K Number
K213199

Validate with FDA (Live)

Manufacturer
Date Cleared
2022-01-11

(104 days)

Product Code
Regulation Number
888.3040
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Asnis® III Cannulated Screw System intended for fracture fixation of small and long bones and of the pelvis. The system is not intended for spinal use.
The Asnis® PRO Cannulated Screw System intended for fracture fixation of small and long bones and of the pelvis. The system is not intended for spinal use.

Device Description

The Asnis® III Cannulated Screw System, previously cleared in K000080, K024060, consists of self-tapping cannulated screws and the corresponding washers. All devices in the system are provided sterile and non-sterile. The thread diameters are 4.0, 5.0. 6.5, and 8.0 mm. They are either fully or partially threaded. All screws are self-drilling and self-tapping. There are corresponding washers to the 4.0 and 5.0 mm screws respectively and one washer fitting for both diameters, 6.5 and 8.0 mm. Screws and washers are made of stainless steel and titanium alloy.
The Asnis® PRO Cannulated Screw System consists of self-tapping cannulated screws. All devices in the system are provided sterile and non-sterile. The thread diameters are 6.5, and 8.0 mm. They are either fully or partially threaded. All screws are self-drilling and self-tapping. Screws are made of stainless steel and titanium alloy.
The subject of this bundled traditional submission is to introduce minor design specification changes; MRI Labeling, shelf-life update of existing package to 10 years, and re-branding Ø6.5/8.0 mm long screws (130-180 mm to Asnis® PRO Cannulated Screw System) to the Asnis® III Cannulated Screw System, cleared in K000080 and K024060 and to introduce a new system Asnis® PRO Cannulated Screw System.

AI/ML Overview

The provided document is a 510(k) Summary for the Asnis® III Cannulated Screw System and Asnis® PRO Cannulated Screw System. It outlines the reasons for the submission and compares the new devices to predicates. However, it explicitly states there was no clinical data or clinical trials needed for this device submission. The changes made (minor design specification changes, MRI labeling, shelf-life update, and re-branding) did not create a new "worst case" scenario that would necessitate mechanical testing beyond what was already performed for the predicate device.

Therefore, the following information regarding acceptance criteria and studies to prove device performance cannot be extracted from this document:

  1. A table of acceptance criteria and the reported device performance: No new performance data is presented, as the document states "No mechanical testing was deemed necessary."
  2. Sample size used for the test set and the data provenance: No new test set was used for mechanical performance, and no clinical data was collected.
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable as there was no test set requiring expert ground truth.
  4. Adjudication method for the test set: Not applicable.
  5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is a medical device (screws), not an AI-assisted diagnostic tool.
  6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
  7. The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not applicable.
  8. The sample size for the training set: Not applicable.
  9. How the ground truth for the training set was established: Not applicable.

What the document does state regarding performance:

  • Performance Data (Nonclinical):
    • "No mechanical testing was deemed necessary as the design modifications and the new system introduction do not create a new worst case. All bench tests performed in accordance with ASTM F543 and previously presented in Asnis® III Cannulated Screw System (K024060), remain true and accurate."
    • Tests performed to establish compatibility with a magnetic resonance environment:
      • Magnetically Induced Displacement per ASTM F2052
      • Magnetically Induced Torque per ASTM F2213
      • RF Heating per ASTM F2182
      • Image Artifacts per ASTM F 2119
  • Clinical Performance and Conclusions: "Clinical data and conclusions were not needed for this device."

In summary, this 510(k) submission relies on the substantial equivalence to its predicate devices and the existing non-clinical performance data for those predicates (K000080, K024060, and K140891), along with new MRI compatibility testing for the modified/new devices. No new trials or studies to establish performance against acceptance criteria in a traditional sense (e.g., diagnostic accuracy, clinical outcomes) were conducted or deemed necessary for this submission.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

January 11, 2022

Stryker GmbH Danielle Jannuzzi Madureira Staff Regulatory Affairs Specialist 325 Corporate Drive Mahwah, New Jersey 07430

Re: K213199

Trade/Device Name: Asnis® III Cannulated Screw System and Asnis® PRO Cannulated Screw System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: HWC. HTN Dated: December 17, 2021 Received: December 20, 2021

Dear Danielle Jannuzzi Madureira:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Shumaya Ali, MPH Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K213199

Device Name Asnis® III Cannulated Screw System

Indications for Use (Describe)

The Asnis® III Cannulated Screw System intended for fracture fixation of small and long bones and of the pelvis. The system is not intended for spinal use.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

| X | Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Indications for Use

510(k) Number (if known)

K213199

Device Name Asnis® PRO Cannulated Screw System

Indications for Use (Describe)

The Asnis® PRO Cannulated Screw System intended for fracture fixation of small and long bones and of the pelvis. The system is not intended for spinal use.

Type of Use (Select one or both, as applicable)

X | Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510 K213199

510(k) Summary

Proprietary Name:Asnis® III Cannulated Screw System and Asnis® PRO Cannulated ScrewSystem
Common Name:Bone Screw (primary)Washer, Bolt Nut
Regulation Description:Smooth or threaded metallic bone fixation fastener (primary)Single/multiple component metallic bone fixation appliances and accessories
Regulation Number:21 CFR 888.3040 (primary), 21 CFR 888.3030
Classification Product Code:HWC (primary), HTN
Device Class:II
Sponsor:Stryker GMBHBohnackerweg 12545 Selzach, Switzerland
Contact Person:Danielle Jannuzzi Madureira, PhDStaff Regulatory Affairs SpecialistDr. Homer Stryker Strasse, 1CH-2545 Selzach, SwitzerlandPhone: +41 79 890 02 89Fax: +41 32 641 66 60
Date:September 24, 2021
Primary Predicate Device:Asnis® III Cannulated Screw System (K000080)
Additional Predicate Device:Biomet Cannulated Screw System (K140891)
Device Description:The Asnis® III Cannulated Screw System, previously cleared in K000080,K024060, consists of self-tapping cannulated screws and the correspondingwashers. All devices in the system are provided sterile and non-sterile. Thethread diameters are 4.0, 5.0. 6.5, and 8.0 mm. They are either fully orpartially threaded. All screws are self-drilling and self-tapping. There arecorresponding washers to the 4.0 and 5.0 mm screws respectively and onewasher fitting for both diameters, 6.5 and 8.0 mm. Screws and washers aremade of stainless steel and titanium alloy.
The Asnis® PRO Cannulated Screw System consists of self-tappingcannulated screws. All devices in the system are provided sterile and non-sterile. The thread diameters are 6.5, and 8.0 mm. They are either fully orpartially threaded. All screws are self-drilling and self-tapping. Screws aremade of stainless steel and titanium alloy.
The subject of this bundled traditional submission is to introduce minordesign specification changes; MRI Labeling, shelf-life update of existingpackage to 10 years, and re-branding Ø6.5/8.0 mm long screws (130-180 mmto Asnis® PRO Cannulated Screw System) to the Asnis® III CannulatedScrew System, cleared in K000080 and K024060 and to introduce a newsystem Asnis® PRO Cannulated Screw System.
Indications for Use:The Asnis® III Cannulated Screw System intended for fracture fixation ofsmall and long bones and of the pelvis. The system is not intended for spinaluse.The Asnis® PRO Cannulated Screw System intended for fracture fixation ofsmall and long bones and of the pelvis. The system is not intended for spinaluse.
Comparison to PredicateDevice:The intended use of the modified devices, as described in its labeling, has notchanged because of the modifications proposed in the present submission.The introduction of minor design specification changes, MRI Labeling, shelf-life update of existing package to 10 years, nor re-branding Ø6.5/8.0 mm longscrews (130-180 mm) to Asnis® PRO Cannulated Screw System nor the newsystem (Asnis® PRO Cannulated Screw System) introduction does not alterthe fundamental scientific technology shared by both the subject devices,Asnis® III Cannulated Screw System and Asnis® PRO Cannulated ScrewSystem, and predicate device, Asnis® III Cannulated Screw System.
Performance Data (Nonclinical):Non-Clinical Performance and Conclusions:No mechanical testing was deemed necessary as the design modifications andthe new system introduction do not create a new worst case. All bench testsperformed in accordance with ASTM F543 and previously presented inAsnis® III Cannulated Screw System (K024060), remain true and accurate.Tests performed to establish compatibility with a magnetic resonanceenvironment:Magnetically Induced Displacement per ASTM F2052 Magnetically Induced Torque per ASTM F2213 RF Heating per ASTM F2182 Image Artifacts per ASTM F 2119
Clinical Performance and Conclusions:Clinical Performance and Conclusions:Clinical data and conclusions were not needed for this device.
Conclusion:The subject devices have the same intended use and indications for use as thepredicate device. The subject devices use the same operating principle,incorporate the same basic design and labeling and are manufactured andsterilized using the same materials and processes as the predicate device.Except for the modifications described in this submission the subject devicesare identical to the predicate device, and the performance data and analysesdemonstrate that:any differences do not raise new questions of safety and

{5}------------------------------------------------

  • enectiveness as established with performance results; and
    the subject devices are at least as safe and effective as the legally
    marketed predicate devices •

§ 888.3040 Smooth or threaded metallic bone fixation fastener.

(a)
Identification. A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.