K Number
K212851

Validate with FDA (Live)

Manufacturer
Date Cleared
2022-03-16

(190 days)

Product Code
Regulation Number
892.1560
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use
  1. EchoSK and EchoSGyn Modules are additional software/hardware modules for EchoS Family devices and they are additions in terms of indication for use of the legally marketed EchoS Family.

  2. EchoSK optional module for EchoS Family devices is intended to provide images of, or signals from, inside the body acquired by an appropriately trained professional in a clinical setting for musculoskeletal applications. The EchoSK module can be used both with a convex or a linear probe.

  3. EchoSGyn optional module for EchoS Family devices is intended to provide images of, or signals from, inside the body acquired by an appropriately-trained professional in a clinical setting for Fetal and OB/GYN (useful for visualization of the ovaries, follicles, uterus and other pelvic structures) applications. The EchoSCyn module can be used both with a convex or with an endocavitary probe.

Device Description

EchoSK and EchoSGyn are optional modules for EchoS Family devices.

EchoSK is an optional module of the EchoStudio software that allows the operator to visualize echographic images for musculoskeletal applications. The EchoSK module can be used both with a convex or a linear probe.

EchoSGyn is an optional module of EchoStudio that allows the operator to visualize echographic images for gynecological/obstetric applications. EchoSGyn module can be used both with a convex or with an endocavitary probe.

AI/ML Overview

The provided FDA 510(k) summary (K212851) does not contain information regarding objective performance criteria that the device (EchoSK and EchoSGyn Modules for EchoS Family devices) meets through a specific clinical study with a test set, ground truth, and expert adjudication.

Instead, the submission focuses on demonstrating substantial equivalence to predicate devices (K202683 and K202514) based on:

  • Intended Use: The new modules extend the existing EchoS Family devices to include musculoskeletal (EchoSK) and OB/GYN (EchoSGyn) applications, which are functions already performed by the primary predicate, the ACUSON Sequoia Diagnostic Ultrasound System.
  • Technological Characteristics: The subject device shares similar hardware (EchoS and EchoStation) and software (EchoStudio) with the additional predicate device (EchoS Family K202514). It also operates similarly to the primary predicate in terms of transmitting, receiving, and displaying ultrasound echo data.
  • Non-Clinical Performance Data: The summary lists non-clinical testing performed, including:
    • Biocompatibility (ISO 10993)
    • Electrical Safety and Electromagnetic Compatibility (EN 60601-1 and EN 60601-1-2)
    • Software Validation (ISO 62304)
    • Usability (IEC 62366-1:2015)

The document explicitly states "PERFORMANCE TESTING - CLINICAL: None" and "ANIMAL STUDIES: None". This indicates that the substantial equivalence determination for K212851 was not based on a clinical performance study with defined acceptance criteria for diagnostic accuracy, sensitivity, specificity, or other performance metrics typically associated with an AI-driven or diagnostic device.

Therefore, the requested information elements related to clinical performance criteria and studies cannot be extracted from this document. The device's acceptance is based on its similarity to legally marketed predicate devices, and the demonstration that any differences do not raise new questions of safety or effectiveness, as supported by the non-clinical testing mentioned above.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym and name on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name, "U.S. Food & Drug Administration," written out next to it.

Echolight S.p.A Giuseppe Criscenti Quality, Certification and Regulatory Manager Via Raffaello Sanzio, 18 Lecce, LE 73100 ITALY

March 16, 2022

Re: K212851

Trade/Device Name: EchoSK and EchoSGyn Modules for EchoS Family devices Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic Pulsed Echo Imaging System Regulatory Class: Class II Product Code: IYO, MUA Dated: February 2, 2022 Received: February 8, 2022

Dear Giuseppe Criscenti:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jessica Lamb, Ph.D. Assistant Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K212851

Device Name

EchoSK and EchoSGyn modules for EchoS Family devices

Indications for Use (Describe)

  1. EchoSK and EchoSGyn Modules are additional software/hardware modules for EchoS Family devices and they are additions in terms of indication for use of the legally marketed EchoS Family.

  2. EchoSK optional module for EchoS Family devices is intended to provide images of, or signals from, inside the body acquired by an appropriately trained professional in a clinical setting for musculosketal applications. The EchoSK module can be used both with a convex or a linear probe.

  3. EchoSGyn optional module for EchoS Family devices is intended to provide images of, or signals from, inside the body acquired by an appropriately-trained professional in a clinical setting for Fetal and OB/GYN (useful for visualization of the ovaries, follicles, uterus and other pelvic structures) applications. The EchoSCyn module can be used both with a convex or with an endocavitary probe.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K212851

Section 5 510(k) Summary

I. SUBMITTER

Applicant InformationEcholight S.p.A.
Via Raffaello Sanzio, 18
73100 Lecce, Italy
ContactDr. Giuseppe Criscenti
(Quality, Certification and Regulatory Manager)
Date Prepared1st September 2021
II.DEVICE
Device Trade NameEchoSK and EchoSGyn Modules for EchoSFamily devices
Common NameMusculoskeletal and gynecologic modules for Ultrasound System
Classification Name:Ultrasonic pulsed echo imaging system
Regulation Number:21 CFR 892.1560
Classification Name:Bone Sonometer
Regulation Number:21 CFR 892.1180
Product CodeIYO, MUA

III. PREDICATE DEVICE

Legally Marketed Primary Predicate Device

510(k)NumberPro CodeTrade NameApplicant
K20268321 CFR 892.1550ACUSON Sequoia Diagnostic Ultrasound System,ACUSON SC2000 Diagnostic Ultrasound System,ACUSON Freestyle Diagnostic Ultrasound System,ACUSON S1000, S2000, S3000 DiagnosticUltrasound Systems, ACUSON P200 DiagnosticUltrasound System, ACUSON P500 DiagnosticUltrasound System, ACUSON NX3, NX3 EliteDiagnostic Ultrasound SystemSiemens MedicalSolutions, USA, Inc.

Legally Marketed Additional Predicate Device

510(k)NumberPro CodeTrade NameApplicant
K20251421 CFR 892.1180EchoS FamilyEcholight S.p.a.

{4}------------------------------------------------

IV. Device Description

EchoSK and EchoSGyn are optional modules for EchoS Family devices.

EchoSK is an optional module of the EchoStudio software that allows the operator to visualize echographic images for musculoskeletal applications. The EchoSK module can be used both with a convex or a linear probe.

EchoSGyn is an optional module of EchoStudio that allows the operator to visualize echographic images for gynecological/obstetric applications. EchoSGyn module can be used both with a convex or with an endocavitary probe.

V. Indication for Use

  1. EchoSK and EchoSGyn Modules are additional software modules for EchoS Family devices and they are additions in terms of indication for use of the legally marketed EchoS Family.

  2. EchoSK optional module for EchoS Family devices is intended to provide images of, or signals from, inside the body acquired by an appropriately trained professional in a clinical setting for musculoskeletal applications. The EchoSK module can be used both with a convex or a linear probe.

  3. EchoSGyn optional module for EchoS Family devices is intended to provide images of, or signals from, inside the body acquired by an appropriately-trained professional in a clinical setting for Fetal and OB/GYN (useful for visualization of the ovaries, follicles, uterus and other pelvic structures) applications. The EchoSGyn module can be used both with a convex or with an endocavitary probe.

VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICES

The subject device and the primary predicate device K202683 are both ultrasound devices intended to provide images of inside the body by an appropriately trained professional in a clinical setting for Fetal, OB/GYN and musculoskeletal applications.

EchoSK optional module is intended to provide images of, or signals from, inside the body by an appropriately trained professional in a clinical setting for musculoskeletal applications as the

{5}------------------------------------------------

analogous function performed by the primary predicate K202683 "ACUSON Sequoia Diagnostic Ultrasound System".

EchoSGyn optional module is intended to provide images of, or signals from, inside the body by an appropriately trained professional in a clinical setting for Fetal and OB/GYN (useful for visualization of the ovaries, follicles, uterus and other pelvic structures) applications as the analogous function performed by the primary predicate K202683 "ACUSON Sequoia Diagnostic Ultrasound System".

Considering the technological characteristics, the subject device and the primary predicate device K202683 are both portable and software controlled diagnostic ultrasound systems that transmit, receive and display ultrasound echo data.

Both subject device and ACUSON Sequoia Diagnostic Ultrasound System are powered by, and communicates with a computer. A transducer is used to generate and transmit ultrasound waves into the tissue. The reflected sound waves are detected and digitized before the data is transferred to the computer. Both devices, use an echographic probe that generates and transmits ultrasound waves into the tissue and the detected analog signals are converted into digital ones by means of an analogto-digital converter.

Both the subject device and the ACUSON Sequoia Diagnostic Ultrasound System, respectively through "EchoSK" and ACUSON proprietary softwares, are used as similar device modules for musculoskeletal applications.

Both the subject device and the ACUSON Sequoia Diagnostic Ultrasound System, respectively through "EchoSGyn" and ACUSON proprietary softwares, are used as similar device modules for Fetal and OB/GYN (useful for visualization of the ovaries, follicles, uterus and other pelvic structures) applications.

In addition, the technological characteristics of the subject device are the same of the additional predicate EchoS Family K202514 because they share the same hardware (EchoS and EchoStation devices) and software (EchoStudio).

EchoSK and EchoSGyn can be considered as additional modules of EchoS Family devices meaning that the subject device includes also the bone sonometer (it is the same of the additional predicate device) and it is intended to be used as bone sonometer and for muscoloskeletal and gynecological ultrasound applications.

{6}------------------------------------------------

Considering the hardware, subject and additional predicate devices share identical hardware. In particular, the beamformer is equipped with three SMA connectors for the transmission of the analog radio frequency (RF) signal and the track synchronization signals (Line Trigger) and frame (Frame Trigger). The model used is the Echo Blaster 128 CEXT-1Z, size 62x215x165 (mm), weighing 1.6 kg and is powered by an external medical grade transformer that provides 12VDC, 4 A. This device is already CE certified, KFDA and FDA 510 (k).

The RF signal is suitably filtered and amplified, before being converted into a digital signal. The designed filter is a bandpass type: the lower cutoff frequency eliminates the noise introduced by the beamformer, the higher cutoff frequency eliminates any signal aliasing phenomena. The amplifier has a gain of 18dB.

The analog signal is converted into digital by means of an analog-to-digital converter (ADC) with a resolution of 16bit and a sampling frequency of 40MHz.

The data transmission block consists of an FPGA and a microcontroller: the first appropriately recomposes the data received from the converter and the second transfers them to the PC using the USB protocol.

The scanned frames are available in proprietary format on the PC.

The filtering, amplification, digitization and data transmission phases have been implemented on a single electronic board, with reduced dimensions. The connection between the beamformer and the designed electronic board takes place by means of coaxial cables with SMA connectors. The voltage from the external medical power supply is used to power both the beamformer and the electronic boards. Inside the ABS cover there is also a USB HUB with 4 ports, on 3 of which are connected: the USB Type A- Type B cable connected to the beamfomer, the USB type B - Mini USB cable connected to the data acquisition card, the encryption USB pendrive.

All the electronic components and the cables are enclosed in a plastic ABS case.

The ultrasound transducer is a convex probe (CE marked) and it has a nominal central operating frequency of 3.5MHz and a radius of curvature of 60 mm (field of view 59°).

The newly added transducers are:

The linear probes (CE marked) with a nominal central operating frequency of 7.5 MHz and a length of 4 cm (field of view of 39 mm) or 6 cm (field of view of 59 mm).

The endocavitary probe (CE marked) with a nominal central operating frequency of 6.5MHz and a radius of 10 mm (field of view 156°).

{7}------------------------------------------------

Considering the software, subject and additional predicate devices share identical software named EchoStudio. The subject device adds two additional software modules:

  • EchoSK optional module for EchoS Family devices is intended to provide images of, or . signals from, inside the body acquired by an appropriately trained professional in a clinical setting for musculoskeletal applications. The EchoSK module can be used both with a convex or a linear probe.
  • . EchoSGyn optional module for EchoS Family devices is intended to provide images of, or signals from. inside the body acquired by an appropriately-trained professional in a clinical setting for Fetal and OB/GYN (useful for visualization of the ovaries, follicles, uterus and other pelvic structures) applications. The EchoSGyn module can be used both with a convex or with an endocavitary probe.

VII. Performance Data

The part in contact with patient has been tested and it is compliant with:

Biocompatibility (according to ISO 10993)

The battery of testing included the following tests:

  • Cytotoxicity -
  • । Sensitization
  • Irritation/Intracutaneous -

Electrical Safety, Electromagnetic Compatibility EMC (according to EN 60601-1 and EN 60601-1-2)

The battery of testing included the following tests:

  • Electrical Safety Test:(see attachment 15_A) -
  • Electromagnetic Compatibility: Test Report (see attachment 15_B and attachment 15_C) -

Software Validation (According to ISO 62304)

Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices". The software for this device was considered as a "moderate" level of concern, since a failure or latent flaw in the software could not directly result in serious injury or death to the patient or operator.

{8}------------------------------------------------

510(k) Premarket NotificationEcholight S.p.A.
EchoS Familyrev.06
Section Vpag. 6 of 6

The documentation of Software Validation can be found in the annexes of Section 14

Usability (According to IEC 62366-1:2015)

The usability of the product was conducted in accordance with IEC 62366-1:2015 and concerns various aspects of the device use.

The documentation can be found in the annexes of Section 16:

Usability File (see Attachment 16 A) -

ANIMAL STUDIES

None

PERFORMANCE TESTING - CLINICAL

None

VIII. Conclusions

Echolight S.p.A. has determined, by using comparisons and tests, that EchoSK and EchoSGYN modules for Echostudio software and EchoS Family devices are substantially equivalent to the listed predicate devices in terms of intended use, typical clinical use, operational characteristics, and fundamental technological characteristics. Any differences are considered minor and do not raise new issues of the safety and effectiveness of EchoSK and EchoSGYN modules for Echostudio software and EchoS Family devices device when compared to the predicate devices.

§ 892.1560 Ultrasonic pulsed echo imaging system.

(a)
Identification. An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.