K Number
K213457

Validate with FDA (Live)

Date Cleared
2022-03-02

(126 days)

Product Code
Regulation Number
880.6880
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Fort Defiance Industries FRONT-LINE Field Sterilizers are autoclaves designed for sterilizing heat- and moisturestable medical, dental, and surgical materials, including wrapped, solid, porous, and hollow items (e.g., dental handpieces, suction pipes) in healthcare facilities.

Device Description

The Fort Defiance Industries FRONT-LINE Field Sterilizers are autoclaves designed for sterilizing heat- and moisture-stable medical, dental, and surgical materials, including wrapped and unwrapped, solid, porous, and hollow items (e.g., dental handpieces, suction pipes) in healthcare facilities. The devices use electrical resistive heaters to produce steam as the sterilizing agent. The devices adhere to sterilization processes and requirements that are universally accepted sterilization standards for Tabletop Steam Sterilizers. A touchscreen controller is used for monitoring and control of the device.

The devices are equipped with multiple safety features. The door includes a safety locking mechanism and door switch. The door switch does not permit cycle operation unless the door is sealed closed and does not permit the door to be opened until the internal chamber pressure is at atmospheric pressure. Additional safety features include electronic overtemperature and overpressure protection, an independent high temperature cutoff switch, and a pressure safety relief valve for the pressure vessel.

Only United States Food and Drug Administration cleared sterilization accessories should be used with these autoclaves.

AI/ML Overview

The Fort Defiance Industries FRONT-LINE Field Sterilizers FL120 and FL135 are autoclaves designed for sterilizing heat- and moisture-stable medical, dental, and surgical materials. The device's performance was evaluated against the ANSI/AAMI ST55:2016 standard.

1. Table of Acceptance Criteria and the Reported Device Performance

Test ProcedureAAMI ST55:2016 RequirementsMeasured Performance
Sterilizer Temperature Control & Pressure Measurement (TP-5070)4.4.2.1 Temperature monitoring4.4.2.2 Relationship between RTD and chamber cold point4.4.2.4 Temperature graduations of 1°C or less4.4.3 -0/+6°F temp control4.4.5.1 Pressure gauge of +/- 3% of full scale, resolution of 1 PSIG or lessPASS
Temperature Accuracy (TP-5240)4.4.2.3 temperature accuracy of +/- 1℃ over designated rangePASS
Air Removal (TP-5080)4.6.1 Air removal for dynamic air removal sterilizersPASS
Software Verification and Validation (TP-5100, TP-5110, TP-5120, TP-5130)4.4.7 Sterilizer fault conditions4.4.8 Cycle documentationPASS
Biological Performance - Textiles (TP-5141)4.5 Biological performance of sterilizersPASS
Biological Performance - Wrapped Instrument (TP-5160)4.5 Biological performance of sterilizersPASS
Biological Performance - Dental Handpieces (TP-5170)4.5 Biological performance of sterilizersPASS
Moisture Retention – Textiles (TP-5190)4.7 Moisture RetentionPASS
Moisture Retention – Wrapped Instrument (TP-5200)4.7 Moisture RetentionPASS
Moisture Retention – Pouches Drying Test (TP-5210)4.7 Moisture RetentionPASS
Endotoxin and Water Quality (TP-5220)A.4.2.6 Water supply reservoirPASS
Labeling (TP-5010) (FDI)AAMI ST55:2016 - 4.1.1 Device MarkingsAAMI ST55:2016 - 4.1.2 Information ManualAAMI ST55:2016 - 4.1.3 Service ManualPASS
Sterilizer Safety (TP-5050) (FDI)AAMI ST55:2016 - 4.3.1 InterlockAAMI ST55:2016 - 4.3.2 Thermal hazardsAAMI ST55:2016 - 4.3.3 Aborting cyclesPASS
General Inspection (TP-5060) (FDI)4.2.1 & 4.2.2 Pressure vessel requirements4.2.4 Corrosion resistance4.2.5 Air filters4.2.6 Water supply reservoir4.4.4 Sterilizer exposure timer4.4.6 Cycle completionPASS
3rd Party Testing Electrical SafetyIEC 61010-1 Edition 3.1 2017-01PASS
3rd Party Testing Electrical/Sterilizer Specific SafetyIEC 61010-2-040:2020PASS
3rd Party Testing EMC/EMIIEC 61326-1:2020PASS

2. Sample size used for the test set and the data provenance

The document does not explicitly state a numerical sample size for the test set. However, a series of performance effectiveness tests were conducted for each cycle type (Immediate Use, Textiles, Wrapped Instruments/Pouches, Handpieces) across both FL120 and FL135 models. These tests were performed "by passing 3 successive cycles at half the programmed exposure time." This indicates that at least three runs were conducted for each sterilization cycle condition for both models.

The data provenance is from Fort Defiance Industries LLC itself, conducting validation studies in accordance with AAMI/ANSI ST55:2016. This suggests that the studies were internal, likely prospective laboratory testing. No information on country of origin of the data is provided beyond Fort Defiance Industries being a US company.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. The ground truth for this device is established through meeting the sterilizing performance criteria defined by the ANSI/AAMI ST55:2016 standard, specifically achieving a sterility assurance level (SAL) of at least 10^-6, and safety standards like IEC and ASME. This is a technical performance measurement rather than an expert interpretation of data.

4. Adjudication method for the test set

Not applicable. The evaluation is based on objective measurements and compliance with established performance standards for sterilization, rather than subjective interpretation requiring adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is a medical sterilizer, not an AI-powered diagnostic or interpretive tool that would involve human readers or AI assistance.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Yes, in concept, the performance testing described is "standalone" for the device, meaning the sterilizer's effectiveness is evaluated on its own based on its ability to sterilize according to predefined biological and physical parameters. There is no human-in-the-loop performance aspect for a sterilizer.

7. The type of ground truth used

The ground truth used for the performance effectiveness of the sterilizer cycles is based on:

  • Biological performance: Achieving a sterility assurance level (SAL) of at least 10^-6, demonstrated by the "overkill method" using biological indicators.
  • Physical and technical parameters: Compliance with predefined temperature accuracy, pressure measurement, air removal, moisture retention, software functionality, and electrical/safety standards as outlined in ANSI/AAMI ST55:2016, IEC 61010-1, IEC 61326-1, IEC 61010-2-040, and ASME Boiler and Pressure Vessel Code.

8. The sample size for the training set

Not applicable. This device is a medical sterilizer and does not involve machine learning or AI models that require a training set.

9. How the ground truth for the training set was established

Not applicable, as there is no training set for this device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION".

March 2, 2022

Fort Defiance Industries LLC Chris Coleman Quality Management Representative 2411 Maremont Parkway Loudon, Tennessee 37774

Re: K213457

Trade/Device Name: Fort Defiance Industries FRONT-LINE Field Sterilizer FL120. Fort Defiance Industries FRONT-LINE Field Sterilizer FL135 Regulation Number: 21 CFR 880.6880 Regulation Name: Steam Sterilizer Regulatory Class: Class II Product Code: FLE Dated: October 26, 2021 Received: October 27, 2021

Dear Chris Coleman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K213457

Device Name

Fort Defiance Industries FRONT-LINE Field Sterilizer FL120, Fort Defiance Industries FRONT-LINE Field Sterilizer FL135

Indications for Use (Describe)

The Fort Defiance Industries FRONT-LINE Field Sterilizers are autoclaves designed for sterilizing heat- and moisturestable medical, dental, and surgical materials, including wrapped, solid, porous, and hollow items (e.g., dental handpieces, suction pipes) in healthcare facilities.

MODELCYCLEMAXIMUMLOADEXPOSURETEMP (F/C)*EXPOSURETIME (MIN)NUMBER OFDYNAMICAIR REMOVALPULSESDRY TIME(MINUTES)
FL120IMMEDIATEUSE (IUSS)20 lbs.270 / 132430
275 / 135330
FL120TEXTILES3 TEXTILEPACKS **250 / 12130360
270 / 1324360
275 / 1353360
FL120WRAPPED INSTRUMENTS/ POUCHES20 lbs.250 / 12130360
270 / 1324360
275 / 1353360
FL120HANDPIECES12 HANDPIECES270 / 1324560
275 / 1353560
FL135IMMEDIATEUSE (IUSS)25 lbs270 /132430
275 / 135330
FL135TEXTILES3 TEXTILEPACKS **250 / 12130360
270 / 1324360
275 / 1353360
FL135WRAPPED INSTRUMENTS/ POUCHES25 lbs250 / 12130360
270 / 1324360
275 / 1353360
FL135HANDPIECES12 HANDPIECES270 / 1324560
275 / 1353560

NOTES:

  • THE ANSI/AAMI ST55:2016 MINIMUM TEMPERATURE IS LISTED FOR ALL CYCLES.

** AAMI STANDARD TEXTILE PACK (9 TOWELS - ST 55 SECTION 5.7.1.1).

{3}------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

  • X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for FDI in orange. The logo is made up of the letters F, D, and I. The letters are all capitalized and in a sans-serif font. The logo is simple and modern. The color is bright and eye-catching.

510(k) Summary for Fort Defiance Industries FRONT-LINE Field Sterilizer FL120 & FL135 K213457

Submitter / 510(k) Owner: Fort Defiance Industries LLC 2411 Maremont Parkway Loudon, TN 37774 Phone: (865) 408-0100 Fax No: (866) 732-8086

Contact: Chris Coleman Quality Management Representative Fort Defiance Industries LLC Phone: (865) 657-6108 (866) 732-8086 Fax:

Submission Date: March 2nd, 2022

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for FDI in orange. The letters are large and bold. There are white lines that cut through the letters, giving the logo a modern look.

1. Device Name

Trade Name:Fort Defiance Industries FRONT-LINE Field Sterilizer FL120, FortDefiance Industries FRONT-LINE Field Sterilizer FL135
Model Name(s):FL120 / FL135
Common/usual Name:Electronic autoclave, steam sterilizer
Classification Name:Steam SterilizerClass II Device - 21 CFR 880.6880 Product Code FLE

2. Predicate Device

K111736, Tuttnauer EZ11 Plus Electronic Tabletop Autoclave, product code FLE, cleared December 8, 2011

3. Description of Device

The Fort Defiance Industries FRONT-LINE Field Sterilizers are autoclaves designed for sterilizing heat- and moisture-stable medical, dental, and surgical materials, including wrapped and unwrapped, solid, porous, and hollow items (e.g., dental handpieces, suction pipes) in healthcare facilities. The devices use electrical resistive heaters to produce steam as the sterilizing agent. The devices adhere to sterilization processes and requirements that are universally accepted sterilization standards for Tabletop Steam Sterilizers. A touchscreen controller is used for monitoring and control of the device.

The devices are equipped with multiple safety features. The door includes a safety locking mechanism and door switch. The door switch does not permit cycle operation unless the door is sealed closed and does not permit the door to be opened until the internal chamber pressure is at atmospheric pressure. Additional safety features include electronic overtemperature and overpressure protection, an independent high temperature cutoff switch, and a pressure safety relief valve for the pressure vessel.

Only United States Food and Drug Administration cleared sterilization accessories should be used with these autoclaves.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for FDI in orange. The logo is made up of the letters F, D, and I. The letters are all capitalized and in a sans-serif font. The logo is simple and modern. The color is bright and eye-catching.

4. Design and Materials

The FRONT-LINE Field Sterilizers, models FL120 and FL135, are steam sterilizers composed of a stainless-steel chamber, an aluminum chamber door, electric resistive heaters, a plumbing system with a clean water tank, control system, and an aluminum frame providing a housing for all components. The devices are electronically controlled, and each device includes 10 pre-programmed sterilization cycles. All materials in the sterilization system and plumbing system are deemed acceptable for steam sterilizers.

న. Technology

The FRONT-LINE series of autoclaves are steam sterilizers composed of a pressure vessel that utilizes electric resistive heaters to produce the sterilizing agent, steam, from a chamber boiler. The autoclaves contain an electronic control system that automatically initiates programmed Steam Flush Pressure Pulse (SFPP) phases to purge atmospheric air from the chamber followed by a timed steam exposure and drying phase. Exhaust air purged during the dry phase passes through a heat exchanger to condense water vapor and reclaim as much condensate as possible for reuse.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for FDI. The logo is orange and features the letters "FDI" in a bold, sans-serif font. The letters are connected to each other, and there are white lines running through the letters, creating a sense of movement. The logo is simple and modern, and the orange color is eye-catching.

6. Intended Use

The Fort Defiance Industries FRONT-LINE Field Sterilizers are autoclaves designed for sterilizing heat- and moisture-stable medical, dental, and surgical materials, including wrapped and unwrapped, solid, porous, and hollow items (e.g., dental handpieces, suction pipes) in healthcare facilities.

ModelCycleMaximumLoadExposureTemperature¹(°F/°C)ExposureTime(minutes)Dry Time(minutes)
FL120Immediate Use(IUSS)20 lbs.270/132275/135430
Textiles3 Textile Packs²250/121270/132275/135304360
WrappedInstruments /Pouches20 lbs.250/121270/132275/135304360
FL120Handpieces12 Handpieces270/132460
275/1353
FL135Immediate Use(IUSS)25 lbs.270/132275/135430
Textiles3 Textile Packs²250/121270/132275/135304360
WrappedInstruments /Pouches25 lbs.250/121270/132275/135304360
FL135Handpieces12 Handpieces270/132460
275/1353

1 The ANSI/AAMI ST55:2016 minimum temperature is listed for all cycles.

2 AAMI Standard Textile Pack (9 towels - ST 55 section 5.7.1.1).

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for FDI. The logo is orange and features the letters "FDI" in a bold, sans-serif font. The "F" and "D" are connected by a horizontal line, and there is a diagonal line cutting through the top right corner of the "I". The logo is simple and modern.

7. Comparison of Technology

The previously cleared device and the FRONT-LINE Field Sterilizers FL120 and FL135 are traditional steam sterilizers of similar chamber size using AAMI approved Steam Flush Pressure Pulse (SFPP) dynamic air removal cycles. The previously cleared device and the FRONT-LINE series sterilizers use electric heaters to generate steam and touchscreen controllers for cycle control. Based on the information provided below, there are negligible differences noted between the FRONT-LINE series and the predicate with regards to chamber size, cycle parameters, and configuration of steam generation. The differences do not affect safety or effectiveness of the new devices compared to the predicate. Table 1 provides a summary of performance and safety characteristics of the proposed new devices and the predicate.

Performanceand/or SafetyCharacteristicNew DeviceFort Defiance IndustriesFRONT-LINE Field SterilizersFL120 & FL135K213457Predicate DeviceElectronic Tabletop AutoclaveTuttnauer EZ11 PlusK111736
Indications forUseThe Fort Defiance IndustriesFRONT-LINE Field Sterilizersare autoclaves designed forsterilizing heat- and moisture-stable medical, dental, andsurgical materials, includingwrapped and unwrapped, solid,porous, and hollow items (e.g.,dental handpieces, suction pipes)in healthcare facilities.The EZ Plus series of autoclaves areautoclaves designed for sterilizingmedical and surgical goods,including both wrapped andunwrapped, solid, hollow, andporous products and goods definedas hollow (e.g. dental handpieces,suction pipes) in ophthalmic, dental,and medical clinics; in first aidrooms; and in small laboratories.
OperatingPrincipleSteam Flush Pressure Pulse(SFPP).Steam is the sterilization agent.(traditional method).Steam Flush Pressure Pulse(SFPP).Steam is the sterilization agent(traditional method).
Performanceand SafetyStandardsFollowedANSI/AAMI ST55ASME Sec. VIII Div 1IEC 61010-1IEC 61326-1IEC 61010-2-040ANSI/AAMI ST55ASME Sec. VIII Div 1UL 61010-1EN 61326-1UL 61010-2-040
Performanceand/or SafetyCharacteristicNew DeviceFort Defiance IndustriesFRONT-LINE Field SterilizersFL120 & FL135K213457Predicate DeviceElectronic Tabletop AutoclaveTuttnauer EZ11 PlusK111736
Factory Pre-ProgrammedSteam Flush Pressure Pulse(SFPP) Conditioning PhaseFactory Pre-ProgrammedSteam Flush Pressure Pulse(SFPP) Conditioning Phase
SterilizationCycles30 min exposure @ 250°F60 min dry time - Default30 min exposure @ 250°F1 min dry time - Default
4 min exposure @ 270°F60 min dry time - Default3 or 4 min exposure @ 270°FDry time dependent on load
3 min exposure @ 275°F60 min dry time - Default
Chamber Size& VolumeFL12012" dia. X 23" lg.1.6 cf. (45.3 L)FL13513.5" dia. X 24" lg.2.00 cf. (56.35 L)11" dia. X 19.8" lg.1.0 cf. (28.5 L)
ChamberDesign andConstructionSingle Wall ChamberASME Section VIII, Div. IcertifiedStainless steel (316L) chamberand aluminum door (6061-T6)Single Wall ChamberASME Section VIII, Div. IcertifiedStainless steel (316L) chamberand stainless-steel door
Shelves / TraysHard anodized aluminum shelfStainless Steel wire trays
Steam SourceInternal chamber boiler viapartitioned sectionInternal chamber boiler alonglength of chamber
SteamGenerationMechanismElectric resistive heaterssubmerged in chamber boilerExternal chamber wall electric bandheaters
Dying ProcessExternal air pulled through 0.3-micron HEPA filter then heatedfor dryingExternal air pulled through 0.2-micron HEPA filter then heatedfor drying
ControlTechnologyElectronic controller touch screen(PLC)Electronic controller touch screen(PLC)
Performanceand/or SafetyCharacteristicNew DeviceFort Defiance IndustriesFRONT-LINE Field SterilizersFL120 & FL135Predicate DeviceElectronic Tabletop AutoclaveTuttnauer EZ11 PlusK111736
ProcessMonitorsChamber pressure transmitter.Dual element chambertemperature sensor.Chamber pressure transmitter.Dual element chambertemperature sensor.
Door FailsafeAutomatic Door InterlockAutomatic Door Interlock
PrimaryOvertemp andOverpressureControlPLC Controller -Utilizing RTDs and PressureTransmitterPLC Controller -Utilizing RTDs and PressureTransmitter
SecondaryOvertempControlThermocouple TemperatureCut-off Switch (TCO)Capillary TemperatureCut-off Switch (TCO)
SecondaryOverpressureControlASME Pressure Safety ValveASME Pressure Safety Valve
ElectricalInput120VAC +/- 10%,1Ø, 50/60 Hz, 15 A230-240VAC +/- 10%,1Ø, 50/60 Hz, 15 A120VAC +/- 5%, 60 Hz

Table 1: Comparison of Technology

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for FDI in orange. The logo is made up of the letters F, D, and I. The letters are all capitalized and in a sans-serif font. There is a white line that cuts through the top of the letters.

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the logo for FDI, which is an acronym. The letters are orange and have a unique design. The letters are connected by a white line that runs through the top of the letters.

8. Summary of Non-Clinical Testing

Fort Defiance Industries conducted validation studies in accordance with AAMI/ANSI ST55:2016 (FDA Recognition Number 14-518). Design outputs and testing demonstrate that the FRONT-LINE Models FL120 and FL135 autoclaves meet all aspects of the standard.

Performance effectiveness of the sterilizer cycles and exposure time recommendations were demonstrated by the overkill method by passing 3 successive cycles at half the programmed exposure time to guarantee a sterility assurance level (SAL) of at least 10 to probability of survival. Fort Defiance Industries validates its sterilization cycles using recommended practices, standards and guidelines developed by independent organizations such as the Association for the Advancement of Medical Instrumentation (AAMI). The Fort Defiance Industries FRONT-LINE Field Sterilizers FL120 and FL135 have been validated to meet the requirements of ANSI/AAMI ST55:2016, Tabletop Steam Sterilizers (FDA Recognition Number 14-518). Table 2 summarizes FRONT-LINE Field Sterilizer FL120 and FL135 effectiveness testing and results performed by Fort Defiance Industries.

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the logo for FDI in orange. The logo is made up of the letters F, D, and I. The letters are connected to each other, and there is a white line running through the top of the letters. The logo is simple and modern.

Table 2: Performance Testing Summary

Test ProcedureAAMI ST55:2016RequirementsAAMI ST55:2016 TestsMeasuredPerformance
Sterilizer TemperatureControl&Pressure MeasurementTP-5070Note:TR-5072 reportsSterilizer Resolutions ofTemperatureMeasurementTR-5075 reportsSterilizer PressureMeasurementTR-5077 reports ColdPoint Relationship4.4.2.1 Temperature monitoring4.4.2.2 Relationship betweenRTD and chamber cold point4.4.2.4 Temperature graduationsof 1°C or less4.4.3 -0/+6°F temp control4.4.5.1 Pressure gauge of +/- 3%of full scale, resolution of 1PSIG or less5.4.2.1 - Verified byinspection5.4.2.2 - Verified byinspection and test 5.4.35.4.2.4 - Verified byinspection5.4.3 - Temperatureinstrumented chamber teststo confirm steady statetemperature control rangeand documentation of coldpoint5.4.5.1 - Verified byinspection and by testingagainst certified standardsPASS
Temperature AccuracyTP-52404.4.2.3 temperature accuracy of+/- 1℃ over designated range5.4.2.3 - Verified by testingagainst calibrated standardsPASS
Air RemovalTP-50804.6.1 Air removal for dynamicair removal sterilizersA.4.6.1 - Alternate test forSteam Flush Pressure PulseSterilizersPASS
Software Verificationand ValidationTP-5100: Sterilizer LogVerificationTP-5110: Sterilizer FaultConditionsTP-5120: GeneralFunctionsTP-5130: Full CycleVerification4.4.7 Sterilizer fault conditions4.4.8 Cycle documentation5.4.7 - Verified by inspection5.4.8 - Verified by inspectionPASS
Biological Performance- TextilesTP-51414.5 Biological performance ofsterilizers5.5.2 Biological performancewith a textile packPASS
Biological Performance- Wrapped InstrumentTP-51604.5 Biological performance ofsterilizers5.5.4 Biological performancewith wrapped instrumentsPASS
Biological Performance- Dental HandpiecesTP-51704.5 Biological performance ofsterilizers5.5.5 Biological performancewith dental handpiecesPASS
Test ProcedureAAMI ST55:2016RequirementsAAMI ST55:2016 TestsMeasuredPerformance
Moisture Retention –TextilesTP-51904.7 Moisture Retention5.7.1 Textile test packsPASS
Moisture Retention –Wrapped InstrumentTP-52004.7 Moisture Retention5.7.2 Wrapped Instr. TesttraysPASS
Moisture Retention –Pouches Drying TestTP-52104.7 Moisture Retention5.7.3 Paper-plastic peelpouchesPASS
Endotoxin and WaterQualityTP-5220A.4.2.6 Water supply reservoirA.4.2.6 Water supplyreservoirPASS
FDI Test Procedure or3rd PartyStandardTestsMeasuredPerformance
LabelingTP-5010(FDI)AAMI ST55:2016 - 4.1.1Device MarkingsAAMI ST55:2016 - 4.1.2Information ManualAAMI ST55:2016 - 4.1.3Service ManualAAMI ST55:2016 - 5.1 Visualinspection onlyPASS
Sterilizer SafetyTP-5050(FDI)AAMI ST55:2016 - 4.3.1InterlockAAMI ST55:2016 - 4.3.2Thermal hazardsAAMI ST55:2016 - 4.3.3Aborting cyclesAAMI ST55:2016 - 5.3.1Interlock functions underconditions describedAAMI ST55:2016 - 5.3.2Thermal hazard testing per UL-61010AAMI ST55:2016 - 5.3.3Verified by inspectionPASS
General InspectionTP-5060(FDI)4.2.1 & 4.2.2 Pressure vesselrequirements4.2.4 Corrosion resistance4.2.5 Air filters4.2.6 Water supply reservoir4.4.4 Sterilizer exposure timer4.4.6 Cycle completionAAMI ST55:2016 - 5.2.1Verified by compliance toASME BPVC Section VIIIDivision 1 Code, requirementsin UL 61010-1 and IEC 61010-2-040AAMI ST55:2016 - 5.2.4Material corrosion resistancereviewAAMI ST55:2016 - 5.2.5 Airfilter inspection necessaryfiltration efficiency and filterreadily accessibleAAMI ST55:2016 - 5.2.6Verified by inspectionAAMI ST55:2016 - 5.4.4Verified by inspection andtimer testing to NIST standardAAMI ST55:2016 - 5.6.6Verified by inspectionPASS
3rd Party TestingElectrical SafetyIEC 61010-1 Edition 3.12017-013rd Party test protocolPASS
3rd Party TestingElectrical/SterilizerSpecific SafetyIEC 61010-2-040:20203rd Party test protocolPASS
3rd Party TestingEMC/EMIIEC 61326-1:20203rd Party test protocolPASS

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image shows the logo for FDI in orange. The logo is made up of the letters F, D, and I. The letters are connected to each other, and the letter D is in the center. There is a white line that cuts through the top of the letters F and D. The logo is simple and modern.

9. Safety

The Fort Defiance FRONT-LINE Field Steam Sterilizers have been designed, constructed, and tested to meet the safety and performance requirements of various national safety codes and standards. The device complies with the following standards:

  • ANSI/AAMI ST55:2016, Tabletop Steam Sterilizers (FDA Recognition Number 14-518).
  • IEC 61010-1 Edition 3.1 2017-01, "Safety Requirements for Electrical Equipment ● for Measurement, Control and Laboratory Use - Part 1 General Requirements". (FDA Recognition Number 19-34)
  • IEC 61326-1:2020, "Electrical Equipment for Measurement, Control, and . Laboratory Use – EMC Requirements – Part 1: General Requirements" (General Use)
  • . IEC 61010-2-040:2020, "Safety Requirements for Electrical Equipment for Measurement, Control and Laboratory Use - Part 2-040: Particular Requirements for Sterilizers and Washer-disinfectors used to Treat Medical Materials" (General Use)
  • ASME Boiler and Pressure Vessel Code, Section VIII, Division I, 2019 Edition, . Rules for Construction of Pressure Vessels (General Use)

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image shows the logo for FDI in orange. The letters are large and bold. A white line cuts through the top right corner of the D and the I.

Table 3 summarizes FRONT-LINE Field Sterilizer FL120 and FL135 safety testing results performed by Fort Defiance Industries or a Third Party where indicated.

Table 3: Safety Testing Summary

{14}------------------------------------------------

Image /page/14/Picture/0 description: The image shows the logo for FDI in orange. The logo is made up of the letters F, D, and I. The letters are all capitalized and in a sans-serif font. The logo is simple and modern. The color is bright and eye-catching.

10. Conclusion

The conclusion drawn from the nonclinical testing demonstrates that the subject device is as safe, as effective and performs as well as or better than the legally marketed predicate device, K111736, Class II (21 CFR 880.6880, Product code FLE).

§ 880.6880 Steam sterilizer.

(a)
Identification. A steam sterilizer (autoclave) is a device that is intended for use by a health care provider to sterilize medical products by means of pressurized steam.(b)
Classification. Class II (performance standards).