K Number
K213674

Validate with FDA (Live)

Device Name
CellFX System
Date Cleared
2022-09-22

(304 days)

Product Code
Regulation Number
878.4400
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CellFX® System is indicated for dermatological procedures requiring ablation and resurfacing of the skin.

Specific Indication: The CellFX System with the CellFX Treatment Tip 2.5mm is indicated for the treatment of sebaceous hyperplasia (SH) in patients with Fitzpatrick skin types I-III.

Device Description

The Pulse Biosciences® CellFX® System is a proprietary platform technology. The CellFX System consists of the CellFX Console, CellFX Handpiece, CellFX Treatment Tips, and CellFX Software. The CellFX System delivers nanosecond duration electrical pulses that disrupt the function of cells leading to cell death, while sparing non-cellular tissue. The CellFX System delivers a series of timed, nanosecond electrical pulses (referred to as a "Cycle") to ablate and resurface tissue areas in dermatologic conditions.

The CellFX Console delivers short electric pulses at amplitudes up to 15 kV and pulse widths up to 700 ns. The electrical energy pulses are applied directly to targeted tissue using sterile Treatment Tips with stainless steel microneedles. The treatment parameters are selected by the user through a user interface on the Touchscreen Display of the CellFX Console.

AI/ML Overview

The provided text describes the CellFX System, an electrosurgical device for dermatological procedures. It outlines its specifications, compares it to a predicate device, and presents performance data from a clinical study to support its safety and effectiveness for treating sebaceous hyperplasia (SH).

Here's an analysis of the acceptance criteria and the study that proves the device meets them:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state "acceptance criteria" in a tabulated format. Instead, it presents co-primary safety and effectiveness endpoints from a clinical study, implying these are the measures against which the device's performance was judged. The study aimed to demonstrate non-inferiority of the CellFX System to Electrodessication.

Here's a table summarizing these endpoints and the reported performance:

Acceptance Criterion (Implied)MeasureCellFX System PerformanceElectrodessication Performance
Safety Endpoint 1: Non-inferiority regarding hyperpigmentation and scarring at 60 days post-treatment.Percentage of lesions showing hyperpigmentation or scarring.19.2%12.1%
Safety Endpoint 2: Non-inferiority regarding skin textural changes (crusting, other textural changes) at 30 days post-treatment.Percentage of lesions showing skin textural changes.6.8%10.3%
Effectiveness Endpoint: Non-inferiority regarding aesthetic improvement at 60 days post-last treatment.Percentage of lesions rated "improved" or "much improved" using the 5-point GAIS (scores of 4 or 5).76.9%75.5%

Interpretation of "Acceptance":
The document states: "Evaluation of primary safety and effectiveness endpoints showed CellFX to be noninferior to Electrodessication." This is the key statement indicating the device met its implied acceptance criteria. While the CellFX System had a slightly higher rate of hyperpigmentation/scarring (19.2% vs 12.1%), the study appears to have concluded non-inferiority based on statistical analysis not fully detailed in this excerpt. The other safety endpoint and the effectiveness endpoint showed comparable or numerically better performance for CellFX.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set:
    • Total subjects: 59
    • Total SH lesions treated with CellFX System: 121
    • Total SH lesions treated with Electrodessication: 107
  • Data Provenance:
    • The study was described as a "prospective, multi-center, randomized, comparison clinical study."
    • The document does not specify the country of origin of the data.
    • Given it's a prospective clinical study, the data is prospective.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • The ground truth for the safety and effectiveness endpoints was established by a "blinded Investigator."
  • The document refers to a single "blinded Investigator" (singular) assessing the endpoints for safety and a "blinded site investigator" (singular) for effectiveness. This phrasing suggests a single expert (per site, if multi-center) or a single central expert. However, in multi-center trials, typically multiple investigators at different sites are involved in assessments. Without further detail, it's hard to definitively state the number of distinct human experts involved across all sites who collectively established the ground truth.
  • Qualifications of Experts: The document identifies them as "Investigator" or "site investigator." It does not provide specific qualifications (e.g., years of experience, specific medical specialization like dermatology).

4. Adjudication Method for the Test Set

  • The document implies that the assessments were made by a "blinded Investigator" for safety endpoints and a "blinded site investigator" for the effectiveness endpoint. It does not describe any adjudication method (e.g., 2+1, 3+1, consensus panel) if there were multiple assessors or concerns about disagreement. It appears that the assessment of the single "blinded Investigator" served as the ground truth.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No, a "multi-reader multi-case (MRMC) comparative effectiveness study" involving human readers assisted by AI vs. without AI assistance was not done.
  • This study was a direct comparison between the CellFX System (device) and Electrodessication (active comparator), both applied by clinicians, and assessed by human investigators. It does not involve AI assistance for human readers.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

  • No, a standalone (algorithm only) performance study was not done.
  • The CellFX System is a physical electrosurgical device, not an AI algorithm. Its performance is intrinsically linked to its application by a human operator.

7. The Type of Ground Truth Used

  • The ground truth was based on expert assessment (by the "blinded Investigator" or "blinded site investigator") of clinical outcomes (pigmentation, scarring, textural changes, aesthetic improvement) at specified follow-up times (30 and 60 days). This could be categorized as clinical assessment/outcomes data derived from expert observation. It is not pathology or another objective measure described as such.

8. The Sample Size for the Training Set

  • The document describes a clinical study to support the safety and effectiveness of the device. This clinical study represents the "test set" or the primary data used for evaluating the device's performance against the specified endpoints.
  • The CellFX System is a hardware device delivering a specific energy type, not an AI/ML algorithm that requires a distinct "training set." Therefore, the concept of a separate "training set" for the device itself is not applicable in this context. The manufacturing and design process for such a device would involve various engineering and preclinical tests, but not a "training set" in the sense of machine learning.

9. How the Ground Truth for the Training Set Was Established

  • As explained above, there is no "training set" in the context of an AI/ML algorithm for which ground truth would need to be established. The "ground truth" (clinical outcomes) established by the blinded investigators in the described clinical trial serves as the basis for the evaluation of the device's performance, not for training it.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.

September 22, 2022

Pulse Biosciences, Inc. Punam Gollamudi Director US Regulatory Affairs 3957 Point Eden Way Hayward. California 94545

Re: K213674

Trade/Device Name: CellFX System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: GEI Dated: August 25, 2022 Received: August 26, 2022

Dear Punam Gollamudi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K213674

Device Name CellFX® System

Indications for Use (Describe)

The CellFX® System is indicated for dermatological procedures requiring ablation and resurfacing of the skin.

Specific Indication: The CellFX System with the CellFX Treatment Tip 2.5mm is indicated for the treatment of sebaceous hyperplasia (SH) in patients with Fitzpatrick skin types I-III.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments reqarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(K) SUMMARY

K213674

510(k) Summary

I. SUBMITTER

Applicant:Pulse Biosciences, Inc.3957 Point Eden WayHayward, CA 94545Phone: (510) 906-4649
Contact Person:Punam GollamudiDirector of US Regulatory AffairsPulse Biosciences, Inc.Phone: (415) 305-4404
Date Prepared:September 19, 2022

II. DEVICE INFORMATION

Trade Name:CellFX® System
Regulation Number21 CFR § 878.4400
Regulation Name:Electrosurgical Cutting and Coagulation Device andAccessories
Regulation Class:Class II
Product Code:GEI
Classification Panel:General and Plastic Surgery

III. PREDICATE DEVICE

K211444 - CellFX® System, Pulse Biosciences, Inc. No reference devices were used in this submission.

IV. DEVICE DESCRIPTION

The Pulse Biosciences® CellFX® System is a proprietary platform technology. The CellFX System consists of the CellFX Console, CellFX Handpiece, CellFX Treatment Tips, and CellFX Software. The CellFX System delivers nanosecond duration electrical pulses that disrupt the function of cells leading to cell death, while sparing non-cellular tissue. The CellFX System delivers a series of timed, nanosecond electrical pulses (referred to as a "Cycle") to ablate and resurface tissue areas in dermatologic conditions.

{4}------------------------------------------------

The CellFX Console delivers short electric pulses at amplitudes up to 15 kV and pulse widths up to 700 ns. The electrical energy pulses are applied directly to targeted tissue using sterile Treatment Tips with stainless steel microneedles. The treatment parameters are selected by the user through a user interface on the Touchscreen Display of the CellFX Console.

V. INDICATIONS FOR USE STATEMENT

The CellFX® System is indicated for dermatological procedures requiring ablation and resurfacing of the skin.

Specific Indication: The CellFX System with the CellFX Treatment Tip 2.5mm is indicated for the treatment of sebaceous hyperplasia (SH) in patients with Fitzpatrick skin types I-III.

VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

The subject device has the same technological characteristics as the predicate device. The device design, materials, fundamental scientific technology, materials and processes for packaging and sterilization have not been changed from the predicate device.

Subject DevicePredicate Device (K211444)
Trade NameCellFX® SystemCellFX® System
Prescription OnlyYesYes
RegulationNumber878.4400878.4400
Product CodeGEIGEI
Indications for Use
Indications for UseStatementThe CellFX System is indicatedfor dermatological proceduresrequiring ablation and resurfacingof the skin.Specific Indication: The CellFXSystem with the CellFX TreatmentTip 2.5mm is indicated for thetreatment of sebaceoushyperplasia (SH) in patients withFitzpatrick skin types I-III.The CellFX System is intended fordermatological proceduresrequiring ablation and resurfacingof the skin.
Subject DevicePredicate Device (K211444)
Technological Characteristics
Mechanism ofActionNano-Pulse Stimulation: ultrafastnanosecond electrical pulses to thetargeted tissue via treatment tipswith an array of microneedlesNano-Pulse Stimulation: ultrafastnanosecond electrical pulses to thetargeted tissue via treatment tipswith an array of microneedles
Pulse Frequency1 Hz -10 Hz1 Hz -10 Hz
PulseAmplitude/Voltage300V to 15kV300V to 15kV
Pulse Width100 ns – 700 ns100 ns – 700 ns
Power Input100-240 VAC, 50/60 Hz, 2A100-240 VAC, 50/60 Hz, 2A
Power Output30 Watts30 Watts
Energy DensityRange2.5 x 2.5 mm – 30-60 mJ/mm³1.5 x 1.5 mm – 110-190 mJ/mm³2.5 x 2.5 mm – 60-110 mJ/mm³5.0 x 5.0 mm – 45-85 mJ/mm³
Use of LocalAnesthesiaYesYes
Sterile TipEO SterilizationEO Sterilization
HandpieceSingle Handpiece for 2-Row TipsSingle Handpiece for 2-Row Tips
System Physical Characteristics
Height132 cm132 cm
Length46 cm46 cm
Width53 cm53 cm
Weight54 kg54 kg
Power Input100-240 VAC, 50/60 Hz, 2A100-240 VAC, 50/60 Hz, 2A
Tip Characteristics
Treatment Tip2-Row2-Row
MicroneedleConfiguration
1.5 x 1.5 mm2.5 x 2.5 mm5.0 x 5.0 mmN/A2 rows of 3 or 4 needles per rowN/A2 rows of 3 needles per row2 rows of 3 or 4 needles per row2 rows of 4 or 6 needles per row
Tip Insertion Depth1 mm and 2 mm1 mm and 2 mm

{5}------------------------------------------------

{6}------------------------------------------------

VII. PERFORMANCE DATA

The following performance data was referenced or provided in support of the substantial equivalence determination.

Biocompatibility Testing

Biocompatibility testing was not required as there were no changes to the device compared to the predicate.

Electrical safety and electromagnetic compatibility (EMC)

Electrical safety and EMC testing was not required as there were no changes to the device compared to the predicate.

Bench Testing

Performance testing was not required as there were no changes to the device compared to the predicate.

Software Verification and Validation Testing

Software verification and validation testing was not required as there were no changes to the device compared to the predicate.

Animal Study

Animal testing was not required to support the change to the Indications for Use.

Clinical Study

A prospective, multi-center, randomized, comparison clinical study was conducted to support safety and effectiveness of the CellFX System for the treatment of sebaceous hyperplasia (SH). A total of 59 subjects treated with the 2.5mm Tip were mostly female (79.7%) and white (91.5%). All treatments were performed at a depth of 2mm on subjects with Fitzpatrick skin types I-III. A total of 121 SH lesions were randomized to the CellFX System and 107 to Electrodessication in a split-face study design.

Safety Endpoints

The study was designed to evaluate co-primary safety and effectiveness endpoints. The first primary safety endpoint of lesions showing hyperpigmentation and scarring and assessed by the blinded Investigator at 60 days after the last treatment demonstrated CellFX to be non-inferior to Electrodessication. This safety endpoint found with 19.2% of lesions treated with CellFX and 12.1% of lesions treated with electrodessication having pigmentary changes or scarring. The second co-primary safety endpoint of skin textural changes (crusting, or other skin textural changes) as assessed by the blinded Investigator at 30 days after the last treatment showed rates of 6.8% of lesions treated with CellFX and 10.3% of lesions treated with Electrodessication.

{7}------------------------------------------------

Effectiveness Endpoint

The primary effectiveness endpoint of improvement at 60-days post-last treatment as assessed "live" by the blinded site investigator using the 5-point GAIS (1 = much worse, 2 = worse, 3 = no change, 4 = improved, 5 = much improved, with scores of 4 or 5 classified as responders) showed rates of aesthetic appearance to be improved or much improved for 76.9% of lesions treated with CellFX and 75.5% of lesions treated with electrodessication.

Adverse Events

There were 3 (2.5%, n=120) treatment-related non-serious adverse events (AEs) in 3 subjects with mild swelling at the lesion site treated with the CellFX 2.5mm treatment tip. None of the events required medical intervention and all resolved by the 7-day follow-up visit. One subject was instructed to use ice for swelling.

Summary

Evaluation of primary safety and effectiveness endpoints showed CellFX to be noninferior to Electrodessication. Adverse events associated with the procedure were relatively mild, transient, and not unexpected for this type of procedure, demonstrating that the risks associated with CellFX and procedure are minimal. There were no device-related adverse events and no unanticipated adverse events. The clinical study demonstrated that the CellFX System using the CellFX Treatment Tip 2.5mm was safe and effective for the treatment of sebaceous hyperplasia (SH).

CONCLUSION

The CellFX System has the same intended use, technological characteristics and principles of operation and similar indications for use as its predicate device. Clinical studies of the CellFX System have demonstrated the safety and effectiveness profile of the device in the intended population. Thus, the CellFX System is substantially equivalent to the predicate device.

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.