K Number
K213520

Validate with FDA (Live)

Device Name
AXIR-CX
Manufacturer
Date Cleared
2022-08-19

(289 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

AXIR-CX is a software package used with general purpose computing hardware to receive, store, distribute and display chest X-ray images and associated data for patient diagnosis. AXIR-CX is a software application that enable the DICOM-compliant chest X-ray image [14 x 17 or 17 x 17 inch size] from DR and CR, and after image displaying the user adds the annotation regarding the diagnosis and print out the patient information or send to another PACS system. AXIR-CX is intended to be used by trained medical professionals including physicians, radiologists, and medical technicians. This device is not indicated for use in mammography.

Device Description

The AXIR software is designed for use by radiologists and radiology technicians for annotation in the Chest X-ray images. The AXIR software is developed to use Radisen Flat Panel DR Detector and Radisen Image Viewer. The purpose of AXIR software is for the doctor to annotate Chest X-ray images and then to print out with patient information or sent to another PACS system.

A client user needs to install AXIR-CX first in the recommended PC environment. After installation, the client user chooses a DICOM format in the uploaded patient list to be annotated, and then annotation is written by user after reviewing of image chosen. After annotation has completed it can be printed out, saved or sent to another PACS system.

AI/ML Overview

This submission, K213520 for AXIR-CX, does not contain the information requested regarding acceptance criteria and performance studies that prove the device meets those criteria.

The document is a 510(k) summary, which focuses on demonstrating substantial equivalence to a predicate device (ExamVue PACS, K162868). It highlights similarities and differences in intended use, performance standards, operating system requirements, and specific functionalities like image archiving, display, patient search, and annotation.

Crucially, this document does not describe a study that validates the performance of AXIR-CX against predefined acceptance criteria for AI/ML-driven diagnostic tasks. The "Performance Standard" mentioned (21 CFR 892.2050) refers to the regulation for Medical image management and processing systems, not specific performance metrics for AI. The "Safety, EMC and Performance Data" section only states that "Safety testing and documentation was performed in accordance with IEEE 1012-2012, Standard for System and Software Verification and Validation," which is a general software engineering standard and does not provide details on specific clinical or AI performance evaluation.

Therefore, I cannot extract the requested information from this document. The device, AXIR-CX, appears to be an image management and processing system (PACS-like) with annotation capabilities, rather than an AI-driven diagnostic tool that would require the rigorous performance study details you've outlined. The differences noted, such as the absence of specific measurement tools, window/level adjustment, and zoom/magnify functions in AXIR-CX compared to the predicate, further suggest it's a foundational image display and management system.

{0}------------------------------------------------

August 19, 2022

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a seal with an abstract design, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The text is arranged in three lines, with "FDA" in a larger font and enclosed in a blue square.

Radisen Co., Ltd. % Dave Kim Medical Device Regulatory Affairs Mtech Group 7505 Fannin St. , Suite 610 HOUSTON TX 77054

Re: K213520

Trade/Device Name: AXIR-CX Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: LLZ Dated: July 20, 2022 Received: July 20, 2022

Dear Dave Kim:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR

{1}------------------------------------------------

  1. for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For

Jessica Lamb, Ph.D. Assistant Director Imaging Software Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K213520

Device Name AXIR-CX

Indications for Use (Describe)

AXIR-CX is a software package used with general purpose computing hardware to receive, store, distribute and display chest X-ray images and associated data for patient diagnosis.

AXIR-CX is a software application that enable the DICOM-compliant chest X-ray image [14 x 17 or 17 x 17 inch size] from DR and CR, and after image displaying the user adds the annotation regarding the diagnosis and print out the patient information or send to another PACS system. AXIR-CX is intended to be used by trained medical professionals including physicians, radiologists, and medical technicians. This device is not indicated for use in mammography.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

__ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

1. Traditional 510(k) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21 CFR Part 807.92.

Date 510K summary prepared: August 18, 2022

Submitter's Name:Radisen Co., Ltd.
Submitter's Address:B-602, Hifield Building, 66, Beolmal-ro, Dongan-gu, Anyang-si,Gyeonggi-do, Republic of Korea, 14058
Submitter's Telephone:Tel:+82-31-8084-9762
Contact person:Mr. John Lim / EVP of DR Business Unit
Official Correspondent:Dave Kim (davekim@mtech-inc.net)
Address:7505 Fannin St. Ste 610, Houston, TX 77054
Telephone:+713-467-2607

Name of the device, including the trade or proprietary name if applicable, the common or usual name and the classification name, if known:

Trade/proprietary name:AXIR-CX
Regulation Name:Medical image management and processing system
Regulation Number:21 CFR 892.2050
Regulatory Class:II
Product Code:LLZ
Predicate Device
Manufacturer:JPI Healthcare Co., Ltd
Device:ExamVue PACS
510(k) Number:K162868
Classification Name:Imaging Processing System, Radiological
Common Name:Picture Archiving and Communication System (PACS)
Regulatory Number:21 CFR 892. 2050
Regulatory Class:II
Product Code:LLZ

2. Device Description

The AXIR software is designed for use by radiologists and radiology technicians for annotation in the Chest X-ray images. The AXIR software is developed to use Radisen Flat Panel DR Detector and Radisen Image Viewer. The purpose of AXIR software is for the doctor to annotate Chest X-ray images and then to print out with patient information or sent to another PACS system.

{4}------------------------------------------------

A client user needs to install AXIR-CX first in the recommended PC environment. After installation, the client user chooses a DICOM format in the uploaded patient list to be annotated, and then annotation is written by user after reviewing of image chosen. After annotation has completed it can be printed out, saved or sent to another PACS system.

3. Indications for Use

AXIR-CX is a software package used with general purpose computing hardware to receive, store. distribute and display chest X-ray images and associated data for patient diagnosis. AXIR-CX is a software application that enable the DICOM-compliant chest X-ray image [14 x 17 or 17 x 17 inch size] from DR and CR, and after image displaying the user adds the annotation regarding the diagnosis and print out the patient information or send to another PACS system. AXIR-CX is intended to be used by trained medical professionals including physicians, radiologists, and medical technicians. This device is not indicated for use in mammography.

4. Summary of Design Control Risk management

After the analysis of risk management, there are no more risk factors and no more actions neede deither. Risk management, during the life cycle from product planning, design process to followup management, identified foreseeable risks in accordance with EN ISO 14971:2012, EN 62304:2006+AC:2008, EN 62366:2008 and EN ISO 13485:2016/AC2016 standards. The validation of risk management was performed by the radiologist who had the experience in clinical field according to the risk management plan. Each item of the validation was checked, and the system had no risks.

5. Comparison with predicate device:

Radisen Co., Ltd, believes that AXIR-CX is substantially equivalent to the predicate device, ExamVue PACS.

Both the subject and predicate devices have the same basic structure (a central server database and associated viewers), function (the storage, display of DICOM images) and follow the DICOM They have similar intended uses and provide similar suites of tools to fulfil their protocol. function.

AXIR-CX differs from the predicate device is user interface and compatible operating system. We believe this does not represent a substantial difference between the two devices, as the change in system requirements reflect the change in computer technology since the release of the predicate device, and the user interface presents the same essential data and supports similar workflow as the predicate device.

{5}------------------------------------------------

6. Substantial Equivalence

CharacteristicProposedAXIR-CXPredicate DeviceExamVue PACSRemark
ManufacturerRadisen Co., LtdJPI Healthcare Co., Ltd
510(k) numberK213520K162868
Intended UseAXIR-CX is a softwareapplication that enable theDICOM-compliant chest X-rayimage [14 x 17 or 17 x 17 inchsize] from DR and CR, and afterimage displaying the user addsthe annotation regarding thediagnosis and print out thepatient information or send toanother PACS system. AXIR-CXis intended to be used bytrained medical professionalsincluding physicians,radiologists, and medicaltechnicians. This device is notindicated for use inmammography.ExamVue PACS is an imagemanagement system intendedto be used by trainedprofessionals, includingphysicians, radiologists, nursesand medical technicians.The software is a softwarepackage used with generalpurpose computing hardwareto receive, store, distribute,process and display imagesand associated data throughouta clinical environment. Thesoftware performs digital imageprocessing, measurement,communication and storage.This device is not indicated foruse in mammography.ExamVue PACS supportsreceiving, sending, printing,storing and displaying studiesreceived from the followingmodality types via DICOM: CRand DX.Similarity
PerformanceStandard21 CFR 892.205021 CFR 892.2050Same
OperatingSystemRequirementWindow 10 andbasedWindowoperating systemWindow 7 orWindow 8 orWindow 10Similarity
ImagingArchiveYes(DICOM 3.0 Standard)Yes(DICOM 3.0 Standard)Same
Image displayYesYesSame
Patient SearchYesYesSame
DistanceandAngleMeasurementNoYesDifference
WindowLevelAdjustmentNoYesDifference
ZoomandMagnifyFunctionNoYesDifference
Line Profile andHistogramNoYesDifference
DICOMDirectoryReadingYesYesSame
DICOMQuery/RetrieveYesYesSame
DICOM ImportNoYesSame
DICOM CD burnNoYesDifference
AnnotationYesYesSame
DICOM PrintYesYesSame
DICOMTagDisplayNoYesDifference
PatientInformationEditingNoYesDifference

{6}------------------------------------------------

There is no significant difference between AXIR-CX and the predicate device that would adversely affect the use of the product. It is substantially equivalent to these devices in design, function and operational principles and intended use.

7. Safety, EMC and Performance Data

Safety testing and documentation was performed in accordance with IEEE 1012-2012, Standard for System and Software Verification and Validation. [AXIR-CX-SVR]

8. Conclusions

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification Radisen Co., Ltd, concludes that AXIR-CX is safe and effective and substantially equivalent to predicate device as described herein.

§ 892.2050 Medical image management and processing system.

(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).