K Number
K213462

Validate with FDA (Live)

Manufacturer
Date Cleared
2022-02-11

(107 days)

Product Code
Regulation Number
892.1720
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

EzRay M18 (Model: VMX-P400) is a portable X-ray system, intended for use by a qualified/trained physician or technician to acquire X-ray images of the desired parts of a patient's anatomy on adult and pediatric patients (including head, chest, abdomen, cervical spine, and extremities).

The device may be used for handheld diagnostic imaging of body extremities.

The system is subject to the following limitations of use when stand-mounted:

  • The device may be used for diagnostic imaging of the head, abdomen, cervical spine, or extremities.

  • The device may be used for imaging of the chest when used without a grid.

This device is not intended for mammography.

Device Description

EzRay M18 (Model: VMX-P400), a medical portable X-ray system, operates on 57.6 Vdc supplied by a rechargeable Li-ion battery pack. The system is composed of an X-ray generating part including an X-ray tube, a device controller, a power controller, a user interface, a beam limiting part, and two optional components: remote controller and stand. The device software supports the EzRay M18 system, and the software is of Moderate level of concern. The device is designed for the human body using image receptors. The image detectors, a necessary component for a fully-functional x-ray system, are not part of this submission.

AI/ML Overview

The provided document is a 510(k) premarket notification for the EzRay M18 (Model: VMX-P400) mobile X-ray system. It establishes substantial equivalence by comparing the device to a predicate device (MinXray, Inc. TR90BH, K182207) and a reference device (OSKO, Inc. XR5, K150663).

However, the document does not contain information about an AI/algorithm component. Instead, it focuses on the hardware and its compliance with established safety and performance standards for X-ray systems. Therefore, many of the requested details related to AI performance, ground truth establishment, expert adjudication, and MRMC studies are not applicable or cannot be extracted from this document.

Here's the information that can be extracted, based on the non-clinical testing performed for this traditional 510(k) submission:

Acceptance Criteria and Reported Device Performance (Non-AI Device)

This device is an X-ray system, not an AI-powered diagnostic tool. The acceptance criteria for such a device primarily revolve around safety, electrical performance, and image quality characteristics, rather than diagnostic accuracy metrics like sensitivity or specificity for a specific condition.

Acceptance Criteria CategoryReported Device Performance (as described in the document)
Safety and Electrical Performance- Complies with IEC 60601-1:2005 (Medical electrical equipment - General requirements for basic safety and essential performance). - Complies with IEC 60601-1-2:2014 (Electromagnetic compatibility - Requirements and tests). - Complies with IEC 60601-2-28:2017 (Medical Electrical Equipment - Exposure conditions - Diagnostic X-ray equipment). - Complies with IEC 60601-2-54:2009, AMD1:2015, AMD2:2018 (X-ray equipment for radiography and radioscopy). - Complies with IEC 62133-2:2017 (Secondary cells and batteries containing alkaline or other non-acid electrolytes – Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications). - Conforms to 21 CFR 1020 Subchapter J (Performance Standards for Ionizing Radiation Emitting Products), specifically 21 CFR 1020.30 (Diagnostic x-ray system) and 21 CFR 1020.31 (Radiographic Equipment).
Image Quality Performance Parameters- MTF (Modulation Transfer Function) results compared with reference device (XR5, K150663) and found equivalent. - Spatial Frequency results compared with reference device and found equivalent. - DQE (Detective Quantum Efficiency) results compared with reference device and found equivalent. - NPS (Noise Power Spectrum) results compared with reference device and found equivalent.
Functional Equivalence- Capable of setting higher mA (20 mA vs 15 mA max for predicate) under same maximum tube voltage (90kV), allowing capture of radiographic images for the same body parts as the predicate device. - User interface (soft touch push buttons) and collimator (continuously adjustable light beam type) are similar to the predicate. - Energy source (rechargeable Li-ion battery pack) is the same as the predicate.
Intended Indications for Use Adherence- Intended for use to acquire X-ray images of desired parts of a patient's anatomy on adult and pediatric patients (head, chest, abdomen, cervical spine, extremities). - May be used for handheld diagnostic imaging of body extremities. - Stand-mounted use limitations: diagnostic imaging of head, abdomen, cervical spine, or extremities; chest imaging without a grid. - Not intended for mammography. (All align with predicate and intended use).

Study Details (Relevant to this Traditional 510(k) X-ray System Submission)

  1. Sample Size used for the test set and data provenance:

    • For image performance testing, the document states: "Image performance testing was performed on the EzRay M18 in comparison with a reference device XR5 (K150663) using a FDA-cleared flat-panel detector, 1417WCC (K171418)."
    • The document does not specify the sample size (e.g., number of images, phantoms, or subjects) used for this image performance testing.
    • The data provenance (country of origin, retrospective/prospective) is not mentioned. These tests are typically bench tests using phantoms or controlled images, not patient data in the context of an X-ray machine clearance.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This is not applicable as the study described is a technical performance comparison of an X-ray machine's image characteristics (MTF, DQE, etc.) with a reference device, rather than a diagnostic accuracy study requiring expert human reads for ground truth. Ground truth for these technical measurements is derived from the physical properties of the test phantoms or from established engineering principles and measurements.
  3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not applicable. There's no human reader interpretation or adjudication described, as the study focuses on the physical image output quality metrics.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This submission is for a mobile X-ray system hardware, not an AI diagnostic algorithm. Therefore, no MRMC study, AI assistance, or human reader improvement associated with AI is discussed or performed.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. There is no AI algorithm being submitted. The device is purely an imaging acquisition system.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For the image performance testing, the "ground truth" for metrics like MTF, DQE, spatial frequency, and NPS would be derived from the physical and mathematical properties of the test phantoms or technical measurement standards used in a controlled laboratory setting. It is essentially comparing objective technical measurements against those of a known, legally marketed device.
  7. The sample size for the training set:

    • Not applicable. This device does not involve a "training set" as it is a hardware device, not a machine learning model.
  8. How the ground truth for the training set was established:

    • Not applicable. As above, there is no training set mentioned or implied for this device.

{0}------------------------------------------------

February 11, 2022

Image /page/0/Picture/1 description: The image shows the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the FDA acronym in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.

Vatech Co., Ltd. % Mr. Dave Kim Contact Title Medical Device Regulatory Affairs Mtech Group 7707 Fannin St., Ste. 200-V111 HOUSTON TX 77054

Re: K213462

Trade/Device Name: EzRay M18 (Model: VMX-P400) Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile x-ray system Regulatory Class: Class II Product Code: IZL Dated: January 7, 2022 Received: January 13, 2022

Dear Mr. Kim:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Laurel Burk, Ph.D. Assistant Director Diagnostic X-ray Systems Team Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K213462

Device Name EzRay M18 (Model: VMX-P400)

Indications for Use (Describe)
----------------------------------

EzRay M18 (Model: VMX-P400) is a portable X-ray system, intended for use by a qualified/trained physician or technician to acquire X-ray images of the desired parts of a patient's anatomy on adult and pediatric patients (including head, chest, abdomen, cervical spine, and extremities).

The device may be used for handheld diagnostic imaging of body extremities.

The system is subject to the following limitations of use when stand-mounted:

  • The device may be used for diagnostic imaging of the head, abdomen, cervical spine, or extremities.

  • The device may be used for imaging of the chest when used without a grid.

This device is not intended for mammography.

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Section 5 – 510(k) Summary K213462

1. Traditional 510(k) Summary

This summary of 510(k) information is being submitted in accordance with requirements of 21 CFR Part 807.92.

2. Date 510K Summary prepared: January 7, 2022

3. Administrative Information

Official Correspondent:Dave Kim / Mtech GroupAddress: 7705 Fannin St., Ste. 610, Houston, TX 77054Tel: +713-467-2607Contact person: Mr. Dave Kim (davekim@mtech-inc.net)
510(k) Submitter:VATECH Co., Ltd.Address: 13, Samsung 1-ro 2-gil, Hwaseong-si, Gyeonggi-do, 18449, KoreaTel: +82-31-379-9492Fax: +82-31-379-9400Contact person: Daniel Kim / Manager (daniel.kim@vatech.com)

4. Device Information

Type of 510(k) Submission:Traditional
Trade or Proprietary Name:EzRay M18 (Model: VMX-P400)
Common or Usual Name:Medical Portable X-ray System
Regulation Classification:Mobile X-ray system (21 CFR 892.1720)
Product Code:IZL
Class of Device:Class II
Panel:Radiology

5. Predicate Device Information

Manufacturer:MinXray, Inc.
Trade or Proprietary Name:TR90BH
Common or Usual Name:Mobile Diagnostic X-Ray System
Regulation Classification:Mobile X-ray system (21 CFR 892.1720)
Product Code:IZL
Class of Device:Class II
Panel:Radiology
510(k) Number:K 182207

{4}------------------------------------------------

6. Reference Device Information

Manufacturer:OSKO, Inc.
Trade or Proprietary Name:XR5
Common or Usual Name:Stationary x-ray system
Regulation Classification:Stationary x-ray system (21 CFR 892.1680)
Product Code:KPR
Class of Device:Class II
Panel:Radiology
510(k) Number:K150663

7. Device Description

EzRay M18 (Model: VMX-P400), a medical portable X-ray system, operates on 57.6 Vdc supplied by a rechargeable Li-ion battery pack. The system is composed of an X-ray generating part including an X-ray tube, a device controller, a power controller, a user interface, a beam limiting part, and two optional components: remote controller and stand. The device software supports the EzRay M18 system, and the software is of Moderate level of concern. The device is designed for the human body using image receptors. The image detectors, a necessary component for a fully-functional x-ray system, are not part of this submission.

8. Indications for use

EzRay M18 (Model: VMX-P400) is a portable X-ray system, intended for use by a qualified/trained physician or technician to acquire X-ray images of the desired parts of a patient's anatomy on adult and pediatric patients (including head, chest, abdomen, cervical spine, and extremities).

The device may be used for handheld diagnostic imaging of body extremities.

The system is subject to the following limitations of use when stand-mounted:

  • The device may be used for diagnostic imaging of the head, abdomen, cervical spine, or extremities.

  • The device may be used for imaging of the chest when used without a grid.

This device is not intended for mammography.

{5}------------------------------------------------

9. Substantial Equivalence Chart

Subject DevicePredicate DeviceReference Device
Device NameEzRay M18 (Model: VMX-P400)TR90BHXR5
Applicant NameVATECH Co., Ltd.MinXray, Inc.OSKO, INC.
510(k) NumberK213462K182207K150663
Device Classification NameMobile X-ray systemMobile X-ray systemStationary X-ray system
Classification Product CodeIZLIZLKPR
Regulation Number21 CFR 892.172021 CFR 892.172021 CFR 892.1680
Regulation ClassIIIIII
Indications for UseEzRay M18 (Model: VMX-P400) isa portable X-ray system, intendedfor use by a qualified/trainedphysician or technician to acquireX-ray images of the desired partsof a patient's anatomy on adultand pediatric patients (includinghead, chest, abdomen, cervicalspine, and extremities).The device may be used forhandheld diagnostic imaging ofbody extremities.The system is subject to thefollowing limitations of use whenstand-mounted:- The device may be used fordiagnostic imaging of the head,abdomen, cervical spine, orextremities.- The device may be used forimaging of the chest when usedwithout a grid.This device is not intended formammography.The TR90BH is a portable X-ray system withfollowing limitations of use:The device may be used for handhelddiagnostic imaging of body extremitiesThe device may be used for stand mounteddiagnostic imaging of head, abdomen, orextremities.The device may be used for stand mountedimaging of the chest when used without agrid.- Not to be used on bariatric patients,unless imaging body extremities- Not for mammography use- The TR90BH is not intended to replace astationary radiographic system, which maybe required for full optimization of imagequality and radiation exposure for differentexam types.The XR5 diagnostic X-ray systemis intended for use on adult andpediatric patients for takingdiagnostics radiographicexposure of all body parts andoperated by a qualified/traineddoctor or technician. The XR5diagnostic X-ray system isdesigned to be used withconventional film/screen, CRcassettes or digital detectors.NOT intended forMammography use.
Size: Body214 x 305 x 309 mm(excluding Skin Distance Bar)219 x 442 x 190 mm1,940 mm (Tube stand height),1,480 mm (Vertical movement),2,540 mm (Longitudinalmovement)
TechnologicalWeight5.88 kg7.5 kg-
CollimatorFour manually and steplesslyadjustable shutters with LEDLightx-ray field indicatorMikasa BLD34LManual (K062788)
User InterfaceSoft touch push buttonsSoft touch push buttonspush buttons
Energy sourceRechargeable 57.6 V DCLi-ion battery packLithium-ion Rechargeable Battery(57.6DC), 300 exposures per charge.110-220 VAC
Exposure times0.02 sec - 2.0 sec: R'20 sec Step(Exposure conditions follow R'20in combination of mAs)0.01 sec - 1.0 sec : 0.01 sec StepHigh Power Mode0.01 sec - 0.3 sec : 0.01 sec Step6.3 sec Max.
mA20 mA @ 40 - 90 kV (1 kV steps)15 mA @ 61 – 90 kV (1 kV steps)10 mA @ 81 – 90 kV (1 kV steps)20 mA @ 40 kVDC – 60 kVDC (2 kVP steps)15 mA @ 62 kVDC – 80 kVDC (2 kVP steps)10 mA @ 82 kVDC – 90 kVDC (2 kVP steps)High Power Mode15 mA @ 82 kVDC – 90 kVDC (2 kVP steps)10 - 500 mA
Memorysettings10 memories via pushbutton5 memories via pushbuttonNo
HF GeneratorHigh FrequencyHigh FrequencyHigh Frequency
kW1.8 kW1.35 kW40 kW
kVp40 – 90kVp40 – 90kVp40-125 kVp
X-ray TubeD-0814D-0814E7239x
FDA Performance StandardCompliesCompliesComplies

{6}------------------------------------------------

VATECH Co., Ltd. Premarket Notification 510(k)=

{7}------------------------------------------------

9. The differences between the subject device and the predicate device

The subject device described in this traditional 510(k) is similar to the predicate device (K182207) in its indications for use and technological application. Both the subject and predicate devices are taking diagnostic Xrays using a variable tube voltage (kVp) and current time product (mAs) or exposure time. The collimator of the EzRay M18 device is a continuously adjustable light beam type similar to the predicate TR90BH. The user can irradiate X-ray by adjusting the X-ray filed size of the desire body part. Exposure parameters are similar for both EzRay M18 and TR90BH using the same Canon X-ray tube model D-However, EzRay 18 has a larger capacity of the X-ray mono tank within the X-ray tube spec, resulting in 0814. higher exposure output. The high power mode of the predicate device, TR90BH, is 15 mA, the output (KW) is 1.35 kW for the maximum tube voltage 90kV. EzRay M18 emits a maximum of 20 mA and the output of 1.8 kW when the maximum tube voltage is 90kV. Because EzRay M18 is capable of setting a higher mA under the same maximum tube voltage of 90kV, it is possible to capture radiographic images of the same body parts intended for the predicate device. For the energy source, battery pack, the subject device receives power from the built-in rechargeable Li-ion battery pack which as same as the predicate device (K182207) uses. Rechargeable Li-ion battery pack has been tested and is in conformity with the standard IEC 62133. Other differences include device design such as exposure time, size and user interface.

10. Non clinical Testing

Testing was performed successfully according to the following standards:

  • IEC 60601-1:2005, COR1:2006, COR2:2007, AMD1:2012 ●
  • . IEC 60601-1-2:2014
  • . IEC 60601-1-3:2008 (Second Edition) + A1:2013
  • IEC 60601-2-28:2017
  • IEC 60601-2-54:2009, AMD1:2015, AMD2:2018
  • IEC 62133-2:2017

Furthermore, the following Specific Guidance Document was utilized in the device development to ensure the safety of this device for both the operators and patients:

"The Content of Premarket Submissions for Software Contained in Medical Devices" "Radiation Safety Consideration for X-ray Equipment Designed for Hand-Held Use"

The device also conforms to the following:

21 CFR 1020 Subchapter J: Performance Standards for Ionizing Radiation Emitting Products

21 CFR 1020.30: Diagnostic x-ray system and their major components

21 CFR 1020.31: Radiographic Equipment

Image performance testing was performed on the EzRay M18 in comparison with a reference device XR5 (K150663) using a FDA-cleared flat-panel detector, 1417WCC (K171418). The results of MTF, spatial frequency, DQE and NPS were compared under the same exposure conditions. The results demonstrated that the general image performance of the subject device is equivalent to the reference device. The difference between the two devices is that EzRay M18 is a portable X-ray system whereas the reference device is a stationary diagnostic X-ray system. It is concluded that EzRay M18 demonstrated adequate image quality performance through comparative study with the reference device.

{8}------------------------------------------------

11. Conclusions

In reference to the comparison information provide in Substantial Equivalence Chart, the subject device and the predicate device have little difference with its technological features. As demonstrated in the performance bench testing, X-ray performance and X-ray Safety of the subject device was tested in accordance with Federal standard 21CFR Part 1020.30 and 31 as well as international standards such as IEC 60601-1, 60601-2-54. Quality assurance procedures are adhered to, and the specifications and functional requirements were met as the test results indicated.

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification. VATECH Co., Ltd. concludes that EzRay M18 (Model: VMX-P400) is substantially equivalent to the predicate device and its image quality performed equivalently in comparison with XR5 (K150663) using 1417WCC, a FDA-cleared digital detector (K171418) under the same exposure condition.

§ 892.1720 Mobile x-ray system.

(a)
Identification. A mobile x-ray system is a transportable device system intended to be used to generate and control x-ray for diagnostic procedures. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.