K Number
K213228

Validate with FDA (Live)

Date Cleared
2022-06-20

(264 days)

Product Code
Regulation Number
882.4305
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

NeuroLine perforators are designed to make holes in the skull bone. The perforators have an integrated safety mechanism that releases automatically after passing through the bone and stops drilling.

Device Description

The NeuroLine Disposable Cranial perforators (NLO-x/y-z) are single-use cutting devices that are packaged sterile and are used to perforate the skull bone. The perforator functions automatically, which means that it employs a clutch mechanism to automatically disenqage once perforation is accomplished and as the drill ceases to find resistance to bone.

The perforator has Hudson type connection and can connect to all the driver units with Hudson connection.

Perforators are offered in three different sizes: 6/9 mm, 7/13 mm and 11/14 mm. Sizes indicate hole diameters to be made by perforators in millimeters.

Perforators are also differentiated based on the thickness of skull bone. The NLO-x/y-3.0 perforators are used on skulls at least 3 mm thick, whereas NLO-x/y-1.5 ones are used on skulls at least 1.5 mm thick.

AI/ML Overview

The provided text describes the regulatory clearance for a medical device, the NeuroLine Disposable Cranial Perforator with Hudson End, and details the performance testing conducted to demonstrate its substantial equivalence to predicate devices. However, this document is a 510(k) summary for a mechanical device (a surgical drill), not an AI/ML-driven device or diagnostic tool.

Therefore, the information required to answer your questions about acceptance criteria and studies proving the device meets them, specifically in the context of AI/ML performance, ground truth establishment, expert adjudication, and MRMC studies, is not present in the provided text.

The document discusses:

  • Acceptance Criteria (for a mechanical device): These are implicitly met by "Pass" results in the performance testing section. The specific criteria are related to Biocompatibility (ISO 10993), Sterilization and Packaging validation (ISO 11137-1/2, ISO 11607-1/2), and Bench Testing for automatic stop function, maximum number of uses, drilling time, and device geometry.

  • Study Proving Device Meets Criteria: This refers to the "Performance Testing" section:

    • Biocompatibility: Tests conducted in accordance with ISO 10993.
    • Sterilization and Packaging Validation: Tests conducted in accordance with ISO 11137-1, ISO 11137-2, ISO 11607-1, and ISO 11607-2.
    • Bench Testing: Evaluated automatic stop function mechanism, maximum number of uses, drilling time, and device geometry.

None of the following questions can be answered from the provided text as they pertain to AI/ML device evaluations, which is not the subject of this document:

  1. A table of acceptance criteria and the reported device performance (for an AI/ML device): The device is not an AI/ML device. The "Pass" results are reported for mechanical and biological performance.
  2. Sample sizes used for the test set and the data provenance: Not relevant for this mechanical device.
  3. Number of experts used to establish the ground truth... and qualifications: Not relevant. No "ground truth" in the AI/ML sense is established. Performance is measured against mechanical specifications.
  4. Adjudication method for the test set: Not relevant.
  5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done: Not relevant.
  6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not relevant.
  7. The type of ground truth used: Not relevant in the AI/ML context. Performance against mechanical specifications is the "truth."
  8. The sample size for the training set: Not relevant. There is no AI/ML training set.
  9. How the ground truth for the training set was established: Not relevant.

In summary, the provided document is about a hardware medical device (a cranial perforator) and its regulatory clearance based on mechanical, sterility, and biocompatibility testing, not an AI/ML or diagnostic software device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

June 20, 2022

EMD Endoszkop Muszer Gyarto es Kereskedelmi Kft. Gergo Ujvari Export Manager Bartok Bela u. 113/B Debrecen, Hajdu-Bihar 4031, Hungary

Re: K213228

Trade/Device Name: NeuroLine Disposable Cranial Perforator with Hudson end Regulation Number: 21 CFR 882.4305 Regulation Name: Powered Compound Cranial Drills, Burrs, Trephines, And Their Accessories Regulatory Class: Class II Product Code: HBF Dated: May 13, 2022 Received: May 16, 2022

Dear Gergo Ujvari:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Adam D. Pierce, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K213228

Device Name

NeuroLine Disposable Cranial Perforator with Hudson End

Indications for Use (Describe)

NeuroLine perforators are designed to make holes in the skull bone. The perforators have an integrated safety mechanism that releases automatically after passing through the bone and stops drilling.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

Date Prepared: June 14, 2022

Applicant:

EMD Kft.

Bartok Bela u. 113/B Debrecen, Hungary 4031

Tel .: +36-52-486-034

Fax.: +36-52-486-035

Contact Person:

Gergo Ujvari – International Sales Manager

E-mail: gergo.ujvari@emd.hu

Trade Name: NeuroLine Disposable Cranial Perforator with Hudson End

Common Name: Disposable Cranial Perforator

Classification name: 21CFR section 882.4305

Product Code: HBF

Device Classification: Class II

Predicate Device:

ACRA-CUT Standard and Disposable Cranial perforator

Reference Device:

EasyDrill Autostop Cranial Perforator

Device Description:

The NeuroLine Disposable Cranial perforators (NLO-x/y-z) are single-use cutting devices that are packaged sterile and are used to perforate the skull bone. The perforator functions automatically, which means that it employs a clutch mechanism to automatically disenqage once perforation is accomplished and as the drill ceases to find resistance to bone.

The perforator has Hudson type connection and can connect to all the driver units with Hudson connection.

Perforators are offered in three different sizes: 6/9 mm, 7/13 mm and 11/14 mm. Sizes indicate hole diameters to be made by perforators in millimeters.

Perforators are also differentiated based on the thickness of skull bone. The NLO-x/y-3.0 perforators are used on skulls at least 3 mm thick, whereas NLO-x/y-1.5 ones are used on skulls at least 1.5 mm thick.

K833266

K141455

{4}------------------------------------------------

Image /page/4/Figure/1 description: This image shows a table with specifications for different models of tools. The table lists the models NLO-6/9-1.5, NLO-7/11-1.5, NLO-9/13-1.5, NLO-11/14-1.5, NLO-6/9-3.0, NLO-7/11-3.0, NLO-9/13-3.0, and NLO-11/14-3.0. For each model, the table provides the dimensions A, B, and C in millimeters, as well as the application for which the tool is intended. The tools with 1.5 in the model name are for use on skulls at least 1.5mm thick, while the tools with 3.0 in the model name are for use on skulls at least 3mm thick.

Indications for Use:

NeuroLine perforators are designed to make holes in the skull bone. The perforators have an integrated safety mechanism that releases automatically after passing through the bone and stop drilling.

Comparison of technology characteristics

DeviceNeuroLine DisposableCranial Perforator withHudson EndACRA-CUT Standard andDisposable CranialPerforatorEasyDrill Autostop CranialPerforator
ManufacturerEMD Endoszkop Muszer Gyartoes Kereskedelmi Kft.Acra-Cut Inc.Micromar Industria e ComercioLtda.
510(k) numberK213228(Subject Device)K833266(Predicate Device)K141455(Reference Device)
Specified holesizes6/9 mm, 7/11 mm, 9/13mm,11/14 mmSimilar6/9 mm, 7/11 mm, 11/14 mm
Indication foruseNeuroLine perforators aredesigned to make holes inthe skull bone. Theperforators have anintegrated safety mechanismthat releases automaticallyafter passing through thebone and stops drilling.SimilarThe EasyDrill Cranial Perforator is asterile, single use cutting deviceintended for performing cranialbone trephination.
OperatingprinciplesSameSameSame
Device MaterialStainless steelSimilarStainless steel

{5}------------------------------------------------

SterilizationSterile(Gamma Radiation)SterileSterile(Gamma Radiation)
Reusable/Single UseSingle UseSimilarSingle Use

Based on the information summarized above, the proposed NeuroLine Disposable Cranial Perforator and with Hudson end are similar to the currently marketed predicate and reference devices in indications for use, design, principle of operation and device material. All perforators are offered sterile.

Performance Testing

The following performance data were provided in support of the substantial equivalence determination:

TestMethodResults
BiocompatibilityThe device is classified as an externally communicatingdevices in contact with tissue/bone for a limited duration(≤24hr). Biocompatibility testing was addressed inaccordance with ISO 10993 as described in FDA'sBiocompatibility Guidance.Pass
Sterilization and packagingvalidation testingSterilization validation testing in accordance with ISO11137-1, ISO 11137-2, ISO 11607-1 and ISO 11607-2.Pass
Bench TestingThe performance of the subject device was evaluated forautomatic stop function mechanism, maximum number ofuses, drilling time and device geometry for establishingsubstantial equivalency to the predicate.Pass

Conclusions

Based on the similarities of the intended use/indications for use, technological and functional characteristics, and the results of the non-clinical performance testing, the NeuroLine Disposable Cranial Perforator with Hudson end, are substantially equivalent to the predicate device, ACRA-CUT Standard and Disposable Cranial perforator.

§ 882.4305 Powered compound cranial drills, burrs, trephines, and their accessories.

(a)
Identification. Powered compound cranial drills, burrs, trephines, and their accessories are bone cutting and drilling instruments used on a patient's skull. The instruments employ a clutch mechanism to disengage the tip of the instrument after penetrating the skull to prevent plunging of the tip into the brain.(b)
Classification. Class II (performance standards).