K Number
K212731

Validate with FDA (Live)

Device Name
Circul8 Pro
Manufacturer
Date Cleared
2021-11-15

(77 days)

Product Code
Regulation Number
870.5800
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Circul8 Pro Vascular Therapy System is intended to be an easy to use portable system, prescribed by a physician, for use in the home or clinical setting to help prevent the onset of DVT in patients by stimulating blood flow in the extremities (stimulating muscle contractions). This device can be used to:

  • · Aid in the prevention of DVT;
  • · Enhance blood circulation;
  • · Diminish post-operative pain and swelling;
  • · Reduce wound healing time;
  • · Aid in the treatment and healing of stasis dermatits, venous stasis ulcers, arterial and diabetic leg ulcers, chronic venous insufficiency and reduction of edema in the lower limbs.
    The unit can also be used as an aid in the prophylaxis for DVT by persons expecting to be stationary for long periods of time.
Device Description

The Circul8 Pro Vascular Therapy System is a lightweight, portable, rechargeable battery powered prescription device. The device is intended to be used in the home or clinical setting by or under the direction of a medical professional by any patient needing venous return due to an increased risk from blood clots.
The subject device is a modification of the predicate device in K193020, which includes:

  • modification of the plastic housing the electronically controlled air pump unit and solenoid valve,
  • relocation of the tactile touch control switch from the front of the housing to the housing,
  • introduction of LED display to replace the tricolored LED for ON, LOW BATTERY, CHARGE COMPLETED indications, and the blue LED for indicating a leak or low pressure alarm,
  • removal of the micro USB port for obtaining usage (patient compliance) data,
  • no longer labeling cuff as either left leg cuff or right leg cuff,
  • cuffs are filled to 55 mmHg pressure
  • introduces Night Mode, which minimizes light disturbances from the LEDs during patient resting times.
  • increased battery life.
    Both the subject device and predicate device utilize pneumatically controlled, single chamber cuffs actuated by an electronically controlled air pump unit and solenoid value of trol components are protectively housed in a plastic case that is permanently attached to the inflatable cuff.
AI/ML Overview

Here's a breakdown of the acceptance criteria and the study information for the Circul8 Pro Vascular Therapy System, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific Acceptance CriterionReported Device Performance
Physical Requirements1. Unit size: No smaller than 140 X 70 X 40 mmComplies (True mean proportion, 5% margin of error, 90% CL)
2. Unit weight: No more than 0.5 kgComplies (True mean proportion, 5% margin of error, 90% CL)
3. Housing material of construction: ABSComplies (True mean proportion, 5% margin of error, 90% CL)
4. On/Off switch location: Top side of the deviceComplies (True mean proportion, 5% margin of error, 90% CL)
5. LED screen displays: "LP" for low pressure, "55mmHg" for operating pressure, " " during cycles, battery statusComplies (True mean proportion, 5% margin of error, 90% CL)
6. Device displays usage hours when powered onComplies (True mean proportion, 5% margin of error, 90% CL)
7. Leg wrap size/shape: Identical (no distinctions between Left/Right cuffs)Complies (True mean proportion, 5% margin of error, 90% CL)
Performance Requirements1. Inflation pressure accuracy: 55 mmHg ± 10%Complies (True mean proportion, 5% margin of error, 90% CL)
2. Night mode: Performs as specified in software design inputComplies (True mean proportion, 5% margin of error, 90% CL)
Electrical Requirements1. Battery: 3.7V, 1800mAh; Power Supply: 100-240 Vac, 50-60 Hz, Output: 5Vdc @ 2.0 AmpComplies (True mean proportion, 5% margin of error, 90% CL)
Electrical Safety StandardsIEC 60601-1:2005+AMD1:2012+AMD2:2020 (Basic safety and essential performance)Complies
IEC 60601-1-2:2014 (EMC)Complies
IEC 60601-1-11:2015+AMD1:2020 (Home healthcare environment)Complies
IEC 62133-2:2017+AMD1:2021 (Lithium systems safety)Complies
Software VerificationConformance with FDA Guidance for Software Contained in Medical DevicesCarried out according to guidance

2. Sample Size Used for the Test Set and Data Provenance

The document mentions "Design verification test results demonstrate the true mean (average) proportion, with a 5% margin of error and at the 90% confidence level," for various physical, performance, and electrical requirements. However, the specific sample size (N number of devices or measurements) for these tests is not provided in the given text.

The document does not provide information on the country of origin of the data or whether the tests were retrospective or prospective. Given the nature of design verification tests, they are typically prospective, conducted on manufactured units.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not provided in the document. The tests described (measurements of size, weight, pressure, electrical characteristics, software function, and compliance with standards) are objective engineering and performance evaluations. There is no indication that expert "ground truth" was established in the clinical sense for the described verification tests.

4. Adjudication Method for the Test Set

This information is not applicable and therefore not provided in the document. The tests performed are objective measurements and compliance checks against predefined specifications and standards, not evaluations requiring expert adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not conducted and is not mentioned in the document. The Circul8 Pro Vascular Therapy System is a physical therapy device (compressible limb sleeve), not an AI-assisted diagnostic or imaging device that would typically involve human readers.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

This question is not applicable to the Circul8 Pro Vascular Therapy System. It is a physical device, and while it has software for controlling its functions, it is not an algorithm that operates in a standalone diagnostic or interpretative capacity. Its performance is evaluated based on its physical, electrical, and functional compliance.

7. The Type of Ground Truth Used

The "ground truth" for the tests described is based on engineering specifications, physical measurements, functional performance parameters (e.g., pressure output, battery voltage), and compliance with established electrical safety and electromagnetic compatibility international standards (IEC 60601 series, IEC 62133-2). There is no "expert consensus," "pathology," or "outcomes data" in the traditional clinical sense being used as ground truth for these design verification tests.

8. The Sample Size for the Training Set

This information is not applicable and therefore not provided. The device is not a machine learning or AI-based system that requires a "training set" in the context of developing an algorithm from data. The software updates mentioned are for controlling the device's functions and user interface, not for learning from a dataset.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable as there is no "training set" in the context of machine learning for this device.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo is composed of two parts: the Department of Health and Human Services logo on the left and the FDA acronym and full name on the right. The Department of Health and Human Services logo is a stylized human figure, while the FDA acronym and full name are written in blue. The full name of the agency is "U.S. Food & Drug Administration".

November 15, 2021

Ortho8 Inc. % John Beasley Senior Consultant MedTech Review, LLC 257 Garnet Garden Street Henderson, Nevada 89015

Re: K212731

Trade/Device Name: Circul8 Pro Vascular Therapy System Regulation Number: 21 CFR 870.5800 Regulation Name: Compressible Limb Sleeve Regulatory Class: Class II Product Code: JOW Dated: October 15, 2021 Received: October 18, 2021

Dear John Beasley:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

for Nicole Gillette Acting Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

Submission Number (if known)

K212731

Device Name

Circul8 Pro Vascular Therapy System

Indications for Use (Describe)

The Circul8 Pro Vascular Therapy System is intended to be an easy to use portable system, prescribed by a physician, for use in the home or clinical setting to help prevent the onset of DVT in patients by stimulating blood flow in the extremities (stimulating muscle contractions). This device can be used to:

  • · Aid in the prevention of DVT;
  • · Enhance blood circulation;
  • · Diminish post-operative pain and swelling;
  • · Reduce wound healing time;

· Aid in the treatment and healing of stasis dermatits, venous stasis ulcers, arterial and diabetic leg ulcers, chronic venous insufficiency and reduction of edema in the lower limbs.

The unit can also be used as an aid in the prophylaxis for DVT by persons expecting to be stationary for long periods of time.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

K212731 page 1 of 3

Contact Details

21 CFR 807.92(a)(1)

Applicant NameOrtho8 Inc.
Applicant Address2217 Plaza Drive Rocklin CA 95765 United States
Applicant Contact Telephone916-289-4002
Applicant ContactMs. Nordeen Taylor
Applicant Contact Emailtaylor@pmpmed.com
Correspondent NameMedTech Review, LLC
Correspondent Address257 Garnet Garden Street Henderson NV 89015 United States
Correspondent Contact Telephone612-889-5168
Correspondent ContactMr. John Beasley
Correspondent Contact Emailjohn@medtechreview.com
Device Trade NameCircul8 Pro Vascular Therapy System
Common NameCompressible Limb Sleeve
Classification NameCompressible Limb Sleeve
Regulation Number21 CFR 870.5800
Product CodeJOW
Predicate #Predicate Trade Name (Primary Predicate is listed first)Product Code
K193020Circul8 Pro Vascular Therapy SystemJOW
K200285VenaOneJOW
The Circul8 Pro Vascular Therapy System is a lightweight, portable, rechargeable battery powered prescription device. The device isintended to be used in the home or clinical setting by or under the direction of a medical professional by any patient needing venousreturn due to an increased risk from blood clots.

The Circul8 Pro Vascular Therapy System is intended to be an easy to use portable by a physician, for use in the home or clinical setting to help prevent the onset of DVT in patients by stimulating blood flow in the extremities (stimulating muscle

{4}------------------------------------------------

contractions). This device can be used to:

· Aid in the prevention of DVT;

  • · Enhance blood circulation;
    · Diminish post-operative pain and swelling;

· Reduce wound healing time;

· Aid in the treatment and healing of stasis dermatitis, venous stasis ulcers, chronic venous insufficiency and reduction of edema in the lower limbs.

The unit can also be used as an aid in the prophylaxis for DVT by persons expecting to be stationary for long periods of time.

Indications for Use Comparison

The indications for use of the predicate device and the subject device are the same.

Technological Comparison

21 CFR 807.92(a)(6)

21 CFR 807 92(a)(5)

The subject device is a modification of the predicate device in K193020, which includes:

  • modification of the plastic housing the electronically controlled air pump unit and solenoid valve,

  • relocation of the tactile touch control switch from the front of the housing to the housing,

  • introduction of LED display to replace the tricolored LED for ON, LOW BATTERY, CHARGE COMPLETED indications, and the blue LED for indicating a leak or low pressure alarm,

  • removal of the micro USB port for obtaining usage (patient compliance) data,

  • no longer labeling cuff as either left leg cuff or right leg cuff,

  • cuffs are filled to 55 mmHg pressure

  • introduces Night Mode, which minimizes light disturbances from the LEDs during patient resting times.

  • increased battery life.

The subject device has the identical performance characteristics to the predicate device in K193020. The circuit board in the subject device is updated to fit within the newly designed plastic case, which does not change any product performance. Additionally, the circuit board is updated to accommodate the relocation of the removal of the micro USB connector and the connection of the power adapter. The circuit board is also updated with a new charging circuit for the battery. Software is updated to accommodate the LED screen and Night Mode. None of these updates change the device performance. The differences between the modified (subject) device and the original (predicate) device do not raise any new issues of safety or efficacy.

The subject device is substantially equivalent to the predicate and reference device in function and operating principles to achieve identical results. The predicate device utilizes a microprocessor controlled pump to deliver approximately 50 mmHg of pressurized air to bladders that are attached to the patient's lower lime of approximately 60 seconds while the subject device is designed to deliver approximately 55 mmHg (reference device: VenaOne, K200285). Each cycle consists of inflation of a bladder, followed by a rest period during which the bladder deflates and the limb relaxes without any compression.

Both the subject device and predicate device utilize pneumatically controlled, single chamber cuffs actuated by an electronically controlled air pump unit and solenoid value of trol components are protectively housed in a plastic case that is permanently attached to the inflatable cuff. The chape of the protective plastic case do not raise any new issues of safety or effectiveness.

Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)

Ortho8 performed the following design verification testing on the modified Circul8 Pro (L10):

(a) Physical Requirements:

  • Test 1. Unit size to be no smaller than 140 X 70X 40 mm
  • Test 2. Unit weight to be no more than Less than 0.5 kg
  • Test 3. Housing material of construction confirmed to be ABS
  • Test 4. New location of On/Off is confirmed located on the top side the device
  • Test 5. LED screen correctly displays "LP" for low pressure alert, "55mmHg" for operating pressure, " " during cycles, and battery status

Test 6. Device correctly displays usage hours when power on

Test 7. Leg wrap size, shape are identical (no distinctions between Left and Right cuffs)

{5}------------------------------------------------

Design verification test results demonstrate the true mean (average) proportion, with a 5% margin of error and at the 90% confidence level, comply with Physical Requirements #1, #2, #3, #4, #5, #6, and #7.

(b) Performance Requirements:

Test 1: Inflation pressure accuracy of the pressure switch and software in the device correctly controls inflation of the wrap to 55 mmHg ± 10% (substantially the same preset pressure used by the predicate devices)

Test 2: Night mode performed as specified in the software design input

Design verification test results demonstrate the true mean (average) proportion, with a 5% margin of error and at the 90% confidence level, comply with Performance Requirements #1, and #2.

(c) Electrical Requirements:

Test 1: Confirm 3.7V, 1800mAh battery and 100 - 240 Vac, 50 - 60 Hz, Output: 5Vdc @2.0 Amp power supply

Design verification test results demonstrate of the true mean (average) proportion, with a 5% margin of error and at the 90% confidence level, comply with Electrical Requirement #1.

Non-clinical tests were performed on the subject device in order to validate the design and to assure conformance with the following voluntary design standards in connection with medical device electrical safety, and electromagnetic compatibility.

(e) IEC 60601-1:2005+AMD1:2012+AMD2:2020, Medical electrical equirements for basic safety and essential performance;

(f) IEC 60601-1-2:2014, Medical electrical equipments for basic safety and essential performance -Collateral Standard: Electromagnetic disturbances - Requirements and tests

(g) IEC 60601-1-11:2015+AMD1:2020, Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirement for medical electrical equipment and medical electrical systems used in the home healthcare environment, and

(h) IEC 62133-2:2017+AMD1:2021, Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications - Part 2: Lithium systems

Design validation test results demonstrate the modified Circul8Pro (L10) device complies with general electrical safety requirements, electromagnetic compatibility requirements for use in home healthcare environment, and requirements for secondary lithium cells and batteries for use in portable applications.

The verification of software used in the modified Circul8Pro (L10) device has been carried out according to the FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices.

These results demonstrate the modified Circul8Pro (L10) device is as safe, as effective, and performs as well as or better than the legally marketed device predicate.

§ 870.5800 Compressible limb sleeve.

(a)
Identification. A compressible limb sleeve is a device that is used to prevent pooling of blood in a limb by inflating periodically a sleeve around the limb.(b)
Classification. Class II (performance standards).