K Number
K213889

Validate with FDA (Live)

Device Name
PicoStar
Date Cleared
2022-04-22

(130 days)

Product Code
Regulation Number
878.4810
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The PicoStar laser system is indicated for the following at the specified wavelength:

532 nm:

Removal of tattoos for Fitzpatrick skin types I-III to treat the following tattoo colors: red, yellow and orange. Indicated for benign pigmented lesions removal for Fitzpatrick skin types I-III.

1064 nm:

Removal of tattoos for all skin types (Fitzpatrick I-V) to treat the following tattoo colors: black, brown, green, blue and purple.

Indicated for benign pigmented lesions removal for Fitzpatrick skin types I-IV.

Only with microspot handpiece, indicated for treatment of wrinkles in Fitzpatrick Skin Types I-IV.

Only with microspot handpiece. indicated for the treatment of acne scars in Fitzpatrick Skin Types II-V.

Device Description

PicoStar is a solid state laser capable of delivering energy at wavelengths of 1064 nm, or 532 nm at short durations of max 400 picoseconds (ps) and repetition rates up to 10 Hz. The device system is composed of a system console, an articulated arm, and attached delivery handpieces (full spot or microspot). The laser output at each wavelength is delivered to the skin through an articulated arm delivery system terminated by a handpiece. The PicoStar is controlled via a touch screen display housed in the front of the device. The control panel enables the user to select the desired energy density level and repetition rate. Emission is triggered by means of a footswitch.

AI/ML Overview

The provided text is a 510(k) summary for the PicoStar laser system. It outlines the device's indications for use, technological characteristics, and a comparison to a predicate device to establish substantial equivalence. However, it does not contain information about acceptance criteria and a study proving a device meets those criteria for an AI/machine learning medical device.

The document states:

  • "Clinical Performance Data: None" on page 6.
  • "Software verification and validation testing was conducted and documentation is provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices". PicoStar passed all the required testing and is in compliance will all applicable sections of the abovementioned performance standards." on page 5. This refers to software engineering practices, not a performance study of an AI algorithm based on clinical data.

The PicoStar device is a laser surgical instrument, and its clearance is based on substantial equivalence to a predicate laser device, supported by non-clinical performance data (electrical safety, EMC, software lifecycle processes, risk management) and biocompatibility, not by a clinical study on AI performance metrics.

Therefore, I cannot extract the requested information regarding acceptance criteria and a study proving an AI device meets those criteria from this document. The document pertains to the regulatory clearance of a laser device, not an AI/ML-driven diagnostic or therapeutic device that would require such performance studies.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

April 22, 2022

Asclepion Laser Technologies GmbH Carolin Kuehling Head of Quality and Regulatory Affairs Bruesseler Strasse 10 Jena, Thuringia 07747 Germany

Re: K213889

Trade/Device Name: PicoStar Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: December 9, 2021 Received: December 13, 2021

Dear Carolin Kuehling:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Purva Pandya, D.Eng. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K213889

Device Name PicoStar

Indications for Use (Describe)

The PicoStar laser system is indicated for the following at the specified wavelength:

532 nm:

Removal of tattoos for Fitzpatrick skin types I-III to treat the following tattoo colors: red, yellow and orange. Indicated for benign pigmented lesions removal for Fitzpatrick skin types I-III.

1064 nm:

Removal of tattoos for all skin types (Fitzpatrick I-V) to treat the following tattoo colors: black, brown, green, blue and purple.

Indicated for benign pigmented lesions removal for Fitzpatrick skin types I-IV.

Only with microspot handpiece, indicated for treatment of wrinkles in Fitzpatrick Skin Types I-IV.

Only with microspot handpiece. indicated for the treatment of acne scars in Fitzpatrick Skin Types II-V.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Asclepion Laser Technologies. The word "Asclepion" is written in a large, gray, sans-serif font. Below it, the words "Laser Technologies" are written in a smaller, gray, sans-serif font. To the right of the word "Asclepion" is a cluster of seven dark blue circles of varying sizes.

510(k) SUMMARY

[As Required by 21 CFR 807.92]

Applicant:ASCLEPION LASER TECHNOLOGIES GmbHBruesseler Str. 1007747 Jena, Germany
510(k) Contact Person:Mrs. Carolin KuehlingHead of Quality and Regulatory AffairsPhone: +49 3641 77 00 213Fax:+49 3641 77 00 102E-mail: carolin.kuehling@asclepion.com
Preparation Date:December 8th, 2021
Common Name:Nd:YAG laser
Device Trade Name:PicoStar
Classification Name:Laser surgical instrument for use in general and plastic surgery and inDermatology
ClassificationClass II
Regulation Number21 CFR 878.4810
Product CodeGEX
Submission Type:Traditional 510(K)

Page 1 of 4

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Asclepion Laser Technologies. The logo features the word "Asclepion" in a gray, sans-serif font, with the words "Laser Technologies" in a smaller font below it. To the right of the word "Asclepion" is a cluster of blue circles of varying sizes, arranged in a semi-circular pattern. The logo is simple and modern, and the blue circles add a touch of visual interest.

Predicate Device Discovery Pico Family (K172376)

Device Description: PicoStar is a solid state laser capable of delivering energy at wavelengths of 1064 nm, or 532 nm at short durations of max 400 picoseconds (ps) and repetition rates up to 10 Hz. The device system is composed of a system console, an articulated arm, and attached delivery handpieces (full spot or microspot). The laser output at each wavelength is delivered to the skin through an articulated arm delivery system terminated by a handpiece. The PicoStar is controlled via a touch screen display housed in the front of the device. The control panel enables the user to select the desired energy density level and repetition rate. Emission is triggered by means of a footswitch.

Indications for Use:

The PicoStar laser system is indicated for the following at the specified wavelength:

  • 532 nm: Removal of tattoos for Fitzpatrick skin types I-III to treat the following tattoo colors: red, vellow and orange. Indicated for benign pigmented lesions removal for Fitzpatrick skin types I-III. 1064 nm: Removal of tattoos for all skin types (Fitzpatrick I-VI) to treat the following tattoo colors: black, brown, green, blue and purple. Indicated for benign pigmented lesions removal for Fitzpatrick skin types I-IV. Only with microspot handpiece, indicated for treatment of wrinkles in Fitzpatrick Skin Types I-IV. Only with microspot handpiece, indicated for the treatment of acne scars in Fitzpatrick Skin Types II-V.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Asclepion Laser Technologies. The word "Asclepion" is written in a gray, sans-serif font. Below it, the words "Laser Technologies" are written in a smaller, gray, sans-serif font. To the right of the word "Asclepion" is a cluster of blue circles of varying sizes, arranged in a pattern that suggests a burst or spray.

Summary of technological characteristics

Predicate DeviceSubject Device
Model NameDiscovery Pico FamilyPicoStar
ManufacturerQuanta System SPAAsclepion Laser Technologies GmbH
Laser TypeNd:YAGNd:YAG
Wavelength (nm)53210645321064
Delivery SystemArticulated armArticulated arm
Max. Pulse duration (ps)up to 400up to 450up to 400up to 400
Spot size (mm)up to 12 roundup to 9x9 squaredup to 5 roundup to 16 round7x7 squared
Max Energy (mJ)300800300800
Pulse Repetition rate (Hz)Max 10 HzMax 10 Hz
Microspot handpieceAvailableAvailable

Non-clinical Performance Data:

The following performance data were applied in support of the substantial equivalence determination:

  • . ISO 60601-1: Medical electrical equipment – Part 1: General requirements for safety and essential performance
  • . ISO 60601-1-2 Medical electrical equipment - Part 1-2: General requirements for safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests
  • IEC 62304 Medical Device Software Software life cycle processes ●
  • IEC 60601-2-22 Medical electrical equipment Part 2: Particular requirements for the safety of ● diagnostic and therapeutic laser equipment
  • . ISO 14971 Medical devices - Application of risk management to medical devices

Software verification and validation testing was conducted and documentation is provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices".

PicoStar passed all the required testing and is in compliance will all applicable sections of the abovementioned performance standards.

Biocompatibility:

The biocompatibility of PicoStar is established based on the predicate device.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Asclepion Laser Technologies. The logo features the word "Asclepion" in gray, with the words "Laser Technologies" in a smaller font size underneath. To the right of the text is a cluster of blue circles of varying sizes, arranged in a scattered pattern.

Clinical Performance Data: None

Comparison with predicate device

The PicoStar and the predicate device have the same intended use with similar indications for use.

The PicoStar Laser System presents the same or similar technological characteristics as its predicate devices, including the laser type, wavelengths, device design, pulse width, frequency, spot sizes and system components. Any minor differences do not present any new types of safety or effectiveness questions since the PicoStar parameters are similar to the ones of the predicates. Further, PicoStar performance has been demonstrated in non-clinical performance data, and results confirm the safety and performance of the device. The PicoStar device and its predicate operate with the same mechanism of action based on selective photothermolysis of pigment particles using laser energy. Therefore, the PicoStar has the same intended use and similar indications for use, technological characteristics, and principles of operation as the predicate devices. The PicoStar is substantially equivalent to the predicate device.

Summary

It is demonstrated that the Picostar performs as intended and is substantially equivalent to the predicate device.

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.