K Number
K213875

Validate with FDA (Live)

Date Cleared
2022-12-21

(373 days)

Product Code
Regulation Number
862.3910
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The DRI™ Tricyclics Serum Tox Assay is a homogeneous enzyme immunoassay intended for the qualitative and/or semiquantitative determination of the presence of tricyclic antidepressants (TCAs) in human serum, plasma, or urine of patients at a cutoff concentration of 300 ng/mL in patients suspected of drug overdose.

The semi-quantitative mode is for the purpose of enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as Liquid Chromatography/Tandem Mass Spectrometry (LC-MS/MS) or permitting laboratories to establish quality control procedures.

The assay provides only a preliminary analytical test result. A more specific alternative chemical must be used to obtain a confirmed analytical result. Liquid Chromatography/Tandem Mass Spectrometry (LC-MS/MS) is the preferred confirmatory method.

Clinical and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.

For In Vitro Diagnostic Use Only.

Device Description

The DRI™ Tricyclics Serum Tox Assay is a homogeneous enzyme immunoassay using ready to-use liquid reagents. Specific tricyclic antibodies were used to detect most tricyclic antidepressants in serum, plasma, or urine. The test is based on the competition of an enzyme, glucose-6-phosphate dehydrogenase (G6PDH), labeled-drug and the drug from the sample for a fixed amount of specific antibody binding sites. In the absence of the drug from the sample, the specific antibody binds the enzyme-labeled drug and the enzyme activity is inhibited. This phenomenon creates a direct relationship between drug concentration in the sample and the enzyme activity. The enzyme activity is determined spectrophotometrically at 340 nm by measuring its ability to convert nicotinamide adenine dinucleotide (NAD) to NADH.

The DRI™ Tricyclics Serum Tox assay is supplied as a two liquid reagent kit (Reagent A and Reagent E). Thev are bottled separately within the same kit, see details below:

  • Antibody/Substrate Reagent (Reagent A): Contains polyclonal anti-tricyclics . antibodies (sheep), glucose-6- phosphate (G6P), and nicotinamide adenine dinucleotide (NAD) in Tris buffer with sodium azide as a preservative.
  • Enzyme Conjugate Reagent (Reagent E): Contains glucose-6-phosphate ● dehydrogenase (G6PDH) labeled with nortriptyline in Tris buffer with sodium azide as a preservative.
AI/ML Overview

The provided document describes the analytical performance of the DRI™ Tricyclics Serum Tox Assay. It details various studies conducted to demonstrate its performance characteristics, but it does not explicitly state specific pass/fail acceptance criteria for each test. Instead, it presents the results of each study, implying that these results met the internal criteria used by the manufacturer for demonstrating sufficient performance and substantial equivalence to the predicate device.

Here's an analysis of the provided information, addressing your points where possible, and noting when the information is unavailable:

1. A table of acceptance criteria and the reported device performance

As mentioned, explicit acceptance criteria (e.g., "Accuracy must be >95%") are not stated in this document. The tables below summarize the reported device performance for several key analytical studies. The implication is that these reported performances were deemed acceptable for the 510(k) clearance.

Study TypeReported Device Performance (Serum)Reported Device Performance (Urine)
Precision (Repeatability)Qualitative Mode (n=80 for each concentration):- 0 ng/mL to 225 ng/mL (-100% to -25% Cutoff): 80/0 (Negative/Positive)- 300 ng/mL (100% Cutoff): 10/70 (Negative/Positive)- 375 ng/mL to 600 ng/mL (+25% to +100% Cutoff): 0/80 (Negative/Positive)Semi-Quantitative Mode (n=80 for each concentration):- 0 ng/mL to 225 ng/mL (-100% to -25% Cutoff): 80/0 (Negative/Positive)- 300 ng/mL (100% Cutoff): 19/61 (Negative/Positive)- 375 ng/mL to 600 ng/mL (+25% to +100% Cutoff): 0/80 (Negative/Positive)Qualitative Mode (n=80 for each concentration):- 0 ng/mL to 225 ng/mL (-100% to -25% Cutoff): 80/0 (Negative/Positive)- 300 ng/mL (100% Cutoff): 68/12 (Negative/Positive)- 375 ng/mL to 600 ng/mL (+25% to +100% Cutoff): 0/80 (Negative/Positive)Semi-Quantitative Mode (n=80 for each concentration):- 0 ng/mL to 225 ng/mL (-100% to -25% Cutoff): 80/0 (Negative/Positive)- 300 ng/mL (100% Cutoff): 73/7 (Negative/Positive)- 375 ng/mL to 600 ng/mL (+25% to +100% Cutoff): 0/80 (Negative/Positive)
Precision (Reproducibility)Qualitative Mode (n=75 for each concentration):- 0 ng/mL to 225 ng/mL (-100% to -25% Cutoff): 75/0 (Negative/Positive)- 300 ng/mL (100% Cutoff): 30/45 (Negative/Positive)- 375 ng/mL to 600 ng/mL (+25% to +100% Cutoff): 0/75 (Negative/Positive)Semi-Quantitative Mode (n=75 for each concentration):- 0 ng/mL to 225 ng/mL (-100% to -25% Cutoff): 75/0 (Negative/Positive)- 300 ng/mL (100% Cutoff): 29/46 (Negative/Positive)- 375 ng/mL to 600 ng/mL (+25% to +100% Cutoff): 0/75 (Negative/Positive)Qualitative Mode (n=75 for each concentration):- 0 ng/mL to 225 ng/mL (-100% to -25% Cutoff): 75/0 (Negative/Positive)- 300 ng/mL (100% Cutoff): 61/14 (Negative/Positive)- 375 ng/mL to 600 ng/mL (+25% to +100% Cutoff): 0/75 (Negative/Positive)Semi-Quantitative Mode (n=75 for each concentration):- 0 ng/mL to 225 ng/mL (-100% to -25% Cutoff): 75/0 (Negative/Positive)- 300 ng/mL (100% Cutoff): 66/9 (Negative/Positive)- 375 ng/mL to 600 ng/mL (+25% to +100% Cutoff): 0/75 (Negative/Positive)
Spike RecoveryLow Control (225 ng/mL, n=20): All 20 Negative (Below Cutoff)High Control (375 ng/mL, n=20): All 20 Positive (Above Cutoff)Low Control (225 ng/mL, n=20): All 20 Negative (Below Cutoff)High Control (375 ng/mL, n=20): All 20 Positive (Above Cutoff)
Dilution Linearity (Recovery)Range 96-115% for expected values 125 ng/mL to 1000 ng/mL (average of 5 replicates)Range 94-118% for expected values 125 ng/mL to 1000 ng/mL (average of 5 replicates)
Method Comparison (Accuracy)Qualitative Mode (n=117):- % Negative Sample Agreement: 98% (60/61)- % Positive Sample Agreement: 100% (56/56)- % Total Sample Agreement: 99% (116/117)Semi-Quantitative Mode (n=117):- % Negative Sample Agreement: 97% (59/61)- % Positive Sample Agreement: 100% (56/56)- % Total Sample Agreement: 98% (115/117)Qualitative Mode (n=100):- % Negative Sample Agreement: 96% (48/50)- % Positive Sample Agreement: 98% (49/50)- % Total Sample Agreement: 97% (97/100)Semi-Quantitative Mode (n=100):- % Negative Sample Agreement: 96% (48/50)- % Positive Sample Agreement: 98% (49/50)- % Total Sample Agreement: 97% (97/100)
Matrix EquivalencySerum: 50 patient samples total (25 positive, 25 negative). Full agreement for K2 EDTA Plasma, K3 EDTA Plasma, Lithium Heparin Plasma, Sodium Citrate Plasma, Potassium Oxalate Plasma across qualitative and semi-quantitative modes.Sodium Heparin Plasma: 45 patient samples (25 positive, 20 negative). Full agreement across qualitative and semi-quantitative modes.Not applicable (serum comparison)
InterferenceSpiking endogenous/exogenous/physiological substances at high concentrations into low and high controls (225 ng/mL and 375 ng/mL) showed accurate detection of controls (negative for low, positive for high), indicating no interference. (See Table 20 for list of compounds and concentrations tested).Spiking endogenous/exogenous/physiological substances and pH variations (pH 3-11) at high concentrations into low and high controls (225 ng/mL and 375 ng/mL) showed accurate detection of controls (negative for low, positive for high), indicating no interference. (See Table 21 for list of compounds and concentrations tested). Specific gravity (1.004-1.030) also showed no interference.
Specificity (Cross-Reactivity)Extensive tables (Table 16, 18, 20) are provided listing cross-reactivity percentages for numerous structurally related and unrelated compounds in serum. The study demonstrated cross-reactivity for various TCA drugs and metabolites (e.g., Amitriptyline 100%, Desipramine 120%, Imipramine 157.9%). Minimal to no cross-reactivity (<0.3% to low single digits) was observed for most structurally unrelated compounds at high concentrations, except for a few exceptions like Chlorpromazine (57.1%), Cyclobenzaprine (66.7%), and Perphenazine (66.7%).Extensive tables (Table 17, 19, 21) are provided listing cross-reactivity percentages for numerous structurally related and unrelated compounds in urine. Similar to serum, cross-reactivity was observed for various TCA drugs and metabolites (e.g., Amitriptyline 100%, Desipramine 120%, Imipramine 136.4%). Minimal to no cross-reactivity (<0.3% to low single digits) was observed for most structurally unrelated compounds at high concentrations, except for a few exceptions like Chlorpromazine (50%), Cyclobenzaprine (60%), and Perphenazine (46.2%).

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Precision (Repeatability): 80 replicates per concentration level for serum and urine.
  • Precision (Reproducibility): 75 replicates per concentration level for serum and urine (across 3 different instruments, 2 reagent lots).
  • Spike Recovery: 20 replicates for low and high controls for both serum and urine.
  • Dilution Linearity: 9 levels, each with 5 replicates, for both serum and urine.
  • Method Comparison and Accuracy:
    • Serum: 61 negative patient samples + 56 positive patient samples = 117 total patient samples.
    • Urine: 50 negative patient samples + 50 positive patient samples = 100 total patient samples.
  • Matrix Equivalency: 50 patient samples for each of 5 plasma types (K2 EDTA, K3 EDTA, Lithium Heparin, Sodium Citrate, Potassium Oxalate), and 45 patient samples for Sodium Heparin plasma. Each run with 2 replicates.
  • Specificity (Cross-Reactivity) & Interference: Tested with known spiked concentrations in drug-free serum/urine and low/high controls. The exact number of individual samples/replicates for each substance is not explicitly stated beyond what is in the tables (e.g., "known amount," "single determinations" implied for cross-reactivity at a given concentration).
  • Data Provenance: The document does not specify the country of origin of the data or whether the patient samples were collected retrospectively or prospectively.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This device is an in vitro diagnostic (IVD) assay for detecting tricyclic antidepressants. The ground truth for the test set (method comparison study) was established by comparison to Liquid Chromatography/Tandem Mass Spectrometry (LC-MS/MS) lab testing. This is a highly accurate analytical chemical method, considered the "preferred confirmatory method" by the device's indications for use. Therefore, human experts were not used to establish the primary ground truth in the way they might be for image-based diagnostics. The LC-MS/MS results serve as the objective reference standard.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable, as the ground truth was based on a reference chemical analytical method (LC-MS/MS), not on human expert adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is an IVD assay, not an AI-based diagnostic tool requiring human interpretation or a multi-reader study.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Yes, the studies presented are standalone performance studies of the device (the assay) itself. There is no human-in-the-loop component for the analytical performance evaluations described. The device provides a preliminary analytical test result independently.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The ground truth for all analytical performance studies (especially method comparison) was Liquid Chromatography/Tandem Mass Spectrometry (LC-MS/MS), which is a gold-standard chemical analytical method.

8. The sample size for the training set

This document describes the analytical validation of an enzyme immunoassay kit, not a machine learning algorithm. Therefore, there is no "training set" in the context of AI/ML. The samples used for various analytical performance studies (precision, spike recovery, linearity, method comparison, etc.) are implicitly part of the validation process, not for "training" an algorithm.

9. How the ground truth for the training set was established

As there is no training set for a machine learning algorithm, this question is not applicable. The assay's performance is based on its biochemical interaction, not on learned patterns from a training dataset.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA text logo on the right. The FDA text logo is in blue and includes the acronym "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a stacked format.

December 21, 2022

Microgenics Corporation Pranjali Shinde Senior Regulatory Affairs Specialist 46500 Kato Road Fremont, CA 94538

Re: K213875

Trade/Device Name: DRI™ Tricyclics Serum Tox Assay Regulation Number: 21 CFR 862.3910 Regulation Name: Tricyclic antidepressant drugs test system Regulatory Class: Class II Product Code: LFH Dated: August 19, 2022 Received: August 23, 2022

Dear Pranjali Shinde:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR

{1}------------------------------------------------

  1. for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Paula --Digitally signed by Caposino - Date: 2022.12.21 Paula Caposino -S 11:51:25 -05'00'

Paula Caposino, Ph.D. Acting Deputy Division Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K213875

Device Name DRI™ Tricyclics Serum Tox Assay

Indications for Use (Describe)

The DRI™ Tricyclics Serum Tox Assay is a homogeneous enzyme immunoassay intended for the qualitative and/or semiquantitative determination of the presence of tricyclic antidepressants (TCAs) in human serum, plasma, or urine of patients at a cutoff concentration of 300 ng/mL in patients suspected of drug overdose.

The semi-quantitative mode is for the purpose of enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as Liquid Chromatography/Tandem Mass Spectrometry (LC-MS/MS) or permitting laboratories to establish quality control procedures.

The assay provides only a preliminary analytical test result. A more specific alternative chemical must be used to obtain a confirmed analytical result. Liquid Chromatography/Tandem Mass Spectrometry (LC-MS/MS) is the preferred confirmatory method.

Clinical and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.

For In Vitro Diagnostic Use Only.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

510(k) Number: K213875

This 510(k) Summary of Safety and Effectiveness is being submitted in accordance with the requirements of Safe Medical Device Act of 1990 and 21 CFR 807.92.

A. Device Information

Sponsor:Microgenics CorporationThermo Fisher Scientific46500 Kato RoadFremont, CA 94538Phone: 510-979-5000FAX: 510-979-5002
Correspondent ContactInformation:Pranjali ShindeMicrogenics CorporationThermo Fisher Scientific46500 Kato RoadFremont, CA-94538Email: Pranjali.Shinde@Thermofisher.comPhone: 213-344-9952Fax: 510-979-5002
Device Common Name:Tricyclics Serum Tox Assay
Trade or Proprietary NameDRI™ Tricyclics Serum Tox Assay
Brand NameDRI™ Tricyclics Serum Tox Assay
Candidate Device ProductCode, Classification,Classification Name & PanelLFH, Class II,21 CFR 862. 3910 – Tricyclic antidepressant drugs testsystem, 91 - Toxicology

Predicate Device Information

Predicate Device:Tricyclic Serum Tox EIA Assay
Predicate Device Manufacturer:Microgenics Corporation
Predicate Device Common NameTricyclic Serum Tox Assay
Predicate Device Premarket Notification #:K983268
Predicate Device Product Code, Classification, Classification Name & PanelLFH, Class II,21 CFR 862. 3910 – Tricyclic antidepressant drugs testsystem, 91 - Toxicology

B. Date Summary Prepared

November 14, 2022

{4}------------------------------------------------

C. Description of Device

The DRI™ Tricyclics Serum Tox Assay is a homogeneous enzyme immunoassay using ready to-use liquid reagents. Specific tricyclic antibodies were used to detect most tricyclic antidepressants in serum, plasma, or urine. The test is based on the competition of an enzyme, glucose-6-phosphate dehydrogenase (G6PDH), labeled-drug and the drug from the sample for a fixed amount of specific antibody binding sites. In the absence of the drug from the sample, the specific antibody binds the enzyme-labeled drug and the enzyme activity is inhibited. This phenomenon creates a direct relationship between drug concentration in the sample and the enzyme activity. The enzyme activity is determined spectrophotometrically at 340 nm by measuring its ability to convert nicotinamide adenine dinucleotide (NAD) to NADH.

The DRI™ Tricyclics Serum Tox assay is supplied as a two liquid reagent kit (Reagent A and Reagent E). Thev are bottled separately within the same kit, see details below:

  • Antibody/Substrate Reagent (Reagent A): Contains polyclonal anti-tricyclics . antibodies (sheep), glucose-6- phosphate (G6P), and nicotinamide adenine dinucleotide (NAD) in Tris buffer with sodium azide as a preservative.
  • Enzyme Conjugate Reagent (Reagent E): Contains glucose-6-phosphate ● dehydrogenase (G6PDH) labeled with nortriptyline in Tris buffer with sodium azide as a preservative.

CI. Intended Use

The DRI™ Tricyclics Serum Tox Assay is a homogeneous enzyme immunoassay intended for the qualitative and/or semiquantitative determination of the presence of tricyclic antidepressants (TCAs) in human serum, plasma, or urine of patients at a cutoff concentration of 300 ng/mL in patients suspected of drug overdose.

The semi-quantitative mode is for the purpose of enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as Liquid Chromatography/Tandem Mass Spectrometry (LC-MS/MS) or permitting laboratories to establish quality control procedures.

The assay provides only a preliminary analytical test result. A more specific alternative chemical method must be used to obtain a confirmed analytical result. Liquid Chromatography/Tandem Mass Spectrometry (LC-MS/MS) is the preferred confirmatory method.

Clinical and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.

For In Vitro Diagnostic Use Only.

{5}------------------------------------------------

E. Comparison to Predicate Device

CharacteristicsCandidate Device:DRI™ Tricyclics Serum Tox Assay(K213875)Predicate Device:Tricyclic Serum Tox EIA Assay(K983268)
Intended UseThe DRI™ Tricyclics Serum ToxAssay is a homogeneous enzymeimmunoassay intended for thequalitative and/orsemiquantitative determination ofthe presence of tricyclicantidepressants (TCAs) in humanserum, plasma, or urine of patientsat a cutoff concentration of 300ng/mL in patients suspected ofdrug overdose.The semi-quantitative mode is forthe purpose of enablinglaboratories to determine anappropriate dilution of thespecimen for confirmation by aconfirmatory method such asLiquid Chromatography/TandemMass Spectrometry (LC-MS/MS)or permitting laboratories toestablish quality controlprocedures.The assay provides only apreliminary analytical test result.A more specific alternativechemical method must be used toobtain a confirmed analyticalresult. LiquidChromatography/Tandem MassSpectrometry (LC-MS/MS) is thepreferred confirmatory method.Clinical and professionaljudgment should be applied to anydrug of abuse test resultSame
particularly when preliminarypositive results are used.
For In Vitro Diagnostic Use Only.
OperatingPrinciple(Technology)DRISame
Measured AnalyteNortriptylineSame
Test MatrixSerum, Plasma, UrineSame
Cut-off Levels300 ng/mLSame
MethodologyHomogeneous EnzymeImmunoassaySame
Reagents FormLiquid ready-to-use.Same
AntibodySheep polyclonal antibodiesMonoclonal antibodies
Storage2-8 °C until expiration date.Same
Principal OperatorTrained professionalsSame
Calibrator Levelsfor Semi-Quant5-point Calibrator5-point Calibrator

{6}------------------------------------------------

F. Test Principle

The DRI™ Tricyclics Serum Tox Assay is a homogeneous enzyme immunoassay using ready-to use liquid reagents. The assay uses specific tricyclic antibodies, which can detect most tricyclic antidepressants in serum, plasma, or urine. The DRI technology is based on the competition of an enzyme glucose-6-phosphate dehydrogenase (G6PDH) labeled-drug and the drug from the serum, plasma, or urine sample for a fixed amount of specific antibody binding sites. In the presence of free drug from the sample, the free drug occupies the antibody binding sites, allowing the drug-labeled G6PDH to interact with the substrate, resulting in enzyme activity. In the absence of drug from the sample, the specific antibody binds to the drug labeled with G6PDH and the enzyme activity is inhibited. This phenomenon creates a direct relationship between the drug concentration in the serum, plasma, or urine and the enzyme activity. The enzyme G6PDH activity is determined spectrophotometrically at 340 nm by measuring its ability to convert nicotinamide adenine dinucleotide (NAD) to NADH.

{7}------------------------------------------------

G. Summary of Supporting Data

Analytical Performance:

The performance was evaluated on Architect c4000 clinical chemistry analyzer.

  • a) Precision
    Precision studies were performed in accordance with CLSI Guideline EP05-A3.

Samples were prepared by spiking Nortriptyline in both drug-free serum and drug-free urine at the cutoff, 25%, 50%, 75%, and 100% above and below the cutoff.

Repeatability:

The study was conducted for 20 days with 2 runs per day, 2 replicates per run in both qualitative and semi-quantitative modes with 1 lot of reagent, calibrators, and controls. There were 80 replicates for each level of precision samples. Serum and urine samples were run separately. The results of the precision study (Repeatability) are presented in the table below.

Table 1: Precision study data (Repeatability) in qualitative and semi-quantitative mode for 300 ng/mL cutoff - Serum

SpikedConcentration(ng/mL)% ofCutoff# ofDeterminantsTotal Precision (n=80)
QualitativeImmunoassay Results(Negative/Positive)Semi-QuantitativeImmunoassay Results(Negative/Positive)
0-1008080/080/0
75-758080/080/0
150-508080/080/0
225-258080/080/0
3001008010/7019/61
375+25800/800/80
450+50800/800/80
525+75800/800/80
600+100800/800/80

{8}------------------------------------------------

SpikedConcentration(ng/mL)% ofCutoff# ofDeterminantsTotal Precision (n=80)
QualitativeImmunoassay Results(Negative/Positive)Semi-QuantitativeImmunoassay Results(Negative/Positive)
0-1008080/080/0
75-758080/080/0
150-508080/080/0
225-258080/080/0
3001008068/1273/7
375+25800/800/80
450+50800/800/80
525+75800/800/80
600+100800/800/80

Table 2: Precision study data (Repeatability) in qualitative and semi-quantitative mode for 300 ng/mL cutoff = Urine

Reproducibility:

The study was conducted for 5 days with 1 run per day, 5 replicates per run in both qualitative and semi-quantitative modes with 2 lots of reagents on 3 different instruments. Serum and urine samples were run separately. The results of the precision study (Reproducibility) are presented in the table below.

Table 3: Precision study data (Reproducibility) in qualitative and semi-quantitative mode for 300 ng/mL cutoff - Serum

SpikedConcentration(ng/mL)% ofCutoff# ofDeterminantsTotal Precision (n=75)
QualitativeImmunoassay Results(Negative/Positive)Semi-quantitativeImmunoassay Results(Negative/Positive)
0-1007575/075/0
75-757575/075/0
150-507575/075/0
225-257575/075/0
3001007530/4529/46
375+25750/750/75
450+50750/750/75
525+75750/750/75
600+100750/750/75

{9}------------------------------------------------

Table 4: Precision study data (Reproducibility) in qualitative and semi-quantitative mode for 300 ng/mL cutoff - Urine

SpikedConcentration(ng/mL)% ofCutoff# ofDeterminantsTotal Precision (n=75)
QualitativeImmunoassay Results(Negative/Positive)Semi-quantitativeImmunoassay Results(Negative/Positive)
0-1007575/075/0
75-757575/075/0
150-507575/075/0
225-257575/075/0
3001007561/1466/9
375+25750/750/75
450+50750/750/75
525+75750/750/75
600+100750/750/75

{10}------------------------------------------------

  • b) Spike Recovery
    The spike recovery study was performed using 20 replicates. Samples were prepared by spiking Nortriptyline in both drug-free serum and drug-free urine at 225 ng/mL, 300 ng/mL, 375 ng/mL. The study demonstrated the accuracy of spike recovery at low control (225 ng/mL) and high control (375 ng/mL) in both qualitative and semiquantitative mode. Serum and urine samples were run separately. The results of the study are presented in the table below.
Table 5: Spike Recovery in qualitative and semi-quantitative mode for 300 ng/mL cutoff -
Serum
ReplicatesLow Control: 225 ng/mL(-25% of cutoff) (n=20)High Control: 375 ng/mL(+25% of cutoff)(n=20)
1NegativePositive
2NegativePositive
3NegativePositive
4NegativePositive
5NegativePositive
6NegativePositive
7NegativePositive
8NegativePositive
9NegativePositive
10NegativePositive
11NegativePositive
12NegativePositive
13NegativePositive
14NegativePositive
15NegativePositive
16NegativePositive
17NegativePositive
18NegativePositive
19NegativePositive
20NegativePositive
OverlapNoNo
Relative to C/OAll 20 below C/OAll 20 above C/O

Table 6: Recovery in qualitative and semi-quantitative mode for 300 ng/mL cutoff -Urine

ReplicatesLow Control: 225 ng/mL-25% of cutoff (n=20)High Control: 375 ng/mL+25% of cutoff (n=20)
1NegativePositive
2NegativePositive
3NegativePositive
4NegativePositive

{11}------------------------------------------------

ReplicatesLow Control: 225 ng/mL-25% of cutoff (n=20)High Control: 375 ng/mL+25% of cutoff (n=20)
5NegativePositive
6NegativePositive
7NegativePositive
8NegativePositive
9NegativePositive
10NegativePositive
11NegativePositive
12NegativePositive
13NegativePositive
14NegativePositive
15NegativePositive
16NegativePositive
17NegativePositive
18NegativePositive
19NegativePositive
20NegativePositive
OverlapNoNo
Relative to C/OAll 20 below C/OAll 20 above C/O

{12}------------------------------------------------

c) Dilution Linearity

The Dilution Linearity study is performed using the CLSI guideline CLSI EP06-A.

To demonstrate the dilution linearity for purposes of sample dilution and quality control of the entire assay range, drug-free serum or drug-free urine were spiked using Nortriptyline and diluted with drug-free serum or drug-free urine, respectively to generate 9 levels between 0 and 1000 ng/mL. Each sample was run in replicates of 5 in semi-quantitative mode and the average was used to determine percent recovery compared to the expected target value. Serum and urine samples were run separately. The study results are presented in the table below.

Level#Expected NortriptylineValues (ng/mL)Serum
Observed Values (ng/mL)Recovery (%)
107N/A
212512197
325024297
4375385103
550047996
6625720115
7750839112
8875913104
910001046105

Table 7: Dilution linearity data - Serum

Level#Expected NortriptylineValues (ng/mL)Urine
Observed Values (ng/mL)Recovery (%)
108N/A
2125148118
3250271108
4375393105
550047094
662562099
775074199
887583896
9100094695

Table 8: Dilution linearity data – Urine

{13}------------------------------------------------

  • d) Method Comparison and Accuracy
    The method comparison study was performed in accordance with CLSI guideline CLSI EP09c-A3.

One (1) replicate of each of the 61 negative patient samples for serum and 50 negative samples for urine and 56 positive patient samples for serum and 50 positive samples for urine were analyzed in both qualitative and semi-quantitative modes. Serum and urine samples were run separately. The results were compared to LC-MS/MS lab testing.

Serum and urine samples were run separately. The results of the study are presented in the tables below.

Table 9: Stratified data comparing immunoassay in qualitative mode with LC-MS/MS –
Serum
Near Cutoff
DRITCA STAssayResultsNegativeby LC-MS/MS< 50% ofCutoffconcentrationby LC-MS/MS (<150 ng/mL)Near CutoffNegative(Between 50%below the cutoffand the cutoffconcentration asdetermined byLC-MS/MS)(150-299 ng/mL)Near CutoffPositive(Between thecutoff and 50%above the cutoffconcentration asdetermined byLC-MS/MS)(300-450ng/mL)HighPositives(Greater than50% abovecutoffconcentration(> 450ng/mL)
Positive01103125
Negative1352400
% Negative Sample Agreement98% or (60/61)
% Positive Sample Agreement100% or (56/56)
% Total Sample Agreement99% or (116/117)

1 Refer to Table 11 for discordant samples of serum.

Table 10: Stratified data comparing immunoassay in semi-quantitative mode with LC MS/MS -

Serum
DRITCA STAssayResultsNegativeby LC-MS/MS< 50% ofCutoffconcentrationby LC-MS/MS (<150 ng/mL)Near CutoffNegative(Between 50%below the cutoffand the cutoffconcentration asdetermined byLC-MS/MS)(150-299ng/mL)Near CutoffPositive(Between thecutoff and 50%above the cutoffconcentration asdetermined byLC-MS/MS)(300-450ng/mL)HighPositives(Greater than50% abovecutoffconcentration(> 450ng/mL)
Positive02103125
Negative1342400
% Negative Sample Agreement97% or (59/61)

{14}------------------------------------------------

% Positive Sample Agreement100% or (56/56)
% Total Sample Agreement98% or (115/117)

1 Refer to Table 11 for discordant samples of serum.

Sample IDQualitative(mA/min)ResultSemi-QuantitativeObserved concentration(ng/mL)Final LC-MS/MSValue(ng/mL)
APP9489-22Negative354 (Positive)110
APP9474-23Positive418 (Positive)120
1Table 11: Discordant Samples - Serum
---------------------------------------

2 Sample APP9489-2 contains Amitriptyline at 8.08 ng/mL, Imipramine at 18.35 ng/mL, Desipramine at 46.35 ng/mL, and Nortriptyline at 17 ng/mL by LC-MS/MS and cross-reacts at 100%, 158%, 120%, and 100% by immunoassay, respectively.

3 Sample APP9474-2 contains Amitriptyline at 9.08 ng/mL, Imipramine at 19.9 ng/mL, Desipramine at 49.8 ng/mL, and Nortriptyline at 19.65 ng/mL by LC-MS/MS and cross-reacts at 100%, 158%, 120%, and 100% by immunoassay, respectively.

Note: In addition, both samples APP9489-2 and APP9474-2 were confirmed by LC-MS/MS to contain Carbamazepine and Carbamazepine Epoxide with concentrations (4275 ng/mL & 5285 ng/mL) and (595 ng/mL & 955 ng/mL), respectively.

Urine
DRITCA STAssayResultsNegativeby LC-MS/MS< 50% ofCutoffconcentrationby LC-MS/MS (<150 ng/mL)Near CutoffNegative(Between 50%below the cutoffand the cutoffconcentration asdetermined byLC-MS/MS)(150-299 ng/mL)Near CutoffPositive(Between thecutoff and 50%above the cutoffconcentration asdetermined byLC-MS/MS)(300-450ng/mL)HighPositives(Greater than50% abovecutoffconcentration(> 450ng/mL)
Positive0024742
Negative221115140
% Negative Sample Agreement96% or (48/50)
% Positive Sample Agreement98% or (49/50)
% Total Sample Agreement97% or (97/100)

Table 12: Stratified data comparing immunoassay in qualitative mode with LC-MS/MS -I Irine

4 Refer to Table 14 for discordant samples of urine.

{15}------------------------------------------------

DRITCA STAssayResultsNegativeby LC-MS/MS< 50% ofCutoffconcentrationby LC-MS/MS (<150 ng/mL)Near CutoffNegative(Between 50%below the cutoffand the cutoffconcentration asdetermined byLC-MS/MS) (150-299 ng/mL)Near CutoffPositive(Between thecutoff and 50%above the cutoffconcentration asdetermined byLC-MS/MS)(300-450ng/mL)HighPositives(Greater than50% abovecutoffconcentration(> 450ng/mL)
Positive0024742
Negative221115140
% Negative Sample Agreement96% or (48/50)
% Positive Sample Agreement98% or (49/50)
% Total Sample Agreement97% or (97/100)

Table 13: Stratified data comparing immunoassay in semi-quantitative mode with LC-MS/MS - Urine

4 Refer to Table 14 for discordant samples of urine.

4Table 14: Discordant Samples - Urine

Sample IDQualitative (mA/min)ResultSemi-QuantitativeObserved concentration(ng/mL)LC-MS/MS Value(ng/mL)
APP6058-25Positive347 (Positive)285
APP6060-26Positive375 (Positive)230
APP6056-27Negative267 (Negative)372

5 Sample APP6058-2 contains Amitriptyline at 146 ng/mL and Nortriptyline at 139 ng/mL by LC-MS/MS and cross-reacts at 100% and 100% by immunoassay, respectively.

6 Sample APP6060-2 contains Amitriptyline at 114 ng/mL, Nordoxepin at 3.05 ng/mL, and Nortriptyline at 115 ng/mL by LC-MS/MS and cross-reacts at 100%, 17.1% and 100% by immunoassay, respectively.

7 Sample APP6056-2 contains Amitriptyline at 186 ng/ mL and Nortriptyline at 186 ng/mL by LC-MS/MS and cross-reacts at 100% and 100% by immunoassay, respectively.

{16}------------------------------------------------

  • e) Matrix equivalency
    Matrix Equivalency is performed in accordance with CLSI Guideline CLSI-EP35.

Two (2) replicates of 50 patient samples of K2 EDTA Plasma, K3 EDTA plasma, Lithium Heparin Plasma, Sodium Citrate Plasma, Potassium Oxalate Plasma and 45 patient samples of Sodium Heparin Plasma were run in both qualitative and semiquantitative modes to demonstrate Nortriptyline concentrations obtained in different test plasma matrices with different anticoagulants are equivalent to those measured in the primary or control matrix, serum. The results of the matrix equivalency study are presented in the table below.

QualitativeSemi-quantitative
Serum MatrixPositiveNegativePositiveNegative
K2 EDTA PlasmaPositive250250
Negative025025
K3 EDTA plasmaPositive250250
Negative025025
Lithium Heparin PlasmaPositive250250
Negative025025
Sodium Citrate PlasmaPositive250250
Negative025025
Potassium Oxalate PlasmaPositive250250
Negative025025
Sodium Heparin PlasmaPositive250250
Negative020020

{17}------------------------------------------------

  • f) Specificity

Cross-Reactivity to structurally related compounds:

The cross-reactivity of Tricyclic compounds and other structurally related compounds were evaluated by adding known amount of each compound into drug-free serum and drug-free urine at concentrations indicated. Serum and urine samples were run separately.

Serum
Structurally Related Compounds(TCA drugs and Metabolites)TestedConcentrations(ng/mL)Positive/NegativeCross-Reactivity (%)
2-Hydroxy Imipramine1,300Positive23.1
7-Hydroxy Quetiapine100,000Negative< 0.3
7-Hydroxy Amoxapine100,000Negative< 0.3
8-Hydroxy Amoxapine100,000Negative< 0.3
Amitriptyline300Positive100
Amoxapine125,000Positive0.2
10-Hydroxyamitriptyline1,000Positive30
Clomipramine300Positive100
Desipramine250Positive120
Dosulepin475Positive63.2
Doxepin600Positive50
Imipramine190Positive157.9
Lofepramine430Positive69.8
N-Desmethyltrimipramine350Positive85.7
Norclomipramine375Positive80
Nordoxepin1,750Positive17.1
Nortriptyline300Positive100
Opipramol350Positive85.7
Protriptyline450Positive66.7
Quetiapine Fumarate50,000Positive0.6
Trimipramine390Positive76.9

Table 16: Cross Reactivity to structurally related compounds in Serum (TCA drugs and Metabolites)

Table 17: Cross Reactivity to structurally related compounds in Urine (TCA drugs and Metabolites)

Urine
Structurally Related Compounds(TCA drugs and Metabolites)TestedConcentrations(ng/mL)Positive/NegativeCross-Reactivity (%)
2-Hydroxy Imipramine1,000Positive30
7-Hydroxy Quetiapine100,000Negative< 0.3
7-Hydroxy Amoxapine100,000Negative< 0.3
8-Hydroxy Amoxapine100,000Negative< 0.3
Amitriptyline300Positive100
Amoxapine100,000Positive0.3

{18}------------------------------------------------

Structurally Related Compounds(TCA drugs and Metabolites)Urine
TestedConcentrations(ng/mL)Positive/NegativeCross-Reactivity (%)
10-Hydroxyamitriptyline700Positive42.9
Clomipramine350Positive85.7
Desipramine250Positive120
Dosulepin425Positive70.6
Doxepin550Positive54.5
Imipramine220Positive136.4
Lofepramine460Positive65.2
N-Desmethyltrimipramine325Positive92.3
Norclomipramine450Positive66.7
Nordoxepin1750Positive17.1
Nortriptyline300Positive100
Opipramol350Positive85.7
Protriptyline400Positive75
Quetiapine Fumarate45,000Positive0.7
Trimipramine400Positive75

Table 18: Cross Reactivity to structurally related compounds in Serum (Other compounds)

Serum
Other Structurally RelatedCompoundsTestedConcentrations(ng/mL)Positive/NegativeCross-Reactivity (%)
Alimemazine12,000Positive2.5
Blonanserin100,000Negative< 0.3
Chlorpromazine525Positive57.1
Clozapine100,000Negative< 0.3
Cyclobenzaprine450Positive66.7
Desloratadine100,000Negative< 0.3
Diphenhydramine60,000Positive0.5
Fluoxetine100,000Negative< 0.3
Fluphenazine2,000Positive15
Haloperidol100,000Negative< 0.3
Loratadine100,000Negative< 0.3
Loxapine100,000Negative< 0.3
Maprotiline100,000Positive0.3
Mianserin100,000Negative< 0.3
Mirtazapine100,000Negative< 0.3
N-Desmethylclozapine100,000Negative< 0.3
Nefazodone100,000Negative< 0.3
N-Desmethylmirtazapine100,000Negative< 0.3
Normaprotiline100,000Positive0.3
Olanzapine100,000Negative< 0.3
Paroxetine100,000Negative< 0.3

{19}------------------------------------------------

Perphenazine450Positive66.7
Phenoxybenzamine100,000Negative< 0.3
Promazine400Positive75
Promethazine20,000Positive1.5
Risperidone100,000Negative< 0.3
Sertraline100,000Negative< 0.3
Thioridazine6,000Positive5
Thiothixene100,000Negative< 0.3
Tianeptine100,000Negative< 0.3
Trazodone100,000Negative< 0.3
Ziprasidone100,000Negative< 0.3

Table 19: Cross Reactivity to structurally related compounds in Urine

(Other compounds)

Urine
Other Structurally RelatedCompoundsTestedConcentrations(ng/mL)Positive/NegativeCross-Reactivity (%)
Alimemazine12,000Positive2.5
Blonanserin100,000Negative< 0.3
Chlorpromazine600Positive50
Clozapine100,000Negative< 0.3
Cyclobenzaprine500Positive60
Desloratadine100,000Negative< 0.3
Diphenhydramine30,000Positive1
Fluoxetine100,000Negative< 0.3
Fluphenazine2,000Positive15
Haloperidol100,000Negative< 0.3
Loratadine100,000Negative< 0.3
Loxapine100,000Positive0.3
Maprotiline100,000Positive0.3
Mianserin100,000Negative< 0.3
Mirtazapine100,000Negative< 0.3
N-Desmethylclozapine100,000Negative< 0.3
Nefazodone100,000Negative< 0.3
N-Desmethylmirtazapine100,000Negative< 0.3
Normaprotiline100,000Positive0.3
Olanzapine100,000Negative< 0.3
Paroxetine100,000Negative< 0.3
Perphenazine650Positive46.2
Phenoxybenzamine100,000Negative< 0.3
Promazine410Positive73.2
Promethazine15,000Positive2
Risperidone100,000Negative< 0.3
Sertraline100,000Negative< 0.3

{20}------------------------------------------------

Urine
Other Structurally RelatedCompoundsTestedConcentrations(ng/mL)Positive/NegativeCross-Reactivity (%)
Thioridazine4,000Positive7.5
Thiothixene100,000Negative< 0.3
Tianeptine90,000Positive0.33
Trazodone100,000Negative< 0.3
Ziprasidone100,000Negative< 0.3

Cross-reactivity to structurally unrelated compounds:

Structurally unrelated compounds and/or concurrently used drugs were spiked at the concentration listed below into low and high controls (225 and 375 ng/mL) in serum and urine. Low control as negative and high control as positive indicate that all the compounds evaluated exhibited minimal cross-reactivity at the concentration tested. Serum and urine samples were run separately. The study results are presented in the table below:

Serum
Structurally Unrelated CompoundsTestedConcentrations(ng/mL)Low Control-25% of cutoff(225 ng/mL)High Control+25% of cutoff(375 ng/mL)
11-nor- $\Delta$ 9-THC-COOH100,000NegativePositive
6-Acetyl morphine75,000NegativePositive
Acetaminophen100,000NegativePositive
Acetylsalicylic acid100,000NegativePositive
Amisulpride100,000NegativePositive
Amoxicillin100,000NegativePositive
Amphetamine100,000NegativePositive
Benztropine Methane Sulfonate3,000NegativePositive
Benzoylecgonine100,000NegativePositive
Brompheniramine3,000NegativePositive
Caffeine100,000NegativePositive
Carbamazepine3,000NegativePositive
Carbamazepine Epoxide10,000NegativePositive
Chloroquine phosphate100,000NegativePositive
Cimetidine100,000NegativePositive
Cocaine75,000NegativePositive
Codeine100,000NegativePositive
Dextromethorphan100,000NegativePositive
Diacetylmorphine (Heroin)100,000NegativePositive
Diazepam100,000NegativePositive
Digoxin100,000NegativePositive
Dihydrocodeine100,000NegativePositive
EDDP Perchlorate100,000NegativePositive
Structurally Unrelated CompoundsSerum
TestedConcentrations(ng/mL)Low Control-25% of cutoff(225 ng/mL)High Control+25% of cutoff(375 ng/mL)
EMDP-HCL100,000NegativePositive
Fentanyl25,000NegativePositive
Glutethimide100,000NegativePositive
Hydrocodone100,000NegativePositive
Hydrocortisone100,000NegativePositive
Hydromorphone100,000NegativePositive
Hydroxyzine3,000NegativePositive
Ibuprofen100,000NegativePositive
Levorphanol-D3100,000NegativePositive
Levothyroxine100,000NegativePositive
Meperidine25,000NegativePositive
Methadone75,000NegativePositive
Methamphetamine100,000NegativePositive
Methaqualone500NegativePositive
Methsuximide75,000NegativePositive
Methylphenidate100,000NegativePositive
Morphine100,000NegativePositive
Morphine-3β-glucuronide100,000NegativePositive
Morphine-6β-glucuronide100,000NegativePositive
Nalbuphine100,000NegativePositive
Nalorphine100,000NegativePositive
Naloxone100,000NegativePositive
Naltrexone100,000NegativePositive
Naproxen100,000NegativePositive
Norcodeine100,000NegativePositive
Nordiazepam100,000NegativePositive
Norethindrone100,000NegativePositive
Norhydrocodone100,000NegativePositive
Noroxycodone100,000NegativePositive
Noroxymorphone100,000NegativePositive
Norpropoxyphene75,000NegativePositive
Oxaprozin100,000NegativePositive
Oxazepam100,000NegativePositive
Oxycodone100,000NegativePositive
Oxymorphone100,000NegativePositive
PCP50,000NegativePositive
Phenobarbital100,000NegativePositive
Phenytoin100,000NegativePositive
Primidone100,000NegativePositive
Procyclidine100,000NegativePositive
Propoxyphene100,000NegativePositive
Secobarbital100,000NegativePositive
Serum
Structurally Unrelated CompoundsTestedConcentrations(ng/mL)Low Control-25% of cutoff(225 ng/mL)High Control+25% of cutoff(375 ng/mL)
Tapentadol100,000NegativePositive
Temazepam100,000NegativePositive
Triprolidine100,000NegativePositive
Valproic Acid100,000NegativePositive
Venlafaxine100,000NegativePositive
Verapamil100,000NegativePositive

Table 20: Cross Reactivity to structurally unrelated compounds in Serum

{21}------------------------------------------------

{22}------------------------------------------------

Table 21: Cross Reactivity to structurally unrelated compounds in Urine

Structurally Unrelated CompoundsTestedConcentrations(ng/mL)Urine
Low Control-25% of cutoff(225 ng/mL)High Control+25% of cutoff(375 ng/mL)
11-nor-Δ9-THC-COOH100,000NegativePositive
6-Acetyl morphine100,000NegativePositive
Acetaminophen100,000NegativePositive
Acetylsalicylic acid100,000NegativePositive
Amisulpride100,000NegativePositive
Amoxicillin100,000NegativePositive
Amphetamine100,000NegativePositive
Benztropine Methane Sulfonate7,000NegativePositive
Benzoylecgonine100,000NegativePositive
Brompheniramine5,000NegativePositive
Caffeine100,000NegativePositive
Carbamazepine5,000NegativePositive
Carbamazepine Epoxide30,000NegativePositive
Chloroquine phosphate100,000NegativePositive
Cimetidine100,000NegativePositive
Cocaine100,000NegativePositive
Codeine100,000NegativePositive
Dextromethorphan100,000NegativePositive
Diacetylmorphine (Heroin)100,000NegativePositive
Diazepam100,000NegativePositive
Digoxin100,000NegativePositive
Dihydrocodeine100,000NegativePositive
EDDP Perchlorate100,000NegativePositive
EMDP-HCL100,000NegativePositive
Fentanyl25,000NegativePositive
Glutethimide100,000NegativePositive
Hydrocodone100,000NegativePositive
Hydrocortisone100,000NegativePositive
Hydromorphone100,000NegativePositive
Hydroxyzine5,000NegativePositive
Ibuprofen100,000NegativePositive
Levorphanol-D3100,000NegativePositive
Urine
Structurally Unrelated CompoundsTestedConcentrations(ng/mL)Low Control-25% of cutoff(225 ng/mL)High Control+25% of cutoff(375 ng/mL)
Levothyroxine100,000NegativePositive
Meperidine25,000NegativePositive
Methadone75,000NegativePositive
Methamphetamine100,000NegativePositive
Methaqualone1,000NegativePositive
Methsuximide75,000NegativePositive
Methylphenidate100,000NegativePositive
Morphine100,000NegativePositive
Morphine-3β-glucuronide100,000NegativePositive
Morphine-6β-glucuronide100,000NegativePositive
Nalbuphine100,000NegativePositive
Nalorphine100,000NegativePositive
Naloxone100,000NegativePositive
Naltrexone100,000NegativePositive
Naproxen100,000NegativePositive
Norcodeine100,000NegativePositive
Nordiazepam100,000NegativePositive
Norethindrone100,000NegativePositive
Norhydrocodone75,000NegativePositive
Noroxycodone100,000NegativePositive
Noroxymorphone100,000NegativePositive
Norpropoxyphene75,000NegativePositive
Oxaprozin100,000NegativePositive
Oxazepam100,000NegativePositive
Oxycodone100,000NegativePositive
Oxymorphone100,000NegativePositive
PCP75,000NegativePositive
Phenobarbital100,000NegativePositive
Phenytoin100,000NegativePositive
Primidone100,000NegativePositive
Procyclidine100,000NegativePositive
Propoxyphene100,000NegativePositive
Secobarbital100,000NegativePositive
Tapentadol100,000NegativePositive
Temazepam100,000NegativePositive
Triprolidine100,000NegativePositive
Valproic Acid100,000NegativePositive
Venlafaxine100,000NegativePositive
Verapamil100,000NegativePositive

{23}------------------------------------------------

{24}------------------------------------------------

  • g) Interference
    The interference study is performed per the CLSI EP07. 3rd edition. The potential interference of endogenous, exogenous, physiological substances, and pH on the recovery of nortriptyline using DRI Tricyclics Serum Tox Assay was assessed. Potentially interfering substances were spiked into the low control, 225ng/mL (-25% of the cutoff concentration of 300 ng/mL) and high controls, 375 ng/mL (+25% of the cutoff concentration of 300 ng/mL). Serum and urine samples were run separately.

As shown in the tables below, the controls were detected accurately. In the presence of the compounds listed below, the controls were detected accurately, indicating that these compounds did not show interference in the assay.

CompoundsTestedConcentrations(mg/dL)Low Control-25% of cutoff(225 ng/mL)High Control+25% of cutoff(375 ng/mL)
Bilirubin (Conjugated)40NegativePositive
Bilirubin (Unconjugated)40NegativePositive
Hemoglobin1000NegativePositive
Albumin7500NegativePositive
y-globulin5000NegativePositive
Rh Factor1300 IUNegativePositive
Triglycerides1500NegativePositive
Cholesterol1400NegativePositive

Table 22: Results of Interfering Substances Testing - Serum

Table 23: Results of Interfering Substances Testing - Urine

CompoundsTestedConcentrations(mg/dL)Low Control-25% of cutoff(225 ng/mL)High Control+25% of cutoff(375 ng/mL)
Acetone500NegativePositive
Ascorbic Acid150NegativePositive
Caffeine5NegativePositive
Creatinine400NegativePositive
Ethanol1000NegativePositive
Galactose5NegativePositive
Glucose1000NegativePositive
Hemoglobin150NegativePositive
Human Serum Albumin(HSA)200NegativePositive
Oxalic acid50NegativePositive
Riboflavin3NegativePositive
Sodium Chloride1000NegativePositive
Urea1000NegativePositive

{25}------------------------------------------------

pH interference for urine only:

The intermediate drug stock at 10.000 ng/mL was used to spike the pH adjusted drug free urine samples to make low control, 225 ng/mL and high control, 375 ng/mL with pH range from 3 to 11. The results are presented in the table below:

Table 24: pH interference study results for urine
pHLow Control-25% of cutoff(225 ng/mL)High Control+25% of cutoff(375 ng/mL)
3NegativePositive
4NegativePositive
5NegativePositive
6NegativePositive
7NegativePositive
8NegativePositive
9NegativePositive
10NegativePositive
11NegativePositive

Table 24: pH interference study results for urine

There is no need to test pH as interfering factor for serum because blood pH range is very narrow. Any continuous pH < 7.3 and > 7.8 are lethal. Varied urine pH is to support in case of sample adulteration.

Specific gravity for urine only:

The specific gravity of multiple negative urine samples from different donors were tested on available clinical chemistry analyzer to select ten (10) negative urine samples with specific gravity range from 1.004 to 1.030.

The selected negative urine samples with a specific gravity range from 1.004 to 1.030 were spiked at control levels of 225 ng/mL and 375ng/mL using intermediate drug stock at 10,000 ng/mL.

Specific GravityLow Control-25% of cutoff(225 ng/mL)High Control+25% of cutoff(375 ng/mL)
1.004NegativePositive
1.006NegativePositive
1.008NegativePositive
1.010NegativePositive
1.011NegativePositive
1.012NegativePositive
1.013NegativePositive
1.016NegativePositive
1.022NegativePositive
1.030NegativePositive
Specific GravityNumber of Patients
1.001-1.01010
1.011-1.02025
1.021-1.03015
>1.0305

{26}------------------------------------------------

H. Conclusion

The information supports a determination of substantial equivalence between DRI™ Tricyclics Serum Tox Assay (K213875) and the predicate device Tricyclic Serum Tox EIA Assay (K983268).

§ 862.3910 Tricyclic antidepressant drugs test system.

(a)
Identification. A tricyclic antidepressant drugs test system is a device intended to measure any of the tricyclic antidepressant drugs in serum. The tricyclic antidepressant drugs include imipramine, desipramine, amitriptyline, nortriptyline, protriptyline, and doxepin. Measurements obtained by this device are used in the diagnosis and treatment of chronic depression to ensure appropriate therapy.(b)
Classification. Class II (special controls). A tricyclic antidepressant drugs test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).