K Number
K213958

Validate with FDA (Live)

Manufacturer
Date Cleared
2022-04-08

(112 days)

Product Code
Regulation Number
888.3030
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Bioabsorbable Anchor is indicated for the fixation of soft tissue grafts in various minimally invasive and open orthopedic surgical procedures in shoulder.

Device Description

The Bioabsorbable Anchor is an absorbable implant device that is intended to provide fixation of a prosthetic material to soft tissue and/or bone. The Bioabsorbable Anchor is composed of polydioxanone (PDO) dyed with D&C Violet No. 2. Two Bioabsorbable Anchors are provided sterile for single use only, packaged preloaded in a disposable anchor inserter instrument, the Dual Anchor Inserter. The preloaded Dual Anchor Inserter is provided in a dual sterile pouch confiquration.

AI/ML Overview

The provided text discusses a 510(k) premarket notification for a medical device called the "Bioabsorbable Anchor." It details the device's description, indications for use, comparison to a predicate device, and performance data required for substantial equivalence determination.

However, the document does not contain information regarding the acceptance criteria and study proving the device meets those criteria in the context of an AI/human-in-the-loop setting, as implied by the detailed questions about sample sizes, expert ground truth, MRMC studies, and standalone algorithm performance.

The "Performance Data" section of the document specifically outlines:

  • Biocompatibility Testing: Conducted in accordance with ISO 10993 standards.
  • Non-Clinical Performance Testing: Included insertion, pullout, and fatigue testing, evaluated in comparison to the predicate device and guided by FDA guidance for "Bone Anchors."
  • In vitro hydrolytic degradation study: Guided by ASTM F1635-16.
  • Absorption Profile: Describes the hydrolytic degradation and tissue response after 13 weeks.

These tests are primarily mechanical, material, and biological evaluations relevant to the physical performance and safety of an implantable device, not an AI/software device that requires extensive clinical validation with human readers or standalone algorithm performance metrics like sensitivity, specificity, or AUC based on expert ground truth.

Therefore, I cannot extract the information required to answer your specific questions from the provided text as it pertains to a different type of medical device evaluation.

{0}------------------------------------------------

April 8, 2022

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.

Embody, Inc. Brianna Schehr Director of Quality, Regulatory & Clinical Affairs 4211 Monarch Way, Suite 500 Norfolk, Virginia 23508

Re: K213958

Trade/Device Name: Bioabsorbable Anchor Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: MAI Dated: March 10, 2022 Received: March 10, 2022

Dear Brianna Schehr:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Laura C. Rose, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K213958

Device Name Bioabsorbable Anchor

Indications for Use (Describe)

The Bioabsorbable Anchor is indicated for the fixation of soft tissue grafts in various minimally invasive and open orthopedic surgical procedures in shoulder.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for Embody. The logo is written in a dark blue font. The "O" in embody is an orange square with white lines inside.

K213958 - 510(k) Summary

SUBMITTER l.

Owner/Submitter:Embody, Inc.4211 Monarch WaySuite 500Norfolk, VA 23508
Phone:236-994-7912
Contact Person:Brianna SchehrDirector of Quality, Regulatory & Clinical Affairs
Date Prepared:April 8, 2022

ll. DEVICE

Name of Device:Bioabsorbable Anchor
Common or Usual Name:Bone Anchor
Classification Name:Fastener, fixation, biodegradable, soft tissue
Regulation Number:21 CFR 888.3030
Regulatory Class:Class II
Product Code:MAI
Medical Specialty:Orthopedic

PREDICATE DEVICE lll.

Rotation Medical Bone Staple (RMB Staple), K131635. This predicate has not been subject to a design-related recall.

ORTHOSORB Resorbable Pins, K111077, is being referenced as a device made of polydioxanone. Therefore, the use of polydioxanone as a fixation material is not new.

DEVICE DESCRIPTION IV.

The Bioabsorbable Anchor is an absorbable implant device that is intended to provide fixation of a prosthetic material to soft tissue and/or bone. The Bioabsorbable Anchor is composed of polydioxanone (PDO) dyed with D&C Violet No. 2. Two Bioabsorbable Anchors are provided sterile for single use only, packaged preloaded in a disposable anchor inserter instrument, the Dual Anchor Inserter. The preloaded Dual Anchor Inserter is provided in a dual sterile pouch confiquration.

V. INDICATION FOR USE

The Bioabsorbable Anchor is indicated for the fixation of soft tissue grafts in various minimally invasive and open orthopedic surgical procedures in shoulder.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the word "EMBODY" in a stylized font. The letters are dark blue, except for the "O", which is inside an orange square with rounded corners. Inside the orange square, there are several diagonal white lines.

Both the subject and predicate devices have the same intended use for the fixation of a prosthetic material.

VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

The Bioabsorbable Anchor is substantially equivalent in terms of indications for use, technological characteristics, and performance characteristics to the predicate device. Fixation is the technological principle for both the subject and predicate devices and is based on the use of a permanent implant in a U-shape polymer strap design with barbed ends, to directly affix a prosthetic material. The Bioabsorbable Anchor is substantially equivalent to the reference device in terms of material composition of an absorbable fixation device.

The following technological differences exist between the subject and predicate devices:

  • The subject device is composed of an absorbable polymer (polydioxanone), whereas the . predicate device is a nonabsorbable polymer (PEEK), and
  • The subject device is optimized for fixation to both soft tissue and bone, whereas the . predicate device is optimized only for fixation to bone.

VII. PERFORMANCE DATA

The following performance data were provided in support of the substantial equivalence determination:

Biocompatibility Testing

The biocompatibility evaluation for the Bioabsorbable Anchor device was conducted in accordance with the FDA quidance document "Use of International Standard ISO 10993-1, 'Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process" dated September 4, 2020, and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1: Evaluation Testing Within a Risk Management Process," as recognized by FDA. The finished device met the requirements of the standards for the following battery of tests; cytotoxicity, sensitization, acute systemic toxicity, systemic toxicity (material-mediated pyrogenicity), bacterial endotoxin testing, genotoxicity, and implantation. Chemical characterization in accordance with ISO 10993-18 and a toxicological risk assessment in accordance with ISO 10993-17 confirmed the Bioabsorbable Anchor is considered safe for its intended use.

The Bioabsorbable Anchor is considered a permanent implant device in contact with tissue/bone, while the anchor inserter instrument is considered tissue/bone contacting for a duration of less than 24 hours.

Non-Clinical Performance Testing

The Bioabsorbable Anchor was evaluated in comparison to the predicate device when applicable in accordance with the FDA guidance document "Bone Anchors—Premarket Notification (510(k)) Submissions" and included the following tests:

  • Insertion Testing .
  • Pullout Testing .
  • Fatigue Testing o

Both the subject and predicate devices had suitable characteristics for their indication for use.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the word "EMBODY" in a stylized font. The letters are primarily dark blue, except for the "BOD" portion, where the "O" is replaced by an orange square with diagonal lines inside. The font is bold and modern, with rounded edges, giving the word a sleek and professional appearance. The overall design is clean and eye-catching.

A sixteen-week in vitro hydrolytic degradation study guided by ASTM F1635-16 "Standard Test Method for in vitro Degradation Testing of Hydrolytically Degradable Polymer Resins and Fabricated Forms for Surgical Implants" was also performed. This study provided characterization of the degradation profile via mass loss.

Absorption Profile

The Bioabsorbable Anchor is hydrolytically degraded in tissues and demonstrated expected local tissue response, histological observations, and degradation after 13 weeks in both intramuscular and femoral condyle implantation. The degradation products are subsequently absorbed by tissues and excreted from the body.

VIII. CONCLUSIONS

Results of the biocompatibility testing and non-clinical performance testing demonstrate that the Bioabsorbable Anchor is as safe and as effective, and substantially equivalent to the predicate device for its intended use.

§ 888.3030 Single/multiple component metallic bone fixation appliances and accessories.

(a)
Identification. Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.