K Number
K212812

Validate with FDA (Live)

Device Name
Surgical Gown
Date Cleared
2022-11-16

(439 days)

Product Code
Regulation Number
878.4040
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Surgical gowns are sterile, disposable, and single use devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. This device is sterilized using Ethylene Oxide (EO) following the Validation and routine control under ANSI/AAMI/ISO 11135 (2014).

Device Description

The Sterile Level 4 Surgical Gown is intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. The proposed devices are single use and EO sterilized. The Sterile Level 4 Surgical Gowns are available in five models including S、 M、L、XL and XXL. The proposed surgical are constructed of a SMS nonwoven material (SMS+Film covering material+Velcro tape+Cuff material) and it is a kind of reinforced surgical gown and blue color.

AI/ML Overview

Here's an analysis of the provided text regarding acceptance criteria and the study that proves the device meets them:

The document is a 510(k) Summary for a "Surgical Gown" (K212812). It focuses on demonstrating substantial equivalence to a predicate device (K121152) through non-clinical performance tests.

1. Table of Acceptance Criteria and Reported Device Performance

The table below summarizes the acceptance criteria and the claimed performance of the subject device, as presented in the "Comparison of technological characteristics with the Predicate" section and the "Test Method" table.

Test MethodPurposeAcceptance CriteriaReported Device Performance
AATCC 127Hydrostatic pressure>50 cmPass
AATCC 42Impact penetration≤1gPass
ASTM D 5034Tensile StrengthLength(lbf): ≥37.6Width(lbf): ≥24.2Pass
ASTM D 5733Tearing strengthLength yarns torn(lbf):14.3Width yarns torn(lbf):8.1Pass
ASTM D 1683Seam strengthArm Opening sleeves: 23.5 lbf (FTS), 24.2lbf (FR)Shoulder: 37.0 lbf (FR)Pass
16 CFR Part 1610Flammability of TextilesClass 1Class 1
ASTM F1671Bacteriophage Phi-X174 penetrationNo penetrationPass
Biocompatibility (ISO 10993-5:2009)CytotoxicityISO 10993-5 (non-cytotoxic)Passed (non-cytotoxic)
Biocompatibility (ISO 10993-10:2002)IrritationISO 10993-10 (non-irritating)Passed (non-irritating)
Biocompatibility (ISO 10993-10:2002)SensitizationISO 10993-10 (non-sensitizing)Passed (non-sensitizing)

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the specific sample sizes used for each individual non-clinical test. It mentions that "Non clinical tests were conducted to verify that the proposed device met all design specifications." Without further detail, it is assumed that appropriate sample sizes, as typically required by the referenced ASTM and ISO standards for material testing, were used.

The data provenance is not explicitly stated in terms of country of origin or whether it's retrospective or prospective. However, given that the submitter is "Xianning Marveldas Protective Articles CO.,Ltd." from China and the tests are for a new device submission, these would be prospective tests performed by the manufacturer, likely at their own facilities or contracted testing labs, to demonstrate compliance.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

This section is not applicable to this submission. The "ground truth" for the performance tests of a surgical gown is established by the specified acceptance criteria of recognized international standards (ASTM, AATCC, ISO, 16 CFR). These standards define objective, measurable physical and chemical properties, rather than subjective interpretations by human experts.

4. Adjudication Method for the Test Set

This section is not applicable. The tests are objective measurements against defined standards. There is no mention of "adjudication" in the context of expert review for establishing ground truth, as the ground truth is the numerical outcome of the physical/chemical tests and comparison against the specified acceptance criteria.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

This section is not applicable. This submission is for a medical device (surgical gown) that is evaluated based on its physical properties and biocompatibility, not on a diagnostic or therapeutic function requiring human-in-the-loop performance or clinical effectiveness studies with human readers. The document explicitly states: "No clinical study is included in this submission."

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

This section is not applicable. The device is a surgical gown, not an AI algorithm or a device with an algorithmic component. The performance evaluated is the physical and biological characteristics of the material.

7. The Type of Ground Truth Used

The ground truth used for these non-clinical performance tests is based on established international and national standards and their defined acceptance criteria. These standards specify the methodology and the expected performance thresholds for various properties of surgical apparel, such as:

  • Physical Property Standards: AATCC 127 (Hydrostatic pressure), AATCC 42 (Impact penetration), ASTM D 5034 (Tensile Strength), ASTM D 5733 (Tearing strength), ASTM D 1683 (Seam strength), 16 CFR Part 1610 (Flammability of Textiles), ASTM F1671 (Bacteriophage Phi-X174 penetration).
  • Biocompatibility Standards: ISO 10993-5 (Cytotoxicity) and ISO 10993-10 (Irritation and Sensitization).

For each test, the "ground truth" is whether the device's measured performance meets or exceeds the specified numerical or categorical acceptance criteria set forth in these standards.

8. The Sample Size for the Training Set

This section is not applicable. There is no "training set" in the context of a surgical gown submission; this term is typically relevant for machine learning or AI algorithm development. The device's characteristics are evaluated directly through testing, not through a model trained on data.

9. How the Ground Truth for the Training Set Was Established

This section is not applicable for the same reasons as point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 16, 2022

Xianning Marveldas Protective Articles CO.,Ltd. % Salon Chen OS Engineer IMD Medical & Drug Technology Service Institutions Room 308, building 11, No. 23, Jinqu Road, Wanjiang Street Dongguan, Guangdong 523069 China

Re: K212812

Trade/Device Name: Surgical Gown Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FYA Dated: October 12, 2022 Received: October 12, 2022

Dear Salon Chen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Bifeng Qian -S

Bifeng Qian, M.D., Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K212812

Device Name Surgical Gown

Indications for Use (Describe)

Surgical gowns are sterile, disposable, and single use devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. This device is sterilized using Ethylene Oxide (EO) following the Validation and routine control under ANSI/AAMI/ISO 11135 (2014).

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary (K212812)

1. Submitter Information

  • Company Name: Xianning Marveldas Protective Articles CO., Ltd.

  • Address: Guishan Road #30, Hi-New Tech Industrial, Xianning, Hubei P R C >
  • Phone: +86-1572066625

  • Contact Person (Title): Zhuxiaokang

  • E-mail: sale@marveldas.com

  • A Date of Preparation: November 10, 2022

2. Subject Device Information

  • Type of 510(k) submission: Traditional

  • Trade Name: Surgical Gown A
  • Common Name: Surgical Gown

  • Classification Name: Surgical Apparel

  • Model: S、M、L、XL、XXL

  • Classification Product Code: FYA

  • Requlation Number: 21 CFR 878.4040

  • Classification: Class II

  • Review Panel: General Hospital

3. Predicate Device Information:

  • 510(k) Number: K121152

  • Predicate Device Name: Surgical Gown

  • Manufacturer: Jiangsu Guangda Medical Material Co., Ltd.

  • This predicate has not been subject to a design-related recall

  • No reference devices were used in this submission.

4. Application Correspondent

  • A Company Name: IMD Medical & Drug technology service institutions
  • Phone: +86-18613190779

  • Fax: +86-755-62809168

  • Contact Person(Title): Salon Chen (System engineer)

  • A E-mail: 33999439@qq.com
  • the address of application correspondent:308, Building 11, Zhongchuang Foreign Exchange Industrial Park, No. 23, Jinqu Road, Wanjiang Street, Dongguan, Guangdong, China

5. Device Description

The Sterile Level 4 Surgical Gown is intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. The proposed devices are single

{4}------------------------------------------------

use and EO sterilized. The Sterile Level 4 Surgical Gowns are available in five models including S、 M、L、XL and XXL.

The proposed surgical are constructed of a SMS nonwoven material (SMS+Film covering material+Velcro tape+Cuff material) and it is a kind of reinforced surgical gown and blue color.

6. Intended Use /Indications for Use

Surgical gowns are sterile, disposable, and single use devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. This device is sterilized using Ethylene Oxide (EO) following the Validation and routine control under ANSI/AAMI/ISO 11135 (2014).

7. Comparison of technological characteristics with the Predicate

Elements of ComparisonPredicate DeviceSubject DeviceComparison
Company NameJiangsu Guangda MedicalMaterial Co., Ltd.Xianning Marveldas ProtectiveArticles CO., Ltd./
510(k) numberK121152K212812/
Device NameSurgical GownSurgical Gown/
Classification Product CodeFYAFYAsame
Regulation21 CFR 878.404021 CFR 878.4040same
Classification NameSurgical ApparelSurgical Apparelsame
Class22same
Prescription or OTCOTCOTCsame
Jiangsu Guangda'sReinforced Surgical Gowns,Model Number GD-SG-01,are non-sterile, single usesurgical gowns intended toprotect surgical patients andoperating room personnelfrom the transfer ofmicroorganisms, body fluids,and particulate material.This product may besterilized using EthyleneOxide (EO) following thevalidation and routine controlunder ANSI/AAMI/ISOSurgical gowns are sterile,disposable, and singleuse devices that are intendedto be worn by operating roomprocedures to protect both thesurgical patient and theoperating room personnel fromtransfer of microorganisms,body fluids, and particulatematerial. This device issterilized using Ethylene Oxide(EO) following the Validationand routine control underANSI/AAMI/ISO 11135 (2014).same
Intended Use & Indicationsfor Use
MaterialsSMS and PE+PP TWO layercompoundSMS+Film coveringmaterial+Velcro tape+Cuffsimilar
Weight Per square(g)45g/m²45g/m²same
DurabilityDisposableDisposablesame
SizeXLS、M、L、XL、XXLdifferent
ColorBlueBluesame
Reinforced areaPE+PP two layer compoundprotective reinforcementwith 100% white Terylenedouble barrier non-woven fabricsame
StyleReinforcedReinforcedsimilar
Hydrostatic pressure:AATCC 127>20cm>50 cmdifferent
Impact penetration:AATCC 42≤1≤1same
Tensile strength: ASTMD5034PassedLength(lbf):37.6Width(lbf):24.2similar
Tearing strength: ASTMD5733PassedLength yarns torn(lbf): 14.3Width yarns torn(lbf): 8.1similar
Seam strength: ASTM D1683PassArm Opening sleeves23.5 lbf (FTS)24.2lbf (FR)Shoulder37.0 lbf (FR)similar
Flammability:16CFR Part 1610Class 1Class 1same
Sterilization methodAlthough sold non-sterile,gown can be EO SterilizedEOsame
Resistance to bacteriophagepenetration ASTM F1671Level 4 per AAMI PB70Level 4 per AAMI PB70same
BiocompatibilityCytotoxicity, ISO10993-5:2009PassedISO 10993-5; Under theconditions of the study, theproposed device extract wasdetermined to be non-cytotoxic
Irritation, ISO10993-10:2002PassedISO 10993-10; Under theconditions of the study, theproposed device extract wasdetermined to be non-irritating
Sensitization, ISO10993-10:2002PassedISO 10993-10; Under theconditions of the study, theproposed device extract wasdetermined to be non-sensitizing
Test MethodPurposeAcceptanceCriteriaResult
AATCC 127Hydrostatic pressure>50 cmPass
AATCC 42Impact penetration≤1gPass
ASTM D 5034Tensile StrengthLength(lbf): ≥37.6Width(lbf): ≥24.2Pass
ASTM D 5733Tearing strengthLength yarnstorn(lbf):14.3Width yarnstorn(lbf):8.1Pass
ASTM D 1683Seam strengthArm Openingsleeves23.5 lbf (FTS)24.2lbf (FR)Shoulder37.0 lbf (FR)Pass
16 CFR Part 1610Flammability of TextilesClass 1Class 1
ASTM F1671Bacteriophage Phi-X174penetrationNo penetrationPass
BiocompatibilityCytotoxicity, ISO10993-5:2009ISO 10993-5ISO 10993-5; Underthe conditions of thestudy, the proposeddevice extract wasdetermined to benon-cytotoxic
Irritation, ISO 10993-10:2002ISO 10993-10ISO 10993-10;Under the conditionsof the study, theproposed deviceextract wasdetermined to benon-irritating
Sensitization, ISO10993-10:2002SO 10993-10ISO 10993-10;Under the conditionsof the study, theproposed deviceextract wasdetermined to benon-sensitizing

{5}------------------------------------------------

{6}------------------------------------------------

Analysis:

  1. In term of key material composition, the surgical gowns are both constructed of a SMS nonwoven material. Other Secondary material is slightly different from those of the predicate device. The proposed device has been passed on the performance and biocompatibility. It is no effect on safety or efficacy.

  2. Although the size is different from those of the predicate device. The proposed device has been passed on the performance testing. It is no effect on safety or efficacy.

  3. In terms of general intended used, performance testing, material composition and configuration, hydrostatic pressure、tensile strength、tearing strength and seam strength are slightly different from those of the predicate device. The proposed device has been tested according to AATCC 127、ASTM D 5034、ASTM D 5733 and ASTM D 1683 respectively, and met the requirement of the standard.

8. Discussion of Non-Clinical Performance Tests Performed

{7}------------------------------------------------

Sterile Level 4 Surgical Gown

Non clinical tests were conducted to verify that the proposed device met all design specifications as was similar to the predicate device. The test results demonstrated that the proposed subject device complies with the following standards:

{8}------------------------------------------------

9. Summary of Clinical Test

No clinical study is included in this submission.

10.Conclusion

The conclusion drawn from the nonclinical tests demonstrates that the subject device in 510(k), Surgical gowns is as safe, as effective, and performs as well as or better than the legally marketed predicate device cleared under K121152.

§ 878.4040 Surgical apparel.

(a)
Identification. Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.(b)
Classification. (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.