(200 days)
The Intelligence Laser Comb (Model: YTLC001-W ) is indicated to promote hair growth in males with androgenic alopecia who have Norwood-Hamilton classifications of IIa-V or females with androgenic alopecia who have Ludwig-Savin Classifications of I-4, II-1, II-2, or frontal and both with Fitzpatrick Skin Phototypes I-IV.
The Intelligence Laser Comb (Model: YTLC001-W) is a home-use comb-shaped low level laser therapy (LLLT) device that emits laser light with the intention to promote hair growth in men and women. The device provides distributed laser to the scalp at 655+/-10nm while the comb teeth simultaneously part the user's hair to ensure the laser light reaches the user's scalp. The Laser Comb is an over-the-counter (OTC) device intended for home-use. The Laser Comb works by providing laser energy to stimulate hair follicles. For optimal results, the laser must not be blocked by the hair and must have an unobstructed path to the scalp. The Laser Comb has teeth that part the hair and allow the maximum amount of laser energy to reach the scalp. The Laser Hair Growth Comb design with plastic Mainframe, it is composed of PCB circuit board, Power Button, Indicator Light, Red Laser Light, Teeth, USB charging port. And the device is equipped with one internal rechargeable lithium battery (3.7 V d.c. 500mA) that can be charged directly by external adapter. The device uses beep and vibrate to alert the user to the next step during treatment.
The provided document is a 510(k) Summary for the Intelligence Laser Comb (Model: YTLC001-W). This document does not describe acceptance criteria for a "device performance" relating to the intended use (hair growth) or a study proving it meets such criteria. Instead, it details that clinical performance is not deemed necessary and focuses on proving substantial equivalence to a predicate device through conformity to recognized international standards for electrical safety, electromagnetic compatibility, and biocompatibility.
Here's a breakdown of the information requested, based on the provided text:
1. A table of acceptance criteria and the reported device performance
The document does not specify quantitative acceptance criteria for the device's efficacy in promoting hair growth, nor does it report on such performance for the subject device. It relies on demonstrating substantial equivalence to a predicate device which does have an indication for hair growth.
The acceptance criteria and reported performance relate to safety and compliance with standards:
| Acceptance Criteria (Standards Conformance) | Reported Device Performance |
|---|---|
| Biocompatibility: | |
| ISO 10993-1 Fifth edition 2018-08 (Evaluation and testing within a risk management process) | Biocompatibility evaluation conducted in accordance with ISO 10993-1. |
| ISO 10993-5:2009 (Tests for In Vitro cytotoxicity) | The battery of testing was performed to, and passed, ISO 10993-5 (In Vitro cytotoxicity). |
| ISO 10993-10:2010 (Tests for irritation and skin sensitization) | The battery of testing was performed to, and passed, ISO 10993-10 (irritation and skin sensitization). |
| Electrical Safety: | |
| IEC 60601-1:2012 (General Requirements for Basic Safety and Essential Performance) | Electrical safety testing was performed to, and passed, IEC 60601-1. |
| IEC 60601-1-11:2015 (Requirements for medical electrical equipment and systems used in the home healthcare environment) | Electrical safety testing was performed to, and passed, IEC 60601-1-11. |
| IEC 60825:2014 (Safety of laser products - Part 1: Equipment classification and requirements) | Electrical safety testing was performed to, and passed, IEC 60825:2014. Classification according to IEC60825-1 is Class 3R, which matches the predicate device. |
| Electromagnetic Compatibility (EMC): | |
| IEC 60601-1-2:2014 (Collateral standard: Electromagnetic compatibility) | EMC safety testing was performed to, and passed, IEC 60601-1-2. |
| Software Verification and Validation: | |
| FDA "Guidance for Premarket Submissions and for Software Contained in Medical Devices" Software Life Cycle Processes | Software verification and validation test was performed to these FDA guidance and IEC 62304:2006+AMD1:2015 Medical Device Software. No specific performance metrics or acceptance criteria are detailed beyond "test was performed" and "conformance" to these. |
| IEC 62304:2006+AMD1:2015 Medical Device Software (Software Life Cycle Processes) | Software verification and validation test was performed to this standard. |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document explicitly states: "Summary for clinical test: Clinical performance is not deemed necessary." Therefore, there is no test set, sample size, or data provenance related to the device's efficacy in promoting hair growth. The data provenance for the safety and compatibility tests is not specified, but the applicant company is located in China.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
Not applicable, as no clinical performance study was conducted or deemed necessary.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable, as no clinical performance study was conducted or deemed necessary.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is a low-level laser therapy device for hair growth, not an AI-based diagnostic or imaging device involving human readers.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This device is a physical laser comb, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
Not applicable, as no clinical performance study was conducted or deemed necessary for hair growth efficacy. For the safety and compliance tests, the "ground truth" is adherence to the specified international standards.
8. The sample size for the training set
Not applicable, as no clinical performance study was conducted or deemed necessary, and this is not an AI/machine learning device that requires a training set for its intended use.
9. How the ground truth for the training set was established
Not applicable.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" arranged in a circular pattern. To the right of the seal, there is the FDA logo in blue, with the words "U.S. FOOD & DRUG" on top and "ADMINISTRATION" below. The logo is simple and professional, reflecting the FDA's role in regulating food and drugs.
June 24, 2022
Yibin Yingtong Intelligent Technology Co.,Ltd. % Iris Fung Official Correspondent SGS-CSTC Standards Technical Services Co., Ltd. 198 Kezhu Road, Scientech Park Guangzhou Economic & Technology Development District Guangzhou, Guangdong 510000 China
Re: K213789
Trade/Device Name: Intelligence LaserComb Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared Lamp Regulatory Class: Class II Product Code: OAP Dated: May 10, 2022 Received: May 11, 2022
Dear Iris Fung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809 medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Jianting Wang Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K213789
Device Name Intelligence Laser Comb (Model: YTLC001-W )
Indications for Use (Describe)
The Intelligence Laser Comb (Model: YTLC001-W ) is indicated to promote hair growth in males with androgenic alopecia who have Norwood-Hamilton classifications of IIa-V or females with androgenic alopecia who have Ludwig-Savin Classifications of I-4, II-1, II-2, or frontal and both with Fitzpatrick Skin Phototypes I-IV.
| Type of Use (Select one or both, as applicable) |
|---|
| ------------------------------------------------- |
| Presentation Use of CFR 221 Subpart C | One-Time Contact Use of CFR 221 Subpart C |
|---|---|
| ------------------------------------------------------------------------------------------------------------------------------------ | --------------------------------------------------------------------------------------------------------------------------------------------------- |
__ Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Date of the summary prepared: June 17, 2022
510(k) Summary
510K number: K213789
This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92. This is a traditional 510(K) submission with no previous application.
1. Submitter's Information
Sponsor
- � Company Name: YIBIN YINGTONG INTELLIGENT TECHNOLOGY CO.,LTD.
- Address: 3F, Building 1, Area A, Intelligent terminal industrial park, Yibin economic development � zone, Sichuan province, China
- � Phone: +86 0831-2020866
- Contact Person (including title): Pengcheng Pan (Manager) �
- E-mail: tony@interthings.cn �
Application Correspondent
- � SGS-CSTC Standards Technical Services Co., Ltd.
- Address: 198 KEZHU Road, SCIENTECH Park Guangzhou Economic & Technology � Development District, Guangzhou, Guangdong, CHINA
- Contact Person: Ms. Iris Fung �
- � Tel: +86-20-32136908
- � Email: Iris.Fung@sgs.com; jianda-lee@foxmail.com
2. Subject Device Information
- ◆ Type of 510(k) submission: Traditional
- ♦ Common Name: Lamp, non-heating, for promotion of hair growth
- ♦ Trade Name: Intelligence Laser Comb
- � Model: YTLC001-W
- ◆ 510(K) Number: K213789
- ♦ Classification Name: Laser, Comb, Hair
- ♦ Review Panel: General & Plastic Surgery
- ◆ Product Code: OAP
- ◆ Requlation Number: 890.5500
- ◆ Regulation Class: II
{4}------------------------------------------------
3. Predicate Device Information
| Predicate Device | |
|---|---|
| Sponsor | OMM IMPORTS INC DBA ZEROGRAVITY |
| Device Name | Laser Hair Therapy |
| Model | Recreo 200 |
| 510(k) Number | K183329 |
| Product Code | OAP |
| Regulation Number | 890.5500 |
| Regulation Class | II |
2. Device Description
The Intelligence Laser Comb (Model: YTLC001-W) is a home-use comb-shaped low level laser therapy (LLLT) device that emits laser light with the intention to promote hair growth in men and women. The device provides distributed laser to the scalp at 655+/-10nm while the comb teeth simultaneously part the user's hair to ensure the laser light reaches the user's scalp. The Laser Comb is an over-the-counter (OTC) device intended for home-use.
The Laser Comb works by providing laser energy to stimulate hair follicles. For optimal results, the laser must not be blocked by the hair and must have an unobstructed path to the scalp. The Laser Comb has teeth that part the hair and allow the maximum amount of laser energy to reach the scalp.
The Laser Hair Growth Comb design with plastic Mainframe, it is composed of PCB circuit board, Power Button, Indicator Light, Red Laser Light, Teeth, USB charging port. And the device is equipped with one internal rechargeable lithium battery (3.7 V d.c. 500mA) that can be charged directly by external adapter. The device uses beep and vibrate to alert the user to the next step during treatment.
3. Intended Use / Indications for Use
The Intelligence Laser Comb (Model: YTLC001-W) is indicated to promote hair growth in males with androgenic alopecia who have Norwood-Hamilton classifications of Ila-V or females with androgenic alopecia who have Ludwig-Savin Classifications of I-4, II-1, II-2, or frontal and both with Fitzpatrick Skin Phototypes I-IV.
4. Test Summary
{5}------------------------------------------------
The Intelligence LaserComb (Model: YTLC001-W) was evaluated for conformance to recognized international standards. The following is a list of these evaluations and tests that were found to be in conformance:
. Electrical safety test
IEC 60601-1:2012, Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance
IEC60601-1-11:2015 medical electrical equipment - part 1-11: general requirements for basic safety and essential performance - collateral standard: requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
IEC 60825:2014 Safety of laser products - Part 1: Equipment classification and requirements
. Electromaqnetic compatibility test
IEC 60601-1-2:2014, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility
. Biocompatibility test
ISO 10993-5:2009, Biological evaluation of medical devices -- Part 5: Tests for In Vitro cytotoxicity, ISO 10993-10:2010, Biological evaluation of medical devices -- Part 10: Tests for irritation and skin sensitization
. Software verification and validation test
FDA "Guidance for Premarket Submissions and for Software Contained in Medical Devices"
- Software Life Cycle Processes
IEC 62304:2006+AMD1:2015 Medical Device Software
5. Comparison to predicate device and conclusion
The subject device Intelligence LaserComb (Model: YTLC001-W) is substantially equivalent to the predicated device based on intended use, design, specifications and performance. The differences between the subject device and predicate devices do not raise new issues of safety or effectiveness. Information for predicate device was obtained from publicly available sources. A technical comparison to the predicate is provided below.
| Elements ofComparison | Subject Device | Predicate Device | Reference Device |
|---|---|---|---|
| 510(k) Number | K213789 | K183329 | K193008 |
| Device Name | Intelligence LaserComb(YTLC001-W) | Laser Hair Therapy /(Model:Recreo 200) | Tricoglam Home USE |
| Product Code | OAP | OAP | OAP |
| Regulation Number | 21 CFR 890.5500 | 21 CFR 890.5500 | 21 CFR 890.5500 |
| Regulation Class | II | II | II |
| Prescription | OTC | OTC | Not disclosed |
| Elements ofComparison | Subject Device | Predicate Device | Reference Device |
| Intended Use | The Intelligence Laser Comb(Model: YTLC001-W) is indicated to promote hair growth in males with androgenic alopecia who haveNorwood-Hamilton classifications of IIa-V or females with androgenic alopecia who have Ludwig-Savin Classifications of I-4, II-1, II-2, or frontal and both with Fitzpatrick Skin Phototypes I-IV. | Laser Hair Therapy is indicated to promote hair growth in males with androgenic alopecia who haveNorwood-Hamilton classifications of IIa-V or females with androgenic alopecia who have Ludwig-Savin Classifications of I-4, II-1, II-2, or frontal and both with Fitzpatrick Skin Phototypes I-IV. | Tricoglam Home Use is indicated to promote hair growth in females with androgenetic alopecia who have Ludwig-Savin Classifications I - II, in males with androgenetic alopecia who haveNorwood Hamilton Classifications IIa - V and for both, Fitzpatrick Classification of Skin Phototypes of I - IV |
| Type of Laser | Visible red light-emitting diodes | Visible red light-emitting diodes | Visible red LED diodes |
| Wavelength | 655nm±10nm | 655nm±5nm | 650 nm +/- 10 nm |
| Amounts of laserdiode | 10 | 12 | 105 |
| Energy of Per LaserDiode | ≤5mW | ≤5mW | 5mW |
| Pow er intensity | 2.5 mW/cm² | Not disclosed | Not disclosed |
| Dosage intensity | 1.2 J/cm² | Not disclosed | 1.2 J/cm² |
| Classificationaccording toIEC60825-1 | Class 3R | Class 3R | Not disclosed |
| Treatment Time | 8 minutes per treatment | 8 minutes per treatment | 20 minutes |
| Treatment Frequency | 3 times per week (every other day) | 3 times per week (every other day) | Continuous every day |
| Applicable People | Norwood Hamilton IIa-V (males)Ludwig-Savin I-4, II-1, II-2, or frontal (females) | Norwood Hamilton IIa-V (males)Ludwig-Savin I-4, II-1, II-2, or frontal (females) | Ludwig-Savin Classification s I - II(females); Norwood Hamilton Classification s IIa - V (males). |
| Applicable Skin | Fitzpatrick Skin Phototypes I-IV | Fitzpatrick Skin Phototypes I-IV | Fitzpatrick Skin Phototypes I-IV |
| Appearance Design | Comb | Comb | Helmet |
| Safety andPerformance Feature | Complied with IEC 60601-1, IEC 60601-1-2, IEC60601-1- 11,IEC 60825-1 | Complied with IEC 60601-1, IEC 60601-1-2, IEC 60825-1, IEC 60601-1-11 | Complied with IEC 60601-1, IEC 60601-1-2, IEC 62471, IEC 60601-2-57, IEC 60601-1-11 |
| Biocompatibility | All patient contacting materials are complied with ISO 10993-5, ISO 10993-10 | All patient contacting materials are complied with ISO 10993-1, ISO 10993-5 | All patient contacting materials are complied with ISO 10993-1, ISO 10993-5 |
{6}------------------------------------------------
{7}------------------------------------------------
Performance Data:
The following performance data were provided in support of the substantial equivalence determination.
- Biocompatibility Testing
The biocompatibility evaluation for the body-contacting components of the Intelligence LaserComb (Model: YTLC001-W) was conducted in accordance with the ISO 10993-1 Fifth edition 2018-08 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process, as recognized by FDA.
The battery of testing was performed to, and passed, including:
ISO 10993-5:2009, Biological evaluation of medical devices -- Part 5: Tests for In Vitro cytotoxicity,
ISO 10993-10:2010, Biological evaluation of medical devices -- Part 10: Tests for irritation and skin sensitization
- Electrical and EMC Safety
Electrical safety and EMC safety testing was performed to, and passed, the following standards:
IEC 60601-1:2012, Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance
IEC60601-1-11:2015 medical electrical equipment - part 1-11: general requirements for basic safety and essential performance - collateral standard: requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
IEC 60825:2014 Safety of laser products - Part 1: Equipment classification and requirements
IEC 60601-1-2:2014, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility
Summary for clinical test
Clinical performance is not deemed necessary.
Conclusion
Based on the above performance as documented in this application, Intelligence Laser Comb (Model: YTLC001-W) was found to have a safety and effectiveness profile that is similar to the predicate devices. Thus, the subject device is substantially equivalent to the predicate devices.
N/A