K Number
K213135

Validate with FDA (Live)

Date Cleared
2022-03-02

(156 days)

Product Code
Regulation Number
876.4300
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Plasma Edge System single use bipolar resection electrodes are used for the ablation and hemostasis of tissues under endoscopic control, in association with endoscopic accessories. They are intended for endoscopic surgeries with saline irrigation, in the field of urology. The use of the Plasma Edge System is restricted to surgeons, specialized in urological surgery, for specific use in: - Transurethral resection of prostate (TURP) for benign prostatic hypertrophy - Transurethral incision of the prostate (TUIP) or bladder neck - Transurethral resection of bladder tumors (TURBT) - Cystodiathermy • Transurethral Vaporization of the prostate (TUVP/TVP) for benign prostatic hypertrophy, and for Transurethral Vaporization of bladder tumors. (MVVS and MVV models only)

Device Description

The Plasma Edge system is a manual surgical device, consisting of a single-use electrode with cable range, an active and passive working element reusable and an adaptor to an HFgenerator compatible. The electrodes consist of an active tip, two wires threaded through ceramic tubes to connect the active tip to the body of the loop, allowing the HF energy to reach the active tip. It has to be used with continuous flow irrigation of saline solution (NaCl 0,9%) that reachesthe operative site through a resectoscope. The HF energy delivered from the generator tothe electrode ionizes the gas of the saline solution, creating a plasma for the cutting, coagulation and vaporization of tissues.

AI/ML Overview

The provided text is a 510(k) Summary for the Plasma EDGE system, which is an endoscopic electrosurgical unit and accessories. It focuses on demonstrating substantial equivalence to a predicate device rather than providing a detailed study of the device's performance against specific acceptance criteria for diagnostic tasks.

Therefore, much of the requested information regarding diagnostic accuracy studies, sample sizes for test and training sets, expert qualifications, and ground truth establishment is not present in this document. The document describes engineering and safety testing.

Here's the information that can be extracted or deduced from the document:

1. A table of acceptance criteria and the reported device performance

The document lists performance tests conducted to ensure the device functions as intended and meets design specifications, based on in-house acceptance criteria derived from ISO-14971:2007 (Risk analysis) and various other standards. Specific quantitative acceptance criteria or reported performance results (e.g., in terms of measurements or thresholds) are not provided in detail in this summary, other than stating that the tests were completed successfully.

Acceptance Criteria CategoryApplied Standard(s)Reported Device Performance
Risk AnalysisISO-14971:2007Carried out in accordance with established in-house acceptance criteria.
BiocompatibilityISO 10993-1:2018, ISO 10993-5, ISO 10993-11, ISO 10993-10Evaluation conducted. Testing included cytotoxicity, acute systemic toxicity, intracutaneous irritation, and sensitization tests. (Results are not specified, only that tests were done).
Electrical SafetyIEC 60601-1:2012, IEC 60601-2-2:2009Tested by an independent laboratory.
Electromagnetic Compatibility (EMC)IEC 60601-1-2:2007Tested by an independent laboratory. Device (Plasma Edge electrode) tested on HF generator Gyrus, with comparison to predicate.
Cleaning & Sterilization Validation (Reusable Working Element)ISO 17664, AAMI TIR N°30, AAMI TIR N°12, ISO 17665Tested by an independent laboratory. Cleaning, disinfection, and steam sterilization (132°C for 4 min) validated.
Sterilization Validation & Shelf-Life (Single Use Electrode)ISO 11135, ISO 11607-1, ISO 11607-2, ISO 11737-1Ethylene oxide sterilization method validated. Shelf life of one (1) year.
Bench Top Validation Testing(Implicit in general performance goals)End of life simulation report, breakdown simulation report, working element compatibility report and test on ex vivo tissues have been tested. Demonstrated product safety and efficiency.

Study that Proves the Device Meets Acceptance Criteria:

The document refers to a series of performance tests and usability studies conducted to ensure the system functions as intended and meets design specifications. These studies are collectively referred to as "Performance testing" (Section iii).

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

This information is not provided in the document. The studies mentioned are engineering and safety tests, not clinical studies involving patient data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not applicable and therefore not provided, as the tests described are technical validations (biocompatibility, electrical safety, sterilization, bench-top) rather than clinical evaluations requiring expert-derived ground truth for diagnostic or therapeutic accuracy.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable, as this is related to expert review of clinical data, which is not described.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is not a diagnostic AI device, but an electrosurgical system.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

Not applicable. This is not an AI algorithm. The performance tests are focused on the device's technical and physical characteristics.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For the technical performance aspects, "ground truth" would be established through adherence to recognized standards (e.g., ISO, IEC), material specifications, and validated processes. For example:

  • Biocompatibility: Results of standardized in-vitro and in-vivo tests according to ISO 10993 series.
  • Electrical Safety/EMC: Compliance with IEC 60601 series, measured electrical parameters within specified limits.
  • Sterilization: Demonstrated sterility assurance level (SAL) according to ISO 11135 and other related standards.
  • Bench-top testing: Physical measurements, visual inspection, and functional verification against design specifications and predicate performance (e.g., cutting efficacy on ex vivo tissues).

8. The sample size for the training set

Not applicable, as this is not an AI/machine learning device.

9. How the ground truth for the training set was established

Not applicable, as this is not an AI/machine learning device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym and name on the right. The FDA part of the logo is in blue, with the acronym "FDA" in a square and the words "U.S. FOOD & DRUG ADMINISTRATION" next to it.

March 2, 2022

Chirurgie Innovation Guillaume Noury CEO Z.A. Les Godets, 3 Rue Des Petits Ruisseaux Verrière le Buisson, 91370 France

Re: K213135

Trade/Device Name: Plasma EDGE system Regulation Number: 21 CFR§ 876.4300 Regulation Name: Endoscopic Electrosurgical Unit and Accessories Regulatory Class: II Product Code: FAS, GEI, FDC Dated: January 27, 2022 Received: January 31, 2022

Dear Guillaume Noury:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Mark J. Antonino, M.S. Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K213135

Device Name

Plasma EDGE system

The Plasma Edge System single use bipolar resection electrodes are used for the ablation and hemostasis of tissues under endoscopic control, in association with endoscopic accessories.

They are intended for endoscopic surgeries with saline irrigation, in the field of urology.

The use of the Plasma Edge System is restricted to surgeons, specialized in urological surgery, for specific use in:

  • · Transurethral resection of prostate (TURP) for benign prostatic hypertrophy
  • · Transurethral incision of the prostate (TUIP) or bladder neck
  • · Transurethral resection of bladder tumors (TURBT)
  • · Cystodiathermy

• Transurethral Vaporization of the prostate (TUVP/TVP) for benign prostatic hypertrophy, and for Transurethral

Vaporization of bladder tumors. (MVVS and MVV models only)

Type of Use (Select one or both, as applicable)
---------------------------------------------------

× Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995,

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

Chirurgie Innovation
SubmitterZ.A. Les Godets, 3 rue des Petits RuisseauxF-91370 Verrières le Buisson FRANCE
ContactsGuillaume NouryCEOregulatory@orange.frPhone : +33 169 20 30 21Fax : +33 160 13 97 47
Date01/27/2022
Trade NamePlasma EDGE system
Common nameBipolar electrodes
Classification NameEndoscopic Electrosurgical Unit and Accessories
ClassII
Product CodeClassification product code : FASSubsequent product code : GEI / FDC
CFR section876.4300 / 878.4400 / 876.1500
Device panelRegulation Medical Specialty : Gastroenterology / Urology510k Review Panel : General & Plastic Surgery
Legally marketedpredicate devicesK163090: Plasma EDGE System - manufactured by ChirurgieInnovation.

Product Description .. "

The Plasma Edge system is a manual surgical device, consisting of a single-use electrode with cable range, an active and passive working element reusable and an adaptor to an HFgenerator compatible.

The electrodes consist of an active tip, two wires threaded through ceramic tubes to connect the active tip to the body of the loop, allowing the HF energy to reach the active tip.

It has to be used with continuous flow irrigation of saline solution (NaCl 0,9%) that reachesthe operative site through a resectoscope. The HF energy delivered from the generator tothe electrode ionizes the gas of the saline solution, creating a plasma for the cutting, coagulation and vaporization of tissues.

{4}------------------------------------------------

a. Electrodes range:

The list of electrodes is divided in 2 ranges: SIDE LOAD and FRONT LOAD.

The SIDE LOAD range is composed by 3 electrodes with 3 actives tips: thin, thick or vaporization. The SIDE LOAD electrodes are dedicated to being used with working element compatible with STORZ and OLYMPUS resectoscope.

The FRONT LOAD range is composed by 3 electrodes with 3 actives tips: thin, thick or vaporization. The FRONT LOAD electrodes are dedicated to being used with working element compatible with OLYMPUS resectoscope.

Each range got is specific way to be assembly into the working element. The only physical difference between both ranges is the electrode cable connexion.

End user can be used to both versions, so we decided to offer the choice between them.

b. Working element range:

The working elements will be divided in 2 families: side load connection and front load connection. According to the working element families, the compatibility with resectoscope brands differs:

  • SIDE LOAD connection: compatible with STORZ and OLYMPUS resectoscope.
  • FRONT LOAD connection: compatible with OLYMPUS resectoscope.

Electrodes are mechanically connected into a working element, used to move the electrode by a linear translation.

  • With an active working element, action is pulling the electrode
  • With a passive working element, action is pushing the electrode

c. Adaptor:

This will allow to connect the Plasma EDGE electrode range to a bipolar HF generator system.

{5}------------------------------------------------

Indications for use ii.

The Plasma Edge System single use bipolar resection electrodes are used for the ablation and hemostasis of tissues under endoscopic control, in association with endoscopic accessories.

They are intended for endoscopic surgeries with saline irrigation, in the field of urology. The use of the Plasma Edge System is restricted to surgeons, specialized in urological surgery, for specific use in:

  • Transurethral resection of prostate (TURP) for benign prostatic hyperplasia
  • . Transurethral incision of the prostate (TUIP) or bladder neck
  • . Transurethral resection of bladder tumors (TURBT)
  • . Cystodiathermy
  • Transurethral Vaporization of the prostate (TUVP/TVP) for benign prostatic hyperplasia, and for Transurethral Vaporization of bladder tumors. (MVVS; MVV models only)

iii. Performance testing

a. Risk analysis:

Risk analysis was carried out in accordance with established in-house acceptance criteria based on ISO-14971:2007

Performance tests were carried out to ensure that the system functions as intended and meets design specifications. The following performance tests and usability studies were conducted.

b. Biocompatibility testing

The biocompatibility evaluation for the subject device was conducted in accordance with the standard ISO 10993-1:2018. The testing included the following tests:

  • Rapport d'évaluation biologique_Résection (Biological evaluation report_Resection) .
  • . ISO MTS cytotoxicity test - ISO 10993-5 (2009): "Biological evaluation of medical devices – Part 5: Tests for in vitro cytotoxicity",
  • . ISO Acute Systemic Toxicity Study in Mice - Two Extracts - ISO 10993-11(2006): "Biological Evaluation of Medical Devices, Part 11: tests for systemic toxicity".
  • ISO intracutaneous Study in Rabbits Two Extracts ISO 10993-10 (2010): . "Biological evaluation of medical devices – Part 10: Tests for irritation and skin sensitization"

{6}------------------------------------------------

  • ISO Guinea Pig Maximization Sensitization Test Two Extracts ISO 10993-10 . (2010): "Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization"

c. Electrical safety

Electrical Safety was tested by an independent laboratory according to IEC 60601-1: 2012; Medical Electrical Equipment - Part 1.1 General requirements for safety and essential performance, and IEC 60601-2-2:2009- Medical Electrical Equipment - Part 2 Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories.

d. Electromagnetic compatibility (EMC)

Electromagnetic compatibility was tested by an independent laboratory according to the standard IEC 60601-1-2:2007. Electrode Plasma Edge was tested on the HF generator Gyrus and a comparison was realized with the assembly Electrode Gyrus on the HF generator Gyrus.

e. Cleaning and sterilization validation for the reusable working element range:

Cleaning and disinfection validation was tested by an independent laboratory following standards ISO 17664 and AAMI TIR N°30: manual cleaning.

Steam sterilization validation was tested by an independent laboratory following standards AAMI TIR N°12, ISO 17664 and ISO 17665: 132°c for 4 min.

f. Sterilization validation and Shelf-Life Discussion for single use electrode

The electrodes are delivered in a sterile state and are intended for single patient use only. The validated sterilization method used is ethylene oxide. The product has a shelf life of one (1) year.

g. Bench top validation testing

End of life simulation report, breakdown simulation report, working element compatibility report and test on ex vivo tissues have been tested on the Plasma Edge System to demonstrate the product safety and the efficiency.

{7}------------------------------------------------

iv. List of standards

StandardsDescription
ISO 10993-1Biological evaluation of medical devices - Part 1: Evaluation and testing withina risk management process
ISO 10993-5Biological evaluation of medical devices -- Part 5: Tests for In Vitro cytotoxicity
ISO 10993-7Biological evaluation of medical devices - part 7: ethylene oxide sterilizationresiduals [including: technical corrigendum 1 (2009)]
ISO 10993-10Biological evaluation of medical devices - Part 10: Tests for irritation and skinsensitization
ISO 10993-11Biological evaluation of medical devices - Part 11: Tests for systemic toxicity
ISO 11135Sterilization of health-care products - ethylene oxide - requirements for thedevelopment, validation and routine control of a sterilization process formedical devices
ISO 11607-1Packaging for terminally sterilized medical devices - Part 1: Requirements formaterials, sterile barrier systems and packaging systems
ISO 11607-2Packaging for terminally sterilized medical devices -Part 2: Validationrequirements for forming, sealing and assembly processes
ISO 17664Sterilization of medical devices - Information to be provided by themanufacturer for the processing of resterilizable medical devices
ISO 17665-1Sterilization of health care products - Moist heat - Part 1: Requirements forthe development, validation and routine control of a sterilization process formedical devices
AAMI TIR N°12Designing, testing, and labeling reusable medical devices for reprocessing inhealth care facilities: A guide for medical device manufacturers
AAMI TIR N°30A compendium of processes, materials, test methods, and acceptance criteriafor cleaning reusable medical devices
IEC 60601-1Medical electrical equipment - Part 1: General requirements for basic safetyand essential performance
IEC 60601-1-2General requirements for basic safety and essential performance - CollateralStandard: Electromagnetic disturbances- Requirements and tests
IEC 60601-1-6Medical electrical equipment - Part 1-6: General requirements for basic safetyand essential performance - Collateral standard: Usability
IEC 60601-2-2Medical electrical equipment - Part 2-2: Particular requirements for the basicsafety and essential performance of high frequency surgical equipment andhigh frequency surgical accessories
IEC 60601-2-18Medical electrical equipment - Part 2-18: Particular requirements for the basicsafety and essential performance of endoscopic equipment
IEC 62366-1Medical Electrical Equipment - Part 1
ISO 14971Medical devices - application of risk management to medical devices
ISO 15223-1Medical devices - symbols to be used with medical device labels, labeling, andinformation to be supplied - part 1 : general requirements
ISO 11737-1Sterilization of health care products - Microbiological methods - Part 1:Determination of a population of microorganisms on products

{8}------------------------------------------------

{9}------------------------------------------------

Proposed devicePlasma EDGE System (K213135)Predicate devicePlasma EDGE System (K163090)Substantial EquivalenceAnalysis
ELECTRODE
Indications foruse- Transurethral resection of prostate(TURP)- Transurethral incision of the prostate(TUIP) or bladder neck- Transurethral resection of bladdertumours (TURBT)- Cystodiathermy-Transurethral electrovaporization of theprostate- Transurethral resection of prostate(TURP)- Transurethral incision of the prostate(TUIP) or bladder neck- Transurethral resection of bladdertumours (TURBT)- Cystodiathermy-Transurethral electrovaporization of theprostateSame
SterilizationEtoEtoSame
Single useYesYesSame
Energy typeHigh frequencyHigh frequencySame
ModeBipolarBipolarSame
User interfaceFootswitchFootswitchSame
Use only inConductiveMediaThe electrode is to be activated onlywhen immersed in a conductive mediasuch as standard saline solutionThe electrode is to be activated onlywhen immersed in a conductive mediasuch as standard saline solutionSame
ElectrodemanipulationWorking elementWorking elementSame
WorkingelementcompatibilityStorzOlympusStorzOlympusSame
Sterilization ofworkingelementSteam sterilization(autoclave)Steam sterilization(autoclave)Same
Proposed devicePlasma EDGE System (K213135)Predicate devicePlasma EDGE System (K163090)Substantial EquivalenceAnalysis
MATERIALS OF ELECTRODES
Active tip ofthin and thickloop-WIRETungsten wire99,95%Tungsten wire99,95%Same
Active wire-WIREPlatinum iridiumPlatinum iridiumSame
InsulationCeramic ZTA10Ceramic ZTA10Same
Wire insulationPTFEAWG28PTFEAWG28Same
Heat-shrinktubingMT 3000MT 3000Same
SkeletonStainless Steel 316LStainless Steel 316LSame
Telescope clipPEEK 150GPEEK 150GSame
GlueLoctite M-121HPLoctite M-121HPSame
Sealing gasketSiliconeSiliconeSame
Inner insulatorPEEK 450GPEEK 450GSame
Connectorover-moulding(SIDE load)PolypropyleneISPLEN PP 080PolypropyleneISPLEN PP 080Same
Connectorover-moulding(FRONT load)PolypropyleneMultiflex G60 A11BPolypropyleneMultiflex G60 A11BSame
Proposed devicePlasma EDGE System (K213135)Predicate devicePlasma EDGE System (K163090)Substantial EquivalenceAnalysis
HF Cables ofthe electrodesSiliconeSiliconeSame
PlugTin-plated copperTin-plated copperSame
Crimp tubeTitaniumTitaniumSame
Packaging material of Electrode
Pouch material- One side coated TYVEK® 1073B® Co7502887C- One side PET/PE 60 Pharma (transparent)- One side PGL 110: 100g/m² MedicalKraft Paper / 10g/m² Grid Lacquer- One side PET/PEVA filmAny differences werevalidated and did not raiseany new risks.
Blister materialPETGAPET CRISTAL / CAROCLEAR MDL 50Any differences werevalidated and did not raiseany new risks.
Packaging process
PackagingMachine:Hawo machine Impulse heat sealer (sealingof the last side of the pouch)Maruani machine (Sealing of 3 sides of thepouch) HAWO scroll welding (Sealing of thelast side of the pouch)Any differences werevalidated and did notraise any new risks.
Sealingtemperature:127°C180°C (for both machine)Any differences werevalidated and did not raiseany new risks.
Sealing speed:sealing time = 1,5stotal times = 10s (sealing dies closing -opening)MARUANI:• Up to 1500 cycles/hourHawo:sealing speed: 10m/minAny differences werevalidated and did not raiseany new risks.
Contactpressure:115NThis parameter is not modified by LNM(only by HAWO)MARUANI:• non communicatedHawo: 100NAny differences werevalidated and did not raiseany new risks.

Comparison of Technological characteristics: V.

{10}------------------------------------------------

{11}------------------------------------------------

{12}------------------------------------------------

At a high level, the subject and predicate devices are based on the same technological principle with the same elements:

  • The indications for use (TURP; TUIP or bladder neck; TURBT; Cystodiathermy; Transurethral electrovaporization)
  • The method of use (compatibility working element, conductive media, etc...) .
  • . The material characteristics
  • The modification of the packaging material and packaging process did not change the ● technological characteristics, the design nor fundamental operating principles of modified device compared to the predicate. The modification of the packaging material and packaging process were validated and did not lead to the identification of any new risk.

Since the indications for use and the technological characteristics are the same between the predicate device and the modified device subject of this Special 510(k), the modified device is substantially equivalent to the predicate device.

Conclusion: vi.

There is no difference between the Plasma EDGE resection system and the predicate devices in terms of intended use, principle of operation, and the technology used for device performance based upon the changes reported in this Special 510(k). In other words, there is no difference technically, clinically and biologically from the predicate.

The Plasma EDGE resection system was subjected to verification testing to confirm device performance. There is no new technology and differences that would raise new or different questions of safety or efficiency. Comparative performance testing demonstrate that the device performed as well as, or better than, the predicate device.

§ 876.4300 Endoscopic electrosurgical unit and accessories.

(a)
Identification. An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.(b)
Classification. Class II (performance standards).