K Number
K213301

Validate with FDA (Live)

Date Cleared
2021-12-21

(81 days)

Product Code
Regulation Number
888.3040
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Nextremity Solutions InCore® Subtalar System is intended for reduction and internal fixation of arthrodeses, osteotomies, and nonunions of the bones and joints of the foot. The three-part construct is specifically intended for internal fixation for Subtalar Joint Arthodesis (also known as Subtalar Joint Fusion).

Device Description

The InCore® Subtalar System consists of a post and two headless compression screws. Posts are available with an 8.0 mm diameter, in an 28mm length, and in right and left orientations. Screws are available in a 5.5 mm diameter and lengths of 50 to 120mm. The post is inserted into the talus and compression screws are inserted into the calcaneus and into the post to maintain apposition of the bones during fusion. A post plug screw is threaded into the top of the post after all components have been implanted to prevent tissue ingrowth into the post and facilitate removal, if needed. All implants are manufactured from color anodized Ti-6Al-4V alloy conforming to ASTM F-136.

The system is provided with a set of accessory instruments, including an implant specific targeting guide, designed for preparation of the implant site and insertion of the implants into bone.

AI/ML Overview

The provided text describes a 510(k) premarket notification for a medical device called the InCore® Subtalar System. This document does not include information about AI/ML device performance, acceptance criteria for an algorithm, or a study that proves a device meets such criteria.

The information provided is typical for a medical device clearance, focusing on:

  • Substantial Equivalence: The primary goal of a 510(k) submission is to demonstrate that the new device is as safe and effective as a legally marketed predicate device.
  • Device Description and Intended Use: What the device is and what it's used for (internal fixation of bone arthrodeses, osteotomies, and nonunions in the foot, specifically subtalar joint arthrodesis).
  • Non-Clinical Testing: Mechanical and biocompatibility tests (e.g., torque to failure, three-point bend tests, endotoxin testing) to ensure the device's physical properties and safety.
  • Clinical Testing: The document explicitly states, "Clinical testing was not necessary to demonstrate substantial equivalence of the InCore® Subtalar System to the predicate device." This means no human clinical trials were conducted for this submission.

Therefore, I cannot extract the requested information regarding acceptance criteria, AI/ML study performance, sample sizes, expert involvement, or ground truth establishment for an AI/ML device from this document. The device described is a physical implant (bone screw system), not an AI algorithm.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.

December 21, 2021

Nextremity Solutions, Inc. Elise Fox Quality and Regulatory Specialist 1195 Polk Drive Warsaw, Indiana 46582

Re: K213301

Trade/Device Name: InCore® Subtalar System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: HWC Dated: September 30, 2021 Received: October 1, 2021

Dear Elise Fox:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Shumaya Ali, MPH Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K213301

Device Name

InCore® Subtalar System

Indications for Use (Describe)

The Nextremity Solutions InCore® Subtalar System is intended for reduction and internal fixation of arthrodeses, osteotomies, and nonunions of the bones and joints of the three-part construct is specifically intended for internal fixation for Subtalar Joint Arthodesis (also known as Subtalar Joint Fusion).

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary - K213301

Prepared:September 30, 2021
Submitter:Nextremity Solutions, Inc.1195 Polk DriveWarsaw, IN 46582
Contact:Elise FoxQuality and Regulatory Specialistelise.fox@nextremity.comPhone: 574-376-2062FAX: 574-966-1396
Proprietary Name:InCore® Subtalar System
Common Name:Bone Screw System
Classification:21 CFR § 888.3040 Smooth or threaded metallic bone fixation fastenerClass II
Product Code:HWC

Substantially Equivalent Devices:

  • Zimmer Biomet, TiMAX™ Cannulated Screw System (Cleared as Biomet Cannulated ● Screw System), K140891-Primary Predicate
  • Nextremity Solutions, Stratum Foot Plating System, K182201-Additional Predicate .

Device Description:

The InCore® Subtalar System consists of a post and two headless compression screws. Posts are available with an 8.0 mm diameter, in an 28mm length, and in right and left orientations. Screws are available in a 5.5 mm diameter and lengths of 50 to 120mm. The post is inserted into the talus and compression screws are inserted into the calcaneus and into the post to maintain apposition of the bones during fusion. A post plug screw is threaded into the top of the post after all components have been implanted to prevent tissue ingrowth into the post and facilitate removal, if needed. All implants are manufactured from color anodized Ti-6Al-4V alloy conforming to ASTM F-136.

The system is provided with a set of accessory instruments, including an implant specific targeting guide, designed for preparation of the implant site and insertion of the implants into bone.

{4}------------------------------------------------

Intended Use / Indications:

The Nextremity Solutions InCore® Subtalar System is intended for reduction and internal fixation of arthrodeses, osteotomies, and nonunions of the bones and joints of the foot. The three-part construct is specifically intended for internal fixation for Subtalar Joint Arthodesis (also known as Subtalar Joint Fusion).

Summary of Technologies/Substantial Equivalence:

The InCore® Subtalar System is substantially equivalent to the predicate devices regarding the intended use and indications, material, design, sizes, and mechanical properties. Differences between the subject device system and the predicate device systems do not raise new types of safety and effectiveness questions.

Non-Clinical Testing:

Endotoxin testing was performed using the Limulus Amebocyte Lysate (LAL) method according to AAMI ST72, USP 161 and USP 85. Results met the Endotoxin limit of ≤20 EU per device. To evaluate the strength of the InCore® Subtalar System and components, torque to failure was performed on worst case compression screws and worst case constructs per ASTM F543. Static three-point bend and dynamic three-point bend tests were performed on worst case constructs per ASTM F382 and ASTM F1264. These tests confirmed that the strength of the InCore® Subtalar System is substantially equivalent to predicate devices with similar indications and is adequate for its intended use.

Clinical Testing:

Clinical testing was not necessary to demonstrate substantial equivalence of the InCore® Subtalar System to the predicate device.

Conclusions/Substantial Equivalence:

The InCore® Subtalar System is substantially equivalent to the predicate devices regarding its intended use, material, design, sizes, and mechanical properties. Differences between the subject device system and the predicate device systems do not raise new types of safety and effectiveness questions.

§ 888.3040 Smooth or threaded metallic bone fixation fastener.

(a)
Identification. A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.