K Number
K213275

Validate with FDA (Live)

Manufacturer
Date Cleared
2021-12-20

(81 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

EchoGo Core is intended to be used for quantification and reporting of results of cardiovascular function to support physician diagnosis. EchoGo Core is indicated for use in adult populations.

Device Description

EchoGo Core 2.0 is a software application manufactured by Ultromics to provide a report of left ventricular cardiac function, in the form of secondary capture DICOM files and/or as a structured DICOM report, to aid interpreting physicians with diagnostic decision-making process. EchoGo Core 2.0 applies to ultrasound images of the heart (echocardiograms).

EchoGo Core 2.0 utilizes artificial intelligence (AI) for the operator-assisted automatic quantification of commonly measured echocardiographic metrics. Independent training, test and validation datasets were used for training and performance assessment of the device.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria were formalized in terms of Root Mean Square (RMS) error against reference values generated using the comparator device, TomTec Arena TTA2. The text states that "the acceptance criteria were formalized such that EchoGo Core 2.0 would produce measures of left ventricular (LV) length, volume at end diastole (ED), and end systole (ES), ejection fraction (EF), stroke volume, cardiac output, global longitudinal strain (GLS) and segmental longitudinal strain (SLS) with an RMS error below a set, pre-determined threshold." While the exact pre-determined thresholds for acceptance are not explicitly listed in numerical values, the "Performance against the comparator device is summarised as follows," implying these are the reported performance values that met the (unspecified) acceptance thresholds.

Left Ventricular MetricRoot Mean Square Error (% RMS)
Length3.06 - 4.59
Volume at End Diastole and End Systole8.57 - 16.59
Ejection Fraction6.69 - 8.50
Stroke Volume10.57 - 13.68
Global Longitudinal Strain3.36 - 4.79
Systolic Segmental Longitudinal Strain5.51 - 9.98

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: 214 previously unseen studies.
  • Data Provenance:
    • Country of Origin: Not explicitly stated, but the submission is from Ultromics Limited in the United Kingdom.
    • Retrospective or Prospective: Retrospective. The study was described as a "formal retrospective, non-interventional validation study."

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

The ground truth for the test set was established using a comparator device, TomTec Arena TTA2 (K150122), not human experts providing a direct ground truth. The acceptance criteria were formalized in terms of Root Mean Square (RMS) error against reference values generated using the comparator device. There is no mention of experts establishing ground truth for the test set; instead, the comparator device served as the reference.

4. Adjudication Method for the Test Set

Not applicable, as the ground truth was established by a comparator device rather than human interpretation requiring adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, and the Effect Size

No, an MRMC comparative effectiveness study where human readers improve with AI vs. without AI assistance was not conducted or reported. The study focused on the device's performance against a comparator device.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) was Done

The primary performance study described seems to align with a standalone assessment against a comparator device. The text states, "EchoGo Core 2.0 would produce measures... with an RMS error below a set, pre-determined threshold" and "Performance against the comparator device is summarised as follows." This suggests the algorithm's output was directly compared to the output of the TomTec Arena TTA2.

However, it's crucial to note the device description also states: "EchoGo Core 2.0 requires an operator at key steps to confirm or relabel automatically labeled acquisition views (if required) and approve the left ventricle segmentations (contours) proposed by the AI." And "The operator will review the report produced and may be asked to approve cautions that are added to the report." This indicates that while the core performance metrics were evaluated in an algorithm-centric manner against a comparator, the device's intended use involves a "human-in-the-loop" for confirmation/approval steps. The reported performance metrics (RMS error) are likely for the algorithm's core measurements before potential human override, as the study compared them to an automated comparator device.

7. The Type of Ground Truth Used

The ground truth was established by comparison to a legally marketed predicate device (TomTec Arena TTA2, K150122). Values generated by the TomTec Arena TTA2 served as the reference values for calculating RMS error.

8. The Sample Size for the Training Set

The sample size for the training set is not specified in the provided text. The text only mentions that "Independent training, test and validation datasets were used for training and performance assessment of the device" and that "Test datasets were strictly segregated from algorithm training datasets."

9. How the Ground Truth for the Training Set was Established

The method for establishing ground truth for the training set is not explicitly detailed in the provided text. It only states that "Independent training, test and validation datasets were used for training and performance assessment of the device."

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo is composed of two parts: the Department of Health & Human Services logo on the left and the FDA acronym and full name on the right. The Department of Health & Human Services logo is a stylized representation of a human figure, while the FDA acronym and full name are written in blue.

Ultromics Limited % Jaco Jacobs Chief Regulatory and Compliance Officer 4630 Kingsgate, Cascade Way Oxford Business Park South Oxford, Oxfordshire OX4 2SU UNITED KINGDOM

December 20, 2021

Re: K213275

Trade/Device Name: EchoGo Core (2.0) Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: QIH Dated: September 27, 2021 Received: September 30, 2021

Dear Jaco Jacobs:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

{1}------------------------------------------------

https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

Submission Number (if known)

K213275

Device Name

EchoGo Core (2.0)

Indications for Use (Describe)

EchoGo Core is intended to be used for quantification and reporting of results of cardiovascular function to support physician diagnosis. EchoGo Core is indicated for use in adult populations.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "ULTROMICS" in bold, dark blue letters. To the left of the word is a circular logo with a gradient of blue and green. The logo appears to be a stylized letter "O" or a ring.

K213275

510(K) SUMMARY

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Ultromics. The logo consists of a circular graphic on the left and the word "ULTROMICS" in bold, dark blue letters on the right. The circular graphic is a gradient of blue and green, creating a swirling effect.

1 Submitter

CompanyUltromics Limited4630 Kingsgate Cascade Way, Oxford Business Park South, Oxford,Oxfordshire, United Kingdom, OX4 2SU
ContactDr. Jaco Jacobs
Date Prepared27 November 2021

2 Subject Device

Product Trade NameEchoGo Core
Model Number2.0
Medical SpecialityRadiology
Regulation892.2050 – Picture Archiving and Communications System
Product CodeQIH - Automated Radiological Image Processing Software
Regulatory ClassII

EchoGo Core is the product trade name and 2.0 is the model number. For the avoidance of doubt, in this submission we combine the product trade name and model number and refer to the subject device as EchoGo Core 2.0.

3 Predicate and Reference Devices

Predicate DeviceEchoGo Core
510(k)K191171
ManufacturerUltromics Limited
Reference DeviceTomTec Arena TTA2
510(k)K150122
ManufacturerTomTec Imaging Systems GmbH

Device Description বা

EchoGo Core 2.0 is a software application manufactured by Ultromics to provide a report of left ventricular cardiac function, in the form of secondary capture DICOM files and/or as a structured DICOM report, to aid interpreting physicians with diagnostic decision-making process. EchoGo Core 2.0 applies to ultrasound images of the heart (echocardiograms).

EchoGo Core 2.0 utilizes artificial intelligence (AI) for the operator-assisted automatic quantification of commonly measured echocardiographic metrics . Independent training, test and validation datasets were used for training and performance assessment of the device.

ECHOGO CORE 2.0

510(k) SUMMARY

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for ULTROMICS. The logo consists of a circular, gradient-colored icon on the left and the word "ULTROMICS" in bold, dark blue letters on the right. The icon has a swirling design with shades of green and blue, creating a sense of movement or flow.

EchoGo Core 2.0 requires an operator at key steps to confirm or relabel automatically labeled acquisition views (if required) and approve the left ventricle segmentations (contours) proposed by the Al.

EchoGo Core 2.0 operates in a sequential workflow manner, that includes:

    1. Automatic view classification and labeling
    1. Operator confirmation or relabeling of view classifications
    1. Contouring of the left ventricle in apical chamber views
    1. Automated heart cycle identification and end-diastolic and end-systolic frame selection
    1. Operator confirmation or rejection of left ventricle contouring and frame selection
    1. Calculation of clinical report metrics

To identify a heart cycle, the heart rate must be known. If an ECG was not taken during acquisition of the echocardiogram, the heart rate may be inferred. The operator is selecting a contoured image clip for each view and approving the selected cycle (and the key end-diastolic (ED) and end-systolic (ES) frames). The operator will review the report produced and may be asked to approve cautions that are added to the report. EchoGo Core 2.0 supports the production of reports for any combination of the three processible views (A2C, A3C and A4C), either embedded as image files in one or more multi-frame true color secondary capture DICOM files, and/or as a DICOM structured report.

5 Indications for Use

EchoGo Core 2.0 is intended to be used for quantification and reporting of results of cardiovascular function to support physician diagnosis. EchoGo Core 2.0 is indicated for use in adult populations.

Both the intended use and indications for use statement for the subject and predicate devices are identical. Both the subject and predicate devices are indicated for use in quantification and reporting of cardiovascular function to support the physician's diagnosis in an adult population.

Comparison of Technological Characteristics o

A comparison of technological differences between the subject and predicate devices follows.

CharacteristicSubject DeviceEchoGo Core 2.0Predicate DeviceEchoGo Core 1.0
Intended useIdenticalQuantification of cardiovascular function from an echocardiogram
Indications for useIdenticalQuantification and reporting of results of cardiovascular functionto support physician diagnosis.EchoGo Core is indicated for usein adult populations
Anatomical siteIdenticalCardiovascular structures
UsersIdenticalAccredited echocardiographersand sonographers at Ultromics
Machine learning-based algorithmYesYes
AlgorithmsAuto-Contouring DICOM HandlingAuto-Contouring DICOM Handling
CharacteristicSubject DeviceEchoGo Core 2.0Predicate DeviceEchoGo Core 1.0
• Technical QC• Technical QC
• Auto-View Classification• Manual Labelling UI
• Manual Labelling UI• Contour Selection andApproval
• Contour Selection andApproval• Metrics Calculation
• Metrics Calculation• Auditing
• Auditing• Reporting
• Service Webhooks
• Reporting
Operates on DICOM clipsYesYes
Average GLS A2C/A4C viewsYesYes
Average GLS A2C/A4C/A3C viewsYesNo
Per view GLS, EF and LV volumeYesNo
Segmental LSYesNo
Echocardiogram images on devicereportYesYes
Horseshoe and bullseyerepresentationsYesNo
Mechanical dispersion plotYesNo
Strain temporal intensityplot/CAMM plotYesNo
Contours overlayed onechocardiogram imagesYesNo
Auto-view classificationYesNo
Automatic ED/ES frame detectionYesYes
EF ReportedYesYes
Automated EF calculationYesYes
Simpson's biplane LV volumecalculationYesYes
Non-Simpson's bi-plane LVvolume calculation methodsYesNo
LV ED/ES volume indexed valuesYesNo
EF results shown with video clipYesYes
User confirmation / rejection ofresultYesYes
Manual editing of automatedresult by userNoNo
LV stroke volumeYesNo
LV lengthYesNo
Operating systemWindowsWindows
Operating platformSingle software applicationplatformSingle software applicationplatform
SoftwareComplies with IEC 62304:2015and developed under an EDA QSRComplies with IEC 62304:2006and developed under an EDA QSR
CharacteristicSubject DeviceEchoGo Core 2.0Predicate DeviceEchoGo Core 1.0
compliant QMS incorporating riskmanagement per ISO 14971:2019compliant QMS incorporating riskmanagement per ISO14971:2007
UsabilityComplies with IEC 62366-1:2020and general use of FDA guidanceon usability engineeringGeneral use of FDA guidance onusability engineering
Performance testingEquivalent to referencecomparator TomTec Arena TTA2(K150122)Equivalent to referencecomparator TomTec Arena TTA2(K150122)

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for ULTROMICS. The logo consists of a circular graphic on the left and the word "ULTROMICS" in bold, dark blue letters on the right. The circular graphic is a gradient of blue and green, creating a swirling effect.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for ULTROMICS. The logo consists of a circular icon on the left and the word "ULTROMICS" in bold, dark blue letters on the right. The circular icon is a gradient of blue and green, creating a swirling effect.

Both the subject and predicate devices have similar technological characteristics and principles of operation. Any technological differences between the subject and predicate devices raise no new concerns with regards to safety and efficacy.

Consensus Standards 7

The following consensus standards were used in the design and manufacture of EchoGo Core 2.0.

StandardRecognitionNumber
ISO 14971:2019 – Medical Devices – Application of Risk Management to Medical Devices5-125
IEC 62304:2015 – Medical Device Software - Software Life Cycle Processes13-79
IEC 62366-1:2020 – Medical Devices – Application of Usability Engineering to MedicalDevices5-129
NEMA PS 3.1 – PS 3.20 (2016) – Digital Imaging and Communications in Medicine (DICOM)Set12-300
IEC ISO 10918-1:1994 – Digital Compression and Coding of Continuous-tone Still Images12-261

Ultromics Limited cites conformity to the voluntary standards above. In addition, EchoGo Core 2.0 was designed and manufactured under a QMS that fully conforms to ISO 13485:2016.

8 Performance Data

8.1 Software Verification and Validation

EchoGo Core software was developed and tested in accordance with Ultromics' Design Control processes and has been subjected to extensive safety and performance testing. Non-clinical verification and validation test results established that the device meets its design requirements and intended use. Specifically, software verification was conducted at unit, module, and system integration levels. Risk management analysis generated multiple risk mitigation measures and verification activities. Formative usability assessments were conducted during formal development. Cybersecurity and data security testing were conducted to verify that data and patient protected health information security measures are included in the design of the software.

Software verification and validation testing were conducted, and documentation was provided as recommended by FDA's Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices. The software for this device is considered as a moderate level of concern, since a failure or latent design flaw could indirectly result in minor injury to the patient through incorrect or delayed information or through the action of a care provider.

510(k) SUMMARY

{8}------------------------------------------------

EchoGo Core 2.0 passed all software verification and validation tests.

Essential Performance 8.2

Performance testing was conducted to demonstrate substantial equivalence to the predicate device, EchoGo Core 1.0, via a reference comparator device, TomTec Arena TTA2. Pre-determined acceptance criteria were formalised in terms of Root Mean Square (RMS) error against reference values generated using the comparator device, for each reported metric. The acceptance criteria were formalised such that EchoGo Core 2.0 would produce measures of left ventricular (LV) length, volume at end diastole (ED), and end systole (ES), ejection fraction (EF), stroke volume, cardiac output, global longitudinal strain (GLS) and segmental longitudinal strain (SLS) with an RMS error below a set, pre-determined threshold.

A formal retrospective, non-interventional validation study was conducted using 214 previously unseen studies. Left ventricle volume and metrics subsequently derived, such as ejection fraction, were calculated. For global longitudinal strain and segmental strain at peak and end-systole, LV contour Euclidean distances were calculated. The performance dataset used was representative of the intended patient population over a clinically representative range of measures. The population was gender-balanced over clinically typical age, weight, and height ranges. Performance testing included acquisitions from multiple ultrasound system manufacturers to validate essential performance across multiple scanner platforms and settings. Test datasets were strictly segregated from algorithm training datasets. Root mean square error was calculated as the primary endpoint. The primary endpoint was met. Performance against the comparator device is summarised as follows.

Left Ventricular Metric% Root Mean Square Error
Length3.06 - 4.59
Volume at End Diastole and End Systole8.57 - 16.59
Ejection Fraction6.69 - 8.50
Stroke Volume10.57 - 13.68
Global Longitudinal Strain3.36 - 4.79
Systolic Segmental Longitudinal Strain5.51 - 9.98

Three operators processed a total of 214 TTE studies, respectively. Inter-operator variability demonstrated acceptable performance with a Deming RMSE of 0 for all report metrics. Similarly, intra-operator variability demonstrated acceptable performance with a Deming RMSE of 0 for all report metrics. The repeatability results for EchoGo Core 2.0 demonstrated acceptable performance.

All measurements produced by EchoGo Core 2.0 were deemed to be substantively equivalent to the predicate device and met acceptable levels of performance. We therefore consider EchoGo Core 2.0 to be substantively equivalent to the predicate device and is therefore deemed to be safe and effective.

9 Conclusions

The subject device, EchoGo Core 2.0 is as safe and as effective as the predicate device, EchoGo Core 1.0, previously cleared under K191171. The subject and predicate devices have identical intended uses and indications for use, and similar technological characteristics and principles of operation. Any minor differences raise no new concerns with regards to safety and effectiveness. Essential performance of both the subject and predicate devices were evaluated against a comparator reference device, TomTec Arena TTA2, previously

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for ULTROMICS. The logo consists of a circular graphic on the left and the word "ULTROMICS" in bold, dark blue letters on the right. The circular graphic is a gradient of blue and green, creating a sense of depth and movement.

cleared under K150122. Performance data demonstrate that EchoGo Core 2.0 and EchoGo Core 1.0 have similar essential performance.

It follows that EchoGo Core 2.0 is substantially equivalent to EchoGo Core 1.0.

§ 892.2050 Medical image management and processing system.

(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).