K Number
K213652

Validate with FDA (Live)

Manufacturer
Date Cleared
2023-06-28

(586 days)

Product Code
Regulation Number
N/A
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Cresilon Hemostatic Gel™ (CHG™) is a hemostatic gel for external use only.

Cresilon Hemostatic Gel™ (CHG™) is indicated for the local management of bleeding wounds such as minor cuts, minor lacerations, and minor abrasions.

Device Description

Cresilon Hemostatic Gel™ (CHG™) is a hemostatic gel for external use only, indicated for the local management of bleeding wounds such as minor cuts, minor lacerations, and minor abrasions.

CHG's hemostatic gel is comprised of poly(N-acetyl-D-glucosamine, D-glucosamine), sodium alginate, and water. CHG is supplied as individually pouched, sterile, pre-filled, single-use syringes. Each syringe is a single 5 mL hemostatic gel application. CHG is packaged in a box containing two (2) CHG applications.

After removal from the pouch, the cap is unscrewed, and the syringe is primed, the hemostatic gel is topically applied directly to the source of bleeding via the syringe.

AI/ML Overview

The provided document is a 510(k) premarket notification from the FDA, specifically concerning the Cresilon Hemostatic Gel (CHG) device. It outlines the device's characteristics, intended use, and a comparison to a predicate device (Gel-E Flex) to demonstrate substantial equivalence.

However, the document does not contain information about:

  • A "study that proves the device meets the acceptance criteria" in the context of an AI/machine learning device.
  • "Acceptance criteria" as typically defined for the performance metrics of an AI/machine learning model (e.g., sensitivity, specificity, AUC).
  • Details on a test set (sample size, provenance), expert ground truth establishment (number of experts, qualifications, adjudication), MRMC studies, standalone algorithm performance, or training set details.

This document pertains to a medical device (a hemostatic gel), not an AI/machine learning product. Therefore, the requested information points (1-9) are largely irrelevant to the content of this FDA 510(k) submission. Medical devices like this gel are typically evaluated based on biocompatibility, performance bench testing (e.g., in vivo animal efficacy), sterilization, and shelf-life, as detailed in the "Non-clinical Testing" section.

Based on the provided text, it is not possible to answer the questions about AI/ML device acceptance criteria and study details. The document is for a physical medical device (hemostatic gel), not a software or AI-powered diagnostic/therapeutic tool.

If the request implies searching for acceptance criteria and study data within the context of a traditional medical device rather than an AI/ML device, then the relevant information from the document would be as follows:

Acceptance Criteria and Device Performance (for a Traditional Medical Device):

The "acceptance criteria" for this traditional medical device are primarily related to its safety and functional performance as observed in non-clinical (biocompatibility and animal efficacy) testing, demonstrating it is "as safe, as effective, and performs as well as the predicate device."

Acceptance Criteria (Biological Endpoint/Test Result)Reported Device Performance (Test Result)
Non-cytotoxicPass
Non-irritating and non-sensitizingPass
Non-pyrogenicPass
Non-toxic (Acute Systemic Toxicity)Pass
Non-hemolyticPass
Functional as intended (In vivo animal efficacy)Functioned as intended; performance as expected

Details of the Study (Non-Clinical Testing for a Traditional Medical Device):

The document describes non-clinical testing, but not in the format of an AI/ML study.

  1. A table of acceptance criteria and the reported device performance: See table above.
  2. Sample size used for the test set and the data provenance:
    • Biocompatibility Testing: Not specified as a "test set" in terms of number of samples, but tests were performed on "representative samples." Data provenance is not explicitly stated beyond being conducted per ISO 10993-1:2009.
    • In Vivo Animal Efficacy Testing: Conducted in a "porcine model of skin laceration." The number of animals or specific "sample size" is not provided. Data provenance is implied to be from this specific animal model, likely controlled lab conditions. "Retrospective or prospective" is not applicable in the AI/ML sense but would generally be a prospective animal study.
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for a physical device's performance in animal models is typically objective (e.g., cessation of bleeding, lack of adverse reactions) or determined by veterinary professionals, not "experts" in the context of radiological interpretation.
  4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
  5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable, as this is not an AI device.
  6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
  7. The type of ground truth used:
    • Biocompatibility: Compliant with ISO 10993-1:2009 standards for biological endpoints (cytotoxicity, irritation, sensitization, pyrogenicity, acute systemic toxicity, hemolysis).
    • In Vivo Animal Efficacy: Demonstrated functional performance (hemostatic effect) in a porcine laceration model. The "ground truth" is whether the device stopped bleeding as intended.
  8. The sample size for the training set: Not applicable, as this is not an AI device that requires a training set.
  9. How the ground truth for the training set was established: Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

June 28, 2023

Cresilon, Inc. Hassaan Ahmad Vice President - Quality & Regulatory Affairs 87 35th Street. Suite 603/604 Brooklyn, New York 11232

Re: K213652/S003 Trade/Device Name: Cresilon Hemostatic Gel, CHG Regulatory Class: Unclassified Product Code: FRO Dated: September 1, 2022 Received: September 2, 2022

Dear Hassaan Ahmad:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

{1}------------------------------------------------

https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

David-Krause -S

David Krause, Ph.D. Deputy Director OHT4. Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) To be assigned by FDA

Device Name Cresilon Hemostatic Gel™ (CHGTM), Cresilon Hemostatic Gel™, CHGTM.

Indications for Use (Describe)

Cresilon Hemostatic Gel™ (CHGTM) is a hemostatic gel for external use only.

Cresilon Hemostatic Gel™ (CHG™) is indicated for the local management of bleeding wounds such as minor cuts, minor lacerations, and minor abrasions.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

1. Submitter Information:

Sponsor and ApplicationCorrespondent (US Agent):Hassaan W. AhmadVice President – Quality & Regulatory Affairs
87 35th Street, Suite 603/604,
Brooklyn, NY 11232
Phone347 435 2226 x103
E-mail:hassaan@cresilon.com
Legal Manufacturer:Cresilon, Inc.
Phone:347 435 2226 x103
Contact Person:Hassaan W. Ahmad
E-mail:hassaan@cresilon.com
Date Prepared:18th Nov 2021

2. Device Identification:

Device Trade Name:
------------------------
CHG TM
Product Code Description:Dressing, Wound, Drug
Device Classification:Unclassified Device (pre-amendment
Review Panel:General & Plastic Surgery
Product Code:FRO
Regulation NumberNot Applicable

3. Predicate Device:

Cresilon Hemostatic Gel™ (CHGTM)

Cresilon Hemostatic Gel™

Device Name510(k) Number
Gel-E Flex Manufactured by Gel-E, Inc. (Now registeredas Medcura, Inc)K180152

4. Device Description

Cresilon Hemostatic Gel™ (CHG™) is a hemostatic gel for external use only, indicated for the local management of bleeding wounds such as minor cuts, minor lacerations, and minor abrasions.

CHG's hemostatic gel is comprised of poly(N-acetyl-D-glucosamine, D-glucosamine), sodium alginate, and water. CHG is supplied as individually pouched, sterile, pre-filled, single-use syringes. Each syringe is a single 5 mL hemostatic gel application. CHG is packaged in a box containing two (2) CHG applications.

After removal from the pouch, the cap is unscrewed, and the syringe is primed, the hemostatic gel is topically applied directly to the source of bleeding via the syringe.

{4}------------------------------------------------

5. Intended Use & Indications for Use

Intended Use: Cresilon Hemostatic Gel™ (CHG™) is a hemostatic gel for external use only.

Indications for Use: Cresilon Hemostatic Gel™ (CHG™) is indicated for the local management of bleeding wounds such as minor cuts, minor lacerations, and minor abrasions.

6. Substantial Equivalence

The subject device Cresilon Hemostatic Gel™ (CHG™) is substantially equivalent to the predicate device Gel-E Flex cleared under the 510(k) number, K180152. The intended use and the indications for use of the subject device are same as that of the predicate and the technological characteristics such as, device design, physical state, mechanism of action and application of the device of the subject device are substantially equivalent to that of the predicate device Gel-E Flex.

Thus, CHG does not give rise to any new safety nor performance questions when compared with Gel-E Flex.

Table 2 – List of Predicate Devices
Device Name510(k) Number
PredicateGel-E FlexK180152
Device Name510(k) Number
PredicateGel-E FlexK180152
ParametersSubject DevicePredicate DeviceComments
NameCresilon Hemostatic GelTM(CHGTM)Gel-E FlexN/A
510(k) numberN/A-To be assigned by FDAK180152N/A
Intended UseCresilon Hemostatic GelTM(CHGTM) is a hemostatic gelfor External use only.The device is a hemostaticgel for External use only.Same
Indications foruseCresilon Hemostatic GelTM(CHGTM) is indicated for thelocal management ofbleeding wounds such asminor cuts, minorlacerations, and minorabrasionsGel-e Flex is indicated forthe local management ofbleeding wounds such asminor cuts, minorlacerations and minorabrasions.Same
Part of thebody to beinteracted withInjured or breached skinInjured or breached skinSame
Single UseYesYesSame
Rx/OTCRxOTCCresilon intends to marketCHGTM for prescription use (Rx).As CHG is intended for use bytrained health care providers,CHG is equivalent to thepredicate, as use error isminimized

Table 3- Substantial Equivalence (Intended Use & Indications for Use)

{5}------------------------------------------------

Cresilon, Inc.

Traditional 510(k) Cresilon Hemostatic Gel™ (CHGTM)

Sr.NoParametersSubject devicePredicate DeviceComments
1.NameCresilon Hemostatic GelTM (CHGTM)Gel-E Flex (Gel)N/A
2.510(k) NumberN/AK180152N/A
3.ManufacturerCresilon Inc.Gel-E, Inc. (Now registered with FDA asMedcura, Inc.)N/A
4.Product CodeFROFROSame as predicate.
5.Regulation NumberUnclassifiedUnclassifiedSame as predicate.
6.Device DescriptionCresilon Hemostatic GelTM (CHGTM) isa gel, composed of poly(N-acetyl-D-glucosamine, D-glucosamine)(Chitosan), sodium alginate, and water.Gel-e Flex is composed of a gel ofpalmitoyl-N-acetylglucosamine (chitosan),dissolved in 0.1M lactic acid in water.Both products have chitosan.Both products are delivered as viscousgels from pre-filled syringes.The animal efficacy testing data andthe biocompatibility testing data donot raise additional questions aboutsafety or efficacy of the subjectdevice.
7.DeviceDesign/TechnologyViscous Gel in Pre-Filled SyringeViscous Gel in Pre-Filled SyringeSame as predicate
8.Volume5 mL syringe5 mL or 10 mL syringes.Same as predicate
9.Sterilization10-6 SAL – Terminally sterilized withgamma radiation10-6 SAL – Terminally sterilized withgamma radiationSame as predicate
10.Mechanism of ActionWhen directly applied to a source ofbleeding, the hemostatic gel rapidlyadheres to the wound site. Thehemostatic gel forms a mechanicalbarrier that stops the flow of bleedingand allows the body to create a naturalclot.Same as subject deviceSame as Predicate
11.Bench TestingBench testing: pH, ViscosityAnimal efficacy testingBench testing: pH, viscosityAnimal efficacy testingSame as predicate.
12.BiocompatibilityCytotoxicityCytotoxicitySimilar
Table 4- Substantial Equivalence (Technological Characteristics)
----------------------------------------------------------------------

{6}------------------------------------------------

Traditional 510(k) Cresilon Hemostatic Gel™ (CHGTM)

Cresilon, Inc.

Sr.NoParametersSubject devicePredicate DeviceComments
SensitizationSensitization
Irritation/Irritation
PyrogenicityPyrogenicity
Systemic ToxicitySystemic toxicity
Hemolysis
13.PackagingBubble TestingSeal Strength TestingDye Penetration TestingBurst Pressure TestingDye Penetration testingSimilar

{7}------------------------------------------------

Traditional 510(k) Cresilon Hemostatic Gel™ (CHGTM)

7. Non-clinical Testing:

The Subject Device has been evaluated through a series of nonclinical studies to determine whether the device meets the acceptance criteria for its intended applications. All the nonclinical tests conducted on the device are summarized below.

a) Biocompatibility Testing

Biocompatibility tests have been performed per the requirements of ISO 10993-1:2009, under the section "Surface devices used on breached or compromised surface with limited contact duration (≤24 hrs) ". The subject device complies with all the tests conducted and complies to the following standards identified in the below table.

BiologicalendpointTest MethodPurposeAcceptancecriteriaTest Result
CytotoxicityISO 10993-5To verify Cytotoxicity potential ofthe subject deviceNon-cytotoxicPass
IrritationandSensitizationISO 10993-10To verify irritation andsensitization potential of thesubject deviceNon-irritatingand non-sensitizingPass
PyrogenicityISO 10993-11To verify the pyrogenicity of thedevice.Non-pyrogenicPass
AcuteSystemicToxicityISO 10993-11:Evaluation of the potential formedical device materials to causeadverse systemic reactions.Non-toxicPass
HemolysisASTM F756, ISO10993-4To verify the hemolytic property ofthe device.Non-hemolyticPass

Table 5 – Summary of Biocompatibility Testing performed

b) Performance Bench Testing

As a part of design verification studies, representative samples of the device underwent testing including packaging validation testing (Bubble Testing, Seal Strength Testing, Dye Penetration Testing and Removal Torque Testing) and in vivo animal efficacy testing.

In vivo animal efficacy testing was conducted in a porcine model of skin laceration to evaluate both the predicate device and Cresilon Hemostatic Gel™ (CHGTM). The porcine model was used due to the vast similarities between pigs and humans when it comes to dermal wound lacerations. Both the predicate and subject devices operate by the same mechanism of action. In all instances, Cresilon Hemostatic GelTM (CHGTM) functioned as intended, device performance was as expected.

8. Sterilization and Shelf Life:

Cresilon Hemostatic Gel™ (CHG™) is terminally sterilized using gamma irradiation to a Sterility Assurance Level (SAL) of 10-6.

Cresilon Hemostatic Gel™ (CHG™) is subjected to shelf-life testing to evaluate the shelflife of the product for in vivo efficacy, container closure integrity, and deployment force.

{8}------------------------------------------------

9. Conclusion:

The intended use and the indications for use of the subject device, Cresilon Hemostatic Gel™, are the same as that of the predicate. The technological characteristics such as device design, physical state, mechanism of action, and application of the subject device are the same as that of the predicate device Gel-E Flex. The nonclinical test data further demonstrates that the subject device is as safe, as effective, and performs as well as the predicate device. Based on the comparison above, the proposed device is determined to be Substantially Equivalent (SE) to the predicate device.

N/A