K Number
K212972

Validate with FDA (Live)

Date Cleared
2022-08-18

(335 days)

Product Code
Regulation Number
880.5570
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The safety needles are intended to be used with a Luer lock syringe for aspiration and injection of fluids for medical purpose. After withdrawal of the needle from the body, the attached needle safety shield can be manually activated to cover the needle immediately after use to minimize risk of accidental needle-stick.

Device Description

The KDL Safety Needle is a hypodermic needle with a hinged safety sheath attached to the needle hub. The safety sheath could be manually activated when the needle is withdrawn from the body before disposal, minimizing the risk of needle sharps injury during disposal. It is activated manually by hand or any solid surface. The self-locking mechanism is positioned within the center and the proximal end of the sheath. The hinge structure could let clinical personnel flexibly adjust the sheath to the designed position for use. The KDL Safety Needle has various colors, needle gauge sizes, and needle length sizes, which could be applied in different clinic-use scenarios. Meanwhile, the KDL Safety Needle is sterilized by ethylene oxide and without any pyrogen or latex.

AI/ML Overview

This document is a 510(k) Premarket Notification from the FDA regarding "KDL Safety Needles". It details the device, its intended use, comparison to a predicate device, and the non-clinical testing performed to establish substantial equivalence.

Based on the provided text, the device in question is a hypodermic needle with a safety shield, not an AI/ML-driven device or a diagnostic aid. As such, the typical "acceptance criteria" and "study" framework for AI/ML medical devices (e.g., related to sensitivity, specificity, reader studies, ground truth establishment by experts, etc.) do not apply to this submission.

The document describes the testing performed for this device as non-clinical testing in conformance with various ISO and ASTM standards, which are primarily related to physical performance, sterility, and biocompatibility. There is an explicit statement: "No clinical testing is available for this device."

Therefore, I cannot provide the information requested in the prompt's numbered list, as it pertains to AI/ML device evaluation criteria that are not relevant to this submission.

The "acceptance criteria" for this device are its ability to meet the specified performance standards (e.g., ISO 7864, ISO 9626, ISO 23908), sterility requirements (e.g., ISO 11135), and biocompatibility requirements (e.g., ISO 10993 series). The "study that proves the device meets the acceptance criteria" refers to the non-clinical tests conducted according to these standards, as enumerated in section 8 of the 510(k) summary. These tests demonstrate the device's physical properties, sterility, and biological safety.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

August 18, 2022

Shanghai Kindly Enterprise Development Group Co., Ltd % Evan Hu Marketing & Technical Manager Shanghai Mind-link Consulting Co., Ltd. 1399 Jiangyue Road, Minhang Shanghai, Shanghai 201114 China

Re: K212972

Trade/Device Name: KDL Safety Needles Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: Class II Product Code: FMI Dated: July 8, 2022 Received: July 18, 2022

Dear Evan Hu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

CAPT Alan M. Stevens Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K212972

Device Name KDL Safety Needles

Indications for Use (Describe)

The safety needles are intended to be used with a Luer lock syringe for aspiration and injection of fluids for medical purpose. After withdrawal of the needle from the body, the attached needle safety shield can be manually activated to cover the needle immediately after use to minimize risk of accidental needle-stick.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K212972 510(k) SUMMARY

1. Preparation date: 8/16/2022

2. Submitter

Manufacturer: Shanghai Kindly Enterprise Development Group Co., Ltd. Address: No.658 Gaochao Road, 201803, Shanghai, China Contact person: Liu Huarong, 86 02169118232, henry_liu@kdlchina.net Submission correspondent: Evan Hu, 86-18616124827, Evan.hu@mind-link.net

3. Device

Trade name: KDL Safety Needles Common name: Safety needle Regulation Number: §21 CFR 880.5570 Classification name: Hypodermic Single Lumen Needle Classification: Class II Product code: FMI

4. Predicate device

TK Safety Needle (K191644), FMI, §21 CFR 880.5570

5. Device description

The KDL Safety Needle is a hypodermic needle with a hinged safety sheath attached to the needle hub. The safety sheath could be manually activated when the needle is withdrawn from the body before disposal, minimizing the risk of needle sharps injury during disposal. It is activated manually by hand or any solid surface.

The self-locking mechanism is positioned within the center and the proximal end of the sheath. The hinge structure could let clinical personnel flexibly adjust the sheath to the designed position for use.

The KDL Safety Needle has various colors, needle gauge sizes, and needle length sizes, which could be applied in different clinic-use scenarios. Meanwhile, the KDL Safety Needle is sterilized by ethylene oxide and without any pyrogen or latex.

Birmingham gaugeLength (mm)Needle cutting edgeNeedle wall
18G25, 32, 38, 50SB, LBRW, TW
19G25, 32, 38, 50SB, LBRW, TW
20G25, 32, 38, 50LBRW, TW
21G25, 32, 38, 50LBRW, TW

Table 1. Device models

{4}------------------------------------------------

22G25, 32, 38, 50LBRW, TW
23G25, 38LBRW, TW
24G20, 25LBRW, TW
25G16, 25, 38LBRW, TW
26G9, 13SB, LBRW, TW
27G6, 8, 9, 12, 13, 15SB, LBRW, TW
28G6, 8, 12, 15SB, LBRW, TW
29G6, 8, 12, 15SB, LBRW, TW
30G6, 8, 9, 12, 13, 15SB, LBRW, TW

The KDL Safety Needles are available in various models, mainly 13 types with different needle lengths and configurations and a total of 120 sub-types, as shown in Table 1.

6. Indications for Use/Intended Use

The safety needles are intended to be used with a Luer slip or Luer lock syringe for aspiration and injection of fluids for medical purpose. After withdrawal of the needle from the body, the attached needle safety shield can be manually activated to cover the needle immediately after use to minimize risk of accidental needle-stick.

7. Comparison of technological characters between proposed and predicate devices

CharactersProposed device(K212972- KDL Safety Needles)Predicate device(K191644- TK Safety Needle)Remark
Product codeFMIFMISame
Indications forUseThe safety needles are intended tobe used with a Luer slip or Luer locksyringe for aspiration and injectionof fluids for medical purpose. Afterwithdrawal of the needle from thebody, the attached needle safetyshield can be manually activated tocover the needle immediately afteruse to minimize risk of accidentalneedle-stick.TK Safety Needle device is intendedfor use in the aspiration andinjection of fluids for medicalpurposes. The TK Safety Needle iscompatible for use with standardluer slip and luer lock syringes.Additionally, after withdrawal of theneedle from the body the attachedneedle safety shield can be manuallyactivated to cover the needleimmediately after use to minimizerisk of accidental needle-stick.Similar 1
Prescription orover-the-counter usePrescriptionPrescriptionSame
Material:Needle hubPolypropylenePolypropyleneSame
NeedleStainless steelStainless steelSame

Table 2. Characters comparison

{5}------------------------------------------------

Needle sheathPolypropylenePolypropylene
AdhesiveEpoxy sizesEpoxy sizes
LubricantSilicone oilSilicone oil
Needle Gaugeand length18G: 25, 32, 38, 50mm16G: 1 to 1 1/2"Different 1
19G: 25, 32, 38, 50mm17G: 1 to 1 1/2"
20G: 25, 32, 38, 50mm18G: 1 to 1 1/2"
21G: 25, 32, 38, 50mm19G: 1 to 1 1/2"
22G: 25, 32, 38, 50mm20G: 1 to 1 1/2"
23G: 25, 38mm21G: 1 to 1 1/2"
24G: 20, 25mm22G: 1 to 1 1/2"
25G: 9, 13, 16, 25, 38mm23G: 1 to 1 1/2"
26G: 9, 13mm24G: 1 to 1 1/2"
27G: 6, 8, 9, 12, 13, 15mm25G: 1 to 1 1/2"
28G: 6, 8, 12, 15mm26G: 1 to 1 1/2"
29G: 6, 8, 12, 15mm27G: 1 to 1 1/2"
30G: 6, 8, 9, 12, 13, 15mm28G: 1/2 to 1"
29G: 1/2 to 1"
30G: 1/2 to 1"
TipconfigurationBevelBevelSame
ProtectivefeatureExtra safety sheathExtra safety sheathSame
Meet ISO 23908Meet ISO 23908
ColorPer ISO 6009Per ISO 6009Same
SterilitySterilized by EOSterilized by EOSame
Single-useSingle-useSingle-useSame
PhysicalperformanceMeet ISO 7864 and ISO 9626Meet ISO 7864 and ISO 9626Same
BiocompatibilityMeet ISO 10993Meet ISO 10993Same

Similar 1: The intended use of the proposed and predicate device is the same that is intended for fluid aspiration and injection. Meanwhile, they both have the same safety protection features. The only difference is that the indications for use of the predicate device includes its trade name. It does not impact the device's safety and effectiveness.

Different 1: 1" is equal to 25.4mm. The needle length range of the predicate device is 13 to 38mm for various gauges. However, the proposed device has models that are longer length (50mm) and shorter length (6~12mm). The difference in needle length could impact the device's performance, so performance testing were conducted to cover the difference. The results demonstrate that device performances of selected models (including 6, 8, 9, 12 and 50mm) meet the requirements of ISO 7864 and ISO 9626. Therefore, the difference in needle length does not impact the device's safety and effectiveness.

{6}------------------------------------------------

8. Non-clinical testing results

The non-clinical tests of this proposed device are tested in conformance with the following standards.

(1) Physical performance testing:

  • (a) ISO 7864:2016, Sterile hypodermic needles for single use.
  • (b) ISO 80369-7:2016, Small-bore connectors for liquids and gases in healthcare applications - part 7: Connectors for intravascular or hypodermic applications.
  • (c) ISO 9626:2016 Stainless Steel Needle Tubing for the Manufacture of Medical Device.
  • (d) ISO 6009:2016 Hypodermic needles for single use Colour coding for identification
  • (e) ISO 23908:2011, Sharps injury protection Requirements and test methods Sharps protection features for single-use hypodermic needles, introducers for catheters and needles used for blood sampling

(2) Sterility:

  • (a) ISO 11135:2014 Sterilization of healthcare products Ethylene oxide -Requirements for the development, validation and routine control of a sterilization process for medical devices.
  • (b) ISO 11138-2:2017 Sterilization of health care products Biological indicators Part 2: Biological indicators for ethylene oxide sterilization processes.
  • (c) ISO 11737-1:2018 Sterilization of health care products Microbiological methods Part 1: Determination of a population of microorganisms on products.
  • (d) ISO 11737-2:2019 Sterilization of health care products Microbiological methods Part 2: Tests of sterility performed in the definition, validation and maintenance of a sterilization process.
  • (e) ISO 10993-7:2008/AMD 1:2019 Biological evaluation of medical devices Part 7: Ethylene oxide sterilization residuals - Amendment 1: Applicability of allowable limits for neonates and infants.
  • (f) EN 868-7:2017, Packaging for terminally sterilized medical devices Part 7: Adhesive coated paper for low temperature sterilization processes - Requirements and test methods
  • (g) EN 868-1:1997, Packaging materials and systems for medical devices which are to be sterilized - Part 1: General requirements and test methods
  • (h) ASTM F1140: 2020, Standard Test Methods for Internal Pressurization Failure Resistance of Unrestrained Packages
  • (i) ASTM F88: 2015, Standard Test Method for Seal Strength of Flexible Barrier Materials
  • (j) ASTM F1929: 2015, Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration

(3) Biocompatibility testing:

  • (a) ISO10993-1:2018, Biological evaluation of medical devices Part 1: Evaluation and testing within a risk management process.
  • (b) ISO 10993-4:2017 Biological evaluation of medical devices Part 4: Selection of tests for interactions with blood.
  • (c) ISO 10993-5:2009, Biological evaluation of medical devices Part 5: Tests for in vitro cytotoxicity.

{7}------------------------------------------------

  • (d) ISO 10993-10:2010, Biological evaluation on medical device Part 10: Test for irritation and Skin Sensitization.
  • (e) ISO 10993-11:2017, Biological evaluation of medical devices Part 11: Tests for systemic toxicity and Pyrogenicity
  • (f) USP <161> Medical Devices—Bacterial Endotoxin test
  • (g) USP <788> Particulate Matter for Injections

9. Clinical testing

No clinical testing is available for this device.

10. Conclusion

The differences between the predicate and the proposed device do not raise any new or different questions of safety or effectiveness. The proposed device is substantially equivalent to the predicate device with respect to indications for use and technological characteristics.

§ 880.5570 Hypodermic single lumen needle.

(a)
Identification. A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.(b)
Classification. Class II (performance standards).