K Number
K213669

Validate with FDA (Live)

Manufacturer
Date Cleared
2023-02-15

(450 days)

Product Code
Regulation Number
878.4810
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The LightScalpel LS-4020 CO2 Laser Systems are intended for use in laser surgery procedures for incision, excision, vaporization, ablation, and/or coagulation of soft tissue in specialties such as: general surgery, dermatology, dentistry and oral surgery, otorhinolaryngology, plastic and reconstructive surgery, neurosurgery, podiatry, and urology.

Device Description

The LightScalpel LS-4020 laser system, similar to the previously released LS-1005 and LS-2010 laser systems are mobile platforms that utilizes a radio frequency (RF) excited carbon dioxide (CO2) laser tube to produce an infrared beam at a nominal 10.6 um wavelength at powers adjustable from 2 - 40 Watts Continuous Wave (CW) and 2 to 20 Watts SuperPulse (SP). The systems differ only in the laser tubes used and pre-programmed values in the controlling software, which allow higher laser tube outputs for the LS-2010 and LS-4020 systems. Laser energy is conducted to the point of application by a flexible fiber waveguide and handpiece assembly. Laser system operation is controlled by operator input on a touchscreen display panel. The RF laser drive is modulated to provide additional pulsed and SuperPulse emission modes selected from the laser system control panel.

AI/ML Overview

The provided document is a 510(k) summary for the LightScalpel LS-4020 CO2 Laser System. This type of submission is for demonstrating substantial equivalence to pre-existing legally marketed devices, rather than proving effectiveness through extensive clinical trials. Therefore, the information requested regarding acceptance criteria, extensive study details, and ground truth establishment, which are typical for novel devices requiring clinical validation, is largely not present in this document.

Here's an analysis based on the available information:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state "acceptance criteria" in the format of a table with specific pass/fail thresholds for clinical performance. Instead, the substantial equivalence is based on technological characteristics and intended use being comparable to predicate devices.

CharacteristicAcceptance Criteria (Implied as Comparable to Predicate)Reported Device Performance (LS-4020)
Laser MediumCO2CO2
Wavelength (μm)10.610.6
Laser Drive SourceRFRF
Output Power (W)Comparable to predicate (2-60W range)2-40
Pulsed Power (W)Comparable to predicate (2-60W range)2-40
Gated PW (ms)Comparable to predicate (5-1000ms range)5-500
Gated Rep. Rate (pps)Comparable to predicate (1-100pps range)1-50
Superpulse (W)Comparable to predicate (1-20W range)2-20
Superpulse Peak (W)Comparable to predicate (50-240W range)100
Superpulse Pulse DurationComparable to predicate (Up to 2ms)Up to 0.8 ms
Beam Delivery SystemFlexible Fiber Waveguide or Articulated ArmFlexible Fiber Waveguide
System HxWxD (in)Similar dimensions to predicate40x15x15
System Weight (lb)Similar weight to predicate66
Mobility4 Wheels & Handle4 Wheels & Handle
Intended UseIncision, Excision, Vaporization, Ablation, and/or Coagulation of Soft Tissue, across various specialtiesIncision, Excision, Vaporization, Ablation, and/or Coagulation of Soft Tissue, across various specialties
Line Voltage - Nom.100-240 VAC100-240 VAC

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: Not applicable in the context of this 510(k) submission. No clinical "test set" of patients or images was used for performance evaluation that is described in the document.
  • Data Provenance: Not applicable. The document states "Non-clinical performance data: Each LS-4020 laser system is tested for electrical safety and output characteristics." This implies testing on the device itself, not patient data.

3. Number of Experts Used to Establish Ground Truth and Qualifications

Not applicable. No ground truth establishment by experts is mentioned as this was not a clinical study requiring such a process.

4. Adjudication Method

Not applicable. No clinical study or expert review requiring adjudication is mentioned.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No. The document explicitly states: "Clinical performance data: None. Clinical testing was determined to be unnecessary, as the performance parameters are consistent with predicate CO2 laser systems and technology." Therefore, no MRMC study or assessment of AI assistance was performed or reported.

6. Standalone Performance Study

Yes, in a sense, a standalone technical performance was assessed. The document states: "Each LS-4020 laser system is tested for electrical safety and output characteristics. Representative data is presented in the Bench Testing section of this report." This refers to the device's technical specifications and safety performance, not its clinical efficacy in a standalone "algorithm only" context.

7. Type of Ground Truth Used

Not applicable for clinical performance. The "ground truth" for the device's substantial equivalence is the technical and operational characteristics of the predicate devices, and the LS-4020's ability to meet those non-clinical specifications.

8. Sample Size for the Training Set

Not applicable. This device is a laser system, not an AI/ML algorithm that requires a training set of data.

9. How the Ground Truth for the Training Set Was Established

Not applicable. As above, this is a physical device, not an AI/ML algorithm trained on data.

Summary of the Study Proving Substantial Equivalence:

The study proving the device meets the acceptance criteria is a technical and comparative analysis against legally marketed predicate devices, specifically the LightScalpel LS-1005 / LS-2010 CO₂ Laser Systems (K141658) and the Lumenis Ultrapulse™ (K151331).

The core of the submission relies on demonstrating that the LightScalpel LS-4020 has substantially equivalent technological characteristics and intended uses to these predicate devices. The document highlights that:

  • The LS-4020 uses the same laser medium (CO2), wavelength (10.6 μm), and laser drive source (RF) as the predicates.
  • Its output power, pulse settings, and superpulse parameters fall within the range of the predicates, or are otherwise comparable.
  • The beam delivery system (flexible fiber waveguide) and physical characteristics (size, weight, mobility) are also considered comparable.
  • Crucially, the Indications for Use for the LS-4020 are directly aligned with and appear to be largely identical to those of the predicate devices across numerous surgical specialties (General Surgery, Dermatology, Dentistry/Oral Surgery, ENT, Plastic/Reconstructive Surgery, Neurosurgery, Podiatry, Urology, Gynecology).

The submission explicitly states that no clinical performance data was deemed necessary because of this consistency with predicate CO2 laser systems and technology. Instead, non-clinical bench testing confirmed the electrical safety and output characteristics of each LS-4020 laser system. This approach is standard for 510(k) clearances where a new device is shown to be similar enough to existing ones that new clinical effectiveness and safety studies are not required.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the agency's name, "U.S. Food & Drug Administration."

February 15, 2023

LightScalpel Inc David Walters Chief Operating Officer 11818 North Creek Parkway N Suite 100 Bothell, Washington 98011

Re: K213669

Trade/Device Name: Lightscalpel Ls-4020 Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: November 1, 2022 Received: November 18, 2022

Dear David Walters:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Jianting Wang -S

Jianting Wang Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K213669

Device Name

LightScalpel® LS-4020 CO2 Laser System

Indications for Use (Describe)

The LightScalpel LS-4020 CO2 Laser Systems are intended for use in laser surgery procedures for incision, excision, vaporization, ablation, and/or coagulation of soft tissue in specialties such as: general surgery, dermatology, dentistry and oral surgery, otorhinolaryngology, plastic and reconstructive surgery, neurosurgery, podiatry, and urology.

Gynecology / Genitourinary:

Incision, excision, ablation, and/or vaporization of soft tissue for treatment of: Conization of the cervix, including cervical intraepithelial neoplasia and vulvar and vaginal intraepithelial neoplasia; Condyloma, including cervical, vulvar, perineal, and Bowenoid papulosa; leukoplakia (vulvar dystrophies); incision and drainage of Bartholin's and nabothian cysts; herpes vaporization; urethral caruncle vaporization; cervical dysplasia; benign and malignant tumors; hemangiomas; benign and malignant lesions of external genitalia; condyloma; phimosis; erythroplasia.

Dermatology / Plastic Surgery:

Incision, Excision, Ablation, and Vaporization of soft tissue for treatment of: laser skin resurfacing; laser dermabrasion: laser burn debridement; wrinkles, rhytids, and furrows; keratosis, including actinic and sebortheic keratosis, seborthoecae vulgares, sebortheica; vermillionectomy of the lip; cutaneous horns; solar/actinic elastosis; cheilitis, including actinic cheilitis; lentigines, including lentigon's malignant freckle; Uneven pigmentation/dyschromia; acne scars; surgical scars; keloids, including acne keloidalis nuchae; hemangiomas, including Buccal, port wine, and pyogenic granuloma, pyogenicum/granuloma telagiectaticum; tattoos; telangiectasias; removal of small skin tumors, including periungual (Koenen) and subungual fibromas; superficial pigmented lesions; adenosebaceous hyperplassa; rhinophyma reduction; cutaneous papilloma (skin tags); milia; debridement of eczematous or infected skin; basal & squamous cell carcinoma, including keratoacanthomas, Bowen's disease (Erythroplasia of Queyrat), and Bowenoid Papulosis (BP) lesions; nevi, including spider, epidermal, and protruding; neurofibromas; laser de-epitheliomas; xanthelasma palpebrarum; syringoma; complete or partial nairixectomy; benign/malignant vascular/avascular skin lesions; Mohs surgery; lipectomy; verrucae and seborrhoecae vulgares, including: paronychial, periungal, and subungual warts; blepharoplasty; and hair transplantation site preparation.

Dental Surgery:

Gingivectomy - Removal of hyperplasias; gingivoplasty; papillectomy; vestibuloplasty; epulis; sulcular debridement; removal of soft tissue, cysts, and fibroma (non-malignant turnor, mucosa, tongue); extraction site hemostasis; treatment of ulcerous lesions, including aphthous ulcers; a heat source to activate tooth bleaching materials; Laser Assisted New Attachment Procedure (cementum-mediated periodontal ligament no the root surface in the absence of long junctional epithelium).

Oral Surgery:

Frenum release/frenectomy; abscess (drainage); biopsy (incisional & excisional); aphthous ulcers (incision & excision); excision & ablation of lesions, benign & malignant lesions, oral cavity tumors, and hemangiomas; salivary gland pathologies; preprosthetic gum preparation, leukoplakia; partial glossectomy; periodontal gum resection; homeostasis; operculectomy; crown lengthening (soft tissue); incision when used with antibiotic therapy; extraction site hemostasis.

{3}------------------------------------------------

General and Thoracic Surgery:

Incision, excision, and vaporization of soft tissue in general and thoracic surgery, including endoscopic and open procedures for: debridement of decubitus ulcers, stasis, diabetic, and other ulcers; mastectomy; debridement of burns; rectal and anal hemorrhoidectomy; breast biopsy, reduction mammoplasty; cytoreduction for metastatic disease, laparotomy and laparoscopic applications; mediastinal and thoracic lesions and abnormalities; skin tag vaporization; atheroma; cysts, including sebaceous and pilar cysts, and mucous cysts of the lips; pilonidal cyst removal and repair; abscesses; other soft tissue applications.

Gyn Laparoscopic Surgery:

Laser incision, excision, vaporization, and photocoagulation of soft tissue for treatment of: endometrial lesions, including ablation for endometriosis; excisions; salpingotomy; oophorectomy/ovariectomy; fimbrioplasty; metroplasty; uterine myomas and fibroids; ovarian fibromas and follicle cysts; uterosacral ligament ablation; hysterectomy.

Otorhinolaryngology / ENT:

Laser incision, excision, ablation, vaporization, and / or photocoagulation of soft tissue in otorhinolaryngology for treatment of: leukoplakia, including oral, larynx, uvula, palatal, and upper lateral pharyngeal tissue; adult and juvenile papillomatosis polyps; lymphangioma removal of recurrent papillomas in the oral cavity, nasal cavity, larynx, pharynx, and trachea, including the uvula, palatal, upper lateral pharyngeal tissue, tongue, and vocal cords, stenosis, including subglottic stenosis; tonsiller tonsillar cryptolysis and neoplasma) and tonsil ablation / tonsillotomy; benign and malignant tumors and fibromas (oral); superficial lesions of the ear, including chondrodermatitis nondularis chronica helices / Winkler's disease; uvulopalatoplasty (LAUP, laser UPPP); turbinectomy and turbinate reduction / ablation; septal spur ablation and septoplasty; partial glossectomy; tumor resection of oral, subfacial, and neck tissues; rhinophyma; verrucae vulgares (warts); gingivoplasty / gingivectomy

Podiatry:

Laser excision, ablation, and /or vaporization of soft tissue in podiatry for the reduction, removal and/or treatment of: Verrucae vulgares / plantar warts, including paronychial, periungual, and subungual warts; fungal nail treatment; matrixectomy - partial and complete; porokeratoma ablation; neuromas/fibromas, including Morton's neuroma removal; ingrown toenail treatment; debridement of ulcers; treatment of other soft tissue lesions.

Orthopedic Surgery:

Laser incision / excision, and /or vaporization of soft tissue in orthopedic surgery. Applications include: meniscectorny; chondromalacia ablation; chondroplasty; ligament release (lateral and other); excision of plica; partial synovectomy; debridement of traumatic wounds; debridement of decubitus & diabetic ulcers; and PMMA removal.

Neurosurgery:

Laser incision / excision, ablation, and /or vaporization of soft tissue in neurosurgery for the treatment of: posterior fossa tumors; peripheral neurectomy; benign and malignant tumors and cysts (e.g. gliomas, meningiomas (including basal tumors), acoustic neuromas, lipomas, and large tumors); arteriovenous malformation; and pituitary gland tumors (transphenoidal approach).

Type of Use (Select one or both, as applicable)

✖ Prescription Use (Part 21 CFR 801 Subpart D)
❏ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{4}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{5}------------------------------------------------

lightScalpel®

510(k) Summary

K213669

510(k) Summary Preparation Date: 23 January 2023

    1. 510(K) Owner:
      LightScalpel® Inc. 11818 North Creek Pkwy N Suite 100 Bothell, WA 98011 866-697-7548 / 425-368-1588 / 425-368-1568 (FAX)
    1. 510(k) Contact:
      David Walters Chief Operating Officer 11818 North Creek Pkwy N Suite 100 Bothell, WA 98011 866-697-7548 / 425-368-1588 / 425-368-1568 (FAX) dwalters@lightscalpel.com
    1. Device Trade Name: LightScalpel® LS-4020
      Common Name: CO2 Laser System

Classification Name: Laser surgical instrument for use in general and plastic surgery and in dermatology (21 CFR 878.4810).

"A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide."

Classification: Class II

Product Code: GEX

    1. Predicate Device(s):
    • 4.1. LightScalpel LS-1005 / LS-2010 CO₂ Laser Systems; K141658
    • 4.2. Lumenis Ultrapulse™; K151331

{6}------------------------------------------------

Image /page/6/Picture/18 description: The image shows the logo for "light Scalpel". The word "light" is in bold, and the word "Scalpel" is in a regular font. A curved line starts at the top of the "l" in "light", goes over the word "Scalpel", and ends in a starburst below the word "light".

5. Device Description and Function

The LightScalpel LS-4020 laser system, similar to the previously released LS-1005 and LS-2010 laser systems are mobile platforms that utilizes a radio frequency (RF) excited carbon dioxide (CO2) laser tube to produce an infrared beam at a nominal 10.6 um wavelength at powers adjustable from 2 - 40 Watts Continuous Wave (CW) and 2 to 20 Watts SuperPulse (SP). The systems differ only in the laser tubes used and pre-programmed values in the controlling software, which allow higher laser tube outputs for the LS-2010 and LS-4020 systems.

The models of the LS family CO2 Laser systems are based on the size and/or quantity of the laser tube installed. The original LS-1005 model cleared on K123037 and K132661 utilizes a shorter (thus lower output) laser tube. The LS-1005 and LS-2010 models cleared on K141568 utilize a standard length all aluminum laser tube. The new LS-4020 utilizes two standard length tubes.

Laser energy is conducted to the point of application by a flexible fiber waveguide and handpiece assembly. Laser system operation is controlled by operator input on a touchscreen display panel. The RF laser drive is modulated to provide additional pulsed and SuperPulse emission modes selected from the laser system control panel.

A "calibration port" on the side of the laser systems allows checking and setting the power emitted from the distal laser aperture to the internal power meter and serves as a check on fiber waveguide transmission efficiency. Laser system power, rates, and durations are adjustable as tabulated below:

The laser systems have safety features complying with requirements in 21 CFR 1040, Performance Standards for Light Emitting Products.

Primary safety features are as follows:

System On-Off Key switch, Emergency Stop Switch, Remote Interlock, Fiber Interlock, Beam Blocking Shutter, Internal Laser Power Detector, RF Power Monitor, and required Laser Safety Labels and Labeling.

Laser system physical characteristics are:

Delivery System: Flexible Fiber Waveguide; ~ 0.75mm ID ; Handpieces with internal focusing lens ("tipless") and with disposable pre-sterilized ceramic tips.

Purge Gas: Internal air pump purge through the Fiber and Handpiece.

System Cooling: Air; two to four thermostatically controlled fans (depending on model) with over-temperature protection.

Mobility: 4 wheels and handgrip on console for convenient system positioning.

{7}------------------------------------------------

Image /page/7/Picture/12 description: The image shows the logo for "light Scalpel". The word "light" is in bold, and the "S" in "Scalpel" is stylized with a curved line extending from the top of the "S" over the word "light" and ending in a starburst symbol below the word "light". The rest of the word "Scalpel" is in a regular font.

6. Intended use(s) of the Device

The LightScalpel CO2 Laser Systems are intended for use in laser surgery procedures for incision, excision, vaporization, and/or coagulation of soft tissue in specialties such as: general surgery, dermatology, gynecology, dentistry and oral surgery, otorhinolaryngology, plastic and reconstructive surgery, orthopedic surgery, neurosurgery, podiatry, and urology.

Gynecology / Genitourinary:

Incision, excision, ablation, and/or vaporization of soft tissue for treatment of: Conization of the cervix, including cervical intraepithelial neoplasia and vaginal intraepithelial neoplasia: Condyloma, including cervical, genital, vulvar, perineal, and Bowenoid papulosa: leukoplakia (vulvar dystrophies); incision and drainage of Bartholin's and nabothian cysts; herpes vaporization; urethral caruncle vaporization; cervical dysplasia; benign and malignant tumors; hemanqiomas; benign and malignant lesions of external genitalia; condyloma; phimosis; erythoplasia.

Dermatology / Plastic Surgery:

Incision, Excision, Ablation, and Vaporization of soft tissue for performance of or treatment of: laser skin resurfacing; laser dermabrasion; laser burn debridement; wrinkles, rhytids, and furrows; keratosis, including actinic and seborrheic keratosis, seborrhoecae vulgares, seborrheci wart, and verruca seborrheica; vermillionectomy of the lip; cutaneous horns; solar/actinic elastosis; cheilitis, including actinic cheilitis; lentigines, including lentigo maligna or Hutchinson's malignant freckle; uneven pigmentation/dyschromia; acne scars; surgical scars; keloids, including acne keloidalis nuchae; hemangiomas, including Buccal, port wine, and pyogenic granuloma, pyogenicum/granuloma telagiectaticum; tatoos; telangiectasias; removal of small skin tumors, including periungual (Koenen) and subungual fibromas; superficial pigmented lesions; adenosebaceous hypertrophy or sebaceous hyperplasia; rhinophyma reduction; cutaneous papilloma (skin tags); milia; debridement of eczematous or infected skin; basal & squamous cell carcinoma, including keratoacanthomas, Bowen's disease (Erythroplasia of Queyrat), and Bowenoid Papulosis (BP) lesions: nevi, including spider, epidermal, and protruding; neurofibromas; laser deepitheliazation; tricoepitheliomas; xanthelasma palpebrarum; syringoma; complete or partial nail matrixectomy; benign/malignant vascular skin lesions; Mohs surgery; lipectomy; verrucae and seborrhoecae vulgares, including; paronychial, periungal, and subungual warts; blepharoplasty; and hair transplantation site preparation.

Dental Surgery:

Gingivectomy - Removal of hyperplasias; gingivoplasty; papillectomy; vestibuloplasty; epulis; sulcular debridement; removal of soft tissue, cysts, and fibroma (non-malignant tumor, mucosa, tongue); extraction site hemostasis; treatment of ulcerous lesions, including aphthous ulcers; a heat source to activate tooth bleaching materials; Laser Assisted New Attachment Procedure (cementum-mediated periodontal ligament new attachment to the root surface in the absence of long junctional epithelium).

{8}------------------------------------------------

lightScalpel®

Oral Surgery:

Frenum release/frenectomy; abscess (drainage); flap surgery; biopsy (incisional & excisional); aphthous ulcers (incision & excision & ablation of lesions, benign & malignant lesions, oral cavity tumors, and hemangiomas; salivary gland pathologies; preprosthetic gum preparation, leukoplakia; partial glossectomy; periodontal gum resection; homeostasis; operculectomy; crown lengthening (soft tissue); incision of infection when used with antibiotic therapy; extraction site hemostasis.

General and Thoracic Surgery:

Incision, excision, and vaporization of soft tissue in general and thoracic surgery, including endoscopic and open procedures for: debridement of decubitus ulcers, stasis, diabetic, and other ulcers: mastectomy: debridement of burns: rectal and anal hemorrhoidectomy: breast biopsy, reduction mammoplasty; cytoreduction for metastatic disease, laparotomy and laparoscopic applications.

midiastinal and thoracic lesions and abnormalities; skin tag vaporization; atheroma; cysts, including sebaceous and pilar cysts, and mucous cysts of the lips; pilonidal cyst removal and repair; abscesses; other soft tissue applications.

Gyn Laparoscopic Surgery:

Laser incision, excision, vaporization, and photocoagulation of soft tissue for treatment of: endometrial lesions, including ablation for endometriosis; excision/lysis of adhesions; salpingotomy; oophorectomy; fimbrioplasty; metroplasty; uterine myomas and fibroids; ovarian fibromas and follicle cysts; uterosacral ligament ablation; hysterectorny.

Otorhinolaryngology / ENT:

Laser incision, excision, ablation, vaporization, and / or photocoagulation of soft tissue in otorhinolaryngology for treatment of: leukoplakia, including oral, larynx, uvula, palatal, and upper lateral pharyngeal tissue; adult and juvenile papillomatosis polyps; lymphangioma removal: removal of recurrent papillomas in the oral cavity, nasal cavity, larvnx, and trachea, including the uvula, palatal, upper lateral pharyngeal tissue, tongue, and vocal cords; stenosis, including subglottic stenosis; tonsillectomy (including tonsilar cryptolysis and neoplasma) and tonsil ablation / tonsillotomy; benign and malignant tumors and fibromas (oral); superficial lesions of the ear, including chondrodermatitis nondularis chronica helices / Winkler's disease; uvulopalatoplasty (LAUP, laser UPPP); turbinectomy and turbinate reduction / ablation; septal spur ablation/reduction and septoplasty; partial glossectorny; tumor resection of oral, subfacial, and neck tissues; rhinophyma; verrucae vulgares (warts); gingivoplasty / gingivectomy

Podiatry:

Laser excision, ablation, and /or vaporization of soft tissue in podiatry for the reduction, removal and/or treatment of: Verrucae vulgares / plantar warts, including paronychial, periungal, and subunqual warts; fungal nail treatment; matrixectomy - partial and complete; porokeratoma ablation; neuromas/fibromas, including Morton's neuroma removal; ingrown toenail treatment; debridement of ulcers; treatment of other soft tissue lesions.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for "light Scalpel". The word "light" is in bold, and the word "Scalpel" is in a smaller, non-bold font. A curved line starts above the "l" in "light", goes over the word "Scalpel", and ends with a starburst symbol below the word "light".

Orthopedic Surgery:

Laser incision / excision, ablation, and /or vaporization of soft tissue in orthopedic surgery. Applications include: menisectomy; chondromalacia ablation; chondroplasty; ligament release (lateral and other); excision of plica; partial synovectomy; debridement of traumatic wounds; debridement of decubitus & diabetic ulcers; and PMMA removal.

Neurosurgery:

Laser incision / excision, ablation, and /or vaporization of soft tissue in neurosurgery for the treatment of: posterior fossa tumors; peripheral neurectomy; benign and malignant tumors and cysts (e.g. gliomas, meningiomas (including basal tumors), acoustic neuromas, lopomas, and large tumors); arteriovenous malformation; and pituitary gland tumors (transphenoidal approach).

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the logo for "light Scalpel". The word "light" is in bold, and the word "scalpel" is in a smaller, non-bold font. A curved line extends from the top right of the "l" in "light", curves over the word "scalpel", and ends in a starburst symbol below the word "light". The registered trademark symbol is located to the right of the word "scalpel".

    1. Technological Characteristics Comparison to Predicate Devices
      The technological characteristics comparison to predicate devices is summarized in the following table.
Predicate 510(k)LS-4020K213669LS-1005 /LS-2010K141658LumenisUltraPulse™K151331
Characteristic------
Laser MediumCO2CO2CO2
Wavelength (μm)10.610.610.6
Laser Drive SourceRFRFRF
Output Power (W)2-402-10 / 2-201 - 60
Pulsed Power (W)2-402-10 / 2-201 - 60
Gated PW (ms)5-5005-50050 - 1000
Gated Rep. Rate (pps)1 - 501 - 501 - 100
Superpulse (W)2-202-5 / 2-101-20
Superpulse Peak (W)10050240
Superpulse Pulse DurationUp to 0.8 msUp to 0.8 msUp to 2 ms
Beam Delivery SystemFlexible FiberWaveguideFlexible FiberWaveguideArticulatedArm or WG
System HxWxD (in)40x15x1540x15x1557x14x20
System Weight (lb)6647270
Mobility4 Wheels &Handle4 Wheels &Handle4 Wheels &Handle
Intended UseIncision,Excision,Vaporization,Ablation, and/orCoagulation ofSoft TissueIncision,Excision,Vaporization,Ablation,and/orCoagulation ofSoft TissueIncision,Excision,Vaporization,Ablation,and/orCoagulation ofSoft Tissue
Line Voltage - Nom.100-240 VAC100-240 VAC100-240 VAC

{11}------------------------------------------------

lightScalpel®

The LS4020 laser system is substantially equivalent to the predicate devices in emission wavelength, control parameters, relative output power, delivery accessories, and physical size.

    1. Non-clinical performance data: Each LS-4020 laser system is tested for electrical safety and output characteristics. Representative data is presented in the Bench Testing section of this report.
    1. Clinical performance data: None. Clinical testing was determined to be unnecessary, as the performance parameters are consistent with predicate CO2 laser systems and technology.
    1. In summary, the LightScalpel LS-4020 CO2 laser systems are equivalent to the predicate laser systems for the indicated uses in the stated medical specialties.

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.