K Number
K212932

Validate with FDA (Live)

Date Cleared
2021-12-18

(94 days)

Product Code
Regulation Number
880.6850
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Nobel Biocare PureSet Trays are used in healthcare facilities to store and organize Nobel Biocare surgical/prosthetic instruments and components during cleaning/sterilization and during implant/prosthetic treatment.

Nobel Biocare PureSet Trays are not intended on their own to maintain sterility; they are intended to be used in conjunction with a legally marketed, validated, FDA cleared sterilization pouch, or sterilization wrap.

Sterilization validations for the worst-case PureSet Tray included surgical instruments such as torque wrenches, implant drivers, direction indicators, drills, screw taps, screwdrivers, and irrigation needles. The PureSet Trays were validated for a maximum load of 1635 grams (Trefoil PureSet Tray), 1122 grams (NobelActive / NobelParallel CC PureSet Tray), 1063 grams (NobelReplace CC PureSet Tray), 454 grams (Nobel Biocare N1™ PureSet Tray), 486 grams (Prosthetic PureSet Tray), 1143 grams (NobelActive Guided PureSet Tray), 1146 grams (NobelParallel CC Guided PureSet Tray), 1176 grams (NobelReplace CC Guided PureSet Tray), and 1035 grams (NobelSpeedy® Groovy / Brånemark System® Mk III TiUnite / Replace SelectTM TC PureSet™ Tray).

MethodSteam Sterilization (Moist Heat Sterilization)for Wrapped Instruments
CycleDynamic-Air-Removal(fractionated vacuum)Gravity-Displacement
Temperature132°C (270°F)132°C (270°F)
Exposure time for a single-usepouched device4 minutes (full-cycle)15 minutes (full-cycle)
Minimum drying times20 minutes30 minutes
Device Description

PureSet Trays are reusable surgical trays to be used in combination with Nobel Biocare surgical instruments and components. PureSet Trays are used to organize and store the instruments and components during both surqical and reprocessing procedures.

PureSet Trays are not intended to maintain sterility on their own; they are intended to be used in conjunction with an FDA cleared sterilization wrap, pouch, or container.

All components of the PureSet Tray are perforated with an evenly-distributed hole pattern and are designed to be used for sterilization via steam sterilization. Because the PureSet Trays are perforated, an FDA-cleared sterilization wrap, pouch, or container must be used during sterilization and storage to maintain the sterility of the contents.

PureSet Trays are designed to be used with standard autoclaves used in hospitals and healthcare facilities.

Principle of Operation / Mechanism of Action:

PureSet Trays consist of multiple components (tray base, lid, and plate) integrated into a single unit which is used to organize instruments during surgical procedures and to protect the instruments during transportation, reprocessing, and storage.

AI/ML Overview

The provided text describes a 510(k) premarket notification for a medical device (Nobel Biocare PureSet Trays), which focuses on demonstrating substantial equivalence to predicate devices rather than proving novel clinical effectiveness. Therefore, the device does not have acceptance criteria in the traditional sense of a clinical trial (e.g., meeting a certain sensitivity/specificity threshold for a diagnostic device). Instead, the "acceptance criteria" here refers to the successful completion of non-clinical performance tests designed to show that the device is as safe and effective as its predicates for its intended use.

Here's an analysis based on the provided text, adapted to the requested format where applicable:

1. Table of Acceptance Criteria and Reported Device Performance

Test Scope / ObjectivesAcceptance CriteriaReported Device Performance
Durability of Tray Handle and Closing Mechanism: Demonstrate that the tray handle and closing mechanism can withstand at least 4 times the maximum weight of a fully loaded tray.• The tray closing mechanism does not open while the tray is lifted with 4 times maximum weight of a fully assembled tray (4140 grams).• The handle does not break loose from the tray, and there is no permanent distortion, cracking or other evidence of failure to the handle or closing mechanism.• Tested to 250N of force (25 kg of weight) without opening.• Handle did not break loose from the tray, and there was no permanent distortion, cracking or other evidence of failure to the handle or closing mechanism.
Packaging Performance: Demonstrate that the packaging system adequately protects the tray against the hazards associated with global distribution.• Product shall be damage free.• The cardboard box and the LD-PE bags containing the PureSet Tray shall be intact.• All box labels shall be present and legible.• Product was damage free.• The cardboard box and LD-PE bags were intact.• All box labels were present and legible.
Simulated Intrafacility Transportation: Demonstrate that the subject PUR0900 PureSet tray retains all instruments in their designated holders during simulated transport of a fully-loaded tray throughout a typical healthcare facility.• The instruments shall stay in the designated location within the tray after each test sequence of each test run.• The assembled Torque Wrench does not become disassembled after transport during each test sequence.• All instruments remained in their designated locations after each sequence during each test run.• The torque wrench remained assembled after each sequence during each test.
Repeated Reprocessing: Demonstrate that the tray and plate components of the PureSet Tray can withstand repeated reprocessing cycles (500 cycles for the tray and 250 cycles for the plate) without unacceptable degradation to the laser marking and printing.• All laser marking on the trays and plates shall remain legible, and the colors on the plate shall remain identifiable and with no detected discoloration or corrosion.• PureSet trays and plates shall remain biocompatible after repeated reprocessing following the methods established by Nobel Biocare.• The PureSet plates met all acceptance criteria after 250 reprocessing cycles.• The PureSet trays met all criteria after both 250 and 500 cycles.• No cytotoxic effects were observed in any of the tested samples.

2. Sample Size Used for the Test Set and the Data Provenance

  • Durability of Tray Handle and Closing Mechanism: The testing was performed on the "worst-case (heaviest) fully-loaded tray in the PureSet Tray family (PUR0100; 1633 grams fully loaded)". The specific number of individual trays tested for this is not explicitly stated, but it implies at least one representative sample.
  • Packaging Performance: "Two group of five sample trays" were packaged and sent for testing. So, a total of 10 trays were tested.
  • Simulated Intrafacility Transportation: A "single example of a fully-loaded PUR0900 PureSet tray" was used. Three test runs were performed for each method by a single operator.
  • Repeated Reprocessing: One plate was subjected to 250 cycles. Two trays were tested; one was subjected to 250 cycles and the other to 500 cycles.
  • Data Provenance: The document does not specify the country of origin for the data collection, but given Nobel Biocare has offices in Sweden and Switzerland, and the FDA submission is for the U.S. market, it's likely conducted in Europe or a certified testing facility globally. The tests are non-clinical (laboratory/simulated), not retrospective or prospective clinical studies.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

This information is not applicable. The device is a sterilization wrap accessory (surgical tray). The "ground truth" for its performance is established through engineering and biological performance testing against recognized standards, not through expert clinical consensus or interpretation of medical images. For example, "Pass/Fail Criteria" are defined by engineering specifications (e.g., "tray closing mechanism does not open", "product shall be damage free", "laser marking... shall remain legible").

4. Adjudication Method for the Test Set

This information is not applicable. Adjudication methods (e.g., 2+1, 3+1) are typically used in clinical studies where expert consensus is needed to establish ground truth from ambiguous or complex data (like interpreting medical images). For the non-clinical testing of this device, the pass/fail criteria are objective and directly measurable.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

This information is not applicable. The device is a surgical tray for sterilization and organizing instruments, not an AI-powered diagnostic or assistive tool for human readers/clinicians.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

This information is not applicable. The device has no algorithmic or AI component.

7. The Type of Ground Truth Used

The "ground truth" for this device's performance is based on:

  • Engineering Specifications and Standards: Defined physical properties (e.g., ability to withstand a specific force, dimensions) and adherence to industry standards (e.g., DIN 58952-3, ASTM D4169-16, ANSI/AAMI ST77:2013, EN ISO 10993-1, EN ISO 17664:2017).
  • Visual Inspection: For integrity, damage, legibility of markings.
  • Functional Testing: To ensure instrument retention and continued functionality after repeated processing.
  • Biocompatibility Testing: To ensure no cytotoxic effects.

8. The Sample Size for the Training Set

This information is not applicable. The device does not involve machine learning or AI, and therefore does not have a training set.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable, as there is no training set for this device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services - USA. To the right of this symbol, there is a blue square with the letters "FDA" in white. Next to the blue square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue.

December 18, 2021

Nobel Biocare Services AG Nick Lewis Senior Regulatory Affairs Manager BOX 5190, SE-402 26, Vastra Hamngatan 1 Goteborg, SE-411 17 Sweden

Re: K212932

Trade/Device Name: Nobel Biocare NobelSpeedy Groovy / Branemark System Mk III TiUnite / Replace Select TC PureSet Tray Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: KCT Dated: November 19, 2021 Received: November 19, 2021

Dear Nick Lewis:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K212932

Device Name

NobelSpeedy® Groovy / Brånemark System® Mk III TiUnite / Replace SelectTM TC PureSet™ Trav

Indications for Use (Describe)

Nobel Biocare PureSet Trays are used in healthcare facilities to store and organize Nobel Biocare surgical/prosthetic instruments and components during cleaning/sterilization and during implant/prosthetic treatment.

Nobel Biocare PureSet Trays are not intended on their own to maintain sterility; they are intended to be used in conjunction with a legally marketed, validated, FDA cleared sterilization pouch, or sterilization wrap.

Sterilization validations for the worst-case PureSet Tray included surgical instruments such as torque wrenches, implant drivers, direction indicators, drills, screw taps, screwdrivers, and irrigation needles. The PureSet Trays were validated for a maximum load of 1635 grams (Trefoil PureSet Tray), 1122 grams (NobelActive / NobelParallel CC PureSet Tray), 1063 grams (NobelReplace CC PureSet Tray), 454 grams (Nobel Biocare N1™ PureSet Tray), 486 grams (Prosthetic PureSet Tray), 1143 grams (NobelActive Guided PureSet Tray), 1146 grams (NobelParallel CC Guided PureSet Tray), 1176 grams (NobelReplace CC Guided PureSet Tray), and 1035 grams (NobelSpeedy® Groovy / Brânemark System® Mk III TiUnite / Replace SelectTM TC PureSet™ Tray).

MethodSteam Sterilization (Moist Heat Sterilization)for Wrapped Instruments
CycleDynamic-Air-Removal(fractionated vacuum)Gravity-Displacement
Temperature132°C (270°F)132°C (270°F)
Exposure time for a single-usepouched device4 minutes (full-cycle)15 minutes (full-cycle)
Minimum drying times20 minutes30 minutes

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Nobel Biocare. The logo consists of a red square with a stylized "N" inside it, followed by the words "Nobel Biocare" in black text. The word "Nobel" is above the word "Biocare", and there is a trademark symbol after the word "Biocare".

510(k) Summary K212932

Nobel Biocare NobelSpeedy®Groovy / Brånemark System® Mk III TiUnite / Replace Select™ TC PureSet™ Tray

    1. Submitter Information:
      Submitter:
Nobel Biocare AB
P.O. Box 5190, 402 26
Västra Hamngatan 1
Goteborg, SE-411 17
Sweden

Submitted By:

Nobel Biocare Services AG
Balz Zimmermann-Str. 7
8302 Kloten
Switzerland
Contact Person:Nick Lewis, M.Sc.
E-Mail:regulatory affairs@nobelbi
E-Mail:regulatory.affairs@nobelbiocare.com
Telephone Number:(+41) 79 941 68 17
Fax Number:(+41) 43 211 42 42

Date Prepared: November 18, 2021

  • Device Name: 2.
    • Proprietary Name: Nobel Biocare NobelSpeedy® Groovy / Brånemark System® Mk III ● TiUnite / Replace SelectTM TC PureSet™ Tray
    • Common Name: Sterilization Wrap Containers, Trays, Cassettes & Other Accessories .
    • Manufacturer: Nobel Biocare Services AG ●
  • CFR Number: 880.6850 ●
    • Classification Name: Sterilization Wrap
    • Device Class: II
    • Product Code: KCT ●

{4}------------------------------------------------

3. Predicate Device

Primary Predicate Device:

  • Proprietary Name: NobelActive / NobelParallel CC PureSet Trav (K181075) ●
  • Common Name: Sterilization Wrap Containers, Trays, Cassettes & Other Accessories .
  • Manufacturer: Nobel Biocare Services AG .
  • CFR Number: 880.6850 ●
  • Classification Name: Sterilization Wrap
  • Device Class: II
  • Product Code: KCT

Predicate Device #2:

  • . Proprietary Name: Nobel Biocare N1 ™ PureSet Tray (K191475)
  • Common Name: Sterilization Wrap Containers, Trays, Cassettes & Other Accessories .
  • Manufacturer: Nobel Biocare Services AG .
  • CFR Number: 880.6850 .
  • Classification Name: Sterilization Wrap ●
  • Device Class: II ●
  • Product Code: KCT ●
    1. Description of Device:

PureSet Trays are reusable surgical trays to be used in combination with Nobel Biocare surgical instruments and components. PureSet Trays are used to organize and store the instruments and components during both surqical and reprocessing procedures.

PureSet Trays are not intended to maintain sterility on their own; they are intended to be used in conjunction with an FDA cleared sterilization wrap, pouch, or container.

All components of the PureSet Tray are perforated with an evenly-distributed hole pattern and are designed to be used for sterilization via steam sterilization. Because the PureSet Trays are perforated, an FDA-cleared sterilization wrap, pouch, or container must be used during sterilization and storage to maintain the sterility of the contents.

PureSet Trays are designed to be used with standard autoclaves used in hospitals and healthcare facilities.

Principle of Operation / Mechanism of Action:

PureSet Trays consist of multiple components (tray base, lid, and plate) integrated into a single unit which is used to organize instruments during surgical procedures and to protect the instruments during transportation, reprocessing, and storage.

{5}------------------------------------------------

Accessories:

There are no accessories for the subject PureSet tray.

    1. Indications for Use:
      Nobel Biocare PureSet Trays are used in healthcare facilities to store and organize Nobel Biocare surgical/prosthetic instruments and components during cleaning/sterilization and during implant/prosthetic treatment.

Nobel Biocare PureSet Trays are not intended on their own to maintain sterility; they are intended to be used in conjunction with a legally marketed, validated, FDA cleared sterilization container, sterilization pouch, or sterilization wrap.

Sterilization validations for the worst-case PureSet Tray included surgical instruments such as torque wrenches, implant drivers, direction indicators, drills, screw taps, screwdrivers, and irrigation needles. The PureSet Trays were validated for a maximum load of 1635 grams (Trefoil PureSet Tray), 1122 grams (NobelActive / NobelParallel CC PureSet Tray), 1063 grams (NobelReplace CC PureSet Tray), 454 grams (Nobel Biocare N1™ PureSet Tray), 486 grams (Prosthetic PureSet Tray), 1143 grams (NobelActive Guided PureSet Tray), 1146 grams (NobelParallel CC Guided PureSet Tray), 1176 grams (NobelReplace CC Guided PureSet Tray), and 1035 grams (NobelSpeedy® Groovy / Brånemark System® Mk III TiUnite / Replace Select™ TC PureSet™ Tray).

MethodSteam Sterilization (Moist Heat Sterilization) forWrapped Instruments
CycleDynamic-Air-Removal(fractionated vacuum)Gravity-Displacement
Temperature132°C(270°F)132°C(270°F)
Exposure time for a single-use pouched device4 minutes(full-cycle)15 minutes(full-cycle)
Minimum drying times20 minutes30 minutes

6. Technological Characteristic Comparison:

Details of the Similarities Between the Subject and Predicate Devices:

The similarities between the NobelSpeedy® Groovy / Brånemark System® Mk III TiUnite / Replace Select™ TC PureSet™ Tray (Subject Device), the Primary Predicate Device NobelActive / NobelParallel CC PureSet Tray (K181075), and Predicate Device #2 Nobel Biocare N1™ PureSet Tray (K191475) as described in Table 5.1 are as follows:

  • The Intended Use statement, design aspects including tray perforation, sterilant . penetration, tray configuration, and reusability, the materials of construction, the

{6}------------------------------------------------

sterilization methods and parameters for reprocessing (including the microbial barriers to be used), the compatibility of the tray materials with the prescribed sterilization methods, and the approach to non-clinical performance testing are all identical for the subject device and both predicate devices.

  • The macro design of the subject device is identical to the primary predicate device. Both ● the subject device and the primary predicate device are single level trays which have an integrated handle. The dimensions of the subject and primary predicate devices are identical, and they have an identical volume to vent ratio.

Details of the Differences Between the Subject and Predicate Devices:

There are no significant differences between the subject and predicate devices but there are minor differences as follows:

  • The Indications for Use for the subject device is similar but not identical to both predicate . devices. It differs from the statement for the primary predicate because it includes the subject device, as well as three trays that were referenced in the Indications statement for predicate device #2. It differs from the statement for predicate device #2 because it includes the subject device, as well as minor updates to the maximum validated load for several of the referenced trays as follows:
    • NobelActive / NobelParallel CC PureSet Tray (PUR0200): 1122 grams instead of O 1082 grams
    • NobelReplace CC PureSet Tray (PUR0300): 1063 grams instead of 945 grams O
    • NobelActive Guided PureSet Tray (PUR0600): 1143 grams instead of 1117 grams O
    • NobelParallel CC Guided PureSet Tray (PUR0700): 1146 grams instead of 1120 O grams
    • NobelReplace CC Guided PureSet Tray (PUR0800): 1176 grams instead of 1156 O grams
  • The macro design of the subject device is similar but not identical to predicate device #2. ● Both trays are single level trays, but the subject tray has an integrated handle whereas predicate device #2 does not.
  • . The dimensions of the subject device are greater than predicate device #2 due to the larger assortment of instruments used with the subject tray. As a result, the volume to vent ratio for the subject device also is greater than predicate device #2.

Table 5.1 presents a comparison of the subject device, primary predicate device, and predicate device #2 with regard to their indications for use, technology, and performance specifications.

{7}------------------------------------------------

Descriptive InformationSubject DevicePrimary Predicate DevicePredicate Device #2Comparison of Subject to Predicate Devices
NobelSpeedy® Groovy / Brånemark System® Mk III TiUnite / Replace Select™ TC PureSet™ TrayNobelActive / NobelParallel CC PureSet Tray (K181075)Nobel Biocare N1 ™ PureSet Tray (K191475)
Indications for UseIndications for UseNobel Biocare PureSet Trays are used in healthcare facilities to store and organize Nobel Biocare surgical/prosthetic instruments and components during cleaning/sterilization and during implant/prosthetic treatment.Nobel Biocare PureSet Trays are not intended on their own to maintain sterility; they are intended to be used in conjunction with a legally marketed, validated, FDA cleared sterilization container, sterilization pouch, or sterilization wrap.Sterilization validations for the worst-case PureSet Tray included surgical instruments such as torque wrenches, implant drivers, direction indicators, drills, screw taps, screwdrivers, and irrigation needles. The PureSet Trays were validated for a maximum load of 1635 grams (Trefoil PureSet Tray), 1122 grams (NobelActive / NobelParallel CC PureSet Tray), 1063 grams (NobelReplace CC PureSet Tray), 454 grams (Nobel Biocare N1 ™ PureSet Tray), 486 grams (Prosthetic PureSet Tray), 1143 grams (NobelActive Guided PureSet Tray), 1146 grams (NobelParallel CC Guided PureSet Tray), 1176 grams (NobelReplace CC Guided PureSet Tray), and 1035 grams (NobelSpeedy® Groovy / Brånemark System® Mk III TiUnite / Replace Select™ TC PureSet™ Tray).The Nobel Biocare PureSet Tray is used in healthcare facilities to store and organize Nobel Biocare surgical instruments and components during cleaning/sterilization and during implant/prosthetic treatment.The Nobel Biocare PureSet Trays are not intended on their own to maintain sterility; it is intended to be used in conjunction with a legally marketed, validated, FDA- cleared sterilization pouch or sterilization wrap.Sterilization validations for the worst-case PureSet Tray included surgical instruments such as torque wrenches, implant drivers, direction indicators, drills, screw taps, screwdriver and irrigation needles. The PureSet Trays were validated for a maximum load of 1635 grams (Trefoil PureSet Tray), 1082 grams (NobelActive/NobelParallel CC PureSet Tray) and 945 grams (NobelReplace CC PureSet Tray).Nobel Biocare PureSet Trays are used in healthcare facilities to store and organize Nobel Biocare surgical/prosthetic instruments and components during cleaning/sterilization and during implant/prosthetic treatment.Nobel Biocare PureSet Trays are not intended on their own to maintain sterility; they are intended to be used in conjunction with a legally marketed, validated, FDA cleared sterilization pouch or sterilization wrap.Sterilization validations for the worst-case PureSet Tray included surgical instruments such as torque wrenches, implant drivers, direction indicators, drills, screw taps, screwdrivers, and irrigation needles. The PureSet Trays were validated for a maximum load of 1635 grams (Trefoil PureSet Tray), 1082 grams (NobelActive / NobelParallel CC PureSet Tray), 945 grams (NobelReplace CC PureSet Tray), 454 grams (Nobel Biocare N1 ™ PureSet), 486 grams (Prosthetic PureSet).The Indications for Use for the subject device is similar to both predicate devices.
MethodCycleTemperatureExposure time for a single-use pouched deviceMinimum drying timesSteam Sterilization (Moist Heat Sterilization)Dynamic-Air-Removal (fractionated vacuum)132°C (270°F)4 minutes (full-cycle)20 minutesGravity-Displacement132°C (270°F)15 minutes (full-cycle)30 minutes
MethodCycleTemperatureExposure time for a single-use pouched deviceMinimum drying timesSteam Sterilization (Moist Heat Sterilization)Dynamic-Air-Removal (fractionated vacuum)132°C (270°F)4 minutes (full-cycle)20 minutesGravity-Displacement132°C (270°F)15 minutes (full-cycle)30 minutes
MethodSteam Sterilization (Moist Heat Sterilization for Wrapped Instruments)
CycleDynamic-Air-Removal (fractionated vacuum)Gravity-Displacement
Temperature132°C (270°F)132°C (270°F)
Exposure time for a single-use pouched device4 minutes (full-cycle)15 minutes (full-cycle)
Minimum drying times20 minutes30 minutes
Descriptive InformationSubject DevicePrimary Predicate DevicePredicate Device #2Comparison of Subject to Predicate Devices
Intended UseNobelSpeedy® Groovy / Brånemark System® Mk III TiUnite / Replace Select™ TC PureSet™ TrayThe Nobel Biocare PureSet Tray is intended for use in healthcare facilities to store and organize Nobel Biocare surgical instruments and components during cleaning/sterilization and during implant/prosthetic treatment. The Nobel Biocare PureSet Trays are not intended on their own to maintain sterility; it is intended to be used in conjunction with a legally marketed, validated, FDA-cleared sterilization container, sterilization pouch, or sterilization wrap. Sterilization validations for the worst-case Nobel Biocare PureSet Tray (276.1 mm x 176 mm x 78 mm) included surgical instruments such as torque wrenches, implant drivers, direction indicators, drills, etc.NobelActive / NobelParallel CC PureSet Tray (K181075)The Nobel Biocare PureSet Tray is intended for use in healthcare facilities to store and organize Nobel Biocare surgical instruments and components during cleaning/sterilization and during implant/prosthetic treatment. The Nobel Biocare PureSet Trays are not intended on their own to maintain sterility; it is intended to be used in conjunction with a legally marketed, validated, FDA-cleared sterilization pouch or sterilization wrap. Sterilization validations for the worst-case Nobel Biocare PureSet Tray (276.1 mm x 176 mm x 78 mm) included surgical instruments such as torque wrenches, implant drivers, direction indicators, drills, etc.Nobel Biocare N1™ PureSet Tray (K191475)The Nobel Biocare PureSet Tray is intended for use in healthcare facilities to store and organize Nobel Biocare surgical instruments and components during cleaning/sterilization and during implant/prosthetic treatment. The Nobel Biocare PureSet Trays are not intended on their own to maintain sterility; it is intended to be used in conjunction with a legally marketed, validated, FDA-cleared sterilization pouch or sterilization wrap. Sterilization validations for the worst-case Nobel Biocare PureSet Tray (276.1 mm x 176 mm x 78 mm) included surgical instruments such as torque wrenches, implant drivers, direction indicators, drills, etc.The Intended Use for the subject device is identical to both predicate devices.
Device Design / Construction / Materials
Representative PictureImage: NobelSpeedy Groovy / Branemark System Mk III TiUnite / Replace Select TC PureSet TrayImage: NobelActive / NobelParallel CC PureSet TrayImage: Nobel Biocare N1 PureSet TrayThe macro design of the subject device is identical to the primary predicate device.
Macro DesignSingle level tray with grommets and a basket for holding tooling in specific locations and covering lid with integrated handle.Single level tray with grommets and a basket for holding tooling in specific locations and covering lid with integrated handle.Single level tray with grommets and a basket for holding tooling in specific locations and covering lid (without handle).
Dimensions (LxWxH)276.1mm x 176mm x 47mm276.1mm x 176mm x 47mm*122.1mm x 115mm x 45.6mmThe dimensions of the subject device are identical to the primary predicate device.
Tray PerforationEvenly distributed hole patternEvenly distributed hole patternEvenly distributed hole patternSubject device is identical to both predicate devices.
Volume to Vent Ratio29.429.4*11.4The volume to vent of the subject device is identical to the primary predicate device.
ConfigurationPerforated bases, lids and PEEK Luvocom grommetsPerforated bases, lids and PEEK Luvocom grommetsPerforated bases, lids and PEEK Luvocom grommetsSubject device is identical to both predicate devices.
ReusableYesYesYesSubject device is identical to both predicate devices.
Sterilization Method• Pre-Vacuum (wrap or pouch)• Gravity Displacement (wrap or pouch)• Pre-Vacuum (wrap or pouch)• Gravity Displacement (wrap or pouch)• Pre-Vacuum (wrap or pouch)• Gravity Displacement (wrap or pouch)Subject device is identical to both predicate devices.
Descriptive InformationSubject DevicePrimary Predicate DevicePredicate Device #2Comparison of Subject to
NobelSpeedy® Groovy / Brånemark System® Mk IIINobelActive / NobelParallel CC PureSet Tray (K181075)Nobel Biocare N1™ PureSet Tray (K191475)Predicate Devices
TiUnite / Replace Select™ TC PureSet™ Tray
Sterilization Parameters• Pre-Vacuum:o Temp 132°C (270° F)o Exposure Time 4 minuteso Pre-vacuum: 4 times < 60 mbaro Drying Time: 20 minuteso Cooling Time: 30 minutes total• Gravity Displacement:o Temp 132°C (270° F)o Exposure Time: 15 minuteso Pre-vacuum: N/Ao Drying Time: 30 minuteso Cooling Time: 30 minutes total• Pre-Vacuum:o Temp 132°C (270° F)o Exposure Time 4 minuteso Pre-vacuum: 4 times < 60 mbaro Drying Time: 20 minuteso Cooling Time: 30 minutes total• Gravity Displacement:o Temp 132°C (270° F)o Exposure Time: 15 minuteso Pre-vacuum: N/Ao Drying Time: 30 minuteso Cooling Time: 30 minutes total• Pre-Vacuum:o Temp 132°C (270° F)o Exposure Time 4 minuteso Pre-vacuum: 4 times < 60 mbaro Drying Time: 20 minuteso Cooling Time: 30 minutes total• Gravity Displacement:o Temp 132°C (270° F)o Exposure Time: 15 minuteso Pre-vacuum: N/Ao Drying Time: 30 minuteso Cooling Time: 30 minutes totalSubject device is identical to bothpredicate devices.
Microbial BarrierFDA cleared sterilization wrap/pouchFDA cleared sterilization wrap/pouchFDA cleared sterilization wrap/pouchSubject device is identical to bothpredicate devices.
Sterilant PenetrationYesYesYesSubject device is identical to bothpredicate devices.
Materials• PureSet Tray:o Tray (including basket / lid / handle / sheet metalparts): Stainless steel (1.4301, 1.4303, 1.4310)o Grommets: PEEK, Stainless steel 1.4310o Tray / Basket Closures: Stainless steel 1.4310,PEEKo Feet: Silicone elastomer• PureSet Tray Plate: Anodized aluminum• PureSet Tray:o Tray (including basket / lid / handle / sheet metalparts): Stainless steel (1.4301, 1.4303, 1.4310)o Grommets: PEEK, Stainless steel 1.4310o Tray / Basket Closures: Stainless steel 1.4310,PEEKo Feet: Silicone elastomer• PureSet Tray Plate: Anodized aluminum• PureSet Tray:o Tray (including lid and sheet metal parts):Stainless steel (1.4301, 1.4303, 1.4310)o Grommets: PEEK, Stainless steel 1.4310o Tray / Basket Closures: Stainless steel 1.4310,PEEKo Feet: Silicone elastomer• PureSet Tray Plate: Anodized aluminumSubject device is identical to bothpredicate devices.
Material Compatibilitywith Sterilization MethodYesYesYesSubject device is identical to bothpredicate devices.
Non-clinical Performance Testing
BiocompatibilityBiocompatibility established via testing performed onrepresentative device (Trefoil PureSet Tray; PUR0100)according to ISO 10993-1:2018, ISO 10993-5:2009, andISO 10993-12:2012Biocompatibility established via testing performed onrepresentative device (Trefoil PureSet Tray; PUR0100)according to ISO 10993-1:2018, ISO 10993-5:2009, andISO 10993-12:2012Biocompatibility established via testing performed onrepresentative device (Trefoil PureSet Tray; PUR0100)according to ISO 10993-1:2018, ISO 10993-5:2009, andISO 10993-12:2012Subject device is identical to bothpredicate devices.
Cleaning / SterilizationCleaning and sterilization method validated via testingperformed on representative device (Trefoil PureSet Tray;PUR0100) according to AAMI TIR12:2010 and ANSI/AAMIST77:2013/(R)2018Cleaning and sterilization method validated via testingperformed on representative device (Trefoil PureSet Tray;PUR0100) according to AAMI TIR12:2010 and ANSI/AAMIST77:2013/(R)2018Cleaning and sterilization method validated via testingperformed on representative device (Trefoil PureSet Tray:PUR0100) according to AAMI TIR12:2010 and ANSI/AAMIST77:2013/(R)2018Subject device is identical to bothpredicate devices.

Table 5.1 Device Comparison Table

{8}------------------------------------------------

{9}------------------------------------------------

" Subsequent to bearne in K18107, he hairling your call CC Pursel Tray was result of the varm, as result of a design charge by helif. Consequenty, the Vourne of the love the was reduced from 40.1 to 29.4.

{10}------------------------------------------------

Non-Clinical Test Data:

The following non-clinical tests were performed as described in Table 5.2 below:

  • . Durability of Tray Handle and Closing Mechanism: The handle and closing mechanism of a representative PureSet tray were able to withstand 250N of force (25 kq of weight) without any permanent distortion, cracking or other evidence of failure to the handle or closing mechanism, which exceeded the 4x weight criteria prescribed in ANSI/AAMI ST77:2013. All test criteria passed.
  • Packaging Performance: Distribution conditioning testing was performed on a . representative PureSet tray according to ASTM D4169-16 and the test results verified the product was damage free and labeling was legible. All test criteria passed.
  • . Simulated Intrafacility Transportation: The subject PUR0900 PureSet tray retained all instruments in their designated holders, and the assembled multi-piece torque wrench remained assembled, during simulated transport of a fully-loaded tray throughout a typical healthcare facility. All test criteria passed.
  • Repeated Reprocessing: The testing demonstrated that the plates and trays can ● withstand at least 250 and 500 reprocessing cycles, respectively, without any loss of functionality or legibility of the markings. There furthermore were no cytotoxic effects observed per EN ISO 10993-1 in any of the tested samples. All test criteria passed.

{11}------------------------------------------------

Test Scope / ObjectivesTest Methods / Applicable StandardsPass/Fail CriteriaTest ResultsConclusion
Durability of Tray Handle andClosing Mechanism:Demonstrate that the tray handleand closing mechanism canwithstand at least 4 times themaximum weight of a fully loadedtray.The testing was performed on the worst-case (heaviest) fully-loadedtray in the PureSet Tray family (PUR0100; 1633 grams fullyloaded). Note that the handle and closing mechanism of the subjectPureSet tray PUR0900 are identical in design to the PUR0100 tray.The base of a closed and latched PureSet tray was fixed in place toa tensile strength testing machine, and the handle of the tray wassubjected increasing force which simulated lifting a tray withincreasing weight.Acceptance criteria were defined by the supplier according to DIN58952-3. Note that the minimum requirements in DIN 58952-3requirements exceed those in ANSI/AAMI ST77:2013, so meetingDIN 58952-3 criteria fulfills ANSI/AAMI ST77:2013 requirements.• The tray closingmechanism does not openwhile the tray is lifted with 4times maximum weight of afully assembled tray (4140grams).• The handle does not breakloose from the tray, andthere is no permanentdistortion, cracking or otherevidence of failure to thehandle or closingmechanism.• Tested to 250N of force (25kg of weight) withoutopening• Handle did not break loosefrom the tray, and therewas no permanentdistortion, cracking or otherevidence of failure to thehandle or closingmechanism. .PASS
Packaging Performance:Demonstrate that the packagingsystem adequately protects the trayagainst the hazards associated withglobal distribution.The testing was performed using the representative worst-case(heaviest) tray (PUR0100). Note that the packaging system for thesubject PureSet Tray PUR0900 is identical to the PUR0100 tray.Packaged trays were placed into a shipping carton and securedwith additional packaging material typical of the Nobel Biocarecentral warehouse.Two group of five sample trays were packaged and sent to the testlaboratory and subjected to climate and distribution conditioningaccording to ASTM D4169-16 using distribution cycle (DC) 13.The conditioned trays were subjected to visual inspection andfunctional testing.• Product shall be damagefree.• The cardboard box and theLD-PE bags containing thePureSet Tray shall beintact.• All box labels shall bepresent and legible.• Product was damage free.• The cardboard box and LD-PE bags were intact.• All box labels were presentand legible.PASS
Simulated IntrafacilityTransportation:Demonstrate that the subjectPUR0900 PureSet tray retains allinstruments in their designatedholders during simulated transportof a fully-loaded tray throughout atypical healthcare facility.The testing was performed using a single example of a fully-loadedPUR0900 PureSet tray, including an assembled torque wrench. TheNobelSpeedy Groovy instrument configuration was employedbecause it contains the most instruments and utilizes all thedifferent types of grommets and holders present in the tray.Three test runs were performed for each method by a singleoperator; two runs were performed a normal walking pace, and onerun at a fast walk approximately twice the normal speed.• The instruments shall stayin the designated locationwithin the tray after eachtest sequence of each testrun.• The assembled TorqueWrench does not becomedisassembled aftertransport during each testsequence.• All instruments remained intheir designated locationsafter each sequence duringeach test run.• The torque wrenchremained assembled aftereach sequence during eachtest.PASS

Table 5.2 Summary of Non-Clinical Test Data

{12}------------------------------------------------

Test Scope / ObjectivesTest Methods / Applicable StandardsPass/Fail CriteriaTest ResultsConclusion
Repeated Reprocessing:Demonstrate that the tray and platecomponents of the PureSet Taycan withstand repeatedreprocessing cycles (500 cycles forthe tray and 250 cycles for theplate) without unacceptabledegradation to the laser markingand printing.The testing was performed using representative PUR0100 TrefoilPureSet trays and plates (all PureSet trays and plates share thesame design, materials, and manufacturing process with regard tolaser marking / printing). One study was performed for the trays,and a separate study performed for the plates.The PureSet plates and trays were repeatedly cleaned andsterilized following the validated methods provided in theInstructions for Use. The plate was subjected to 250 cycles. Twotrays were tested; one was subjected to 250 cycles and the other to500 cycles.The processed trays and plates were then visually inspected andcompared to trays and plates that were not subjected to cleaningand sterilization. The testing was designed to fulfill requirements fordevices subjected to repeated reprocessing described in thefollowing standards:· ANSI/AAMI ST77:2013 (R2018) (Containment devices forreusable medical device sterilization)EN ISO 10993-1 (Biological evaluation of medical devices – Part1: Evaluation and testing within a risk management process)EN ISO 17664:2017 (Processing of health care products -Information to be provided by the medical device manufacturerfor the processing of medical devices)· All laser marking on thetrays and plates shallremain legible, and thecolors on the plate shallremain identifiable and withno detected discoloration orcorrosion.PureSet trays and platesshall remain biocompatibleafter repeated reprocessingfollowing the methodsestablished by NobelBiocare.· The PureSet plates met allacceptance criteria after250 reprocessing cycles.The PureSet trays met allcriteria after both 250 and500 cycles.No cytotoxic effects wereobserved in any of thetested samples.PASS

{13}------------------------------------------------

Clinical Performance Data:

Clinical performance data is not required.

    1. Conclusion:
      The conclusions drawn from the nonclinical test that demonstrate that the subject device (NobelSpeedy® Groovy / Brånemark System® Mk III TiUnite / Replace Select™ TC PureSet™ Tray) is as safe, as effective, and performs as well as or better than the legally marketed primary predicate device (NobelActive / NobelParallel CC PureSet Tray; K181075) and predicate device #2 (Nobel Biocare N1™ PureSet Tray; K191475).

§ 880.6850 Sterilization wrap.

(a)
Identification. A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.(b)
Classification. Class II (performance standards).