K Number
K212905

Validate with FDA (Live)

Device Name
Sotair Device
Date Cleared
2022-08-18

(339 days)

Product Code
Regulation Number
868.5915
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Sotair™ device is intended for use with a manual resuscitator with air-tight connections, in non-breathing patients that require flow controlled ventilation using ambient air or a supplemental oxygen source. The Sotair device is a single use, disposable device added to a manual resuscitator and can be used for in-hospital, emergency, and transport care. The adult Sotair device comprises a flow-limiting valve that limits the inspiratory flow enabling providers to ventilate approximately 55 liters-per-minute (LPM). The flow-limiting valve is intended to minimize gastric inflation during manual ventilation. The Sotair device can be disabled by removing the manual resuscitator to its conventional operation. The Sotair device is intended for adult use only.

Device Description

The Sotair™ device is a universal safety accessory device that is compatible with all manual resuscitators and airway interfaces that have standard ISO 5356-1:2015 15/22 mm connections. The Sotair device's flow-limiting function has been tested with the following bag resuscitators: [List of compatible resuscitators provided in the document]. The Sotair device is not compatible with manual resuscitators that have a built-in flow limiting device. The Sotair device uses a flapper valve with a blade design to limit the flow rate.

AI/ML Overview

The provided text describes the Sotair™ Device, a flow-limiting valve for manual resuscitators, and its equivalence to a predicate device. However, the text does not include a table of acceptance criteria and reported device performance in the specific format requested. It also does not contain information on sample sizes, data provenance, ground truth establishment methods, or the involvement of experts in establishing ground truth for a test set. There's no mention of a multi-reader multi-case (MRMC) comparative effectiveness study or a standalone algorithm performance study, as this is a medical device, not an AI/software product in the traditional sense described by those elements of the prompt.

Based on the available information, here's a breakdown of what can be extracted:

1. Table of Acceptance Criteria and Reported Device Performance:

The document describes various performance tests conducted. While it states that the testing "met the requirements" or "demonstrated that the Sotair device met the product specifications and also performed equivalent to the predicate device," it does not provide specific acceptance values or the quantitative results from these tests in a table format.

Test CategoryAcceptance Criteria (Implied)Reported Device Performance
BiocompatibilityMet requirements of ISO 10993 and ISO 18562"Overall, the testing met the requirements of ISO 10993 and ISO 18562."
Mechanical TestingMet product specifications and performed equivalent to predicate"The performance data demonstrated that the Sotair device met the product specifications and also performed equivalent to the predicate device in the Valve Functionality Testing..."
Human Factors / UsabilityMet criteria for use with intended users, uses, and environments"The study concluded that the Sotair device met the criteria for use with the intended users, uses, and use environments. The overall result... confirm that the Sotair device is designed, functions and meets the expectations and need for use."

2. Sample size used for the test set and the data provenance:

  • Test Set Sample Size: Not explicitly stated for any of the tests.
  • Data Provenance: The studies were conducted by SafeBVM Corporation (submitter) and its contracted consultants/labs. The country of origin for the data is not specified beyond the submitter's address in Chesterfield, MO, USA. The studies are described as "performance testing" and "simulated-use Human Factors Engineering / Usability Engineering (HFE/UE) Summative / Validation Study," implying prospective testing rather than retrospective data analysis.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

This information is not provided. The "ground truth" for the device's performance appears to be established through adherence to engineering standards (ISO) and functional testing against design specifications and a predicate device, rather than expert consensus on a diagnostic output.

4. Adjudication method for the test set:

Not applicable or not described in the document, as it pertains to a physical medical device's performance against predefined criteria and standards.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

Not applicable. The Sotair™ Device is a mechanical medical device (flow-limiting valve), not an AI-powered diagnostic or assistive tool for human readers. Therefore, an MRMC study related to AI assistance is not relevant to this submission.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

Not applicable. The device is a mechanical component of a manual resuscitator, not an algorithm. Its performance is inherent to its physical design and function.

7. The type of ground truth used:

The ground truth used for evaluating the Sotair™ device's performance is based on:

  • Biocompatibility Standards: Adherence to ISO 10993 and ISO 18562.
  • Mechanical and Functional Specifications: Meeting predetermined product specifications for various mechanical properties and functionality (e.g., flow rate, leakage, resistance, valve function).
  • Equivalence to Predicate Device: Direct comparison testing to demonstrate equivalent performance to the legally marketed predicate device (Smart Bag MO) in Valve Functionality Testing.
  • Human Factors/Usability Criteria: Meeting usability criteria for intended users, uses, and use environments as per FDA guidance.

8. The sample size for the training set:

Not applicable. The device is a mechanical product, not an AI model requiring a training set. The descriptions relate to performance testing and usability studies.

9. How the ground truth for the training set was established:

Not applicable. There is no training set for this type of medical device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Safe BVM Corporation % Sara Toyloy President Fabrica Consulting, LLC 73 Lincoln Drive Sausalito, California 94965

Re: K212905

Trade/Device Name: Sotair™ Device Regulation Number: 21 CFR 868.5915 Regulation Name: Manual Emergency Ventilator Regulatory Class: Class II Product Code: BTM Dated: July 14, 2022 Received: July 18, 2022

Dear Sara Toyloy:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Brandon Blakely, Ph.D. Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory, and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K212905

Device Name Sotair™ Device

Indications for Use (Describe)

The Sotair™ device is intended for use with a manual resuscitator with air-tight connections, in non-breathing patients that require flow controlled ventilation using ambient air or a supplemental oxygen source. The Sotair device is a single use, disposable device added to a manual resuscitator and can be used for in-hospital, emergency, and transport care. The adult Sotair device comprises a flow-limiting valve that limits the inspiratory flow enabling providers to ventilate approximately 55 liters-per-minute (LPM). The flow-limiting valve is intended to minimize gastric inflation during manual ventilation. The Sotair device can be disabled by removing the manual resuscitator to its conventional operation. The Sotair device is intended for adult use only.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

5. 510(k) Summary

1. Submitter:SafeBVM, CorporationChesterfield, MO 63017 USA
Phone Number:+1 (314) 973-6606
Sponsor Contact:Haris Shekhani, MDPresident, SafeBVMEmail: haris@safebvm.comPhone: +1 (314) 973-6606
Correspondent:Sara ToyloyFabrica Consulting, LLCEmail: sara@fabricaconsulting.comPhone: +1 (415) 297-8826
Date Prepared:August 17, 2022
    1. Device:
Proprietary Name:Sotair™ Device
Common Name:Flow-Limiting Valve
Regulation:21 CFR § 868.5915
Classification:Manual emergency ventilator
Device Class:Class 2
Device Code:BTM
Review Panel:Anesthesiology

3. Predicate Device:

Smart Bag MO (K021328), O-Two Systems International, Inc. No reference devices were used in this submission

4. Device Description:

The Sotair™ device is a universal safety accessory device that is compatible with all manual resuscitators and airway interfaces that have standard ISO 5356-1:2015 15/22 mm connections. The Sotair device's flow-limiting function has been tested with the following bag resuscitators:

Model NumberDescription/Brand NameManufacturer
520 211 00X, 520 211 00XXAmbu Spur II, adultAmbu
2442-BVMPADCuraplex Resuscitation Bags, adult premiumBVMCuraplex

{4}------------------------------------------------

Model NumberDescription/Brand NameManufacturer
301-558XEACuraplex® VentiSure2™ BVM Manual Resuscitator, AdultCuraplex
04-2K80XXAirLife Adult Manual ResuscitatorVyaire Medical/Carefusion
AF5140MBX, AF5140MBXXSunmed/ventlab AirFlow Standard, Size 5Sunmed/Ventlab
10564XX, 10585XXMercury Medical CPR 2 - Small Adult BagMercury Medical
1056XXX, 10582XXAdult CPR-2 BagMercury Medical
850XX1st Response Adult Manual Resuscitator with Oxygen Reservoir BagSmith's Medical
8451X1The BAG II Resuscitator Adult w/ Mask #5Laerdal
562013Adult BagEasy Resuscitator w/ MaskWestmed
L670-0X0XXDisposable Bag Mask ResuscitatorAllied Healthcare Products
53710XRUSCH, Manual Pulmonary Resuscitator with oxygen reservoir bag, AdultTeleflex

The Sotair device is not compatible with manual resuscitators that have a built-in flow limiting device.

5. Indications for Use:

The Sotair device is intended for use with a manual resuscitator with air-tight connections, in nonbreathing patients that require flow-controlled ventilation using ambient air or a supplemental oxygen source. The Sotair device is a single use, disposable device added to a manual resuscitator and can be used for in-hospital, emergency, and transport care. The adult Sotair device comprises a flow-limiting valve that limits the inspiratory flow enabling user to ventilate approximately 55 liters-per-minute (LPM). The flow-limiting valve is intended to minimize gastric inflation during manual yentilation. The Sotair device can be disabled by removing the device, thereby returning the manual resuscitator to its conventional operation. The Sotair device is for adult use only.

6. Comparison of Indications for Use and Technological Characteristics with the Predicate Device:

Comparison of Indications for Use:

The Sotair device is a flow-limiting valve with the primary function of minimizing gastric inflation during manual ventilation. This intended use statement is the same as the predicate device with regard to the Smart Bag MO's flow-limiting-valve which provides the same primary function of minimizing gastric inflation during manual ventilation. Small differences exist between the two intended uses wherein the Sotair device has a statement indicating that "the device is intended for

{5}------------------------------------------------

use in conjunction with a manual resuscitator which does not have a flow-limiting valve" whereas the predicate Smart Bag MO indicates that "the Smart Bag MO includes a flow-limiting valve that limits the inspiratory flow". Additional differences exist between the two devices including the maximum ventilation rate of ~55 LPM and the targeted patient population (adult use only) for Sotair, however, these differences do not impact the substantial equivalence of the Sotair device as compared to the predicate device.

Comparison of Technological Characteristics:

The Sotair device (subject device) differs from the predicate device with regard to the actual mechanism that is used to limit the flow rate. Specifically, the Sotair device uses a flapper valve with a blade design whereas the Smart Bag MO (predicate device) uses a t-shaped piston with spring design. While the actual morphology differs between the two devices, the mechanism of action is the same for both devices, i.e., the flow-limiting valve becomes activated when the provider squeezes the bag forcefully (i.e., too hard). The Sotair device and the Smart Bag MO mechanisms are both actuated with excessive force and enable the provider to limit inspiratory flows delivered to the patient. When activated, the flow-limiting valve mechanisms of both devices move within their respective housing to limit flow. Both valves are brightly colored to aid in visualization, and their plastic components consist of similar medical-grade polycarbonate material.

7. Performance Data:

To demonstrate that the differences in technological characteristics specific to the flow-limiting valve mechanism do not impact the substantial equivalence of the Sotair device (subject device) to the Smart Bag MO device (predicate device), performance testing was completed to support the determination of substantial equivalence as follows:

Biocompatibility Testing:

Biological safety testing on the final, finished Sotair device was conducted in accordance with ISO 10993 and ISO 18562. The Sotair device has contact with the humidified breathing gas pathway for a limited duration (<24 hours) according to ISO 10993-1 and ISO 18562-1. While the actual clinical use is typically brief, under extreme conditions, the device could be used for up to 6 hours. The Sotair device was evaluated for the following endpoints through testing: Cytotoxicity, Sensitization, Intracutaneous Reactivity, Acute Systemic Toxicity, Material-mediated Pyrogenicity, Volatile Organic Compound (VOC) emissions, Particulates, and Leachable Substances in condensate via ISO 10993-18. Toxicological Risk Assessments were conducted according to ISO 10993-17 and ISO 18562-1. Overall, the testing met the requirements of ISO 10993 and ISO 18562.

Mechanical Testing:

The following mechanical tests were completed at Time 2-years (accelerated) to support the substantial equivalence determination:

  • Auditory Feedback, Bounce Back and Valve Penetration Testing ●
  • Forward and Backward Air Leakage and Expiratory and Inspiratory Resistance Testing ●

{6}------------------------------------------------

  • . Drop Testing
  • . Water Immersion Testing
  • Valve Functionality Testing
  • Bag Valve Connectors* ●
  • . Maximum Threshold Testing
  • Dead Space Testing* ●
  • Shelf Life Testing (comprised of the applicable the mechanical testing described above) ●
  • Resuscitator Compatibility Testing ●
  • Accessory Compatibility Testing .

*Time Zero testing only.

The Sotair device performance testing considered the applicable sections of the following two standards where appropriate:

  • ISO 5356-1:2015 Anaesthetic and respiratory equipment conical connectors . - Part 1: Cones and sockets.
  • ISO 10651-4:2002 Lung ventilators Part 4: Particular requirements for operator-● powered resuscitators.

Device Risk Assessment:

The Sotair device risk assessment was conducted in accordance with ISO 14971 Medical Devices - Application of Risk management to Medical Devices.

Human Factors / Usability Testing:

SafeBVM conducted a simulated-use Human Factors Engineering / Usability Engineering (HFE/UE) Summative / Validation Study (summative study) on the Sotair device. This study was conducted in accordance with the Guidance for Industry and Food and Drug Administration Staff - Applying Human Factors and Usability Engineering to Medical Devices; February 3rd, 2016. The study concluded that the Sotair device met the criteria for use with the intended users, uses, and use environments. The overall result of the Usability Studies confirm that the Sotair device is designed, functions and meets the expectations and need for use.

8. Conclusions:

The performance data demonstrated that the Sotair device met the product specifications and also performed equivalent to the predicate device in the Valve Functionality Testing which included direct comparison to the Smart Bag MO. The simulated-use Human Factors Engineering / Usability Engineering (HFE/UE) Summative / Validation Study (summative study) demonstrated that the critical tasks and severity level were considered acceptable with the analysis of users, uses and use environments. Overall, this data demonstrates that the Sotair device (subject device) is substantially equivalent to the Smart Bag MO (predicate device).

§ 868.5915 Manual emergency ventilator.

(a)
Identification. A manual emergency ventilator is a device, usually incorporating a bag and valve, intended to provide emergency respiratory support by means of a face mask or a tube inserted into a patient's airway.(b)
Classification. Class II (performance standards).