K Number
K212994

Validate with FDA (Live)

Date Cleared
2022-01-17

(119 days)

Product Code
Regulation Number
878.4040
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The medical face masks are intended to be worn to protect both the patient and healthcare persomel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device, provided non-sterile.

Device Description

The Medical Face Masks are single use, flat-pleated, ear loop masks that are provided in blue. The medical face masks are available in three types, which are Level 1, Level 2 and Level 3 based on ASTM F2100-19, depending on marketing reasons. However, Level 2 and Level 3 masks are exactly the same. The outer and inner layers of the mask are made of spunbond polypropylene. The middle filter layer is made of meltblown polypropylene filter. The ear loops are made of nylon and spandex. The ear loops are held in place over the users' mouth and nose by two elastic ear straps. The nose of PE and galvanized iron wire. Users can adjust the nose clip according to the shape of the nose and fix the mask on the bridge of the nose to prevent the mask from falling off. The proposed devices are provided in non-sterile.

AI/ML Overview

This document is a 510(k) summary for a Medical Face Mask (K212994) by Guangzhou Quantum Laser Intelligent Equipment Co., Ltd. It compares the proposed device to a predicate device (K160269) to demonstrate substantial equivalence.

Here's the breakdown of the acceptance criteria and the study that proves the device meets them:

1. Table of Acceptance Criteria and Reported Device Performance

The non-clinical performance characteristics evaluated, their acceptance criteria, and the reported results for the proposed device (Level 3 mask, as Level 1, 2, and 3 are physically the same) are presented in the following table:

Performance CharacteristicsTest MethodReported Device PerformanceAcceptance CriteriaConclusion
Fluid Resistance PerformanceASTM F1862/F1862M: 2017Pass at 160 mmHgNo penetration at 160 mmHgMeet the requirement
Particulate Filtration EfficiencyASTM F2299/F2299M-03 (2017)Pass at 99.2%≥98%Meet the requirement
Bacterial Filtration EfficiencyASTM F2101: 2019Pass at 99.2%≥98%Meet the requirement
Differential Pressure (Delta-P)EN 14683: 2019, Annex CPass at 3.8 mm H2O/cm²<6.0 mmH2O/cm²Meet the requirement
Flammability class16 CFR Part 1610Class 1Class 1Meet the requirement
Biocompatibility (Cytotoxicity)ISO 10993-5:2009Non-cytotoxicNon-cytotoxicMeet the requirement
Biocompatibility (Sensitization)ISO 10993-10:2010Non-sensitizingNon-sensitizingMeet the requirement
Biocompatibility (Irritation)ISO 10993-10:2010Non-irritatingNon-irritatingMeet the requirement

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the sample sizes used for each specific non-clinical test. However, it indicates that the tests were conducted on the "proposed device" which is a "Medical Face Mask." The provenance of the data is not explicitly stated as country of origin, but the submission is from Guangzhou Quantum Laser Intelligent Equipment Co., Ltd., China. The tests are non-clinical (laboratory-based performance tests) and are inherently retrospective in the sense that they are conducted on finished products for regulatory submission.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

Not applicable. These are non-clinical performance tests, not studies requiring expert interpretation of clinical data. The "ground truth" for these tests is defined by the standardized test methods (e.g., ASTM, EN, ISO standards) themselves, which specify objective measurements and criteria.

4. Adjudication Method for the Test Set

Not applicable. As described above, these are objective, standardized non-clinical tests, not subjective assessments requiring adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This document is for the clearance of a medical device (face mask) based on non-clinical performance tests and substantial equivalence to a predicate device, not for evaluating the comparative effectiveness of human readers with or without AI assistance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Yes, standalone non-clinical performance testing was done for the device. The results reported in the table (e.g., fluid resistance, filtration efficiency, differential pressure, flammability, biocompatibility) are all based on the performance of the device itself, without any human-in-the-loop component for evaluating its primary function.

7. The Type of Ground Truth Used

The ground truth used for these non-clinical tests is based on established industry standards and regulatory requirements. For example:

  • Fluid Resistance: No penetration at 160 mmHg, as per ASTM F1862/F1862M.
  • Particulate/Bacterial Filtration Efficiency: ≥98%, as per relevant ASTM standards (F2299/F2101) and ASTM F2100.
  • Differential Pressure: <6.0 mmH2O/cm², as per EN 14683 Annex C and ASTM F2100.
  • Flammability: Class 1, as per 16 CFR Part 1610.
  • Biocompatibility: Non-cytotoxic, non-sensitizing, non-irritating, as per ISO 10993-5 and ISO 10993-10.

8. The Sample Size for the Training Set

Not applicable. This is a medical device (face mask) submission, not an AI/ML algorithm requiring a training set.

9. How the Ground Truth for the Training Set Was Established

Not applicable. As this is not an AI/ML algorithm, there is no training set and therefore no ground truth establishment for a training set.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.

January 17, 2022

Guangzhou Quantum Laser Intelligent Equipment Co., Ltd. % Diana Hong General Manager Mid-Link Consulting Co., Ltd P.O. Box 120-119 Shanghai. 200120 China

Re: K212994

Trade/Device Name: Medical Face Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: May 18, 2021 Received: September 20, 2021

Dear Diana Hong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For Clarence W. Murray III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known)

K212994

Device Name Medical face mask

Indications for Use (Describe)

The medical face masks are intended to be worn to protect both the patient and healthcare persomel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device, provided non-sterile.

Type of Use (Select one or both, as applicable)

] Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section 807.92.

The assigned 510(k) Number: K212994

  • Date of Preparation: 12/23/2021 1.
    1. Sponsor Identification

Guangzhou Quantum Laser Intelligent Equipment Co., Ltd.

Building B27 Huachuang Animation Industrial Park, Jinshan Village, Shiqi Town, Panyu District, Guangzhou, Guangdong, China

Establishment Registration Number: 3016792608.

Contact Person: Hao Zheng Position: Business Manager Tel: +86-20-39969520 Email: 2281976947@qq.com

    1. Designated Submission Correspondent
      Ms. Diana Hong (Primary Contact Person) Ms. Jinlei Tang (Alternative Contact Person)

Mid-Link Consulting Co., Ltd.

P.O. Box 120-119, Shanghai, 200120, China

Tel: +86-21-22815850, Fax: 360-925-3199 Email: info@mid-link.net

{4}------------------------------------------------

4. Identification of Proposed Device

Trade Name: Medical Face Mask Common Name: Mask

Regulatory Information Classification Name: Mask, Surgical Classification: II; Product Code: FXX; Regulation Number: 21CFR 878.4040 Review Panel: General Hospital

Intended Use/Indications for Use:

The medical face masks are intended to be worn to protect and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device, provided non-sterile.

Device Description:

The Medical Face Masks are single use, flat-pleated, ear loop masks that are provided in blue. The medical face masks are available in three types, which are Level 1, Level 2 and Level 3 based on ASTM F2100-19, depending on marketing reasons. However, Level 2 and Level 3 masks are exactly the same. The outer and inner layers of the mask are made of spunbond polypropylene. The middle filter layer is made of meltblown polypropylene filter. The ear loops are made of nylon and spandex. The ear loops are held in place over the users' mouth and nose by two elastic ear straps. The nose of PE and galvanized iron wire. Users can adjust the nose clip according to the shape of the nose and fix the mask on the bridge of the nose to prevent the mask from falling off. The proposed devices are provided in non-sterile.

  • Identification of Predicate Device ર.
    510(k) Number: K160269 Product Name: Surgical Face Masks (Ear loops and Tie-on)

{5}------------------------------------------------

Summary of Technological characteristics 6.

ITEMProposed DevicePredicate Device K160269Remark
Product CodeFXXFXXSame
Regulation No.21 CFR 878.404021 CFR 878.4040Same
ClassIIIISame
IndicationsforUseThe medical face masks are intended to beworn to protect both the patient andhealthcare personnel from transfer ofmicroorganisms, body fluids andparticulate material. These face masks areintended for use in infection controlpractices to reduce the potential exposureto blood and body fluids. This is a singleuse, disposable device, provided non-sterile.The surgical face masks are intended to beworn to protect both the patient andhealthcare personnel from transfer ofmicroorganisms, body fluids, and particulatematerial. These face masks are intended foruse in infection control practices to reducethe potential exposure to blood and bodyfluids. This is a single-use, disposabledevice, provided non-sterile.Same
Mask styleFlat-pleatedFlat-pleatedSame
ASTMF2100LevelLevel 1, Level 2, Level 3Level 1, Level 2, Level 3Same
Design featureEar loopEar loop and Tie-onDifferent
ColorBlueBlue, WhiteDifferent
DimensionMask body: 175mm×95mmNose Clip: 100±5mmEar loop: 170±5mm175×90mmNose Clip: 100±5mmEar loop: 180×90mmDifferent
SterilityNon-SterileNon-SterileSame
UseSingle Use, DisposableSingle Use, DisposableSame
ParticulatefiltrationefficiencyPass at 99.2%Level 1: Pass at 99.6%Level 2: Pass at 99.6%Level 3: Pass at 99.7%Different
BacterialfiltrationefficiencyPass at 99.2%Level 1: Pass at 98%Level 2: Pass at 98%Level 3: Pass at 99%Different
DifferentialpressurePass at 3.8 mm H2O/cm²Level 1: Pass at 2.0 mmH2O/cm²Level 2: Pass at 1.6 mmH2O/cm²Level 3: Pass at 2.5 mmH2O/cm²Different
FlammabilityClass 1Class 1Same
Fluid resistancePass at 160 mmHgLevel 1: Pass at 80 mmHgLevel 2: Pass at 120 mmHgLevel 3: Pass at 160 mmHgDifferent
Label/LabelingComplied with 21 CFR part 801Complied with 21 CFR part 801Same
Patient Contacting Material
Outer facing layerSpunbond PolypropylenePolypropyleneDifferent
Middle layerMeltblown Polypropylene Filter1. Polypropylene spunbond2. Polypropylene meltblown
Inner facing layerSpunbond PolypropylenePolypropylene
Nose clipPE and Galvanized iron wirePolyethylene coated steel wireSame
Ear loopNylon and SpandexPolyester, polyurethane
Biocompatibility
CytotoxicityUnder the conditions of the study, the proposed device was non-cytotoxic.Under the conditions of the study, the subject device was non-cytotoxic.
SensitizationUnder the conditions of the study, the proposed device was non-sensitizing.Under the conditions of the study, the subject device was non-sensitizing.
IrritationUnder the conditions of the study, the proposed device was non-irritating.Under the conditions of the study, the subject device was non-irritating.

Table 1 Comparison of Medical Face Mask

{6}------------------------------------------------

Different - Design feature

The proposed masks are ear-loop masks. The design features of the proposed device can be covered by the design features of the predicate device. Thus, this difference will not affect the safety and effectiveness of the proposed device.

Different - Color

The proposed device is blue and the predicate device is provided in two colors, the proposed device can be covered by the predicate device. In addition, the biocompatibility test has been conducted and test results did not show any adverse effects. Thus, this difference will not affect the safety and effectiveness of the proposed device.

Different - Dimension

The dimension for the proposed device is different from the predicate device. However, the difference does not affect the indication for use and will not raise safety issues. Thus, this difference will not affect the safety and effectiveness of the proposed device.

Different - Particulate filtration efficiency

The test result for particulate filtration efficiency for the proposed device is different from the predicate

{7}------------------------------------------------

device. The proposed devices are available in Level 1, Level 3 based on ASTM F2100-19, depending on marketing reasons. However, Level 1, Level 3 masks are exactly the same. Therefore, the particulate filtration efficiency test was conducted on Level 3 masks and the test result can meet the requirements based on ASTM F2100-19. Thus, this difference will not affect the safety and effectiveness of the proposed device.

Different - Bacterial filtration efficiency

The test result for bacterial filtration efficiency for the proposed device is different from the predicate device. The proposed devices are available in Level 1, Level 3 based on ASTM F2100-19, depending on marketing reasons. However, Level 1, Level 2 and Level 3 masks are exactly the same. Therefore, the bacterial filtration efficiency test was conducted on Level 3 masks and the test result can meet the requirements based on ASTM F2100-19. Thus, this difference will not affect the safety and effectiveness of the proposed device.

Different - Differential pressure

The test result for differential pressure for the proposed device is different from the predicate device. The proposed devices are available in Level 1, Level 2 and Level 3 based on ASTM F2100-19, depending on marketing reasons. However, Level 1, Level 2 and Level 3 masks are exactly the same. Therefore, the differential pressure test was conducted on Level 3 masks and the test result can meet the requirements of the standard ASTM F2100-19 which was recognized by FDA. Thus, this difference will not affect the safety and effectiveness of the proposed device.

Different - Fluid resistance

The test result for fluid resistance for the proposed device is different from the predicate device. The proposed devices are available in Level 1, Level 2 and Level 3 based on ASTM F2100-19, depending on marketing reasons. However, Level 1, Level 2 and Level 3 masks are exactly the same. Therefore, the fluid resistance test was conducted on Level 3 masks and the test result can meet the requirements based on ASTM F2100-19. Thus, this difference will not affect the safety and effectiveness of the proposed device.

Different - Patient Contacting Material

The patient contacting material for the proposed device is different from the predicate device. However, biocompatibility test has been conducted on the proposed device and the result does not show any adverse effect. Thus, this difference will not affect the safety and effectiveness of the proposed device.

{8}------------------------------------------------

7. Summary of Non-Clinical Test

Non clinical tests were conducted to verify that the proposed device met all design specifications as was same/similar to the predicate devices. The test results demonstrated that the proposed device complies with the following standards:

  • 16 CFR Part 1610 Standard for the Flammability of Clothing Textiles;

  • ASTM F1862/F1862M: 2017 Standard Test Method for Resistance of Medical Face Masks to Penetration by Synthetic Blood (Horizontal Projection of Fixed Volume at a Known Velocity)

  • ASTM F2299/F2299M-03 (2017) Standard Test Method for Determining the Initial Efficiency of Material Used in medical Face Masks to Penetration by Particulates using Latex Spheres

  • ASTM F2101: 2019 Standard Test Method for Evaluating the Bacterial Filtration Efficiency (BFE) of Medical Face Mask Materials, Using a Biological Aerosol of Staphylococcus aureus

  • ASTM F2100: 2019 Standard Specification for Performance of Materials Used in Medical Face Masks

  • EN 14683: 2019, Annex C, Medical face masks- Requirements and test methods

  • ISO 10993-5:2009 Biological evaluation of medical device- Part 5: Tests for in vitro cytotoxicity

  • ISO 10993-10:2010 Biological evaluation of medical device- Part 10: Tests for irritation and skin sensitization

  • ISO 10993-1:2018 Biological evaluation of medical devices - Part 1 Evaluation and testing within a risk management process.

PerformanceCharacteristicsTest MethodTest resultAcceptance CriteriaConclusion
Fluid ResistancePerformanceASTMF1862/F1862M:2017Pass at 160mmHgNo penetration at 160mmHgMeet the requirement
Particulate FiltrationEfficiencyASTMF2299/F2299M-03(2017)Pass at 99.2%≥98%Meet the requirement
Bacterial FiltrationEfficiencyASTM F2101:2019Pass at 99.2%≥98%Meet the requirement
Differential Pressure(Delta-P)EN 14683: 2019,Annex CPass at 3.8 mmH2O/cm2<6.0 mmH2O/cm2Meet the requirement
Flammability class16 CFR Part 1610Class 1Class 1Meet the requirement

8. Clinical Test Conclusion

No clinical study is included in this submission.

9. Conclusion

The conclusion drawn from the non-clinical tests demonstrates that the medical face mask is as safe, as

{9}------------------------------------------------

effective, and performs as well as or better than the legally marketed predicate device cleared under K160269.

§ 878.4040 Surgical apparel.

(a)
Identification. Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.(b)
Classification. (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.