K Number
K212726

Validate with FDA (Live)

Date Cleared
2022-03-23

(208 days)

Product Code
Regulation Number
878.4040
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Disposable Medical Surgical Mask is intended to be worn to protect both the patient and healthcare personnel from the transfer of microorganisms, body fluids, and particulate material. The Disposable Medical Surgical Mask is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single-use, disposable device(s), provided sterile.

Device Description

The Disposable Medical Surgical Mask is intended to be worn to protect both the patient and healthcare personnel from the transfer of microorganisms, body fluids, and particulate material. The Disposable Medical Surgical Mask is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single-use, disposable device(s), provided sterile.

The proposed device(s) are Blue color, and Flat Pleated type mask(s), utilizing Ear Loops to secure to the users head and a pliable Nose Piece for fitting the facemask around the nose.

The proposed device(s) are manufactured with three layers, the inner and outer layers are made of polypropylene spunbond fabric, and the middle layer is made of polypropylene melt-blown fabric.

The Medical Surgical Masks-Sterile is held in place over the users' mouth and nose by two elastic ear loops affixed to the facemask. The elastic ear loops are made with nylon or spandex.

The nose piece contained in the proposed device(s) is contained within the layers of the facemask to allow the user to fit the facemask around their nose, which is made of polypropylene and metal wire.

The proposed device(s) are sold sterile and are intended to be single-use, disposable devices.

AI/ML Overview

The provided document is a 510(k) Summary for a medical device, a "Disposable Medical Surgical Mask". It details the device's technical characteristics, performance, and biocompatibility compared to a predicate device, as well as the non-clinical tests performed to demonstrate substantial equivalence.

Here's a breakdown of the requested information based on the provided text:

1. A table of acceptance criteria and the reported device performance

Test MethodPurposeAcceptance CriteriaReported Device Performance
ASTM F1862Resistance to penetration by synthetic blood120 mm Hg120 mm Hg (Table 2 and Test Method/Results table, page 7)
ASTM F2299Sub-micron particulate filtration efficiency at 0.1 micron≥ 98%≥ 98% (Table 2 and Test Method/Results table, page 7)
ASTM F2101Bacterial Filtration Efficiency≥ 98%≥ 99% (Table 2 and Test Method/Results table, page 7)
EN 14683 Annex CDifferential Pressure< 6.0 mm H2O/cm²< 5.7 mm H2O/cm² (Table 2 and Test Method/Results table, page 7)
16 CFR 1610FlammabilityClass 1Class 1 (Table 2 and Test Method/Results table, page 7)
ISO 10993-10IrritationNo irritation effectUnder the conditions of the study, no irritation effect (Table 3 and Test Method/Results table, page 7)
ISO 10993-10SensitizationNo sensitization effectUnder the conditions of the study, no sensitization effect (Table 3 and Test Method/Results table, page 7)
ISO 10993-5CytotoxicityNo cytotoxicity effectUnder the conditions of the study, no cytotoxicity effect (Table 3 and Test Method/Results table, page 7)
ISO 10993-7EtO and ECH ResidualThe average daily dose of EO to patient shall not exceed 4 mg;The average daily dose of ECH to patient shall not exceed 9 mg;EO average daily dose (mg/d) < 0.006;ECH average daily dose (mg/d) < 0.03 (Test Method/Results table, page 7)
ASTM F1980Shelf-Life ValidationMeet the requirements of Level 2 barrier performance at the end of shelf-life claimedMeet the requirements of Level 2 barrier performance after 2 years accelerated aging. The aging temperature is 60°C and the aging time is 56 days. (Test Method/Results table, page 7)

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document does not explicitly state the sample sizes for each specific test (e.g., number of masks tested for fluid resistance). It refers to the tests being conducted per the cited ASTM, EN, and ISO standards, which would typically specify sample sizes. The data provenance is not explicitly mentioned as "country of origin of the data" or "retrospective/prospective," but the manufacturer is based in China: Hubei Huaqiang High-Tech CO., LTD., No.499 DongLin Road, Yichang, Hubei Province, 443000, China. The testing described is non-clinical performance and biocompatibility testing of the device, not a human study.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not applicable to the type of testing described in the document. The tests performed are objective, standardized laboratory tests (e.g., measuring filtration efficiency, fluid resistance) whose "ground truth" is determined by the physical properties of the device and the methodologies of the test standards, not by expert interpretation.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. This is for direct performance measurements of physical characteristics, not expert adjudication of clinical outcomes.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is a medical surgical mask, not an AI-powered diagnostic tool. The document explicitly states: "No clinical study is included in this submission."

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable, as this device is not an algorithm or AI system.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The "ground truth" for this device's performance is established by the specified international and national standards (ASTM, EN, ISO, CFR) and their defined test methodologies. The results reflect the physical and biological properties of the mask as measured by accredited laboratory procedures, not by expert consensus, pathology, or outcomes data in a clinical setting.

8. The sample size for the training set

Not applicable. This device is not an AI/ML device that requires a training set.

9. How the ground truth for the training set was established

Not applicable for the same reason as above.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 23, 2022

Hubei Huaqiang High-Tech Co., Ltd. % Ray Wang General Manager Beijing Believe-Med Technology Service Co., Ltd. Rm.912, Building #15, XiYueHui, No.5, YiHe North Rd. FangShan District, Beijing 102401 China

Re: K212726

Trade/Device Name: Disposable Medical Surgical Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: March 8, 2022 Received: March 11, 2022

Dear Ray Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Bifeng Qian, M.D., Ph.D. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known)

K212726

Device Name Disposable Medical Surgical Mask

Indications for Use (Describe)

The Disposable Medical Surgical Mask is intended to be worn to protect both the patient and healthcare personnel from the transfer of microorganisms, body fluids, and particulate material. The Disposable Medical Surgical Mask is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single-use, disposable device(s), provided sterile.

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.

{3}------------------------------------------------

510(k) Summary K212726

This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.

    1. Date of Preparation: 2022/03/23
    1. Sponsor Identification

Hubei Huaqiang High-Tech CO., LTD. No.499 DongLin Road, Yichang, Hubei Province, 443000, China

Contact Person: Jing Yang Position: Vice Director of Tech Center Tel: +86-13545751006 Fax: +86-717-6347336 Email: 409281039@gq.com

    1. Designated Submission Correspondent
      Mr. Ray Wang

Beijing Believe-Med Technology Service Co., Ltd. Rm.912, Building #15, XiYueHui, No.5, YiHe North Rd., FangShan District, Beijing, 102401, China

Tel: +86-18910677558 Fax: +86-10-56335780 Email: Ray.Wang@believe-med.com

    1. Identification of Proposed Device
      Trade Name: Disposable Medical Surgical Mask Common Name: Mask, Surgical

Regulatory Information Classification Name: Mask, Surgical Classification: II Product Code: FXX Regulation Number: 878.4040 Review Panel: General Hospital

{4}------------------------------------------------

Indication for use Statement:

The Disposable Medical Surgical Mask is intended to be worn to protect both the patient and healthcare personnel from the transfer of microorganisms, body fluids, and particulate material. The Disposable Medical Surgical Mask is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single-use, disposable device(s), provided sterile.

Device Description

The Disposable Medical Surgical Mask is intended to be worn to protect both the patient and healthcare personnel from the transfer of microorganisms, body fluids, and particulate material. The Disposable Medical Surgical Mask is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single-use, disposable device(s), provided sterile.

The proposed device(s) are Blue color, and Flat Pleated type mask(s), utilizing Ear Loops to secure to the users head and a pliable Nose Piece for fitting the facemask around the nose.

The proposed device(s) are manufactured with three layers, the inner and outer layers are made of polypropylene spunbond fabric, and the middle layer is made of polypropylene melt-blown fabric.

The Medical Surgical Masks-Sterile is held in place over the users' mouth and nose by two elastic ear loops affixed to the facemask. The elastic ear loops are made with nylon or spandex.

The nose piece contained in the proposed device(s) is contained within the layers of the facemask to allow the user to fit the facemask around their nose, which is made of polypropylene and metal wire.

The proposed device(s) are sold sterile and are intended to be single-use, disposable devices.

    1. Identification of Predicate Device(s)
      510(k) number: K202904 Device Name: Surgical Face Mask Manufacturer: Jiangxi Feilikang Medical Technology Co., Ltd.

Technological Characteristics Comparison 6.

Table 1 General Comparison

ITEMProposed Device K212726Predicate Device K202904Remark
ASTM F2100 Level 2ASTM F2100 Level 2
Intended Use /Indication for UseThe Disposable Medical Surgical Maskis intended to be worn to protect boththe patient and healthcare personnelfrom the transfer of microorganisms,body fluids, and particulate material.The Disposable Medical Surgical Maskis intended for use in infection controlpractices to reduce the potentialexposure to blood and body fluids. Thisis a single-use, disposable device(s),provided sterile.Surgical Face Mask is intended forsingle use by operating room personneland other general healthcare workers toprotect both patients and healthcareworkers against transfer ofmicroorganisms, blood and body fluids,and particulate materials.Similar
Basic DesignEar Loops, Flat Pleated, 3 layersEar Loop, Flat-pleated, 3 layersSAME
MaterialsOuter FacingLayerpolypropylene spunbond fabricPolypropylene Non-Woven FabricDifferencesaddressed byperformance
Middle Layerpolypropylene melt-blown fabricPolypropylene Melt-Blown Fabric
Inner Facing Layerpolypropylene spunbond fabricPolypropylene Non-Woven Fabricand
Nose Piecepolypropylene and metal wirePolypropylene and Iron Wirebiocompatibility
Ear Loopsnylon or spandexNylon and Spandextesting
ColorBlueBlueSAME
Dimension (Length)$17.5 \pm 5%$ cm17.5 cmSAME
Dimension (Width)$9.5 \pm 5%$ cm9.5 cmSAME
OTC useYesYesSAME
Single UseYesYesSAME
SterileSterileSterileSAME
Sterilization methodEOEOSAME
Sterilization residuesEtO and ECH residues meet therequirements of ISO 10993-7EtO and ECH residues meet therequirements of ISO 10993-7SAME
Shelf Life2 yearsUnknownDifferenceaddressedbyShelf-Lifevalidation
Packaging10 pcs/bagUnknownSAME
PackagingConfiguration andmaterialsMedical dialysis paper (Tyvek) and PP(polyethylene and polypropylene) film.

{5}------------------------------------------------

Table 2 Performance Characteristic Comparison

ITEMProposed DeviceK212726Predicate DeviceK202904ASTM F2100RequirementsRemark
ASTM F2100 LevelLevel 2Level 2Level 2SAME
Fluid ResistancePerformance ASTMF1862120 mmHg120 mmHg120 mmHgBoth meetrequirements.No new
Particulate Filtration≥98%Average 98.74% at≥ 98%issues of

{6}------------------------------------------------

Efficiency ASTMF22990.1 µmsafety oreffectiveness
Bacterial FiltrationEfficiency ASTMF2101≥99%Average 99.65%≥ 98%
Differential Pressure(Delta P)EN 14683-2019+AC:2019 Annex C< 5.8mmH2O/cm²Average 4.6mmH2O/cm²< 6.0 mmH2O/cm2
Flammability16 CFR 1610Class 1Class 1Class 1SAME

Table 3 Biocompatibility Comparison

ITEMProposed Device K212726Predicate Device K202904Remark
CytotoxicityNo CytotoxicityNo CytotoxicitySAME
IrritationNo IrritationNo IrritationSAME
SensitizationNo SensitizationNo SensitizationSAME

7. Non-Clinical Test Conclusion

The test results demonstrated that the proposed device complies with the following standards:

  • ISO 11135-2014 Sterilization of health-care products - Ethylene oxide - Requirements for the development, validation and routine control of a sterilization process for medical devices

  • ISO 10993-5: 2009 Biological Evaluation Of Medical Devices – Part 5: Tests For In Vitro Cytotoxicity

  • ISO 10993-7: 2008 Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals

  • ISO 10993-10: 2010 Biological Evaluation Of Medical Devices Part 10: Tests For Irritation > And Skin Sensitization.
  • ASTM F2100-19, Standard Specification For Performance Of Materials Used In Medical Face Masks.

  • ASTM F1862-17, Standard Test Method For Resistance Of Medical Face Masks To Penetration By Synthetic Blood (Horizontal Projection Of Fixed Volume At A Known Velocity)

  • EN 14683-2019+AC:2019 Annex C, Medical face masks – Requirements and test methods;

  • ASTM F2101-19, Standard Test Method For Evaluating The Bacterial Filtration Efficiency A (Bfe) Of Medical Face Mask Materials, Using A Biological Aerosol Of Staphylococcus Aureus;
  • ASTM F2299-03, Stand test method for determining the initial efficiency of materials used in medical face masks to penetration by particulates using latex spheres;

{7}------------------------------------------------

Test MethodPurposeAcceptance CriteriaResults
ASTM F1862Resistance to penetration by synthetic blood120 mm Hg120 mm Hg
ASTM F2299Sub-micron particulate filtration efficiency at 0.1 micron≥ 98%≥ 98%
ASTM F2101Bacterial Filtration Efficiency≥ 98%≥ 99%
EN 14683Annex CDifferential Pressure< 6.0 mm H2O/cm²< 5.7 mm H2O/cm²
16 CFR 1610FlammabilityClass 1Class 1
ISO 10993-10IrritationSensitizationNo irritation effectNo sensitization effectUnder the conditions of the study, no irritation effectUnder the conditions of the study, no sensitization effect
ISO 10993-5CytotoxicityNo cytotoxicity effectUnder the conditions of the study, no cytotoxicity effect
ISO 10993-7EtO and ECH ResidualThe average daily dose of EO to patient shall not exceed 4 mg;The average daily dose of ECH to patient shall not exceed 9 mg;EO average daily dose (mg/d) < 0.006;ECH average daily dose (mg/d) < 0.03
ASTM F1980Shelf-Life ValidationMeet the requirements of Level 2 barrier performance at the end of shelf-life claimedMeet the requirements of Level 2 barrier performance after 2 years accelerated aging. The aging temperature is 60°C and the aging time is 56 days.
  • 16 CFR 1610, Standard for the Flammability of clothing textiles;

    1. Clinical Test Conclusion
      No clinical study is included in this submission.
    1. Conclusion
      The conclusions drawn from the nonclinical tests demonstrate that the subject device, K212726, Disposable Medical Surgical Mask, is as safe, as effective, and performs as well as or better than the legally marketed device, K202904, Surgical Face Mask by Jiangxi Felikang Medical Technology Log., Ltd.

§ 878.4040 Surgical apparel.

(a)
Identification. Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.(b)
Classification. (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.