K Number
K213825

Validate with FDA (Live)

Date Cleared
2023-04-28

(506 days)

Product Code
Regulation Number
868.5440
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

This oxygen concentrator is intended for using as an oxygen supplement device in the professional healthcare facility and home healthcare environment. It provides high concentration of oxygen to persons requiring oxygen therapy. This device is to be used as an oxygen supplement and is NOT considered life-supporting or life-sustaining.

Device Description

The Yuyue Oxygen Concentrator models 8F-3A/8F-5A is a portable concentrator which could utilize a molecular sieve and pressure swing adsorption to produce the oxygen. The device could separate nitrogen by absorbing through the molecular sieve when the room air enters the device, and allows the enriched oxygen to be collected. The subject device is intended use for adult population.

The subject device contains primary electrical components including main unit, compressor, fan, flowmeter and circuit breaker, the software is moderate level according to FDA software guideline.

AI/ML Overview

This submission is for an Oxygen Concentrator, models 8F-3A and 8F-5A. The provided document does not contain information related to a study that proves the device meets acceptance criteria in the context of AI/ML performance, human-in-the-loop studies, or expert ground truth adjudication.

Instead, the document focuses on demonstrating substantial equivalence to a predicate device (L4 Oxygen Concentrator (EverFlo), K061261) through non-clinical performance data for regulatory clearance. This involves comparing technical specifications and performance against recognized standards, rather than evaluating AI performance with acceptance criteria typically found in machine learning contexts.

Therefore, many of the requested categories for acceptance criteria and study details cannot be filled from the provided text.

Here's an analysis based on the available information:

1. A table of acceptance criteria and the reported device performance

The document provides a comparison table for substantial equivalence, which includes various technical characteristics and a comparison to the predicate device. These characteristics serve as "acceptance criteria" for showing equivalence to the predicate, rather than explicit performance metrics against a defined threshold specific to a new AI/ML device.

CharacteristicAcceptance Criteria (Subject Device Performance)Predicate Device Performance K061261Comparison
Oxygen Concentration(8F-3A): ≥90%;(8F-5A): 95.5%-87%90-96% from 1 to 5 LPMEquivalent
Flow Rate(8F-3A): 1-3 LPM;(8F-5A): 1-5 LPMfrom 1 to 5 LPMEquivalent
Output Pressure≤70 kPa44.8 kPa (6.5PSIG)Equivalent
IP classificationIP21IP21Same
Complies with ISO 10993-1YesYesSame
Complies with ISO 80601-2-69YesYesSame
Complies with ASTM D4169-16YesYesSame
Complies with ES 60601-1 Electrical SafetyYesYesSame
Complies with IEC 60601-1-2 EMC SafetyYesYesSame
Indication for UseThis oxygen concentrator is intended for using as an oxygen supplement device in the professional healthcare facility and home healthcare environment. It provides high concentration of oxygen to persons requiring oxygen therapy. This device is to be used as an oxygen supplement and is NOT considered life-supporting or life-sustaining.The Respironics L4 Oxygen Concentrator is intended to provide supplemental oxygen to persons requiring oxygen therapy. The device is not intended to be life supporting or life sustaining. The Respironics L4 Oxygen Concentrator is intended for use in the home or hospital/institutional environment.Equivalent

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

Not applicable. The provided document details non-clinical laboratory testing and compliance with various standards, not performance results from a test set of patient data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable. Ground truth establishment by experts is not relevant to this type of device and submission, which focuses on hardware performance and safety standards.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. Adjudication methods are not relevant to this type of device and submission.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is an oxygen concentrator, not an AI/ML-driven diagnostic or interpretative tool.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is a medical device, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

Not applicable. Ground truth for AI/ML performance is not relevant here. The "ground truth" for this device's performance is its ability to meet the specified technical parameters (e.g., oxygen concentration, flow rate) and safety standards through physical testing and engineering measurements.

8. The sample size for the training set

Not applicable. This is not an AI/ML device; therefore, no training set is mentioned.

9. How the ground truth for the training set was established

Not applicable. This is not an AI/ML device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym and full name on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name, "U.S. Food & Drug Administration," written out next to it.

April 28, 2023

Jiangsu Yuyue Medical Equipment & Supply Co., Ltd. % Randy Jiang Sr. consultant Emergo Global Consulting, LLC 2500 Bee Cave Road, Building 1, Suite 300 Austin, Texas 78746

Re: K213825

Trade/Device Name: Oxygen Concentrator model 8F-3A, Oxygen Concentrator model 8F-5A Regulation Number: 21 CFR 868.5440 Regulation Name: Portable Oxygen Generator Regulatory Class: Class II Product Code: CAW Dated: March 27, 2023 Received: March 28, 2023

Dear Randy Jiang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Ting Song -S

For

James J. Lee, Ph.D. Director, Division of Sleep Disordered Breathing, Respiratory and Anesthesia Office of Health Technologies 1, Ophthalmic, Anesthesia, Respiratory, ENT & Dental Devices Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K213825

Device Name

Oxygen Concentrator models 8F-3A&8F-5A

Indications for Use (Describe)

This oxygen concentrator is intended for using as an oxygen supplement device in the professional healthcare facility and home healthcare environment. It provides high concentration of oxygen to persons requiring oxygen therapy. This device is to be used as an oxygen supplement and is NOT considered life-supporting or life-sustaining.

Type of Use (Select one or both, as applicable)

❌ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

K213825

1. Submission Sponsor

Jiangsu Yuyue Medical Equipment& Supply Co., Ltd Yunyang Industrial Park Danyang Jiangsu China Wei Yan, RD/RA Email: weiyan@yuyue.com.cn Phone number: 025-85582701-6225

2. Submission Correspondent

Emergo Global Consulting, LLC 2500 Bee Cave Road Building 1, Suite 300 Austin, TX78746 Office Phone:(512) 327-9997 Contact: Randy Jiang Title: Senior Technical Consultant

3. Date Prepared

April 24, 2023

4. Device Identification

Trade/Proprietary Name:Oxygen Concentrator models 8F-3A/8F-5A
Common/Usual Name:Portable oxygen generator
Classification Name:Portable oxygen generator
Regulation Number:21 CFR 868.5440
Product Code:CAW
Class:II
Classification Panel:Anesthesiology

5. Legally Marketed Predicate Device

Predicate Device:
Trade/Proprietary Name:L4 Oxygen Concentrator (EverFlo)
Common/Usual Name:Portable oxygen generator

{4}------------------------------------------------

Classification Name:Portable oxygen generator
Regulation Number:21 CFR 868.5440
Product Code:CAW
Class:II
Classification Panel:Anesthesiology
510(k) Number:K061261
Manufacturer:RESPIRONICS, INC.

6. Indication for Use Statement

This oxygen concentrator is intended for using as an oxygen supplement device in the professional healthcare facility and home healthcare environment. It provides high concentration of oxygen to persons requiring oxygen therapy. This device is to be used as an oxygen supplement and is NOT considered lifesupporting or life-sustaining.

7. Device Description

The Yuyue Oxygen Concentrator models 8F-3A/8F-5A is a portable concentrator which could utilize a molecular sieve and pressure swing adsorption to produce the oxygen. The device could separate nitrogen by absorbing through the molecular sieve when the room air enters the device, and allows the enriched oxygen to be collected. The subject device is intended use for adult population.

The subject device contains primary electrical components including main unit, compressor, fan, flowmeter and circuit breaker, the software is moderate level according to FDA software guideline.

The patient contacting components are listed in following table, the biocompatibility test and gas pathway biocompatibility test was conducted to subject device against ISO 10993-10, ISO 18562-1, ISO 18562-2, ISO 18562-3.

{5}------------------------------------------------

Name of ComponentType and Duration of ContactDirect or IndirectContact
CompressorExternal communicating Device,Long term, > 30 daysIndirect (gas pathway)
Combination valveExternal communicating Device,Long term, > 30 daysIndirect (gas pathway)
Silicone tubeExternal communicating Device,Long term, > 30 daysIndirect (gas pathway)
Sieve bedExternal communicating Device,Long term, > 30 daysIndirect (gas pathway)
Oxygen tankExternal communicating Device,Long term, > 30 daysIndirect (gas pathway)

8. Substantial Equivalence Discussion

The following table compares the Oxygen Concentrator to the predicate device with respect to indications for use, and technological characteristics, forming the basis for the determination of substantial equivalence. The subject device does not raise any new questions of safety or effectiveness as compared to the predicate device.

{6}------------------------------------------------

CharacteristicsSubject DevicePredicate Device K061261Comparison
Device nameoxygen concentrator models8F-3A/8F-5AL4 OxygenConcentrator(EverFlo)N/A
Classification22Same
Product CodeCAWCAWSame
RegulationNumber21 CFR 868.544021 CFR 868.5440Same
Intended UseProvide supplemental oxygenProvide supplemental oxygenSame
Indication forUseThis oxygen concentrator isintended for using as anoxygen supplement device inthe professional healthcarefacility and home healthcareenvironment. It provides highconcentration of oxygen topersons requiring oxygentherapy. This device is to beused as an oxygen supplementand is NOT considered life-supporting or life-sustaining.The Respironics L4 OxygenConcentrator is intended toprovide supplemental oxygento persons requiring oxygentherapy. The device is notintended to be life supportingor life sustaining.The Respironics L4 OxygenConcentrator is intended foruse in the home orhospital/institutionalenvironment.Equivalent
Environment ofUseprofessional healthcare facilityand home healthcareenvironment.home or hospital /institutional environment.Same
IP classificationIP21IP21Same
PatientInterfacenasal oxygen cannulanasal oxygen cannulaSame
TechnologyPressure Swing Adsorptionwith molecular sievePressure Swing Adsorptionwith molecular sieveSame
OxygenConcentration(8F-3A): ≥90%;(8F-5A): 95.5%-87%90-96% from 1 to 5 LPMEquivalent
CharacteristicsSubject DevicePredicate Device K061261Comparison
Flow Rate(8F-3A): 1-3 LPM;(8F-5A): 1-5 LPMfrom 1 to 5 LPMEquivalent
OutputPressure≤70 kPa44.8 kPa (6.5PSIG)Equivalent
User InterfaceButtonsButtonsSame
LCD Display and Indicator lightLCD Display and Indicator light
AlarmsE 1 low pressureSystem malfunctionSimilar
E 2 high pressure
E 3 low current
E 4 high current
Low oxygen concentrationalarmLow oxygen condition (opi unit only)
E 5 high temperature insidethe deviceN/A
N/Ahigh oxygen flow condition
LL low flowrateLow oxygen flow condition
Power interrupt alarmPower interrupt alarm
Complies withISO 10993-1YesYesSame
Complies withISO 80601-2-69YesYesSame
Complies withASTM D4169-16YesYesSame
Complies withES 60601-1Electrical SafetyYesYesSame
CharacteristicsSubject DevicePredicate Device K061261Comparison
Complies withIEC 60601-1-2EMC SafetyYesYesSame

Table 5A – Comparison of Characteristics

{7}------------------------------------------------

{8}------------------------------------------------

Comparison Discussion Summary:

The Oxygen Concentrator models 8F-3A&8F-5A has the same intended use and the same or similar technological characteristics and functionality as the predicate, the slight differences do not raise new questions of safety and effectiveness as compared to the predicate device.

9. Non-Clinical Performance Data

To demonstrate safety and effectiveness of Oxygen Concentrator models 8F-3A/8F-5A and to show substantial equivalence to the original device, Jiangsu Yuyue Medical Equipment& Supply Co., Ltd completed the following non-clinical tests. Results confirm that the design inputs and performance specifications for the device are met. The device passed the testing in accordance with internal requirements, national standards, and international standards shown below, supporting its safety and effectiveness, and its substantial equivalence to the predicate device:

  • Electrical safety testing per ES 60601-1
  • Electromagnetic Disturbance (EMD) testing per IEC 60601-1-2
  • . General requirements, tests and guidance for alarm systems per IEC 60601-1-8
  • General requirements for basic safety and essential performance in the home healthcare environment per IEC 60601-1-11
  • Particular requirements for the basic safety and essential performance of oxygen concentrator equipment per ISO 80601-2-69
  • Biocompatibility in gas pathway per ISO 18562-1, ISO 18562-2, ISO 10993-5, ISO 10993-10.
  • Software verification and validation per IEC 62304/FDA Guidance – results /conclusion
  • . Shelf Life Testing - Supports the claimed shelf life
  • . Transportation Testing per ASTM D4169 – Demonstrates package integrity maintained
  • . Altitude limit 2000m device performance test
  • Accessory compatibility test including connect security, accessory performance

10. Statement of Substantial Equivalence

The Oxygen Concentrator models 8F-3A/8F-5A has the same intended use as the predicate device, and similar technological characteristics. The differences in technological characteristics do not raise new or different questions of safety and effectiveness. Performance testing has demonstrated the Oxygen Concentrator models 8F-3A/8F-5A is as safe and effective as the predicate device. Therefore, the Oxygen Concentrator models 8F-3A/8F-5A is substantially equivalent to the predicate device.

§ 868.5440 Portable oxygen generator.

(a)
Identification. A portable oxygen generator is a device that is intended to release oxygen for respiratory therapy by means of either a chemical reaction or physical means (e.g., a molecular sieve).(b)
Classification. Class II (performance standards).