth.Light Cure orthodontic adhesive intended foruse in bonding appliances for orthodontictreatment. Adhesive is already pre-applied toappliance. Once part is positioned, visible lightis used for curing the adhesive and by this affixpart to tooth
Delivery SystemPre-appliedPre-applied
Curing MechanismOrthodontic Curing LightOrthodontic Curing Light
Curing RecommendationsIFU recommends following curing lightmanufacturer's directions.Instructions are provided if using a 300mW/cm2and average wavelength of 400-500nm curinglightIFU recommends following curing lightmanufacturer's directions.Instructions are provided if using 3M's OrtholuxLuminous Curing Light (1600 mW/cm2)
Resin BaseMethacrylate MonomersMethacrylate Monomers
Filler CompositionSilica and GlassSilica, Glass and Quartz.
CatalystCamphoroquinoneCamphoroquinone
Prescription (Rx) or Over-the-Counter (OTC)RxRx

{6}------------------------------------------------

Table 5.2. TP Orthodontics Adhesive Comparison – TPO® Pre-Applied Adhesive

{7}------------------------------------------------

CharacteristicTP Orthodontics EtchantAcid Etchant
DeviceProposed DevicePredicate Device
510(k)To be determinedK172953
ManufacturerTP Orthodontics, Inc.American Orthodontics
Regulation Number21 CFR 872.375021 CFR 872.3200
Device ClassificationNameBracket Adhesive Resin and Tooth ConditionerAgent, tooth bonding, resin
Product CodeDYH, KLEKLE
Device ClassClass IIClass II
Indications for UseTP Orthodontics Etchant is indicated for etchingenamel and dentin to promote adhesion ofprimer/bonding agent adhesives to toothstructure and restorative materials.Acid Etchant is a phosphoric acid etchantused for etching enamel and dentin topromote adhesion of primer/bondingagent adhesives to tooth structure andrestorative materials.
Dentition to be UsedMixed and PermanentMixed and Permanent
Mode of UseApplied to tooth surface to prepare it fororthodontic bonding.Applied to tooth surface to prepare it fororthodontic bonding.
Delivery SystemSyringeSyringe
% Phosphoric Acid37% (CAS 7664-38-2)37% (CAS 7664-38-2)
Material CompositionWater, phosphoric acid, silica, dyeWater, phosphoric acid, silica, dye
ConsistencyThixotropic Gel/Thickened FluidThixotropic Gel/Thickened Fluid
Prescription (Rx) orOver-the-Counter (OTC)RxRx

Table 5.3. TP Orthodontics Adhesive Comparison - TPO Enamel Etchant

Clinical Performance Testing

No clinical performance testing was conducted on TP Orthodontics Light Cure Adhesive System, TP Orthodontics Pre-Applied Adhesive, and TP Orthodontics Etchant.

Non-Clinical Performance Testing

TP Orthodontics Light Cure Adhesive System, TP Orthodontics Pre-Applied Adhesive and TP Orthodontics Etchant bench testing was performed. The proposed device passed testing and performed similar to its predicate devices (test reports are attached to this submission).

Biocompatibility Testing

Biocompatibility testing TP Orthodontics Light Cure Adhesive System and TP Orthodontics Pre-Applied Adhesive was performed per guidance "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" as discussed in the Biocompatibility section of this application.

TP Orthodontics Etchant and the predicate device (American Orthodontics' Acid Etchant) have similar chemical composition, properties and utilize phosphoric acid CAS# 7664-38-2. The biocompatibility of these devices is equivalent; therefore, biocompatibility testing of

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for TP Orthodontics, Inc. The logo consists of a stylized globe with an arrow wrapping around it. The text "TP Orthodontics, Inc." is placed to the right of the globe image. The text is in a simple, sans-serif font.

TP Orthodontics Etchant was not performed.

Substantial Equivalence Conclusion

TP Orthodontics Light Cure Adhesive System and TP Orthodontics Pre-Applied Adhesive present substantially equivalent indication for use, technological characteristics, mechanism of action, composition, and biocompatibility as its predicate devices. There are slight differences in the lanquage of the instructions for use, however these differences do not impact safety and efficacy of the device. TP Orthodontics Light Cure Adhesive System has similar components as its reference predicate device, being comprised of an adhesive paste and liquid sealant/ primer.

TP Orthodontics Etchant presents substantially equivalent indication for use, technological characteristics, mechanism of action, composition, and biocompatibility as its predicate devices. It is concluded that TP Orthodontics Light Cure Adhesive System, TP Orthodontics Pre-Applied Adhesive and TP Orthodontics Etchant are substantially equivalent to their predicate devices.

§ 872.3750 Bracket adhesive resin and tooth conditioner.

(a)
Identification. A bracket adhesive resin and tooth conditioner is a device composed of an adhesive compound, such as polymethylmethacrylate, intended to cement an orthodontic bracket to a tooth surface.(b)
Classification. Class II.