K Number
K212981

Validate with FDA (Live)

Manufacturer
Date Cleared
2022-04-22

(217 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The MirrorMe3D Modeling System is intended for use as an image processing system for the transfer of 3D medical images. The MirrorMe3D Modeling System is also intended as a visualization system for measuring and treatment planning for aesthetic facial soft tissue. The input data is processed by the System using off-the-shelf modeling software and the result is an output data file that may then be provided as a digital model or used as input for the additive manufacturing of a physical anatomic model, which is not for diagnostic use. The MirrorMe3D Modeling System should only be used in conjunction with expert clinical judgment and is not intended for diagnostic use. MirrorMe3D trained personnel will use off-the-shelf software to assist users in creating the 3D virtual (or digital) model that depicts the surgeon's intended outcome. The anatomic models are not for diagnostic use.

Device Description

The MirrorMe3D Modeling System is image processing software that enables the input and visualization of 3D medical imaging with output files that can be virtual or physical 3D anatomic models. The Modeling System software is used for visualization of preoperative treatment planning options, with measurement functionality, for surgery of the aesthetic facial soft tissue.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study information you requested, based on the provided document:

Acceptance Criteria and Device Performance

The document describes that for each device produced by MirrorMe3D, there are checks to ensure the system conforms to specifications and is fit for its intended use. There isn't a single, aggregated table of numerical acceptance criteria for the entire device as one might find for a standalone AI algorithm. Instead, the "acceptance criteria" are tied to individual product validation and process auditing.

However, based on the text, we can infer some key performance aspects being tested.

1. Table of Acceptance Criteria (Inferred) and Reported Device Performance

Acceptance Criteria CategorySpecific Acceptance Criteria (Inferred)Reported Device Performance/Verification Method
Input Data IntegrityInput imaging data is free from corruption and suitable for processing."MirrorMe3D checks the integrity of the input imaging data."
Treatment Option Visualization AccuracyThe visualization of patient-specific treatment options accurately reflects the intended plan and is acceptable to the doctor."validates the visualization of the patient specific treatment options through doctor and staff review, conducts testing and a verification of the model design files." Additionally, "The approval of the design of the anatomic model depicting the intended treatment outcome by the Doctor is required and indicates design acceptance."
Model Design File VerificationThe digital model design files adhere to specifications and accurately represent the intended outcome."conducts testing and a verification of the model design files."
Physical Product QualityPhysical 3D printed models are free from visual defects and meet established quality protocols."visually inspections all physical products using a quality protocol."
Geometric Accuracy of Additive ManufacturingAdditively manufactured outputs maintain geometric accuracy within an established tolerance range."The model production process is tested on a monthly basis to confirm the additively manufactured outputs meet conformance standards and maintain geometric accuracy within an established tolerance range."
Off-The-Shelf (OTS) Software PerformanceThe off-the-shelf software programs maintain acceptable tolerances and reasonable measurement parameters for modeling."Software testing is periodically conducted to determine if the modeling maintains acceptable tolerances and is within reasonable measurement parameters and documentation was provided as recommended by the FDA Guidance for 'Off-The-Shelf Software Use in Medical Devices'." Risk analysis determined "the severity of the harm that could result from a software failure is a minor level of hazard to patients."

Important Note: The document describes a workflow and quality control process for each individual product rather than a large-scale statistical study of the device as a whole against predefined performance metrics like sensitivity/specificity for a diagnostic device. The MirrorMe3D Modeling System is framed as an "image processing system" and "visualization system" that relies heavily on human expertise and approval.


Study Information

The document describes process validation and quality control measures for the manufacturing and utilization of the MirrorMe3D Modeling System for each specific case, rather than a single, large-scale clinical performance study with a test set in the traditional sense of an AI/diagnostic device submission.

Thus, many of your requested items for a "study" (like test set sample size, expert qualifications for ground truth establishment, MRMC studies) are not directly applicable or explicitly detailed in the provided text, as the "study" is more of an ongoing quality assurance and product-specific validation process.

Let's address the points as best as possible given the provided text:

2. Sample size used for the test set and the data provenance:

  • Sample Size: Not applicable in the context of a single retrospective/prospective test set. The document states their process: "For each device, MirrorMe3D checks..." indicating that every individual case processed by MirrorMe3D undergoes a series of checks and reviews.
  • Data Provenance: Not specified. The input is "3D medical images," but the source (country, hospital) is not mentioned. Given it's a service, the data would come from individual clinicians/patients. The document also doesn't specify if the input images for the internal process validation were retrospective or prospective.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Number of Experts: For each case, at least one surgeon (Doctor) is involved in establishing the "ground truth" (i.e., the intended treatment outcome). MirrorMe3D trained personnel also assist.
  • Qualifications of Experts: The primary "expert" establishing the desired outcome (which acts as the ground truth for the model's design) is the "Doctor" (Surgeon). Specific qualifications like years of experience or board certification are not detailed in this document but are implicitly understood for a surgeon using such a system for treatment planning. MirrorMe3D also employs "trained personnel" to assist.

4. Adjudication method for the test set:

  • Adjudication Method: "The approval of the design of the anatomic model depicting the intended treatment outcome by the Doctor is required and indicates design acceptance." This suggests a single-expert approval model for each case. There is no mention of a multi-reader consensus or 2+1/3+1 adjudication for a test set.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was NOT done according to the provided text. The device is described as a "visualization system for measuring and treatment planning" and "image processing software" where "MirrorMe3D trained personnel will use off-the-shelf software to assist users in creating the 3D virtual (or digital) model that depicts the surgeon's intended outcome." This process does not describe an AI-assisted diagnostic workflow with human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • No, a standalone algorithm-only performance study was NOT done. The system explicitly requires "MirrorMe3D trained personnel" and "expert clinical judgment" (from the Doctor/Surgeon) to create the models and approve the design. It's a human-in-the-loop system. The document states, "MirrorMe3D trained personnel will use off-the-shelf software to assist users in creating the 3D virtual (or digital) model that depicts the surgeon's intended outcome."

7. The type of ground truth used:

  • The primary "ground truth" for the treatment plan and subsequent model design is the "surgeon's intended outcome" or "expert clinical judgment" as approved by the Doctor/Surgeon. For the physical models, the ground truth is adherence to the approved digital design and established geometric accuracy tolerances.

8. The sample size for the training set:

  • Not applicable as this document describes a quality control and process validation framework for a system that uses "off-the-shelf software" and relies on human intervention, rather than a deep learning AI model that requires a dedicated training set. The "off-the-shelf software" would have been developed and "trained" by its respective manufacturers, but that is not part of this submission for the MirrorMe3D system.

9. How the ground truth for the training set was established:

  • Not applicable. As explained in point 8, the MirrorMe3D system itself doesn't refer to a "training set" in the context of an AI algorithm that learns from data.

{0}------------------------------------------------

April 22, 2022

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

MirrorMe3D, LLC % Jordan Mills Chief Executive Officer 222 W 37th Street. Suite 1501 NEW YORK NY 10018

Re: K212981

Trade/Device Name: MirrorMe3D Modeling System Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: LLZ Dated: March 21, 2022 Received: March 22, 2022

Dear Jordan Mills:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K212981

Device Name MirrorMe3D Modeling System

Indications for Use (Describe)

The MirrorMe3D Modeling System is intended for use as an image processing system for the transfer of 3D medical images. The MirrorMe3D Modeling System is also intended as a visualization system for measuring and treatment planning for aesthetic facial soft tissue. The input data is processed by the System using off-the-shelf modeling software and the result is an output data file that may then be provided as a digital model or used as input for the additive manufacturing of a physical anatomic model, which is not for diagnostic use. The MirrorMe3D Modeling System should only be used in conjunction with expert clinical judgment and is not intended for diagnostic use. MirrorMe3D trained personnel will use off-the-shelf software to assist users in creating the 3D virtual (or digital) model that depicts the surgeon's intended outcome. The anatomic models are not for diagnostic use.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY

Image /page/3/Picture/2 description: The image contains the logo for MirrorMe3D. The logo consists of a circular icon on the left and the text "MirrorMe3D" on the right. The icon features a stylized, geometric shape resembling a folded paper or origami design in light blue and white, set against a muted blue background.

212-376-4546 hello@mirrorme3d.com

222 W 37th Street Suite 1501 New York, NY 10018

A. SUBMITTER

Submitter NameMirrorMe3D, LLC
Submitter Address222 W 37th St., Suite 1501, New York, NY 10018
Contact Person:Jordan Mills (jordan@mirrorme3d.com)
Phone/Fax Numbers:212-376-4546 / 914-350-4697
Date of Submission:April 19, 2022

B. DEVICE IDENTIFICATION & MANUFACTURER

Name of DeviceMirrorMe3D Modeling System
Common NameMirrorMe Modeling
Classification Name21 CFR 892.2050 Medical image management andprocessing system
Classification CodeLLZ
Regulatory ClassII

C. PRIMARY, SECONDARY PREDICATE AND REFERENCE DEVICES

Primary Predicate Device: K183105, Materialise Mimics Medical Secondary Predicate: K161634, ISP CaseDesigner Reference Device: K192282, MedCAD AccuPlan® System The predicate, the secondary predicate and the reference devices have not been subject to a design-related recall.

D. DEVICE DESCRIPTION

The MirrorMe3D Modeling System is image processing software that enables the input and visualization of 3D medical imaging with output files that can be virtual or physical 3D anatomic models. The Modeling System software is used for visualization of preoperative treatment planning options, with measurement functionality, for surgery of the aesthetic facial soft tissue.

MirrorMe3D trained personnel will process the input data using off-the-shelf software and assist doctors in visualizing treatment plan options. The digital models created under the direction of doctors can then be additively manufactured or 3D printed to provide the surgeon with physical replicas of the digital models. The anatomic models

510(k) Summary Page 1

{4}------------------------------------------------

are not for diagnostic use, do not have any contact with the patient and are not intended for use in a surgical environment. Patient-specific documentary reports will contain screenshots of the virtual models and depict dimensional differences.

E. INDICATIONS FOR USE

The MirrorMe3D Modeling System is intended for use as an image processing system for the transfer of 3D medical images. The MirrorMe3D Modeling System is also intended as a visualization system for measuring and treatment planning for aesthetic facial soft tissue. The input data is processed by the System using off-the-shelf modeling software and the result is an output data file that may then be provided as a digital model or used as input for the additive manufacturing of a physical anatomic model, which is not for diagnostic use. The MirrorMe3D Modeling System should only be used in conjunction with expert clinical judgment and is not intended for diagnostic use. MirrorMe3D trained personnel will use off-the-shelf software to assist users in creating the 3D virtual (or digital) model that depicts the surgeon's intended outcome. The anatomic models are not for diagnostic use.

F. PREDICATE DEVICES

The intended use of the MirrorMe3D Modeling System and its predicate devices, the Materialise Mimics Medical (K183105) and ISP CaseDesigner (K161634) are substantially equivalent. The MedCAD AccuPlan® System (K192282) has been included as a reference device to address the minor differences in technological characteristics between the MirrorMe3D System and the primary and secondary predicates. The subject device, predicates and reference systems are intended for use in the transfer and processing of medical data and the visualization of treatment plan options, with 3D models as the output.

{5}------------------------------------------------

MirrorMe3D ModelingSystemPrimary Predicate Device:Materialise Mimics Medical,K183105Secondary PredicateDevice: ISP CaseDesignerK161634Reference Device: MedCADAccuPlan®, K192282TREATMENTPLANNING/ APPROVALSurgeons virtually visualizeand plan treatment options.Virtual models are created bytrained MirrorMe3D employeesand must be approved by thesurgeon before beingmanufactured.Users create the digital 3Dmodels and should be used inconjunction with expertclinical judgmentSurgeons (or clinicians)virtually plan the surgeryUtilized by trained MedCADemployees with the approvalof the surgeon before beingmanufactured.
REGULATORYCLASSIFICATIONII892.2050LLZII892.2050LLZII892.2050LLZII892.2050 and 872.4120DZJ, LLZOUTPUTAdditively manufacturedanatomic Models for visual useonly; documentary reportsAdditively or traditionallymanufactured physicalreplicas for diagnostic usePatient specific 3DSurgical models andsurgical splintsRapid prototyping ofanatomical models, surgicalquides and dental splints;case reports
INDICATIONSFOR USEThe MirrorMe3D ModelingSystem is intended for use asan image processing systemfor the transfer of 3D medicalimages. The MirrorMe3DModeling System is alsointended as a visualizationsystem for measuring andtreatment planning foraesthetic facial soft tissue.The input data is processedby the System using off-the-shelf modeling software andthe result is an output datafile that may then beprovided as a digital model orused as input for the additivemanufacturing of a physicalanatomic model, which is notfor diagnostic use. TheMirrorMe3D Modeling Systemshould only be used inconjunction with expertclinical judgment and is notMimics Medical is intendedfor use as a softwareinterface and imagesegmentation system for thetransfer of medical imaginginformation to an output file.Mimics Medical is alsointended for measuring andtreatment planning. TheMimics Medical output can beused for the fabrication ofphysical replicas of the outputfile using traditional oradditive manufacturingmethods. The physical replicacan be used for diagnosticpurposes in the field oforthopedic, maxillofacial andcardiovascular applications.Mimics Medical should beused in conjunction withexpert clinical judgement.IPS CaseDesigner isindicated for use as asoftware and imagesegmentation system forthe transfer of imaginginformation from ascanner such as a CTscanner. It is alsoindicated to support thediagnostic and treatmentplanning process ofcraniomaxillofacialprocedures. IPSCaseDesigner facilitatesthe service offering ofindividualized surgicalaids.The MedCAD® AccuPlan®System is intended for use asa software system and imagesegmentation system for thetransfer of imaginginformation from a medicalscanner such as a CT basedsystem. The input data file isprocessed by the MedCAD®AccuPlan® System and theresult is an output data filethat may then be provided asdigital models or used asinput to a rapid prototypingportion of the system thatproduces physical outputsincluding anatomical models,surgical guides and dentalsplints for use inmaxillofacial surgery. Thesurgical guides and dentalsplints are intended to beused for the maxillofacialbone in maxillofacial surgery.The MedCAD® AccuPlan®System is also intended as apre-operative software toolSOFTWARE FORPLANNINGCommercial off-the shelfsoftware.Image processing softwareImage processingsoftwareCommercial off-the shelfsoftware
intended for diagnostic use.MirrorMe3D trained personnelwill use off-the-shelf softwareto assist users in creating the3D virtual (or digital) modelthat depicts the surgeon'sintended outcome. Theanatomic models are not fordiagnostic use.for evaluating surgicaltreatment options.Diagnostic UseNoYesYesNo
DEVICE TYPE,USAGEPatient specific,Single UsePatient specific,Single UsePatient specific,Single UsePatient specific,Single Use
SURGICALSITESExternal maxillofacial softtissueOrthopedic, maxillofacial, andcardiovascular applicationsCranio-maxillofacialincluding soft tissueMaxillofacial surgery
INPUT FILEFORMATS.OBJ, .MTL and other formats(such as TIFF, JPG, .PNG andraw imagesDICOM and other formats(such as BMP, TIFF, JPG, andraw images)DICOM from CT/CBCT, STLCT based
DEVICEFUNCTION-ALITYTransferring and viewingmedical imaging;Processing to output file;Measuring and treatmentplanning on 3D modelsTransferring and viewingmedical imaging;Image segmentation;Processing to output file;Measuring and treatmentplanning on 3D modelsTransferring and viewingmedical imaging;Image segmentation;Processing to output file;Measuring and treatmentplanning on 3D modelsTransferring and viewingmedical imaging;Image segmentation;Processing to output file;Measuring and Treatmentplanning on 3D models

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

Summary

    1. The subject device, predicate, secondary predicate, and reference systems are intended for use in the transfer and processing of medical data with 3D models as the output.
    1. The subject device, predicate, secondary predicate, and reference device are 3D visualization systems for measuring and treatment planning.
    1. The subject device, predicate, secondary predicate, and reference device start with medical imaging provided by a doctor then use software for manipulation in order to create virtual 3D models that reflect the doctor's intended treatment plan.
    1. The subject device, primary predicate, and reference systems produce an output file that is used for the additive manufacture of anatomic models, with the difference that the physical models by the subject device are not for diagnostic use.
    1. The outputs of all are prescriptive, patient-specific, and single use devices intended for use solely by Doctors, not Patients.
    1. The subject device and the reference systems include documentary or case reports as output.
    1. The MirrorMe3D System is more limited in its application to the facial soft tissue, as opposed to Mimics Medical (maxillofacial, orthopedic and cardiovascular applications), ISP CaseDesigner (craniomaxillofacial including soft tissue) and AccuPlan (maxillofacial).
    1. MirrorMe3D System starts with the input of 3D medical imaging of soft tissue, while the other devices start with files that must be segmented prior to 3D modeling.

MirrorMe3D believes its Modeling System is safe and effective and that it is substantially equivalent to legally marketed devices. The MirrorMe3D System has the same intended use as Mimics Medical, ISP CaseDesigner and the AccuPlan Systems. The differences in technical characteristics do not add any questions of safety or effectiveness to the subject device.

G. SUMMARY OF PERFORMANCE TESTING

MirrorMe3D's Modeling System consists of patient-specific output so to provide objective evidence that the system conforms to specifications and is fit for its intended use, MirrorMe3D intends to show that every product is tested to meet acceptance criteria and we periodically audit the processes. MirrorMe3D has tried to identify the risks associated with every step of the design and manufacture process and to mitigate all of those risks.

Testing For each device, MirrorMe3D checks the integrity of the input 1. imaging data, validates the visualization of the patient specific treatment options

510(k) Summary Page 6

{9}------------------------------------------------

through doctor and staff review, conducts testing and a verification of the model design files, and visually inspections all physical products using a quality protocol. The approval of the design of the anatomic model depicting the intended treatment outcome by the Doctor is required and indicates design acceptance. The model production process is tested on a monthly basis to confirm the additively manufactured outputs meet conformance standards and maintain geometric accuracy within an established tolerance range.

  1. Biocompatibility and Sterilization: The models are not sterile and are not intended for any physical contact with the patient during the procedure.

Off-The-Shelf Software: MirrorMe3D uses only off-the-shelf software 3. programs to design and manufacture its products. Software testing is periodically conducted to determine if the modeling maintains acceptable tolerances and is within reasonable measurement parameters and documentation was provided as recommended by the FDA Guidance for "Off-The-Shelf Software Use in Medical Devices" issued September 27, 2019. MirrorMe3D has analyzed the risks and determined that the severity of the harm that could result from a software failure is a minor level of hazard to patients.

Summary

The intention of these tests is to provide objective evidence that the system conforms to specifications, is fit for its intended use, and that its performance is substantially equivalent to the predicate device. All validations, verifications and testing demonstrate that MirrorMe3D has identified and evaluated the potential risks, and its design and production processes have been conformed to meet its quality standards.

The MirrorMe3D Modeling System is substantially equivalent to and has the same intended use as Mimics Medical but different tarqet areas, as well as slightly different technological characteristics that do not raise different questions of safety and effectiveness, and the information submitted herein demonstrates that the system is as safe as the legally marketed predicate device.

§ 892.2050 Medical image management and processing system.

(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).