The device is intended for medical purposes to provide mobility to persons restricted to a sitting position.
A mechanical wheelchair is a chair with wheels, designed to be a replacement for walking, where it is propelled by the seated occupant turning the rear wheels by hand. There are also handles behind the seat for someone else to do the pushing. Wheelchairs are used by people for whom walking is difficult or impossible due to illness, injury, or disability.
The device is mainly composed of painted frame, fireproof and breathable seat cushion, armrest, backrest, hand push tube, detachable leg, 8-inch front wheel, 24-inch rear wheel, brake group and anti-overturn device.
The provided document describes the safety and performance testing for a manual wheelchair and not an AI or imaging device. Therefore, many of the requested fields are not applicable to this submission.
Here's the information that can be extracted and a note for non-applicable fields:
1. A table of acceptance criteria and the reported device performance
| Description | Acceptance Criteria | Reported Device Performance |
|---|---|---|
| Static stability | Meets ISO 7176-1:2014 | Meets ISO 7176-1:2014 |
| Effectiveness of brakes | Meets ISO 7176-3:2012 | Meets ISO 7176-3:2012 |
| Dimensions, mass and maneuvering space | Meets ISO 7176-5:2008 | Meets ISO 7176-5:2008 |
| Seating and wheel dimensions | Meets ISO 7176-7:1998 | Meets ISO 7176-7:1998 |
| Static, impact, and fatigue strengths | Meets ISO 7176-8:2014 | Meets ISO 7176-8:2014 |
| Information disclosure, documentation and Labeling | Meets ISO 7176-15:1996 | Meets ISO 7176-15:1996 |
| Resistance to ignition | Meets ISO 7176-16:2012 | Meets ISO 7176-16:2012 |
| General requirements for wheelchairs | Meets ISO 7176-11:2012 | Tests performed, met requirements |
| Determination of friction of test surface | Meets ISO 7176-13:1989 | Tests performed, met requirements |
| Set-up procedures for wheelchairs | Meets ISO 7176-22:2014 | Tests performed, met requirements |
| Biocompatibility (risk management, cytotoxicity, irritation/hypersensitivity) | Meets ISO 10993-1:2009, ISO 10993-5:2009, ISO 10993-10:2010 | Tests performed, met requirements |
| Usability | User Manual understood and device operability | 15 participants understood the User Manual and operated the device |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Test set sample size: Not explicitly stated for each specific ISO test, but a "Usability Study" was conducted with 15 participants.
- Data provenance: Not specified in the provided text. The manufacturing company is in China, which might suggest where the testing occurred, but it's not explicitly stated for the data provenance. It is a non-clinical/physical performance test, so "retrospective or prospective" is not directly applicable in the same way it would be for clinical data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This is not applicable as the document describes physical performance testing of a device against international standards, not expert-derived ground truth for an AI system. The "ground truth" here is the adherence to the published ISO standards and internal specifications.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This is not applicable. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies with human assessors, not for determining mechanical compliance with ISO standards.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This is not applicable. The device is a manual wheelchair, not an AI or imaging device that would involve human readers or AI assistance.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This is not applicable. The device is a manual wheelchair, not an algorithm. Performance tests were conducted on the physical device itself.
7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)
The "ground truth" for the performance tests is adherence to international consensus standards (ISO 7176 series and ISO 10993 series) and the manufacturer's internal specifications.
8. The sample size for the training set
This is not applicable. There is no AI model or "training set" for a manual wheelchair.
9. How the ground truth for the training set was established
This is not applicable. There is no AI model or "training set" for a manual wheelchair.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 14, 2022
Mobility Source Medical Technology Co., Ltd. % Shanfeng Jiang Regulation Control Manager Guangzhou Junyi Information Technology Co., Ltd. Room 304, Building A, 62 Nanyun2 Road Huangpu District, Science City, Guangdong 510663 China
Re: K213790
Trade/Device Name: Manual wheelchair SYIV 100-LQXAL2482-407 Regulation Number: 21 CFR 890.3850 Regulation Name: Mechanical Wheelchair Regulatory Class: Class I, reserved Product Code: IOR Dated: December 6, 2021 Received: December 6, 2021
Dear Shanfeng Jiang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Heather Dean, PhD Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K213790
Device Name Manual Wheelchair SYIV1OO-LQXAL2482-407
Indications for Use (Describe)
The device is intended for medical purposes to provide mobility to persons restricted to a sitting position.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary
Date of Summary Preparation: Nov,08, 2021
1. Submitter's Identifications
Submitter's Name: Mobility Source Medical Technology Co., Ltd.
-
- Address: 160 Heheng Road, Yanghe Town, Gaoming District, Foshan City, Guangdong Province, China, 528513
2 ) Contact Person: Evan Wong Contact Title: QA Manager Contact E-mail Address: evan@lqxmedical.com Telephone: +86- 757-88807188 Fax: +86-757-88809768 3 ) Manufacturer: Guangdong Shunde LQX Health Technology Inc. Address: Gujian Industrial Zone, Liangle Road Daliang Town, Shunde District Foshan City, Guangdong Province, China, 528000 Tel: +86-757-28091601 Fax: +86-757-28091602 Email: evan@lqxmedical.com
- Address: 160 Heheng Road, Yanghe Town, Gaoming District, Foshan City, Guangdong Province, China, 528513
2. Correspondent's Identifications
Correspondent's Name: Guangzhou Junyi Information Technology Co., Ltd. Address: Room 304, building A, 62 Nanyun2 Road, Science City, Huangpu District, Guangzhou China ZIP Code: 510663 Contact Person: Shanfeng Jiang Contact Title: Regulation Control Manager Contact E-mail Address:jiang13620586569@126.com Telephone: +86-20-82329549 Fax: +86-20-82329549
3. Name of the Device
Device Classification Name: Wheelchair, Mechanical Product Name: Manual wheelchair SYIV100-LQXAL2482-407 Trade Name: Aluminum back folding wheelchair Model: SYIV100-LQXAL2482-407 Classification Panel: Physical Medicine Regulation Number: 21 CFR 890.3850 Product Code: IOR Device Classification: Class I
4. The Predicate Devices
{4}------------------------------------------------
AST Model MA012 and MS019 Rehab Wheelchair K181795
5. Device Description
Operating principle
A mechanical wheelchair is a chair with wheels, designed to be a replacement for walking, where it is propelled by the seated occupant turning the rear wheels by hand. There are also handles behind the seat for someone else to do the pushing. Wheelchairs are used by people for whom walking is difficult or impossible due to illness, injury, or disability.
【Structure composition
The device is mainly composed of painted frame, fireproof and breathable seat cushion, armrest, backrest, hand push tube, detachable leg, 8-inch front wheel, 24-inch rear wheel, brake group and anti-overturn device.
【Main materials
Frame: high strength aluminum alloy 6061 - T6, external diameter of Ø 25.4 mm, wall thickness of 2.0 mm.
Armrest: high strength aluminum alloy 6061 - T6, external diameter of Ø 19 mm, wall thickness of 1.5 mm
Brake assembly: aluminum alloy
Seat and backrest: flame resistant fabrics
Tire: PU
6. Intended Use of Device
The device is intended for medical purposes to provide mobility to persons restricted to a sitting position.
7. Summary of Substantial Equivalence
Table 1 Comparison to Predicate Device
| Proposed Device | Predicate device | Comparison | |
|---|---|---|---|
| 510k Number | K213790 | K181795 | ------ |
| Product Code | IOR | IOR | Same |
| Proprietary Name | ManualWheelchairSYIV100-LQXAL2482-407 | AST Model MA012 andMS019 Rehab Wheelchair | ------ |
| Model | SYIV100-LQXAL2482-407 | MA012 and MS019 | ------ |
| Manufacturer | Guangdong Shunde LQXHealth Technology Inc. | Sichuan AST MedicalEquipment Co., Ltd. | ------ |
| Indications for Use | The device is intendedfor medical purposes toprovide mobility topersons restricted to a | The AST Model MA012 andMS019 Rehab Wheelchairsare to provide mobility topersons limited to a sitting | Same |
| sitting position. | position. | ||
| Type of Use | Over-the-counter | Over-the-counter | Same |
| Basic Design | 1. The wheelchairadopts aluminum alloytube frame andergonomic design, thesurface of the pianopaint treatment, toachieve high strength,high load bearing, lightweight, small volume,and at any time foldingstructure, do not take upspace, easy to carry;2. Seat depth andbackrest angle can beadjusted according tohuman comfort.3. Backrest height canbe adjusted.4. The armrest of thewheelchair can be liftedback, the height of thearmrest can be adjusted,the front and back canslide, the footplate canrotate, and the angle canbe adjusted.5. Front and rear wheelsadopt PU solid tire,fireproof and breathableseat cushion andbackrest.6. Backrest can be backfolding.7. The device can beequipped with drumbrake, easy for thecaregiver control anddown the steep slopecontrol flexibility andstability. | The AST Model MA012 andMS019 Rehab Wheelchair aremanual wheelchairs.They have adjustablearmrests, and multiple axleposition. The casters are6"/7"/8" PU wheels withheight adjustable forks andthe rear wheels are20"/22"/24"*1-3/8"polyurethane (MA012 andMS019).Aluminum frame with liquidcoated and folding backrestFlip-up and detachablearmrest PU padDetachable footrestFront casters PU tires andquick release rear wheel PUtiresNylon cushion | SimilarThe structureand materialsof mainframe of theproposeddevice andpredicatedevice aresame, onlyindividualmaterials aredifferent,such as backupholsteryand seatupholstery,etc.Thedifferencesdo not raisesafety andeffectivenessof theproposeddevice. |
{5}------------------------------------------------
{6}------------------------------------------------
| Materials | with aluminum alloybrake assembly.9. There is Equippedwith anti-overturndevice.Frame: high strengthaluminum alloy 6061 -T6, external diameter of$Ø$ 25.4 mm, wallthickness of 2.0 mm.Armrest: high strengthaluminum alloy 6061 -T6, external diameter of$Ø$ 19 mm, wallthickness of 1.5 mmBrake assembly:aluminum alloySeat and backrest: flameresistant fabricsTire: PU | AST Model MA012 RehabWheelchair: Aluminum framewith liquid coated and foldingbackrestFlip-up and detachablearmrest PU padDetachable footrestFront casters PU tires andquick release rear wheel PUtiresNylon cushion | SimilarThe Mainstructuralmaterials andwheelsmaterials aresame.Thedifferencesdo not raisesafety andeffectivenessof theproposeddevice. | ||
|---|---|---|---|---|---|
| Components | The device is mainlycomposed of paintedframe, fireproof andbreathable seat cushion,armrest, backrest, handpush tube, detachableleg, 8-inch front wheel,24-inch rear wheel,brake group andanti-overturn device. | AST Model MA012 RehabWheelchair: Main frame,handle sleeve, backupholstery, seat upholstery,armrest pad, armrest, sidepanel, front casters, rearwheel, legrest, footplate,brake, front fork, cross-brace. | SameAlthough theexpressionsare different,the maincomponentsare same. | ||
| Control Mode | Mechanical | Mechanical | Same | ||
| Length | 1070mm | Model MA012:1173mm( $\pm$ 1mm) | SimilarThe twophysical | ||
| Dimension | Width | 625mm | Model MA012:645mm ( $\pm$ 1mm) | dimensionsare different.The | |
| Height | 1035mm | Model MA012:892mm ( $\pm$ 1mm) | differencedoes notaffect theeffectivenessand safety |
{7}------------------------------------------------
| Net weight | 38.3 lbs/17.4kg | Model MA012:17.2kg/38lbs | SimilarTheproposeddeviceisheavier thanthe predicatedevice. | |
|---|---|---|---|---|
| Weight capacity | 300lbs/136kg | Model MA012:300lbs/136kg | Same | |
| Seat width | 446mm(17.6") | Model MA012:16"(406mm)18"(457mm)20"(508mm) | SimilarDifferentseat sizes donot raises thesafetyandeffectivenessofthedevice. | |
| Seat height | 518mm(20.4") | Model MA012:19.7"(500mm) | ||
| Seat depth | 495mm(19.5") | Model MA012:16"-20"(406mm-508mm) | ||
| Frame type | foldable | foldable | Same | |
| Cross-braceconfiguration | 18" | 14", 16", 18", 20" or 22" | SameCross-bracesizesofpredicatedeviceinclude withthatofproposeddevice. | |
| Back style | Fixed | Model MS019: FixedModel MA012: Adjustable | Same | |
| Anti-tippers | Optional | Optional | Same | |
| Wheelconstruction | Quick release | Quick release | SameWheelconstructionof proposeddeviceissameaspredicatedevice modelSIVFH2A102. | |
| Tires | Front:200mm (8")Rear: 600mm(24") | Front: 6",7",8"Rear: 20",22",24" | SameThetires | |
| sizes ofproposeddevice areincluded tothose ofpredicatedevice. | ||||
| Armrest | Flip back armrest | Model MA012: HeightAdjustable desk lengtharmrest, Flip backModel MS019: Fixed oradjustable height; desk or fulllength; removable | SimilarThe height ofarmest ofproposeddevice cannot beadjustable. | |
| Foot rest | Optional/ swing away | Optional/ swing away | Same | |
| Rear Axle Position | Single | Multiple | SimilarThe rear axleof proposeddevice hasonly onesize. | |
| Frame Material | high strength aluminumalloy | Model MA012: AluminumModel MS019: Steel | SameThe framematerial ofproposeddevice issame aspredicatedevice modelMA012. | |
| Safety Feature | Manual Wheel Lock | Manual Wheel Lock | Same | |
| Standard | ISO7176-1ISO7176-3ISO7176-5ISO7176-7ISO7176-8ISO7176-11ISO7176-13ISO7176-15ISO7176-16ISO7176-22ISO10993-1ISO10993-5 | ISO7176-1ISO7176-3ISO7176-5ISO7176-7ISO7176-8ISO7176-11ISO7176-13ISO7176-15ISO7176-16ISO7176-22ISO10993-1ISO10993-5 | Same | |
| ISO10993-10 | ISO10993-10 | |||
{8}------------------------------------------------
{9}------------------------------------------------
8. Substantial Equivalence discussion:
Our device and the predicate device are almost identical in terms of all areas described in the above table (Table 1). There are some minor differences with the predicate device don't affect the safety or effectiveness of the subject device.
The following table (Table 2) shows similarities and differences of the performance between our device and the predicate device. Tests were conducted following the recommended procedures outlined in the respective consensus standards, and results for "Manual Wheelchair SYIV100-LQXAL2482-407" met all relevant requirements in the test standards, our internal specifications, and are comparable to the predicate device.
| Description | Proposed Device | Predicate device |
|---|---|---|
| Static stability | Meets ISO 7176-1:2014 | Meets ISO 7176-1:2014 |
| Effectiveness of brakes | Meets ISO 7176-3:2012 | Meets ISO 7176-3:2012 |
| Dimensions, mass andmaneuvering space | Meets ISO 7176-5:2008 | Meets ISO 7176-5:2008 |
| Seating and wheeldimensions | Meets ISO 7176-7:1998 | Meets ISO 7176-7:1998 |
| Static, impact, and fatiguestrengths | Meets ISO 7176-8:2014 | Meets ISO 7176-8:2014 |
| Information disclosure,documentation andLabeling | Meets ISO 7176-15:1996 | Meets ISO 7176-15:1996 |
| Resistance to ignition | Meets ISO 7176-16:2012 | Meets ISO 7176-16:2012 |
Table 2: Comparison of Performance Testing
The tests were performed following the general requirements outlined in ISO 7176-11:2012, ISO 7176-13:1989, ISO 7176-22:2014.
A brief discussion of the non-clinical testing data that was submitted, referenced or relied on to demonstrate that the Proposed Device is safe and effective, and whose performance meets the requirements of its user-defined acceptance criteria and intended uses:
"Mechanical Wheelchair SIVFH2A102" meets performance requirements per ISO 7176-1:2014, ISO 7176-3: 2012, ISO 7176-5:2008, ISO 7176-7:1998, ISO 7176-8:2014, ISO 7176-11:2012, 7176-13: 1989, ISO 7176-15: 1996, ISO 7176-16:2012 and ISO 7176-22:2014. It is safe and effective, and its performances meet the requirements of the pre-defined acceptance criteria and intended use.
A brief discussion of the clinical submitted, reference, or relied on in the premarket notification submission for a determination of substantial equivalence.
According to the non-clinical test results, the proposed devices are as safe, as effective and perform as well as the predicate device. So the proposed device is
{10}------------------------------------------------
Substantially Equivalent (SE) to the predicate device which is US legally market device.
9. Non-Clinical Tests Performed:
following testing was performed on the "Mechanical Wheelchair The SYIV100-LQXAL2482-407" in accordance with the requirements of the design control regulations and established quality assurance procedures.
Safety and performance:
ISO 7176-1:2014 Wheelchairs - Part 1: Determination of static stability
ISO 7176-3:2012 Wheelchairs - Part 3: Determination of effectiveness of brakes
ISO 7176-5:2008 Wheelchairs - Part 5: Determination of dimensions, mass and maneuvering space
ISO 7176-7:1998 Wheelchairs - Part 7: Measurement of seating and wheel dimensions
ISO 7176-8:2014 Wheelchairs - Part 8: Requirements and test methods for static, impact and fatigue strengths
ISO 7176-11:2012 Wheelchairs - Part 11: Test dummies
ISO 7176-13:1989 Wheelchairs – Part 13: Determination of friction of test surface
ISO 7176-15:1996 Wheelchairs - Part 15: Requirements for information disclosure, documentation and labeling
ISO 7176-16:2012 Wheelchairs - Part 16: Resistance to ignition of postural support devices
ISO 7176-22:2014 Wheelchairs - Part 22: Set-up procedures
Biocompatibility:
ISO 10993-1:2009 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
ISO 10993-5:2009 Biological evaluation of medical devices — Part 5: Tests for in vitro cytotoxicity
ISO 10993-10:2010 Biological evaluation of medical devices -- Part 10: Tests for irritation and delayed-type hypersensitivity
10. Usability Study:
The 15 participants read the User Manual before initial use of the device. They can understand the questions and answer them. The 15 participants can operate the device referring to the User Manual.
Based on the verification & validation testing records provided, we concluded that the Manual Wheelchair, Model SYIV100-LQXAL2482-407 is considered to meet the usability requirement as defined in the verification & validation test report.
These non-clinical tests demonstrate that the labeling for the Manual Wheelchair, SYIV100-LQXAL2482-407, Model by Guangdong Shunde LQX Health Technology Inc. and bv US-Distributors is comprehensive, was well understood by a variety of "patients", and provided sufficient information for safe and proper use of the device.
{11}------------------------------------------------
11. Conclusion:
Based on the comparison of the proposed device of SYIV100-LQXAL2482-407 is determined to be Substantially Equivalent (SE) to the predicate device of AST Model MA012 and MS019 Rehab Wheelchair, in respect of safety and effectiveness.
N/A