K Number
K212758

Validate with FDA (Live)

Device Name
Autoplaque
Date Cleared
2023-05-19

(626 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Autoplaque is intended to provide an optimized non-invasive application to analyze coronary anatomy and pathology and aid in determining treatment paths from a set of Computed Tomography (CT) Angiographic images.

Autoplaque is a workstation-based post processing application. It is a non-invasive diagnostic reading software intended for use by cardiologists and radiologists as an interactive tool for viewing and analyzing cardiac CT data for determining the presence and extent of coronary plaques and luminal stenoses.

The software is not intended to replace the skill and judgment of a qualified medical practitioner and should only be used by people who have been appropriately trained in the software's functions, capabilities and limitations. Users should be aware that certain views make use of interpolated data. This is data that is created by the software based on the original data set. Interpolated data may give the appearance of healthy tissue in situations where pathology that is near or smaller than the scanning resolution may be present.

Autoplaque must be installed on a suitable commercial computer platform. It is the user's responsibility to ensure the monitor quality and ambient light conditions are consistent with the clinical applications.

Typical users of Autoplaque are trained medical professionals, including but not limited to radiologists, clinicians, technologists, and others.

Device Description

Autoplaque 3.0. a stand-alone software, performs post-processing of coronary Computed Tomography Angiography (CTA) images and measurements of components of images using computerized algorithms.

Autoplaque 3.0 aids the physician with measurement of coronary artery stenosis and provides measurements for coronary plaque and coronary artery remodeling. Autoplaque 3.0 does not replace standard clinical practice or clinician decision making.

Autoplaque 3.0 allows for standardized characterization of plaque and stenosis from DICOM image data (loaded from the local computer hard drive) and includes the following features:

  • Review of heart and coronary vessels in Multiplanar Reformatting (MPR), curved MPR, and straightened MPR views:
  • Measurement of vessel diameter and area;
  • Characterization and measurement of plaque parameters; and
  • Measurement of lumen diameter, area, and luminal stenosis.

Autoplaque 3.0 includes automated vessel, plaque and lumen segmentation, which is reviewed and can be edited, if necessary, by the clinician.

Autoplaque 3.0 can run on Windows or Mac computer platforms. The minimum hardware specifications are specified in the user manual.

AI/ML Overview

Here's an analysis of the acceptance criteria and the studies conducted for Autoplaque 3.0, based on the provided text:

Acceptance Criteria and Device Performance for Autoplaque 3.0

The acceptance criteria for Autoplaque 3.0 focused on establishing concordance and agreement for various plaque and stenosis measurements compared to expert reader measurements and the predicate device. The primary statistical measures used were the intraclass correlation coefficient (ICC) and correlation coefficient. While specific numerical thresholds for "excellent agreement" or "excellent correlation" are not explicitly stated as acceptance criteria, the studies report achieving this level of agreement.

1. Table of Acceptance Criteria and Reported Device Performance

MetricAcceptance Criteria (Implied)Reported Device Performance
For Plaque Volume Measurements (Total Plaque, Calcified Plaque, Non-Calcified Plaque) and Diameter Stenosis:
Agreement with Expert Reader (US Population)Excellent agreement (ICC and correlation coefficient)Excellent agreement shown (ICC and correlation coefficient)
Agreement with Expert Reader (Non-US Population)Excellent agreement (ICC and correlation coefficient)Excellent agreement shown (ICC and correlation coefficient)
Agreement with Predicate DeviceExcellent correlation and intraclass correlation agreementExcellent correlation and intraclass correlation agreement
Analysis Time Per Lesion (Compared to Experts)Significantly faster than experts (for plaque analysis)5.7 seconds (Autoplaque 3.0) vs. 25-30 minutes (experts)
Analysis Time Per Lesion (Compared to Predicate)Significantly faster than predicate device (for plaque analysis)<2 seconds (Autoplaque 3.0) vs. 30 seconds (predicate)

2. Sample Sizes and Data Provenance

Test Set - Comparison with Expert Reader (US Population):

  • Sample Size: 201 patients, with a total of 781 lesions analyzed.
  • Data Provenance: Retrospective, clinically indicated coronary computed tomography angiography (CCTA) images from multiple U.S. sites. Ethnicity information was available for 80 patients: White alone (not Hispanic or Latino) 18%, Black 19%, Hispanic or Latino 61%, and American Indian and Alaska native alone 1%.

Test Set - Comparison with Expert Reader (Non-US Population):

  • Sample Size: 175 patients, with a total of 1081 lesions analyzed.
  • Data Provenance: Retrospective, clinically indicated coronary computed tomography angiography (CCTA) images from multiple non-U.S. external sites.

Test Set - Comparative Study with Predicate Device:

  • Sample Size: 27 patients, with a total of 30 lesions analyzed.
  • Data Provenance: Not explicitly stated as retrospective or prospective, but derived from clinically indicated CCTA images obtained from a commercial Siemens CT scanner.

Training Set:

  • Sample Size: Not specified in the provided text.
  • Data Provenance: Not specified in the provided text.

3. Number and Qualifications of Experts for Ground Truth

For US Patient Population Study Validation:

  • Number of Experts: One
  • Qualifications: Board-certified imaging cardiologist.

For Non-US Patient Population Study Validation:

  • Number of Experts: Two cardiologists and one additional highly experienced radiologist.
  • Qualifications: Two cardiologists, one highly experienced radiologist.

4. Adjudication Method for the Test Set

For US Patient Population Study Validation:

  • Adjudication Method: Not explicitly stated, as only one expert was mentioned for ground-truthing. This suggests a direct comparison method.

For Non-US Patient Population Study Validation:

  • Adjudication Method: Discrepancies between the two cardiologists were resolved by an additional highly experienced radiologist. This implies a 2+1 adjudication model.

For Comparative Study with Predicate Device:

  • Adjudication Method: Not applicable, as this study compared the two devices rather than devices to human ground truth.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No MRMC comparative effectiveness study was specifically described where human readers improved with AI vs. without AI assistance. The clinical validation studies focused on comparison of the device's output to expert measurements, rather than assessing an AI-assisted workflow for human readers. However, the significantly reduced analysis time for the device compared to experts (5.7 seconds vs. 25-30 minutes) implies a substantial efficiency improvement when using the AI tool.

6. Standalone Performance Study

Yes, a standalone performance study was conducted. Both the "Comparison with Expert Reader Measurements in a Multicenter US Patient Population" and "Comparison with Expert Reader Measurements in a Non-US Patient Population" evaluated the Autoplaque 3.0 device's performance independently against expert ground truth. The device generates its output (plaque volumes, diameter stenosis) without human intervention in the measurement process, although the output can be reviewed and edited by clinicians.

7. Type of Ground Truth Used

For comparison with expert readers (both US and Non-US populations):

  • Ground Truth Type: Expert Consensus/Measurements. The ground truth was established by human expert readers (board-certified imaging cardiologist, cardiologists, and highly experienced radiologist) directly performing the measurements on the CCTA images.

8. Sample Size for the Training Set

The sample size for the training set is not explicitly mentioned in the provided text. The document focuses on the validation studies performed after the device was developed.

9. How Ground Truth for Training Set was Established

The method for establishing ground truth for the training set is not explicitly mentioned in the provided text. However, given that the device uses "Deep learning-based vessel, plaque and lumen segmentation," it is highly probable that the training data's ground truth would have been established through expert manual annotations or segmentations on CCTA images, similar to how the validation ground truth was established. The text only mentions that the deep learning-based segmentation was "validated with ground truth by clinician expert readers."

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Cedars-Sinai Medical Center % Philip Won Hyman, Phelps & McNamara, P.C. 700 Thirteenth Street NW, Suite 1200 WASHINGTON DC 20005

Re: K212758

May 19, 2023

Trade/Device Name: Autoplaque 3.0 Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: QIH, LLZ Dated: April 24, 2023 Received: April 25, 2023

Dear Philip Won:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jessica Lamb

Jessica Lamb, Ph.D. Assistant Director Imaging Software Team DHT 8B: Division of Radiological Imaging Devices and Electronic Products OHT 8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K212758

Device Name Autoplaque 3.0

Indications for Use (Describe)

Autoplaque is intended to provide an optimized non-invasive application to analyze coronary and pathology and aid in determining treatment paths from a set of Computed Tomography (CT) Angiographic images.

Autoplaque is a workstation-based post processing application. It is a non-invasive diagnostic reading software intended for use by cardiologists and radiologists as an interactive tool for viewing and analyzing cardiac CT data for determining the presence and extent of coronary plaques and luminal stenoses.

The software is not intended to replace the skill and judgment of a qualified medical practitioner and should only be used by people who have been appropriately trained in the software's functions, capabilities and limitations. Users should be aware that certain views make use of interpolated data. This is data that is created by the software based on the original data set. Interpolated data may give the appearance of healthy tissue in situations where pathology that is near or smaller than the scanning resolution may be present.

Autoplaque must be installed on a suitable commercial computer platform. It is the user's responsibility to ensure the monitor quality and ambient light conditions are consistent with the clinical applications.

Typical users of Autoplaque are trained medical professionals, including but not limited to radiologists, clinicians, technologists, and others.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(K) SUMMARY

Cedars-Sinai Medical Center's Autoplaque 3.0

In accordance with 21 CFR § 807.92 the following summary of information is provided:

I. Submitter

Cedars-Sinai Medical Center 8700 Beverly Blvd. Los Angeles, CA 90048

Contact Person: Damini Dey, Ph.D.

Title: Director, Quantitative Image Analysis Program, Biomedical Imaging Research Institute (BIRI), Department of Biomedical Sciences, Cedars Sinai Medical Center Cedars-Sinai Medical Center Phone: 310-423-1517

Fax: 310-423-8396

Date of 510(k) Summary Preparation: II.

May 2, 2023

III. Device

Device Proprietary Name:Autoplaque 3.0
Common or Usual Name:Image Processing System, Radiological
Classification Name:Automated Radiological Image Processing SoftwareRadiological Imaging Processing System
Regulation Number:21 CFR § 892.2050
Product Code:QIH, LLZ
Device ClassificationII

IV. Predicate Device

Primary Predicate Device:

  • Autoplaque add-on ORS Visual, K122429, Object Research Systems (ORS) Inc. .

{4}------------------------------------------------

V. Device Description

Autoplaque 3.0. a stand-alone software, performs post-processing of coronary Computed Tomography Angiography (CTA) images and measurements of components of images using computerized algorithms.

Autoplaque 3.0 aids the physician with measurement of coronary artery stenosis and provides measurements for coronary plaque and coronary artery remodeling. Autoplaque 3.0 does not replace standard clinical practice or clinician decision making.

Autoplaque 3.0 allows for standardized characterization of plaque and stenosis from DICOM image data (loaded from the local computer hard drive) and includes the following features:

  • Review of heart and coronary vessels in Multiplanar Reformatting (MPR), curved MPR, . and straightened MPR views:
  • Measurement of vessel diameter and area; .
  • Characterization and measurement of plaque parameters; and
  • . Measurement of lumen diameter, area, and luminal stenosis.

Autoplaque 3.0 includes automated vessel, plaque and lumen segmentation, which is reviewed and can be edited, if necessary, by the clinician.

Autoplaque 3.0 can run on Windows or Mac computer platforms. The minimum hardware specifications are specified in the user manual.

VI. Indications for Use

Autoplaque is intended to provide an optimized non-invasive application to analyze coronary anatomy and pathology and aid in determining treatment paths from a set of Computed Tomography (CT) Angiographic images.

Autoplaque is a workstation-based post processing application. It is a non-invasive diagnostic reading software intended for use by cardiologists and radiologists as an interactive tool for viewing and analyzing cardiac CT data for determining the presence and extent of coronary plaques and luminal stenoses.

The software is not intended to replace the skill and judgment of a qualified medical practitioner and should only be used by people who have been appropriately trained in the software's functions, capabilities and limitations. Users should be aware that certain views make use of interpolated data. This is data that is created by the software based on the original data set. Interpolated data may give the appearance of healthy tissue in situations where pathology that is near or smaller than the scanning resolution may be present.

{5}------------------------------------------------

Autoplaque must be installed on a suitable commercial computer platform. It is the user's responsibility to ensure the monitor quality and ambient light conditions are consistent with the clinical applications.

Typical users of Autoplaque are trained medical professionals, including but not limited to radiologists, clinicians, technologists, and others.

VII. Comparison of Technological Characteristics

Autoplaque 3.0 and the predicate device are identical with respect to intended use and technological characteristics with the exception of the following:

  • Compatibility of the subject device with Mac operating system (OS);
  • Provision of the subject device as stand-alone software versus an add-on; and ●
  • . Deep learning-based vessel, plaque and lumen segmentation, which is reviewed and can be edited, if necessary, by the clinician.

The table below compares the key technological features between the subject and predicate devices.

ParameterSubject Device: Autoplaque 3.0(K212758)Predicate Device: Autoplaque add-onORS Visual(K122429)
Computer operating systemWindows OSMac OSWindows OS only
Stand-alone softwareYesNo, ORS Visual add-on
DICOM complianceDICOM 3.1Same
2D ImagingReview of coronary vessels in 2DMPR, curved MPR, and straightenedview.Same
2D Measurement2D measurement tools of vesseldiameter and contourSame
3D ImagingReview of structures in 3DSame
Maximum intensityprojection (MIP)MIP with interactive controlSame
Multiplanar reformatting(MPR)MPR with oblique slicing andvariable slab thicknessSame
Quantitative MeasurementsNCP volume: non-calcifiedplaque volumeYesSameCP volume: calcified plaquevolumeYesSameLD-NCP volume: low-density noncalcified plaquevolumeYesSameTotal plaque volumeYesSameVessel volumeYesSame
Quantitative Measurements
NCP volume: non-calcifiedplaque volumeYesSame
CP volume: calcified plaquevolumeYesSame
LD-NCP volume: low-density noncalcified plaquevolumeYesSame
Total plaque volumeYesSame
Vessel volumeYesSame

{6}------------------------------------------------

ParameterSubject Device: Autoplaque 3.0(K212758)Predicate Device: Autoplaque add-onORS Visual(K122429)
NCP burden: noncalcified plaque volume/analyzed vessel volumeYesSame
LD-NCP burden: low-density noncalcified plaque volume/analyzed vessel volumeYesSame
CP burden: calcified plaque volume/analyzed vessel volumeYesSame
Total plaque burden: total plaque volume/analyzed vessel volumeYesSame
Plaque composition NCP: Noncalcified plaque composition (NCP volume /total plaque volume)YesSame
Plaque composition CP: Calcified plaque composition (CP volume/total plaque volume)YesSame
Plaque composition LD-NCP: low-density noncalcified plaque composition (LD-NCP volume/ NCP volume)YesSame
Diameter stenosis: maximal diameter stenosis, with respect to proximal and distal referencesYesSame
QCAD: Maximal diameter stenosis, with respect to proximal and distal referencesYesSame
Remodeling index: ratio of maximum vessel area/proximal and distal referencesYesSame
Area stenosis: maximum area stenosis, with respect to proximal and distal referencesYesSame
Plaque length: diseased vessel lengthYesSame
Contrast density difference: maximum difference in contrast density over lesion with respect to proximalYesSame
MLD: minimal luminal diameter over lesionYesSame
ParameterSubject Device: Autoplaque 3.0(K212758)Predicate Device: Autoplaque add-onORS Visual(K122429)
MLA: minimum luminalarea over lesionYesSame
Vessel profile: Area,maximum diameter,minimum diameter measuredfrom selected vessel crosssectionYesSame
Lumen profile: Area,maximum diameter,minimum diameter measuredfrom selected lumen crosssectionYesSame
Vessel, plaque, and lumensegmentationDeep learning based, validated withground truth by clinician expertreaders, with full option to edit.Algorithm based, validated with groundtruth by clinician expert readers, with fulloption to edit.

{7}------------------------------------------------

Discussion: As shown above, there are slight technological differences between the subject and predicate devices with respect to computer operating system, provision of the product as standalone software, and vessel, plaque, and lumen segmentation method; however, these differences do not raise new questions of safety and effectiveness, and the differences are addressed through software verification and validation as well as clinical performance studies.

VIII. Performance Data

Software verification and validation were conducted on the subject device to demonstrate that the subject device meets specifications and works as intended.

In addition, clinical validation studies were undertaken to verify and validate that Autoplaque 3.0 performs as intended. A summary of these studies is provided below.

1. Comparison with Expert Reader Measurements in a Multicenter US Patient Population

Performance data was collected on a data set of clinically indicated coronary computed tomography angiography (CCTA) images from multiple U.S. sites. The patient population consisted of 201 patients (53% men, 47% women) with a mean age of 60 ± 12 years. A total of 781 lesions from 201 patients were analyzed. Ethnicity available in 80 patients was as follows: White alone (not Hispanic or Latino) 18%, Black 19%, Hispanic or Latino 61%, and American Indian and Alaska native alone 1%. The CT scanners to collect images included commercial CT scanners from Siemens, Philips and GE. Ground truthing was performed by a board-certified imaging cardiologist. This study established concordance and agreement between clinician expert reader and Autoplaque 3.0 measurements for total plaque volume, calcified plaque volume, non-calcified plaque volume, and diameter stenosis using the acceptance criteria of an intraclass correlation coefficient and correlation coefficient. An excellent agreement was

{8}------------------------------------------------

shown between the subject device and expert reader measurements for total plaque volume, noncalcified plaque volume, calcified plaque volume, and diameter stenosis.

2. Comparison with Expert Reader Measurements in a Non-US Patient Population

Performance data was collected on clinically indicated coronary computed tomography angiography (CCTA) images from multiple non-U.S. external sites. The patient population consisted of 175 patients (53% men, 47% women) with a mean age of age 56 ± 9 years. A total of 1081 lesions from 175 patients were analyzed. The CT scanners to collect images included commercial CT scanners from Siemens, Philips and Canon. Ground truthing was performed by two cardiologists with one additional highly experienced radiologist to resolve discrepancies. This study established concordance and agreement between clinician expert reader and Autoplaque 3.0 measurements for total plaque volume, calcified plaque volume, non-calcified plaque volume, and diameter stenosis using the acceptance criteria of an intraclass correlation coefficient and correlation coefficient. An excellent agreement was shown between the subject device and expert reader measurements for total plaque volume, noncalcified plaque volume, calcified plaque volume, and diameter stenosis. In addition, the average per-patient plaque analysis time was 5.7 seconds for the subject device versus 25-30 minutes taken by experts.

3. Comparative Study with the Predicate Device

To establish that the subject device provides clinical performance for total plaque volume, calcified plaque volume, non-calcified plaque volume, and diameter stenosis that is substantially equivalent to the predicate device, a total of 30 lesions (from 27 patients) throughout the coronary artery tree were analyzed using both the subject and predicate devices. The patient population were 60% men and 40% women with a mean of age 64.5 ± 10.1 years. The CT scanner used to collect the images was a commercial Siemens CT scanner. There was an excellent correlation between the subject and predicate devices for total plaque volume, noncalcified volume, calcified plaque volume, and diameter stenosis. These measurements also demonstrated excellent intraclass correlation agreement. It took less than 2 seconds per lesion for the subject device Autoplaque 3.0 to analyze, which was much faster than the predicate device (30 seconds per lesion).

IX. Conclusion

The information provided above supports that Autoplaque 3.0 is as safe and effective as the predicate device. Although there are minor differences between the subject and predicate devices, the differences are addressed through software verification and validation studies, and performance testing. Therefore, it is concluded that Autoplaque 3.0 is substantially equivalent to the predicate device.

N/A