K Number
K212825

Validate with FDA (Live)

Manufacturer
Date Cleared
2021-10-13

(40 days)

Product Code
Regulation Number
882.4560
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

CoreLink Navigation Instruments are intended to be used in the preparation and placement of CoreLink screws (Tiger and Entasis) during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. These instruments are designed for use with the Medtronic StealthStation® System S8 (V1.2.20), which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as vertebra, can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks of the anatomy.

Device Description

CoreLink Navigation Instruments are non-sterile, re-usable manual surgical instruments made from stainless steel. These instruments are designed to interface with the Medtronic StealthStation® Navigation System to assist surgeons in precisely locating anatomical structures.

The purpose of this submission is to clear the CoreLink Navigation Instruments for use with additional versions of the StealthStation software and to add instruments to the previously cleared CoreLink Navigation Instruments set.

AI/ML Overview

This is a 510(k) premarket notification for CoreLink Navigation Instruments.
Here's a breakdown of the requested information based on the provided document:

1. Table of Acceptance Criteria and Reported Device Performance

The document mentions "confirmatory positional accuracy testing per ASTM F2554" but does not explicitly state the acceptance criteria (e.g., a specific tolerance in mm) or the numerical results of this performance testing. It only states that the testing was done to "demonstrate that the system modifications provide adequate and substantially equivalent accuracy for their intended use."

Acceptance Criteria (Not Explicitly Stated in Document)Reported Device Performance
Positional Accuracy per ASTM F2554 (Specific metric not provided)Adequately and substantially equivalent accuracy for intended use (Specific numerical results not provided)

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify the sample size used for the positional accuracy testing. It also does not explicitly state the data provenance (e.g., country of origin, retrospective or prospective).

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

This information is not provided in the document. The testing described focuses on device performance against a standard (ASTM F2554) rather than human expert-established ground truth.

4. Adjudication Method for the Test Set

This information is not provided in the document. Adjudication methods are typically relevant for studies involving human interpretation or subjective assessments, which isn't explicitly described here.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study is not mentioned. The document describes testing of the navigation instruments' accuracy, not a study comparing human reader performance with and without AI assistance.

6. If a Standalone Study (Algorithm Only Without Human-in-the-Loop Performance) Was Done

Yes, a standalone study of the device's performance was done. The document states that CoreLink, LLC "conducted confirmatory positional accuracy testing per ASTM F2554 with Medtronic StealthStation® System S8 Version 1.2.0 (1.2.0-20)." This refers to testing the device itself for accuracy, independent of a human surgeon's interaction in a clinical setting for the purpose of this particular performance data section.

7. The Type of Ground Truth Used

The ground truth for the positional accuracy testing would be established by the specifications and methodology of ASTM F2554, which defines how to measure and assess the accuracy of navigated surgical instruments. This standard typically involves objective measurements against a known reference or ideal position, rather than expert consensus, pathology, or outcomes data.

8. The Sample Size for the Training Set

This device is a set of manual surgical instruments that interface with a navigation system, not an AI or machine learning algorithm that requires a training set in the conventional sense. Therefore, the concept of a "training set" is not applicable here.

9. How the Ground Truth for the Training Set Was Established

As explained above, a training set is not applicable for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health and Human Services and the U.S. Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 13, 2021

CoreLink, LLC % Nathan Wright Engineer & Regulatory Specialist Empirical Testing Corp. 4628 Northpark Drive Colorado Springs, Colorado 80918

Re: K212825

Trade/Device Name: CoreLink Navigation Instruments Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: September 2, 2021 Received: September 3, 2021

Dear Nathan Wright:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA

Page 2

has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICESForm Approved: OMB No. 0910-0120
Food and Drug AdministrationExpiration Date: 06/30/2020
Indications for UseSee PRA Statement on last page.
510(k) Number (if known)K212825
Device NameCoreLink Navigation Instruments
Indications for Use (Describe)
CoreLink Navigation Instruments are intended to be used in the preparation and placement of CoreLink screws (Tiger and Entasis) during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. These instruments are designed for use with the Medtronic StealthStation® System S8 (V1.2.20), which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as vertebra, can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks of the anatomy.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995. DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (7/17)

{3}------------------------------------------------

K212825 - 510(K) Summary

Submitter's Name:CoreLink, LLC
Submitter's Address:2072 Fenton Logistics ParkSt. Louis, Missouri 63026
Submitter's Telephone:888-349-7808
Contact Person:Nathan Wright MSEmpirical Testing Corp.719-351-0248nwright@empiricaltech.comImage: Empirical Testing Corp. Logo
Date Summary was Prepared:September 2, 2021
Trade or Proprietary Name:CoreLink Navigation Instruments
Common or Usual Name:Orthopedic Stereotaxic Instruments
Classification:Class II per 21 CFR §882.4560
Product Code:OLO
Classification Panel:Orthopedic

DESCRIPTION OF THE DEVICE SUBJECT TO PREMARKET NOTIFICATION:

CoreLink Navigation Instruments are non-sterile, re-usable manual surgical instruments made from stainless steel. These instruments are designed to interface with the Medtronic StealthStation® Navigation System to assist surgeons in precisely locating anatomical structures.

The purpose of this submission is to clear the CoreLink Navigation Instruments for use with additional versions of the StealthStation software and to add instruments to the previously cleared CoreLink Navigation Instruments set.

INDICATIONS FOR USE

CoreLink Navigation Instruments are intended to be used in the preparation and placement of CoreLink screws (Tiger and Entasis) during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. These instruments are designed for use with the Medtronic StealthStation® System S8 (V1.2.0), which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as vertebra, can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks of the anatomy.

TECHNOLOGICAL CHARACTERISTICS

The subject and predicate devices have nearly identical technological characteristics and the minor differences do not raise any new issues of safety and effectiveness. Specifically, the following characteristics are identical between the subject and predicates:

  • Indications for use ●
  • Materials of manufacture .

{4}------------------------------------------------

  • Principles of operation ●

Table 5-1 Predicate Devices

510k NumberTrade or Proprietary orModel NameManufacturerPredicateType
K182731CoreLink NavigationInstrumentsCoreLink, LLCPrimary
"K161210"; "K170679"Medtronic NavigatedReusable InstrumentsMedtronic SofamorDanekAdditional

Performance Data

In support of this Special 510(k) submission, CoreLink, LLC has conducted confirmatory positional accuracy testing per ASTM F2554 with Medtronic StealthStation® System S8 Version 1.2.0 (1.2.0-20) to demonstrate that the system modifications provide adequate and substantially equivalent accuracy for their intended use.

CONCLUSION

The overall technology characteristics and performance data lead to the conclusion that the CoreLink Navigation Instruments are substantially equivalent to the predicate device.

§ 882.4560 Stereotaxic instrument.

(a)
Identification. A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.(b)
Classification. Class II (performance standards).